Sei sulla pagina 1di 7

163

Clinical and Electrophysiological Difference Between Idiopathic


Right Ventricular Outflow Tract Arrhythmias
and Pulmonary Artery Arrhythmias
YOSHIHIRO YAMASHINA, M.D.,∗ TETSUO YAGI, M.D., Ph.D.,∗ AKIO NAMEKAWA, M.D.,∗
AKIHIKO ISHIDA, M.D.,∗ HIROKAZU SATO, M.D.,∗ TAKASHI NAKAGAWA, M.D.,∗
MANJIROU SAKURAMOTO, M.D.,∗ EIJI SATO, M.D.,∗ and TOMOYUKI YAMBE, M.D., Ph.D.†
From the ∗ Division of Cardiology, Sendai City Hospital, Sendai, Japan; and †Department of Medical Engineering and Cardiology, Institute
of Development, Aging and Cancer, Tohoku University, Sendai, Japan

Characteristics of Pulmonary Artery Arrhythmias. Introduction: The precise incidence and


characteristics of ventricular arrhythmias originating from the pulmonary artery have not been fully
described. The purpose of this prospective study was to clarify these points.
Methods: Thirty-three consecutive patients with an idiopathic left bundle branch block and inferior-axis
deviation type ventricular arrhythmia were included. All patients underwent detailed electroanatomical
mapping (CARTO, Biosense-Webster, Diamond Bar, CA, USA) during sinus rhythm prior to the catheter
ablation. The precise location of the catheter tip at the successful ablation site was confirmed by both
electroanatomical mapping and contrast radiography. The clinical and electrophysiological data were
compared between the right ventricular outflow tract (RVOT) arrhythmia patients (RVOT group) and PA
arrhythmia patients (PA group).
Results: Eight patients (8/33 patients: 24.2%) had their ventricular arrhythmias successfully ablated
within the PA. The local bipolar electrogram at the successful ablation sites in the PA group exhibited
a significantly greater duration (P < 0.05) and lower amplitude (P < 0.05) than did those in the RVOT
group (n = 19). In the PA group, all patients exhibited a multicomponent electrograms composed of a
spiky potential and a dull potential, which might have consisted of near-field PA activation and a far-field
ventricular activation, respectively, at the successful ablation site. Direct ablation to the spiky electrogram
was able to eliminate the arrhythmias in all the PA group patients.
Conclusions: PA arrhythmias may be more common than previously recognized. Careful mapping and in-
terpretation of low amplitude and multicomponent electrograms are important for recognizing ventricular
arrhythmias originating from the PA. (J Cardiovasc Electrophysiol, Vol. 21, pp. 163-169, February 2010)

right ventricular outflow tract, pulmonary artery, ablation, ventricular tachycardia, premature ventricular
complex

Introduction was undertaken to clarify these points prospectively in the


consecutive patients.
Recently, a small number of cases with idiopathic ven-
tricular tachycardia (VT)/premature ventricular complexes
(PVCs) have been reported to have originated from the pul- Methods
monary artery (PA).1-5 However, most of the previous reports Study Population
were retrospective studies in selected cases and verified the
anatomical locations of the successful ablation sites only The study population consisted of 33 consecutive patients
by means of two-dimensional right ventriculography or pul- (mean age 45.5 ± 17.3 years, 12 males) with symptomatic
monary arteriography without any real-time monitoring of VT/PVCs, left bundle branch block and an inferior-axis de-
the actual catheter position. Therefore, the precise incidence, viation in whom radiofrequency catheter ablation (RFCA)
exact location of the successful ablation site, and charac- was attempted between 2007 and 2008 at our institute. No
teristics of the PA-VT/PVCs remain unknown. This study patients had any structural heart disease. All patients pro-
vided their written informed consent for the mapping and
ablation. The study was approved by the institutional review
board.
No disclosures were made. Electrophysiological Evaluation, Mapping, and Ablation
Address for correspondence: Tetsuo Yagi, M.D., Ph.D., Division of Car- All patients were studied in the absence of any antiar-
diology, Sendai City Hospital, Wakabayashi-ku, Shimizukouji3–1, Sendai, rhythmic drug therapy. Prior to the conventional electro-
Japan. Fax: 81-022-211-8972; E-mail: tetsuo.yagi@nifty.com
physiological evaluation and ablation, all patients underwent
Manuscript received 30 May 2009; Revised manuscript received 30 July
detailed electroanatomical voltage mapping focused on the
2009; Accepted for publication 3 August 2009. right ventricular outflow tract (RVOT) and PA with a CARTO
system (Biosense-Webster, Diamond Bar, CA, USA) during
doi: 10.1111/j.1540-8167.2009.01601.x sinus rhythm (SR), as previously reported.6-10 In brief, a 7-F
164 Journal of Cardiovascular Electrophysiology Vol. 21, No. 2, February 2010

