Intrinsically safe level probe /


Eigensichere Pegelsonde /
Sonde de niveau scurit intrinsque /
Sonda de nivel de seguridad intrnseca

IL-10

Operating instructions
Betriebsanleitung
Mode demploi
Manual de instrucciones

8.
9.
10.
11.

7.

1.
2.
3.
4.
5.
6.

Page 3-16

Seite 17-30

8.
9.
10.
11.

7.

1.
2.
3.
4.
5.
6.

Page 31-44

D
F
F

IL-10

Detalles importantes para su informacin


Resumen rpido para usted
Signos, smbolos y abreviaciones
Funcin
Para su seguridad
Embalaje
Puesta en servicio, funcionamiento
Mantenimiento, accesorios
Eliminacin de perturbaciones
Almacenaje, eliminacin de desechos
Declaracin de conformidad CE

Pagins 45-58

Informations inportantes
Aperu rapide
Explication des symboles,abrviations
Fonction
Pour votre scurit
Emballage
Mise en service, exploitation
Entretien, accessoires
Elimination de perturbations
Stockage, mise au rebut
Dclaration de confeamit CE

Contenido

8.
9.
10.
11.

7.

1.
2.
3.
4.
5.
6.

Contenu

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Wichtiges zu Ihrer Information


Der schnelle berblick fr Sie
Zeichenerklrungen, Abkrzungen
Funktion
Zu Ihrer Sicherheit
Verpackung
Inbetriebnahme, Betrieb
Wartung, Zubehr
Strbeseitigung
Lagerung, Entsorgung
EG-Konformittserklrung

GB

Important details for your information


A quick overview for you
Signs, symbols and abbreviations
Function
For your safety
Packaging
Starting, operation
Maintenance, accessories
Trouble shooting
Storage, disposal
EC declaration of conformity

Inhalt

8.
9.
10.
11.

7.

1.
2.
3.
4.
5.
6.

Contents

Contents / Inhalt / Contenu / Contenido

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

Certificate ATEX:
Level probe for operation in hazardous areas in compliance with the respective certificate (see
attached EC-type examination certificate DMT 00 ATEX E 045 X).
ATEX Approval ratings:
Gases and mist: Mounting to Zone 0; Installation in Zone 0, Zone 1 and Zone 2.
Dust: Mounting to Zone 20; Installation in Zone 20, Zone 21 and Zone 22.
Mining Category M1, M2.

Use of the product in accordance with the intended use IL-10:


Use the intrinsically safe level probe to transform the hydrostatic pressure into an electrical
signal in hazardous areas.

WIKA level probes are carefully designed and manufactured using state-of-the-art technology.
Every component undergoes strict quality and environmental inspection before assembly and
each instrument is fully tested prior to shipment. Our environmental management system is
certified to DIN EN ISO 14001.

Read these operating instructions before installing and starting the level probe. Keep the
operating instructions in a place that is accessible to all users at any time.
The following installation and operating instructions have been compiled by us with great care
but it is not feasible to take all possible applications into consideration. These installation and
operation instructions should meet the needs of most pressure measurement applications. If
questions remain regarding a specific application, you can obtain further information:
n Via our Internet address www.wika.de / www.wika.com
n The product data sheet is designated as PE 81.23
n Contact WIKA for additional technical support (+49) 9372 / 132-295
If the serial number on the product label gets illegible (e.g. by mechanical damage or repainting), the retraceability of the instrument is not possible any more.

1. Important details for your information

1. Important details for your information

GB

Notice, important information, malfunction.

Instructions for hazardous


areas: Potential danger of
life or of severe injuries.

Caution

Warning

IL-10

Potential danger of burns


due to hot surfaces.

Potential danger of life or


of severe injuries due to
catapulting parts.

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Warning

Warning

Potential danger of life or of


severe injuries.

3. Signs, symbols and abbreviations

If you want to get a quick overview, read Chapters 3, 5, 7 and 10. There you will get some
short safety instructions and important information on your product and its starting. Read
these chapters in any case.

2. A quick overview for you

Knowledge required
Install and start the level probe only if you are familiar with the relevant regulations and directives of your country and if you have the qualification required. You have to be acquainted
with the rules and regulations on hazardous areas, measurement and control technology
and electric circuits, since this level probe is electrical equipment as defined by EN 50178.
Depending on the operating conditions of your application you have to have the corresponding knowledge, e.g. of aggressive media.

