Model QW-400
Vertical Wall Frame
Service Manual
Chapter 1 Specifications
PHYSICAL SPECIFICATIONS ................................................................................. 1-3
ELECTRICAL SPECIFICATIONS .............................................................................. 1-4
SYSTEM OPTIONS ............................................................................................... 1-4
SYSTEM OPERATING ENVIRONMENT .................................................................... 1-5
NON-OPERATING ENVIRONMENT ......................................................................... 1-5
Chapter 5 Diagrams
FIGURE 5-1. MODELS QW-420, QW-420-D, QW-420-S VERTICAL WALL
STAND/FRAME WIRING DIAGRAM ..................................................................... 5-3
FIGURE 5-2. MODEL QW-400 VERTICAL WALL FRAME WIRING DIAGRAM ........... 5-4
SPECIFICATIONS
1 -1
1-2
Chapter 1 Specifications
PHYSICAL SPECIFICATIONS
The following are physical specifications of the Vertical Wall Stand/Frame (see Figures
1-1 through 1-6):
Distance from Wall 7.36 inches 12.75 inches 13.25 inches 8.0 inches
to Front of (187 mm) (324 mm) (337 mm) (203 mm)
Receptor:
Distance from Front N/A 4.5 inches 4.5 inches 8.0 inches
of Column to Front (114 mm)* (114 mm)* (203 mm)
of Receptor:
Front Cover Density: <1 mmAl <1 mmAl N/A <1 mmAl
Receptor Types: 14" x 17" grid 14" x 17" Canon 14" x 17" grid
or bucky, grid or CXDI-40G, CMT or bucky
14" x 36" grid bucky, SmartRad, IDC
or bucky 14" x 36" XplorerTM 1500
grid or bucky
Weight: Approx. 100 lbs Approx. 225 Approx. 200 lbs Approx. 225
(45.4 kg) lbs (98.0 kg) (90.7 kg) lbs (98.0 kg)
*Note: Up to 8.0 inches (196 mm) from front of column to front of receptor with
optional extension QW-EX8
ELECTRICAL SPECIFICATIONS
Mode of Operation: Continuous
Locking System Type: Electromagnetic, Fail-Safe
Locking System Power: 24 VDC/1.0A
Power Supply Type: Double Insulated (or reinforced); must meet con-
struction/performance criteria set forth in UL
2601-1
Note: Systems equipped with a bucky require separate 120/220-240 VAC power
supply.
SYSTEM OPTIONS
QW-HG20: Patient “Side-Mounted” Handgrips (Used on Model
QW-420 only)
QW-HG30: Patient “Overhead” Handgrip (Used on Models
QW-420 and QW-420-S only)
NON-OPERATING ENVIRONMENT
Ambient Temperature: -18°C to +70°C
Relative Humidity: 20 to 95%, non-condensing
Atmospheric Pressure: 500 hPa to 1060 hPa
Figure 1-1. Model QW-420 (with 14" x 17" Receptor) Dimensional Diagram
Figure 1-2. Model QW-420 (with 14" x 36" Receptor) Dimensional Diagram
4"
WALL
1/2"
1/2"
1"
plane
film
8"
[203
13/16
1 "
[46] 3 1/4" 34"
[83] [864]
1"
Verti-Q
Verti-Q
1" [25]
MIN.
