DIGITAL pH METER
DPH-507
Version: 1.2X
OPERATION MANUAL

GLOBAL ELECTRONICS,
3-5-331, Narayanguda,
Hyderabad 500029, India
Ph: +91 40 2322 1391, 2322 6687, FAX: 23222484
E-mail info@ge-india.com
URL: www.ge-india.com

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 1 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

GENERAL SAFETY
Review the following safety precautions to avoid injury and prevent damage to this
product or any products connected to it. To avoid potential hazards, use this product
only as specified. Only qualified personnel should perform service procedures. To Avoid
Fire or personal Injury
Use Proper Power Cord: Use only the power cord specified for this product and certified
for the country of use.
Connect & Disconnect Properly: Do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Ground the Product: This product is grounded through the grounding conductor of the
power cord. To avoid electric shock, the grounding conductor must be connected to earth
ground. Before making connections to the input or output terminals of the product, ensure
that the product is properly grounded.
Connect the Probe Properly: The probe ground lead is at ground potential. Do not
connect the ground lead to an elevated voltage.
Observe all Terminal Ratings: To avoid fire or shock hazard, observe all ratings and
marking on the product. Consult the product manual for further ratings information before
making connections to the product.
Do not operate without Covers: Do not operate this product with covers or panels
removed.
Use Proper Fuse: Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry: Do not touch exposed connections and components when
power is present.
Do not Operate with Suspected Failures: If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Provide Proper Ventilation: Refer to the manuals installation instructions for details on
installing the product so it has proper ventilation.
Do not operate in Wet/Damp Conductions.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 2 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

MANUAL HISTORY:
Status:
Date Last Modified:
Document Owner:
Distribution List:

Public Circulation
23-08-2014
MP
Client: DPH-507 Users
Global: Marketing, Service

Date

Description

Person

Version

14-09-2009
30-04-2012
10-05-2012

Basic Version
Change calibration
Change calibration display

GS/MP
GS/MP
GS/MP

1.20
1.21
1.22

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 3 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

INDEX
Page no
1. Introduction5
2. Salient features.5
3. Theory of operation.6
4. Specifications7
5. Front panel keys operation8
6. Functional flow diagram.9
7. Program mode flow diagram...10
8. Configuration mode flow diagram.10
9. Display...11
10. Front panel LED indications..11
11. Modes of operation..12
12. Run mode...12
13. Configuration mode.13
13.1Password change...14
13.2Baud rate setting.15
13.3calibration standard...16
14. Program mode...17
14.1Calibration mode.17
14.2Service mode....25
15. Load default mode....29
16. Error description...29
Warranty certificate....30

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 4 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

1. Introduction:
GLOBAL
Microprocessor
based
Digital pH, Model DPH 507 is a
precise and rugged instrument
designed to meet the needs of serious
pH and temperature measurement in
Laboratories & Industries. This
instrument
features
Automatic
temperature compensation & Manual
temperature compensation for pH, in
addition to measuring Temperature of
the solution.
This instrument incorporates the latest
state of Art Technology and is housed in a specially treated stress relieved metal cabinet.
All the controls in regular use are provided on the front panel for operational ease. It has
non-blinking bright 7-segment LEDs display of 4 digits for pH/mV and 3 digits for temp.
This instrument has a built-in voltage stabilizer and is capable of operating at 230V-50Hz.

2. Salient feature:
Microprocessor based design with `Self Diagnostics Program the system
Checking itself, the electrode and temperature sensor for faults.
Truly versatile stand along instrument.
Extremely user-friendly keyboard and prompting display for easy operation.
Easy calibration from keyboard
Display of pH and Temperature with selectable mV display for ORP measurement.
Auto / Manual Temperature Compensation with digitally selected temperature.
Poor Electrode condition warning.
Totally environment proof keyboard.
Data stored in NON VOLATILE memory. Totally unaffected by power
Interruptions.
Automatic Display/decimal adjustment depending upon the range.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 5 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

