Sei sulla pagina 1di 4

SERVO-i/s

Disinfection Washer-Disinfector
Wiping and Discarding

Rinse

Refer to the Cleaning and Maintenance Users Manual for


detailed information.

Washer-Disinfector

Whenever possible, clean and


disinfect the equipment immediately
after use. Do not leave foreign
particles, such as blood or saliva, to
dry on the equipment.

85-95C (185-203F)

Drying alternatives
Shake/Tilt

Run the cassette


in a SERVO-i/s with
a testlung

5-7 times

Important!
When Servo Duo Guard filter (or
equivalent) is used, the only cleaning
needed is to wipe the ventilator and
removable parts with a soft lint-free
cloth moistened in soap & water or
detergent-based disinfectant.
Note. In case of more contaminated
surfaces use ethyl alcohol (70%) or
isopropyl alcohol (70%). Avoid
contact with electrical contacts.

If rinsing before disinfection is


not included in the washer
disinfector program:
Rinse the parts thoroughly in
water (<35C/95F).
Let the water flow through
the parts.

Important!
Rinsing the cassette in water
(<35C/95F) immediately after
use may be a useful alternative
to disinfection. Immediate
rinsing can remove particles
and reduces the risk of cross
contamination between
patients.

Caution!
All personnel should be aware of the risk
of parts being infected when disassembling
and cleaning the ventilator.

Position and flow

Note.
The water pressure in the
washer-disinfector should not
exceed 1.5 bar.
The maximum water flow
should not exceed 10 l/min.
The water must have free
passage through the cassette.

Drying cabinet

1 hour
maximum 70C (158F).

10 min

Room air

12-24 hours

If available, connect a 22 mm
hose to expiratory inlet.

Caution!
Never dry the cassette by applying
high pressure air as you may
damage the internal tubing.

0123

Maquet Critical Care AB 2012. All rights reserved. MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014
Printed in Sweden 1201 Rev. 09 English Routine wall cleaning diagram 3.2.0 English

GETINGE GROUP is a leading global provider of products and


systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
specializes in solutions, therapies and products for surgical
interventions and intensive care

SERVO-i/s
Disinfection Manually
Wiping and Discarding

Refer to the Cleaning and Maintenance Users Manual for


detailed information.

Disinfectant

Rinse

Whenever possible, clean and


disinfect the equipment immediately
after use. Do not leave foreign
particles, such as blood or saliva, to
dry on the equipment.

Drying alternatives
Shake/Tilt

Run the cassette


in a SERVO-i/s with
a testlung

Let the parts soak in a disinfectant:


Ethyl- or isopropyl alcohol 70%
Cidex OPA
Hexanios G+R, Aniosyme DD1,
Gigazyme Plus or Anioxyde 1000.

Rinse the parts thoroughly in


water (<35C/95F). Let the
water flow through the parts.

Important!
Follow the disinfectant manufacturers
recommendations and instructions.

5-7 times

Drying cabinet

10 min

Room air

Rinse
Important!

Important!

When Servo Duo Guard filter (or


equivalent) is used, the only cleaning
needed is to wipe the ventilator and
removable parts with a soft lint-free
cloth moistened in soap & water or
detergent-based disinfectant.

Rinsing the cassette in water


(<35C/95F) immediately after
use may be a useful alternative
to disinfection. Immediate
rinsing can remove particles
and reduces the risk of cross
contamination between
patients.

Note. In case of more contaminated


surfaces use ethyl alcohol (70%) or
isopropyl alcohol (70%). Avoid
contact with electrical contacts.

Caution!
All personnel should be aware of the risk
of parts being infected when disassembling
and cleaning the ventilator.

Important!
Rinse the parts thoroughly in
water to remove all traces of
disinfectant. Let the water flow
through the parts. Carefully shake
and tilt the cassette, turn upside
down and repeat, at least 5-7
times.

1 hour
maximum 70C (158F).

12-24 hours

If available, connect a 22 mm
hose to expiratory inlet.

