Sei sulla pagina 1di 41

VISUCAM lite

(effective from Oct. 2003)

Fundus Camera

Service Instructions

Service

SM-30-4044-A1-en

Service

Service
Service

Issue:
Replaces issue:

22.09.2003
-

page

of 41

MS - TS 1

Contents
Seite

Contents............................................................................................................................... 2
Update information.............................................................................................................. 5
General information............................................................................................................. 6
Cleaning Care and Servicing of the unit......................................................................... 6
Tools and test devices ........................................................................................................ 6
VISUCAM lite - Unit................................................................................................................ 7
System overview .............................................................................................................. 7
Design............................................................................................................................... 7
VISUCAM lite connector panel ......................................................................................... 7
Overview of functions ..................................................................................................... 8
Help for troubleshooting ................................................................................................. 8
Help for troubleshooting ................................................................................................. 9
Questions about mechanical problems ........................................................................... 9
Questions about electrical or system problems............................................................... 9
VISUCAM lite - Component Service............................................................................... 10
Replacing the blinking diode of the fixation light............................................................ 10
Opening and closing the camera head ......................................................................... 10
Replacing the halogen bulb .......................................................................................... 10
Separating the camera head from the instrument base................................................. 11
Replacing the VISUCAM lite control cable...................................................................... 11
Replacing the complete lamp housing .......................................................................... 12
Replacing Lamp Control PCB VISUCAM lite .................................................................. 12
Overview connectors lamp control PCB .................................................................... 12
Replacing the fan cooling the lamp housing.................................................................. 13
Replacing the temperature sensor ................................................................................ 13
Replacing Shutter ......................................................................................................... 13
Replacing the light guide .............................................................................................. 14
Exciter filter assembly................................................................................................... 14
Installing / removing the Exciter filter assembly ......................................................... 14
Replacing the stepper motor of the exciter filter assembly......................................... 15
Replacing the opto-coupler of the exciter filter assembly........................................... 15
Replacing the filters of the exciter filter assembly...................................................... 15
Checking the exciter filter drive ................................................................................. 15
Barrier filter assembly ................................................................................................... 16
Installing / removing the barrier filter assembly.......................................................... 16
Replacing the stepper motor of the barrier filter assembly......................................... 16
Replacing the opto-coupler of the barrier filter assembly ........................................... 16
Replacing the filters of the barrier filter assembly ...................................................... 16
Checking the barrier filter drive.................................................................................. 17
Finder camera .............................................................................................................. 17
Replacing the finder camera ..................................................................................... 17
Positioning the finder camera.................................................................................... 17
Adjusting the focus of the finder camera ................................................................... 17
Checking and adjusting the image position of the finder camera ............................... 18
Camera module ............................................................................................................ 19
Service Instructions VISUCAM lite
SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

2 of 41

Replacing the cameras ............................................................................................. 19


Connectors of the Camera module............................................................................ 19
Stepper motor for camera shift .................................................................................. 19
Adjusting the horizontal image positions of the cameras. .......................................... 20
Replacing the stepper motor of the camera module .................................................. 20
Replacing the opto-coupler of the camera module .................................................... 20
Replacing the VISUCAM lite control PCB....................................................................... 21
Opening and closing the control base........................................................................... 21
Opening and closing the control base........................................................................... 22
Replacing the headrest................................................................................................. 23
Components of the control base ................................................................................... 24
Exchanging power supply............................................................................................. 24
Exchanging Motherboard MEDP3................................................................................. 25
Motherboard connectors ........................................................................................... 25
Exchange hard disk ...................................................................................................... 25
Exchange GIF instrument interface............................................................................... 25
Connections of the GIF instrument interface ............................................................. 25
System settings and work with internal programs ...................................................... 26
Activating the system settings....................................................................................... 26
Driver installation from CD ........................................................................................ 26
Mouse ....................................................................................................................... 26
Keyboard .................................................................................................................. 27
Printer ....................................................................................................................... 27
Network..................................................................................................................... 28
Date / Time ............................................................................................................... 30
Regional settings....................................................................................................... 30
Opening the Windows Explorer program ...................................................................... 31
Closing an active window ............................................................................................. 31
Opening watch files (Logserver) ................................................................................... 31
Enter BIOS ................................................................................................................... 31
Password for BIOS.................................................................................................... 31
Installing, uninstalling CD-writer.................................................................................... 31
About window ................................................................................................................ 32
Opening the Test Tools program .................................................................................. 32
Overview Test Tools program....................................................................................... 32
Test and adjustment of mechanical and electrical functions ..................................... 33
Testing focus drive ....................................................................................................... 33
Test Focus drive for +25dpt....................................................................................... 33
Test Focus drive for 25dpt....................................................................................... 33
Testing grabber function and quality of the sensors...................................................... 33
Data input fields............................................................................................................ 34
Driver information field .............................................................................................. 34
Enter defaults............................................................................................................ 34
Save data.................................................................................................................. 34
Temperature levels field ............................................................................................ 34
Sensor field ............................................................................................................... 34
Input and output signals of the electronic system.......................................................... 35
Checking the potentiometer input.............................................................................. 35
Checking the temperature sensor ............................................................................. 35
Checking the voltages with A/D converter. ................................................................ 35
Service Instructions VISUCAM lite
SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

3 of 41

Checking of LED white, LED IR, fan, shutter camera supply ..................................... 35
Checking the joystick switch for right/left recognition and the opto-couplers.............. 35
Checking brightness adjustment using the service program...................................... 35
Checking the flash function ....................................................................................... 35
Measuring the operating voltages................................................................................. 36
Voltages present at Interface GIF PCB ..................................................................... 36
Voltages present at the VISUCAM lite lamp Control PCB ........................................... 36
Voltages present at the VISUCAM lite control PCB..................................................... 37
Voltages and LEDs on the camera control PCB ........................................................ 37
Software VISUCAM lite .................................................................................................. 38
Folder Database....................................................................................................... 38
Folder Images .......................................................................................................... 38
Folder Sequences .................................................................................................... 38
Folder TNIs.............................................................................................................. 38
VISUCAM Save and Update folder........................................................................... 39
VISUCAM Installs folder........................................................................................... 39
VISUCAM HP Driver folder....................................................................................... 39
System check ................................................................................................................. 40
Mechanical check ......................................................................................................... 40
Electrical function check ............................................................................................... 40
Adjustment Circle .......................................................................................................... 41