CLEARANCE 83.00
23.50 [2108]
74.25 [597]
MAX
WALL
[1886]
13.75 MIN
Verti-Q
Verti-Q
[349]
DC80-019-1 REV. C
NOTES:
2"
.312 [8] DIA [51 3/4"
MOUNTING HOLES [19]
REF
84.00
4" [2134]
[102 REF
plane
film
8"
[203]
34"
13/16
1 " [864
[46] 3 1/4" 3/4
[83] " [19] REF
1.00
[25.4]
Verti-Q
Verti-Q
83.00
23.50 [2108]
[597]
74.25 MAX
[1886]
WALL
13.75 MIN
Verti-Q
Verti-Q
[349]
DC80-019 REV. D
NOTES:
MOUNTING
STRAP
34.94
[887]
SEE NOTE 1
26.63
[676]
SEE NOTE 1 C
Chiro - X
A A
39.88
[1013]
Ø.31 MOUNTING
HOLE
2X SECTION A-A LEG 2X
ASSEMBLY &
INSTALLATION
2-1
2-2
Chapter 2 Assembly & Installation
OVERVIEW
This chapter describes the steps required to assemble, install and align the Vertical
Wall Stand/Frame and how to install the completed assembly in your facility, including
making all required electrical connections.
UNPACKING
The Vertical Wall Stand/Frame is shipped in separate pieces as follows:
• Vertical Wall Stand/Frame Assembly
• Associated system components, depending on the model and options - a
Base Cover, Top Cap, trim weights, cable covering (with bucky and/or Ion
Chamber equipped systems only), cassette tray, power input cable,
hardware bag (including coil cord tie-down clamp and ground stud), patient
NOTE handgrips (optional), and Operator and Service Manuals.
Open the crate or carton marked "packing list enclosed" first. Locate and remove the
Examine all car-
tons and crates
packing list. Do not dispose of packing material until packing list is matched with
carefully at time actual parts received. If any damaged parts are found, notify the shipping or freight
of delivery. If company immediately (the manufacturer is relieved of any responsibility for damage
damage is appar- during shipment after unit is picked up by the carrier). Should there be a shortage of
ent, have deliv-
parts, notify Quantum Medical Imaging’s Service Department.
ery driver write a
"Damaged Ship-
ment Note" on REQUIRED TOOLS
copies of freight
bill, sign it, and The following tools and materials are required to complete the assembly and installa-
file appropriate tion procedures:
carrier claim.
Should you dis- • Socket Wrench Set
cover concealed
damage, immedi- • Assorted Phillips and slot blade screwdrivers
ately notify the
transporting • Drill (and masonry bit if concrete construction)
agent and ask
for an • Bubble level
"Inspection of
Damage". Carrier
will not accept
ASSEMBLY/INSTALLATION INSTRUCTIONS
concealed dam- The following paragraphs describe how to assemble the Verical Wall stand/Frame.
age claim if filed
after 15 days Assembly instructions for Model QW-400 are described separately because of the dif-
from date of ferences in this device’s design and construction. Assembly instructions for Models
receipt of mer- QW-420, QW-420-D and QW-420-S are very similar in design and construction and
chandise. are therefore combined into one set of instructions with model differences noted
where necessary.
MODELS QW-420, QW-420-D, QW-420-S ASSEMBLY/INSTALLATION
To assemble the Model QW-420, QW-420-D or QW-420-S Vertical Wall Stand,
refer to Figures 2-1 through 2-6 and proceed as follows:
1. Rest the bottom end of the Vertical Column on soft support material.
2. Remove shipping bolt located in Flat Base Mounting Plate (see Figure 2-1).
b. While pulling the Index Plunger out, insert the Patient Handgrip into
mounting hole in Patient Handle Mount Bracket and then release
plunger (see Figure 2-2).
c. Tighten the small black thumbscrew located on the Patient Handle
Mount Bracket.
INDEX
PLUNGER
PATIENT
HANDGRIP
THUMBSCREW
SIDE
MOUNTED
HANDGRIP
SPACER
REQUIRED
HERE IF
OVERHEAD
HANDGRIP
ALSO
INSTALLED
COLUMN
CENTER
COVER
COLUMN
CENTER
COVER
SCREWS
(2X)
c. Adjust the vertical position of the Receptor Cabinet so the top of the
counterweight is visible through the Access Opening.
d. Place trim weights (each weighs approximately 1.5 lbs.) inside com-
partment on top of counterweight (see Figure 2-5) until proper bal-
ance (motion) is achieved.
e. Re-install Column Center Cover mounting screws.
f. Re-check counterbalancing and adjust if necessary.
