MULTIX Swing

Ergonomic, reliable, economical


Data sheet

www.siemens.com/healthcare

MULTIX Swing
Ergonomic, reliable, economical

MULTIX Swing is a compact, costefficient radiography workplace


with a cassette catapult bucky.
It is prepared for use with film-cassette
or phosphor storage acquisition
systems.
It is especially suited for all acquisitions
of skeletal radiography of the
recumbent and seated patient:
orthopedic diagnostics

An easy-to-use system with many


advantages

Generator

Table

High-frequency X-ray generator with


multi-pulse waveform for diagnostic
acquisition procedures on radiographic
workplaces.
The multi-pulse waveform enables
high beam quality, precise reproducibility
and short exposure times.
The generator is integrated into the
table base to save space.

Patients with body height of up to 190 cm


(75) can be conveniently radiographed
from head to toe without repositioning
(with short tabletop* for smaller rooms:
170 cm / 67).
Additionally, obese patients are imaged
easily, with a table weight capacity of up
to 450 kg (1,000 lbs).

trauma, routine and ER diagnostics


thorax diagnostics
MULTIX Swing system consists of a
patient table with a floating tabletop,
a catapult bucky and an integrated
tube support with tube assembly and
collimator to meet all clinical imaging
requirements.
This system is a universal radiographic
workplace suitable for private practice,
clinics or radiology departments of
hospitals with different types of
examinations utilizing conventional
acquisition systems.

Tabletop
The floating tabletop (standard length:
220 cm / 86.6) provides transverse
movements of 12.0 cm ( 4.7)
and longitudinal movements of
60 cm / + 50 cm ( 23.6 / + 19.7).
The optional short tabletop
(length: 200 cm / 78.7) provides
longitudinal movements of
50 cm / + 40 cm (19.7 / 15.7).

POLYDOROS ITS X-ray generator:

Bucky wall stand*


For radiography of the seated, standing
or recumbent patient with horizontal
beam projection.
Counterweighted, height-adjustable
catapult bucky with oscillating grid (wall
or floor mounting available).

Tube assembly stand


Integrated tube assembly stand for
vertical, oblique, horizontal and lateral
acquisitions.
Synchronous horizontal movement of the
X-ray tube and bucky tray for constant
beam centering.

* Option

MULTIX Swing
Technical Data

Radiographic table
Radiographic table with floating tabletop, transparent grid and IONTOMAT three-field chamber.
Table height

75 cm

Tabletop dimensions

220 cm x 80 cm (standard), 200 cm x 80 cm (short tabletop*)

Radiation absorption

0.6 mm Al eq at 100 kV

Tabletop travel

longitudinal:
60 cm / + 50 cm (standard),
50 cm / + 40 cm (short tabletop*);
transverse: 12 cm

Max. patient load

450 kg (1,000 lbs)

Cassette sizes

13 cm x 18 cm to 35 cm x 43 cm; 5 x 7 to 14 x 17

Scatter radiation grid in bucky

Pb 10 : 1; 40 lines/cm; focused at 115 cm; moving grid

Travel of bucky

30 cm / + 20 cm; movement arrested by electromagnetic brakes

Tabletop-to-film distance

7 cm

Automatic exposure control

standard

Bucky wall stand VERTIX Swing (option)


Bucky wall unit with height-adjustable catapult bucky cabinet, transparent grid and IONTOMAT three-field chamber.
Left- or right-handed operation of the catapult bucky.
Vertical movement

152 cm

Central beam height

38 cm to 190 cm above floor

Cassette sizes

13 cm x 18 cm to 35 cm x 43 cm; 5 x 7 to 14 x 17

Scatter radiation grid

Pb 10 : 1; 40 lines/cm, focus range 110 cm 180 cm, moving grid

Front plate-to-film distance

< 4.5 cm

Radiation absorption

0.55 mm Al eq

Automatic exposure control

standard

Column stand

Longitudinal travel

123 cm; movement arrested by electromagnetic brakes

Vertical travel

149 cm; movement arrested by electromagnetic brakes

Central beam height

36 cm to 185 cm

Rotation of tube around vertical axis

90; lock positions: 0, 90

X-ray tube rotation

120; lock positions: 0, 90

Source-image distance

up to 116 cm at table

* Option

X-ray tube assembly


X-ray tube

OPTIPHOS 135/30/55R

Nominal voltage (IEC 60613)

135 kV

Nominal output (IEC 60613)

30/55 kW

Focal spot, nominal (IEC 60336)

0.6 / 1.2

Anode heat storage capacity

230,000 HU (170,000 J)

Anode operating frequency

50 Hz / 150 Hz

Radiation absorption (IEC 60601-1-3)

2.5 mm Al at 70 kV

Generator
Generator

POLYDOROS ITS 55

POLYDOROS ITS 30

Output (IEC 60601-2-7)

55 kW
(550 mA at 100 kV)

30 kW
(300 mA at 100 kV)

Exposure voltage

40 kV to 133 kV

40 kV to 133 kV

Shortest switching time


with automatic exposure control

2 ms
1 ms

2 ms
1 ms

mAs range

0.5 mAs to 800 mAs

0.5 mAs to 320 mAs

Installation data
Line connection

3-phase, 400 V 15% / + 10%, 50/60 Hz;


3-phase, 440/480 V, 50/60 Hz with optional pre-transformer*

Power consumption

max. 96 kVA (ITS 55) / max. 51 kVA (ITS 30)

Nominal load

25 kVA at 400 V;
27 kVA at 440 V with optional pre-transformer*;
29 kVA at 480 V with optional pre-transformer*

Room height

min. 245 cm

Ambient conditions (operation)

Temperature range:
Relative humidity:
Barometric pressure:

* Option

+ 10C to + 40C
20% to 75% below dew point
700 hPa to 1060 hPa

MULTIX Swing
Technical Data

Accessories
MULTIX Swing
Two handgrips for insertion into the accessory rails of the patient positioning tabletop
An external manual release for exposure
Compression belt*
Lateral cassette holder for lateral acquisitions*
Patient positioning mattress*
VERTIX Swing
Lateral handgrips*
Patient stretch grip*
Babix holder*

* Option

Floor plan (mm) examples

*1 depending on installation (fix SID)


7

The information in this document contains


general descriptions of the technical options
available and may not always apply in
individual cases.
The required features should therefore be
specified in each individual case at completion
of contract.
Siemens reserves the right to modify the design
and specifications contained herein without
prior notice. Please contact your local Siemens
sales representative for the most current
information.
Original images always lose a certain amount
of detail when reproduced.
In the interest of complying with legal
requirements concerning the environmental
compatibility of our products (protection of
natural resources and waste conservation),
we recycle certain components.
Using the same extensive quality assurance
measures as for factory-new components,
we guarantee the quality of these recycled
components.

Global Business Unit Address


Siemens AG
Medical Solutions
Angiography, Fluoroscopic and
Radiographic Systems
Siemensstrasse 1
D-91301 Forchheim
Germany
Telephone +49 9191 18-0
www.siemens.com/healthcare

Global Siemens Headquarters

Global Siemens Healthcare Headquarters

Legal Manufacturer

Siemens AG
Wittelsbacherplatz 2
D-80333 Muenchen
Germany

Siemens AG
Healthcare Sector
Henkestrasse 127
D-91052 Erlangen
Germany
Telephone +49 9131 84-0
www.siemens.com/healthcare

Siemens AG
Wittelsbacherplatz 2
D-80333 Muenchen
Germany

www.siemens.com/healthcare
Order No. A91AX-30001-11T3-7600 | Printed in Germany | AX CRM MC 07083 | 07.2008 Siemens AG