Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Revision 1.2
21/08/2007
- ii -
Table of Contents:
MANUFACTURERS WARRANTY STATEMENT........................................................................................viii
SECTION 1
1.1
PRINCIPLE OF OPERATION .................................................................................................................... 1-1
1.2
802 COMPONENTS IN BRIEF .................................................................................................................. 1-1
1.2.1
Electronics Enclosure .................................................................................................................. 1-2
1.2.2
Chassis......................................................................................................................................... 1-2
1.2.3
Keypad ......................................................................................................................................... 1-2
1.2.4
Total Sulphur furnace .................................................................................................................. 1-2
1.2.5
Sample Conditioning System........................................................................................................ 1-2
1.3
GRAPHICAL USER INTERFACE .............................................................................................................. 1-2
SECTION 2
2.1
INSTALLATION CONSIDERATIONS ......................................................................................................... 2-3
2.2
BASIC START-UP PROCEDURE............................................................................................................... 2-4
2.3
INSTALLATION OF TAPE ........................................................................................................................ 2-5
2.4
LEAK TESTING OF COMPRESSION HEAD ............................................................................................... 2-7
2.5
INITIATING CALIBRATION, REFERENCE, AND ZERO ............................................................................ 2-10
2.5.1
Manual Initiation of Reference .................................................................................................. 2-10
2.5.2
Manual Initiation of Calibration................................................................................................ 2-10
2.5.3
Manual Initiation of Zero........................................................................................................... 2-11
2.5.4
Manual Quick Calibration / Zero .............................................................................................. 2-11
2.5.4.1
2.5.4.2
Quick Calibration....................................................................................................................................2-12
Quick Zero ..............................................................................................................................................2-12
2.6
TOTAL SULPHUR PROCEDURES ........................................................................................................... 2-13
2.6.1
Replacing Quartz Reaction Tube ............................................................................................... 2-13
2.6.2
Leak Checking Total Sulphur Furnace ...................................................................................... 2-14
SECTION 3
3.1
ELECTRONICS ENCLOSURE ................................................................................................................. 3-16
3.1.1
Front Panel ................................................................................................................................ 3-16
3.1.1.1
3.1.1.2
LCD Display...........................................................................................................................................3-17
Front Panel LEDs....................................................................................................................................3-19
3.1.2
Electronics Assembly ................................................................................................................. 3-20
3.1.3
AC to DC Power Supply ............................................................................................................ 3-20
3.1.4
Intrinsically Safe Barrier ........................................................................................................... 3-20
3.2
CHASSIS .............................................................................................................................................. 3-21
3.2.1
Sample Chamber Assembly........................................................................................................ 3-21
3.2.1.1
Read/Rate Aperture Strip........................................................................................................................3-23
3.2.1.2
Sensor Block...........................................................................................................................................3-24
3.2.1.2.1 Sensor Block Calibration ..................................................................................................................3-24
3.2.1.2.2 Other Sensor Block Functions ..........................................................................................................3-25
3.2.2
3.2.2.1
3.2.2.2
3.2.2.3
3.2.2.4
3.2.2.5
3.2.3
3.2.3.1
3.2.3.2
3.2.3.3
3.2.3.4
3.3
KEYPAD .............................................................................................................................................. 3-31
3.3.1
Operator Menu .......................................................................................................................... 3-32
Revision 1.2
iii
21/08/2007
3.3.2
Maintenance Menu..................................................................................................................... 3-34
3.3.3
Analyzer Alarm List ................................................................................................................... 3-36
3.3.4
Display ....................................................................................................................................... 3-36
3.4
TOTAL SULPHUR OPTION .................................................................................................................... 3-37
3.5
SAMPLE CONDITIONING SYSTEM ........................................................................................................ 3-38
SECTION 4
4.1
EVENT SEQUENCE .............................................................................................................................. 4-40
4.2
STANDARD ANALYSIS ........................................................................................................................ 4-40
4.2.1
Sample Delay Standard Analysis ............................................................................................ 4-40
4.2.2
Sample Interval Standard Analysis ......................................................................................... 4-41
4.2.3
Tape Saturation Cut-off (TSC)................................................................................................... 4-41
4.2.4
Calculation of Concentration Value .......................................................................................... 4-41
4.3
TIMED ANALYSIS ................................................................................................................................ 4-41
4.3.1
Sample Delay Timed Analysis................................................................................................. 4-41
4.3.2
Sample Interval Timed Analysis.............................................................................................. 4-42
4.4
ANALYSIS MODES .............................................................................................................................. 4-42
4.4.1
Normal Run................................................................................................................................ 4-42
4.4.1.1
4.4.1.2
4.4.2
4.4.2.1
4.4.3
4.4.3.1
4.4.4
4.4.4.1
4.5
Alarm 1 ...................................................................................................................................................4-42
Predictive Alarm Analysis (PAA)...........................................................................................................4-42
SECTION 5
5.1
INSTALLATION AND CONNECTION ...................................................................................................... 5-46
5.1.1
System Requirements ................................................................................................................. 5-46
5.1.2
Software Installation.................................................................................................................. 5-46
5.1.3
Connecting the 802 to the PC .................................................................................................... 5-46
5.1.3.1
5.1.3.2
5.1.3.3
5.2
INTERFACE AND ICONS ....................................................................................................................... 5-52
5.2.1
Toolbar Buttons ......................................................................................................................... 5-53
5.2.1.1
5.2.1.2
5.2.1.3
5.2.2
5.2.2.1
5.2.2.2
5.2.2.3
5.2.2.4
5.2.3
5.2.3.1
5.2.3.2
5.3
APPLICATION SCREENS ....................................................................................................................... 5-58
5.3.1
Watch Window ........................................................................................................................... 5-59
5.3.1.1
5.3.1.2
5.3.1.3
5.3.1.4
5.3.1.5
5.3.1.6
5.3.2
Revision 1.2
Analysis Alarms......................................................................................................................................5-60
Analog Outputs .......................................................................................................................................5-60
Relays .....................................................................................................................................................5-60
Solenoids ................................................................................................................................................5-60
Discrete Inputs ........................................................................................................................................5-61
General Indicators...................................................................................................................................5-61
iv
21/08/2007
5.3.2.1
Graphing Data.........................................................................................................................................5-62
5.3.2.2
Filtering Data ..........................................................................................................................................5-64
5.3.2.2.1 Filtering Data by Date......................................................................................................................5-64
5.3.2.2.2 Filtering Data Numerically...............................................................................................................5-65
5.3.2.3
Exporting Data........................................................................................................................................5-66
5.3.2.4
Log Setup................................................................................................................................................5-66
5.3.3
5.3.4
5.3.4.1
General....................................................................................................................................................5-69
5.3.4.1.1 Tape Box ...........................................................................................................................................5-70
5.3.4.2
Global Alarms.........................................................................................................................................5-71
5.3.5
5.3.5.1
All Stream Parameters ............................................................................................................................5-73
5.3.5.1.1 Sample Timing Method .....................................................................................................................5-73
5.3.5.2
Normal Run ............................................................................................................................................5-74
5.3.5.2.1 Switching...........................................................................................................................................5-74
5.3.5.2.2 Analog Output Managers..................................................................................................................5-75
5.3.5.2.3 Alarms...............................................................................................................................................5-76
5.3.5.3
Calibration ..............................................................................................................................................5-76
5.3.5.4
Reference ................................................................................................................................................5-78
5.3.5.5
Zero.........................................................................................................................................................5-78
5.3.6
5.3.7
5.3.8
5.3.9
Input........................................................................................................................................... 5-79
Output ........................................................................................................................................ 5-80
Ports........................................................................................................................................... 5-81
Modbus ...................................................................................................................................... 5-82
5.3.9.1
5.3.9.2
5.3.9.3
5.3.9.4
5.3.10
5.3.10.1
5.3.10.2
5.3.10.3
5.3.10.4
5.3.10.5
Enron ......................................................................................................................................................5-83
Modicon 16.............................................................................................................................................5-83
Modicon with Floating Point ..................................................................................................................5-83
Typical Modbus Setup ............................................................................................................................5-84
Maintenance............................................................................................................................... 5-85
Calibrating Analog Outputs................................................................................................................5-86
Relays and Solenoids..........................................................................................................................5-87
Tape Encoder Test..............................................................................................................................5-87
ArcNet Threshold (mV)......................................................................................................................5-87
Firmware Upgrade..............................................................................................................................5-87
SECTION 6
SECTION 7
SECTION 8
CONFIGURATION............................................................................................................. 8-96
SECTION 9
SECTION 10
SECTION 11
11.1
11.2
SECTION 12
Revision 1.2
21/08/2007
Figures
FIGURE 1-1: SCHEMATIC OF 802 FLOW PATH ....................................................................................................... 1-1
FIGURE 2-1: 802 POWER SUPPLY CONNECTIONS (FROM LEFT, 24VDC DIV. 2, AC DIV. 2, AC DIV. 1)................ 2-3
FIGURE 2-2: KEYPAD CONNECTOR LOCATION...................................................................................................... 2-4
FIGURE 2-3: STAIN APPEARANCE ......................................................................................................................... 2-5
FIGURE 2-4: TAPE ENCLOSURE PARTS REFERENCE .............................................................................................. 2-6
FIGURE 2-5: INSERTION OF TAPE END INTO TAKE-UP REEL.................................................................................... 2-6
FIGURE 2-6: CORRECT INSTALLATION OF TAPE .................................................................................................... 2-7
FIGURE 2-7: MINIHELIC INDICATING A GOOD SEAL (L); CONNECTED TO VENT ELBOW (R) ................................... 2-8
FIGURE 2-8: SAMPLE CHAMBER ASSEMBLY EXPLODED VIEW ............................................................................. 2-8
FIGURE 2-9: LEAK TESTING THE HUMIDIFIER ....................................................................................................... 2-9
FIGURE 2-10: TOTAL SULPHUR FURNACE ENCLOSURE - EXTERIOR (L) AND INTERIOR (R) ................................ 2-13
FIGURE 2-11: TOTAL SULPHUR QUARTZ REACTION TUBE ................................................................................. 2-13
FIGURE 2-12: EXPLODED VIEW OF ULTRA-TORR FITTING .................................................................................... 2-14
FIGURE 2-13: LOCATION OF CAP FOR FURNACE LEAK CHECK ............................................................................. 2-15
FIGURE 3-1: 802 DIVISION 2 ELECTRONICS ENCLOSURE .................................................................................... 3-16
FIGURE 3-2: 802 FRONT PANEL .......................................................................................................................... 3-17
FIGURE 3-3: INTRINSICALLY SAFE BARRIER CONNECTIONS ............................................................................... 3-20
FIGURE 3-4: 802 DIVISION 2 CHASSIS (ALL COVERS REMOVED)........................................................................ 3-21
FIGURE 3-5: SAMPLE CHAMBER ASSEMBLY EXPLODED VIEW ........................................................................... 3-22
FIGURE 3-6: SAMPLE FLOW ................................................................................................................................ 3-23
FIGURE 3-7: 802 SENSOR BLOCK (FROM LEFT: TOP VIEW, SIDE VIEW, INSTALLED VIEW) ................................ 3-24
FIGURE 3-8: 802 TAPE TRANSPORT SYSTEM ...................................................................................................... 3-26
FIGURE 3-9: PULSE COUNTER ............................................................................................................................. 3-27
FIGURE 3-10: 802 COMPRESSION HEAD.............................................................................................................. 3-28
FIGURE 3-11: NAFION HUMIDIFIER..................................................................................................................... 3-29
FIGURE 3-12: 802 CHASSIS ROTAMETER ............................................................................................................ 3-30
FIGURE 3-13: EDUCTOR ...................................................................................................................................... 3-30
FIGURE 3-14: 802 KEYPAD ................................................................................................................................. 3-31
FIGURE 3-15: OPERATOR MENU SCREEN ............................................................................................................ 3-32
FIGURE 3-16: EDIT PARAMETER SCREEN ............................................................................................................ 3-33
FIGURE 3-17: ANALYZER ALARM LIST ............................................................................................................... 3-36
FIGURE 3-18: DISPLAY MENU ............................................................................................................................ 3-37
FIGURE 3-19: TOTAL SULPHUR FURNACE........................................................................................................... 3-38
FIGURE 3-20: TYPICAL SAMPLE CONDITIONING SYSTEM ................................................................................... 3-39
FIGURE 5-1: 802 DISPLAY BOARD WITH COMMUNICATIONS SELECT DIPSWITCH ............................................... 5-47
FIGURE 5-2: WINDOWS XP USB INSTALLATION SCREEN 1 ................................................................................ 5-48
FIGURE 5-3: WINDOWS XP INSTALLATION SCREEN 2......................................................................................... 5-48
FIGURE 5-4: WINDOWS LOGO TESTING WARNING ............................................................................................. 5-49
FIGURE 5-5: WINDOWS 2000 USB INSTALLATION SCREEN 1 ............................................................................. 5-49
FIGURE 5-6: WINDOWS 2000 INSTALLATION SCREEN 2...................................................................................... 5-50
FIGURE 5-7: WINDOWS 2000 INSTALLATION SCREEN 3...................................................................................... 5-50
FIGURE 5-8: WINDOWS 2000 INSTALLATION SCREEN 4...................................................................................... 5-51
FIGURE 5-9: PORTS IDENTIFIED IN DEVICE MANAGER ....................................................................................... 5-52
FIGURE 5-10: 802 GUI PRIOR TO LOGON ........................................................................................................... 5-53
FIGURE 5-11: CLOSE FIRMWARE UPDATE MESSAGE BOX .................................................................................. 5-53
FIGURE 5-12: CHANGE UPDATE PASSWORD DIALOGUE BOX ............................................................................. 5-55
FIGURE 5-13: RESET UPDATE PASSWORD CONFIMATION DIALOGUE BOX ......................................................... 5-56
FIGURE 5-14: ABOUT WINDOW .......................................................................................................................... 5-56
FIGURE 5-15: COMMUNICATIONS SETUP WINDOW ............................................................................................. 5-57
FIGURE 5-16: LOGON PROGRESS INDICATOR ...................................................................................................... 5-57
FIGURE 5-17: SELECT MODE DIALOGUE BOX .................................................................................................... 5-57
FIGURE 5-18: UPDATE MODE LOGON DIALOGUE BOX ....................................................................................... 5-58
FIGURE 5-19: WATCH WINDOW ......................................................................................................................... 5-59
Revision 1.2
vi
21/08/2007
Tables
TABLE 3-1: RUN CODES .................................................................................................................................... 3-17
TABLE 3-2: STATE CODES ................................................................................................................................ 3-18
TABLE 3-3: MENU EXPLANATIONS ..................................................................................................................... 3-19
TABLE 3-4: FRONT PANEL LED FUNCTIONS....................................................................................................... 3-19
TABLE 3-5: SAMPLE CHAMBER ASSEMBLY PARTS LIST ..................................................................................... 3-22
TABLE 3-6: READ/RATE APERTURE STRIP SIZES AND ASSOCIATED RANGES ..................................................... 3-23
TABLE 3-7: KEYPAD KEY FUNCTIONS ................................................................................................................ 3-31
TABLE 3-8: OPERATOR MENU ............................................................................................................................ 3-33
TABLE 3-9: MAINTENANCE MENU...................................................................................................................... 3-34
TABLE 3-10: TOTAL SULPHUR REACTIONS ......................................................................................................... 3-37
TABLE 4-1: EVENT SEQUENCE ............................................................................................................................ 4-40
TABLE 5-1: COMPUTER REQUIREMENTS FOR 802 GUI ....................................................................................... 5-46
TABLE 5-2: COMMUNICATE TOOLBAR BUTTONS ................................................................................................ 5-54
TABLE 5-3: OPERATION TOOLBAR BUTTONS...................................................................................................... 5-54
TABLE 5-4: GENERAL TOOLBAR BUTTONS ......................................................................................................... 5-55
TABLE 5-5: APPLICATION SCREEN FUNCTIONS................................................................................................... 5-58
TABLE 5-6: WATCH WINDOW FIELDS................................................................................................................. 5-59
TABLE 5-7: CHART BUTTON FUNCTIONS ............................................................................................................ 5-63
TABLE 5-8: CHART MANIPULATION OPTIONS..................................................................................................... 5-64
TABLE 5-9: DATA SORTING OPERATORS ............................................................................................................ 5-65
TABLE 6-1: TROUBLESHOOTING GUIDE .............................................................................................................. 6-89
TABLE 10-1: 802 TECHNICAL SPECIFICATIONS ............................................................................................... 10-103
Revision 1.2
vii
21/08/2007
Operation of the product in unfavourable environments, especially high temperature, high humidity,
Disclaimer
No other warranty is expressed or implied. The manufacturer specially disclaims the implied warranties of
merchantability and fitness for a particular purpose.
Caution
The manufacturer shall not be liable for personal injury or property damage suffered in servicing the product.
The product should not be modified or repaired in a manner at variance with procedures established by the
manufacturer.
Revision 1.2
viii
21/08/2007
Section 1
802 Overview
The Galvanic Applied Sciences Inc. 802 Lead Acetate Tape analyzer (henceforth referred to as the
802) is an analyzer that has been designed to measure relatively low levels of hydrogen sulphide or
total sulphur in natural gas streams. The 802 is capable of measuring up to 2000 parts per million
(ppm) of hydrogen sulphide by volume without the requirement of sample dilution. With a new patentpending analysis algorithm, the 802 has a larger dynamic range and better linearity and repeatability
than previous models of lead acetate tape analyzers.
1.1
Principle of Operation
When hydrogen sulphide, H2S, reacts with lead acetate, the previously white lead acetate begins
to turn brown due to a chemical reaction. The reaction is as follows:
Pb(CH3COO)2 + H2S PbS + 2 CH3COOH
The colour change is caused by the conversion of lead acetate into lead sulphide. When a paper
tape that has been impregnated with lead acetate is exposed to hydrogen sulphide, brown stains
are seen on the tape. The rate of colour change on the tape is directly proportional to the
concentration of hydrogen sulphide in the gas stream. Thus, if the rate of colour change is
measured, the concentration of hydrogen sulphide in the gas stream can be determined. The
ASTM Method used by the 802 is D 4084-82 Analysis of Hydrogen Sulphide in Gaseous Fuels
(Lead Acetate Reaction Rate Method). For more information on how the analysis is carried out,
see sections 4.2 and 4.3.