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

4 of 41

Update information
Issue

Comments

22.09.2003

Issue 1

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

5 of 41

General information
Use the User Manual, SAP No.: 1165-216, as additional literature on the VISUCAM lite fundus
camera. You must be familiar with the contents of the user manual to be able to work with the
service instructions.
The service manual mainly covers the defined spare parts and the description of how the most
important components are replaced.
The replacement modules have been chosen in such a way that replacing them should be
possible without a special training. If in-depth work on the unit is necessary, you should attend
the appropriate service training.
The VISUCAM lite fundus camera does not have to be installed by service staff. For this reason,
no Service Report is necessary.
No fastening components are allocated to and supplied with the spare parts. For this reason,
the fastening components provided in the unit must be re-used when modules are replaced.
Caution:

After the unit has been opened, hazardous voltages are accessible. Take the
necessary precautions against the coming into contact with hazardous voltages and
comply with the applicable work instructions.

Cleaning Care and Servicing of the unit


Also see the Care and Maintenance chapter in the user manual.
The unit requires almost no servicing. We recommend having the unit serviced once a year on
the basis of an appropriate support contract.
The points listed in the "System check" chapter represent a work plan which must be completed
when a unit is serviced.
The optical surfaces can be cleaned using an ether-methylated spirits mixture (9:1), a special
optics cleaning mixture (SM-30-8005-32) or methanol.

Tools and test devices


Material number Benennung deutsch
1239-573 Servicekit VISUCAMlite

English description
Service kit VISUCAMlite

1.

1212-348 Testtool Fokusebene

Test tool for focus plane

2.

1194-369 Schlssel fr Sucherkamera

Key for Finder Camera adjust

3.

0118-751 Sechskantschlssel Zoll 3/32


(Netzteildemontage)

Hex key, inch 3/32


(removal of power supply unit)

4.

1242-737 Recovery CD VISUCAMlite R1.02

Recovery CD VISUCAMlite R1.02

5.

1240-531 Justierfarbtafel

Color test picture

6.

0449-685 Graufilter 2,5%

Gray filter 2.5%

7.

1046-391 Halter fr Demoauge

Holder for Demo Eye

1-6 ist Teil des Servicekits


1177-617 Demoauge mit Halter

Service Instructions VISUCAM lite


SM-30-4044-A1-en

1-6 is part of the service kit


Demo eye with clamp

Issue: 22.09.2003
Replaces issue:

Page

6 of 41

VISUCAM lite - Unit


Notes on servicing:
Always disconnect the power plug before opening the unit.
After servicing, check all functions and, if necessary, the electrical safety of the unit.

System overview
1
2
3
4
5
6
7

Cap of fixation light


Fixation light
Headrest with chinrest
Monitor
USB Flash Drive (Option)
Keyboard, Mouse (PS2)
Instrument base with joystick and
potentiometer for brightness adjustment
8 Camera head
9 Control base
9

Design
The base, mechanical stage, instrument head and headrest form one functional unit.
The connector panel on the base provides the following connections:
1 The VISUCAM lite is switched on/off using power switch (1).
2 The power cord is connected to power inlet (2).
3 The display is connected to VGA connector (3). (The display has its own power supply
and is not switched on/off using the power switch.)
4 2 USB Ports (4) are available for USB devices (Printer).
5 The camera head is connected to DEVICE connector (5).
6 The Mouse is connected to PS2 Mouse connector (6.
7 The Keyboard is connected to PS2 Keyboard connector (7).
8 Label for connector layout
9 A printer can be connected to LPT1 parallel port (9).
10 COM1 serial port (10) is provided for the connection of serial units.
11 NET network connector (11) for integration of the VISUCAM lite in a network.
12 Warning label
VISUCAM lite connector panel

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

7 of 41

Overview of functions
Camera head

Instrument base
+5V

X3
X7

Brightness

X4

X20

Release

X8

X50 Programming

M Stepper
Exciter filter drive
Opto-coupler

X5

HalogenTemperaturbulb
sensor
Fan
+

M Stepper
Barrier filter drive
Opto-coupler

Illumination white

Control of
VISUCAM lite

Shutter

Illumination IR
+12V

X19
X42

X45

X47

X46

X9

X11

X43

M Stepper
Camera change
Opto-coupler
Camera module

Signals

Lamp control

X6

+12V La

X41

+5V1

+12V

+5V
+12V

Line

Insul.
Transf.

Finder camera
IR

X12

115/230V
AC
115VA

X141

X16

Control B/W

X18
X17

+12V B/W

Signals

+12V

+12V
Cameras

X1

X2

X12

X14

X102

X112

X114

X101

MUX

+12V Col.

St9
Camera
St66 Control
St11
PCB
St8
Signal
Y/C

X131

1/2Color Cam.

St4
Signal
B/W

X132
Line

Line

S1
Power
inlet
100...240V
AC

Printer

Headrest
DISPLAY

Fixation

Keyboard

USB Flash Drive


Optional

Mause

Instr.
Table
115/230V,
AC

Connector
panel

DEVICE

X7
+5V, +12V

Power
supply

1/2 B/W Cam.

X5
X6

GIF
Instument
Interface
X4

VGA
VGA
X6

X1

PS2 PS2 LPT1 COM1 NET USB1 USB2


X2
X5
X1
X3
X4

X21
X22

Computer PCB
X17

X19
Primary IDE

+5V

Fan

CD-ROM and
CD -writer

Position sensor
left - right

Hard disk
lite

VISUCAM
Computer Base

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

8 of 41

Help for troubleshooting


Questions about mechanical problems
Can the camera head be swiveled all the way to the right and left?

Can the VISUCAM lite fundus camera be easily moved and locked on the base within the
entire working range?

Is the ophthalmoscope lens clean?

Questions about electrical or system problems


Have all connections been made and all plugs been properly inserted?

Does the computer start up after the power switch has been switched on?

Can the operating software be started?

Can the brightness of the halogen lamp be adjusted?

Does the drive for the exciter filter initialize?

Does the drive for the barrier filter initialize?

Does the illumination of the finder camera function properly for white light and IR light?

Does the finder camera function properly and is the image provided by it displayed on the
monitor?

Can the B/W camera sensor (1/2) be selected for observation?

Can the color camera sensor (1/2) be selected for observation?

