Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
How long does it take to reach sterilization temperature (250 0F)? Does this
differ under a load?
Are there any cold or hot spots within the autoclave?
Is it able to hold the target temperature with minimal temperature variations?
Does the internal temperature coincide with the temperature on the display,
gauge or dial?
Is there adequate pressure? Is there too much pressure?
Is the timer accurate? Does the timer start at the beginning of the cycle or
upon reaching the target temperature?
Is the autoclave able to hold a sufficient amount of water to complete the
cycle?
What cycle time is required to achieve adequate sterilization of a typical
load?
Copyright BME Labstore
Figure 2 is a graph showing the temperature throughout a sterilization cycle. From this
graph, it is possible to determine the following performance characteristics:
a. Come-up time for the cycle was approximately 20 minutes.
b. There does not appear to be any hot or cold spots as indicated by the consistency of
all of the probes.
c. The autoclave was able to hold the target temperature with minimum fluctuations in
temperature throughout the cycle.
d. The autoclaves timer, temperature, and pressure gauges can be validated against the
certified data logger data.
Copyright BME Labstore
Figure 3 is a graph showing the pressure throughout a sterilization cycle. From this graph, it
is possible to determine the following performance characteristics:
a. The autoclave was able to reach and hold an acceptable pressure level.
b. There was more initial pressure variation in the back of the autoclave than in the
front. This variation was caused by the probe being closer to the vent.
c. The vent appears to be working properly. Air and steam are able to be exhausted at
the beginning, throughout and end of the cycle.
Figure 4 is a graph showing the F values or heat equivalency throughout a sterilization cycle.
From this graph, it is possible to determine the following performance characteristics:
a. An F value of 8 (heat equivalent of 8 minutes at 250 0F) was reached after about 10
minutes. An F value of 8 is normally adequate to sterilize glassware and stainless
utensils. This information can be helpful in determining the required cycle time for
adequate sterilization.
b. The F values will vary greatly depending on the size and content of the load. A
validation test with a typical load provides more accurate data for determining
adequate sterilization time and temperatures.
BME Labstore provides lab equipment sales and service throughout Minnesota and western Wisconsin. For
questions or comments about this article, you may contact Tom Holm @ 651-646-5339 or
tom.holm@bmelabstore.com.
NELAC 2003
QUALITY SYSTEMS APPENDIX D
ESSENTIAL QUALITY CONTROL REQUIREMENTS
D.3
MICROBIOLOGYTESTING
D.3.8 Constant and Consistent Test Conditions
a)
LaboratoryFacilities
Floors and work surfaces shall be non-absorbent and easy to clean and disinfect.
Work surfaces shall be adequately sealed. Laboratories shall provide sufficient storage
space, and shall be clean and free from dust accumulation. Plants, food, and drink shall be
prohibited from the laboratory work area.
b)
LaboratoryEquipment
1)
TemperatureMeasuringDevices
Autoclaves