1.2
The 802 consists of 3 major components the electronics enclosure, the chassis, and the keypad.
The total sulphur version of the 802 has an additional component - the total sulphur furnace.
Many 802s are also equipped with an external sample conditioning system. A schematic of the
802 system is shown in Figure 1-1.
Revision 1.2
1-1
21/08/2007
All of the components in the 802 are described briefly here, and in depth in Section 3.
1.2.2 Chassis
The chassis is the black aluminium box that occupies the top half of the 802. It contains the
motor used to advance the tape, the optics block (detector), the lead acetate tape, the pulse
counter used to accurately advance the tape, the flow meter, and the sample humidifier
(bubbler). The chassis is described in detail in section 3.2.
1.2.3 Keypad
The keypad is attached to the chassis. This intrinsically safe keypad allows the user to review
the configuration of the 802 through the display screen, as well as to make any changes to the
parameters that can be changed through the keypad. Because the keypad is intrinsically safe,
it is able to be used even in classified areas. The keypad is described in detail in section 3.3.
1.3
The Graphical User Interface (GUI) that is included on the 802 configuration disk that is shipped
with the analyzer allows the user to configure and monitor the 802 from a PC. The GUI allows the
user to change many more parameters than are accessible from the keypad. As well, it allows the
user to download, review, and export archived data. Further, it allows the user to configure and
calibrate both analog outputs, the four digital inputs, and all eight outputs, both relays and
solenoids. Finally, it allows the user to monitor the status of the entire analyzer from one screen,
something that is not possible on the 802s LCD screen. The GUI is described in detail in Section
5.
Revision 1.2
1-2
21/08/2007
2.1
Installation Considerations
The 802 should be mounted in a location where it is not exposed to excessive vibration. In
addition, it should be mounted in a location where the ambient pressure remains relatively
consistent over the course of any given day, and where the ambient temperature does not
fluctuate wildly from day to night and winter to summer.
The vent line cannot have any backpressure on it, as backpressure can seriously impact readings.
The vent line should have no vertical bends, and should be as short as possible to allow the
unrestricted flow of sample out to the vent.
Allow approximately 15 on the left hand side of panel-mounted 802s to allow the removal of the
tape cover on the side of the chassis, and to allow easy access for tape installation and removal.
Allow 6 on the left side of cabinet-mounted 802s for the installation of a vent line.
The seals on Class 1 Division 1 802s are not poured at the factory. Before starting use of the 802
in a classified area, the seals must be poured. All explosion proof covers (motor box, electronics
enclosure, 110VAC to 24VDC power supply, and total sulphur furnace enclosure) must be closed
and sealed before powering up the analyzer in a classified area.
For total sulphur 802s, the furnace enclosure contains packing material inside that is used to
prevent damage to the system during shipping. Remove this packing material before powering up
the furnace.
Connect the correct power source. If the 802 has no AC to DC power supply, connect a 24VDC
power supply as shown in the leftmost image in Figure 2-1. If the 802 is a Division 2 AC-powered
unit, connect the AC power source to the AC to DC power supply as shown in the centre image in
Figure 2-1. If the 802 is a Division 1 AC-powered unit, connect the AC power source to the AC to
DC power supply unit as shown in the rightmost image in Figure 2-1. The power terminal block on
the division 1 terminal block is on the left when looking up from underneath the 802. The terminal
block with the red and black wires is the 24VDC output block. Do not turn the power on yet.
Figure 2-1: 802 Power Supply Connections (From left, 24VDC Div. 2, AC Div. 2, AC Div. 1)
Using the wiring diagrams in Section 7, connect the analog outputs as desired. Remember that
the analog outputs on the 802 require loop power, as the 802 does not provide power to them.
Connect any other digital inputs, solenoids, or relay connections as desired.
The dimensions of the 802, as well as the full wiring diagrams for both Class 1 Division 1 and
Class 1 Division 2 are shown in Section 7.
Revision 1.2
2-3
21/08/2007
2.2
Fill the
solution
into the
the two
3. Remove the tape cover on the left side of the 802 chassis. Install the tape as outlined in
section 2.3.
4. Connect sample gas to the 802. If the 802 is equipped with a sample system, set the
pressure regulator to 15psig. If not, ensure that the sample gas is clean and is regulated
to a pressure of 15psig before connecting it to the 802. If the 802 is a total sulphur model,
set the sample flow meter to a flow of 0.5 and the hydrogen flow meter to a flow of 1.5.
Test the furnace for leaks by following the procedure in section 2.6.2.
5. Power up the 802. Wait for the 802s LCD screen to display 0.00ppm. Using the keypad,
enter the correct length of the reel of tape installed (section 3.3.1) 300 foot reels of tape
are included in the crate with the 802.
6. If the 802 is a total sulphur unit, connect to the unit with the GUI (section 5.1) and ensure
that the Furnace Pulse Width in the Global screen (section 5.3.4.1) is set to the correct
value for the AC Voltage being used (50% for 110VAC, 12.5% for 220VAC). WARNING:
Failure to ensure that the furnace pulse width is set to the correct value could result in
failure of the furnace element or melting of the quartz reaction tube.
Revision 1.2
2-4
21/08/2007
7. Follow the leak test procedure in section 2.4 to ensure that the tape is well sealed against
the sample chamber by the compression head.
8. Allow the 802 to run on sample gas for approximately 1 hour. After 1 hour, inspect the
stains on the tape and compare them to those in Figure 2-3. If the stains are crisp and
well spaced, the compression head is sealing the tape well against the sample chamber. If
the stains have fuzzy edges, check the seal again as in step 7, and also check to ensure
that the tape is flat against the sample chamber (i.e. not creased) and is correctly
installed.
2.3
Installation of Tape
The correct operation of the 802 is dependent on the correct installation of the lead acetate tape.
Follow this procedure exactly to ensure correct installation. Figure 2-4 shows a photo of the tape
enclosure with all components numbered these numbers will be referenced in the procedure.
The system MUST be in BYPASS mode prior to starting this procedure for replacement of tapes.
1. Remove the screw-on disks (not shown) from the supply reel (1) and take-up reel (2).
2. Take a new reel of tape and remove the adhesive tape securing the end. Place the tape reel
on the supply reel. Make sure the tape is installed so that the loose end hangs down on the
left side of the reel, not the right.
3. Thread the tape over the lower sample chamber guide pin (3). Push the compression head (4)
back and slide the tape behind the compression head. Ensure that the tape is flat against the
groove in the sample chamber (8) and has no creases.
Revision 1.2
2-5
21/08/2007
Revision 1.2
2-6
21/08/2007
7. Replace the screw-on disks onto the supply and take-up reel. The larger disk goes on the
take-up reel.
8. Take up any slack in the tape by rotating the tape reel on the supply reel clockwise until the
tape is tight.
Other components of interest in Figure 2-4 are the sensor block cover (8), the vent bulkhead (9),
and the low tape sensor window (10). Once these steps have been followed, the tape should be
correctly installed on the tape transport system. Figure 2-6 shows the correct installation of tape
on the 802. The image on the left shows the completed tape installation without the anti-coning
disks and the one on the right shows the tape installation with the disks.
2.4
To ensure that repeatable data is obtained, it is essential to have a good seal between the
compression head, the tape, and the sample chamber. To test the seal, Galvanic recommends a
Dwyer Minhelic II gauge, available from Galvanic Applied Sciences. The gauge measures a range
of pressure from 0 to 5 inches of water column.
The following procedure is used to leak-check the 802 compression head and to obtain a proper
seal. Make sure the flow indicated on the chassis rotameter is approximately 2.0 before following
this procedure. The system MUST be in BYPASS mode prior to carrying out this procedure.
1. Remove the vent line from the elbow at the base of the chassis and attach the Minihelic to the
elbow, as shown in Figure 2-7.
2. Check the pressure indicated on the gauge. If the pressure reads 5 or above, as shown in
Figure 2-7, the tape is well sealed against the sample chamber by the compression head.
Press TAPE ADVANCE on the keypad and ensure that the pressure does not drop below 4
even when the tape is moving. It should return to 5 by the time the tape stops moving.
3. If the pressure is NOT 5 or greater, there is a problem with the seal somewhere in the
chassis. The first step will be to check and see if the problem is in the sample chamber itself.
Revision 1.2
2-7
21/08/2007
Figure 2-7: Minihelic indicating a good seal (L); connected to vent elbow (R)
4. Press BYPASS on the keypad to put the 802 into Bypass mode. Remove the compression
head. Pull the tape away from the sample chamber and put a thumb over the hole in the
sample chamber. If the pressure indicated on the Minihelic goes to 5 or greater, this will
indicate that the leak problem is caused by the compression head itself. If the pressure
indicated on the Minihelic with a thumb over the hole is close to zero, this is indicative of an
improperly installed gasket and/or window in the sample chamber assembly. Go to step 6. If
the pressure indicated is greater than 1, this is indicative of a leak in the humidifier. Go to step
7.
5. If the problem is with the compression head, there are a couple of possible resolutions. First,
take the compression head and rub the surface that actually contacts the tape on the paper
wrapper from the lead acetate tape to try and smooth it down. Reinstall the compression head
and check the pressure again. If the pressure is still not 5, remove the compression head
again, remove the setscrew from one side of the compression head and move it to the other
side to allow the compression head to be installed the other way in. Reinstall the compression
head, press TAPE ADVANCE on the keypad, and check the pressure again. If it is still not 5,
please call the service department at Galvanic Applied Sciences.
6. To check the sample chamber assembly for an improperly installed rear window and/or
gasket, remove the sensor block cover (number 9 in Figure 2-4) and unscrew the two screws
retaining the sample chamber in the chassis. Disconnect the sample inlet line (top) and the
sample vent line (bottom). Figure 2-8 shows an exploded view of the sample chamber
assembly for reference in taking the chamber assembly apart.
Revision 1.2
2-8
21/08/2007
Check that the two screws holding the sensor block clips (6) in position are as tight as
possible and then put a thumb back over the hole to check the pressure again. If the pressure
is still less than 5, remove the sensor block (4) entirely by disconnecting the two (or three)
electronics connectors on the sensor block, loosening the two screws holding the sensor block
retaining clips, and turning the clips outwards so that the sensor block is no longer retained.
Remove the sensor block. Ensure that the gasket (5) underneath the plastic window (3) is
correctly in position and that the plastic window is flat against the surface of the gasket, and
properly located within the indentation in the sample chamber. Check that the aperture strip
(2), if present, is stuck to the inside of the sample chamber with silicone grease. Clean out any
lint or dust noted in the square hole in the bottom of the sample chamber, as this may cause
the sensor block to have a high sensor current. Place the sensor block back on the sample
chamber, making sure not to knock the window out of position. Tighten it back down as far as
possible using the retaining clips. Reconnect the sample lines and check the pressure again
by placing a finger over the hole in the sample chamber it should be 5. If it is close to zero,
follow the procedure in this step again to check if the window was knocked out of position
when the sensor block was reinstalled. If it is greater than 1, go to step 7 to leak check the
humidifier. Reinstall the sample chamber back in the chassis and reinstall the compression
head, making sure that the tape is flat against the sample chamber. Check the pressure again.
If it is not 5, follow step 5 to check the problems with the compression head.
7. To test the humidifier for leaks, remove the Minihelic from the vent elbow and slide the
Minihelics rubber tubing over the end of the outlet barb on the humidifier, as shown in Figure
2-9.
Revision 1.2
2-9
21/08/2007
greater than 5, there may be a problem with the tubing or one of the fittings. Please contact
Galvanic Applied Sciences Inc.s service department for assistance. If the pressure is greater
than 5, reattach the top to the humidifier and reconnect the Tygon tubing inside the chassis to
the outlet barb on the humidifier.
2.5
The procedures for running a calibration, reference, and zero run are slightly different, so they will
be covered separately for greater concision. A zero run should be performed before a calibration
run if both are to be done at the same time. NOTE: These runs will NOT initiate if an analysis
alarm (Alarm 1 or PAA) is currently triggered on the Normal run. If the 802 is in Alarm 1 or
PAA, the calibration, reference, or zero run will be entered into the run queue and will only
run once the alarm condition has cleared.
Revision 1.2
2-10
21/08/2007
the certificate value of the calibration gas concentration, and then press ENTER. Press
EXIT to return to the default screen.
4. If the sample system has a manual 3-way selection valve to choose between calibration
and sample streams, turn the valve to select the calibration stream. Make sure the chassis
rotameter still reads a flow of 2.0. For total sulphur 802s, make sure the sample flow
meter still reads 0.5 and the hydrogen still reads 1.5.
5. To initiate the calibration run, press the F4 key to access the MNT (maintenance) menu.
Using the F4 (PGDN) key, scroll down through the maintenance menu until the line that
says Cal OFF is seen. Use the F2 and F3 keys if necessary to move the black highlight
bar to highlight this line. Press the EDIT key. Pressing the ENTER key will toggle the 802
to put a calibration run into the run queue.
6. Once the calibration run is complete, run a reference to ensure that the calibration is valid.
7. Once satisfied with the calibration, return the 3 way valve to the sample position.
Revision 1.2
2-11
21/08/2007
procedures in sections 2.5.2 and 2.5.3. Ensure that the 802 is in BYPASS mode before
following these procedures.
2.5.4.1
Quick Calibration
2.5.4.2
Quick Zero
Revision 1.2
2-12
21/08/2007
6. Once satisfied with the zero offset, return the 3-way valve to the sample position.
2.6
There are a couple of procedures that must be followed for maintenance and testing of totalsulphur furnace equipped 802s. These procedures are related to replacing the quartz reaction
tube and checking the furnace for hydrogen leaks after replacing the reaction tube.
IMPORTANT: It is essential that the power be turned off to the 802 before
carrying out either of these procedures. The furnace enclosure must be
allowed to cool for at least 45 minutes prior to initiating these procedures.
Do NOT remove the end caps of the furnace enclosure until the furnace is
cool attempting to remove the end caps while the furnace is still hot could
result in an explosion if even a small amount of hydrogen has leaked into
the furnace!
Figure 2-10 shows a numbered photo of the exterior and interior of the total sulphur furnace, for
ease of identifying all the parts discussed in these procedures.
Figure 2-10: Total Sulphur Furnace Enclosure - Exterior (L) and Interior (R)
The total sulphur furnace is housed inside an explosion proof enclosure (1). There are two screw
on end caps (2,3) that screw onto large threads (10) at each end of the enclosure. A conduit (6)
carries the power wiring (11,12) and the temperature sensor wiring (not visible) into the enclosure.
The sample flows through flame arrestors (4,5) on the way into and out of the furnace. The
reaction takes place in a quartz reaction tube (8) inside the actual total sulphur furnace, which is
housed in a smaller aluminium enclosure (7). The sample lines are attached to the quartz tube by
means of ultra-torr elbows (9) that have finger-tightened nuts to create an adequate seal.
Revision 1.2
2-13
21/08/2007
In some situations it may be necessary to replace the quartz reaction tube. Depending on the
composition of the sample gas, the tube may coke up over time and cause flow impediment.
Alternatively, if the tube cokes up significantly, it may reduce the catalytic behaviour of the
reaction tube that helps the total sulphur reactions occur. The following procedure is to be
followed to replace a quartz reaction tube. Refer back to Figure 2-10 for assistance in
identifying parts in the total sulphur enclosure assembly.
1. As mentioned previously, turn off the power to the 802. Allow the furnace to cool down for
at least 45 minutes before removing the end caps.
2. Turn off the hydrogen and sample flow at the rotameters on the sample conditioning
system.
3. Unscrew the nuts all the way on the reaction tube sides of the two ultra-torr elbows inside
the furnace assembly.
4. Pull the ultra-torr elbows away from the reaction tube. Slide the old tube out of the furnace
assembly. Slide the nuts, ferrules and O-rings off the old tube.
5. Carefully slide the new tube into the furnace. Slide the old nuts and ferrules onto the two
ends of the tube, as well as two new kal-rez high temperature rubber o-rings. The correct
order is shown in the exploded view in Figure 2-11.
Revision 1.2
2-14
21/08/2007
Turn the sample and hydrogen flow back on. Ensure hydrogen is flowing at 1.5 and sample is
flowing at 0.5 on the rotameters.
Place a cap on the top side of the flame arrestor on the right hand side of the total sulphur
furnace, as shown in Figure 2-13. This will block flow through the total sulphur furnace.
Watch the flow on the hydrogen and sample rotameters on the sample conditioning system. If
there is a good seal in the furnace, the flow on both rotameters should drop to zero within 1
minute.
If the flow does not drop to zero, use liquid leak detector (e.g. Snoop) to check all external fittings
for leaks. If there are none, remove the end caps from the furnace enclosure and ensure that all
the fittings on the four ultra-torr elbows inside the enclosure, especially the two on the reaction
tube itself, are as tight as possible. Again, be careful not to break the tube while tightening the
fittings.
Once the leak has been fixed, replace the end caps, remove the cap, and reattach the sample
line to the top of the flame arrestor. Ensure that this fitting is tight.
Revision 1.2
2-15
21/08/2007
3.1
Electronics Enclosure
The Electronics enclosure is the brain of the 802. It contains the main processor, the system
memory, the power supply, the LCD display, the analog outputs, the relays, the solenoids, and the
communications ports USB, RS-232, RS-485, and high speed LAN. In Class 1 Division 1 (XP)
units, the electronics enclosure also contains the intrinsically safe barrier through which pass all
electrical connections between the electronics enclosure and the chassis. The interior of a Division
2 electronics enclosure is seen in Figure 3-1.
