FIA8600 Quantitative Immunoassay Analyzer

Cheryl
Overseas Technical
Support
GP Biotech, Inc.

Content
8000

2. Technical specification

Content
8000

1. Appearance of FIA8000(New Version)


Micro Printer

Main interface

Keypad

SD cassete
slot

Test slot
Adapter

Switch
USB port

RJ45 port

Com port

Power
3

Keypad

2. Technical specification

Dimensions (LWH)

250mm x 250mm x 120mm

Weight

1.60 kg

Power voltage

AC 100~240V, 50~60 Hz; DC 12V 5A, supplied via AC/DC adapter in package

Power

60 VA

Operating wavelengths

540 5nm

Measurement range

0mV ~4000mV

Voltage resolution

1mV

Keypad display

4 buttons

Screen size

5.6 inch touch screen, 640*480

Printer

Thermal printer

Data storage

10, 000 data can be stored

3. Operating the FIA8000(New Version)

Turn on the power switch of the analyzer. The system will check its hardware automatically. The self-check
procedure shows in following Figure. After self-check is finished, the system will enter the main interface.

Press the ENT button or touch the Measure icon on the main interface and then enter the measure
interface as shown in following figure.

3.1 Sample queries

In the main interface, click the "Search" icon and enter the search interface. Searching has two ways:
No. and ID number. Number query is system default.

3.2 System settings

Click on the "Settings" icon in the main interface and enter "Settings" interface.

3.2.1 Date & Time setting

When time is not accurate, you can choose "Date & Time" icon to adjust the time.

10

3.2.2 Reset
If setup parameters have a difference or when you need to debug, you can choose to restore the "Reset".

11

3.2.3 Screen Calibration


Screen Calibration is used for re-calibrate the screen when the analyzer is used for a long time as
the subtitles position may shift.

12

3.2.4 About

Click on the "About" icon to check the program version.

13

3.3 Printer test

14

3.4 QC card check

15

QC card demonstration
3

16

3.5 SD card check

SD QC

17

QC (SD)

418

3.6 Quality control

19

Operation Demonstration
4

20

3.7 Update Procedure


3

21

3.7 Update Procedure


7

22

3.8 Switch Language


3

23

3.8 Switch Language


7

24

10

3.8 Switch Language


11

12
13

25

4. Troubleshooting

Err 01: Display board is invalid and C line on the strip is too shallow or
no C Line; the solution: Change a new card.
Err 02: Inconsistent measurement items and samples; the solution:
switch sample mode, re-measure.
Err 03: The barcode on the card surface failed to be identified; the
solution: Change a new card with a clear barcode.
Err 05: Internal parameters error; the solution: Re-QC (SD) or Reset.

8000 8600

26

Summary

2. Technical specification

Content
5. The use of Barcode scanner
6. How to apply the FIA8000
test kits on the FIA8600

27

THANK YOU
FOR YOUR LISTENING!