Sei sulla pagina 1di 56

Model QW-400
Vertical Wall Frame

Models QW-420, QW-420-S,


and QW-420-D
Vertical Wall Stand/Frame

Service Manual

Manual Part No. DC30-040 Revision C


This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Quantum
Medical Imaging, LLC. (QMI)

Copyright© 2007 QMI

Quantum Medical Imaging, LLC Made in U.S.A.


2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800
Fax: (631) 567-5074
E-mail: info@qmiteam.com
www.quantummedical.net
Revision History

REVISION DATE TYPE OF MODIFICATION


A 5/24/04 Initial Release

B 10/5/04 Incorporated ECO# 1664, updated to new address

C 6/22/07 Incorporated ECO# 1761, 1788, 1824

Page Page Page


Number Rev Number Rev Number Rev
i - iv C

1-1 thru 1-12 C

2-1 thru 2-14 C

3-1 thru 3-4 C

4-1 thru 4-16 C

5-1 thru 5-4 C

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC i
Revision History

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


ii Quantum Medical Imaging, LLC
Table of Contents

Chapter 1 Specifications
PHYSICAL SPECIFICATIONS ................................................................................. 1-3
ELECTRICAL SPECIFICATIONS .............................................................................. 1-4
SYSTEM OPTIONS ............................................................................................... 1-4
SYSTEM OPERATING ENVIRONMENT .................................................................... 1-5
NON-OPERATING ENVIRONMENT ......................................................................... 1-5

Chapter 2 Assembly & Installation


OVERVIEW .......................................................................................................... 2-3
UNPACKING ........................................................................................................ 2-3
REQUIRED TOOLS ............................................................................................... 2-3
ASSEMBLY/INSTALLATION INSTRUCTIONS ........................................................... 2-3
MODELS QW-420, QW-420-D, QW-420-S ASSEMBLY/INSTALLATION ................... 2-3
MODEL QW-400 ASSEMBLY/INSTALLATION ....................................................... 2-9
POST INSTALLATION CHECK ........................................................................... 2-13
Chapter 3 Theory of Operation
OVERVIEW .......................................................................................................... 3-3
MECHANICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S ................. 3-3
MECHANICAL OPERATION - MODEL QW-400 ..................................................... 3-3
ELECTRICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S ................... 3-3
ELECTRICAL OPERATION - MODEL QW-400 ....................................................... 3-3

Chapter 4 Service Instructions


OVERVIEW .......................................................................................................... 4-3
SERVICE MAINTENANCE ................................................................................... 4-3
Cleaning .................................................................................................... 4-3
Visual Inspection ....................................................................................... 4-3
REMOVAL/REPLACEMENT PROCEDURES ............................................................ 4-4
Replacing the Vertical Pulley Assembly (Models QW-420, QW-420-D,
QW-420-S Only) ........................................................................................ 4-4
Cassette Tray Removal Procedure ............................................................... 4-8
REPLACEMENT PARTS AND ORDERING INFORMATION .......................................... 4-8

Chapter 5 Diagrams
FIGURE 5-1. MODELS QW-420, QW-420-D, QW-420-S VERTICAL WALL
STAND/FRAME WIRING DIAGRAM ..................................................................... 5-3
FIGURE 5-2. MODEL QW-400 VERTICAL WALL FRAME WIRING DIAGRAM ........... 5-4

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC iii
Table of Contents

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


iv Quantum Medical Imaging, LLC
Chapter
1

SPECIFICATIONS

1 -1
1-2
Chapter 1 Specifications

PHYSICAL SPECIFICATIONS
The following are physical specifications of the Vertical Wall Stand/Frame (see Figures
1-1 through 1-6):

Model Model Model Model


QW-400 QW-420 QW-420-D QW-420-S

Height: 85.0 inches 84.0 inches 84.0 inches 84.0 inches


(2159 mm) (2134 mm) (2134 mm) (2134 mm)

Width: 26.5 inches 27.5 inches 25.19 inches 37.25 inches


(673 mm) (699 mm) (640 mm) (947 mm)

Distance from Wall 7.36 inches 12.75 inches 13.25 inches 8.0 inches
to Front of (187 mm) (324 mm) (337 mm) (203 mm)
Receptor:

Distance from Front N/A 4.5 inches 4.5 inches 8.0 inches
of Column to Front (114 mm)* (114 mm)* (203 mm)
of Receptor:

Receptor Front Fiber Resin Fiber Resin N/A Fiber Resin


Cover Material: (phenolic) (phenolic) (phenolic)

Front Cover Density: <1 mmAl <1 mmAl N/A <1 mmAl

Receptor Types: 14" x 17" grid 14" x 17" Canon 14" x 17" grid
or bucky, grid or CXDI-40G, CMT or bucky
14" x 36" grid bucky, SmartRad, IDC
or bucky 14" x 36" XplorerTM 1500
grid or bucky

Focal Spot-to-Floor 25.0 inches 13.75 inches Receptor- 13.75 inches


Distance (with 14" x (635 mm) with (349 mm) dependent (349 mm)
17" receptor): fixed grid
26.75 inches
(679 mm) with
bucky

Focal Spot-to-Floor 34.875 inches 16.0 inches N/A N/A


Distance (with 14" x (886 mm) with (406 mm)
36" receptor): fixed grid
37.875 inches
(962 mm) with
bucky