Revision 1.2
3-16
21/08/2007
3.1.1.1
LCD Display
The LCD display on the 802 is much larger than the displays found on previous Galvanic
lead acetate tape analyzers. The default display can be seen clearly in Figure 3-2. In the
default display mode, there are three separate zones on the LCD display. At the top of
the screen are 4 status indicators, marked mV, RUN, STATE, and TAPE. The middle of
the screen shows the calculated concentration value. The bottom of the screen shows
four menus that can be accessed by using the keypad.
The mV indicator indicates the current output of the sensor block. This number is adjusted
to zero at the beginning of an analysis, and becomes larger as the tape gets darker on
exposure to H2S. The larger this number is, the darker the stain on the tape is.
The RUN indicator indicates which analysis mode the 802 is currently in. The four
possible values of this indicator are shown in Table 3-1.
Code
NR0 to NR9
CR0 to CR9
RR0 to RR9
ZR0 to ZR9
The four run types are explained in detail in section 4.4. The numeral in the code indicates
which analysis cycle in the sequence the 802 is currently carrying out.
Revision 1.2
3-17
21/08/2007
The STATE code indicates what state the 802 is currently in. There are seven possible
codes that may be seen in this indicator. They are explained in Table 3-2.
Code
INIT
IDLE
TA
CSENS
SD
SI
PRG
The INIT, IDLE, and CSENS codes are explained in more detail in section 3.2.1.2, which
describes the operation of the sensor block. The SD and SI codes are explained in more
detail in sections 4.2 and 4.3, which explain the standard analysis cycle and the timed
analysis cycle.
The TAPE indicator displays the length of tape currently remaining on the reel. This
allows the user to quickly determine the amount of tape left without having to open the
tape enclosure. This number MUST be reset every time the tape is changed to ensure
that this value is correct. The tape length can be reset through the keypad (Section 3.3)
or through the GUI (Section 5.3.4.1.1).
The centre of the LCD display is occupied by a large concentration display. This displays
the last calculated concentration value for the current stream. That is to say, if the
analyzer switches streams, the screen will switch to show the last calculated value for
that stream. If that stream has not been run since the 802 was powered up, the display
will read 0.00. The concentration value will update live if the 802 is in PAA Alarm (section
4.4.1.2) or if the user has configured the LCD to display live concentration values, either
through the keypad or through the GUI. The units value, by default set to PPM, can be
configured to display any units of the users choosing through the GUI (section 5.3.5.1). If
the 802 is configured to display the live concentration reading, the concentration units will
have an L in front of them. For example, PPM would become LPPM.
The bottom portion of the LCD is taken up by four menus. The codes F1, F2, F3, and F4
indicate the keys on the keypad used to access these menus. The purpose of each menu
is outlined in Table 3-3, and described in detail in section 3.3.
Revision 1.2
3-18
21/08/2007
Code
OPER
ALRM
DISP
MNT
3.1.1.2
There are a total of nine coloured LEDs on the front panel of the 802. These LEDs
indicate specific status and communications related parameters. The functions of the 9
LEDs are explained in Table 3-4.
Table 3-4: Front Panel LED Functions
Revision 1.2
LED Label
Colour
Function
LAN
ACCESS
Green
(blinks)
LAN DATA
Green (solid)
Tx USB
Green
(blinks)
Rx USB
Red (blinks)
ALARM
Red (solid)
BYPASS
Yellow
(blinks)
This LED blinks green if the unit is being communicated to via the
high speed LAN port.
This LED indicates that there is communication between the
various components of the LAN electronics.
This LED blinks green if the unit is transmitting data via the USB
connection.
This LED blinks red if the unit is receiving data via the USB
connection.
When lit, this LED indicates that at least one alarm has been
triggered.
This LED blinks yellow if the unit is in bypass mode (section 4.5)
FAIL
Red (solid)
Red (blinks)
KEYPD
Green
SERIAL
COM
Green
3-19
21/08/2007
Revision 1.2
3-20
21/08/2007
sensor block and the pulse counter, and signal wiring to carry communications between the
sensor block and the electronics assembly. The intrinsically safe barrier is mounted behind the
electronics assembly, so for space reasons the power supply for Class 1 Division 1 802s must
be mounted externally.
3.2
Chassis
The Chassis is the component of the 802 that contains all of the mechanical parts of the 802. It
contains the tape-advance motor, the pulse counter used to accurately advance the tape, the lead
acetate tape spool, and all components related to the actual flow of sample gas and the
measurement of the concentration of H2S in the sample gas stream. A photo of the chassis of a
Class 1 Division 2 802 with all the covers removed is shown in Figure 3-4.
Revision 1.2
3-21
21/08/2007
Qty.
Description
Part Number
Sample Chamber
AF0509
2*
Varies
Rear Window
MC0406
Sensor Block
SA2466
MC0592
AF0917
The Rate/Read Aperture strip and the Rear Window are both made of a clear plastic material
to allow the light from the sensor block to pass through to the tape to correctly measure the
rate of colour change on the tape. The sensor block is held onto the sample chamber by
means of clips. By tightening the screws on the clips, the rear window is pressed into the rear
window rubber gasket and sealed to the sample chamber by pressure from the sensor block
above it. The rate/read aperture strip is sealed to the inside of the sample chamber using
silicone grease. The sample flow through the sample chamber assembly is shown in Figure
3-6, on the following page.
The clean, humidified, and flow-regulated sample enters the sample chamber from the sample
inlet. It then passes over the Rate/Read Aperture strip, behind the rear window. Some of the
gas passes through the aperture strip and onto the lead acetate tape, causing a measurable
colour change. The tape is held firmly against the sample chamber by the compression head.
The amount of gas that passes onto the tape is dependent on the size of the aperture strip
(section 3.2.1.1). The gas then passes out to the vent.
Revision 1.2
3-22
21/08/2007
3.2.1.1
The Read/Rate aperture strip is used to control the quantity of gas that is actually passes
onto the lead acetate tape. If the sample gas H2S concentration is high, it is necessary to
reduce the quantity of gas that passes onto the tape to react with the lead acetate,
otherwise the rate of colour change will be too fast for the 802 to correctly analyze, and
poor results will be obtained. Likewise, if the sample gas H2S concentration is low, it is
necessary to allow more sample gas to pass onto the tape to allow the rate of colour
change to be high enough to produce decent results. Thus, for high concentrations, the
size of the aperture in the read/rate aperture strip is small, and for low concentrations it is
large. Certain ranges do not even require an aperture strip. There are several possible
read/rate aperture strip sizes. The most common aperture strip sizes are listed in Table
3-6 along with the maximum ranges they are capable of measuring and their correct part
numbers. For other ranges, please contact Galvanic Applied Sciences.
Table 3-6: Read/Rate Aperture Strip Sizes and Associated Ranges
Maximum Range (ppm)
0-5
0-10
0-20
0-50
0-200
Part Number
N/A
MC0412
MC0410
MC0409
MC0570
Note that each read/rate aperture strip size is capable of measuring any range between
the maximum range of the next largest aperture strip size and the maximum range of the
given aperture strip size. For example, a 1/8 aperture strip would be capable of
measuring 0-5 and 0-10ppm.
Revision 1.2
3-23
21/08/2007
3.2.1.2
Sensor Block
The sensor block consists of two electronics boards, a light emitting diode, and a
photodiode detector. The 802s sensor block is shown in Figure 3-7.
Figure 3-7: 802 Sensor Block (From Left: Top View, Side View, Installed View)
The primary job of the sensor block is to measure the intensity of light reflected off of the
lead acetate tape. To do so, the light emitting diode emits a beam of red light that passes
through the rear window and the rate/read aperture strip and reflects off of the lead
acetate tape. The reflected light then passes back through the aperture strip and rear
window to reach the photodiode detector, where it is measured. As the stain on the tape
gets darker and darker while exposed to H2S, the amount of light reflected back to the
photodiode detector is reduced. The rate of change in the amount of light reflected back to
the photodiode detector is directly proportional to the concentration of H2S in the sample
gas. The sensor block communicates with the electronics assembly via an RS-485
connection. If the connection between the sensor block and the electronics assembly is
broken for any reason, an alarm will be triggered.
3.2.1.2.1
The measured reading at the photodiode detector is referenced to the sensor output
obtained on white tape at the beginning of every analysis. At the end of every cycle,
the tape is advanced a certain distance to ensure that there is a new section of white
tape to be analyzed. The sensor block then adjusts the current to the light emitting
diode such that the photodiode detector has an output of 1000 millivolts. The current
required to produce a reading of 1000mV is called the Sensor Current. The
differential sensor output is obtained by subtracting the reference output of 1000mV
from the sensor output obtained on tape that has been exposed to H2S. The
differential sensor output is calculated continuously during an analysis.
As the sensor ages, the amount sensor current will increase. In addition, if the sensor
chamber gets dirty and blocked up with lint or fuzz from the tape, the sensor current
will also increase. To clean out this lint, follow the procedure found in Step 6 in section
2.4. As such, the 802 can be set up to trigger an alarm should the sensor current go
above a user-selected value. The set point for this alarm can be configured through
the GUI (Section 5.3.4.2)
Revision 1.2
3-24
21/08/2007
3.2.1.2.2
The sensor block has three other secondary functions. First, it contains a temperature
sensor that measures the ambient temperature near the sensor block. This
temperature sensor can be used to trigger alarms should the ambient temperature
rise above or drop below set amounts, values that can be set in the GUI (Section
5.3.4.2).
Secondly, the sensor block plays a key role in the tape advance algorithm. While the
tape advance motor is running, the sensor block receives pulses from the pulse
counter (section 3.2.2.3). When the sensor block receives the number of pulses set in
the GUI (section 5.3.4.1.1), it sends a message to the 802s main processor. The
processor then sends a message to the motor to stop advancing the tape.
Finally, the sensor block also receives signals from the low tape sensor (Section
3.2.2.4). If the sensor block receives an alarm signal from the low tape sensor, it
sends this signal to the main processor, which then triggers a low tape alarm.
These other functions are not permitted to interfere with the sensor blocks primary
purpose of measuring the light reflected off the lead acetate tape. As such, they are
only updated during the time that the tape is advancing. Running these other functions
during an analysis cycle might introduce noise into the analog-to-digital converter
(ADC), which could negatively affect the concentration results obtained by the 802, so
they are turned off during an analysis cycle.
Revision 1.2
3-25
21/08/2007
The tape transport system consists of the supply and take-up reels, the drive motor, the
compression head, the pulse counter, and the optional low-tape sensor.
3.2.2.1
The supply and take-up reels are the key components in the tape transport system. The
supply reel, which has a small screw-on retaining disk, is found on the bottom of the
chassis, and can hold reels of tape up to 350 feet in length. The tape is threaded over the
lower tape guide, past the sample chamber assembly and compression head, over the
two upper tape guides, and finally over the pulse counter pulley before reaching the takeup reel. The take-up reel, which has a large screw-on retaining disk, is driven by the
motor, and is responsible for advancing the tape at the end of every cycle. For more
information on the correct installation of tape into the tape transport system, please see
section 2.3.
3.2.2.2
Motor
The drive motor is a small 5 volt electric motor that is controlled by the main processor of
the 802. It is either installed inside the chassis, as in the Division 2 802, or in an explosion
proof enclosure on the outside of the chassis, as in the Division 1 802. The motor turns
the take-up reel to advance the tape at the beginning of every analysis cycle to ensure
that the next analysis has a section of pure white tape to react with the hydrogen sulphide.
The length of time that the motor runs for is controlled by the 802s processor in
conjunction with the pulse counter.
Revision 1.2
3-26
21/08/2007
3.2.2.3
Pulse Counter
The pulse counter is the component of the tape transport system responsible for ensuring
even spacing of stains on the lead acetate tape. The pulse counter is shown in Figure 3-9.
3.2.2.4
The optical low tape sensor is an optional low tape sensor that is mounted just above the
supply reel. It shines a beam of light at the tape on the supply reel and looks for light
reflected off of the tape on the supply reel. If the amount of tape remaining on the supply
reel is below the level of the optical low tape sensor, there will not be any reflection read
by the low tape sensor. This will result in the low tape sensor triggering a low tape alarm.
Typically the low tape alarm triggers when there is approximately 15 feet of tape
Revision 1.2
3-27
21/08/2007
remaining on the supply reel. The low tape alarm can be disabled or enabled through the
GUI (Section 5.3.4.2).
3.2.2.5
Compression Head
The compression head is the component in the tape transport system that seals the lead
acetate tape up against the sample chamber so that the lead acetate can react with the
hydrogen sulphide in the gas stream. The compression head is shown in Figure 3-10.
3.2.3.1
Critical Orifice
A critical orifice is a Swagelok fitting that has a very small hole passing through the
fitting. Because the hole is so small, the maximum flow rate through the fitting is limited,
regardless of the inlet pressure. Thus, using a critical orifice ensures that the sample flow
rate through the analyzer remains consistent, even if there are fluctuations in the pressure
of the sample gas.
It is essential to have a consistent flow rate because the concentration reading is
dependent on sample flow rate. The higher the flow rate, the more hydrogen sulphide
Revision 1.2
3-28
21/08/2007
comes into contact with the lead acetate tape in a given period time, and thus the faster
reaction and higher concentration reading.
A total sulphur 802 does not contain a critical orifice. The sample flow through a total
sulphur 802 is controlled by two rotameters on the external sample system.
3.2.3.2
Sample Humidifier
The reaction between lead acetate and hydrogen sulphide is improved if the hydrogen
sulphide-containing sample gas is humidified prior to coming into contact with the lead
acetate. The standard Nafion humidifier for the 802 is shown in Figure 3-11.
3.2.3.3
Rotameter
The rotameter is a device that indicates the flow rate through the 802. It is mounted on the
upper access door on the chassis. The standard rotameter on the 802 chassis is shown in
Figure 3-12, on the following page.
It is important to note that the chassis rotameter does not control the flow rate of the
sample gas it only measures the flow rate of the gas. The flow rate through the 802 is
controlled either by the critical orifice or by the two controllable rotameters (one for
sample, one for hydrogen) on a 802 total sulphur units external sample conditioning
system. The flow rate indicated on the chassis rotameter is approximately 2.0, but this will
vary from unit to unit due to variances in the size of the hole in the critical orifice.
Revision 1.2
3-29
21/08/2007
3.2.3.4
The power vent, also known as the eductor, is a device that produces a positive pressure
at the exhaust while allowing the analyzer vent to remain at ambient pressure. An eductor
is shown in Figure 3-13.
Revision 1.2
3-30
21/08/2007
the tape and the sample chamber, resulting in poorly shaped stains that give poor results.
It can also cause more air to be introduced into the sample chamber, diluting the sample
gas and producing lower-than-expected results.
Secondly, if the 802 is located in a cold climate, the humidified sample gas passing out of
the vent can freeze, blocking off the flow through the sample system. By using an eductor,
the residence time of the sample gas within the 802 is reduced, making it less likely that
the moist air will freeze inside the analyzer.
3.3
Keypad
The keypad is the component of the 802 that allows the user to make changes to the analyzer
without the need for attaching a computer with the Graphical User Interface software installed. The
keypad is intrinsically safe, meaning that it can be used even in a class 1 division 1 area without
first having to declassify the area. This means that changes can be made to the analyzer quickly
and easily on site. The 802s keypad is shown in Figure 3-14.
F1
Revision 1.2
Function
Depends on current screen displayed. From default screen, allows the
user access to the Operator (OPER) menu. In Operator, Maintenance
(MNT), and Alarm (ALRM) screens it is used as the page-up (PGUP)
function key to scroll through multiple menu screens. In Display (DISP)
screen, is used to toggle between the raw sensor output and the net
sensor output (section 3.2.1.2.1).
3-31
21/08/2007
F2
F3
F4
BYPASS
TAPE ADVANCE
EDIT
EXIT
ENTER
The functions, if any, of keys F1 through F4 are always displayed underneath the labels for these
keys on the 802s LCD screen. In certain screens, some of these keys may not have any function.
Revision 1.2
3-32
21/08/2007
contains editable data. If a parameter does not have a value in the second column, it cannot
be edited from the keypad. The third column contains values that cannot be edited from the
keypad. Four parameters can be displayed on each page. Pressing F1 PGUP or F4 PGDN
will show the previous or next four parameters in the menu. Table 3-8 explains the Operator
Menu, in order from top to bottom of the menu. Parameters that are listed in the table as
being editable can be edited by moving the black highlight bar up and down using the F2 and
F3 keys until the chosen parameter is highlighted. Once highlighted, press the EDIT key and a
screen like the one shown in Figure 3-16 will appear.
Parameter Displayed
Editable?
Alarm
Alarm 1 Setpoint
Yes
PAA
Predictive Alarm
Analysis (PAA)
setpoint
Yes
PPM_AO
Analog Output 1
Range
No
TAPE_AO
Analog Output 2
Range
No
Date
Current Date
No
Time
Current Time
No
Current
Sensor Current
No
Temp.
Sensor Temperature
No
Revision 1.2
Explanation
This is the setpoint value for alarm 1 on the Normal run. If the
calculated concentration value at the end of any cycle exceeds
this value, Alarm 1 will be triggered. See section 4.4.1.1 for
more information on Alarm 1. This value can also be set in the
GUI see section 5.3.5.2.3.
This is the setpoint for PAA on the normal run. If the calculated
value at any time exceeds this value, the PAA will be triggered.
See section 4.4.1.2 for more information on the PAA. This value
can also be set in the GUI see section 5.3.5.2.3.
This parameter displays the name and range associated with
analog output 1. By default, this output is set to output the
concentration value from the normal run. The name and range
of this value can be set in the GUI (Section 5.3.5.2.2)
This parameter displays the name and range associated with
analog output 2. By default, this output is set to output the
length of tape remaining. The name and range of this parameter
can be set in the GUI (Section 5.3.5.2.2)
This parameter displays the 802s current date stamp. This can
be set through the GUI using the Synchronize Time button
see section 5.2.1.2
This parameter displays the 802s current time stamp. This can
be set through the GUI using the Synchronize Time button
see section 5.2.1.2
This parameter displays the 802s sensor current value (section
3.2.1.2.1).