14. Secure the Vertical Column to the floor through the Flat Base Mounting Plate
using appropriate hardware (must be capable of supporting 100 lbs. pull-
load).
15. Secure the Vertical Column to the wall using the Wall-Mounting Bracket (and
optional Wall-Mounting Bar) with appropriate hardware (must be capable of
supporting 60 lbs. pull load) See Figure 2-6.
PULLEY/ WALL
BRAKE MOUNTING
ASSEMBLY BRACKET
20. If equipped, dress bucky and AEC cables using corrugated tubing and cable
tie-wraps as appropriate.
21. Attach Base Cover to Flat Base Mounting Plate using two (2) 6-32 x 1/4” pan
head screws.
22. On Models QW-420 and QW-420-S only: Attach Front Cover to Receptor
Cabinet and align centerline using collimator light.
MODEL QW-400 ASSEMBLY/INSTALLATION
To assembleand install the Model QW-400 Vertical Wall Frame, refer to Figures
2-7 through 2-& and proceed as follows:
1. Lift off the fiber-resin (phenolic) Front Cover from the Vertical Wall Frame.
2. Typically, the Receptor Cabinet is pre-installed into the Vertical Wall Frame at
the factory. However, instructions for assembly and installation of the Recep-
tor Cabinet are provided in steps 2a through 2k if needed for service or
replacement purposes:
a. Place the Receptor Cabinet on a flat surface with front side facing up,
top end away, and bottom end toward you.
b. Loosely install the two (2) Spring-Loaded Bearing Brackets on the top
and bottom non-film loading side of the Receptor Cabinet using two
(2) 1/4-20 x 3/4” screws on each bracket.
c. Loosely install Bearing Bracket (without spring), part number ME10-
013, on the bottom film-loading side of the Receptor Cabinet using
two (2) 1/4-20 x 3/4” screws.
d. Loosely install Bearing Bracket (without spring), part number ME20-
014, and Control Handle Mounting Bracket (part number ME20-012)
on the top film-loading side of Receptor Cabinet using two (2) 1/4-20
x 3/4” screws.
e. Install Receptor Cabinet by sliding it onto Bearing Rails in the Vertical
Wall Frame.
f. The slotted mounting holes in the Bucky Bearing Brackets are pro-
vided to adjust tension on bearings (wheels). When adjusted prop-
erly, the wheels on all four brackets should rotate throughout the
entire travel distance of the Receptor Cabinet.
g. Slide Receptor Cabinet all the way down to bottom of Vertical Wall
Frame.
h. Lift the counterweight so that the Pulley Cables slip over the Pulleys.
i. Secure the Pulley Cables to two top Bucky Bearing Brackets on
Receptor Cabinet using 1/4-20 x 1" hex head bolts and flat washers.
j. Slowly allow the counterweight to pull Receptor Cabinet to top of Ver-
tical Wall Frame.
k. Check bumper stops making sure that the Receptor Cabinet does not
contact Vertical Wall Frame.
COIL CORD
TIE-WRAPPED
TO FRAME
Figure 2-8. Model QW-400 Coil Cord Cable Routing (Right Hand Load Systems)
COIL CORD
TIE-WRAPPED
TO FRAME
Figure 2-9. Model QW-400 Coil Cord Cable Routing (Left Hand Load Systems)
12. Secure the Vertical Wall Frame to the floor through the Floor Mounting Brack-
ets using the appropriate hardware for floor construction (must be capable of
supporting 100 lbs. pull-load) as shown in Figure 2-10.
MOUNTING
BRACKET
HOLES (2X)
MOUNTING
BRACKET
HOLES (2X)
15. Connect red and black wires from Quantum generator (Interface Board) or
from a double (or reinforced) insulated 24 VDC power source (must meet
construction/performance criteria per UL 2601-1) to Terminal Board TB1 ter-
minals 1 and 2, respectively (see Model QW-400 Electrical Wiring Diagram in
Chapter 5, DIAGRAMS). Use cable ties, as required, to manage and secure
cable.
