Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Heidi Hanes
Document Number:
Equ24-02
17 July 2008
Company:
MU-JHU
Jo Shim
Review date
9-Feb-12
SMILE Comments: This document is provided as an example only. It must be revised to accurately reflect your labs
specific processes and/or specific protocol requirements. Users are directed to countercheck facts when considering
their use in other applications. If you have any questions contact SMILE.
Effective Date:
Jan 15, 2005
Revision #1
Supersedes Rev# 0
Revised/Prepared By
Date
Approved By
Signature
Date
Signature
Date
Annual Review By
(Lab Supervisor/Lab Management)
Jan 2006
Rev# 1
Page 2 of 12
Rev# 1
Page 3 of 12
Incorrect cleaning materials can damage the Integra. Organic solutions damage cuvettes,
tubing, and other plastic materials. Use only 70% propanol solution to clean the equipment.
Never remove a rack from the rack area if the rack status indicators are red.
Never place any objects on top of the instrument
Rev# 1
Page 4 of 12
Filtered -0.22 um- normal human serum is used for ISE activator.
Filtered -0.22um- fetal bovine serum pooled human serum or reconstituted Roche
controls (ie. Preicpath or Precinorm) may also be used as activator if normal human
serum is not available.
4.5.2. Activator container should be cleaned out using DI water and replaced at the end of
each workday with 5 ml of serum (see 4.5.1 above).
4.5.3. Once per week, clean activator bottle out with 10% bleach and then rinse with plenty of
DI water.
4.5.4. Activator is placed in plastic bottles, such as those previously used for ISE Etcher or ISE
Solutions.
4.6. Loading cleaners, diluents, and ISE solutions
4.6.1. Check for any expired or low level reagents.
4.6.2. Load the necessary cleaners and diluents on their correct rack positions according to the
Begin of Day report (BOD)
4.6.3. Insert the racks
Note: After loading or replacing ISE solutions on the ISE rack, the ISE rack replaced
dialog box is displayed. On the dialog box select the ISE solutions that you have
replaced and leave the other boxes blank.
4.7. Loading and Mixing Cassettes
4.7.1. Check for any expired or low level cassettes
4.7.2. Click on the Cassettes tab in the status work area to check what cassettes you need to
load for the daily routine.
4.7.3. Unwrap the cassettes (pierce the aluminum foil on the bottles-as needed), using the tip
of the unlock rack tool.
4.7.4. Place the required cassettes on a cassette rack, and insert the rack into a free slot in the
cassette area.
4.7.5. To mix cassettes: Click the New Messages tab in the Messages work area to check for
messages of the format Cassette <name> on rack <number>, Position <position>
requires mixing.
4.7.6. Click the Cassettes tab in the Status work area and check for cassettes with the Req.
Mixing status.
4.7.7. Wait until the rack status light is green and remove the racks containing the cassettes to
be mixed.
4.7.8. Place the reconstituted cassettes on the external mixer (use the Plate-form shaker) for
10 minutes.
4.7.9. Place the cassettes needing periodical mixing on an external
mixer for one minute.
4.7.10. Re-insert the rack into the cassette area.
4.7.11. After the required mixing time is elapsed, remove the rack again
and put the reconstituted cassettes on the rack, and re-insert the
rack into the cassette area.
4.8. CHECKING RESOURCES:
4.8.1. If you need to take some action to keep the system running
efficiently, the background color and the picture on the Status
button change: for example, if cuvettes are running low, you see
the cuvette icon with a yellow background.
4.8.2. The background color of the Status button tells you whether you
need to take action or not:
I. Red (running man)
Immediate action Required
Jan 2006
Rev# 1
Page 5 of 12
4.8.3. The various status buttons will change color as above when action is needed.
4.8.4. Replacing the Cleaner
a. When the cleaner is nearly finished, the system warns you by displaying the cleaner icon
with a yellow background.
b. The system automatically detects that cleaner has been replaced and resets the icons to
their normal states.
c. If the system completely runs out of cleaner, or if the cleaner temperature exceeds
37.5C, the icon turns red and the system stops.
d. To replace the cleaner bottle
I. Lift the main cover
II. Remove the cap (B) from the existing cleaner bottle and
place the level detection sensor and tubing carefully to
one side.
III. Discard the used cleaner bottle
IV. Remove the cap from the new cleaner bottle and place
the bottle in the slot provided (C).
V. Reinsert the sensor and the tubing. The cap should fit
loosely on the top of the bottle.
VI. Close the main cover.
4.8.5. Refilling the Cuvette reservoir
a. When the cuvette icon turns yallow you must refill the cuvette reservoir before the
system runs out of cuvettes. The number of cuvettes that are still available for use is
displayed in the Analyzer tab in the Status work area.
b. You can perform this task at any time while the system is running.
c. The cuvette reservoir can hold a maximum of 1400 cuvettes. Each pack contains
approximately 1000 cuvettes.
d. Make sure that you do not touch the surface of any cuvettes. Result accuracy may be
affected.
e. To refill the cuvette reservoir
I. Open the cuvette reservoir access panel (A)
II. Remove the cuvette reservoir together with the covering slide.
DO not remove the slide from the reservoir.
III. Turn the reservoir upside down.
IV. Cut open the cuvette pack and, taking care not to touch the
cuvettes, pour the cuvettes into the reservoir.
V. Reinsert the reservoir and the slide back into position.
VI. Remove the slide, taking care not to spill any cuvettes
VII. Replace the slide at the top of the reservoir.
VIII. The system automatically detects that the cuvette reservoir has been refilled.
4.8.6. Refilling the external Water reservoir
a. When the water supply icon turns yellow, you must refill the external water reservoir.
When the icon is red, the system stops.
b. To refill the reservoir
I. Remove the tubing adapter from the reservoir and put it temporarily into a clean
empty plastic bucket.
Jan 2006
Rev# 1
Page 6 of 12
5. CALIBRATION:
5.1. Calibrations are performed at regular intervals, as defined for each analyte/lot#, each reagent
cassette has the calibration intervals information pre-programmed by the manufacturer into its
bar code, and interrupted automatically by the instrument.
5.2. Additional calibrations also have to be performed when certain physical events occur. These
include:
I. A change in cassette
II. A change in cassette reagent lot
III. A change in the defined interval.
IV. Failure of quality control to meet required specifications
V. Major service of equipment
Jan 2006
Rev# 1
Page 7 of 12
Rev# 1
Page 8 of 12
Rev# 1
Page 9 of 12
Rev# 1
Page 10 of 12
Page 11 of 12
Jan 2006
Rev# 1
Page 12 of 12