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 1-3
Chapter 1 Specifications

Model Model Model Model


QW-400 QW-420 QW-420-D QW-420-S

Collimator Manual, Manual, Manual, Manual,


Compatability: Selectable, Selectable, Selectable, Selectable,
Automatic Automatic Automatic Automatic

Weight: Approx. 100 lbs Approx. 225 Approx. 200 lbs Approx. 225
(45.4 kg) lbs (98.0 kg) (90.7 kg) lbs (98.0 kg)

Image Receptor 42.5 inches 60.5 inches Receptor- 59.5 inches


Travel (14" x 17" (1080 mm) (1537 mm) dependent (1511 mm)
cabinet) with fixed grid with fixed grid
40.75 inches 60.5 inches
(1035 mm) (1537 mm)
with bucky with bucky

Image Receptor 22.75 inches 46.5 inches N/A N/A


Travel (14" x 36" (578 mm) with (1181 mm)
cabinet) fixed grid with fixed
19.75 inches grid
(502 mm) with 42.0 inches
bucky (1067 mm)
with bucky

*Note: Up to 8.0 inches (196 mm) from front of column to front of receptor with
optional extension QW-EX8

ELECTRICAL SPECIFICATIONS
Mode of Operation: Continuous
Locking System Type: Electromagnetic, Fail-Safe
Locking System Power: 24 VDC/1.0A
Power Supply Type: Double Insulated (or reinforced); must meet con-
struction/performance criteria set forth in UL
2601-1
Note: Systems equipped with a bucky require separate 120/220-240 VAC power
supply.

SYSTEM OPTIONS
QW-HG20: Patient “Side-Mounted” Handgrips (Used on Model
QW-420 only)
QW-HG30: Patient “Overhead” Handgrip (Used on Models
QW-420 and QW-420-S only)

Revision C Wall Stand/Frame - Service Manual


1-4 Quantum Medical Imaging, LLC
Chapter 1 Specifications

QW-EX8: 8 inch Bucky Extension Bracket (Used on Models


QW-420 and QW-420-D only)
QW-AT: Auto-Tracking Kit for synchronization with of Wall
Stand with Siemens Motorized Tubestand

SYSTEM OPERATING ENVIRONMENT


Ambient Temperature: +10°C to +40°C
Relative Humidity: 30 to 75%, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa

NON-OPERATING ENVIRONMENT
Ambient Temperature: -18°C to +70°C
Relative Humidity: 20 to 95%, non-condensing
Atmospheric Pressure: 500 hPa to 1060 hPa

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 1-5
Chapter 1 Specifications

Figure 1-1. Model QW-420 (with 14" x 17" Receptor) Dimensional Diagram

Revision C Wall Stand/Frame - Service Manual


1-6 Quantum Medical Imaging, LLC
Chapter 1 Specifications

Figure 1-2. Model QW-420 (with 14" x 36" Receptor) Dimensional Diagram

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 1-7
Chapter 1 Specifications

4"
WALL

1/2"

1/2"
1"
plane
film

8"
[203
13/16
1 "
[46] 3 1/4" 34"
[83] [864]
1"
Verti-Q
Verti-Q

1" [25]
MIN.
CLEARANCE 83.00
23.50 [2108]
74.25 [597]
MAX
WALL

[1886]

13.75 MIN
Verti-Q
Verti-Q

[349]

DC80-019-1 REV. C

NOTES:

Figure 1-3. Model QW-420-S Dimensional Diagram


(Parallel Wall Mount Configuration)

Revision C Wall Stand/Frame - Service Manual


1-8 Quantum Medical Imaging, LLC
Chapter 1 Specifications

2"
.312 [8] DIA [51 3/4"
MOUNTING HOLES [19]
REF

84.00
4" [2134]
[102 REF
plane
film

8"
[203]
34"
13/16
1 " [864
[46] 3 1/4" 3/4
[83] " [19] REF

1.00
[25.4]

Verti-Q
Verti-Q

83.00
23.50 [2108]
[597]
74.25 MAX
[1886]
WALL

13.75 MIN
Verti-Q
Verti-Q

[349]

DC80-019 REV. D

NOTES:

Figure 1-4. Model QW-420-S Dimensional Diagram


(Perpendicular Wall Mount Configuration)

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 1-9
Chapter 1 Specifications

DETAIL A Ø.38 X .75


SLOT HOLE
2X
18.25
[464]

MOUNTING
STRAP

34.94
[887]
SEE NOTE 1
26.63
[676]
SEE NOTE 1 C

WALL FRAME 19.63


QW-400 [498]

FOCAL 8'-9" MAX


SPOT [2667]
7'-9" MIN
[2362]
76
[1929]
Chiro - X

Chiro - X

A A

39.88
[1013]

Ø.31 MOUNTING
HOLE
2X SECTION A-A LEG 2X

Figure 1-5. Model QW-400 Dimensional Diagram

Revision C Wall Stand/Frame - Service Manual


1-10 Quantum Medical Imaging, LLC
Chapter 1 Specifications

Figure 1-6. Model QW-420-D Dimensional Diagram

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 1-11
Chapter 1 Specifications

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


1-12 Quantum Medical Imaging, LLC
Chapter
2

ASSEMBLY &
INSTALLATION

2-1
2-2
Chapter 2 Assembly & Installation

OVERVIEW
This chapter describes the steps required to assemble, install and align the Vertical
Wall Stand/Frame and how to install the completed assembly in your facility, including
making all required electrical connections.

