Direction 2137747
Revision 0
Operating Documentation
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent,
have notation damage in shipment written on all copies of the freight or express
bill before delivery is accepted or signed for by a General Electric representative or
a hospital receiving agent. Whether noted or concealed, damage MUST be reported
to the carrier immediately upon discovery, or in any event, within 14 days after
receipt, and the contents and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an inspection is not requested
within this 14 day period.
Call Traffic and Transportation, Milwaukee, WI (414) 7855052/8*3235052
immediately after damage is found. At this time be ready to supply name of carrier,
delivery date, consignee name, freight or express bill number, item damaged and
extent of damage.
Complete instructions regarding claim procedure are found in Section S of the
Policy & Procedure Bulletins.
3/12/92
GE Medical Systems
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
GE Medical Systems Europe: Telex 261794
Shortlands, Hammersmith, London W6 8BX U.K.
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2128126
D THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
WARNING
AVERTISSEMENT
WARNUNG
AVISO
1 of 2
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2128126
ATENO
AVVERTENZA
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2137747
DATE
PAGE
REV
PAGE
REV
PAGE
REV
PAGE
REV
Title Page . . . . . . . . . . . 0
Damage In Trans. . . . .
Direction 2128126 . . . 0*
A ..................0
B . . . . . . . . . . . . . . blank
11 to 19 . . . . . . . . . . 0
GE Logo . . . . . . . . . . . 0
* This revision number corresponds to the indicated documents revision control system.
PAGE
REV
GE MEDICAL SYSTEMS
Blank
DIRECTION 2137747
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2137747
11 Overview
This document provides the GE Medical Systems usage guidelines and installation instructions for the RJ11 Multiplexer.
The following RJ11 Multiplexers are available:
o 4line, 115VAC/60Hz (46328475P1)
o 8line, 115VAC/60Hz (46328475P2)
o 4line, 220VAC/50Hz (46328475P3)
o 8line, 220VAC/50Hz (46328475P4)
1-1
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2137747
1-2
GE MEDICAL SYSTEMS
REV 0
DIRECTION 2137747
131
132
Switch 1 Setting
The switch settings on the RJ11 Multiplexer can enable the multiplexer to have a fax machine attached to EXTENSION 1. The factory default settings are for no fax connection. Therefore if you are not connecting a fax machine to
the multiplexer (and this is a new installation), skip to Section 133, Multiplexer Connections.
Use the appropriate procedure to set the switches on the RJ11 Multiplexer.
Note
Do not connect a fax machine to the multiplexer on sites in Canada or the United States.
On a 4line multiplexer use the following steps to set switch 1.
1.
Remove the two Phillips screws on the bottom of the multiplexer. Keep these screws for later use.
2.
Turn the multiplexer over, and remove the top cover of the multiplexer case.
3.
If you are connecting a fax machine to EXTENSION 1, push switch 1 (on the S1 switches) down toward the
numeral 1. See Illustration 11.
If you are not connecting a fax machine to the multiplexer, push switch 1 down toward OPEN. See Illustration
11.
4.
Slide on the top cover of the multiplexer case. If the sides do not meet exactly, rotate the top by 180 degrees.
5.
Replace the two Phillips screws on the bottom of the multiplexer. Go to Section 133, Multiplexer Connections.
1-3
GE MEDICAL SYSTEMS
REV 0
132
DIRECTION 2137747
(TOP VIEW)
46328475P1
46328475P3
SWITCH 1
Loosen the two Phillips screws from the switch plate on the bottom of the multiplexer. See Illustration 12.
(BOTTOM VIEW)
46328475P2 AND 46328475P4
SWITCH
PLATE
1-4
GE MEDICAL SYSTEMS
REV 0
132
DIRECTION 2137747
Switch 1 Setting (continued)
2.
3.
If you are connecting a fax machine to EXTENSION 1, push switch 1 toward ON. See Illustration 13.
If you are not connecting a fax machine to the multiplexer, push switch 1 toward the numeral 1. See Illustration
13.
4.
Replace the switch plate and tighten the screws. Go to Section 133, Multiplexer Connections.