This parameter displays the current ambient temperature, in
degrees Celsius, experienced by the temperature probe on the
sensor block.
3-33
21/08/2007
Tape
Firmware
Sensor
Tape Length
Remaining
Yes
Firmware Revision
Sensor block firmware
revision
No
No
MODBUS
Add.
MODBUS Address
No
FURN
Digital Input 1
No
PRESS
Digital Input 2
No
CAL
Digital Input 3
No
DI4
Digital Input 4
No
Parameter
Displayed
Alarm
Alarm 1
Setpoint
PAA
Predictive
Alarm Analysis
(PAA) setpoint
PPM_AO
Analog Output
1 Range
TAPE_AO
Analog Output
2 Range
Gain
Calibration
Gain
Revision 1.2
Explanation
This is the setpoint value for alarm 1 on the Normal run. If the calculated
concentration value at the end of any cycle exceeds this value, Alarm 1 will be
triggered. See section 4.4.1.1 for more information on Alarm 1. This value can also be
set in the GUI see section 5.3.5.2.3.
This is the setpoint for PAA on the normal run. If the calculated value at any time
exceeds this value, the PAA will be triggered. See section 4.4.1.2 for more
information on the PAA. This value can also be set in the GUI see section 5.3.5.2.3.
This parameter displays the name and range associated with analog output 1. By
default, this output is set to output the concentration value from the normal run. The
range can be set using the keypad, while the name and range of this parameter can
also be set in the GUI (Section 5.3.5.2.2)
This parameter displays the name and range associated with analog output 2. By
default, this output is set to output the length of tape remaining. The range can be set
using the keypad, while the name and range of this parameter can also be set in the
GUI (Section 5.3.5.2.2)
This is the calibration gain, which is the number that the rate of change is multiplied
by to obtain a concentration value. When this number is changed, the display and the
analog output immediately update to reflect the new calculated concentration value.
For more information on the gain and the calculation of the concentration value, see
section 4.4.2. This value can also be set through the GUI (section 5.3.5.3).
3-34
21/08/2007
GCalga
Calibration Gas
Concentration
Offset
Zero Offset
OCalga
Zero Offset
Gas
Concentration
FURN,
PRESS,
CAL, DI4
DI_POL
Polarity of
Digital Inputs
ALM1,
ALM2,
MAINT,
Relay4
FS_POS
Fail Safe
Position of
Relays
ALM1,
ALM2,
MAINT,
Relay4
LCH_EN
Cal
Initiate
Calibration Run
Ref
Initiate
Reference Run
Revision 1.2
This is the calibration gas concentration, from which the calibration gain is calculated
during a calibration run (section 4.4.2). This value can also be set through the GUI
(section 5.3.5.3).
This is the concentration that the analyzer reads while analyzing gas that contains no
hydrogen sulphide. The zero offset is calculated during a zero run (section 4.4.3).
When the gain is updated, the value of this parameter is automatically changed to
reflect the new gain. This value can also be set through the GUI (section 5.3.5.5).
This is the zero gas concentration, from which the zero offset is calculated during a
zero run. This value is typically zero, and should not be set to any other number
unless the zero gas is known to contain a certain quantity of hydrogen sulphide. This
value can also be set through the GUI (section 5.3.5.5).
The next four parameters indicate the polarity of the digital inputs. If the value is set to
OFF, it means that the digital input will be considered triggered if the switch
connected to the digital input is off. This is used for the Total Sulphur furnace
temperature switch and the low pressure switches, both of which turn on when
conditions are normal and turn off when in an alarm condition. If the value is set to
ON, it means that the digital input will be considered triggered if the switch connected
to the digital input is on. This is used for run initiation switches which are off when the
switch is not depressed, and turn on when the switch is pressed. The polarity of
digital inputs, as well as their displayed names, can also be set in the GUI (Section
5.3.6).
These parameters indicate the fail safe position of the four relays. The fail-safe
position is the position that the relay will default to when a) the 802 is in bypass mode
and b) the relay has not been triggered by an alarm or any other event. If the value is
set to OFF (which all four relays are set to by default) the relay will be off (i.e. set to
the normally open position) when the 802 is in bypass mode and when the relay has
not been triggered by any event. If the value is set to ON, the relay will be on (i.e. set
to the normally closed position) when the 802 is in bypass mode and when the relay
has not been triggered by any event. The Fail Safe Position of the relays can also be
set through the GUI (Section 5.3.7).
These parameters indicate whether or not the four relays will latch when triggered by
an event. If the value is set to OFF, the relay will return to fail safe position as soon as
the event that triggered the relay has cleared. If the value is set to ON, the relay will
remain in the non fail safe position until the latch has been cleared by the user, either
through the GUI or through the keypad. The latching behaviour of the relays can also
be set through the GUI (section 5.3.7).
This parameter allows the user to initiate a calibration cycle. By default this is set to
OFF. A calibration run can be initiated by highlighting this parameter, pressing the
EDIT key, then pressing ENTER to toggle the setting to ON. This will put a calibration
run into the run queue. The setting will return to OFF once the calibration run has
completed. Note that a calibration run will not initiate if Alarm 1 or the PAA are
currently triggered it will wait in the run queue until the alarm condition has
completely cleared. Ensure that the calibration gas concentration has been correctly
entered and that the calibration gas is hooked up to the calibration port prior to
initiating a calibration run. The Calibration run can also be initiated, and the displayed
name for this run can be changed from the GUI (5.3.5.3).Note than a Calibration run
can initiated from the keypad ONLY if the Conditional checkbox in the Calibration
stream page in the GUI has a checkmark in it.
This parameter allows the user to initiate a reference run to check the 802s
calibration. By default this is set to OFF. A reference run can be initiated by
highlighting this parameter, pressing the EDIT key, then pressing ENTER to toggle
the setting to ON. This will put a reference run into the run queue. The setting will
return to OFF once the reference run has completed. Note that a reference run will
not initiate if Alarm 1 or the PAA are currently triggered it will wait in the run queue
until the alarm condition has completely cleared. Ensure that the calibration gas is
hooked up to the calibration port prior to initiating a reference run. The reference run
can also be initiated, and the displayed name for this run can be changed from the
GUI (5.3.5.4). Note than a Reference run can initiated from the keypad ONLY if the
Conditional checkbox in the Reference stream page in the GUI has a checkmark in
it.
3-35
21/08/2007
Zero
Initiate Zero
Run
Ack_all
Acknowledge
all Alarms
This parameter allows the user to initiate a zero run. By default this is set to OFF. A
zero run can be initiated by highlighting this parameter, pressing the EDIT key, then
pressing ENTER to toggle the setting to ON. This will put a zero run into the run
queue. The setting will return to OFF once the zero run has completed. Note that a
zero run will not initiate if Alarm 1 or the PAA are currently triggered it will wait in the
run queue until the alarm condition has completely cleared. Ensure that the zero gas
concentration has been correctly entered and that the zero gas is hooked up to the
correct port prior to initiating a zero run. The zero run can also be initiated, and the
displayed name for this run can be changed from the GUI (5.3.5.5). Note than a Zero
run can initiated from the keypad ONLY if the Conditional checkbox in the Zero
stream page in the GUI has a checkmark in it.
This parameter allows the user to acknowledge all current alarms, and by doing so
clear all latches at the same time. Clearing all latches can be accomplished by
highlighting this parameter, pressing the EDIT key, and then pressing ENTER to
toggle the setting to ON. This setting will immediately return to OFF. Clearing all
latches at the same time can be accomplished through the GUI by pressing the Reset
Latches button (section 5.2.1.2)
3.3.4 Display
The Display Menu allows the user to choose how they wish to have their sensor output and
concentration values displayed. To access this menu, press F3 from the default display. This
will bring up a screen as shown in Figure 3-18.
Revision 1.2
3-36
21/08/2007
3.4
The 802 is available with the option of being capable of measuring the total sulphur content of a
sample gas stream. It is not possible, however, to measure total sulphur and hydrogen sulphide
content on the same 802. The sample gas is mixed with hydrogen in a quartz reaction tube at a
high temperature of 900C. At this temperature, the hydrogen reacts with all sulphur components
to form hydrogen sulphide, and most hydrocarbons to form methane. The hydrocarbon reactions
are known as cracking reactions. The conversion of total sulphur to hydrogen sulphide is covered
in ASTM Method D4468-85: Total Sulphur in Gaseous Fuels by Hydrogenolysis and Rateometric
Colorimetry. A few common reactions are shown in Table 3-10.
Table 3-10: Total Sulphur Reactions
Carbonyl Sulfide
COS + 4H2
900C
Ethyl Sulfide
(C2H5)2S + 4H2
H2S + 4CH4
900C
Revision 1.2
3-37
21/08/2007
Methyl Sulfide
(CH3)2S + 2H2
H2S + 2CH4
900C
Example of cracking:
Butane
C4H10 + 3H2
4CH4
900C
Because the sulphur compounds are all converted into hydrogen sulphide, a measurement of the
hydrogen sulphide content formed by these hydrogenation reactions will be directly equivalent to
the total sulphur content in the sample gas stream.
The total sulphur reaction takes place in an external total sulphur furnace that is mounted below
the electronics enclosure. The total sulphur furnace is shown in Figure 3-19.
3.5
The sample conditioning system is an optional accessory that filters the sample gas and lowers it
to a pressure suitable for use in the 802. A typical sample conditioning system is seen in Figure
3-20.
Revision 1.2
3-38
21/08/2007
Revision 1.2
3-39
21/08/2007
Section 4 Operation
The 802 is capable of measuring hydrogen sulphide using two measurement methods. The first
method is the same time-based analysis method used in previous models of Galvanic lead acetate
tape analyzers. The second method, the standard analysis, is a new patent-pending analysis method
that improves repeatability, linearity, and dynamic range. The 802 is has four total analysis modes
normal analysis, which is analysis of the customers sample gas; calibration, which produces a new
gain factor for the 802; zero reference, which determines the zero offset of the 802; and reference,
which checks the current calibration of the 802. Each analysis mode also has certain alarming
capabilities, regardless of which analysis method is chosen.
4.1
Event Sequence
There is a set sequence of events that occur during an analysis, regardless of the analysis method
used and the type of stream being analyzed. These events are outlined in Table 4-1.
Table 4-1: Event Sequence
Time (sec)
Event
State
Code
Explanation
0 to value in Purge
Time field in GUI
Purge
PRG
Prior to Analysis
Tape
Advance
TA
Prior to Analysis
Calibrate
Sensor
Sample
Delay
CSENS
Sample
Interval
SI
This event only occurs if the 802 has switched analysis run
types (e.g. from normal to calibration) directly prior to the
current analysis. The sample system is purged for a given
amount of time, set in the GUI (section 5.3.4.1) prior to
initiating analysis on the new run type.
The tape is advanced until a new section of white tape is
available for analysis. Sets the photodiode detector output
to 1000mV by adjusting the sensor current.
Sets the 1000mV standard photodiode detector output on
white tape equal to 0mV.
This event differs between the Standard Analysis and the
Time-based analysis. The two versions of the Sample Delay
will be covered in their respective sections standard
analysis (Section 4.2.1) and timed analysis (Section 4.3.1)
The concentration is calculated. At the end of the sample
interval the analog outputs and MODBUS registers are
updated, and any alarm conditions reported.
0 to value in Sample
Delay field in GUI
Value in Sample
Delay field to value
in Sample Interval
field in GUI
4.2
SD
Standard Analysis
The standard analysis method produces highly linear results with a large dynamic range. The
standard analysis has two parts the Sample Delay and the Sample Interval. The cycle time is
generally the sum of the sample delay time and the sample interval time, though there is one other
parameter, the Tape Saturation Cutoff (TSC) that may affect the length of the analysis.
Revision 1.2
4-40
21/08/2007
prevents this noise from negatively affecting results. The sample delay can be set in the GUI
(section 5.3.5.1.1) to any time period of the users choosing, but it is recommended that this
value be left at the default value for the best possible results.
4.3
Timed Analysis
The timed analysis method is the same analysis method that was found on previous Galvanic lead
acetate tape analyzer models the 801 and 902. This method is useful for customers who desire
static and well defined analysis times above all else. The algorithm used to calculate the
concentration in the timed analysis is different from the one used in the standard analysis.
Revision 1.2
4-41
21/08/2007
the live concentration value exceeds the PAA alarm set point, the PAA alarm is immediately
triggered. However, if the live concentration does not exceed the PAA alarm set point, the live
concentration value is not stored in memory and is instead discarded. The sample delay can
be set in the GUI (section 5.3.5.1.1) to any time period of the users choosing, though the
default value of 140 seconds is usually ideal for a timed analysis.
4.4
Analysis Modes
The 802 has four analysis modes, or run types that can be selected for operation at any time.
They are normal, calibration, reference, and zero. The length of the sample delay and sample
interval, as well as the magnitude of the tape saturation cut-off point, can be set independently for
each type of stream. These are set through the GUI see the Streams section 5.3.5.1.1. This will
allow either tape conservation or rapid updates (with the downside of increased tape consumption)
in the normal analysis and shorter analysis cycles for calibration, reference, and zero, reducing
both down-time for non-normal analysis cycles and reducing calibration gas consumption.. Each
stream type also has certain alarm types associated with it.
4.4.1.1
Alarm 1
Alarm 1 is an alarm that triggers when the calculated concentration at the end of a cycle
is higher than the Alarm 1 set point. The alarm 1 set point can be set either through the
OPER menu from the keypad (section 3.3.1) or through the GUI (section 5.3.5.2.3). The
802 compares the final calculated concentration at the end of the sample interval with the
alarm 1 set point value. If the value of the calculated concentration is higher than the
alarm set point value, the alarm will be triggered, the alarm LED on the front panel will turn
on, the event will be listed in the Event Log and on the Analyzer Alarm list on the 802s
LCD screen, and any outputs associated with this alarm will also be triggered. Once the
concentration of the normal stream drops back below the Alarm 1 set point, the alarm LED
will turn off, Alarm 1 will be cleared from the Analyzer Alarm List, and any unlatched
outputs associated with Alarm 1 will return to their non-triggered state. Any latched
outputs associated with this alarm will have to be manually cleared through the GUI
(section 5.3.7).
4.4.1.2
The Predictive Alarm Analysis (PAA) alarm mode is an alarm mode that triggers when the
live calculated concentration at any point during the analysis cycle (except for the sample
delay on the standard analysis method) exceeds the PAA alarm set point. When the 802
enters PAA mode, the concentration displayed on the 802s LCD screen immediately
begins to update with the live concentration value once per second. In addition, any
Revision 1.2
4-42
21/08/2007
analog output or MODBUS register that is currently outputting the stream 1 normal
concentration value will also output the live value of the concentration once per second
while the 802 is in PAA mode. The PAA mode will clear as soon as the concentration of
the gas stream drops back below the PAA alarm set point. The PAA alarm set point can
be set either through the OPER menu from the keypad (section 3.3.1) or through the GUI
(section 5.3.5.2.3). When the 802 is in PAA mode, this alarm will be listed in the Analyzer
Alarm list, as well as in the Event Log, and the red ALARM LED on the front panel of the
802 will also be lit. Any outputs associated with this alarm mode will also be triggered.
Once the concentration of the normal stream drops back below the PAA set point, the
alarm LED will turn off, PAA will be cleared from the Analyzer Alarm List, and any
unlatched outputs associated with PAA will return to their non-triggered state. Any latched
outputs associated with this alarm will have to be manually cleared through the GUI
(section 5.3.7).
4.4.2.1
The Deviation alarm is an alarm that triggers if the newly calculated calibration gain
differs from the calibration gain in the 802s memory by more than a user-selected
percentage value (set in the GUI). If the calibration gain changes by more than this
amount, the % Deviation alarm will be triggered. This alarm will trigger any output
associated with it, and it will be logged in the Event Log and in the Analyzer Alarm List on
the 802s LCD screen. The new calculated gain factor will not be stored in memory, and
the 802 will continue to operate using the previous gain factor. Possible causes of a %
deviation alarm being triggered include an empty calibration gas cylinder, a calibration
cylinder that is not present, or an incorrectly entered calibration gas concentration value. If
Revision 1.2
4-43
21/08/2007
4.4.3.1
Like the calibration % deviation alarm, the Zero % Deviation alarm is triggered if the
newly calculated zero offset differs from the zero offset stored in memory by more than a
user-set percentage value, set in the GUI. If the zero offset changes by more than this
amount, the % Deviation alarm will be triggered. This alarm will trigger any output
associated with it, and it will be logged in the Event Log and in the Analyzer Alarm List on
the 802s LCD screen. The new calculated zero offset will not be stored in memory, and
the 802 will continue to operate using the previous zero offset. Possible causes of a %
deviation alarm being triggered include an empty zero gas cylinder, a zero gas cylinder
that is not present, or use of a gas cylinder for zero that contains a measurable
concentration of hydrogen sulphide. If a Deviation Alarm is observed, a manual zero run
should be performed on-site to verify the performance of the 802.
Revision 1.2
4-44
21/08/2007
though the number of cycles to average and the total number of analysis cycles per reference
run are user configurable through the GUI.
Like the calibration and zero runs, the reference run can be initiated in several ways. It can be
initiated from the MNT menu using the keypad (Section 3.3.2), remotely via a button
connected to a digital input (section 5.3.6), or manually or on a timed schedule via the GUI
(section 5.3.5.4). The Condition checkbox in the Reference stream tab in the Stream window
(section 5.3.5.4) must be checked in order to initiate a calibration by any of these methods.
Note that regardless of how the analysis is triggered, it is essential that a gas cylinder
containing a known concentration of hydrogen sulphide be hooked up to the correct port,
usually the calibration port, on the 802 sample system.