UNPACKING
The Vertical Wall Stand/Frame is shipped in separate pieces as follows:
• Vertical Wall Stand/Frame Assembly
• Associated system components, depending on the model and options - a
Base Cover, Top Cap, trim weights, cable covering (with bucky and/or Ion
Chamber equipped systems only), cassette tray, power input cable,
hardware bag (including coil cord tie-down clamp and ground stud), patient
NOTE handgrips (optional), and Operator and Service Manuals.
Open the crate or carton marked "packing list enclosed" first. Locate and remove the
Examine all car-
tons and crates
packing list. Do not dispose of packing material until packing list is matched with
carefully at time actual parts received. If any damaged parts are found, notify the shipping or freight
of delivery. If company immediately (the manufacturer is relieved of any responsibility for damage
damage is appar- during shipment after unit is picked up by the carrier). Should there be a shortage of
ent, have deliv-
parts, notify Quantum Medical Imaging’s Service Department.
ery driver write a
"Damaged Ship-
ment Note" on REQUIRED TOOLS
copies of freight
bill, sign it, and The following tools and materials are required to complete the assembly and installa-
file appropriate tion procedures:
carrier claim.
Should you dis- • Socket Wrench Set
cover concealed
damage, immedi- • Assorted Phillips and slot blade screwdrivers
ately notify the
transporting • Drill (and masonry bit if concrete construction)
agent and ask
for an • Bubble level
"Inspection of
Damage". Carrier
will not accept
ASSEMBLY/INSTALLATION INSTRUCTIONS
concealed dam- The following paragraphs describe how to assemble the Verical Wall stand/Frame.
age claim if filed
after 15 days Assembly instructions for Model QW-400 are described separately because of the dif-
from date of ferences in this device’s design and construction. Assembly instructions for Models
receipt of mer- QW-420, QW-420-D and QW-420-S are very similar in design and construction and
chandise. are therefore combined into one set of instructions with model differences noted
where necessary.
MODELS QW-420, QW-420-D, QW-420-S ASSEMBLY/INSTALLATION
To assemble the Model QW-420, QW-420-D or QW-420-S Vertical Wall Stand,
refer to Figures 2-1 through 2-6 and proceed as follows:
1. Rest the bottom end of the Vertical Column on soft support material.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-3
Chapter 2 Assembly & Installation

2. Remove shipping bolt located in Flat Base Mounting Plate (see Figure 2-1).

Figure 2-1. Removing the Shipping Bolt


3. On Model QW-420-D only: Refer to the Digital Image Receptor Mounting
Instruction Sheet for specific information for the receptor type (DC30-067 -
CMT SmartRad, DC30-068 - Canon CXDI-40G, or DC30-069 - IDC Xplorer).
4. Carefully raise the Vertical Column to an upright position. The Receptor Cab-
inet will rise up the column because of the pull of the counterweight.
5. Connect red and black wires from Quantum generator (Interface Board) or
from a double (or reinforced) insulated 24 VDC power source (must meet
construction/performance criteria per UL 2601-1) to Terminal Board TB1 ter-
minals 1 and 2, respectively (see Models QW-420, QW-420-D, QW-420-S
Electrical Wiring Diagram in Chapter 5, DIAGRAMS). Use cable ties, as
required, to manage cable.
6. Install ground stud (supplied with system) into threaded hole on back side of
Vertical Column. Connect #10AWG yellow/green ground wire between Verti-
cal Wall Stand/Frame’s ground stud (located on back side of Vertical Column
near base) and x-ray generator.
7. On systems equipped with a moving grid (bucky), connect bucky power cable
to terminal block on bucky (refer to bucky manufacturer’s documentation,
provided with system, for specific wiring instructions) and tie-wrap cable to
inside of Receptor Cabinet along the bottom surface. Route cable through
opening in rear side of Receptor Cabinet.
8. If Patient “Overhead” Handgrip option (QW-HG30) is supplied, install as fol-
lows:
a. Loosen the small black thumbscrew located on the Patient Handle
Mount Bracket until threads are no longer visible in mounting hole for
Patient Handgrip.

Revision C Wall Stand/Frame - Service Manual


2-4 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

b. While pulling the Index Plunger out, insert the Patient Handgrip into
mounting hole in Patient Handle Mount Bracket and then release
plunger (see Figure 2-2).
c. Tighten the small black thumbscrew located on the Patient Handle
Mount Bracket.

INDEX
PLUNGER

PATIENT
HANDGRIP

THUMBSCREW

Figure 2-2. QW-HG30 Patient "Overhead" Handgrip Mounting


9. If Patient “Side-Mounted” Handgrips option (QW-HG20) is supplied, mount
each Side Handgrip to Receptor Cabinet using six (6) 10-32 x 1/2” pan head
phillips screws and star washers.

Note: If Patient “Overhead” Handgrip option


(QW-HG30) is also installed, add a Spacer (part
number ME11-122, provided with handgrips)
between the lower end of Side Handgrip that
attaches to the Overhead Patient Handle Mount
Bracket. Secure handgrip on this side using six
(6) 10-32 x 5/16” pan head phillips screws (see
Figure 2-3).