Navi-Star (Biosense-Webster) catheter with a 4-mm distal Comparison Between RVOT and PA Arrhythmias
tip electrode was introduced into the right ventricle via the
femoral vein under fluoroscopic guidance and was used as the Each patient with successful right-sided ablation in the
mapping/ablation catheter. The catheter was placed at mul- present study population was classified into the 2 groups.
tiple sites on the endocardial surface to record bipolar and The RVOT group consisted of patients with VT/PVCs ablated
unipolar electrograms from the RVOT and PA. The bipolar successfully below the pulmonary valve. The PA group con-
electrogram signals (filtered at 10 to 400 Hz and displayed sisted of patients with VT/PVCs ablated successfully above
at 100 mm/sec on the CARTO system) were analyzed with the pulmonary valve. The patient characteristics, form of
regard to their amplitude, duration, relation to the surface clinical ventricular arrhythmia, characteristics of the 12-lead
QRS, and presence of multiple components. A recording was ECG during the clinical VT/PVCs, and characteristics of
accepted and integrated into the map when the variability in the local electrogram at the successful ablation site during
the cycle length, local activation time stability, and maximum SR and VT/ PVCs were analyzed and compared between the
beat-to-beat difference of the location of the catheter were 2 groups. A quantitative and morphological analysis of the lo-
<2%, <3 ms, and <4 mm, respectively. Those parameters cal electrogram at the successful ablation site was performed
were used to exclude any signals with a low amplitude due to distinguish the characteristics of the local electrograms
to poor endocardial catheter contact. In addition, adequate between the RVOT group and PA group. As a previous re-
catheter contact was confirmed by (1) a concordant catheter port described,11 the electrographic amplitude (in mV) was
tip motion with the cardiac silhouette that was verified by defined as the maximum upward to maximum downward de-
multiplane fluoroscopy using mainly a 30 degree right an- flection on the bipolar electrogram, and the electrographic
terior oblique projection and 50 degree left anterior oblique duration (in ms) was defined as the time from the earli-
projection and (2) by a significant ST segment elevation on est electrical activity to the termination of the electrogram
the unipolar recording. in the bipolar electrogram. The amplification and duration
After obtaining the 3-dimensional geometry of the RVOT measurements were combined to give an amplitude/duration
and PA during SR, activation mapping was performed in ratio to allow for an equal emphasis to be placed on each
the RVOT region and site above the pulmonary valve dur- of those values. The local activation time from the surface
ing spontaneous VT or PVCs. If no clinical VT or PVCs QRS onset of the VT or PVCs was measured in each pa-
appeared spontaneously, ventricular stimulation or an intra- tient. The intracardiac electrograms were analyzed at a paper
venous isoproterenol infusion (1 to 5 μg/min) was admin- speed of 100 mm/sec using a computerized recording system
istered to provoke the arrhythmia. The choice of the abla- (Cardiolab System, Prucka Engineering Inc., Houston, TX,
tion site was based on a conventional activation mapping USA). The bipolar electrograms were filtered at a frequency
method and/or pace mapping method. The activation map of 30–500 Hz.
was created by mapping several points during the ventric- In the present study, a description of the electrogram mor-
ular arrhythmias while using a surface electrocardiogram phology at the successful ablation site was defined as follows.
(ECG) lead as a reference. The activation times were as- A “Single” electrogram was a potential consisting of 3 elec-
signed on the basis of the onset of the bipolar electrograms trical deflections and a total electrogram duration of <50 ms
and displayed as color gradients on a 3-dimensional activa- (Fig. 1A). A “multicomponent” electrogram was a poten-
tion map. A suitable target for ablation was selected based on tial that consisted of over 4 electrical deflections and a total
the earliest endocardial activation times during the arrhyth- electrogram duration of ≥50 ms (Figs. 1B, 2, and 3).
mia and confirmed by pace mapping that exhibited ≥11 of
12 matches between the paced and spontaneous QRS com- Analysis of the Electrograms Above the Pulmonary Valve
plexes. Ablation was performed with a 4-mm tip Navi-Star Several investigators have reported supravalvar near-field
ablation catheter using radiofrequency energy (target tem- potentials as PA potentials separated from far-field ventric-
perature 50◦ C, maximal power 40 W) delivered for up to ular potentials in PA arrhythmias.1,3-5 In the present study,
120 seconds. After the elimination of the arrhythmia by a the electrograms at the successful ablation sites in the PA
successfully delivered radiofrequency application, success- group were analyzed for the following points. If the local
ful ablation sites were recorded on a 3-dimensional elec- electrogram was composed of 2 electrical components, the
troanatomical map during SR. The catheter position at the sequence pattern of the 2 components at the successful ab-
successful ablation sites was evaluated by pulmonary ar- lation site was noted during both SR and VT/PVCs (Figs. 2
teriography or right ventriculography while the catheter and 3), and the duration from the onset of the first compo-
position was monitored in real-time on a 3-dimensional elec- nent to the second component was measured in each case
troanatomical map during SR. The minimum distance be- during SR (Figs. 2 and 3). The difference in the morphol-
tween the tip of the catheter at the successful ablation site ogy of the local electrogram at the successful ablation site
and the attachment of the nearest leaflet of the pulmonary was also noted before and after a successful ablation (Figs. 2
valve was measured. and 3).
Successful ablation was defined as follows: (1) absence
of any spontaneous or induced clinical RVOT arrhythmias, Statistical Analysis
both in the absence and presence of isoproterenol, at the end
of the procedure; (2) no recurrence of any clinical ventric- The continuous variables are expressed as the mean ±
ular arrhythmias during continuous ECG monitoring in the SD. Mann-Whitney’s U tests were used to compare the
hospital for at least 72 hours after the ablation procedure; 2 groups. A frequency analysis was performed with a Fisher’s
and (3) no recurrence of any symptomatic arrhythmias in the exact test. A Pearson’s correlation coefficient was used to de-
absence of any antiarrhythmic drug therapy during at least termine the relationship between the 2 parameters. P value
2 months of follow-up. <0.05 was considered significant.
Yamashina et al. Characteristics of Pulmonary Artery Arrhythmias 165