Certificate FM/CSA:
Level probe for operation in hazardous areas in compliance with the respective certificate (see
Control drawing No. 2323880).
FM / CSA Approval ratings:
Intrinsically Safe with entity approval for Class I, II and III Division 1, Groups A, B, C, D, E, F, G
and Class I, Zone 0, AEx ia IIC
Dust-ignitionproof for Class II and III, Division 1, Groups E, F, and G.
Non-incendive for Class I Division 2 Groups A, B, C and D

1. Important details for your information / 2. A quick overview for you / 3. Signs, symbols

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

GB

CSA
Canadian Standard Association
The product was tested and
certified by CSA International.
It complies with the applicable
Canadian and US-American
standards on safety (including
explosion protection).

FM
Factory Mutual
The product was tested and
certified by FM Approvals. It
complies with the applicable USAmerican standards on safety
(including explosion protection).

Two connection lines are intended for the voltage supply.


The supply current is the measurement signal.

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

The hydrostatic pressure prevailing within the application is transformed into a standardised
electrical signal through the deflection of the diaphragm, which acts on the sensor element
with the power supply fed to the level probe. This electric signal changes in proportion to the
hydrostatic pressure and can be evaluated correspondingly

4. Function

2-wire

ATEX
European guideline for explosion
protection (Atmosphre=AT,
Explosion=EX)
The product complies with the
requirements of the European
directive 94/9/EC (ATEX) on
explosion protection.

The product complies with the


applicable European directives.

3. Signs, symbols and abbreviations / 4. Function

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

Information about material consistency against corrosion and diffusion can be found in our
WIKA-Handbook, 'Pressure and Temperature Measurement'.
Consider the details given in the EC-type examination certificate as well
as the respective country specific regulations for installation and operation
in hazardous areas (e.g.: IEC 60079-14, NEC, CEC). If you do not observe
Warning
these stipulations, serious injuries and/or damage can occur.

n Select the appropriate level probe with regard to scale range, performance
and specific measurement conditions prior to installing and starting the
instrument.
Warning
n Observe the relevant national regulations (e.g.: EN 50178, NEC, CEC) and
observe the applicable standards and directives for special applications.
If you do not observe the appropriate regulations, serious injuries and/
or damage can occur!
n Please make sure that the level probe is only used within the overload threshold limit all
the time!
n Observe the ambient and working conditions outlined in section 7 Technical data.
n Ensure that the level probe is only operated in accordance with the provisions i.e. as
described in the following instructions.
n Do not interfere with or change the level probe in any other way than described in these
operating instructions.
n Remove the level probe from service and mark it to prevent it from being used again accidentally, if it becomes damaged or unsafe for operation
n Take precautions with regard to remaining media in removed level probes. Remaining media in the pressure port may be hazardous or toxic!
n Have repairs performed by the manufacturer only.

5. For your safety

5. For your safety

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

6. Packaging / 7. Starting, operation

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Warning

IL-10

n Pay attention to any liquid leaking out, for this points to a diaphragm
damage.
n Use the pressure transmitter only if the diaphragm is undamaged.
n Use the pressure transmitter only if it is in a faultless condition as far as the
safety-relevant features are concerned.

Diaphragm test for your safety


It is necessary that before starting the level probe you test the diaphragm, as this is a safetyrelevant component.

7. Starting, operation

n Inspect the level probe for possible damage during transportation. Should there be
any obvious damage, inform the transport company and WIKA without delay.
n Keep the packaging, as it offers optimal protection during transportation (e.g. changing installation location, shipment for repair).
n Ensure that the connection contacts will not be damaged.

Has everything been supplied?


Check the scope of supply:
n Completely assembled level probes
n EC-type examination certificate and Control Drawing (FM, CSA)

6. Packaging

7. Starting, operation

(A)

Mechanical connection

(B)

S#

P#

Coded manufacture date


Signal
Power Supply
Product No.
Order code
Serial No.
PIN assignment

Product label (example)

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Warning

IL-10

n Protect the diaphragm against any contact with abrasive substances and
pressure peaks and do not touch it with tools. If you damage the diaphragm,
no intrinsic safety can be guaranteed (ATEX, FM, CSA)!
n Observe the technical data for the use of the pressure transmitter in connection with aggressive/corrosive media and for the avoidance of mechanical
hazards.

n An additional strain relief is not necessary because the cable has a max. tensile
strength of 1000 N (500 N with FEP).
n The protection cap (A) protects the secluded diaphragm (B) from damaging the
level probe during transport and immersion. Remove the protection cap if used with
viscous or contaminated media.