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-5
Chapter 2 Assembly & Installation

SIDE
MOUNTED
HANDGRIP

SPACER
REQUIRED
HERE IF
OVERHEAD
HANDGRIP
ALSO
INSTALLED

Figure 2-3. QW-HG20 Patient "Side-Mounted" Handgrips Installation


10. Turn on the collimator light.
11. Align the collimator cross hair lines with the centerline on grid, cassette tray
or digital receptor.
12. On Models QW-420 and QW-420-S only: Insert a 10 x 12 film cassette
into tray in Receptor Cabinet.
13. Verify the counterweight is balanced properly so it provides a smooth up and
down motion of the Receptor Cabinet. If additional counterweight is neces-
sary, add trim weights as follows:
a. Bring the Receptor Cabinet down to minimum vertical travel position.
b. Remove two screws securing the Column Center Cover to top of the
column (see Figure 2-4) and carefully tilt the center cover outward
enough to gain access to Access Opening (see Figure 2-10).

Revision C Wall Stand/Frame - Service Manual


2-6 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

COLUMN
CENTER
COVER
COLUMN
CENTER
COVER
SCREWS
(2X)

Figure 2-4. Column Center Cover Top Mounting Screws

Figure 2-5. Trim Weight Access Opening Location

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-7
Chapter 2 Assembly & Installation

c. Adjust the vertical position of the Receptor Cabinet so the top of the
counterweight is visible through the Access Opening.
d. Place trim weights (each weighs approximately 1.5 lbs.) inside com-
partment on top of counterweight (see Figure 2-5) until proper bal-
ance (motion) is achieved.
e. Re-install Column Center Cover mounting screws.
f. Re-check counterbalancing and adjust if necessary.
14. Secure the Vertical Column to the floor through the Flat Base Mounting Plate
using appropriate hardware (must be capable of supporting 100 lbs. pull-
load).
15. Secure the Vertical Column to the wall using the Wall-Mounting Bracket (and
optional Wall-Mounting Bar) with appropriate hardware (must be capable of
supporting 60 lbs. pull load) See Figure 2-6.

PULLEY/ WALL
BRAKE MOUNTING
ASSEMBLY BRACKET

Figure 2-6. Vertical Wall Mounting Bracket


16. On top of the Vertical Column, remove the two (2) phillips head screws from
standoffs. Install the Column Top Cap over top of column and secure by re-
installing the two (2) phillips head screws.
17. Move Receptor Cabinet through its entire vertical movement, making sure the
light beam remains on the center line on grid, cassette tray or digital receptor.
18. On Models QW-420 and QW-420-S only: Insert a film cassette into cas-
sette tray in Receptor Cabinet.
19. Take a projection into the cassette/receptor for verification of compliance.
(Refer to the calibration section of the Collimator manual.)

Revision C Wall Stand/Frame - Service Manual


2-8 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

20. If equipped, dress bucky and AEC cables using corrugated tubing and cable
tie-wraps as appropriate.
21. Attach Base Cover to Flat Base Mounting Plate using two (2) 6-32 x 1/4” pan
head screws.
22. On Models QW-420 and QW-420-S only: Attach Front Cover to Receptor
Cabinet and align centerline using collimator light.
MODEL QW-400 ASSEMBLY/INSTALLATION
To assembleand install the Model QW-400 Vertical Wall Frame, refer to Figures
2-7 through 2-& and proceed as follows:
1. Lift off the fiber-resin (phenolic) Front Cover from the Vertical Wall Frame.
2. Typically, the Receptor Cabinet is pre-installed into the Vertical Wall Frame at
the factory. However, instructions for assembly and installation of the Recep-
tor Cabinet are provided in steps 2a through 2k if needed for service or
replacement purposes:
a. Place the Receptor Cabinet on a flat surface with front side facing up,
top end away, and bottom end toward you.
b. Loosely install the two (2) Spring-Loaded Bearing Brackets on the top
and bottom non-film loading side of the Receptor Cabinet using two
(2) 1/4-20 x 3/4” screws on each bracket.
c. Loosely install Bearing Bracket (without spring), part number ME10-
013, on the bottom film-loading side of the Receptor Cabinet using
two (2) 1/4-20 x 3/4” screws.
d. Loosely install Bearing Bracket (without spring), part number ME20-
014, and Control Handle Mounting Bracket (part number ME20-012)
on the top film-loading side of Receptor Cabinet using two (2) 1/4-20
x 3/4” screws.
e. Install Receptor Cabinet by sliding it onto Bearing Rails in the Vertical
Wall Frame.
f. The slotted mounting holes in the Bucky Bearing Brackets are pro-
vided to adjust tension on bearings (wheels). When adjusted prop-
erly, the wheels on all four brackets should rotate throughout the
entire travel distance of the Receptor Cabinet.
g. Slide Receptor Cabinet all the way down to bottom of Vertical Wall
Frame.
h. Lift the counterweight so that the Pulley Cables slip over the Pulleys.
i. Secure the Pulley Cables to two top Bucky Bearing Brackets on
Receptor Cabinet using 1/4-20 x 1" hex head bolts and flat washers.
j. Slowly allow the counterweight to pull Receptor Cabinet to top of Ver-
tical Wall Frame.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-9
Chapter 2 Assembly & Installation

k. Check bumper stops making sure that the Receptor Cabinet does not
contact Vertical Wall Frame.