Figure 1. Representative cases of ventricular tachycardia (VT) successfully ablated within the right ventricular outflow tract (RVOT) below the pulmonary
valve (RVOT group). (A) Data from the successful ablation site in a 33-year-old male. The intracardiac electrogram revealed a “single” electrogram during
both sinus rhythm (SR) and premature ventricular complexes (PVCs) at the success site. The radiogram and electroanatomical voltage map show the
ablation catheter position at the success site. The white arrow indicates the pulmonary valve position. The successful ablation site was located 4 mm below
the pulmonary valve. (B) Data from the successful ablation site in a 53-year-old male. The intracardiac electrogram appeared to be a “multicomponent”
electrogram consisting of 2 components (arrow and arrowhead). The sequence of those components did not change during SR and PVCs. The successful
ablation site was located 6 mm below the pulmonary valve. A = anterior; ABL = ablation catheter; d = distal; F = free-wall side; P = posterior; p =
proximal; S = septum side; RAO = right anterior oblique projection.

Results group (Table 1). Although we compared the 12-lead surface


ECG characteristics of the clinical VT/PVCs between the
Ablation Outcome and Prevalence of PA-VT/PVCs 2 groups, no significant differences were observed in the R-
Among the total 33 patients, 29 patients (29/33: 87.8%) wave amplitude, QRS duration, Q-wave ratio of aVR/aVL,
met the successful ablation criteria. In the successful ablation R/S ratio in leads V1, V2, and V3, and precordial transitional
group patients, 8 patients (8/33 patients: 24.2%) had their zones (data are not shown). On the other hand, there were
VT/PVCs successfully ablated within the PA (PA group). some significant differences in the intracardaic electrograms
Among the remaining 21 patients, 19 patients (19/33 pa- between the RVOT group and PA group. The local bipolar
tients: 57.5%) had their VT/PVCs successfully ablated from electrograms at the successful ablation sites in the PA group
within the RVOT (RVOT group) and 2 (2/33 patients: 6.0%) exhibited a significantly greater duration and lower amplitude
had their PVCs ablated from within the left ventricular during both SR and VT/PVCs than did those in the RVOT
outflow tract (LVOT). In the unsuccessful ablation group group (Table 2). The mean amplitude/duration ratio was also
(4/33 patients: 12.1%), 2 patients exhibited recurrences of significantly lower in the PA group than in the RVOT group
their clinical VT/PVCs during the follow-up period. Al- (Table 2). Moreover, there was a higher incidence of mul-
though both of the patients had their VT ablated temporar- ticomponent electrograms at the successful ablation sites in
ily from within the RVOT, they had relapses of their PVCs the PA group than in the RVOT group (Table 2). Although a
after the procedure. In the remaining 2 patients in the un- pacing stimuli from a conventional pacing unit using a max-
successful ablation group, 1 procedure was stopped be- imal output (9.9 V/2.0 ms) could be captured in all patients
cause no arrhythmias could be induced in the laboratory. of the RVOT group, in 3 of 8 patients (37.5%) in the PA
Although another 1 was mapped and ablated from both group it did not capture even with maximal output pacing at
the RVOT side and LVOT side, the targeted PVCs were the successful ablation sites (Table 3). There were no signif-
not affected at all by the RFCA during the procedure. No icant differences in the mean local activation time from the
procedure-related complications were observed in any of the QRS onset during the VT/PVCs between the RVOT group
patients. and PA group. The mean distance of the successful ablation
sites from the pulmonary valve in the RVOT group was 6.2 ±
Comparison of the Clinical, Electrocardiographic and 3.7 mm (range 1–14 mm) below the pulmonary valve. The