3. Theory of operation
When a pair of electrodes, namely pH sensitive glass and reference electrodes is dipped
in an aqueous solution, they generate an EMF, which is proportional to the pH of the
solution. The magnitude of EMF is also dependent on the temperature of the solution.
Beside this temperature dependent EMF, a certain potential is generated by the
electrode pair and this remains fairly constant for a given pair of electrodes. Since the pH
sensitive surface is glass, the internal resistance between the EMF from such a high
resistance source, an ordinary VTVM cannot be used.
Therefore an electrometer amplifier with high input impedance is used to ensure
high accuracy and a very good performance.
S.No
1
2
3
Display:

Unit
pH
mV
Temp

Range
-2.00 to 16.00
1999
-50 C to + 200 C

7 Digit, High brightness 7 -Segment LED &


8 Ann LEDs
4 Digits for pH/mV,3 Digits for Temp, All 7 for Messages

Annunciation LEDs: ATC, MTC, C, RD, pH, mV, PGM, COM.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 6 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

4. Specifications
pH
Range

: -2.00 to16.00 pH

Resolution

: 0.01 pH

mV
+/- 1999mV
0.1 (-199.9 to +999.9)
1 < -199.9 & > 999.9

Accuracy

: 0.1% 1 digit

0.1% 1 digit

Reproducibility

: 0.01

0.1/1

Temperature Manual
Compensation
: MANUAL /ATC

Temp
-50 to +200C
0.1(-19.9 to 99.9)
1 C <-19.9 & >99.9
0.1% 1 digit
0.1/1

-50C to +200C

Input impedance

: Greater than 1013 Ohms

Input bias current

: Less than 1 pA @ 25C

AP Range

: 75mV

Slope adjustment

: 75% to 125%

Calibration

: pH: Single, Two- point & Three - point


Using any standard of USA: 4.01, 7.00 & 10.00 or
NIST: 4.01, 6.86 & 9.18
Temperature Single & Two point at any temperature

Programming

: Direct Keyboard Entry: 5 Keys

Data entry

: Through user-friendly keyboard and Prompting display.

Options

Dimensions

: 190 x 130 x 60 (w x d x h)

Power

: 230V AC, 50 Hz

RS232C

*Specifications and Data subject to change without prior notice.


*Please contact factory for details of the Display range, Accuracy & Sensor Scaling.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 7 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

5. Front Panel Keys operation:


Program/Calibration/Escape key
Modes: RUN, Configuration, Load default, Program.
If this key is pressed and the instrument powered up while still

pressed, the instrument Configuration mode is initiated.


During normal operation Run mode, if this key is pressed, Program
mode is initiated.
Return from any mode without saving changes (Escape)
Return back to previous screen (Back).

Scroll key
Modes: RUN, Configuration, Load default, Program.
Selects next digit, Load Default Mode (Power ON).
Select ATC or MTC
Select next function in Configuration, Service or calibration mode.
Select next digit when user is entering any numerical data (e.g.
Password entry, Calibration values, etc.)
Increment key
Modes: RUN, Configuration, Load default, Program.
Increments selected digit during data entry.
For each keystroke increments one-value up to 9 and for the next
keystroke roll back to 0 and again starts incrementing.
Decrement key
Modes: RUN, Configuration, Load default, Program.
Decrements selected digit during the data entry.
For each keystroke decrements one value up to 0 and for
next keystroke again starts decrementing from 9.

(PH/mV)/ENT key
Modes: RUN, Configuration, Load default, Program.
This key is used to enter into any of the modes and check pH value in
mV (displayed value equals to the value set in the mV calibrator).
Accepts and stores displayed value.
Accept and store program value changes.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 8 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

6. Functional Flow Diagram:

DPH-507
Power ON starts

ALL Leds are ON

PGM
key
pressed ?

Yes
Configuration Mode

No

SCRL
key
pressed ?

Yes
Default load Mode

No
Enter Run Mode

PGM
key
pressed ?

Prepared by: SS/TS

Reviewed by: MP

Yes
Service,Calibration Modes

File: DPH507V12X OPMANUAL REVA

Page: 9 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

7. Program Mode Flow Diagram:


PROGRAM MODE

Service

Calibration

pH

1pt

2pt

Temp

3pt

1pt

2pt

Disp

Adc

Key

EEPROM

8. Configuration Mode Flow Diagram:

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 10 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

9. Display:
Two digit power ON display.

This will be displayed during the Initial power


up for the device.

