Insertion of the Intra-aortic Balloon Catheter

ABBREVIATEDGUIDE FOR SENSATION AND SENSATION PLUS

Preparing the IAB Catheter

1
Push then twist to firmly attach
the one-way valve to the male
luer fitting of the IAB catheter.

IAB Catheter Insertion

Using the provided 30cc syringe


apply a 30cc vacuum.

Twist and remove the syringe,


keeping the one-way valve attached
to the male luer fitting of the IAB
catheter.

Make a small incision at the exit


of the guidewire.

Insert needle at 45 angle or less,


then insert guidewire (7.5Fr./8Fr.
IAB: 0.025" (0.06cm) / 7Fr. IAB:
0.018" (0.05cm) - sheathless,
0.035" (0.09cm) - sheathed.

For Sheathless Insertion

Insert the vessel dilator over the


guidewire, tapered end first.

For Sheathed Insertion

9a

Carefully spread the


subcutaneous tissue
to facilitate sheathless
insertion.

Remove the IAB catheter


from the T-handle by
pulling STRAIGHT out to
avoid damaging it.

10

Advance the IAB catheter over the guidewire


into the artery in short
1" strokes until correct
placement is achieved.

Advance the sheath seal


as close to the insertion
site as possible and secure
the IAB catheter to the
patients leg.

Pressure monitoring set up

Insert the introducer


dilator into the sheath
hub and secure.

8
1 Advance sheath over
guidewire into artery
using a rotary motion.
2 Withdraw introducer
dilator leaving sheath
in place.

9b

10

11

7Fr. IAB only:


9a Remove 0.035"
(0.09cm) guidewire
and
9b Replace with 0.018"
(0.05cm) guidewire.

Remove the IAB catheter


from the T-handle by
pulling STRAIGHT out
to avoid damaging it.

Advance the IAB catheter in


short 1" strokes through the
sheath into the artery until
correct placement is achieved.
Push the sheath seal into the
hub of the sheath and secure
the IAB catheter to the
patients leg.

Connection to IABP

OR

12

13

14

15

16

17

17

Remove the guidewire and


aspirate 3cc of blood from the
inner lumen.

Perform a manual flush of the


inner lumen with 3-5cc of flush
solution.

Attach a standard arterial fluid


pressure monitoring apparatus.

Twist and pull to remove the


one-way valve from the IAB
catheter.

Connect the IAB catheters male


luer fitting to the female luer
fitting of the catheter extender.

Connect male luer fitting of


catheter extender to Safety Disk
of CS300. Insert fiber-optic
sensor connector into IABPs
sensor input receptacle until
it clicks.

Connect male luer fitting of


catheter extender to Pneumatic
Module of CARDIOSAVE. Insert
fiber-optic sensor connector into
IABPs sensor input receptacle
until it clicks.

MAQUET
1300 MacArthur Blvd.
Mahwah, NJ 07430, USA
Phone: +1 201 995 8700 or

1 800 777 4222

MAQUET Medical Systems USA


45 Barbour Pond Drive
Wayne, NJ 07470, USA
Phone: 1 888 880 2874
http://ca.maquet.com
www.maquet.com

PN: 0002-08-1003 R3 (U.S.) MCV00013298 REVA (Intl.) MCV-PT-80000311-EN-03 MAQUET is a registered trademark of MAQUET GmbH The following are U.S. registered trademarks of Datascope Trademark Corp.:
SENSATION, SENSATION PLUS, CARDIOSAVE CS300 is a trademark of Datascope Trademark Corp. STATLOCK is a registered trademark of C. R. Bard, Inc. Copyright 2013 Datascope Corp. or its affiliates All rights
reserved 01/13 Caution: Federal (U.S.A.) Law restricts this device to sale by or on the order of a physician.
Refer to Instructions for Use for current indications, warnings, contraindications, and precautions.