Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
PHASED ARRAY
ULTRASONIC INSPECTION
GENERAL PROCEDURE
FOR WELDS
CONTENTS
DESCRIPTION
PAGE
COVER PAGE
CONTENTS
SECTION
1. INTRODUCTION
3. REFERENCES
4. PERSONNEL QUALIFICATION
6. EQUIPMENT
7. CALIBRATION
9. ACCEPTANCE CRITERIA
12
10.
12
REPORTING
This procedure describes the contact ultrasonic method in which the probe makes direct
contact with the material and the ultrasonic instrument is a phased array unit. The probes
addressed by the instrument may be phased array or single element probes.
2. PURPOSE AND SCOPE
This procedure defines the requirements of mechanized pulse echo mode contact ultrasonic
examination of welds. This procedure shall apply to the general aspects of ultrasonic
inspection of welds using the phased array techniques and equipment.
These aspects include;
Details of scanning shall be found in the associated technique sheets. Scanning for transverse
flaws may be done using either manual, fully automatic or mechanized techniques.
This procedure is applicable to ultrasonic examination of welded components for new and in
service welds.
Provisions are incorporated into this procedure and the special adaptations unique to their
implementation are described in this procedure and the associated standard practices
referenced by this procedure. Since the qualification of this procedure and associated standard
practices and techniques is unique to the EPOCH-1000i Phased Array instrument, this
procedure is not valid with any other instrument.
This procedure is valid for only those techniques that are identified and are recorded in the
latest revision of this procedure.
3. REFERENCES
1. Surface Preparation
The scanning area shall be rendered free of dirt, loose scale, weld splatter, machining
particles, or corrosion products. Where the surface cannot be made suitable for ultrasonic
inspection by light brushing and wiping, the Client representative and IES-Industrial
Engineering Service shall be advised. The base material or weld shall be prepared as
needed to permit examination.
2. Weld Marking and Inspection Area Location.
Numbers allocated to welds being examined shall designate inspection areas.
Each weld designated for inspection shall be marked with a reference index and position
numbers marked on the examination surface during initial examination. This shall include a
method for identifying the location of the weld centerline. If welds are to be permanently
marked, low stress stamps or vibra-tools may be used such that the marks are not deeper
than 1mm.
Unless specifically noted otherwise, all circumferential measurement shall be made clockwise
from the zero reference position on the weld when looking in the direction of product flow.
6. EQUIPMENT
Examination Equipment
1. Ultrasonic Test Instruments
This procedure shall use the EPOCH-1000i phased array ultrasonic inspection unit exclusively.
The ultrasonic flaw detection instrument shall be used in the pulse echo or pitch-catch mode
with an A-Scan presentation used for calibration. The instrument is equipped with a stepped
gain control calibrated in units of 1dB or less. The equipment is capable of amplitude control
over its Useful range and Time Corrected Gain (TCG) shall be used in lieu of a simple Distance
Amplitude Correction (DAC) curve.
2. Search Units
Search units used for weld examinations shall operate at a frequency between 5 MHz. either
phased array or single element probes may be used. Phased array linear array configurations
shall have 16 elements or more having element width of 6mm to 15mm. Element spacing will
be search unit design specific and based on the calculations for element pitch to provide beam
steering of +/- 20. For typical shear wave applications the phased array probe will be placed
on a refracting wedge of suitable material and refracting angle as to be compatible with the
requirement to provide beam steering of +/- 20.
In addition to phased array search units, pulse-echo single element probes may be used where
geometry or resolution or other factors dictate that there is advantage to do so such as for
thickness assessments.
3. Couplant
Couplant used shall contain no ingredients deemed harmful to the test piece. For automated
or mechanised scanning, water shall be used. Alternatively, Ultragel or Glycerine or Lub oil or
soft grease may be used and diluted with water to reduce viscosity so as to facilitate its
movement through irrigation lines. Ultragel or Glycerine is also approved for use on the work
piece when performing manual transverse scans.
4. Calibration Blocks
Weld inspections using the EPOCH-1000i Ultrasonic Flaw detector shall be calibrated using
standard calibration blocks and blocks fabricated from material that is acoustically similar to
that being inspected and which has been subjected to minimum heat treatment required by
the material specification. The surface of calibration blocks is finished to approximately 10
micron Ra. At no time shall the calibration block surface be inferior to the test piece surface.
Calibration blocks are of 2 types
a. Range setting calibration blocks
b. Sensitivity setting calibration blocks.
a. Range Setting Calibration Blocks
Blocks used to set instrument range include:
1. International Institute of Welding (IIW) (BS A2) for range calibration
2. Din 54-122 (V2) (BS A4) alternative for range calibration
3. Standard Thickness Calibration Block (ladder step wedge)
Other range setting blocks may be used provided they are demonstrated to provide the
accuracy required for the materials examined.
Thickness of
Material to be
inspected (mm)
Thickness of block
(mm)
10< t50
40 or t
50 < 100
75 or t
Diameter Hole
Distance of Hole
from Surface
2mm (+/-0.2mm)
125 or t
175 or t
allowed and
225 or t
recommended
t > 2500
275 or t
6mm (+/-0.2mm)
For examinations in materials with diameters greater than 500 mm, a block of essentially the
same curvature, or alternatively, a flat basic calibration block, may be used.