Does the focusing system function properly?

Does the B/W camera sensor function properly and can images be captured?

Does the color camera sensor function properly and can images be captured for 45/35?

Can you print with the printer (option) connected?

Have the screen settings (contrast, brightness) been adjusted as required?

Have the screen values (1024x768, 24bit, true color) been adjusted?

Does external archiving function (optional network or USB Flash Drive) properly?

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page

9 of 41

VISUCAM lite - Component Service


Note: Easy-to-exchange components are not described in detail.
Replacing the blinking diode of the fixation
light

Screw out the blinking diode (1) from its


socket(2).
Screw the new blinking diode into the socket.
Check whether it is properly functioning.

1 Blink diode
2 Socket of Blink diode

Opening and closing the camera head

Disconnect the VISUCM C from line


power.
Remove the lamp housing cover.
(Snap lock)
Remove Focus Drive left side and right
side.
Loosen 4 x hex socket head screws
(2.5 mm) in the left-hand cover.
Remove the left-hand cover.
Loosen 4 x hex socket head screws
(2.5 mm) in the right-hand cover.
Remove the right-hand cover.
Note: When removing and attaching the
covers, make sure that the finder
camera cover plate is in the grooves
and not damaged and take care not
to squeeze any cables.
Assemble in reverse order.
Replacing the halogen bulb
Caution: The halogen bulb may be hot. Risk of burns!
Allow the halogen bulb to cool down.
Open the camera head
(see: Opening and closing the camera head).
Remove the halogen bulb including socket
from the lamp mount.
Take care not to bend the mount!
Remove the defective halogen bulb from the
socket.
Insert the new halogen bulb into the socket.
Insert the halogen bulb including socket into
the mount.
Close the camera head
(see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.
Note: The lamp need not be readjusted
after bulb replacement.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page10 of 41

Separating the camera head from the instrument


base
Open the camera head
(see: Opening and closing the camera head).
Disconnect the cable for the camera head X14,
X20 from the connector panel.
Disconnect the 2 cables leading to the
instrument base from the VISUCAM lite control
PCB. Remove the cable ties.

Push the cables through the connecting holes.

Securing ring

Remove the securing ring.


Close the camera head (see: Opening and closing the camera head).
Pull the camera head from the pin and lay it aside.
Mount the new camera head accordingly.
Switch on VISUCAM lite and check its functions.

Replacing the VISUCAM lite control cable


Open the camera head (see: Opening and
closing the camera head).
Disconnect the control cable from the
connector panel.
Disconnect the control cable from the
VISUCAM lite control PCB, the cameras
control PCB and the lamp control PCB.
Remove cable ties and cable clamps from
the cable.
Route the new VISUCAM lite control cable
as per wiring diagram and connect it.
Close the camera head (see: Opening and
closing the camera head).

Cameras control PCB

Control PCB
VISUCAM lite

Control cable

Switch on VISUCAM lite and check its


functions.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page11 of 41

Replacing the complete lamp housing


Open the camera head (see: Opening
and closing the camera head).
Remove the light guide from the top.
Disconnect 2 cables to VISUCAM lite
control PCB.
Remove various cable mounts on the lamp
housing.
Remove fastening screws.
Remove the lamp housing including PCB.
Install the new lamp housing. Observe the
guide slots inside both covers.
Connect the cables.
Secure the light guide in position at the top.
Re-attach the mounts for the other cables.
Close the camera head (see: Opening
and closing the camera head).
Switch on VISUCAM lite and check its functions.

Guide slot
inside cover
Light guide

Securing screw
Light guide, top

Replacing Lamp Control PCB VISUCAM lite


Remove the complete lamp housing (see: Replacing the
complete lamp housing).
Remove the lamp control PCB.
Install the new lamp control PCB.
fan X45
Install the complete lamp housing (see:
X46
Replacing the complete lamp housing).
Temperature sensor

Overview connectors lamp control PCB


Connectors at solder side (access from lamp
housing):
X42 Halogen lamp
X45 Fan
X46 Temperature sensor
X47 Shutter
Connectors at component side:
X41 +12V Lamp supply
X43 Signals

X42

X47

Halogen lamp

Shutter

X43

X41
+12V lamp supply
X141 from DEVICE X1
control cable

Cable to
control PCB VISUCAM lite
X11

Temperature sensor
X46
fan
X45

Halogen lamp
X42
X47
Shutter

Temperatur sensor

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page12 of 41

Replacing the fan cooling the lamp housing


Note: Fan blows outward. Observe markings on the fan!

Open the camera head (see: Opening and closing


the camera head).
Disconnect the fan cable from X45 on the lamp
control PCB.
Screw off the defective fan.
Mount the new fan and connect it to X45
on the lamp control PCB.
Close the camera head
(see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.

Replacing the temperature sensor


Disconnect VISUCAM lite from line power.
Open the camera head (open left half only)
(see: Opening and closing the camera head).
Disconnect the cable of the temperature sensor
from X46 on the lamp control PCB.
Screw off the defective temperature sensor.
Mount the new temperature sensor and connect
it to X46 on the lamp control PCB.
Close the camera head
(see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.

Replacing Shutter
VISUCAM lite vom Netz trennen.
Disconnect VISUCAM lite from line power.
Open the camera head (open left half only)
(see: Opening and closing the camera head).
Disconnect the cable of shutter from X47 on the lamp
control PCB.
Screw off the defective shutter. Shutter wing is still on
place.
Mount the new shutter and connect it to X47 on the lamp
control PCB.
Close the camera head
(see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Shutter
X47

Page13 of 41

Replacing the light guide


Open the camera head (see: Opening and
closing the camera head).
Loosen the upper mount of the light guide
and push the light guide through the hole.
Press out the light guide from the clip.
Loosen the lower mount of the light guide
and remove the light guide.
Insert the new light guide accordingly,
push it all the way into the mounts and
lock it in position.
Close the camera head (see: Opening
and closing the camera head).
Switch on VISUCAM lite and check
its functions.