Figure 2-7. Receptor Cabinet Installed in QW-400 Wall Frame


3. Install the Control Handle with bracket to right side Bucky Bearing Bracket
(for right-hand film loading) or left side Bucky Bearing Bracket (for left-hand
film loading).
4. Connect the black coil cable wires to Terminal Board TB1 terminals 2 and 3 as
shown in Model QW-400 Electrical Wiring Diagram in Chapter 5, DIAGRAMS.
5. On systems configured for right-hand film loading, tie-wrap coil cord cable as
shown in Figure 2-8.
6. On systems configured for left-hand film loading, tie-wrap coil cord cable as
shown in Figure 2-9.
7. Re-attach the Front Cover on the Vertical Wall Frame.
8. Stand the Vertical Wall Frame against wall in proper position. Level Vertical
Wall Frame front-to-back and left-to-right using bubble level (use shims to
level left-to-right, as necessary).
9. Turn on collimator light.
10. Align the collimator cross-hair lines to the center line on the Front Cover of
the Vertical Wall Frame. Adjust position of Vertical Wall Frame as required.
11. Drill holes in the floor using the holes in the Vertical Wall Frame’s floor mount-
ing bracket as a guide.

Revision C Wall Stand/Frame - Service Manual


2-10 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

COIL CORD
TIE-WRAPPED
TO FRAME

COIL CORD TIE-


WRAPPED TO
COUNTERWEIGHT
CABLE

Figure 2-8. Model QW-400 Coil Cord Cable Routing (Right Hand Load Systems)

COIL CORD
TIE-WRAPPED
TO FRAME

Figure 2-9. Model QW-400 Coil Cord Cable Routing (Left Hand Load Systems)

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-11
Chapter 2 Assembly & Installation

12. Secure the Vertical Wall Frame to the floor through the Floor Mounting Brack-
ets using the appropriate hardware for floor construction (must be capable of
supporting 100 lbs. pull-load) as shown in Figure 2-10.

MOUNTING
BRACKET
HOLES (2X)

Figure 2-10. Vertical Wall Frame Floor Mounting Brackets


13. Remove Front Cover from the Vertical Wall Frame.
14. Secure the Vertical Wall Frame to the wall through mounting plate in rear of
Vertical Wall Frame using appropriate hardware (must be capable of support-
ing 60 lbs. pull-load).

MOUNTING
BRACKET
HOLES (2X)

Figure 2-11. Vertical Wall Frame Wall Mounting Bracket

Revision C Wall Stand/Frame - Service Manual


2-12 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

15. Connect red and black wires from Quantum generator (Interface Board) or
from a double (or reinforced) insulated 24 VDC power source (must meet
construction/performance criteria per UL 2601-1) to Terminal Board TB1 ter-
minals 1 and 2, respectively (see Model QW-400 Electrical Wiring Diagram in
Chapter 5, DIAGRAMS). Use cable ties, as required, to manage and secure
cable.
16. Ground the Vertical Wall Frame by connecting ground wire from system
ground to the ground stud located on the left leg (rear side) of the Vertical
Wall Frame Assembly.
17. If using a moving grid (bucky), connect bucky power cable to terminal block
on bucky (refer to bucky manufacturer’s documentation, provided with sys-
tem, for specific wiring instructions). Use cable ties to secure bucky cable
along top of Receptor Cabinet and to Control Handle Coil Cable (leave a ser-
vice loop to enable full travel of Receptor Cabinet).
18. Remove grid. Insert cassette into tray in Receptor Cabinet.
19. Move Receptor Cabinet to top of Vertical Wall Frame. Make sure that the Fail-
Safe Brake is de-energized by pressing the LOCK RELEASE button on the con-
trol handle attached to the Receptor Cabinet.
20. Line up the image receptor with collimator light field.
21. Tighten Bucky Bearing Brackets slightly.
22. Move Receptor Cabinet to the bottom of the Vertical Wall Frame.
23. Re-check alignment of the image receptor with the collimator light field.
24. Ensure that control handle coil cord remains in Vertical Wall Frame groove.
Check for smoothness of Receptor Cabinet motion throughout range of travel.
Adjust Bucky Bearing Brackets if necessary.
25. Fully tighten the Bucky Bearing Brackets.
26. Install grid.
27. Take projection into cassette for verification of compliance. (Refer to Collima-
tor Manual for alignment procedures).
28. Re-attach Front Cover to Vertical Wall Frame Assembly.
POST INSTALLATION CHECK
All mounting procedures must be approved before, and checked after, installation
by the hospital or facility engineering department and a registered, professional,
structural engineer. It is recommended that a leakage current measurement be
taken following installation to verify system meets allowable levels.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 2-13
Chapter 2 Assembly & Installation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