For examinations in materials with diameters 500 mm and less, a curved block shall be used.
Unless otherwise specified, a single curved basic calibration block may be used for
examinations in the range of curvature from 0.9 to 1.5 times the basic calibration block
diameter.
7. CALIBRATION
1. Equipment Calibration
a. Instrument Linearity Checks
All instruments are subject to the following linearity checks before use. Details of the steps
involved in the set-up of equipment for calibration can be found in the Standard Guide for Use
of EPOCH-1000i Ultrasonic Phased array. The results of these checks are recorded in the
Ultrasonic Instrument Linearity
Report forms (see standard practice for Verification of EPOCH-1000i Ultrasonic Phased Array
Linearity Performance) and are kept as part of the Inspection Record.
Instrument Linearity Checks are conducted:
a) on initial receipt of equipment
b) at the commencement of an inspection campaign
c) on a twelve-month cycle
d) after any electronic repair or physical damage
e) or for any other reason deemed appropriate at intervals less than 12 months
Screen Height Linearity Ultrasonic Phased array EPOCH-1000i shall provide vertical
amplitude presentation linear to within +/- 3% of full screen height over the range 10%
to 100% of full screen height. The method for evaluating screen height linearity is
outlined in standard practice for Verification of Ultrasonic Phased array EPOCH-1000i
Linearity Performance.
Sweep Range Linearity (Horizontal Linearity) - The ultrasonic instrument must provide
sweep range linearity accurate to +/-2% of the total base line. The method for
evaluating sweep range linearity is outlined in standard practice for Verification of
Ultrasonic Phased array EPOCH-1000i Linearity Performance.
exit point and nominal angle do not apply. As well, probes using curved wedges to fit the
diameter of pipes, and components will not permit measurements of nominal angle and exit
points using the standard calibration blocks (IIW and DIN-54) as they cannot be used on flat
surfaces. Standard Practice for Verification of Performance of Probes used with EPOCH- 1000i
will be used as appropriate to assess the suitability of the probe and focal laws applied to
phased array probes. Guidance for Probe Performance is also found in this Standard Practice
for assessment of single element probes when used with the EPOCH- 1000i phased array
ultrasonic instrument.
Search unit Checks are conducted:
a) on initial receipt of equipment
b) at the commencement of an inspection campaign
c) on a one-month cycle
d) after any electronic repair or physical damage
e) or for any other reason deemed appropriate at intervals less than 1 month
2. Calibration - Weld Examination
1. Range and Signal Amplitude (TCG) Calibration
The EPOCH- 1000i is calibrated for each examination specified on the detailed technique for
both range and signal amplitude.
Range setting for both shear and longitudinal wave examination is detailed in Standard
Practice for Setting Range using TD Handy Scan
Time Corrected Gain is applied as detailed in Standard Practice for Distance Amplitude
Correction using EPOCH- 1000i
2. Calibration Confirmation
A system calibration confirmation shall be used to verify the sweep range calibration and DAC
and or TCG curve at the start and finish of each examination and at a minimum of once each
day during an examination session using the appropriate calibration block
Any change in search units, shoes, couplants, cables, ultrasonic instruments, personnel or
other parts of the examination system is cause for a calibration check.
Calibration checks are performed as follows:
The sensitivity control settings are adjusted to match those recorded for the calibrated
reference sensitivity of the middle focal law of a single angle Phased Array scanning
technique or the single element probe for non-phased array probe applications.
The search unit is positioned to detect the first and last points on the TCG using the
reference reflectors on the appropriate basic calibration block. The maximum signal
response amplitudes and horizontal sweep positions are observed.
If either point along the time base using the TCG has decreased by greater than 2 dB in
amplitude, all recorded data sheets since the last calibration check are marked void. A
new calibration is made and recorded and the voided examination areas re-examined.
If any point along the time base using the TCG has increased by greater than 2 dB in
amplitude, only recorded indications taken since the last valid calibration check need be
reexamined with the correct calibration and their values changed on the data sheets or
files rescanned.
If during system confirmation of calibration, any point on the time base has moved on
the sweep line more than 5% of the total base line, correct the sweep range calibration.
If recordable reflectors are noted on the data sheets, these data sheets are voided, a
new calibration is recorded and the voided examination is repeated.
3. Transfer Values
Loss of sound transmission due to surface geometry, or surface condition (roughness or
coatings), can be compensated using a transfer value. Transfer value shall be determined
using Standard Practice for Determining Acoustic Compatibility using EPOCH- 1000i.
Any difference between the transfer value obtained on the basic calibration block and the test
piece is added to the Primary Reference and this noted in the report. If a negative dB value
results from the transfer value determination, no adjustment to the primary Reference shall be
made.
4. Temperature Control
For contact examination, the temperature differential between the calibration block and the
examination surface shall be within +/-10C.
5. Instrument Controls
Any controls affecting the linearity of the instrument shall be in the same position for
calibration, calibration checks, instrument linearity checks and examinations.