16. Ground the Vertical Wall Frame by connecting ground wire from system
ground to the ground stud located on the left leg (rear side) of the Vertical
Wall Frame Assembly.
17. If using a moving grid (bucky), connect bucky power cable to terminal block
on bucky (refer to bucky manufacturer’s documentation, provided with sys-
tem, for specific wiring instructions). Use cable ties to secure bucky cable
along top of Receptor Cabinet and to Control Handle Coil Cable (leave a ser-
vice loop to enable full travel of Receptor Cabinet).
18. Remove grid. Insert cassette into tray in Receptor Cabinet.
19. Move Receptor Cabinet to top of Vertical Wall Frame. Make sure that the Fail-
Safe Brake is de-energized by pressing the LOCK RELEASE button on the con-
trol handle attached to the Receptor Cabinet.
20. Line up the image receptor with collimator light field.
21. Tighten Bucky Bearing Brackets slightly.
22. Move Receptor Cabinet to the bottom of the Vertical Wall Frame.
23. Re-check alignment of the image receptor with the collimator light field.
24. Ensure that control handle coil cord remains in Vertical Wall Frame groove.
Check for smoothness of Receptor Cabinet motion throughout range of travel.
Adjust Bucky Bearing Brackets if necessary.
25. Fully tighten the Bucky Bearing Brackets.
26. Install grid.
27. Take projection into cassette for verification of compliance. (Refer to Collima-
tor Manual for alignment procedures).
28. Re-attach Front Cover to Vertical Wall Frame Assembly.
POST INSTALLATION CHECK
All mounting procedures must be approved before, and checked after, installation
by the hospital or facility engineering department and a registered, professional,
structural engineer. It is recommended that a leakage current measurement be
taken following installation to verify system meets allowable levels.
THEORY
OF OPERATION
3-1
3-2
Chapter 3 Theory of Operation
OVERVIEW
Models QW-400, QW-420, QW-420-D and QW-420-S feature sufficient vertical travel to
accommodate all upright exams, and fail-safe electromagnetic locks for secure image
receptor positioning.
MECHANICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a Vertical Trolley, which rides on eight spring-loaded
radial ball bearings along with four metal shielded guide bearings bolted to it.
MECHANICAL OPERATION - MODEL QW-400
Smooth vertical (upward and downward) image receptor movement is achieved by
attachment of the receptor to a set of fourspring-loaded Receptor Bearing Guides.
ELECTRICAL OPERATION - MODELS QW-420, QW-420-D, QW-420-S
See Figure 5-1 in Chapter 5, DIAGRAMS.
Two fail-safe brakes (24 VDC electromagnets) are located at the top of the vertical
column (part of the Vertical Pulley/Brake Assembly). When the electromagnets are
de-activated (i.e., the Receptor Cabinet Lock Release Button is not pressed), the fail-
safe brakes "grab" the counterweight cable pulleys securely, locking the vertical
motion of the receptor cabinet. Pressing the Receptor Cabinet Lock Release Button
energizes the 24 VDC electromagnets, releasing the fail-safe brakes and allowing the
counterweight cable pulleys to spin freely thereby enabling vertical motion of the
receptor cabinet.
ELECTRICAL OPERATION - MODEL QW-400
See Figure 5-2 in Chapter 5, DIAGRAMS.
A fail-safe brake (24 VDC electromagnets) is located at the top right corner of the ver-
tical frame. When the electromagnet is de-activated (i.e., the Receptor Cabinet Lock
Release Button is not pressed), the fail-safe brake "grabs" the counterweight cable
pulley securely, locking the vertical motion of the receptor cabinet. Pressing the
Receptor Cabinet Lock Release Button energizes the 24 VDC electromagnet, releasing
the fail-safe brake and allowing the counterweight cable pulley to spin freely thereby
enabling vertical motion of the receptor cabinet.