2-14 Quantum Medical Imaging, LLC
Chapter
3

THEORY
OF OPERATION

3-1
3-2
Chapter 3 Theory of Operation

OVERVIEW
Models QW-400, QW-420, QW-420-D and QW-420-S feature sufficient vertical travel to
accommodate all upright exams, and fail-safe electromagnetic locks for secure image
receptor positioning.
MECHANICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a Vertical Trolley, which rides on eight spring-loaded
radial ball bearings along with four metal shielded guide bearings bolted to it.
MECHANICAL OPERATION - MODEL QW-400
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a set of fourspring-loaded Receptor Bearing Guides.
ELECTRICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S
See Figure 5-1 in Chapter 5, DIAGRAMS.
Two fail-safe brakes (24 VDC electromagnets) are located at the top of the vertical
column (part of the Vertical Pulley/Brake Assembly). When the electromagnets are
de-activated (i.e., the Receptor Cabinet Lock Release Button is not pressed), the fail-
safe brakes "grab" the counterweight cable pulleys securely, locking the vertical
motion of the receptor cabinet. Pressing the Receptor Cabinet Lock Release Button
energizes the 24 VDC electromagnets, releasing the fail-safe brakes and allowing the
counterweight cable pulleys to spin freely thereby enabling vertical motion of the
receptor cabinet.
ELECTRICAL OPERATION - MODEL QW-400
See Figure 5-2 in Chapter 5, DIAGRAMS.
A fail-safe brake (24 VDC electromagnets) is located at the top right corner of the ver-
tical frame. When the electromagnet is de-activated (i.e., the Receptor Cabinet Lock
Release Button is not pressed), the fail-safe brake "grabs" the counterweight cable
pulley securely, locking the vertical motion of the receptor cabinet. Pressing the
Receptor Cabinet Lock Release Button energizes the 24 VDC electromagnet, releasing
the fail-safe brake and allowing the counterweight cable pulley to spin freely thereby
enabling vertical motion of the receptor cabinet.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 3-3
Chapter 3 Theory of Operation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


3-4 Quantum Medical Imaging, LLC
Chapter
4

SERVICE
INSTRUCTIONS

4-1
4-2
Chapter 4 Service Instructions

OVERVIEW
This chapter provides instructions to assist the service technician in maintaining the
smooth operation of the Vertical Wall Stand/Frame. The Vertical Wall Stand/Frame
will function reliably when maintained according to the instructions provided in this
chapter. Only properly trained service personnel should service or maintain this
equipment. Safe equipment performance requires the attention of service personnel
who are specifically trained and experienced with medical x-ray apparatus. Applicable
preventive maintenance or any repair service should be performed by these skilled
individuals. Failure to follow manufacturer’s or service personnel’s recommendations
may result in serious injury.

WARNING! Personnel engaged in maintenance


activities should exercise normal caution and care
while working with electromechanical equipment.
Only properly trained service personnel should
service or maintain this equipment. Failure to fol-
low manufacturer's or service personnel's recom-
mendations may result in serious injury. Before
removing or opening any electrical power panels
or covers, verify that the incoming power supply is
turned OFF. In the event maintenance procedures
require power to be supplied to the unit, extreme
care MUST be exercised to insure the safety of
service or other personnel in the area.

SERVICE MAINTENANCE

WARNING! Always disconnect the equipment


from the main power supply prior to any cleaning.
Verify that the equipment is properly grounded
before attempting any electrical operation. Never
use abrasive polish on this equipment.
Cleaning
The Vertical Wall Stand/Frame’s Front Cover surface should be wiped
clean. Painted metal surfaces should be cleaned using a clean cloth
slightly moistened in warm soapy water (use mild soap). Wipe dry with a
clean wet cloth.
Visual Inspection
A complete series of inspections and functional checks was conducted at
installation to insure proper operation of the system. However, periodic
routine inspection and maintenance of the Vertical Wall Stand/Frame
should be performed by a qualified service technician. The following
inspection and adjustment procedures are recommended to maintain the
system in its original operating condition:

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-3
Chapter 4 Service Instructions

Perform Every Six Months


• Visually inspect for wear and cleanliness
• Check for evidence of loose hardware or loose wires
• Check all bearings and bearing surfaces for cleanliness and corrosion
• Clean Bearing Tracks
• Conduct a general inspection for worn or damaged parts
• Check for smoothness of Receptor Cabinet motion
• Verify that all ground conductors are properly and securely installed
and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying
• Check external covers for proper fit; inspect for scratches that may
indicate cover misalignment or damage caused by collisions with
footstools, trolleys, or other equipment
• Verify vertical electomagnetic lock holding force exceeds 35 lbs.
(Models QW-420, QW-420-S, QW-420-D) or 18 lbs. (Model QW-400).
REMOVAL/REPLACEMENT PROCEDURES
The following procedures provide step-by-step instructions for removal and
replacement of the repairable components in the Vertical Wall Stand/Frame.
Replacing the Vertical Pulley Assembly (Models QW-420, QW-
420-D, QW-420-S Only)
1. Remove the Top Cap and Base Cover from the Vertical Column
(see Figure 4-1).
COLUMN
CENTER COVER

TOP CAP

NOTE: MODEL
QW-420 IS SHOWN
IN THIS DIAGRAM, COLUMN BASE
MODELS QW-420-D COVER
AND Q W-420-S
HAVE IDENTICAL
PART LOCATIONS

Figure 4-1. Top Cap, Base Cover and Center Column Cover Locations
2. Remove the Column Center Cover from the Vertical Column (see
Figure 4-1).

