Technical Overview
Introduction
Attenuated total reectance (ATR) is the most widely used sampling methodology for
FTIR spectroscopy. This popularity is due to the ability to quickly and easily measure a
broad range of sample types including liquids, solids, powders, semisolids, and pastes.
Depending on the specic application or sample, different ATR sensors are employed.
The Agilent Cary 630 FTIR accommodates a wide selection of ATR sensors, and features
the ability to switch from one ATR sensor to another instantaneously.
For most applications, single reection zinc selenide (ZnSe), diamond, and germanium
(Ge) ATR sampling accessories are available for the Cary 630. These accessories are
used with a sampling press, and are excellent for analyzing solid materials, as well as
liquids, pastes, and gels. The diamond sensor is highly durable, and the best choice
for harder materials. The ZnSe sensor is a good choice for softer solids, and is more
affordable than the diamond ATR. The Ge sensor, with a shorter pathlength, is the best
choice for highly absorbing samples. In addition, the ve reection ZnSe ATR sampling
accessory is a great option for the analysis of liquids, pastes, and gels, where additional
sensitivity is needed.
ZnSe ATR
Wavelength range
5,100600cm1
Effective pathlength
1.1 m at 4,000 cm1
2.6 m at 1,700 cm1
7.3 m at 600 cm1
Sample types
Soft solids (for example,
textile bers, polymer lms),
neat liquids, pastes, and gels
Wavelength range
5,100 cm1 to 600 cm1
Effective pathlength
5.5 m at 4,000 cm1
13.0 m at 1,700 cm1
36.5 m at 600 cm1
Sample types
Aqueous and nonaqueous
solutions, viscous and
nonviscous liquids, gels, and
pastes
Diamond ATR
Ge ATR
Effective pathlength
0.15 m at 4,000 cm1
0.36 m at 1,700 cm1
1.02 m at 600 cm1
Wavelength range
4,500 to 2,300 cm1
1,800 to 650 cm1
Samples types
Carbon black lled polymers
Effective pathlength
1.1 m at 4,000 cm1
2.6 m at 1,700 cm1
7.3 m at 600 cm1
Samples types
All liquids, solids, gels, and hard
materials
Wavelength range
5,100 to 600 cm1
ATR sampling technology available for the Agilent Cary 630 FTIR
ATR element
Wavelength range
1
Effective pathlength
1
Sample type
Pressure clamp
One-reection ZnSe
5,100600 cm
1.1 m at 4,000 cm
2.6 m at 1,700 cm1
7.3 m at 600 cm1
Yes
Five-reection ZnSe
5,100600 cm1
No
One-reection Diamond
4,5002,300 cm1
1,800650 cm1
One-reection Ge
5,100600 cm1
Yes
www.agilent.com/chem
This information is subject to change without notice.
Agilent Technologies, Inc., 2016
Published in the USA, May 11, 2016
5991-6858EN