IRIDEX
CORPORATION
Manufacturing
Procedure
Rev.:
MIP60502
Page #:
Page 14 of 16
Title:
7.0 CALIBRATION:
7.1
Preparation:
7.1.1 Turn off the calibration system (FTS60665) key. Place assembly 60502 on the IQ 810
calibration fixture with the fiber port to the left and the PC board facing front. Attach
the ribbon cable from the fixture to J24 on 60502. Place a calibration jumper (60660)
on J23. Attach an EndoProbe to the fiber port and focus the probe onto an AN/2 power
meter.
7.1.2 Turn on the key switch and when the Splash screen comes on quickly hold down the
top and bottom left buttons and the middle right button to enter the Diagnostic 1 menu.
7.2 Set Clock
7.2.1 Turn the knobs at the bottom to enter password 255 and press Clock Setting
7.2.2 Select 24 using middle knob.
7.2.3 Adjust left knob for correct hour.
7.2.4 Adjust right knob for correct minute.
7.2.5 Press lower left button for correct month. Select (+) or (-) with upper right button as
needed to increment or decrement the value.
7.2.6 Press lower right button for correct day/date. Select (+) or (-) with upper right button
as needed to increment or decrement the value.
7.2.7 Press middle right button for correct year. Select (+) or (-) with upper right button as
needed to increment or decrement the value.
7.2.8 Press Ok to return to the Diagnostics 1 screen.
7.3
Procedure #:
IRIDEX
CORPORATION
Manufacturing
Procedure
Rev.:
MIP60502
Page #:
Page 15 of 16
Title:
7.6
7.7
7.8
7.9
7.10
Procedure #:
IRIDEX
CORPORATION
Manufacturing
Procedure
Rev.:
MIP60502
Page #:
Page 16 of 16
Title:
INSPECTION: QA is to review and confirm the calibration values and Diagnostics 2 information on
FRM990469 per QTP60502. Log per FRM990496.
9.0
TRAINING: NA