Electrophysiological Characteristics Between the RVOT mean distance of the successful ablation sites from the pul-
Group and PA Group monary valve in the PA group was 7.5 ± 4.8 mm (range
4–15 mm) above the pulmonary valve. The distribution of
There were no significant differences in the gender, age, the successful ablation sites for each distance from the pul-
or form of the arrhythmias between the RVOT group and PA monary valve is shown in Figure 4.
166 Journal of Cardiovascular Electrophysiology Vol. 21, No. 2, February 2010

Figure 2. Representative cases of VT/PVCs arising from the pulmonary artery (PA) (PAgroup). (A) Data from the successful ablation site in patient number
4 in Table 3. The intracardiac electrogram exhibited a multicomponent electrogram with a tiny dull potential (arrowhead) and spiky potential (arrow). The
spiky potential was almost buried in the dull potential during SR (duration between each potential = 30 ms), and the sequence of the 2 components became
reversed during the PVC. (A’) The intracardiac electrograms after a successful ablation. The spiky potential became diminished leaving a dull potential. A
radiogram and electroanatomical voltage map showing the ablation catheter position at the success site. The white arrow indicates the pulmonary valve
position. The successful ablation site was located 4 mm above the pulmonary valve. (B) Ablation data from patient number 5 (6 mm above the pulmonary
valve). The asterisk indicates the far-field atrial potential. (B’) After the successful ablation. (C) Ablation data from patient number 6 (8 mm above the
pulmonary valve). (C’) After the successful ablation. (D) Ablation data from patient number 7 (15 mm above the pulmonary valve). This patient had coexisting
complete right bundle branch block during SR (shown in V1), and depolarization of the right ventricle might have been delayed and fused with the dull and
sharp potentials during SR. (D’) After the successful ablation. The abbreviations are as shown in the prior figure. See the text and Table 3 for a detailed
analysis in each patient.

Analysis of the Electrograms in the PA Group group, 7 of 8 patients exhibited tiny dull potentials preceding
the spiky potentials during SR (dull/spiky pattern) (Table 3,
The electrophysiological characteristics of the PA- Fig. 2). In those patients, all exhibited a reversed sequence
VT/PVC cases are summarized in Table 3. Among the 8 (spiky/dull pattern) during the VT/PVCs. Although patient
PA-VT/PVCs cases, 7 had their arrhythmias ablated success- number 8 had a single potential during the late phase of the
fully along the septal side of the PA and 1 on the posterior QRS complex at the successful ablation site during SR, in-
free-wall side. As for morphological characteristics of the creasing the gain revealed a very tiny dull potential preceding
local electrograms at the successful ablation site in the PA the spiky potential (a detailed analysis is shown in Fig. 4).
Yamashina et al. Characteristics of Pulmonary Artery Arrhythmias 167

Figure 3. Ablation data from patient number 8. (A) An intracardiac electrogram from a failed ablation site above the pulmonary valve. This site corresponds
to a failed site [ABL(A)] on the electroanatomical voltage map. At that site, ABL-d exhibited a dull (arrowhead)/spiky (arrow) potential during SR and had
a reversed sequence during the PVC. (B) Intracardiac electrograms at the successful ablation site (8 mm above the site in panel A). During SR, the dull
potential is nearly gone, and the spiky potential is 10 ms later from the QRS than the spiky potential in panel A. An increased gain (square) revealed a dull
potential preceding the spiky potential. The radiogram shows the successful ablation site. The white arrow indicates the pulmonary valve position. LAO =
left anterior oblique projection. The other abbreviations are as shown in the prior figures.