Revision C Wall Stand/Frame - Service Manual


4-4 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

3. Remove all trim weights from counterweight tray.


4. Move the receptor cabinet to upper most vertical position.
5. Turn off power to Vertical Wall Stand. Disconnect +24 VDC power
cable from terminal block TB1 located on top of Vertical Column.
6. Disconnect ground wire from ground lug on back side of Vertical
Column.
7. If wall stand is equipped with a bucky and/or ion chamber, free
up bucky/ion chamber cable drape to allow enough slack for wall
stand to be laid down on floor. If sufficient drape cannot be
attained, disconnect bucky power and/or ion chamber cables
from wall stand.
8. Remove all remaining wires from terminal block TB1.

TOP BRACKET
TERMINAL
MOUNTING
BLOCK TB1
SCREWS (4X)

Figure 4-2. Removing Vertical Pulley Assembly Mounting Screws


9. Remove four (4) ¼-20 x 3/4” hex head screws securing the Verti-
cal Column Wall Mount Bracket to Vertical Column (see Figure
4-2). Retain hardware for re-assembly.
10. Remove hardware securing the Vertical Column Base to floor.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-5
Chapter 4 Service Instructions

CAUTION! The Vertical Column requires two per-


sons to lift. Before lowering the column, prepare a
resting location for laying the column on; lay
down two pieces of 2” x 4” lumber about three
feet apart (cover lumber with soft cloth or cush-
ioning material to prevent damage to column or
scratching of painted surfaces).

11. Carefully, slide the Vertical Column forward until clear of Wall
Mount Bracket, then slowly lower column until horizontal on floor.
12. Remove two (2) ¼-20 x 3/4” screws securing Vertical Pulley
Assembly to Vertical Column (see Figure 4-3).

NOTE: Before disconnecting counterweight


cables, tag the cable on the left side “LEFT” and
the cable on the right side “RIGHT” (as viewed
from top). This will ensure correct position of
cables on replacement Vertical Pulley Assembly
during re-installation.

Figure 4-3. Removing Vertical Pulley Assembly Mounting Screws

Revision C Wall Stand/Frame - Service Manual


4-6 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Figure 4-4. Tilting the Vertical Pulley Assembly to Access Counterweight Cables
13. Tilt the Vertical Pulley out from Vertical Column enough to gain
access to counterweight cable mounting hardware (see Figure
4-4).
14. Remove two (2) ¼-20 x 7/8” screws securing each counterweight
cable to Trolley Assembly.
15. Remove the Vertical Pulley Assembly from column.
16. Position the replacement Vertical Pulley Assembly on the Vertical
Column and tilt forward as shown in Figure 4-4.
17. Route the LEFT and RIGHT counterweight cables over the appro-
priate ends of the Vertical Pulley Assembly. (Do not cross left and
right cables!). Secure each cable to Trolley Assembly using two 1/
4-20 x 7/8” screws. Ensure screws are tightened securely.
18. Insert the Vertical Pulley Assembly fully into Vertical Column and
secure to column using two (2) ¼-20 x 3/4” screws.

CAUTION! The Vertical Column requires two per-


sons to lift.

19. Carefully lift Vertical Column to vertical position and slide into the
location on floor where it was originally installed.
20. Install four (4) ¼-20 x 3/4” hex head screws securing the Vertical
Column Wall Mount Bracket to Vertical Column (see Figure 4-2).
21. Install hardware to secure the Vertical Column Base to floor.

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-7
Chapter 4 Service Instructions

22. Return all previously removed trim weights to counterweight tray.


23. Re-connect wiring to terminal block TB1.
24. Return-bucky and/or ion chamber cables to original draping posi-
tions. If disconnected, re-connect the bucky and/or ion chamber
cables to wall stand.
25. Connect ground wire to ground lug on back side of Vertical Column.
26. Re-connect +24 VDC power cable to terminal block TB1.
27. Install the Column Center Cover, Top Cap and Base Cover on the
Vertical Column (see Figure 4-1).
28. Turn on power to Vertical Wall Stand. Verify proper receptor vertical
motion and locking functionality.
Cassette Tray Removal Procedure
To remove the cassette tray from the receptor cabinet housing for cleaning
or repair, proceed as follows:
1. Pull the cassette tray out by the handle until it stops.
2. Grasp both sides of the tray with your right hand placed near the
vertical wall stand/frame and your left hand towards the tray front.
3. Locate the finger hole on the right side of the tray bottom. A
spring-loaded plate covers this hole. Press the plate with your right
index finger, lift the cassette tray up slightly and pull the tray away
from the housing.

REPLACEMENT PARTS AND ORDERING INFORMATION


Table 4-1 below provides a list of replaceable parts for the Vertical Wall Stand/Frame.
See Figures 4-5 through 4-11 for locations of replaceable parts. Use only original
replacement parts, as supplied or recommended by Quantum Medical Imaging, LLC.
ORDERING INFORMATION
To order replacement parts for the Vertical Wall Stand/Frame, contact the Service
Department at;
Quantum Medical Imaging, LLC.
2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800 (x2)
Fax: (631) 567-5074
e-mail: service@qmiteam.com
When ordering replacement parts, supply the following information:
•Model and serial number of equipment
•Part number

Revision C Wall Stand/Frame - Service Manual


4-8 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

•Part description
•Quantity required
•P.O. Number
•Shipping Instructions
When ordering components or parts not listed in Table 4-1, a complete description
of the part, including its function and location should be provided with the model
number and serial number of the unit.
Table 4-1. Replaceable Parts