The upper row value 7.00 indicates pH and 25.0 in


the lower row indicates temperature in degree
centigrade (C). ATC, MTC can select using scroll
key.

10. Front Panel LED indications:


The Front Panel is having 2 rows of 4 digits LEDs display for pH/mV and 3 digits LEDs
for temperature, including 8 -Annunciation LEDs for ATC, MTC, C, RD, pH, mV, PGM,
COM.

pH LED
mV LED
READ LED
C LED
ATC LED
MTC LED

PGM LED

An ATC LED glows when Temperature is set in ATC mode. It is selected using scroll
key with Sensor connected (Automatic Temperature compensation).
MTC LED glows when the temperature is set in MTC (Manual Temperature
Compensation) mode.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 11 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

Celsius LED indicates temperature is in C (ATC / MTC mode).


Read LED glows when pH is stable while calibration.
pH LED glows when the pH value is displayed.
mV LED glows when a pH value is indicated in mV.
PGM LED glows in other than RUN mode.

11. Modes of Operation:


1. RUN Mode
2. Program mode
CALIBRATION Mode.
SERVICE Mode
3. CONFIGURATION Mode.
4. Load Default Mode

12. Run Mode:


Whenever the device is switched ON and if no key pressed during Power ON, it
goes into RUN mode, it remains in RUN mode until any key press.
The upper row displays the pH value.
The lower row displays temperature value. User can
change MTC temperature in run Mode,only by using
INCR or DECR key if it is in MTC mode

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 12 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

13. Configuration Mode: (Can be activated only while switching ON device)


To enter into CONFIG mode, first press PGM key & then switch ON the device do
not leave PGM key till all LEDs are switched OFF except PGM LED. It displays
Configuration.
Enter correct password (three digits numeric 000 to 999) the device will enter
Configuration mode. In case if you forget the password, all settings including the
password can be set to default setting by loading default during power on set-up. (Refer
to Default Load section).

In Configuration mode, following functions can be set:


1. Configuration & Calibration modes Password can be changed.
2. BAUDRATE can be set for communication.
3. Calibration Stranded selection

Note: Any change in data is saved ONLY after pressing ENT key.
Enter correct password. (In this screen first digit blinks)

/: Increment/decrement selected digit by 1 for


each keystroke, 1st digit.
: Select next digit of password, selected digit
blinks.
ENT: If the entered password is correct then after
pressing the ENT key enters into the Configuration
Mode.
PGM: Abort CONFIG mode & go to RUN mode.
(Note: Default configuration password is 900)

If the entered password is wrong, it will show the


error (ConF Err) screen for 3 sec and return back
to configuration screen. This will repeat until the
correct password is entered.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 13 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

Note: In case if you forget the password, all settings including the password can be set to
default setting by loading default during power on. (Refer to Default Load section).

13.1. Password change


The starting function of configuration mode is Password CHange. Password can be
changed in between the range 000 to 999.

: Goes into baud rate setting screen.


ENT: Enters into Password CHange for
Configuration mode.
PGM: Abort Configuration mode.

13.1.1 Password change for Configuration:


: Goes into password change for Calibration
mode.
ENT: Enters into password CHange for
ConFiguration mode.
PGM: Goes to Password Change screen.
Always 000 will display in Password screen (default).

/: Increment / decrement selected digit by 1 for


each keystroke, 1st digit.
: Select next digit of password, selected digit
blinks.
ENT: New password will be saved & the
Configuration Screen automatically will appear.
PGM: Previous password remains same & new
password will not save.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 14 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

13.1.2. Password change for Calibration mode:


Password change for Calibration mode:

: Goes password change for Configuration mode.

ENT: Enters into password change for


CALibration mode.
PGM: Return to the Password change Screen.

/: Increment/decrement selected digit by 1 for


each keystroke 1st digit.
: Select next digit of password, selected digit
blinks.
ENT: New password will be saved & the Previous
Screen will appear.
PGM: Previous password remain same & new
password will not save and password change
screen will appear.

13.2. Baud rate Setting:


: Goes to the Calibration Standard Setting
screen
ENT: Enter into baud rate Setting.
PGM: Abort Configuration mode.