The duration from the onset of the dull potential to the spiky 16.2% in idiopathic ventricular arrhythmias that were treated
potential during SR correlated with the distance between the by RFCA.2,3 However, most previous reports were retro-
ablation site within the PA and base of the pulmonary valve spective studies in selected cases and evaluated the rela-
(r = 0.77, P < 0.05). After the successful ablation within the tionship between the pulmonary valve position and success-
PA, a spiky potential during the late phase of the QRS in SR ful ablation sites only by means of contrast enhancement
was still recorded in 3 (37.5%) patients but disappeared in without real-time monitoring of the catheter position. This
the remaining 5 (62.5%) (Fig. 2). prospective study using both an electroanatomical mapping
system and contrast enhancement demonstrated that approxi-
Discussion mately 25% of VT/PVCs were ablated successfully above the

The Prevalence of PA-VT/PVCs


This prospective study demonstrated that the prevalence TABLE 2
of PA-VT/PVCs might be more common than previously Comparison of the Local Bipolar Electrograms at the Successful Ablation
recognized. Although it is well known that most idiopathic Sites Between the Groups
VT or PVCs originate from the RVOT just below the pul-
RVOT PA P
monary valve,12-15 a small number of cases of idiopathic Variables Group Group Value
VT/PVCs have been recently reported to have originated
from the PA.1-5 Several previous reports explained that PA- Amplitude (mV)
During SR 1.31 ± 0.50 0.64 ± 0.35 <0.01
VT/PVCs were rare entities and their prevalence was 4 ∼ During VT/PVC 1.42 ± 0.88 0.63 ± 0.39 <0.05
Duration (ms)
During SR 37 ± 7 71 ± 10 <0.01
During VT/PVC 49 ± 15 75 ± 19 <0.05
TABLE 1 Amplitude/Duration Ratio (mV/ms)
Comparison of the Clinical Characteristics Between the Groups During SR 0.029 ± 0.012 0.0085 ± 0.0036 <0.01
During VT/PVC 0.032 ± 0.021 0.0082 ± 0.0048 <0.01
Variables RVOT Group PA Group P Value Morphology During SR
Single electrogram 16 (84%) 1 (12.5%)
Age (year) 45.8 ± 15.3 44.6 ± 21.5 NS Multicomponent electrogram 3 (15%) 7 (87.5%) <0.01
Gender (M/F) 6/13 4/4 NS Morphology During VT/PVC
VT/PVC 9/10 3/5 NS Single electrogram 14 (73%) 0 (0%)
Multicomponent electrogram 5 (26%) 8 (100%) <0.01
PA = pulmonary artery; PVC = premature ventricular contraction;
RVOT = right ventricular outflow tract; VT = ventricular tachycardia. SR = sinus rhythm. The other abbreviations are as shown in Table 1.
168 Journal of Cardiovascular Electrophysiology Vol. 21, No. 2, February 2010

TABLE 3
Electrophysiological Characteristics of the Successful Ablation Sites in the PA Arrhythmia Cases
Maximal Output Component Pattern
Pacing Component Distance
Patient Location (9.9V/2.0 ms) SR VT/PVC Interval (ms) AT (ms) (mm)

1 PS Failure Dull/spiky Spiky/dull 36 −19 4


2 AS Captured Dull/spiky Spiky/dull 33 −28 4
3 MS Captured Dull/spiky Spiky/dull 34 −20 4
4 PS Failure Dull/spiky Spiky/dull 30 −39 4
5 PS Captured Dull/spiky Spiky/dull 40 −32 6
6 PF Captured Dull/spiky Spiky/dull 56 −40 8
7 AS Failure Dull/spiky Spiky/dull 45 −45 15
8 PS Captured (Dull)/spiky Spiky/dull 61 −42 15
AS = anterior-septal side; AT = activation time from the QRS onset of the VT or PVC; Component Pattern = Pattern of the sequence between the far-field
potential (dull potential) and near-field potential (spiky potential). Examples are shown in Figures 2 and 3. Component Interval = intercomponent time from
the onset of the dull potential to spiky potential during SR. Distance = distance from the pulmonary valve. MS = midseptal side; PF = posterior-free wall
side; PS = posterior-septal side. The other abbreviations are as shown in the prior tables.