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

Electrical connection
n Ground the cable screen at one end, preferably in the safe, thus non-Ex,
area (EN 60079-14). For devices with flying leads, the screen is connected
to the housing. The simultaneous connection of housing and cable screen
Warning
to ground is only permitted if ground loop problems between the screen
connection (e.g. at the power supply) and housing can be excluded
(see EN 60079-14).
n If the housing of the transmitter is not grounded and is likely to be subjected
to any electrostatic charging (EN 600079-14, 6.3 Paragraph 4) or if the level
probe is to be operated in Zone 0 (DMT 00 ATEX E 045 X), the cable screen
should be connected to the equipotential bonding.
n Supply the level probe from an intrinsically safe current circuit (Ex ia).
n Consider both the internal capacitance and inductance.
n Cover flying leads with fine wires by an end splice (cable preparation).
n Consider that cables for use in zones 1 and 2 must be checked with a test voltage
between conductor/earth, conductor/screen of more than 500V (AC).

Warning

n Include the screen/sheath of the connection line used as a suspension rope


in the equipotential bonding of the container.
n For seperation of the zones the bulkhead cable gland in the wall must have
the ingress protection according to IEC 60 529 namely for zone 0: IP 67 and
for zone 20: IP 6X.
n Integrate an overvoltage protection, if the mounting position is less than 1 m
away from the transition into Zone 0.

Installation in / mounting to zone 0 and zone 20


(In general Zone 0 is given when the pressure transmitter is surrounded by a mixture of explosive gases more than 1.000 hours per year = continuous hazard).

7. Starting, operation

Load (e.g. display)

Screen/case, grey

green

brown

Hazardous
(classified) area

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Non hazardous
area

2-wire

Positive supply / measurement connection


Negative supply / measurement connection

Power supply

IL-10

With a line transformer you realise the mandatory galvanic isolation of the voltage
and current supply between hazardous and non-hazardous areas and ensure the
safety connection data.

Flying leads, conducter cross


section up
to max. 0.25 mm /AWG 24 with
end splices, conducter outer
diameter 7.5 mm,
IP 68 (Please note "Over pressure
safety" in Chapter 7 Specification)

UB+/Sig+
OV/Sig-

10

GB

n Operate the level probe with a shielded cable and earth the shield at least on one
side of the cable, if the cable is longer than 30m (2-wire), or if it is run outside of the
building.
n There mus be no differences in potential between medium/tank and the grounding of
the junction box and the control cabinet when the shield of the cable is applied.
n Ingress protection per IEC 60529
n Please make sure that the ends of cables with flying leads do not allow any ingress of
moisture.

7. Starting, operation

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

bar

Burst pressure

0.1

0.16

0.25

0.4

0.6

25
42

35

{0.25} 1) 2)
2)

-14 ... +140 F

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Medium 3) 4) 5)

(at reference conditions)

(BFSL) according to IEC 61298-2

-10 ... +60 C

0.2

% of span
Permissible temperature of

1-year stability

% of span

Non-repeatability

0.2
0.1

% of span

Non-linearity

IL-10

11

Including non-linearity, hysteresis, zero point and full scale error (corresponds to
error of measurement per IEC 61298-2).
2)

Adjusted in vertical mounting position with lower pressure connection.

Accuracy { } for pressure ranges 0.25 bar


1)

(BFSL)

{0.125} 1)

0.25
0.5

Insulation complies with EN 50020, 6.4, 12

Accuracy

% of span

Insulation voltage

4 20 mA, 2-wire

42

35

16

RA (UB 10 V) / 0.02 A - (length of flying leads in m x 0.14 Ohm)

42

35

10

4.8

maximum ohmic load RA

42

35

2.4

GB

Signal output and


RA in Ohm

10 < UB 30

Power supply UB

Stainless steel {Hastelloy}


Synthetic oil

Internal transmission fluid

n Case

Stainless steel {Hastelloy}

20.5

17

2.4

PUR {FEP up to 10 bar}

12

10

2.5

1.5

Protection cap

12

10

1.6

Cable

n Wetted parts

UB in VDC

bar

Over pressure safety


Materials

bar
bar

Pressure ranges

Over pressure safety


Burst pressure

bar
bar

Pressure ranges

Model IL-10

Specifications

7. Starting, operation

% of span

n Mean TC of range

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

UB+ towards UB-

n Reverse polarity protection

12

Sig+ towards UB-

n Short-circuit proofness

Wiring protection

kV

Burst

Environmental Category C, F, EMC 1

10

V/m

RF-immunity

IL-10

94/9/EC

n ATEX-Directive ATEX of equipment intended for use in potentially explosive atmospheres

Approval German Lloyd GL

2004/108/EEC, EN 61 326 Emission (Group 1, Class B) and Immunity


(industrial locations)

