Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Ophthalmologica 2004;218:5763
DOI: 10.1159/000074568
of Ophthalmology, Johannes Gutenberg University, and b Max Planck Institute for Polymer Research,
Mainz, Germany; c John A. Moran Eye Center, University of Utah Health Sciences Center, Salt Lake City, Utah, USA
Abstract
Introduction: With the recent introduction of small-incision cataract surgery, requirements for intraocular lens
(IOL) flexibility, strength and hydrophilicity have rapidly
evolved. The IOL surface, however, remains a critical factor influencing uveal biocompatibility. Purpose: To objectively quantify factors of material properties of various IOLs using contact angle measurements, differential
scanning calorimetry, dynamic-mechanical measurements and scanning electron microscopy. Material and
Methods: In our study, 17 currently available IOLs were
investigated using contact angle measurements to assess hydrophilicity and biocompatibility, as well as differential scanning calorimetry for the estimation of glass
transition temperature. Mechanical capacity and flexibility were investigated using dynamic-mechanical measurements. Additional analysis of the IOL surface was
performed using scanning electron microscopy. Results:
The contact angle measurements of the studied IOLs
revealed similar values within each group. The silicone
IOLs had values between 106 and 119. The PMMA IOLs
ABC
Fax + 41 61 306 12 34
E-Mail karger@karger.ch
www.karger.com
Mana Tehrani, MD
Johannes Gutenberg University
Department of Ophthalmology, Langenbeckstrasse 1
DE55131 Mainz (Germany)
Tel. +49 6131 173130, Fax +49 6131 175566, E-Mail manatehrani@hotmail.com
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
Key Words
Material properties W Intraocular lens W Hydrogel
Cellular deposition on the IOL surface is a good indicator of postoperative inflammatory reaction and biocompatibiliy [46]. The extent of this cellular reaction depends on the material, design and biocompatibility of the
IOL used [7]. There is no doubt that material properties
influence the biological response, and the IOL surface
may be the most important factor influencing biocompatibility, due to its direct interaction with intraocular tissue, proteins and inflammatory mediators.
The purpose of this study is to objectively quantify the
material characteristics of currently available IOLs, using
the methods described below. This information is potentially relevant to the surgeon faced with the task of choosing the most appropriate IOL for a clinical or surgical situation.
Table 1. Characteristics of 17 commercially available IOLs of various manufacturers and contact angle results of the
representative IOLs
IOL type/manufacturer
Chemical composition
IOL design
IOL material
CeeOn 811C/Pharmacia
SC60B-UV/MDR (f)
ACR6D/Corneal (f)
92L/Morcher (f)
Hydriol 28C/PhysIOL (f)
U940A/Mentor Ophthalmics (f)
Easacryl 1/Bausch & Lomb (f)
PS 101 A/Allergan
MA60BM/Alcon (f)
Siflex 4/Bausch & Lomb
Flex 60/Bausch & Lomb
CeeOn 740P/Pharmacia
AR-40/Allergan (f)
SI-40 NB/Allergan (f)
Soflex 2/Bausch & Lomb (f)
C11UB/Bausch & Lomb (f)
Silens 6/Bausch & Lomb (f)
HSM, PMMA
HEMA
HEMA (26% water)
MMA, HEMA (28% water)
HEMA
MMA, HEMA, EGDMA
HEMA (26% water)
PMMA
PEA, PEMA
PMMA
PMMA
PMMA
EA, EMA, TFEMA
PDMDPS
PDMS
PDMS
PDMS
one-piece
one-piece
one-piece
three-piece
one-piece
three-piece
one-piece
one-piece
three-piece
one-piece
one-piece
one-piece
three-piece
three-piece
three-piece
one-piece
three-piece
HSM, PMMA
HEMA
HEMA
PMMA
HEMA
polypropylene
HEMA
PMMA
PMMA
PMMA
PMMA
PMMA
PMMA
PMMA
PMMA
PDMS
PMMA
Contact angle
degrees
56.5B3.8
59.2B2.5
64.3B1.8
64.6B1.2
64.9B2.4
65.2B3.1
69.1B1.7
73.2B1.4
73.3B2.4
75.4B1.2
75.7B2.9
77.1B4.2
81.7B1.3
106.2B1.4
112.9B1.7
115.4B1.3
199.0B3
Ophthalmologica 2004;218:5763
Tehrani/Dick/Wolters/Pakula/Wolf
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
58
Results
Ophthalmologica 2004;218:5763
59
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
4, C11UB and Flex 60 (Bausch & Lomb), CeeOn 740P and CeeOn
811C (Pharmacia), Memorylens U940A (Mentor Ophthalmics), 92L
(Morcher), Hydriol 28C (PhysIOL), ACR6D (Corneal) and the
SC60B-UV (Medical Development Research).