Caution!
Never dry the cassette by applying
high pressure air as you may
damage the internal tubing.

0123

Maquet Critical Care AB 2012. All rights reserved. MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014
Printed in Sweden 1201 Rev. 09 English Routine wall cleaning diagram 3.2.0 English

GETINGE GROUP is a leading global provider of products and


systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
specializes in solutions, therapies and products for surgical
interventions and intensive care

SERVO-i/s
Sterilization procedure
(Note. Autoclaving reduces the lifetime of the expiratory cassette)

Wiping and Discarding

Rinse

Refer to the Cleaning and Maintenance Users Manual for


detailed information.

Autoclave

Whenever possible, clean and


disinfect the equipment immediately
after use. Do not leave foreign
particles, such as blood or saliva, to
dry on the equipment.

Shake/Tilt
4 min 134C (273F)

Rinse the cassette in water


(<35C/95F). Let the water flow
through the parts

5-7 times

Shake/Tilt

Important!
Before placing the expiratory cassette
in an autoclave, make sure that no
water remains inside the cassette.

When Servo Duo Guard filter (or


equivalent) is used, the only cleaning
needed is to wipe the ventilator and
removable parts with a soft lint-free
cloth moistened in soap & water or
detergent-based disinfectant.
Note. In case of more contaminated
surfaces use ethyl alcohol (70%) or
isopropyl alcohol (70%). Avoid
contact with electrical contacts.

5-7 times

Drying cabinet

Room air

Run the cassette


in a SERVO-i/s with
a test lung

15 min 121C (250F)

Dry

Important!

Drying alternatives

Note.
Sterilization is normally not necessary
for the expiratory cassette but, when
applied, use validated processes
only.
Instrument parts are typically
autoclaved at a temperature of 134C

Drying cabinet

10 min

Room air

1 hour
12-24 hours
maximum 70C (158F).
If available, connect a 22 mm
hose to expiratory inlet.

(273F) for 4-7 minutes.


1 hour
12-24 hours
maximum 70C (158F).
If available, connect a 22 mm
hose to expiratory inlet.

Caution!
All personnel should be aware of the risk
of parts being infected when disassembling
and cleaning the ventilator.

Rubber parts are typically autoclaved at


a temperature of 121C (250F) for 15
minutes.

Caution!
Never dry the cassette by applying
high pressure air as you may
damage the internal tubing.

0123

Maquet Critical Care AB 2012. All rights reserved. MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014
Printed in Sweden 1201 Rev. 09 English Routine wall cleaning diagram 3.2.0 English

GETINGE GROUP is a leading global provider of products and


systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
specializes in solutions, therapies and products for surgical
interventions and intensive care

SERVO-i/s

Hazardous
waste
(infectious)

Caution!
All personnel should be aware of the risk of
parts being infected when disassembling and
cleaning the ventilator.

Whenever possible, clean and disinfect the equipment


immediately after use. Do not leave foreign particles,
such as blood or saliva, to dry on the equipment.

Hazardous
waste
(infectious)

Wipe off/discard

When Servo Duo Guard filter (or equivalent) is used, the


only cleaning needed is to wipe the ventilator and
removable parts with a soft lint-free cloth moistened in
soap & water or detergent-based disinfectant.
Note. In case of more contaminated surfaces use ethyl
alcohol (70%) or isopropyl alcohol (70%). Avoid contact
with electrical contacts

Important!

Refer to the Cleaning and Maintenance Users Manual for detailed


information.

Cleaning procedure when


Servo Duo Guard filter is used

0123

Maquet Critical Care AB 2012. All rights reserved. MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014
Printed in Sweden 1201 Rev. 09 English Routine wall cleaning diagram 3.2.0 English

GETINGE GROUP is a leading global provider of products and


systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. We operate under the three
brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh
focuses on patient mobility and wound management solutions.
GETINGE provides solutions for infection control within healthcare
and contamination prevention within life sciences. MAQUET
specializes in solutions, therapies and products for surgical
interventions and intensive care

Potrebbero piacerti anche