Group A, B, C, D, E, F, G and Class I, Zone 0 AEx ia II C


Read the operating conditions and safety-relevant data in the EC-type
examination certificate in any case (DMT 00 ATEX E 045 X)

n EMC directive

CE- conformitiy

5)

Instrinsic safe Class I, II, III Division 1,

Class I, II and III

Ignition protection type

Ex-protection

Category 5) 1G (IIA), 1/2 G, 2G, 1D, 1/2D, 2D, M1, M2

0.2 / 10 K

EEx ia I/II C T4, EEx ia I/II C T5, EEx ia I/II C T6

FM, CSA

32 ... +122 F

0.2 / 10 K (< 0.4 for pressure range 250 mbar)

0 ... +50 C

Ignition protection type

ATEX

% of span

Ex-protection

-14 ... +140 F

{-14 ... +185 F with FEP-cable}

GB

Also complies with EN 50178, Tab. 7, Operation (C) 4K4H, Storage (D) 1K4,
Transport (E) 2K3
4)
Other temperature ranges are possible, depending on the electrical connection;
see EC-type examination certificate and table page 13)

3)

-10 ... +60 C

{-10 ... +85C with FEP-cable}

Model IL-10

n Mean TC of zero

Temperature coefficients within


compensated temp range

Compensated temp range

Storage 3)

Specifications

7. Starting, operation

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

Approx. 0.08 per m cable

Approx. 0.2

-30 ... +60 C


-30 ... +105 C
-30 ... +80 C

1 G (IIA), 2 G (IIA), M1, 1 D, 2 D


1 G (IIA)
2 G (IIA), M1
1 D, 2 D

PUR-cable
FEP-cable

-22 ... +176 F

-22 ... +221 F

-22 ... +140 F

14 ... +140 F

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

n WIKA level probes require no maintenance.


n Have repairs performed by the manufacturer only.
Accessories: For details about the accessories, please refer to WIKAs price list, WIKAs
product catalog on CD or or contact our sales department.

8. Maintenance, accessories

Caution

13

n Observe the ambient and working conditions outlined in section 7


Technical data.
n Please make sure that the level probe is only used within the over load
threshold limit at all times!
When touching the level probe, keep in mind that the surfaces of the instrument components might get hot during operation.

Functional test
The output signal must be proportional to the pressure. If not, this might point to a
damage of the diaphragm. In that case refer to chapter 9 Troubleshooting.

-10 ... +60 C

Category

Electrical connections

Warning

GB

Ambience-/ Medium temperature range

Permissible temperature ranges depending on electrical connections

{ } Items in curved brackets are optional extras for additional price.

kg

kg

Weight
Cable

Model IL-10

Specifications

7. Starting, operation / 8. Maintenance, accessories

Contact the manufacturer and replace


the instrument

Diaphragm is damaged, e.g. through


impact, abrasive/agressive media;
corrosion of diaphragm/pressure
connector; transmission fluid missing.

Moisture present (e.g. at the cable tail) Install the cable correctly

Signal span dropping off/too small

Signal span drops off

Diaphragm is damaged, e.g. through


impact, abrasive/agressive media;
corrosion of diaphragm/pressure
connector.

14

Install the cable correctly,


filter element insertion

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Zero point signal unstable/ too low/high Moisture present (e.g. at the cable tail)

IL-10

Ensure permissible temperatures as per


the Operating Instructions

Replace instrument

Medium or ambient temperature too


high/too low

Abnormal zero point signal

Working temperature too high/too low

Control the internal temperature of the


instrument within the permissible range;
observe the allowable temperature error
(see Operating Instructions)

Working temperature too high/too low

Signal span incorrect

Abnormal zero point signal

Ensure permissible temperatures as per


the Operating Instructions

Working temperature too high/too low

Signal span erratic

Ensure permissible temperatures as per


the Operating Instructions

Procedure

Possible cause

Failure

Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the
right voltage supply and the right type of wiring (2-wire) has been chosen?

Do not insert any pointed or hard objects into the pressure port for cleaning to prevent
damage to the diaphragm.

!
Warning

GB

n Take precautions with regard to remaining media in removed level probes.


Remaining media in the pressure port may be hazardous or toxic!
n Remove the level probe from service and mark it to prevent it from being
used again accidentally, if it becomes damaged or unsafe for operation.
n Have repairs performed by the manufacturer only.

9. Trouble shooting

9. Trouble shooting

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

Hot instrument case surface

Make sure that after the assembly the unit is working properly. In case the error continues to exist send in the instrument for
reparation (or replace the unit).