Regardless of how the reference run was triggered, it will NOT run if the 802 Normal run is
currently in Alarm 1 or PAA alarm condition. It will be entered into a run queue and will only
run once the alarm condition(s) on the normal run has cleared.
4.4.4.1
The Reference % Deviation alarm is triggered if the current reference reading varies
from the previous reference reading by more than a user-set percentage value. The
reference deviation alarm is reported on the Analyzer Alarm List on the 802s LCD screen,
as well as in the Event Log. It is important to note, however, that if the concentration of
gas used for reference in the current reference run differs significantly (i.e. more than the
percentage deviation value) from the concentration of gas used for the previous reference
run, the reference deviation alarm will be triggered even though the calibration on the 802
may still be valid. For this reason, it is very important that the concentration of gas used
for reference purposes remain as consistent as possible.
4.5
Bypass Mode
The Bypass mode allows users to make changes to the 802 changes such as changing the
tape, cleaning the sample chamber, or anything else that could cause an erroneously high
concentration reading. When Bypass is enabled, either through the GUI (section 5.2.1.2) or the
keypad (section 3.3), all outputs switch to their fail-safe positions. The analog outputs will switch to
outputting their fail-safe current. The BYPASS LED on the front panel of the 802 will also blink
yellow as a visual indication that the 802 is in bypass mode. The reason for this is straightforward.
Replacing tape and working around the sample chamber can cause the 802 to read false high
concentration readings. If the 802 were not in bypass mode, these false high readings could cause
a plant shut-in when the alarm outputs related to these false high readings were triggered. This is
not a desirable situation.
Revision 1.2
4-45
21/08/2007
5.1
Serial/USB Port
Requirement
Microsoft Windows 2000 or XP are the only operating systems
supported by the GUI.
Minimum 512MB
A CD ROM drive is required to read the installation disk, and a minimum of
20 megabytes of space is required for installation on the PC hard drive.
More space will be required to save analyzer configurations and archive
data.
The 802 can be connected to the PC via either a 9-pin male serial
connector or via a standard USB cable (included with the analyzer). The
COM port used to connect to the 802 cannot be used for any other
purposes; otherwise connecting to the 802 will not be possible.
Revision 1.2
5-46
21/08/2007
5.1.3.1
Once the USB connection has been made and the analyzer is powered on, Windows will
immediately detect new hardware. Ensure that the configuration disk that came with the
analyzer is in the PCs CD-ROM drive. If the PC is running on Windows XP, a balloon
notifying the user that Windows has found new hardware (802 H2S Analyzer) will appear
in the taskbar, and then a window like the one shown in Figure 5-2 will appear.
Revision 1.2
5-47
21/08/2007
Revision 1.2
5-48
21/08/2007
5.1.3.2
The installation of the drivers for Windows 2000 is similar to the Windows XP Procedure.
When the USB connection has been made and the 802 powered on, a window like the
one shown in Figure 5-5 will appear.
Revision 1.2
5-49
21/08/2007
Revision 1.2
5-50
21/08/2007
5.1.3.3
Many modern computers have more than one COM port, so it is essential to know the
correct COM port to use to connect to the 802. The method of determining the correct
COM port is the same for both Windows 2000 and Windows XP. To determine the correct
port, go to the Control Panel of the PC. Double-click on System, and then open the
Hardware tab. Then click on the Device Manager Button. Scroll down in the device
manager until the node called Ports (COM & LPT) is seen. Expand this node by pressing
the (+) icon. Something like the display seen in Figure 5-9 should be seen.
Revision 1.2
5-51
21/08/2007
5.2
The GUI is a Windows-based, point and click interface. When the software is first opened, the
screen shown in Figure 5-10 is seen.
Revision 1.2
5-52
21/08/2007
Revision 1.2
5-53
21/08/2007
ONLY, which means that the 802 is in Read Only mode (section 5.2.3.1); and UPDATE, which
means that the 802 is in Update mode (Section 5.2.3.2).
5.2.1.1
Communicate Toolbar
The Communicate Toolbar contains the buttons that control the communications
between the GUI and the 802, as well as allow the user to set the communications link
parameters. Table 5-2 shows the buttons in the Communicate Toolbar.
Table 5-2: Communicate Toolbar Buttons
Button
5.2.1.2
Name
Logon
Function
Establishes a communication link between the GUI and the 802.
Logoff
Communications
Settings
Operation Toolbar
The Operation toolbar contains several buttons that control various aspects of the
operation of the 802 and the GUI. Table 5-3 shows the buttons in the Operation Toolbar.
Table 5-3: Operation Toolbar Buttons
Button
Name
Function
Temporary Write
to Analyzer
Permanent Write
Configuration to
Analyzer
Updates the on-board real time clock in the 802 to the time
and date set on the linked computer.
Acknowledge All
Latches
Bypass Alarms
Tape Advance
5.2.1.3
General Toolbar
The General Toolbar contains several buttons that allow the user to load and save
configuration files, print configurations, and open the context selective help file. Table 5-4
shows the buttons in the General Toolbar.
Revision 1.2
5-54
21/08/2007
Name
Open
Function
Opens a saved 802 configuration file
Save Current
Configuration
Save
Configuration to a
new file
Print
Context Help
Exit
5.2.2.1
File Menu
The File menu contains two options Open and Exit. Clicking on Open allows the user to
load a previously saved configuration file. Clicking on Exit causes the GUI to close.
5.2.2.2
View Menu
The View menu allows the user to decide which toolbars to display. Toolbars with a
checkmark beside them are displayed, while those without checkmarks beside them are
not displayed. Putting a checkmark beside Upgrade Window brings up the upgrade
window shown in Figure 5-10.
5.2.2.3
Tools Menu
The Tools menu allows the user to configure certain aspects of the GUI. If Update Mode
is checked, then the GUI is in Update Mode. If it is unchecked, the GUI is in Read-Only
mode. For more information on these modes, see section 5.2.3.1 and 5.2.3.2. Selecting
Change Password allows the user to change the password for the Update Mode. The
original password must be entered, and the new password entered twice, in the dialogue
box shown in Figure 5-12.
Revision 1.2
5-55
21/08/2007
Reset Password will reset the Update Mode password to the factory default value. The
user must first confirm this change in the dialogue box shown in Figure 5-13.
5.2.2.4
Help Menu
There is only one option under the Help menu About. Pressing About will bring up a
window that shows the version number of the GUI currently in use. This may be useful for
troubleshooting. The About window is shown in Figure 5-14.
Revision 1.2
5-56
21/08/2007
Revision 1.2
5-57
21/08/2007
5.2.3.1
Read Only mode is a mode of the 802 GUI that does not allow the user to write any
changes to the analyzer. It allows the user to look at all the screens and review all the
settings of the analyzer, but prevents the user from writing any changes to these settings
down to the 802.
5.2.3.2
Update Mode
Update mode is a mode of the 802 GUI that allows the user to make changes to any
setting in the GUI and write these changes to the analyzer. This mode is password
protected to prevent unauthorized changes to the analyzer. If the user originally logs into
the 802 in read only mode, it is possible to enter update mode by going to the Tools menu
and selecting Update. This will bring up the password request dialogue box. If the correct
password is entered, the GUI will then be in Update mode.
5.3
Application Screens
There are several main screens within the 802 GUI. Access to these screens is obtained by
pressing the labelled buttons at the bottom of the screen. There are 10 screens in total. The
functions of each screen are listed in Table 5-5.
Table 5-5: Application Screen Functions
Screen Name
Watch Window
Archive
Event Log
Global
Stream
Input
Output
Ports
Modbus
Maintenance
Revision 1.2
Function
Allows the user to see the current status of the 802, including concentration
readings and input and output status. (Section 5.3.1)
Allows the user to access the numerous archives that store live data, hourly and
daily averages, as well as other user configurable archives. (Section 5.3.2)
Allows the user to access the Event Log, which keeps track of all the configuration
changes and situations that the 802 has experienced. (Section 5.3.3)
Allows the user to change settings that affect the global operation of the 802, such
as stream purge times. (Section 5.3.4)
Allows the user to change settings specific to each stream; normal, zero,
reference, and calibration. (Section 5.3.5)
Allows the user to configure the 802s four digital inputs. (Section 5.3.6)
Allows the user to configure the behaviour of the 802s relays and solenoids.
(Section 5.3.7)
Allows the user to view the communications port settings of all of the 802s
communications ports RS-232, RS-485, and USB. (Section 5.3.8)
Allows the user to configure Modbus lists, including the type of Modbus list and the
data points that are outputted to Modbus. (Section 5.3.9)
Allows the user and service personnel to have greater control over the electronics
board, for testing analog outputs, relays, and solenoids. (Section 5.3.10)
5-58
21/08/2007
Revision 1.2
Explanation
Shows the current date and time on the analyzer. It is important that this be
set correctly so that the archived data (section 5.3.2) has the correct time
stamp. If this is not correct, ensure that the PCs time and date is correct,
and then press the Synchronize Time button.
The 802s serial number
Shows the site name that the 802 is installed at.
Shows the specific location within the site that the 802 is installed at.
Shows the current state of the 802. The various analysis states are shown
in Table 3-2.
Shows the current run type being performed by the 802 normal,
calibration, reference, or zero.
Shows the number of seconds elapsed since the beginning of the current
analysis cycle.
5-59
21/08/2007
Run Number
Trended Reading
PPM
Raw mV
dV/dT
mV
Zero Offset
Calibration Gain
Sensor LED Current
Sensor Temperature
Tape Remaining
Run Counter
Shows the current number of runs of the current analysis type, as well as
the total number of runs of this type to be performed.
Shows a trended concentration reading that takes the values of several
past concentration readings of the same run type into account.
Shows the current calculated stream concentration reading.
Shows the current raw photodiode detector output, in millivolts. This
reading starts at around 1000mV at the start of an analysis, and drops as
the analysis proceeds.
Shows the rate of change in sensor output.
Shows the absolute value of the sensor output, increasing from zero at the
beginning of an analysis.
Shows the value subtracted from all concentration readings. This value is
the parameter that must be subtracted from a reading taken on gas
containing no hydrogen sulphide in order to make the 802 read
approximately 0.00ppm.
Shows the value that the sensor output is multiplied by in order to produce
a valid concentration reading.
Shows the current used to ensure a reading of 1000mV is obtained by the
sensor block on pure white tape.
Shows the current ambient temperature at the sensor block, in degrees
Celsius.
Shows the length of tape, in feet, remaining in the reel. See section
5.3.4.1.1 for more information on how this is calculated.
Shows the current number of runs of all types carried out by the 802 since
it was first started up.
The GUI automatically polls the analyzer every 2 seconds, causing this page to be updated
continuously. This allows the user to be able to monitor the status of the 802 analyzer live.
5.3.1.1
Analysis Alarms
At the bottom of the screen is a box labelled Analysis Alarms. Within this box are 4
smaller boxes that are entitled Normal Run, Calibration Run, Reference Run, and Zero
Reference. Each box contains 3 or 4 labels next to circles. These circles are either green
or red. Green means that the given alarm is not on, while red means that the given alarm
has been triggered. Stream alarms are covered in sections 5.3.5.2.3 (Normal run alarms),
5.3.5.3 (Calibration alarms), 5.3.5.4 (Reference Alarms), and 5.3.5.5 (Zero Alarms).
5.3.1.2
Analog Outputs
The box labelled Analog Outputs shows the current output, from 4 to 20mA, of the two
analog outputs on the 802 electronics board. Analog outputs can only be configured from
the GUI, not from the keypad. Calibration of the analog outputs is covered in section
5.3.10.1, while configuring analog outputs to output certain data points and configuring
ranges is covered in section 5.3.5.2.2.
5.3.1.3
Relays
The box labelled Relays shows the current status of the four solid-state relays. The circle
beside each of the relay titles indicates the status. A red colour means that the relay has
been triggered, be it by an alarm or some other trigger, while a grey colour means that the
relay has not been triggered. Setting up the latching behaviour of relays is described in
section 5.3.7. Associating relays with various alarms and stream types is described in
sections 5.3.4 (Global) and 5.3.5 (Streams).
5.3.1.4
Solenoids
The box labelled Solenoids shows the current status of the four solenoid drivers on the
802 electronics board. If the analyzer is not configured with any solenoids, this box can be
Revision 1.2
5-60
21/08/2007
ignored. A grey circle beside a solenoid label indicates that the solenoid is de-energized,
while a red circle indicates that the solenoid is energized. Setting up the latching
behaviour of solenoids is described in section 5.3.7. Associating solenoids with various
alarms and stream types is described in sections 5.3.4 and 5.3.5.
5.3.1.5
Discrete Inputs
The box labelled Discrete Inputs shows the current status of the four discrete inputs on
the 802 electronics board. These discrete inputs can be such things as low-pressure
switches, oven temperature sensors (on total sulphur analyzers), or piezo buttons used to
initiate specific runs. A red circle beside a discrete input label indicates that the particular
discrete input has been triggered, while a green circle indicates that the particular discrete
input has not been triggered. Setting up discrete inputs is described in section 5.3.6.
5.3.1.6
General Indicators
The box labelled General Indicators indicates the current status of non-analysis related
alarms, as well as the status of relay latches. As before, a green circle beside each label
indicates that the alarm has not been triggered, while a red circle indicates that the alarm
has been triggered. If a latch has a red circle beside it, it means that that relay or solenoid
has been latched. Latches must be manually cleared (section 5.3.7) before the circle
returns to grey. See section 5.3.4.2 for more information on non-analysis related alarms,
and section 5.3.7 for information on setting up latches.
5.3.2 Archive
The archive is where data related to the analyzers operation, including live concentration
data, hourly averages, and other important data, is stored. When this screen is first accessed,
the screen appears as in Figure 5-20.
Revision 1.2
5-61
21/08/2007
first row in the archive has data labels that identify exactly what data is in each column. The
column widths can be automatically scaled to fit the data by pressing the Auto-Size Columns
(
5.3.2.1
Graphing Data
The data in the white columns can be graphed by right clicking and dragging down to
select the desired portion of a given data column. Once the data of interest has been
selected, click on the button at the top of the screen with the label Graph Selected. This
will create a graph that shows the trend in the selected data, an example of which is
shown in Figure 5-22.
Revision 1.2
5-62
21/08/2007
Name
Zoom Mode
Auto-Scale
Show Entire
Graph
Function
Toggles Zoom mode on and off. When the button is clicked, left
clicking on the chart will zoom in on a section of the chart,
centred on the clicked location. Right clicking will zoom out on a
section of the chart, again centred on the clicked location. When
zoomed in on the data, the user can scroll left, right, up and
down using the arrow keys on the computers keyboard.
Automatically adjusts the y-axis scale so that all data within the
zoomed in selection can be seen.
Cancels all zooms and returns to the auto-scaled graph
originally produced when the data was first graphed.
If the user right clicks on the chart, a pop-up menu appears on the chart. The options for
this menu are shown in Table 5-8.
Revision 1.2
5-63
21/08/2007
Set Scale
Print
Copy to
Clipboard
Sub-options
Scale to Data
Scale Manually
N/A
N/A
Set Trace Colour
Set Background
Colour
Reset Default
Colours
Invert Scrolling
Preferences
Print Orientation
Print Sizing
Show Statistics
5.3.2.2
Function
Scales the y-axis to the data. This is the default scaling
when a graph is first made.
Brings up a dialogue box that allows the user to set the
minimum and maximum value for the y-axis scale.
Prints the chart to the printer of the users choice
Copies the chart to the clipboard as an image to allow the
user to paste it into another application, such as Microsoft
Word
Allows the user to set the colours used for both the data
and the statistics lines on the chart.
Allows the user to set the charts background colour.
Resets all colours to their defaults: red for data, blue for
statistics, and grey for the background.
When this option is selected, inverts the function of the
arrow keys while zoomed in on data (i.e. down key moves
up, left moves right, etc).
Allows the user to select whether the chart will be printed
in landscape or portrait mode. The default is landscape.
Allows the user to select whether the printed chart will be
scaled to fit the selected size of paper, or kept in the onscreen proportions. The default is scaled to fit.
Toggles statistics labels on the chart on and off. The
default is on.
Filtering Data
The data table can be filtered to show only certain data by clicking on the funnel ( ) icon
in the header on all but the Log Name column. Data can be filtered by date in the
timestamp column, and by using one of six numerical operators in any column containing
numerical data. The two types of filtering will be covered separately.
5.3.2.2.1
Clicking on the funnel icon on the right hand side of the Timestamp column header
will cause a pull-down menu to appear. This menu contains the following options:
[All] Selecting [All] will turn off any current filtering of data by date.
[Custom] Selecting [Custom] will bring up the dialogue box shown in
Figure 5-23.
Revision 1.2
The dialogue box allows the user to choose a date range for the data that
they wish to display, starting at a certain date and ending at a certain date.
Today shows only data obtained on the current date.
Yesterday shows only data obtained yesterday.
This Month shows only data obtained in the current month.
Last Month shows only data obtained in the previous month.
5-64
21/08/2007
IMPORTANT: The date and time must be correctly set on the 802 for this filtering to
work correctly, as the timestamp on each row of data is obtained from the analyzers
internal clock.
5.3.2.2.2
This method of filtering works best in the sequence number column, although it can be
used in any column that contains numerical data. In the header of the column to be
filtered, clicking on the funnel icon on the right hand side of the header will cause a
pull down menu to appear. The menu contains the options [All] and [Custom]. As
before, selecting [All] will cause any current numerical filtering of data to be cleared.
Selecting [Custom] will bring up the dialogue box shown in Figure 5-24.
Revision 1.2
Operator
Name
>
Greater
Than
<
Less Than
Equal to.
Ending
Digits
And
&
Both
functions
fulfilled
Function
When this operator is used in conjunction with a number, the data will
be filtered such that only data greater in value than the number
entered will be displayed. For example, if the expression >1000 was
entered into the dialogue box, the column will be filtered such that only
data with values greater than 1000 will be displayed.