: baud rate is changed to 19.2


: baud rate is changed to 4.8
ENT: New baud rate (9.6) is saved then baud rate
screen automatically will appear.
PGM: Old baud rate remains same and new baud
rate is not saved then again baud rate setting
screen will appear.
Note: Default baud rate setting is 9.6
Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 15 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

13.3. Calibration standard:


There are two kinds of standards:

USA standard
NIST standard

13.3.1. USA Standard:


: Goes password change Screen.
ENT: Enter into CALibration Standard mode.
PGM: Abort Configuration mode.

: Select Nist standard using the scroll key.


ENT: Accepts the Calibration in USA Standard
and previous Calibration standard screen will
appear .
PGM: Previous Calibration standard remain same,
new Calibration standard will not save and it will
come back to the previous screen.
Note: If USA standard mode is selected calibration will be done for buffers 4.01, 7.00 and
10.01 pH.

13.3.2. NIST Standard:


: Select USA standard using the scroll key.
ENT: Accepts the Calibration in nist Standard
and previous Calibration standard screen will
appear.
PGM: Previous Calibration standard remain same,
new Calibration standard will not save and it will
come back to the previous screen.
Note: If niSt standard mode is selected calibration will be done for buffers 4.01, 6.86 and
9.18 pH.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 16 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

14. Program mode:


When PGM key is pressed in RUN mode following modes can be accessible.

Calibration mode
Service mode

Enter correct password for Programming mode i.e Calibration mode (Password can
be changed in Configuration mode).
In RUN mode(After switch ON the instrument) press PGM key then enters in to the
Calibration mode. It will ask password for Calibration mode as shown in the below
screen.

14.1. Calibration Mode:


: Goes for service mode.
ENT: Displays CAL screen with password blinking.
PGM: Abort the calibration mode & go to RUN
mode.
Enter correct password. (In this screen first digit blinks.)

/: Increment / decrement selected digit by 1


for each Keystroke, 1st digit.
: Select next digit of password, selected digit
blinks.
ENT: Goes into pH calibration mode.
PGM: Abort Program Function & go to RUN mode.

If the entered password is wrong, it will show the


error (CAL Err) screen for 3 sec and return back to
previous screen. This will repeat until the
correct password is entered.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 17 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

Note: In case if you forget the password, all settings including the password can be set to
default setting by loading default during power on. (Refer to Default Load section).
In CALibration mode, two parameters should be calibrated.
They are:
1. pH calibration.
2. Temperature calibration.
Both pH and temperature calibration can be classified further based on number of
points.
Single point
Two point
Three point (pH only)

14.1.1PH calibration:
: It goes for temperature Calibration Screen.
ENT: Enter into pH Single point calibration mode.
PGM: Goes back to calibration mode screen.

14.1.1.1 Single point calibration:


pH
mV
Read
C
ATC

: Goes to 2pt/3pt point calibration.


ENT: Enters into single (1) Point PH calibration.
PGM: Abort the single point calibration.

MTC

: Selects next buffer value 4.01/ 9.18


pH.
PGM: Abort the single point calibration mode & go
to the pH calibration screen

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 18 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

ENT: The display alternates between current pH and required pH value.

ENT: Calibrate for selected pH & busy screen indicates device is validating calibration.
Calibration is valid then device will display Previous screen i.e busy 1p1 otherwise it will
display busy error.

Note: The above procedure is Single point calibration. The same procedure should be
followed for Two point & Three point Calibration.

14.1.1.2 .Two point calibration:


To enter into Two (2) Point PH calibration mode user has to press key after
Single point calibration screen.

: Goes for 3pt/1pt point calibration.


ENT: Enter into two (2) Point calibration mode and
go to first point entry i.e. 6.86
PGM: Abort two-point calibration and goes to pH
calibration screen.

: Selects other buffer values 4.01 / 9.18 (nist) to


be calibrated.
PGM: Abort calibration and goes back to pH
calibration screen.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 19 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

ENT: The display alternates between current pH and required pH value.

ENT: Accept the Current pH value to required pH value & displays the screen of 2P2
(Second point).

: Selects other buffer values 9.18(nist) to be


calibrated.
PGM: Aborts and goes back to pH Calibration
display.