Securing screw

Clamp

Hole

Clamp screw

Exciter filter assembly


Installing / removing the Exciter filter assembly
Open the camera head (see: Opening
and closing the camera head).
Disconnect the stepper motor cable at X8
and the opto-coupler cable at X5 from the
control PCB and remove the cable mounts.
Screw out 2 fastening screws and remove
the complete module.
Stepper motor, opto coupler or filters can be
replaced.
Re-insert the module and secure it in
position.
Re-route and connect the cables for the
opto-coupler and the stepper motor.
Close the camera head (see: Opening and
closing the camera head).
Switch on VISUCAM lite and check its functions.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page14 of 41

Replacing the stepper motor of the exciter filter assembly

Disconnect the stepper motor cable


X8 from the control PCB and remove
the cable mounts.
Screw out 2 fastening screws and
remove the stepper motor.
Replace the stepper motor by a new
component.
Re-route and connect the cable for
the stepper motor to X8 control PCB.
Close the camera head (see: Opening
and closing the camera head).
Switch on VISUCAM lite and check its
functions.

Stepper

Opto-coupler

Green filter
IR filter

Fluo filter

Replacing the opto-coupler of the exciter


filter assembly
Blue/IR-Cut-filter
Disconnect the opto-coupler cable X5
from the control PCB and remove the
cable mounts.
Filter cover plate
Screw out 2 fastening screws and
remove the opto-coupler.
Replace the opto-coupler by a new component.
Re-route and connect the cable for the opto-coupler to X5 control PCB.
Close the camera head (see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.
Replacing the filters of the exciter filter assembly
Screw out 2 fastening screws and remove the complete exciter filter drive.
Remove the filter cover plate.
Replace the filter you want.
Tighten the filter cover plate.
Replace the exciter filter drive.
Switch on VISUCAM lite and check its functions.
Checking the exciter filter drive
The function of the stepper motor and opto-coupler can be checked using the service software.

Start the service software (see "Opening the


Test Tools program").
Select the "step motors" tab.
Select the "exciter" tab.

Position selection 0-3 permits you to select


the different filters

If you press the "set home position" button,


the 0 position is set.
Manually turn the filter wheel until the slit is
located in the opto-coupler.

The  mark indicates the transmitted light


function of the opto-coupler.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page15 of 41

Barrier filter assembly


Installing / removing the barrier filter assembly
Open the camera head (see: Opening
and closing the camera head).
Loosen the lamp housing and move it
slightly to the side.  (See also:

Replacing the complete lamp housing).


Disconnect the stepper motor cable from
X7 and the opto-coupler cable from X4 on
the VISUCAM lite control PCB and remove
the cable mounts.
Undo 2 fastening screws and remove the
complete module.
Insert and secure the new module.
Re-route and connect the cables for the
opto-coupler and stepper motor.
Mount the lamp housing.
Close the camera head (see: Opening and
closing the camera head).
Switch on VISUCAM lite and check its
functions.

Replacing the stepper motor of the barrier filter assembly


Disconnect the stepper motor cable X7 from the control PCB and remove the cable mounts.
Screw out 2 fastening screws and remove the stepper motor.
Replace the stepper motor by a new component.
Re-route and connect the cable for the stepper motor to
Stepper motor
X7 control PCB.
Close the camera head (see: Opening and closing the
camera head).
Switch on VISUCAM lite and check its functions.
Replacing the opto-coupler of the barrier filter assembly
Disconnect the opto-coupler cable X4 from the control
PCB and remove the cable mounts.
Screw out 2 fastening screws and remove the optocoupler.
Replace the opto-coupler by a new component.
Re-route and connect the cable for the opto-coupler to
X4 control PCB.
Close the camera head (see: Opening and closing the
camera head).
Switch on VISUCAM lite and check its functions.

Opto-coupler
Safety clamp

Red filter

Mounting hole
Fluo filter

Lense
IR-Cutfilter

Replacing the filters of the barrier filter assembly


Unscrew and remove the stepper motor of the barrier filter (see also Replacing the stepper
motor of the barrier filter assembly).
Lift the cover plate from the safety clamp and turn it so that it is clear of the clamp.
Turn the mounting hole of the cover plate to the filter to be replaced.
Replace the filter.
Turn the cover plate back to its initial position making sure that the safety clamp engages in
the hole.
Install the stepper motor of the barrier filter.
Switch on VISUCAM lite and check its functions.
Service Instructions VISUCAM lite
SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page16 of 41

Checking the barrier filter drive


The function of the stepper motor and opto-coupler can be checked using the service software.

Start the service software (see "Opening the


Test Tools program").
Select the "step motors" tab.
Select the "barrier" tab.

Position selection 0-3 permits you to select


the different filters

If you press the "set home position" button,


the 0 position is set.
Manually turn the filter wheel until the slit is
located in the opto-coupler.

The  mark indicates the transmitted light


function of the opto-coupler.
Finder camera
Replacing the finder camera
Open the camera head (see: Opening
and closing the camera head).
Unplug the connections of the finder
camera from the control PCB.
Remove the cable mounts.
Remove the angled bracket including the
finder camera.
Mount the new finder camera, connect the
cables and secure them in position.

Focus ring
IR LED

Positioning the finder camera


Mount one of the two covers.
Move the bracket including the finder
camera in such a way that it rests slightly
against the inner side of the housing.
Tighten fastening screws.

White LED

Adjusting the focus of the finder camera


Switch on VISUCAM lite and start the program (also see the user manual).
Create a test patient file and start the image capture mode.

The finder camera's field of view is visible on the bottom left of the
screen.
Attach a piece of paper displaying centering circles to the headrest.
Adjust brightness as required for observation. Center and focus the
illuminated ring on the centering circles. You can see then structures
of fiber optic.

The rings must be displayed sharply defined. If the rings are


blurred, turn the focusing ring on the finder camera until you have
obtained a sharply defined image.
This completes the mechanical alignment.
Close the camera head (see: Opening
and closing the camera head).

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page17 of 41

Checking and adjusting the image position of the finder camera


Camera Field is max. 768 x 576 pics.
Maximum possible settings are:
a+c 768
b+d 576
b (y)
Finder camera

d
a (x)

c
Illuminated ring

b > image
a > image
d > image
c > image

b < image
a < image
d < image
c < image

The rings should be centered on the field of view. The shadow (visible at the top of the picture)
should not be larger.
If you detect any deviations, the position of the rings can be adjusted in the x and y directions.
The height and width of the image can be adjusted as required.
Open the Windows Explorer (see: Opening the Windows Explorer).
Click the VISUCAM lite software window to bring it to the front of the screen.
Using the ALT + F4 key combination, close the VISUCAM lite program.

Now the Explorer is visible.