Fig. 1. Graphic representation of the results of dynamic-mechanical measurements. The IOLs are ordered according
to their mechanical capacity between rubber and glass. log E) = Elasticity modulus, which is defined as the relationship of elastic pressure in phase to the applied force and is dependent on the ability of a substance to store energy
elastically [1]; log E)) = viscosity modulus, which describes the viscous part of a viscoelastic system; the higher the
viscosity, the higher the value of the viscosity modulus [1]; Tg = glass transition temperature.
60
Ophthalmologica 2004;218:5763
Discussion
Biocompatibility is defined as the capability of a prosthesis implanted in the body to exist in harmony with tissue without causing deleterious changes [911]. Further,
uveal and capsular biocompatibility have been differentiated. The foreign-body cellular reaction can be seen as
the most important parameter for uveal biocompatibility.
Thus, IOLs causing relatively less foreign-body reaction
can be described as uveally biocompatible. The main
determinants of capsular biocompatibility are lens epithelium cell outgrowth, anterior and posterior capsular opacification and capsular contraction [9, 10].
Tehrani/Dick/Wolters/Pakula/Wolf
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
Ophthalmologica 2004;218:5763
61
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
Fig. 2. Scanning electron image of an onepiece PMMA IOL (PS 101 A/Allergan). The
image from the side view revealed fibrin
accumulation in height. Magnification
!52.
pacity and flexibility. A descriptive analysis of lens surface interaction with fibrin was performed using scanning
electron microscopy.
The contact angle measurements revealed similar angles within the material groups, but differences between
material groups. Minor differences could be seen in the
silicone group with values between 106.2 B 1.4 and 119.0
B 3. Within the PMMA/acrylic group, the values again
had minor differences: PS 101 A, AcrySof MA60BM,
Siflex 4 and the Flex 60 had similar values (73.2, 73.3,
75.4 and 75.7). The AR-40 from Allergan had a higher
value, 81.7. In the hydrophilic group, the values ranged
between 59.2 B 2.5 (SC60B-UV/MDR) and 69.1 B 1.7
(Easacryl 1/Bausch & Lomb). The lowest value of any lens
was seen in the heparin-surface-modified IOL (56.5 B
3.8).
Contact angle readings in particular assess surface
hydrophilicity, which can be used as a reliable parameter
for predicting uveal and capsular biocompatibility [12].
The heparin-surface-modified IOL revealed the lowest
contact angle values. Tognetto and Ravalico [13] described less anterior segment inflammation in terms of
reduction of postoperative inflammatory cell response,
fibrinoid reaction and synechia formation in eyes which
had received heparin-surface-modified PMMA lenses
compared to those with normal PMMA lenses. Miyake et
al. [14] found that IOLs with hydrophobic surfaces induced higher postoperative inflammation than lenses
with hydrophilic surfaces. Amon et al. [9] have described
a significantly lower incidence of foreign-body reactions
on hydrophilic than on hydrophobic IOL surfaces and
conclude that the best approach to improve biocompatibility of IOL material is to reduce protein adhesion, e.g.
with hydrophilic IOLs. Especially for eyes with preoperative risk factors for foreign-body reaction, IOLs with
hydrophilic surfaces should be recommended [9].