ENT: The display alternates between current pH and required pH value.

ENT: Calibrate for selected pH & busy screen indicates device is validating calibration.
Calibration is valid then device will display Previous screen i.e busy 2p1 otherwise it will
display busy error.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 20 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

14.1.1.3 Three point calibration:


To enter into Three (3) Point CALibration mode, user has to press key after display of
two point calibration on screen.

: Goes for single point /two point calibration.


ENT: Enter into three (3) Point CALibration mode
and go to first point entry i.e. 6.86
PGM: Aborts 3-point calibration and goes back to
pH calibration screen.

: Selects next buffer values 4.01 /9.18(nist) to be


calibrated.
PGM: Abort calibration and goes back to pH
calibration screen.

ENT: The display alternates between current pH and required pH value.

ENT: Accept the Current pH value to required pH value & displays the screen of 3P2
(Second point).

: Selects other buffer values 9.18/6.86 (nist) to be


calibrated.
PGM: Abort calibration and goes to pH calibration
screen.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 21 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

ENT: The display alternates between current pH and required pH value.

ENT: Accept the Current pH value to required pH value & displays the screen of 3P3
(Third point).
: Selects other buffer values 6.86/4.01 (nist)
to be calibrated
PGM: Aborts and goes back to pH calibration
Screen.

ENT: The display alternates between current pH and required pH value.

ENT: Calibrate for selected pH & busy screen indicates device is validating calibration.
Calibration is valid then device will display Previous screen i.e busy 3p1 otherwise it will
display busy error.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 22 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

14.1.2 Temperature calibration:


Note: For temperature calibration sensor should be connected & it should be in ATC
Mode only.
1. This screen appears when the user tries to
calibrate temperature in MTC mode.
2. To select ATC mode user should go to run
mode & Press key.

: Goes pH calibration screen


ENT: Enter into tEMP CALibration function. i.e.
single point calibration
PGM: Aborts temperature calibration.

14.1.2.1 Single point calibration:


: Goes for two point calibration.
ENT: Enter into Single (1) Point CALibration
mode.
PGM: Aborts temperature calibration.
/: Increment /decrement selected digit by 1 for
each Keystroke selected (digit blinks).
: Select next digit selected digit blinks.
ENT: Accept the entered value and then ok screen
will appear as shown in the below screen
PGM: Abort the Single point calibration mode.
The first point entry 0 (zero) is to be loaded followed
bby enter key press which starts Calibration as
shown below.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 23 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

If Calibration is done wrong, in that case the display


will shown Error screen.

14.1.2.2 Two point calibration:


: Goes for single point calibration.
ENT: Enter into two (2) Point CALibration
mode.
PGM: Aborts temperature calibration.
/: Increment /decrement selected digit by 1 for
each Keystroke selected (digit blinks).
: Select next digit selected digit blinks.
ENT: Accept the entered value and go for first
point entry and then Second point entry screen will
appear.
PGM: Abort the two-point calibration mode.
/: Increment/decrement selected digit by 1 for
each keystroke (last digit blinks).
: used to select next digit, selected digit blinks.

ENT: Accept the entered value and then ok screen


will appear as shown in the below screen.
PGM: Abort the two-point calibration mode.

If Calibration done correctly The busy 2P1


screen will appear.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 24 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

If Calibration done wrong in that case the display


Will shown Error screen.

14.2 Service Mode:


In service mode the various functions provided for the user are

Dsp
Adc
Key
EEPROM

: Select next mode of program function calibration


mode.
ENT: Goes into service mode.
PGM: Aborts service mode.

14.2.1 Service Display:


In service mode each bit of 7-seg Leds are tested in a particular fashion.

: Goes for service AdC mode


ENT: Check whole display (every single LED will
glow one by one )
PGM: Goes for service mode screen.

After checking display, goes to service display


screen.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 25 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

14.2.2. Service ADC:


The ADC mode displays 4 parameters in mV, which are
1. pH value.
2. Temperature value.
3. Reference.
4. Ground.

: Selects Service Key mode.


ENT: The value of pH in milli-volts will
display.
PGM: Abort from Service screen.