SERVICE
INSTRUCTIONS
4-1
4-2
Chapter 4 Service Instructions
OVERVIEW
This chapter provides instructions to assist the service technician in maintaining the
smooth operation of the Vertical Wall Stand/Frame. The Vertical Wall Stand/Frame
will function reliably when maintained according to the instructions provided in this
chapter. Only properly trained service personnel should service or maintain this
equipment. Safe equipment performance requires the attention of service personnel
who are specifically trained and experienced with medical x-ray apparatus. Applicable
preventive maintenance or any repair service should be performed by these skilled
individuals. Failure to follow manufacturer’s or service personnel’s recommendations
may result in serious injury.
SERVICE MAINTENANCE
TOP CAP
NOTE: MODEL
QW-420 IS SHOWN
IN THIS DIAGRAM, COLUMN BASE
MODELS QW-420-D COVER
AND Q W-420-S
HAVE IDENTICAL
PART LOCATIONS
Figure 4-1. Top Cap, Base Cover and Center Column Cover Locations
2. Remove the Column Center Cover from the Vertical Column (see
Figure 4-1).
TOP BRACKET
TERMINAL
MOUNTING
BLOCK TB1
SCREWS (4X)
11. Carefully, slide the Vertical Column forward until clear of Wall
Mount Bracket, then slowly lower column until horizontal on floor.
12. Remove two (2) ¼-20 x 3/4” screws securing Vertical Pulley
Assembly to Vertical Column (see Figure 4-3).
Figure 4-4. Tilting the Vertical Pulley Assembly to Access Counterweight Cables
13. Tilt the Vertical Pulley out from Vertical Column enough to gain
access to counterweight cable mounting hardware (see Figure
4-4).
14. Remove two (2) ¼-20 x 7/8” screws securing each counterweight
cable to Trolley Assembly.
15. Remove the Vertical Pulley Assembly from column.
16. Position the replacement Vertical Pulley Assembly on the Vertical
Column and tilt forward as shown in Figure 4-4.
17. Route the LEFT and RIGHT counterweight cables over the appro-
priate ends of the Vertical Pulley Assembly. (Do not cross left and
right cables!). Secure each cable to Trolley Assembly using two 1/
4-20 x 7/8” screws. Ensure screws are tightened securely.
18. Insert the Vertical Pulley Assembly fully into Vertical Column and
secure to column using two (2) ¼-20 x 3/4” screws.
19. Carefully lift Vertical Column to vertical position and slide into the
location on floor where it was originally installed.
20. Install four (4) ¼-20 x 3/4” hex head screws securing the Vertical
Column Wall Mount Bracket to Vertical Column (see Figure 4-2).
21. Install hardware to secure the Vertical Column Base to floor.
•Part description
•Quantity required
•P.O. Number
•Shipping Instructions
When ordering components or parts not listed in Table 4-1, a complete description
of the part, including its function and location should be provided with the model
number and serial number of the unit.
Table 4-1. Replaceable Parts
*Note: On QW-420-D systems interfaced with IDC Xplorer receptor, use Handle
Assembly part number AY20-201.
4 4
6
Figure 4-7. Replaceable Parts Location Diagram -
Models QW-420, QW-420-D, QW-420-S
11
14 15
13
10 10
12
16
17
18 2
19
20
DIAGRAMS
5-1
5-2
Chapter 5 Diagrams
BLACK
BLACK
A9 INTERFACE BOARD
RED
RED
JR2
1 24 VDC RED D2
APTB2 -
2 24 RTN BLACK
-
1N4005 D1
1N4005
RED
BLACK
BLACK BLACK
BLACK
HAND SWITCH
SW1 6
5
4 GROUND STUD
COM N/O
3
BLACK 2
1
RED
SC50-006T REV. C