Part Model(s) Qty


Item Description Number Used On Used in
System
1 Switch, Gray Push button EL50-002 QW-400 1
QW-420 1
QW-420-D 1
QW-420-S 1

2 Handle Assembly AY20-022 QW-400 1


QW-420 1
QW-420-D* 1
QW-420-S 1

3 R6ZZ Bearing ME40-018 QW-420 4


QW-420-D 4
QW-420-S 4

4 1628-ZZ Bearing ME40-017 QW-420 8


QW-420-D 8
QW-420-S 8

5 Vertical Pulley Assembly AY20-081 QW-420 1


QW-420-D 1
QW-420-S 1

6 Brake, Fail-Safe EL80-002 QW-400 1


QW-420 2
QW-420-D 2
QW-420-S 2

7 Trim Weight (not shown) ME10-105 QW-420 As Req’d


QW-420-D
QW-420-S

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-9
Chapter 4 Service Instructions

Part Model(s) Qty


Item Description Number Used On Used in
System
8 24-Volt Power Cable (not EL75-005 QW-400 35 feet
shown) QW-420
QW-420-D
QW-420-S

9 Chin Rest ME30-002 QW-420 1


QW-420-S 1

10 Front Panel (phenolic), 14" x ME31-002 QW-420 1


17" QW-420-S 1

11 Cover, Cabinet, Top ME10-048 QW-420 1


QW-420-S 1

12 Cover, Cabinet, Bottom ME10-047 QW-420 1


QW-420-S 1

13 Cover, Cabinet, Non-Load ME10-049 QW-420 1


Side QW-420-S 1

14 Cover, Cabinet, Load Side, ME10-051 QW-420 1


Left Load Bucky QW-420-S
OR ME10-194
Cover, Cabinet, Load Side,
Left Load Grid Cabinet

15 Cover, Cabinet, Load Side, ME10-032 QW-420 1


Right Load Bucky QW-420-S
OR
Cover, Cabinet, Load Side, ME10-195
Right Load Grid Cabinet

16 Switch, Pushbutton, Auto- EL50-030 QW-420 with 1


Track Enable QW-AT Option

17 Potentiometer, 5K, 10 Turn, EL20-024 QW-420 with 1


Precision QW-AT Option

18 Front Panel (phenolic) ME31-004 QW-400 1

19 12" Control Handle Retract- EL75-001 QW-400 1


able Coil Cord

20 Receptor Bearing Guides ME30-008 QW-400 4

*Note: On QW-420-D systems interfaced with IDC Xplorer receptor, use Handle
Assembly part number AY20-201.

Revision C Wall Stand/Frame - Service Manual


4-10 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

Figure 4-5. Replaceable Parts Location Diagram - All Models

4 4

Figure 4-6. Replaceable Parts Location Diagram -


Models QW-420, QW-420-D, QW-420-S

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-11
Chapter 4 Service Instructions

6
Figure 4-7. Replaceable Parts Location Diagram -
Models QW-420, QW-420-D, QW-420-S

Figure 4-8. Replaceable Parts Location Diagram -


Models QW-420, QW-420-S

Revision C Wall Stand/Frame - Service Manual


4-12 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

11

14 15
13

10 10

12

LEFT SIDE LOAD RIGHT SIDE LOAD

Figure 4-9. Replaceable Parts Location Diagram -


Models QW-420, QW-420-S

16

Figure 4-10. Replaceable Parts Location Diagram -


Model QTW-420 (with QW-AT Option)

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-13
Chapter 4 Service Instructions

17

Figure 4-11. Replaceable Parts Location Diagram -


Model QTW-420 (with QW-AT Option)

18 2

Figure 4-12. Replaceable Parts Location Diagram - Model QW-400

Revision C Wall Stand/Frame - Service Manual


4-14 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

19

20

Figure 4-13. Replaceable Parts Location Diagram - Model QW-400

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 4-15
Chapter 4 Service Instructions

THIS PAGE INTENTIONALLY LEFT BLANK

Revision C Wall Stand/Frame - Service Manual


4-16 Quantum Medical Imaging, LLC
Chapter
5

DIAGRAMS

5-1
5-2
Chapter 5 Diagrams

24 VDC FAIL SAFE VERTICAL BRAKES

BUCKY WALL STAND


X-RAY GENERATOR

BLACK

BLACK
A9 INTERFACE BOARD
RED

RED
JR2
1 24 VDC RED D2
APTB2 -
2 24 RTN BLACK
-
1N4005 D1
1N4005

RED
BLACK
BLACK BLACK

BLACK
HAND SWITCH
SW1 6
5
4 GROUND STUD
COM N/O
3
BLACK 2
1
RED

GREEN/YELLOW TO X-RAY GENERATOR MAIN GROUND STUD

SC50-006T REV. C

Figure 5-1. Models QW-420, QW-420-D, QW-


420-S Vertical Wall Stand/Frame Wiring
Diagram

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 5-3
Chapter 5 Diagrams

Figure 5-2. QW-400 Vertical Wall Frame


Wiring Diagram

Wall Stand/Frame - Service Manual Revision C


Quantum Medical Imaging, LLC 5-4