SWITCHES
1
ON
2 3 4 5 6 7 8
WITHOUT
FAX
WITH
FAX
1-5
GE MEDICAL SYSTEMS
DIRECTION 2137747
Multiplexer Connections
1.
Place the RJ11 Multiplexer in a secure, dry area (also outside the 10 Gauss line for MR systems). The 8line
units may be rackmounted in the customers phone or utility room using the included brackets.
2.
Connect each modem (and fax machine, if any) to its designated extension. See Illustrations 14 and 15 for
connection guidelines.
Note
Connect no more than a total of two Cath Lab, CT, MR, or PET systems on a single multiplexer, and do
not connect a fax machine to the multiplexer on sites in Canada or the United States.
MULTIPLEXER
5X MR SYSTEM
MODEM FOR
INSITE
EXTENSION 1
EDM MAGNET
MONITORING
MODEM FOR
MAGNET MONITORING
EXTENSION 2
CT SYSTEM
MODEM
EXTENSION 3
XRAY SYSTEM
MODEM
EXTENSION 4
LINE
INCOMING LINE
MULTIPLEXER
FAX MACHINE
EXTENSION 1
5X MR SYSTEM
MODEM FOR
INSITE
EXTENSION 2
EDM MAGNET
MONITORING
MODEM FOR
MAGNET MONITORING
EXTENSION 3
PET SYSTEM
MODEM
EXTENSION 4
LINE
1-6
INCOMING LINE
GE MEDICAL SYSTEMS
REV 0
133
3.
DIRECTION 2137747
Write down the exact configuration on the extension identification sticker (see Illustration 16) that came with the
RJ11 Multiplexer. Print the system ID of each room on the line for the appropriate extension.
Attach this sticker to the top of the multiplexer.
Ext.
1
2
3
4
System
System ID
Connect the incoming phone line to the LINE jack on the multiplexer.
Note
Do not connect anything to the CONTROL jack on the multiplexer.
5.
Verify that the power supply has the correct voltage rating (115VAC/60Hz or 220VAC/50Hz) for the site.
6.
Connect the circular connector of the power supply to the multiplexer, and plug the power supply line cord into a
receptacle that meets the multiplexers power and grounding requirements. Refer to Section 136, Specifications.
1-7
GE MEDICAL SYSTEMS
DIRECTION 2137747
Installation Checkout
Call the OnLine Center (OLC) at 8003217937 to complete the multiplexer installation.
CAUTION
Failure to follow this checkout procedure will cause InSite to be unable to connect to your system,
which will adversely affect customer satisfaction.
Talk to the support group for EACH modality that has a system connected to the RJ11 Multiplexer. Each OLC engineer needs the following information:
o The System ID for each room
o The extension jack to which each system is connected
Note
If a system was not connected previously to InSite, you need to perform the standard InSite check
out procedure.
The OLC engineers will update each phone number database, and then attempt to manually connect to each system
to test that the installation and the database are both functioning properly.
135
Indications
The RJ11 multiplexers have the following indicators available on the front panel to aid in troubleshooting:
INDICATOR
POWER
LINE
EXTENSIONS
CONDITION
ON
FLASHING
ON
ON
MEANING
Power on
Ring on line
Call in program
Extension off hook
1-8
GE MEDICAL SYSTEMS
REV 0
136
DIRECTION 2137747
Specifications
8 line
* Height
2.5 inches
1.75 inches
* Width
8 inches
16.5 inches
* Length (deep)
6.5 inches
7.9 inches
* Weight
2.5 pounds
6 pounds
o Connectors
* The LINE and all EXTENSIONS have RJ11 modular connectors.
* POWER has a 5pin DIN jack to wall mounted supply.
o Telephone Line Interface
* FCC Registration is DNM64S15658VPE
* Ringer Equivalence is 0.3B
* FCC Part 15 Sub B was tested and certifed by CCL
* Canadian Models have Certification CD4, # DOC 1794 4702 A,
Family Approval for CD8 Certificate # 8978,
Load Number; 19
* Insertion Loss is less than 3 dBm.
1-9