: The value of Temperature in millivolts will display


PGM: Abort from Service Adc screen.
Eg: pH 000.4mV

: The value of reference in millivolts will display


PGM: Abort from Service Adc screen
Eg: Temperature 250.3 mV

: The value of Ground in millivolts will display


PGM: Abort from Service Adc screen.
Eg: Reference Voltage 0998.8 mV

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 26 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

: The value of pH in milli- volts will display


PGM: Aborts from service Adc screen.
Eg: Ground 0000.0 mv

14.2.3 Service Key:


Service mode is to check the Validity of each key.

: If we press SCRL key then EEPROM mode will


be selected.
ENT: Goes into keypad checking.

Device will display which key is press.

ENT: Indicates that enter key is pressed.

PGM: Indicates that Program key is pressed.

SCRL: Indicates that Scroll key is pressed.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 27 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

DECR: Indicates that decrement key is pressed.

INCR: Indicates that increment key is pressed.

Note: To exit from this function do not press any key for 1min.
14.2.4. Service EEPROM:
: Selects Service dispaly mode.
ENT: If EEPROM is working then display ok. If it
is not working properly then screen will display err
PGM: Goes to Service screen.

This screen will display for a while and again goes


to Service EEPROM screen.

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 28 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

15. Load Default Mode: (can be activated only during Power ON)
To enter into Load default mode, first press SCROLL key & then switch ON the
device, keep key pressed until all LEDs are switched OFF except PGM LED & it
displays load default.
If the user wants to load default values press enter key. Here if ATC mode is already
Selected ATC LED glows while MTC LED is off.

PGM: When PGM key is pressed it aborts


from load default function.
ENT: When enter key is pressed default
values are loaded.

Parameter Default Value Table:


S.No

Parameter

Default Value

Configuration Password

900

2
3
4

Calibration Password
CAL Standard
Baud rate

900
USA
9.6

16. Error Description:


Error

CORRECTION

ADC Error

Bring instrument to factory

EEPROM Error

Change memory chip


Again calibrate with proper calibration
procedure
Enter correct password
while calibration of pH keep temperature
between 0 to 61C

Calibration Error
PASSWORD Error
Temp < 0C or > 61C
while calibration
Entered number is not
within limit

Prepared by: SS/TS

Reviewed by: MP

Please check limit of enter digit

File: DPH507V12X OPMANUAL REVA

Page: 29 of 30

Global Electronics, Hyderabad 500029 India.

Project: DPH 507

Date: 23-08-14

Warranty certificate
Product: :
Model
:
Serial No. :

Digital pH Meter
DPH 507

GLOBAL ELECTRONICS warrants its products, when properly applied and operated
under normal conditions, to be free from faulty workmanship or defective materials for a
period of ONE YEAR from the date of original shipment. Global products repaired or
replaced pursuant to this warranty shall be warranted for the un expired portion of the
warranty applying to the original products. The liability of Global under this warranty shall
exist provided that
a) Global is promptly notified upon discovery of such defects by Buyer; and
b) The defective product is returned to Global, transportation paid by Buyer or at
Globals discretion the said product is inspected and repaired at Buyers site; and
c) Globals examination of such product disclose that there was defect in Global material
or workmanship; and
d) Global nameplates and markings have not been altered or removed.
e) The safety procedures mentioned in this manual are strictly adhered to.
f) The unit is always used within the specifications and environment mentioned in this
manual and never misused.
In the event the defect is determined to be within the terms of this warranty, Global
agrees to repair the product or the defective portion at no charge to Buyer and to pay for
transportation back to Buyer. This warranty does not apply to consumable items like
electrodes, conductivity cells, temperature sensors, fuses etc. or parts which in Globals
opinion have been subject to misuse, neglect, altered or damaged by accident or
abnormal conditions of operation or handling.
Global reserves the right to make changes in design and specifications without
notice and without any obligation of such changes on units purchased. Until superseded
or in contractual form, this warranty is expressly in lieu of all other liabilities and
obligations on the part of Global Electronics.

Place: Hyderabad

for GLOBAL ELECTRONICS

Date:

Prepared by: SS/TS

Reviewed by: MP

File: DPH507V12X OPMANUAL REVA

Page: 30 of 30