Note: The file VECtrl.ini has to be copied and saved before due any changes in it.

Open the VISUCAM


directory.
Open the VECtrl.ini file.

By changing the values of


the marked positions
(arrows), you can change
both the position and the
size of an image.
Tip: changes in the image
do not become visible until
you change the values by 10
to 20.

Save VECtrl.ini.
Start VISUCAM lite
program by clicking the
main.exe file inside the
VISUCAM directory.
Check the position of the finder camera picture.
If it is necessary, repeat the procedure again.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page18 of 41

Camera module
Replacing the cameras
Open the camera head (see: Opening
and closing the camera head)
Remove the cable connections.
Screw out 3 fastening screws and
remove the camera module.
Insert the new module and align it with
the right-hand edge of the housing.
Make electrical connections.
Align the module using the right-hand
housing edge in such a way that the
vertical position of the image on the
monitor is correct.

B/W camera

color camera
Housing edge

Opto-coupler
cameras

Stepper

Housing edge

St5

St66

Connectors of the Camera module


Control PCB VISUCAM lite, +12V, B/W sensor, X18
Control PCB VISUCAM lite, B/W control, X16
Control cable VISUCAM lite, Signals Y/C, X131
Control cable VISUCAM lite, B/W signal, X132
Control PCB VISUCAM lite, +12V, Color sensor, X17

St9

St8

St11
St7

St4

Stepper motor for camera shift


The distance between the opto-coupler and B/W camera ranges between >5 and <50 steps.
Outside this range, an error message is displayed (Overflow position error).
Settings: Before starting the service program, deactivate the date and set the translucency to
bright. Defocus the illuminated ring. No external light should reach the camera
sensors. Darken the room.
Positions of the cameras
>5 steps <50

Opto coupler

B/W camera sensor

color camera sensor

If the positions of both cameras have been


correctly adjusted, both images are centered on
the screen. This is indicated by the black corners.
The extended field of view has an even, circular
shape.
Start the service software (see "Opening the Test
Tools program").
Select the step motors tab.
Select the shift camera tab.
There are two alternatives for positioning:
- Click on 0 or set 1 position without camera function.
Click on 0 or set 1 position with camera function.
Adjust the stepper motor by +1/-1 until the optocoupler is no longer activated.
The  mark indicates the transmitted light
function of the opto-coupler.
If you press the set home position button, the
basic position is resumed (B/W camera sensor,
position 0, color camera sensor position 1). If the
opto-coupler functions correctly, HP-okay is
displayed in the Notification box.
Note: The test image function is not used.
Service Instructions VISUCAM lite
SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Extended field of view

Page19 of 41

Adjusting the horizontal image positions of the cameras.


Note:
The vertical image positions of the cameras are adjusted during the installation of the
camera modules by upward and downward shifting along the housing edge.
First, the image position of B/W camera sensor is checked and adjusted.

Start the service software (see "Opening the Test Tools program").
Select the step motors tab.
Select the shift camera tab.
Click the set home position button
The basic position is set (position of B/W camera sensor).
Click on the 0 button (pos. and power)
B/W camera sensor is switched on.
Check the image position on the monitor. The 4 black corners should have the same size.
To correct the image position, press the +/- steps buttons until the corners are the same
size (+1 step, image moves to ; - 1 step, image moves to ).
Press the use values button.
Save the new position in the EPROM of the VISUCAM lite control system.
Click on the 1 button (pos. and power)
Color camera sensor is switched on. (If the color camera sensor is not switched on, +1 step
button and 1 step button should be clicked to switch the color camera sensor to on).
Check the image position on the monitor. The 4 black corners should have the same size.
To correct the image position, press the +/- steps buttons until the corners are the same
size (+1 step, image moves to ; - 1 step, image moves to ).
Press the use values button.
Save the new position in the EPROM of the VISUCAM lite control system.
Check the positions of the color cameras in the image capturing mode of the normal
program.
Replacing the stepper motor of the camera module
Open the camera head (see: Opening and closing the camera head).
Disconnect the stepper motor cable X9 from the control PCB and remove the cable mounts.
Screw out 2 fastening screws and remove the stepper motor.
Manually move the carriage of the cameras to the basic position (flag in the opto-coupler).
Replace the stepper motor by a new component.
Slide the pilot pin of the motor spindle through the mounting hole.
Firmly secure the stepper motor with 2 fastening screws.
Secure the motor spindle with nuts.
Re-route and connect the cables for the stepper motor at X9 control PCB.
Close the camera head (see: Opening and closing the camera head).
Switch on VISUCAM lite and check the function of the stepper motor.
Check and, if necessary, adjust the horizontal image positions of both camera sensors (see
Adjusting the horizontal image positions of the cameras).
Replacing the opto-coupler of the camera module
Disconnect the opto-coupler cable X6 from the control PCB and remove the cable mounts.
Screw out 2 fastening screws and remove the opto-coupler.
Replace the opto-coupler by a new component.
Re-route and connect the cable for the opto-coupler to X6 control PCB.
Close the camera head (see: Opening and closing the camera head).
Switch on VISUCAM lite and check its functions.
Check and, if necessary, adjust the horizontal image positions of color cameras1 and 2 (see
Adjusting the horizontal image positions of the color cameras).

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page20 of 41

Replacing the VISUCAM lite control PCB


Note:
When exchanging the control PCB, re-enter and save the serial number of the
instrument. Also enter and save the instrument type (see Data input fields).

Open the camera head (see: Opening


and closing the camera head).
Unplug all cable connections.
Screw out 4 fastening screws and remove
the PCB.
Insert the new VISUCAM lite control PCB
and secure it in position.
Re-connect the cables.
Close the camera head (see: Opening
and closing the camera head).
Switch on VISUCAM lite and check its
functions.

X21

X19

Camera
sensor

Barrier filter
stepper motor
X7

Barrier filter
sensor

Camera
stepper motor

Exciter filter
stepper motor

Exciter filter
sensor

Illumination
IR and white
Control Camera

Instrument-base pot

X14 (X114)

X8
X9

X3

X4

X5

X11

X6

X50
X12
(X112)

X1 (X101)
X10

X131 signals Y/C


Color camera sensor
camera pcb St 8

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Instrument base release

X20

X2 (X102)

DEVICE
(X1 interface)
control cable

Lamp-control X43

Programming
X18

X17
X16

+12V,
B/W camera sensor
camera pcb St 66

Control
B/W camera sensor
camera pcb St 9

+12V,
Color camera sensor
camera pcb St 11

X132 signal
B/W camera sensor
camera pcb St 4

Issue: 22.09.2003
Replaces issue:

Page21 of 41

Opening and closing the control base

Removing the camera head including


instrument base.
- Disconnect the cable for the camera
head.
- Loosen the screws in the toothed
rack.
(2.5mm hex socket head)
- Push out the camera head including
instrument base to one side.
- Place the camera head including
instrument base on the table.