When this operator is used in conjunction with a number, the data will
be filtered such that only data lower in value than the number entered
will be displayed. For example, if <1000 was entered into the
dialogue box, the column will be filtered such that only data with
values less than 1000 will be displayed.
When this operator is used in conjunction with a number, the data will
be filtered such that only data that is EXACTLY equal to the number
entered will be displayed. For example, if =1000 was entered into the
dialogue box, only data exactly equal to 1000 will be displayed.
When this operator is used in conjunction with a number, the data will
be filtered such that only data ending in the number entered will be
displayed. For example, if *25 was entered, the column will be
filtered such that only data ending in 25 (e.g. 9.2929292925,
1.00000025, etc) will be displayed.
When this operator is used to join a string of numbers together, the
data will be filtered such that only data being exactly equal to each
number in the string will be displayed. For example, if the expression
1^5^7^12 was entered into a list of data containing the whole
numbers from 1 to 20, the only numbers in the list that would pass
through this filter would be the numbers 1, 5, 7, and 12.
When this operator is used to join two or more other operators in a
string, the data will be filtered such that only data that fulfills all of the
operators conditions will be displayed. For example, if the user wished
to display all of the data with values between 7 and 8, they could type
>7&<8 into the dialogue box. Note, of course, that this would filter out
values of exactly 7 and exactly 8.
5-65
21/08/2007
5.3.2.3
Exporting Data
Right clicking on the archive data table gives the user several options for exporting data,
by printing or saving to a file. The file export options are listed in Table 5-10:
Load From File loads a previously saved proprietary data file (.dcar).
Save to File saves the data to a Galvanic proprietary format (.dcar).
Save as Packet CSV
Save as Legal CSV
Save as XLS - saves the data to a Microsoft Excel File
Copy Selection To Clipboard allows the user to copy a selection from the data
table to the clipboard to be pasted into another program, such as Microsoft Excel.
Copy Image to Clipboard copies an image of the table to the clipboard that
can be pasted into an image-editing program or into a Microsoft Word Document.
Hide Selected Column(s) hides one or more selected columns. If a table is
exported to a Microsoft Excel file while columns are hidden, those hidden columns
will NOT be exported to the file.
Unhide All Columns displays any columns of data that may have been
previously hidden.
5.3.2.4
Log Setup
To configure the 802 to log values in data points other than those set up at the factory,
press the Log Setup button at the top of the Archive screen. This will cause a pop-up
window as seen in Figure 5-25 to appear.
Revision 1.2
5-66
21/08/2007
Log On Run the values of the data points logged in this log will update after
every completed analysis cycle.
Log Hourly the values of the data points logged in this log will update every
hour. The log is updated at the top of every hour.
Log Daily the values of the data points logged in this log will update every day.
The time at which the log is updated can be set in the Hour Taken column. The
log can only be updated at the top of the hour. Enter a value of 0 (midnight) to 23
(11pm) in this column.
Log Periodic 1 to 3 these three logs are empty by default when the analyzer is
set up at the factory. Any data point of the users choosing can be logged in each
of these three logs, at a frequency of the users choosing. The frequency of
logging, in seconds, is entered into the column entitled Frequency (Seconds).
Please note that the frequency of logging for each data point within a given
periodic log must be the same; that is to say, if one data point in Log Periodic 1
has a logging frequency of 1000 seconds, all the data points in Log Periodic 1 will
have a logging frequency of 1000 seconds.
On the left hand side of the Log Setup window is a tree that contains 7 nodes. Within
each of these nodes are data points that can be logged. Many of these data points are
static or rarely changed values that should not need to be continuously logged. The data
points of the most interest are stored in the two nodes called Non Persistent Analysis
Status and Non Persistent Status Entries.
Non-Persistent Status Entries contain data points such as the current
concentration readings for all four stream types (normal, calibration, reference,
and zero), the gain, status of various alarms and latches, and the status of digital
inputs and outputs.
Non-Persistent Analysis Status contains data points such as the live sensor
output, hourly and daily minima, maxima, and averages, and the tape remaining.
To expand any node, simply left click on the (+) icon to the left of the node label. This will
expand the node to show all of the data points within that node. To put any data point into
a log, left click on the data point of interest, and while holding the right mouse button
down, drag the data point into the desired log. Once the left mouse button has been
released, the data point should become a row in the table for the log, and the change to
the log configuration will be automatically written to the analyzer. If the data point is in Log
Periodic 1-3, ensure that a frequency is entered for the data point. Note that changing one
frequency time within a given periodic log will cause all the frequency times in that log to
change to the same value. To remove a data point from any log configuration, simply right
click on the row to be removed, and then select the Delete Item option from the popup
menu that appears. Once all the desired changes are made, click on the Read from Unit
to ensure that the changes have correctly written to the unit, and then click on the (x) in
the top right hand corner to close the log configuration window.
Revision 1.2
5-67
21/08/2007
5.3.4 Global
The Global screen shows parameters that affect all four analyzer streams (sample, calibration,
reference, and zero), as well as general parameters that are unaffiliated with the actual
analysis. The Global screen is shown in Figure 5-27.
Revision 1.2
5-68
21/08/2007
5.3.4.1
General
The General box contains fields that affect the overall configuration of the analyzer. There
are 4 smaller boxes within this box, as well as several changeable fields and 4
checkboxes. Each field or box will be explained separately. Whenever any change is
made to a changeable field, the change must be written to the unit by pressing the ( )
button. If more than one change is made on the Global screen, it is not necessary to write
after every change, but all changes to the Global screen MUST be written before leaving
the screen.
Revision 1.2
Live PPM Enable - Putting a checkmark in this box will cause the analyzers LCD
screen to update live with the gas concentration as it is calculated within the
analyzer. When this checkbox is left un-checked, the screen will only update with
a new concentration value when an analysis cycle has completed. If Live PPM is
enabled, the concentration units displayed on the 802s LCD screen will have an L
in front of them i.e. if the concentration units are PPM, then the screen will read
LPPM if Live PPM is enabled.
Live mV Enable - Putting a checkmark in this box will cause the analyzers LCD
screen to show the raw sensor output, counting down from approximately
1000mV, rather than the sensor output that is adjusted to count up from 0mV. If
the Live mV is enabled, the units of the sensor output displayed on the units LCD
screen will change from mV to LmV.
Backlight Lock On - The LCD on the 802 has a backlight that causes the screen
to light up and be easier to read. If this checkbox is checked, the backlight on the
screen will be continuously turned on. This should only be checked if power
consumption on the analyzer is not a serious concern.
5-69
21/08/2007
Timed Analysis Enable Checking this box will disable the standard method of
analysis (section 4.3) and enable the timed analysis method that was used on
previous Galvanic lead acetate tape analyzers. Because the 802 standard
analysis method is significantly more linear and has a broader dynamic range
than the timed analysis method, it is highly recommended that this box remain
unchecked. However, if a consistent cycle length is of the utmost importance,
timed analysis is the best way to ensure this.
Site Identification - This box contains three fields that contain the units unique
identification. The ID field contains the analyzers serial number. The Site field
tells the operator which site the analyzer is installed at, and the location tells the
operator where in the site the analyzer is installed. Do not change the ID field
under any circumstances unless instructed to by Galvanic service personnel.
Version The values in the three unchangeable fields in this box show the
current versions of firmware installed in the analyzer, the sensor block, and the
Rosetta that helps the GUI identify the data it reads from the analyzer. These
values are useful for diagnosing problems.
Bypass Timeout Bypass mode is a mode of the analyzer that disables any
outputs that may be triggered by alarms. This mode is useful for doing any kind of
service work that may result in inadvertent triggering of alarms. However, for
normal operation, it is essential that bypass mode be disabled. Thus, the 802 is
equipped with a bypass timeout. When enabled, by putting a checkmark in the
Enable box, bypass mode will automatically be exited by the analyzer after a
certain number of seconds after the last keypad button push. The length of the
bypass timeout, in seconds, can be set by the user in the Bypass Timeout (s)
field. By default, this is set to 450 seconds, and is disabled. When the 802 is in
bypass mode, the output from both analog outputs goes to the fail-safe output to
ensure that the user recognizes that the analyzer is in bypass mode. See section
4.5.
Furnace Pulse Width This field is only relevant for total sulphur measurement
(section 3.4). The furnace pulse width is the percentage of time that power is
delivered to the total sulphur furnace, and serves as a temperature control
mechanism for the furnace. IMPORTANT: It is essential that this number be set at
50% for analyzers running on 110VAC, and 12.5% for 220-240VAC. Any values
higher than these numbers could result in failure of the furnace element or melting
of the quartz reaction tube within the furnace. This field should not be changed
until after consultation with the Galvanic Applied Sciences service department.
Trend Filter Length This field determines the weighting the current reading has
in the calculation of the historical trend. The larger this number is, the smaller the
weighting the current value has in the calculation of the historical trend. The trend
is a form of data filtering that smoothes the long-term data to show a long-term
trend in concentration. This value should not be changed except after consultation
with the Galvanic Applied Sciences service department.
Purge Time This field determines the length of time the analyzer purges after a
stream switch. When a stream switch occurs, the analyzer will advance the tape,
then purge with the new gas for as many seconds as the number in this field, then
advance the tape again before starting to analyze the new stream. The longer this
number is, the better the results will be on the next stream. A minimum time in this
field should be about 30 seconds.
5.3.4.1.1
Tape Box
There are 4 fields in the Tape box that control the way the analyzer advances tape
between analyses. The first field, Stain Width, shows the width, in feet or metres, of
one stain on the tape. This field is calculated based on values in the next two fields.
Revision 1.2
5-70
21/08/2007
The second field, which is called Stain Width per Pulse, shows the distance the tape
advances every time the pulse counter (section 3.2.2.3) receives a pulse.
Calculating Stain Width per Pulse should only need to be done if the pulley on the
pulse counter needs to be replaced for some reason. Measure the diameter, in
inches, of the pulse counter pulley using an accurate set of callipers. Multiply this
number by pi to get the circumference of the pulley. Divide this number by 12 to get
the circumference of the pulley in feet. The Pulse Counter receives 256 pulses every
time it makes one full revolution, so dividing the circumference of the pulley in feet by
the number of pulses per revolution will give the length of tape advanced when the
pulse counter receives one pulse. This is expressed in the following formula:
Stain Width Per Pulse (feet) = ((Diameter (inches) x ) / 12)/256
The formula can also be expressed in metric as follows:
Stain Width Per Pulse (metres) = ((Diameter (millimetres) x )/1000)/256
If a metric value is entered into this field, the value in the Stain Width field will also be
in metric.
The Stain Width In Pulses box allows the user to choose how many pulses the pulse
counter should receive before stopping a tape advance. The larger the number in this
field, the greater the stain spacing will be. The smaller the number in this field, the
more overlap there will be between stains. By default, this number is set to around 40.
If the spacing is too great with a value of 40, reduce this number to around 37 and
check stain spacing again. Too much overlap between subsequent stains will cause
poor results. Too much white space between subsequent stains will cause faster than
desired tape usage. The maximum value that can be entered into this field is 99. If the
user enters a value higher than this into the field and attempts to write it to the 802,
the field will turn yellow, indicating that an invalid value has been entered. Enter a
value between 1 and 99 into this field.
The Tape Remaining field shows the amount of tape, in feet or metres, remaining on
the reel. This value is used for the Tape Length alarm (section 1.3.4.2) and for
outputting of tape length remaining on analog outputs and the analyzers LCD display.
It is ESSENTIAL that this field be manually updated whenever a tape is replaced to
ensure accuracy of the Tape display on the analyzers LCD and analog outputs. For
example, if the user is using a 300 foot lead acetate tape, it is essential that he or she
update this field to a value of 300 every time a new tape is installed. The value in this
field is correct within 1% on a 350 foot tape.
5.3.4.2
Global Alarms
Global alarms are non-analysis related alarm parameters that affect all streams
identically. These alarms are related to hardware and other physical parameters that
affect the operation of the analyzer. In many cases, the set point for these global alarms is
user configurable. Each alarm can be configured to a) be active or not active and b) to
trigger certain outputs, either solenoids or relays.
Sensor Current If the sensor current (section 3.2.1.2.1) rises above the setpoint value, the alarm is triggered, as there is likely a problem with the sensor
block. The set-point for this alarm is configured at the factory and should not be
changed.
Revision 1.2
5-71
21/08/2007
5.3.5 Stream
The Stream screen shows parameters that affect each individual stream, and all parameters
that directly affect the analysis. The Stream screen is shown in Figure 5-28.
Revision 1.2
5-72
21/08/2007
5.3.5.1
On the left hand side of the screen are fields and boxes that tell the analyzer how to
calculate the concentration, as well as how long to run each cycle for, what type of
analysis to run, and what concentration units to display on the screen.
The Gain Factor is the parameter that the rate of change is multiplied by to get a
concentration. For more detailed information on the analysis and how
concentration is calculated, see section 4.2.4.
The Zero Offset is the value that is subtracted from all concentration readings to
give corrected readings. Zero offset is obtained by running the Zero stream,
which is usually run on a gas that contains no hydrogen sulphide, such as pure
nitrogen. For more information on the Zero, see section 4.4.3.
The Analysis Units field allows the user to change the units that are displayed on
the screen. This can be PPM, PPB, Mol%, or any other unit the user chooses.
Please note that changing the value in this field does NOT change the calibration
of the analyzer if the user wishes to change the calibration of the unit to be in a
different unit, they must use a calibration gas with a known concentration in the
unit they wish to display.
5.3.5.1.1
Each individual stream (Normal, Calibration, Reference, and Zero) can be configured
with a different Sample Delay, Sample Interval, and TSC (mV). In this way,
Calibration, Reference, and Zero runs can be shorter than the Normal Run, allowing
for faster calibration, referencing, and zeroing. This is particularly useful if the analyzer
is not auto-calibrated, as the technician does not need to be on-site for as long to
Revision 1.2
5-73
21/08/2007
calibrate the analyzer. It also allows for conservation of calibration and zero gas, as
these gases do not need to be run for as long for each cycle. The sample timing
method is actually adjusted in each individual stream, accessed via the four tabs at
the top of the screen.
At the top of each individual stream screen are three fields Sample Delay, Sample
Interval, and TSC (mV). The combined total of Sample Delay and Sample Interval
fields, in seconds, is the maximum length an analysis can be. By default, the delay is
set to 10 seconds, and the interval is set to 230 seconds. Thus, the default maximum
cycle time on the 802 is 240 seconds, or 4 minutes. This can be changed by the user
to whatever maximum analysis cycle time they desire. If the Timed Analysis Enable
checkbox is checked in the Global page, the values in these fields should be set to
140 for the delay and 30 for the interval. The TSC (mV) is the sensor output, in
millivolts, at which the analyzer will finish a cycle and advance to a new piece of white
tape. TSC is an acronym for Tape Saturation Cut-off. If the sensor output reaches the
tape saturation cut-off before the maximum cycle time is reached, the analyzer will
consider the analysis cycle to be complete. Thus, the actual cycle time may be shorter
than the maximum cycle time, depending on the magnitude of the tape saturation cutoff, as well as the concentration of gas being analyzed and the configuration of the
analyzer.
5.3.5.2
Normal Run
A Normal Run is an analysis that is carried out on the sample stream. The Normal Run
screen is shown in Figure 5-29.
5.3.5.2.1
Switching
The Switching box allows the user to control what outputs are associated with the
normal run. There are 4 relays and 4 solenoids that can be controlled by the analyzer.
Revision 1.2
5-74
21/08/2007
Putting a checkmark in any of the relay or solenoid boxes will cause that relay or
solenoid to energize when a normal run is running. Note that relays 1 to 3 generally
are used for alarm indications, so it is inadvisable to use any of these relays to
indicate a given run.
5.3.5.2.2
The Analog Output Managers box allows the user to set up each of the two analog
outputs on the 802 board to output a given parameter, as well as the range that each
of the two 4-20mA outputs is scaled to.
If the Enabled box is not checked for the given analog output, that analog output will
not be active for the normal stream.
The Description field allows the user to enter a text description for each analog
output that will be displayed on the 802s LCD screen.
The Function field is a pull down menu that allows the user to choose from a list of
parameters to output to each analog output. By default, analog output 1 is set to
display the live concentration reading and analog output 2 is set to display the amount
of tape remaining. The following parameters can be output:
The Gain Factor.
The Zero Offset.
The Stream 1 Cal. Gas concentration.
The Stream 1 Ref. Gas concentration for the reference run.
The Stream 1 Zero Gas concentration for the zero run.
The Stream 1 Norm Reading, which is the current concentration of
H2S measured in the sample stream.
The Stream 1 Cal Reading, which is the last result obtained on a
calibration run.
The Stream 1 Ref Reading, which is the last result obtained on a
reference run.
The Stream 1 Zero Reading, which is the last result obtained on a
zero run.
The amount of Tape Remaining.
The Sensor Temperature, in degrees Celsius.
The Sensor LED Current, in milliamps.
The Stream 1 Trended Reading.
The Stream 1 Hourly Average, which is the average concentration
reading taken over the last hour.
The Stream 1 Hourly Minimum, which is the minimum concentration
reading observed over the last hour.
The Stream 1 Hourly Maximum, which is the maximum
concentration reading observed over the last hour.
The Stream 1 Daily Average, which is the average concentration
reading taken over the last day.
The Stream 1 Daily Minimum, which is the minimum concentration
reading observed in the last day.
The Stream 1 Daily Maximum, which is the maximum concentration
reading observed in the last day.
To the right of each pull down menu are two fields which tell the analyzer how to scale
each of the analog outputs on the 802 electronics board. The left box field is entitled
Minimum and the right field is entitled Range. The value in the Minimum field is the
minimum numerical value that will be output on the analog output, and the value in the
Range field is the maximum numerical value that will be output on the analog output.