6. Encoding
Automated scanning shall use position encoders that are verified to provide position accuracy
of scans within +/-1mm in 100mm of scan length. All encoders shall be verified daily to be
correctly calibrated. Moving the scanning apparatus over a fixed distance (e.g. 150mm) and
comparing the start and end readings is sufficient to verify the calibration of the encoder. All
encoded distances recorded by the data acquisition system shall be accurate to within 2mm
over 100mm of actual travel. Encoders failing to maintain this tolerance shall be re-calibrated.
8. EXAMINATION AND RECORDING CRITERIA
Inspection Area
Each inspection area is identified with a unique "weld" number. The width of the inspection
area extends beyond the point of maximum width of the final weld preparation for a minimum
distance of one-half the material thickness or
25 mm, whichever is less. The weld zone is defined as the weld and the base metal up to 10
mm (Heat Affected Zone -HAZ), from the edge of the final weld preparation, as illustrated in
Figure 1. The base material on either side of the inspection area of pipe welds need not be
examined for laminations that would interfere with shear wave examination if these materials
have been inspected for laminar inclusions at some previous stage in fabrication.
HAZ HAZ
Non-Relevant Indications
In respect to weld configuration: With a single V a root indication will be masked by a surface indication
as per 10.0 note 5.
In a 50/50 double V TOFD can support findings like LOP / LOF / center line cracking.
Automated weld inspection for laminar flaws is carried out using a phased array linear raster scan using
focal laws providing a 0 beam in the inspection volume when possible. Restrictions due to geometries
of surroundings shall be reported.
Automated weld inspection for axial flaws is carried out using a phased array linear raster scan using
focal laws providing angled beam examination of the inspection volume. Restrictions due to geometries
of surroundings shall be reported.
Manual or automated transverse scans in two directions either side of the weld cap will also be carried
out where access permits. Restrictions due to geometries of surroundings shall be reported.
All automated scanning shall be position encoded at sampling intervals of 1mm.
Recording
1. General
The fundamental criterion for the evaluation of the existence of an ultrasonic examination indication is
the amplitude of the indication in relation to a specified reference level. The reference level is
established during calibration of the test instrument on the sensitivity calibration block and set so that
the amplitude from the applicable notches or holes are set to 80% full screen height using TCG.
Scanning shall be done at the Primary Reference level plus any gain required for transfer value.
EPOCH- 1000i Phased Array ultrasonic instrument is a digital instrument. Analogue to Digital
Conversion will be at a rate suitable to ensure that the recorded peak signal is within 1dB of the
analogue peak. This shall be considered achieved when the digitization rate is set to at least 5 times the
nominal frequency of the probe used1. Actual digitization settings shall be specified in the Technique.
2. Sequence of Examination Reporting
The Site Examination Record is a record prepared by the ultrasonic technician containing the data
pertaining to an examination of a particular weld. This is the raw data collected in the EPOCH- 1000i
software.The Ultrasonic Inspection Data Record Sheet 2 is the official source of ultrasonic records on
which observed reflectors are plotted, and all inspection results are permanently recorded.
1.
The examination equipment shall be calibrated as described in item 8.0 herein to verify linearity and set
range and sensitivity.
2. Thickness Measurements
Thickness Measurements - Weld Inspection At the examination site, the ultrasonic operator takes an
initial thickness measurement. The thickness is compared to the thickness indicated on the detailed
technique to confirm that the
technique and related range settings are applicable. This thickness assessment may be made using a
separate zero degree probe with EPOCH- 1000i unit or as part of any required scan results from a 0
scan.
Thickness measurements are to be noted at the edge of the weld cap and at 50 mm either side of the
weld at 3 points around the circumference of the weld. The minimum thickness reading is recorded.
3. Shear Wave Scanning and Recording
The instrument is calibrated as described in Standard Practice for Verification of Performance of Probes
used with EPOCH- 1000i, Standard Practice for Setting Range using EPOCH- 1000i, Standard
Practice for Distance Amplitude Correction using EPOCH- 1000i and Standard Practice for
Determining Acoustic Compatibility using EPOCH- 1000i. Calibration checks are performed when
required as outlined in item 8.0 herein.
4. Scanning Sensitivity
Scanning of all welds is performed at a gain setting equal to the Primary Reference Sensitivity plus any
transfer value. On the Data Acquisition display an evaluation threshold is set using a colour threshold at
6dB below the Primary Reference level. Typically Primary Reference level is set at 80% Full Screen
Height and the evaluation level is then 40% FSH.
5. Scanning Normal to the Weld (Axial Scan)
The probe is directed at 90 to the weld centre line and mechanically moved so that the ultrasonic
beam passes through all of the inspection area. This is accomplished by an electronic raster or S-scan.
Where possible, the weld is scanned from both sides of the same surface. The speed of scanning is such
as to ensure that coupling and data acquisition integrity is maintained.
Data recorded will ensure encoded samples along the weld are at 1mm intervals and not greater than
10% of data points shall be missed in a weld scan with not more than 3 adjacent samples missing in a
single scan.
6. canning Parallel to the Weld (Transverse Scan)