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

15

Storage
Mount the protection cap when storing the pressure transmitter in order to prevent any
damage to the diaphragm.
Disposal
Dispose of instrument components and packaging materials in accordance with the
respective waste treatment and disposal regulations of the region or country to which
the instrument is supplied.

Warning

When storing or disposing of the level probe, take precautions with regard
to remaining media in removed level probes. We recommend cleaning the
level probe properly and carefully. Remaining media in the pressure port
may be hazardous or toxic!

10. Storage, disposal

Process material certificate (Contamination declaration for returned goods)


Purge / clean dismounted instruments before returning them in order to protect our employees
and the environment from any hazard caused by adherent remaining media.
Service of instruments can only take place safely when a Product Return Form has been
submitted and fully filled-in. This Return Form contains information on all materials with which
the instrument has come into contact, either through installation, test purposes, or cleaning.
You can find the Product Return Form on our internet site (www.wika.de / www.wika.com).

USA, Canada
If the problem continues, contact WIKA or an authorized agent for assistance. If the level
probe must be returned obtain an RMA (return material authorization) number and shipping
instructions from the place of purchase. Be sure to include detailed information about the
problem. Level probes received by WIKA without a valid RMA number will not be accepted.

If the problem persists, contact our sales department.

*)

In case of unjustified reclamation we charge the reclamation handling expenses.

Ensure permissible ambient/medium


Permissible ambient or medium tempetemperature limits are observed (see
rature exceeded
Operating Instructions)

Procedure

Possible cause

GB

Failure

9. Trouble shooting / 10. Storage, disposal

16

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

WIKA reserves the right to alter these technical specifications.

Original see Page 59

Text

11. EC declaration of conformity

11. EC declaration of conformity

IL-10

GB

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

IL-10

17

Zulassung ATEX:
Druckmessgert zur bestimmungsgemen Verwendung in explosionsgefhrdeten Bereichen
(siehe beiliegende EG-Baumusterprfbescheinigung DMT 00 ATEX E 045 X).
ATEX Zulassungseigenschaften:
fr Gase und Nebel: Anbau an Zone 0; Einbau in Zone 0, Zone 1 und Zone 2
Stube: Anbau an Zone 20; Einbau in Zone 20, Zone 21 und Zone 22
Bergbau: Kategorie M1, M2.

Bestimmungsgeme Produktverwendung IL-10:


Verwenden Sie die eigensichere Pegelsonde, um in explosionsgefhrdeten Bereichen hydrostatischen Druck in ein elektrisches Signal zu wandeln.

Lesen Sie diese Betriebsanleitung vor Montage und Inbetriebnahme der Pegelsonde.
Bewahren Sie die Betriebsanleitung an einem fr alle Benutzer jederzeit zugnglichen Ort auf.
Die nachfolgenden Einbau- und Betriebshinweise haben wir mit Sorgfalt zusammengestellt.
Es ist jedoch nicht mglich, alle erdenklichen Anwendungsflle zu bercksichtigen. Sollten Sie
Hinweise fr Ihre spezielle Aufgabenstellung vermissen, knnen Sie hier weitere Informationen
finden:
n ber unsere Internet-Adresse www.wika.de / www.wika.com
n Die Bezeichnung des zugehrigen Datenblattes ist PE 81.23
n Anwendungsberater: (+49) 9372/132-295
Wird die Seriennummer auf dem Typenschild unleserlich (z. B. durch mechanische Beschdigung oder bermalen), ist eine Rckverfolgbarkeit nicht mehr mglich.
Die in der Betriebsanleitung beschriebenen WIKA-Druckmessgerte werden nach den
neuesten Erkenntnissen konstruiert und gefertigt. Alle Komponenten unterliegen whrend der
Fertigung strengen Qualitts- und Umweltkriterien. Unser Umweltmanagementsystem ist nach
DIN EN ISO 14001 zertifiziert.

1. Wichtiges zu Ihrer Information

1. Wichtiges zu Ihrer Information

58

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

WIKA se reserva el derecho de modificar las especificaciones detalladas.

Original vase Pagina 60

Text

11. Declaracin de conformidad CE

IL-10

Eliminacin de los desechos


Elimine los desechos de componentes de instrumentos y materiales de embalaje segn
el reglamento respectivo del tratamiento de residuos y eliminacin de desechos de la
regin o del pas donde el instrumento se ha suministrado.

10. Almacenaje, eliminacin de desechos / 11. Declaracin de conformidad CE

2146455.08 GB/D/F/E 08/2009

2146455.08 GB/D/F/E 08/2009

GB

WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio

Original EC declaration of conformity

IL-10

59