Revision 1.2
5-75
21/08/2007
Note that the scaling for the analog outputs depends on the value being output. If the
user wishes to change the values output on either analog output, he or she must
ensure that the scaling is correct for the chosen output. If the user is uncertain what
range to set a given output value to, please consult Galvanic Applied Sciences for
assistance. To calibrate and test analog outputs, please refer to Section 5.3.10.1.
5.3.5.2.3
Alarms
The Alarms box contains the configuration for the analysis alarms on the normal
stream. There are four types of alarm Alarm 1, PAA Alarm, %Deviation Alarm.
Each alarm can be enabled by putting a checkmark in the check box next to the alarm
label. By default, Alarm 1 and the PAA alarm are enabled for the normal stream. Both
alarm types have an associated field that allows the user to enter a set-point
concentration for each alarm. In addition, both alarm types can be configured to
trigger specific relays and/or outputs. To enable specific outputs for each alarm,
simply put a checkmark in the desired checkboxes in the Actions box.
The PAA, or predictive alarm analysis, is outlined in detail in section 4.4.1.2. The
%Deviation alarm is triggered when consequent runs vary from each other by a
greater percentage than the value set in this field. This alarm is generally disabled for
the normal stream, as it is of little value for normal runs it is expected that some
degree of variation in H2S concentration from analysis to analysis on the normal
sample stream will occur.
5.3.5.3
Calibration
The Calibration Run is set up very similarly to the Normal run, with a few notable
differences. The calibration run is set up to automatically calibrate the analyzer and
calculate a new gain factor. This calibration run can be triggered to start from either a
digital input or on a timed schedule. Note that the analyzer must be equipped with the
auto-calibration option in order to use remote calibration or timed schedule initiation of
calibration. The Calibration screen is shown in Figure 5-30.
Revision 1.2
5-76
21/08/2007
and if the user wishes to have the calibration run initiate on a timed schedule, the Time
checkbox should be checked. If the Time checkbox is checked, the fields in the
Frequency field must be filled in. The Day of Week field tells the analyzer what day of the
week to run the calibration on. This value can be from 0 to 7, with zero telling the analyzer
to run the calibration every day, 1 being Monday, and 7 being Sunday. In the Oclock
field, an hour from 0 (midnight) to 23 (11pm) must be entered. In the Minutes field, a
number of minutes, from 0-60, can be entered. So, for example, if the number 6 was
entered in the Day of Week field, 21 was entered in the Oclock field, and 15 was entered
in the Minutes field, the analyzer would run a calibration every Saturday at 9.15pm.
IMPORTANT: If the user is using a remote or timed calibration, it is essential that the
analyzer have calibration gas correctly hooked up to the analyzer calibration port, and that
the correct calibration gas concentration value be entered in the Cal. Gas Concentration
field directly above the Frequency box (see below for more information). If not, the
analyzer will not calibrate correctly.
Below the three checkboxes at the top of the screen are four fields entitled Name,
Consecutive Runs, Runs to Average, and Cal. Gas Concentration. In the Name field
the user can enter a 7 character or less description of the Calibration run. This will be
displayed in the Current Analysis field in the Watch Window, as well as on the analyzers
LCD Display. The Name field for the Calibration run is set to Cal by default. In the
Consecutive Runs field, the user can set how many cycles to run during each calibration.
This is set to 4 by default. In the Runs to Average field, the user can set how many of the
consecutive runs to average. For example, if the Consecutive Runs field is set to 4, and
the Runs to Average is set to 3, the analyzer will average the values of the last 3
calibration runs to calculate the new gain factor. See section 4.4.2 for more information on
the operation of the calibration run and the nature of the gain calculation. In the Cal. Gas
Concentration field, enter the value of the concentration that is listed on the certificate
that came with the calibration cylinder. Getting a valid calibration is dependent on the
calibration gas concentration being entered correctly, so it is ESSENTIAL that the
calibration gas value be entered precisely as it is written on the calibration gas certificate.
A value of zero cannot be entered into this field if a value of zero is entered into this field
and the user attempts to write this value to the unit, the field will turn yellow, indicating that
an invalid value has been entered into the field. In addition, this value will not have been
written to the 802.
The Switching and Analog Output Managers boxes are set up identically to the way the
normal run is set up. Generally, Solenoid 1 is used as the auto-calibration solenoid, so the
Solenoid 1 checkbox should be checked for Calibration. By default, the analog outputs are
disabled for all runs except the normal runs, to prevent operator confusion. However,
should the user wish to output values from the calibration to analog output, they can do so
via the pull down menus in the two analog output configurations.
The Alarms box is configured similarly to the way it is configured for the Normal run. To
avoid accidentally triggering an alarm and shutting in the plant while the analyzer is in
calibration, it is advised that Alarm 1 and Alarm 2 be disabled for the Calibration run. The
%Deviation Enable is the alarm that is of most importance for the Calibration run. When
enabled, this alarm compares the current value for the gain to the new value of the gain
calculated during the analysis. If the new gain varies from the old gain by more than plus
or minus the percentage value entered in this field, the analyzer will trigger an alarm and
the new gain will not be accepted by the analyzer. This alarm is particularly useful for
analyzers that are calibrated remotely without a service person on site, as it will prevent
incorrect gains caused by such things as empty calibration cylinders or empty hydrogen
(on total sulphur units) from being entered.
As with the normal run, the alarms on the calibration run can be configured to trigger any
digital output the user desires. To initiate a calibration from the GUI, press the button
Revision 1.2
5-77
21/08/2007
marked Stream 1 Calibration. When this button is pressed, a confirmation box, shown in
Figure 5-31, appears to ask for confirmation of the intent to run a calibration.
5.3.5.4
Reference
The Reference run is a way to check whether or not the analyzer is correctly calibrated. It
is configured in much the same way as the Calibration run, although this run only
displays the result of 1 run (or the average of 2 or more runs) on the same calibration gas
as used for the calibration run. It does not recalculate the gain as the calibration run does.
To determine whether or not the analyzer needs to be calibrated, run a reference. If the
reading obtained by running the reference is close to the certificate for the calibration
cylinder analyzed, then the analyzer is still correctly calibrated. If not, a calibration should
be run immediately afterwards.
The %Deviation Enable alarm on the Reference run is different from that on the
calibration run. The reference deviation alarm is based on the percentage difference in
reading from the previous reference reading. If the current reference reading differs by
more than the percentage value entered in this field, a % deviation alarm is triggered. This
indicates that a calibration should be run. To ensure that there is a good reference to
compare the % deviation alarm to, a reference should be run immediately after a
calibration.
All other settings are configured the same way as for the Calibration run. The solenoid for
the reference should be the same as the one for the Calibration, as the two runs both use
the same gas cylinder. The Reference run can also be initiated from the GUI by pressing
the Stream 1 Reference button and confirming the request in the pop-up dialogue box,
just as in the Calibration run.
Please note that if any analysis alarms have been triggered on the Normal run (i.e. Alarm
1 or PAA), any reference run triggered from the GUI will be put into the run queue and not
run until the alarm condition(s) has/have cleared.
5.3.5.5
Zero
The Zero run is a way to check what the analyzer will read on a gas that contains no H2S.
Running a zero run will result in a zero offset being calculated for the analyzer. The zero
offset can be set up similarly to the Calibration and Reference runs, as either a remotely
triggered or timed scheduled analysis. The Zero run is generally associated with a
different solenoid than the Calibration and Reference runs use by default, the Zero is
set to use solenoid 2. The Zero run can be configured like the Calibration run in that it
can be set to run several analyses and take the average to calculate the zero offset.
Revision 1.2
5-78
21/08/2007
It is important to note that a suitable zero gas is used for the Zero run. A suitable zero gas
can be ultra-pure nitrogen, or instrument air that is known to contain no H2S. If the zero
gas contains a significant amount of H2S, the ability of the instrument to operate correctly
can be impaired. If a gas with no H2S is unavailable, it is essential that a value reflecting
the accurate H2S concentration of the gas used for the zero be entered into the Cal. Gas
Concentration field. If the gas does not contain any H2S, the value in this field should be
set to zero.
The %Deviation Enable alarm on the Zero run is similar to the one on the reference run
it is deviation relative to the previously obtained zero value.
The Zero run can also be initiated from the GUI in the same way as the Calibration and
Reference runs. Please note that if any analysis alarms have been triggered on the
Normal run (i.e. Alarm 1 or PAA), any zero run triggered from the GUI will be put into the
run queue and not run until the alarm condition(s) has/have cleared.
5.3.6 Input
There are 4 digital inputs on the 802. The Input screen allows the user to configure these
digital inputs in any manner. These inputs can be configured to trigger a certain type of run, or
to monitor external pressure or temperature switches, for example. The Input screen is shown
in Figure 5-32. Each digital input is configured in the same manner. At the top of each digital
input is a field called Discrete Input (x) Name. The value entered in this field (alphanumeric,
maximum 6 characters) is displayed on the Watch Window screen. In addition, whenever a
digital input is triggered, this value will be displayed on the Alarm screen on the analyzers
LCD screen (See Section 3.3.3). Ensure the value entered in each name field is
understandable.
Below the name field are two checkboxes, one labelled Enable and one labelled Alarm. If the
Enable checkbox is enabled, the digital input is enabled and monitored. If not, the digital input
is disabled and will not respond to any inputs. If the Alarm checkbox is checked, the condition
of each digital input will be logged in the Event Log (section 5.3.3). Below the two
checkboxes for each digital input are three boxes, Polarity, Run Selector, and Actions. The
Polarity box contains two radio buttons labelled Active On and Active Off. When Active On is
selected, the input will be considered triggered if an ON (binary 1) signal is received from the
digital input. If Active Off is selected, the input will be considered triggered if an OFF (binary 0)
signal is received from the digital input. The Run Selector box contains 4 checkboxes that
allow the user to select what run should be associated with each digital input. By default,
Discrete Input 3 is configured to initiate a calibration run.
Revision 1.2
5-79
21/08/2007
5.3.7 Output
The Output screen allows the user to configure the behaviour of the 8 outputs 4 relays and
4 solenoids - on the analyzer main electronics board. The Output screen is shown in Figure
5-33.
Revision 1.2
5-80
21/08/2007
5.3.8 Ports
The Ports screen shows the configuration of the 3 main communications ports on the 802
analyzer the Front Panel Serial Port (RS-232), the ISO 485 Port (RS-485), and the
MODEM Port (RS-232). The Ports screen is shown in Figure 5-34, on the following page.
The configurations of these communications ports are not user configurable. The purpose of
this screen is to show the user the communications settings for each of the three
communications ports to allow for successful communication between the analyzer and the
PC. Ensure that the settings on this screen match the settings entered in the communications
setup box in Section 5.2.3.
Revision 1.2
5-81
21/08/2007
5.3.9 Modbus
The Modbus screen allows the user to set up the 802 analyzer to output data in 3 varieties of
MODBUS communication protocol Enron, Modicon 16, and Modicon Floating Point. As there
are a large number of possible configurations for Modbus, and every user may desire a
slightly different configuration, this manual will only cover the basics of Modbus configuration.
If the user requires assistance in setting up a Modbus configuration, the Service department at
Galvanic Applied Sciences Inc. will assist in configuring the unit as desired. To set up a new
Modbus list, press the New Modbus List button at the bottom of the screen. This will bring up
the dialogue box shown in Figure 5-35.
Revision 1.2
5-82
21/08/2007
5.3.9.1
Enron
An Enron Modbus list contains 4 nodes on the Modbus tree. They are Coils, Short
Integers, Long Integers, and Floating Point. Coils are Boolean data points that is to
say, they have a value of either 0 or 1. To that end, data points entered into the Coils
node would usually be data points that are simple status indicators, indicating the status of
a given alarm, input, or output. If the value of the data point is 0, the status of the logged
alarm, relay, etc, would be off, and if the value is 1, the status would be on. Short integers
are 16 bit whole numbers with either positive or negative sign. Long integers are 32 bit
whole numbers with either positive or negative sign. Floating point values are also 32 bit
numbers, but unlike the integers they do not have a sign, but they do have decimal points.
5.3.9.2
Modicon 16
A Modicon 16 Modbus list contains 4 nodes on the Modbus tree. They are Output Status,
Input Status, Input Register, and Output Register. The Input and Output Status nodes
contain Boolean data points. Data points in the Output Status node are able to be written
to, so the Output Status node can contain data points such as stream requests. If the
value in a given stream was changed from 0 to 1, the analyzer would then initiate that
stream. Data points in the Input Status node are read-only, so this node would contain
data points such as alarm, input, and output status that cannot be changed remotely.
Input Registers are data points that are read-only outputs of analyzer data, such as
calculated concentration. Please note that Modicon 16 only outputs data as 16 bit
numbers, so the display of decimal points in this type of Modbus list is not possible.
Output Register contains non-Boolean data points that can be written to remotely. These
would include such things as tape length, calibration gas concentration, and gain factor.
5.3.9.3
Just like the Modicon 16 Modbus list, the Modicon with Floating Point (MODICON 32) list
contains 4 main nodes on the Modbus tree. However, the Input and Output Registers
contain sub-nodes. They are Register Short, Register Long, and Register Float. Thus,
the input and output registers can output data in 16-bit, 32-bit, or 32-bit with floating point
in the Modicon with Floating Point Modbus list.
Revision 1.2
5-83
21/08/2007
5.3.9.4
The default Modbus setup for the 802 uses the Modicon with Floating point format. Data
points are added to the Modbus list in much the same way that data points are added to
the Archive (see Section 5.3.2.4 Log Setup). Figure 5-37, on the following page, shows
a typically configured Modicon with Floating Point Modbus list.
On the left hand side of the screen is a narrower field entitled Available Points. In this
field are several expandable nodes. The nodes of most interest are Non-Persistent
Status Entries and Non-Persistent Analysis Status. These contain the data points that
are most likely to be output in a Modbus list, as they are the most relevant to the operation
and status of the analyzer. The following data points are output by default:
Stream 1 Normal Reading, which is found in the Non-Persistent Status Entries
node (expanded by pressing on the (+) sign to the left of the node label). This
value is the current value of the gas concentration in the sample stream
calculated by the analyzer. This value will update live every second if the analyzer
is in PAA alarm mode.
Gain Factor, which is found in the Non-Persistent Analysis Status node. This
value is the current value of the calibration gain, which is used to calculate the
concentration of the gas being analyzed.
System Alarm Status, which is found in the Non-Persistent Status Entries node.
The value in this entry, either 0 or 1, indicates whether or not any of the global
alarms (Section 1.3.4.2) have been triggered. If the value is 0, no global alarms
have been triggered. If the value is 1, a global alarm has been triggered.
Revision 1.2
5-84
21/08/2007
There are several columns in the Modbus list table. The first is Modbus Items which
shows the Modbus list tree and the various data points that are in each node. Register is
the most important, as it shows the address of each data point. These addresses are
essential for setting up the device that will read the Modbus list, so it knows which data
point is output to which register. Type shows the type of each data point. Possible values
in this column include Boolean, unsigned long, and float. Most data points are either
Boolean or Float. Floats can be output in the short and long nodes, but the decimal points
will be lost. Certain data points have specific units associated with them, and these will be
shown in the Units column. If a data point does not have units associated with it, this
column will show a value of Unknown for that data point. In the Value column, Boolean
data points will have either a value of True (if on) or False (if off). Numerical data points
will show the value currently entered in the given data point. The Permissions column
shows whether a given data point is Read Only or Writeable. Writeable data points are
the only data points that can be written to remotely through Modbus.
5.3.10
Maintenance
The Maintenance screen is a password-protected screen that allows the user to test and
configure certain parts of the analyzers hardware. The Maintenance screen is shown in
Figure 5-38.
Revision 1.2
5-85
21/08/2007
Revision 1.2
5-86
21/08/2007
Press the Output Zero button. The multimeter should read exactly the value entered in
the Scaled Value field in the Zero column, usually exactly 4 milliamps. If the multimeter
does not read exactly this value, use the up and down arrow buttons next to the raw value
field in the zero column to adjust the current output to the analog output until it does read
the exact value of the scaled value field. Once the multimeter does read this exact value,
press the Output Span button and repeat the process. This time, the meter should be
made to read exactly the value in the scaled value field in the span column, usually
exactly 20 milliamps.
Finally, the fail-safe position should be calibrated. The fail-safe position is the value that
the analog output will output when the analyzer is in bypass mode, and is a value outside
the standard scaled range of the analog output. For example, if the analog output is
scaled to output 4 to 20 milliamps, the fail-safe value could be 3 or 21 milliamps. To
calibrate the fail-safe position, press the Output Fail-Safe button. Using the arrow buttons
beside the Fail-Safe Position field, adjust the value in this field until the meter reads 3
milliamps, or whatever the chosen fail-safe value is.
Once the analog output is correctly calibrated, press the Capture Calibration button to
write this new configuration to the analyzer. Then, test that the output works correctly
using the Analog Output Test box. Enter a value between 0 and 100 into the Test
(Percent) field and ensure that the reading on the multimeter is the same as the reading
in the Test mA field. If it is not, the calibration is not correct, and needs to be changed. If it
is, the calibration is correct. Press the Calibrate AO 1 button again to close this field, and
then press the Calibrate AO 2 button and repeat the process for analog output 2. Make
sure to move the meter to analog output 2. Once both analog outputs are calibrated, press
the Permanent Configuration Write to Analyzer (
new Analog Output configuration to the 802.
Revision 1.2
5-87
21/08/2007
reads back the current configuration and saves it to a backup file. It will also bring up the
Firmware Update screen shown in Figure 5-39 as an overlay to the Maintenance screen.
Revision 1.2
5-88
21/08/2007
Section 6 Troubleshooting
The following is a list of commonly observed problems that may be encountered during normal
operation and servicing of the 802.
Table 6-1: Troubleshooting Guide
Problem
Cause
Solution
1.
2.
3.
4.
5.
6.
Total Sulphur readings are
erratic.