Hollick et al. [15] described a reduced inflammatory
cell reaction in hydrophilic lenses compared with PMMA
and silicone IOLs but with many more lens epithelial cells
on the anterior surface. In contrast, Miyake et al. [14]
found a correlation between the hydrophilicity of an IOL
and the severity of postoperative inflammation (alteration of the blood aqueous barrier) and the speed of anterior capsule opacification. Hydrophobic lenses induced
greater postoperative inflammation and more rapid anterior capsule opacification [16].
Overall, these findings suggest that the contact angle of
a material, as a value for hydrophilicity, is a reliable basic
parameter for predicting uveal and capsular biocompatibility. Together with clinical observations, this informa-
62
Ophthalmologica 2004;218:5763
Perhaps equally as important, however, in postoperative cellular reactions is IOL design [17]. Does IOL material play only a minor role in posterior chamber opacification while the design of the IOL is the major determinant?
In fact, several studies have established that a sharp optic
edge design is associated with significantly reduced posterior chamber opacification [18, 19].
In summary, there are many parameters influencing
the biocompatibility of IOls. IOL material, and in particular its surface properties [20], as well as its design/shape
are important in determining the clinical behavior of the
lens in vivo. Taken together with known preoperative risk
factors, such as glaucoma, diabetes mellitus, pseudoexfoliation or uveitis [21], proper IOL selection may result in
superior postoperative results. This has been borne out by
a reduced inflammatory response for routine glaucoma,
uveitis and diabetes patients with the heparin-surfacemodified IOLs, as compared with non-heparin-surfacemodified lens implants.
In the future, IOL materials and designs may be routinely tested in vitro to understand their interactions with
cells, proteins and tissues, thus streamlining the process of
product development and making surgery safer and more
predictable for patients [1, 12, 20].
Conclusion
Tehrani/Dick/Wolters/Pakula/Wolf
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM
tion can guide the choice of IOL based on the clinical situation.
Differential scanning calorimetry, which measures
glass transition temperature, gives us additional information about the IOLs. The classic PMMA lenses have a
glass transition temperature of approximately 110 C so
that at room temperature this material is in a brittle, glass
status. Our analysis revealed mean glass transition temperatures of 113.5 C (Flex 60 from Bausch & Lomb) and
118.8 C (CeeOn 811C from Pharmacia, surface modified) for modern PMMA lenses.
Glass transition temperatures are especially important
for foldable IOLs. The hydrogel IOLs are as rigid as Plexiglas when dehydrated and are related to PMMA lenses, as
they are acrylate/methacrylate polymers. At room temperature, in a dehydrated state, hydrogel lenses are rigid
and fragile. These lenses become soft and elastic when
hydrated. The measurements had to be taken in the dehydrated state, so these values likely differ from that of the
hydrated state.
The acrylic lenses had a mean glass transition temperature of 14.0 C. Thus, by warming, a faster unfolding of
the lens can be achieved. Alternatively, implantation at
room temperature allows slower unfolding within the eye.
The silicone lenses revealed rubber-like characteristics at
room temperature. This correlates with clinical intraoperative experiences of very rapid unfolding within the eye.
The dynamic-mechanical measurements revealed that
the PMMA lenses remained in the glass state. The hydrogel lenses were in the half-elastic state. The acrylic lenses
had values between the half-elastic and glass state. They
unfold slowly within a temperature range of 10 C, which
allows a controlled unfolding process. The silicone lenses
demonstrated a high returning resilience which contributes to a powerful unfolding during implantation. The
returning resilience was independent of the temperature.
Scanning electron microscopy confirmed that fibrin, a
non-water-soluble protein, withdraws from hydrophilic
surfaces and adheres to hydrophobic surfaces. In the
hydrophobic PMMA and acrylic groups, an intense adhesion could be observed, and in the silicone IOLs, the
fibrin was spread widely. In the SI-40 NB and the C11UB,
the fibrin was peeled off from the surface.
Biological interactions are influenced by material surface properties. Property studies can yield information for
a better understanding of material interactions [12].
Our findings are results from measurements in an artificial environment but correlate with empirical experience and provide new, comparative information for the
different IOLs investigated.
References
Ophthalmologica 2004;218:5763
63
Downloaded by:
Siriraj Medical Library, Mahidol University
202.28.191.34 - 2/17/2015 12:14:13 PM