Opening the control base


- Put the base including headrest to one
side.
- Remove 8 x 3 mm hex socket head
screws.

- Remove 4 x Phillips screws from


the connector panel.
- Remove the base cover and lay it
aside.

Caution: Take care not to tear off the fixation lamp cable. (see also replacing the headrest!).

Close the base in reverse order.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page22 of 41

Replacing the headrest


Remove the fixation lamp including the blinking
diode.
Loosen the locking screws for the headrest.
Open the base (see: Opening and closing the
control base).
Disconnect the cable for the fixation lamp from X7
on the instrument interface.
Pull the headrest from the clamping mounts.
Push the new headrest into the clamping mounts.
Secure the headrest in position by tightening the
locking screws.
Plug the cable for the fixation lamp into X7 on the
interface.
Close the base (see: Opening and closing the
control base).

Clamping mount
Cable of
fixation lamp

X7
fixation lamp

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page23 of 41

Components of the control base


GIF instrument interface


PS2 Interface
Mouse, Keyboardr

Med PIII Motherboard


Inductive approximation
switch

+5V, +12V
power supply

12VDC fan

Line module

Hard disk

Connection
field


CD-ROM / drive, writer

Notes on the removal and the installation of the base components:


All interchangeable components can be removed and installed without difficulty.
The continuous external saving of images and patient data has to be done by the
customer itself. Both a hard disk with an operating system and a VISUCAM lite program
and a blank hard disk are available.
Exchanging power supply
For exchanging the power supply there is an Allen key 3/32 inch included in the Service kit.
red
black
yellow

+5V
ground
+12V

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Power input

Issue: 22.09.2003
Replaces issue:

Page24 of 41

Exchanging Motherboard MEDP3


Note: To be able to exchange motherboard MEDP3, you first have to remove instrument
interface GIF.
Motherboard connectors
Battery

Note: Do not fix the motherboard


here with a screw!

Hard disk
CD-ROM / burner

X19

X10
X17

Power
supply

X9
X6
VGA Display
Con. panel

User interface
X21
User interface
X22

X5

PS2 Mouse key


board Con. panel

Exchange hard disk


Note: If possible, before exchanging the hard disk a backup of the patient data, network and
printer settings should be done.
Exchange the hard disk (see: components of the control base).
Restore the operating system and the data base using the recovery CD-ROM.
(SM-30-4045-A2).
Reinstall printer driver and run it (optional).
Check the position of the finder camera window (see: checking and adjusting the image
position of the finder camera).
Restore saved patient data.
Reinstall the network settings (optional).
Exchange GIF instrument interface
For exchange see: components of the control base.
Connections of the GIF instrument interface
X1

Camera head

Fixation lamp, head rest


X10 Motherboard

X21

X7
X22

X9 Motherboard

X2
X4

Sensor
left/right movement

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Fan base

X6

Issue: 22.09.2003
Replaces issue:

X3
X5

Power
supply

Page25 of 41

System settings and work with internal programs


Note: The system settings are only accessible to authorized service staff and may only be
changed by this staff. The access to the system settings is protected by a password.
All system settings have been adjusted ex factory as required in the country of
destination.
Activating the system settings
VISUCAM lite is on and in the patient or image

capture mode.
Switch on VISUCAM lite.
Click File in the program bar.

The overview window opens.


Select and click System.

The password window opens.


Enter password iris and click OK.
Note: This Password is only available to service
staff and is provided during service training.

The system selection window opens.


Select and click on the function required
Driver installation from CD
This selection menu permits you to install a driver
from a CD. For installation, select the program
provided for this purpose, e.g. setup.exe.

Mouse
3 menus are available for defining the settings for the mouse.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page26 of 41

Keyboard
2 menus are available for defining the settings for the keyboard.
In the Speed menu, you can change the mode of operation.
In the Language menu, you can adapt the regional settings to the keyboard used.

Printer
Via this selection, you can
access the printer menu. Here,
a printer can be selected as a
Default Printer. You can also
install new printers here.

Selecting printer features


Click File and then
Properties to open the
settings window for the
printer selected.
You can find the data on the
driver and the port
assignment (e.g. USB) in
the Details register.

The changes you will make


in the properties are
archived.

Before removing printer


drivers all applications have
to be closed.

Connect and start up USB


printers during operation. This permits the printers to be recognized by the USB interface,
and the printer drivers can then be installed.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page27 of 41

Network
In the Network menu, you can adjust the settings for the network connection.
The Configuration window shows the protocols used.
In the Identification window, you can enter the computer name assigned by the network
administrator (e. g. ZO01Q73J) and the workgroup or domain used (e. g. DEOKO). You can
also enter a name for the unit connected (e.g. VISUCAM lite) in the Computer Description box.

If a special IP address must be used, open the TCP/IP protocol and select Properties. You can
specify an IP address in he IP Address box.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page28 of 41

Logon to a domain it can be setup in network configuration window.

Click Client for Microsoft


Networks.
Click Properties.
Properties General

opens.
Mark Logon to Windows NT
domain.
Type in domain name (in
this case zeiss).
Select Network logon
options.
Finish with OK.

Restart VISUCAM lite

If you have selected the logon


to a domain, after restart the
system asks you for your login
data of the domain server
(zeiss). This data you get from
the domain administrator.

Note:
If the program asks for new
drivers you can find drivers in
windows directory
c:\WIN98SE\WinInst.
Everery times reboot is
necessary

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page29 of 41

Date / Time
In the Date/Time Properties menu you can
change the time currently set.

The time zone can be adapted to the current


location.

Regional settings
In the Regional Settings menu, you can set the
language zones.

In the Date menu you can set the date.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page30 of 41

Opening the Windows Explorer program


VISUCAM lite is on and in the patient or image

capture mode.
Click the ? in the program bar.