Revision 1.2
6-89
21/08/2007
Section 7
Drawings
Revision 1.2
7-90
21/08/2007
Revision 1.2
7-91
21/08/2007
Revision 1.2
7-92
21/08/2007
Revision 1.2
7-93
21/08/2007
Revision 1.2
7-94
21/08/2007
Revision 1.2
7-95
21/08/2007
Section 8 Configuration
Calibration Data
Customer:
Address:
City:
St/Prov:
Zip/PC:
Country:
Contact:
Phone:
Installation Site:
Site contact:
P.O. No:
Analyzer S.N.
S.O. No:
Sensor #
Issued:
Completed:
Shipped:
Firmware Version:
Rosetta Version:
Sensor Firmware
Version:
H2S / TS
Standard Analysis or
Timed Analysis?
Sample Delay
Sample Interval
Included Options
Sample System:
Y/N
Enclosure:
Y/N
Gauge Make:
GO PR-1
Gauge Range:
0-30 psig
Filter Type:
BA0075
Filter Element:
MC0477
PSV:
Y/N
Set:
15 psig
Flow Meter:
Model #65-AV1
Tube:
#A1128
Auto-Cal Solenoid
Y/N
Flow Setting:
Revision 1.2
8-96
21/08/2007
802 Checklist
Aperture strip size Correct for Range
range
ppm
ppm
AO 1 Range
AO 1 Value Output
AO 2 Range
AO 2 Value Output
TSC (mV) same for all streams?
If No
Y/N
mV
mV
mV
mV
Y/N
Seconds
Chart included
Zero Offset
Gain
Date of Test:
Technician:
Revision 1.2
8-97
21/08/2007
Revision 1.2
DataBase Config
Rosetta Version
Firmware Version
Unit ID
Unit Site
Unit Location
Operator Password
Superviser Password
Factory Password
Backlight Lock On
Black Tape Detect Volts (mV)
Black Tape Detect Enable
Purge Time
Max Analysis Time
ARCNET Threshold
Battery Alarm Enable
Battery Alarm Actions
Low Tape Alarm Enable
Low Tape Alarm Actions
Sensor Comm Alarm Enable
Sensor Comm Alarm Actions
Sensor Cal Alarm Enable
Sensor Cal Alarm Actions
Tape Break Alarm Enable
Tape Break Alarm Actions
High Temp. Alarm Enable
High Temp. Alarm Limit
High Temp. Alarm Actions
Low Temp. Alarm Enable
Low Temp. Alarm Limit
Low Temp. Alarm Actions
Sensor Current Alarm Enable
Sensor Current Alarm Limit
Sensor Current Alarm Actions
Max Tape Adv. Alarm Enable
Max Tape Adv. Alarm Limit
Max Tape Adv. Alarm Actions
Tape Length Alarm Enable
Tape Length Alarm Limit
Tape Length Alarm Actions
Max Analysis Time Alarm Enable
Max Analysis Time Alarm Limit
Max Analysis Time Alarm Actions
Bypass Timeout Enable
Bypass Timeout
MODBUS Address
Live PPM Enable
Live mV Enable
Trend Filter Length
Analysis Units
Gain Factor
Zero Offset
Timed Analysis Enable
PAA Enable
PAA Level
PAA Actions
Time Enable
9-98
131585
66063
SNXXXX802XXX
XXXXXXX
XXXXXXX
0
0
0
True
200
False
30
660
250
True
00000100
False
00000100
True
00000100
True
00000100
True
00000100
True
40
00000100
True
0
00000100
True
14
00000100
True
60
00000100
True
10
00000100
True
630
00000100
False
450000
255
False
False
20
PPM
25.0602
0.600001
False
True
8.5
00000010
False
21/08/2007
43
25
26
24
45
46
47
48
49
50
51
419
52
53
54
55
420
56
57
58
59
60
61
421
66
304
305
403
68
69
27
28
70
71
72
73
74
422
75
76
77
78
423
79
80
81
82
83
84
30
31
464
424
89
404
405
406
91
92
32
33
93
94
95
Revision 1.2
Condition Enable
Sample Delay
Sample Interval
Tape Saturation Cutoff (mV)
Name
Consecutive Runs
Switching
Analog Output 1 Enable
Analog Output 1 Description
Analog Output 1 Function
Analog Output 1 Range
Analog Output 1 Minimum
Analog Output 2 Enable
Analog Output 2 Description
Analog Output 2 Function
Analog Output 2 Range
Analog Output 2 Minimum
Deviation Alarm Enable
Deviation Alarm Limit
Deviation Alarm Actions
Alarm 1 Enable
Alarm 1 Limit
Alarm 1 Actions
Time Enable
Condition Enable
Sample Delay
Sample Interval
Tape Saturation Cutoff (mV)
Name
Consecutive Runs
Runs To Average
Stream 1 Cal. Gas (PPM)
Switching
Analog Output 1 Enable
Analog Output 1 Description
Analog Output 1 Function
Analog Output 1 Range
Analog Output 1 Minimum
Analog Output 2 Enable
Analog Output 2 Description
Analog Output 2 Function
Analog Output 2 Range
Analog Output 2 Minimum
Deviation Alarm Enable
Deviation Alarm Limit
Deviation Alarm Actions
Alarm 1 Enable
Alarm 1 Limit
Alarm 1 Actions
Scheduled Day Of Week
Scheduled Hour
Scheduled Minute
Time Enable
Condition Enable
Sample Delay
Sample Interval
Tape Saturation Cutoff (mV)
Name
Consecutive Runs
Runs To Average
Stream 1 Ref. Gas (PPM)
Switching
Analog Output 1 Enable
Analog Output 1 Description
9-99
True
10
170
2500
Norm
1
00000000
True
PPM_AO
450
10
0
True
TAPE_AO
520
350
0
False
20
00000000
True
7
00000001
False
True
10
170
2500
Cal
4
3
5.103
00010000
False
PPM_AO
451
50
0
False
TAPE_AO
520
350
0
False
20
00000000
False
16
00000000
1
1
0
False
True
10
170
2500
Ref
4
3
15
00010000
False
PPM_AO
21/08/2007
96
97
425
98
99
100
101
426
102
103
104
105
106
107
35
36
465
427
112
510
521
522
114
115
37
38
116
117
118
119
120
428
121
122
123
124
429
125
126
127
128
129
130
40
41
466
386
285
470
471
472
483
484
247
248
249
250
251
252
253
254
255
256
257
Revision 1.2
9-100
452
50
0
False
TAPE_AO
520
350
0
False
20
00000000
False
16
00000000
1
1
0
False
True
10
170
2500
Zero
4
3
0
00100000
False
PPM_AO
453
50
0
False
TAPE_AO
520
350
0
False
20
00000000
False
0
00000000
1
1
0
10
10
True
1000
15
600
600
ALM1
False
ALM2
False
MAINT
False
Relay4
False
CAL
False
ZERO
21/08/2007
258
259
260
261
262
263
540
533
264
265
442
266
541
534
267
268
443
269
542
535
270
271
444
272
543
536
273
274
458
275
278
279
280
281
282
409
544
283
286
287
288
289
290
445
545
291
292
293
294
295
296
297
298
299
300
301
516
517
306
307
308
309
310
311
Revision 1.2
9-101
False
Sol 3
False
Sol 4
False
False
False
True
FURN
00000100
00000000
False
False
True
PRESS
00000100
00000000
False
True
True
CAL
00000000
00000010
False
False
False
DI 4
00000000
00000000
7833
10426
51777
4
20
41351
10426
3
7876
10504
52518
4
20
42014
10504
3
50
False
False
False
False
False
False
False
False
True
0
0.000480647
60
5
8
0
1
False
5
21/08/2007
312
313
314
315
316
317
318
319
320
321
322
323
324
325
Revision 1.2
Data Bits
Parity
Stop Bits
Envelope
Baud Rate
Data Bits
Parity
Stop Bits
Envelope
Baud Rate
Data Bits
Parity
Stop Bits
Envelope
8
0
1
False
5
8
0
1
False
5
8
0
1
False
9-102
21/08/2007
Section 10
Technical Specifications
Total Sulphur
Linearity
Repeatability
Ambient
Cycle Time
Tape Life
Zero Drift
Power Consumption
Power Input Range
Analog Outputs
Other Outputs
Inputs
Electronics
Back Pan
Dimensions
Weight
Revision 1.2
16 bits of resolution
Isolation: 500V input to output or output to 24 VDC supply; 2kV output to power line
4 relay contact SPDT (form C), 2 Amps, 220VDC maximum - function is user programmable
512 Kbytes non-volatile memory for data and event logging (up to 10 months of H2S / total
sulphur concentration data logs)
Tape length sensor to accurately measure tape movement and usage (reported length is
within 1% of actual length)
10-103
21/08/2007
Section 11
11.1
MSDS Information
Acetic Acid
Anachemia
Emergency Numbers:
255 Norman
H8R 1A3
WHMIS
Protective Clothing
TDG Road/Rail
CI#
Not available.
Chemical formula:
Not applicable
CAS#
Not applicable
Synonyms:
Code
R-0020
Supplier:
Anachemia Canada
255 Norman
Ville St. Pierre (Montreal). Que.
H8R 1A3
Formula weight
Not applicable
Material uses
Supersedes
Revision 1.2
CAS#
64-19-7
7732-18-5
ACETIC ACID, GLACIAL
%
1-5
Balance
TLV
Exposure limits: ACGIH
TWA 10 ppm (25 mg/m3);
3
STEL 15 ppm (37 mg/m )
Not established by ACGIH
ORAL (LD50):
Acute:
DERMAL (LD50):
Acute:
VAPOR (LC50):
hour(s)).
Acute:
UNREPORTED (LD50):
Acute:
INTRAVENOUS (LD50):
Acute:
11-104
21/08/2007
PH (1% soln/water)
Odor threshold
>0.21 ppm
Percent volatile
100% (V/V)
Freezing point
Not available.
Boiling point
Not available.
Specific gravity
Vapor density
Vapor pressure
Not available
Evaporation rate
Not available
Solubility
Miscible in water
Flammable limits
Not available.
Auto-ignition temperature
Not available.
Fire
extinguishing
procedures
Use DRY chemical, carbon dioxide, alcohol-resistant foam or water spray. Wear
adequate personal protection to prevent contact with material or its combustion
products. Self-contained breathing apparatus with a full facepiece operated in a
pressure demand or other positive pressure mode. Cool containing vessels with
flooding quantities of water until well after fire is out.
Fire
and
Hazards
Explosion
Effects of Acute
Exposure
Harmful by inhalation, in contact with skin and if swallowed. Target organs: liver kidneys,
eyes, skin, cardiovascular system, respiratory system, lungs, teeth.
Eye
Causes irritation.
Skin
Inhalation
Material is irritating to mucous membranes and upper respiratory tract. May cause
central nervous system depression (e.g., headache, nausea, vomiting), coughing,
dyspnea, thoracic pain, bronchopneumonia, chemical pneumonitis, pulmonary edema.
Ingestion
Effects of Chronic
Overexposure
Revision 1.2
Conjunctivitis, dark coloration of the skin, dermatitis, dental erosion, pharyngitis, chronic
bronchitis, hyperkeratosis, pyrosis, constipation, kidney damage. May cause allergic skin
reaction. Detected in maternal milk in animals. To the best of our knowledge the chronic
toxicity of this substance has not been fully investigated.
11-105
21/08/2007
Immediately flush eyes with copius quantities of water for at least 15 minutes holding
lids apart to ensure flushing of the entire surface. Call a physcian.
Skin contact
Immediately flush skin with plenty of water for at least 15 minutes while removing
contaminated clothing and shoes. If irritation occurs or persists seek medical attention.
Wash contaminated clothing before reusing.
Inhalation
Remove patient to fresh air. Administer approved oxygen supply if breathing is difficult.
Administer artificial respiration or CPR if breathing has ceased. Call a physician.
Ingestion
If conscious, wash out mouth with water. Have conscious person drink several glasses
of water to dilute. DO NOT induce vomiting. Call a physician. Never give anything by
mouth to an unconscious person. Guard against aspiration into lungs. If spontaneous
vomiting occurs, have victim lean forward with head down to avoid breathing in of
vomitus, rinse mouth and administer more water.
Not available
Incompatibility
Reaction Products
Wear self-contained breathing apparatus, rubber boots and heavy rubber gloves. Full
suit.
Evacuate the area. Cover with dry soda ash or lime. Absorb on sand or vermiculite and
place in a closed container for disposal. Ventilate area and wash spill after material
pick up is complete. DO NOT empty into drains. DO NOT touch damage container or
spilled material. Adequate is required for soda ash due to release of carbon dioxide
gas.
Waste disposal
According to applicable regulations. This material and its container must be disposed of
in a safe way.
Store in a cool place away from heated areas, sparks, and flame. Store in a wellventilated area. Store away from incompatible materials. Do not add any other material
to the container. Do not wash down the drain. Do not breathe gas/fumes/vapor/spray.
In case of insufficient ventilation, wear suitable respiratory equipment. Keep away from
direct sunlight or strong incandescent light. Keep container tightly closed and dry. Keep
away from combustible and oxidizing materials. Manipulate in a well-ventilated area or
under an adequate fume hood. Empty containers may contain a hazardous residue.
Handle and open container with care. Take off immediately all contaminated clothing.
This product must be manipulated by qualified personnel. Do not get in eyes, on skin,
or on clothing. Wash well after use. In accordance with good storage and handling
practices. Do not allow smoking and food consumption while handling.
Revision 1.2
11-106
21/08/2007
Splash goggles. Impervious neoprene gloves, apron, coveralls, and/or other resistant
protective clothing. Sufficient to protect skin. A OSHA/MSHA jointly approved respirator
is advised in the absence of proper environmental controls. If more than TLV, do not
breathe vapor. Wear self-contained breathing apparatus. Do not wear contact lenses.
Make eye bath and emergency shower available. Ensure that eyewash station and
safety shower is proximal to the workstation location.
Engineering controls
Revision 1.2
Irritant! Do not breathe vapor. Avoid all contact with the product. Avoid prolonged or
repeated exposure. Manipulate in a well-ventilated area or under an adequate fume
hood. Handle and open container with care. Container should be opened only by a
technically qualified person. Harmful to aquatic life at very low concentrations. Can be
dangerous if allowed to enter drinking water intakes. Do not contaminate domestic or
irrigation water supplies, lakes, streams, ponds or rivers.
RTECS NO: AF1225000 (ACETIC ACID, GLACIAL).
11-107
21/08/2007
11.2
SECTION I MANUFACTURER
MANUFACTURES NAME
Galvanic Applied Sciences Inc.
WHMIS CLASSIFICATION
CLASS D2A
CHEMICAL FAMILY
FORMULA
PB (C2H3O2) 2H2O
APPROXIMATE
CONCENTRATIONS %
LEAD ACETATE
1-5
C.A.S.
NOS.
LD50/LC50 SPECIFY
SPECIES & ROUTE
EXPOSURE LIMITS
301-04-2
FREEZING POINT C
DECOMPOSES AT 200C
PHYSICAL STATE
SOLID
pH
5.5 6.5
Revision 1.2
11-108
21/08/2007
FLAMMABLE LIMITS
Not applicable
LEL
UEL
Not applicable
Not applicable
EXTINGUISHING MEDIA
FOAM CO2, DRY CHEMICAL
SPECIAL FIRE FIGHTING PROCEEDURES
FLAMABILITY
PAPER TAPE WILL BURN (LEAD ACETATE NOT COMBUSTIBLE)
CONDITIONS
OPEN FLAME
SPECIAL PROCEDURES
WEAR SCBA, DUE TO PRESENCE OF TOXIC LEAD FUME
LOWER EXPLOSIVE LIMIT
N.App.
SENSITIVITY TO IMPACTsed)
N.App.
NO
TDG CLASSIFICATION
9.2 SUBSTANCE HAZARDOUS TO THE ENVIRONMENT
Revision 1.2
11-109
CONDITIONS
HEAT, STRONG SUNLIGHT
21/08/2007
SKIN CONTACT
INHALATION: ACUTE
YES
CHRONIC
NO
EYE CONTACT
YES IRRITATION
INGESTION
YES
YES
IRRITANCY
SYNERGISTIC MATERIALS
CERTAIN NITROSO- AND AMIDE CHEMICALS MAY
INCREASE KIDNET TOXICITY.
EYE IRRITANT
CARCINOGENICITY
REPRODUCTIVE EFFECTS
POSSIBLY
TERATOGENICITY
YES
MUTAGENICTY
POSSIBLY
Revision 1.2
YES
11-110
21/08/2007
PERSONAL
PROTECTIVE
EQUIPMENT
GLOVES
NITRILE, NEOPRENE, PVC, NATURAL
RUBBER
EYE
NOT REQUIRED *
RESPIRATORY
NOT NECESSARY FOR ROUTINE HANDLING
OF TAPE
FOOTWEAR
NO SPECIAL REQUIREMENTS *
CLOTHING
NO SPECIAL REQUIREMENTS *
WASTE DISPOSAL
DISPOSE OF IN ACCORDANCE WITH ALL LOCAL, PROVINCIAL, STATE AND FEDERAL REGULATIONS
HANDLING PROCEDURES & EQUIPMENT
WEAR IMPERVIOUS GLOVES (NOT LEATHER OR COTTON SEE ABOVE). AVOID ADBRASION TO TAPE
(WHICH MIGHT GENERATE DUST.)
STORAGE REQUIREMENTS
STROE IN A COOL, DRY, WELL VENTILATED AREA AWAY FROM HEAT, STRONG SUNLIGHT AND
IGNITION SOURCES.
SPECIAL SHIPPING INFORMATION
UN 1616, SUBSTANCE HAZARDOUS TO THE ENVIRONMENT, CLASS 9.2. (NOTE: NOT REGULATED IN
QUANTITIES LESS THAN 230 KG)
PREPARATION DATE
PHONE NUMBER
EXPIRY DATE
(403) 252-8470
Revision 1.2
11-111
21/08/2007
Section 12
Revision History
Revision 1.2
12-112
21/08/2007