The window with Help Topics and Info opens .


Click Info

The About windows opens.


This window contains important data on
VISUCAM lite including the serial number.
Press the ??? +? key combination.
Note: This key combination is only available to
service staff and is provided during service
training.

The Windows Explorer opens,


providing access to the system settings.

Closing an active window


Use the ALT + F4 key combination to close an active window.
Opening watch files (Logserver)
Use the ALT + w key combination to open newest watch file (watchx.lgg) of the logserver.
Two more watch files are in the folder C:\VISUCAM\Watch(1-3).lgg. To open watchx.lgg is not
possible. X can be no. 1-3. The oldest watchx.lgg will be deleted after start, opened new by
logserver and get the newest data if VISUCAM lite program is active.
Enter BIOS
To enter BIOS, press F2 several times during booting the instrument.
Password for BIOS
The Password is ????? (only available to service staff!)
Installing, uninstalling CD-writer
Install / uninstall new drivers for CD-writer (Plexwriter)
Uninstall: 1. IN CD program
Install:
2. Driver for the CD-writer
Note:

1. Driver for the CD-writer


2. IN CD program

If it is not possible to read any data from a CD-RW, you have to install the
UDF- Reader program on your computer to read files from CD-RW.

Drivers are local at c:\ installs


Service Instructions VISUCAM lite
SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page31 of 41

About window
The About window contains the actual information
about the installed software version, Drivers and the
inserted serial number.

Opening the Test Tools program


VISUCAM lite is running in the patient or image
capture mode.
Click the ? in the program bar.

The window with Help Topics and Info opens.

Click Info

The About window opens.


Press the ??? + ? key combination.
Note: This key combination is only available to
service staff and is provided during service
training.

The Test Tools selector window opens,


providing access to different modules of the
Test Tool.

Note: Service module is designed for service use.


Other modules are not necessary to use.

Click Service

The Test Tool windows opens.

Overview Test Tools program


There are 3 windows to check VISUCAM lite functions or setup data:
in / out window
checks the Input Output
functions

Service Instructions VISUCAM lite


SM-30-4044-A1-en

stepper motors window


stepper motor of exciter filter
stepper motor of barrier filter
stepper motor of camera selector

Issue: 22.09.2003
Replaces issue:

Data window
shows setup or data of
factory settings

Page32 of 41

Test and adjustment of mechanical and electrical functions


Testing focus drive
Test Focus drive for +25dpt
Focus the illumination ring to the first area by moving the
VISUCAM lite camera by hand. You should
see the structure off the light guide inside
the illuminated ring.
Focus with focus drive to the area with the
vertical lines and expect a sharp picture at
the screen.
Note: The distance between the front area and
the back area is 40mm and it should be
possible to focus in this area.
Test Focus drive for 25dpt
Cover front lens with dust cap.
Adjust the brightness of illumination in such a way
you can see a picture at the screen.
Focus to the inside of the dust cap.
It should be possible to get a sharp picture.

Testing grabber function and quality of the sensors


Use the VISUCAM lite in a dimmed area and switch off the room light.
Put the gray filter in between the light guide and the halogen lamp.
Use capture mode.
Focus to the gray field of the color table.
Grey filter
Setup Angle 45
Fluo mode, barrier and exciter filter
b/w sensor
Select maximum brightness of the halogen lamp
Inspect the image on screen. No lines only
snow or noise should be visible.
Snap the image and save it to a test patient to
compare it later time.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page33 of 41

Data input fields


Start the service software (see "Opening
the Test Tools program").
Select the data tab.
Driver information field
type HW
instrument type selection.
DateEE
EEPOM programming date.
Serial Nr. entry of the serial number.
Date VCGIF Grabber driver date.
Enter defaults
Enter the default values (factory settings) for
temperature levels and sensor.
After changing the entries,
click on the save data button.
Save data
The data is saved in the EEPROM of the VISUCAM lite control system.
Temperature levels field
These preset values must not be changed.
high temperature
critical temperature

fixed value 085 corresponds with 50C


fixed value 105 corresponds with 60C
The halogen lamp will be switched off.
An error message will be displayed.

Sensor field
This field shows the camera positions currently stored.
HP-Pos0
B/W camera sensor position in steps (steps between light barrier and B/W camera
sensor)
P0-P1
color camera sensor position in steps (steps between B/W camera sensor and
color camera sensor).

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page34 of 41

Input and output signals of the electronic system


Start the service software (see "Opening the Test Tools program").
The in / out tab is displayed.
By clicking on the relevant function,
you can check the function or switch it
ON and OFF.
The display box shows the respective
values..
Checking the potentiometer input
Click on Rheostat (brightness
potentiometer) and turn the
potentiometer.
Display value ~1 254.
Checking the temperature sensor
Click on Temperature
The display values correspond to:
20C38, 50C85, 60C-105.
From 50C: max. fan speed.
From 60C: instrument shutdown.
Checking the voltages with A/D converter.
Click on the voltage value.
+12V display value 200 5, 5% (12V from power supply unit)
+ 5V display value 50 5, 5% (5V from power supply unit)
+ 5V1 display value 200 5, 5% (5V1 from the VISUCAM lite contro system)
Checking of LED white, LED IR, fan, shutter camera supply
By clicking on the relevant function, you can switch it ON or OFF.
Checking is done visually, acoustically or by measuring the supply voltage.
Checking the joystick switch for right/left recognition and the opto-couplers
Operate the joystick switch.
Display  RHEO of P/O input signal
Move the camera to the left.
Display OS/OD of P/O input signal
Move the camera to the right.
Display  OS/OD of P/O input signal
Light barrier of barrier filter not activated (0 position)
Display  BAR of P/O input signal
Light barrier of exciter filter not activated (0 position)
Display  EXC of P/O input signal
Light barrier of camera shift not activated (0 position)
Display  CAM of P/O input signal
Checking brightness adjustment using the service program
Click on Brightness lamp.
Use the slide to adjust the brightness level. Display level 0-200.
Checking the flash function
Click on Brightness flash.
The slider permits you to select the flash brightness. Display value 0-255.
Click on Flash. to trigger the flash.
Note:
When the flash is triggered, the lamp is set to the selected flash brightness. The
maximum flash brightness is the rated voltage (8V) of the lamp.
The flash duration and flash delay can also be selected.
Further functions not described here are not supported by the program.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page35 of 41

Measuring the operating voltages


Input voltage of the power supply unit (X4)

+12V
+12V
0V
0V
+5V

Voltage supplied to the instrument interface


(GIF)
Voltage supplied to the motherboard

+5V
+5V
+5V
0V
0V
0V
+12V
+12V
0V

0V
0V
+5V
+5V

Output voltage of the power supply unit


+5V (red)
+12V (yellow)
0V (black)
Voltages present at Interface GIF PCB

C13
X21

X4 Sensor left right


movement

U1

X22

1
+5V

2
0V

3
4
P22 +12V

1
+5V

2
0V

3
4
P23 +12V

C2

D2

R2

X1

D3
R3
C4

R18

L5

R1

X4

44
R13

X3

C3

R12

R11

Display from patient sight


left
OS
right OD

C6
C8
C7
C5

R5*

R8
R10
R14

Sensor P22 pin 3


+5.2V without metal
+0.8V with metal and LED on
at the sensor

R4

C9

X7 C14

X6

X2

C11
C12
R7

C10
X8*

R17

X5
1

X7 Fixation lamp

R6

2
0V

L6

1
+5V

+5V logic
0V
0V
+12V lamp
+12V cameras

X6 Fan
X5 Voltage supply
camera head

2
+7V

1
0V

Voltages present at the VISUCAM lite lamp Control PCB


GND
Temperature sensor Fan, max. +11.8V, normal +7V
Shutter, ON +3,6V, OFF 0V
GND
Brightnes, max. 1,44V, min. 88mV

0V 0V 0V +12V +12V

+5V1

X41

X43
1

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Signals

Issue: 22.09.2003
Replaces issue:

Page36 of 41

Voltages present at the VISUCAM lite control PCB


To GIF A/D Converter
X14

1
OUT
2A

2
OUT
2B

2
OUT

3
GND

4
+5V1

5
Not
used

Lamp control

Light barriers

X4
X5
X6
X7
X8
X9

1
GND

1
+5V
+R

3
OUT
1A

2
GND

3
IN

4
GND

5
+5V

1
+5V1

X11

2
Lamp
out

3
GND

4
OUT
1B

4
GND

5
GND

6
GND

7
Fan
out

1
2
3
4
AUDIO VIDEO VIDEO GND
GND
IN

8
GND

5
+12V

9
Temp
in

10
GND

6
7
8
9
+12V LED_IR +12V LED_WT
ON/OFF
ON/OFF

+12V Finder Camera VISUCAM lite / C


+-

Measuring
Points

GIF A/D Base-Poti

+12V
+5V1
+5V
GND

- in+
Lamp control
Exciter

Base-Switch

Cameras

GND
+5V

Finder
Camera
VISUCAMlite

User interface
control cable

(bevor Oct.03)

+
+12V

+-

++12 V Control
B/W camera
sensor

GIF

+12V
Color
Camera
sensor

For VISUCAM lite connection


1
+5V

2
3
4
GND VIDEO VIDEO
IN
GND

7
8
+12V LEDS
ON/OFF

X16

1
2
3
4
5
Gamma DGND FieldFr DGND Gain

6
AGND

Voltages and LEDs on the camera control PCB

LED ON if +12V for


B/W camera sensor
are on.

+12V
+ -

+12V
+ -

LED ON if +12V for


color camera sensor
are on.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page37 of 41

Software VISUCAM lite


Folder Database

The Database folder is on drive D: (partition2) and contains:


the Visucam.mdb database,
the Image folder for all stored images,
the Sequences folder for the video streams
the TNI`s folder for all thumbnails.
Folder Images

The folder Images contains all images (black and white and color) in BMP format.
Folder Sequences

The folder Sequences contains all video streams in AVI format.


Folder TNIs

The folder TNI`s contains all the thumbnails of all stored images in JPG normat.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page38 of 41

VISUCAM Save and Update folder

In the Save folder the files Main.ini and visucam.mdb are saved as the last correct files during
start up. The copies of this files are stored every times if the system is starting up.
In the Update folder there are temporarily data during software update. Normally empty!
VISUCAM Installs folder
A lot of different drivers are placed in this installs folder.

VISUCAM HP Driver folder

In the Hawlett-Packard/HPZ/Glue folder, the install driver of the HP Deskjet 940c are placed.
See also printer installation.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page39 of 41

System check
Mechanical check
Check that the camera head can be swiveled all the way from the left to the right. ............ 
Check that VISUCAM lite fundus camera be easily moved and securely locked on the base
within the entire working range........................................................................................... 
Check that the ophthalmoscope lens is clean .................................................................... 
Check that the connectors are firmly seated. ..................................................................... 
Check that the built-in filters and optics are clean. ............................................................. 
Check all cables for any damage or traces of wear ............................................................ 
Check that all labels and lettering are legible and complete. .............................................. 
Check the glide plate for any traces of wear....................................................................... 
Electrical function check

Check that all electrical connections have been made and securely mounted. .................. 
After switching on VISUCAM lite, check the booting process of the computer. ................... 
Listen for any unusual noise made by the fan and the hard disk. ....................................... 
Check whether the operating software is starting properly ................................................. 
Check whether the exciter filter drive is initialized ............................................................. 
Check whether the barrier filter drive is initialized............................................................... 
Check the brightness control of the halogen lamp and its remaining service life. ............... 
Check the illumination for the finder camera. ..................................................................... 
Check the proper functioning of the finder camera and the image produced by it on the
monitor............................................................................................................................... 
Check that the B/W camera sensor and color camera sensor can be selected ................. 
Check that the focusing system is functioning properly and moving easily . ....................... 
Check whether images can be captured using the B/W camera sensor and whether the
images can be saved ........................................................................................................ 
Check whether images can be captured using the color camera sensor and whether the
images can be saved . ....................................................................................................... 
Check whether archiving on an external unit is possible
(Option: network, USB Flash Drive, CD writer)................................................................... 
Check if a print out of images with the connected printer is possible (optional) .................. 
Check the screen settings (contrast, brightness, resolution) ............................................. 
Check the screen resolution setting 1024 x 768, 24bit, True Color .................................... 

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page40 of 41

Adjustment Circle
This Circle can be used for finder camera adjustment.

Service Instructions VISUCAM lite


SM-30-4044-A1-en

Issue: 22.09.2003
Replaces issue:

Page41 of 41