pulmonary valve in consecutive patients who were treated It has been previously demonstrated that in animals dur-
with catheter ablation. ing the embryogenesis of the heart the distal part of the out-
Commonly, an ablation catheter is inserted into the RVOT flow tract loses its myocardial phenotype and becomes the
region antegradely. Therefore, the ablation effect from under proximal part of the ascending aorta and pulmonary trunk.16
the pulmonary valve might affect the origin above the pul- In contrast to the myocardium of the atrial and ventricular
monary valve in some cases. It is suggested that some PA ar- chambers, this myocardium retains its embryonic features
rhythmias have been inadvertently treated as RVOT arrhyth- (i.e., slow propagation of the depolarizing impulse owing to
mias in previous studies. We believe that detailed PA mapping the poor intercellular coupling of the cardiac muscle cells).
should be considered when high-power energy or repeated Myocardial regression in the outflow tract continues until af-
RFCA applications are needed to eliminate VT/PVCs below ter birth, as revealed by the disappearance of the myocardial
the pulmonary valve. In the case of the PA mapping, careful cuff surrounding the semilunar sinuses.16 Incomplete my-
mapping must be performed because even high-output pac- ocardial regression might be the reason for the presence of
ing might not capture the optimal ablation site, which would a myocardial sleeve in the main stem of the PA and the aor-
lead to difficulty in performing pace mapping. tic sinuses, which were connected to the rest of the related
ventricles. Recently, an anatomic study has shown that ven-
tricular myocardial extensions extend into the PA in humans.
The study demonstrated that ventricular myocardial exten-
sions into the PA were relatively common, with a prevalence
of 17% in patients without idiopathic VT/PVCs.17 These
anatomical studies suggested that the myocardial extensions
above the pulmonary valve might play an important role as
arrhythmogenic origins for PA-VT/PVCs.

Characteristics of the PA-VT/PVCs


In the present study, there were no significant differences
between the RVOT group and PA group in terms of the age,
sex, form of arrhythmia, and ECG characteristics. One previ-
ous report demonstrated that PA arrhythmias were well rec-
ognized among young patients.3 However, Hasdemir et al.
revealed that no age or gender differences were observed in
the prevalence of ventricular myocardial extensions, which
might be an arrhythmogenic source of PA-VT/PVCs, into the
pulmonary artery.17 As for the 12-lead ECG characteristics,
1 previous report demonstrated that the R-wave amplitude in
the inferior ECG leads, aVL/aVR ratio of the Q-wave am-
plitude, and R/S ratio in lead V2 were statistically larger in
PA-VT/PVCs than RVOT-VT/PVCs.2 However, other previ-
ous reports demonstrated no specific ECG findings for PA-
VT/PVCs.1,3 It might be difficult to discriminate between
RVOT-VT/‘PVCs and PA-VT/PVCs using those ECG pa-
rameters because the RVOT and PA are a contiguous three-
dimensional tube-like structure.
Figure 4. Distribution of the successful ablation sites. PV = pulmonary On the other hand, some significant characteristics were
valve. The black dots indicate the ablation sites. observed in the local electrogram at the successful ablation
Yamashina et al. Characteristics of Pulmonary Artery Arrhythmias 169

sites in the PA group. The local ventricular bipolar elec- creased success rate of the RFCA and will help avoid
trograms recorded at the successful ablation sites showed complications.
a significantly lower amplitude and greater duration in the
PA group than those in the RVOT group. Most of the local References
electrograms at the successful ablation sites in the RVOT
group exhibited a single component right ventricle electro- 1. Timmermans C, Rodriguez LM, Crijns HJ, Moorman AF, Wellens HJ:
Idiopathic left bundle-branch block-shaped ventricular tachycardia may
gram (Fig. 1A). All patients in the PA group exhibited a originate above the pulmonary valve. Circulation 2003;108:1960-1967.
multicomponent electrogram composed of a dull potential 2. Sekiguchi Y, Aonuma K, Takahashi A, Yamauchi Y, Hachiya H,
and spiky potential above the pulmonary valve (Figs. 2 and Yokoyama Y, Iesaka Y, Isobe M: Electrocardiographic and electro-
3). These multicomponent electrograms in the PA group had physiologic characteristics of ventricular tachycardia originating within
the pulmonary artery. J Am Coll Cardiol 2005;45:887-895.
unique characteristics: (1) a spiky potential was seen after 3. Tada H, Tadokoro K, Miyaji K, Ito S, Kurosaki K, Kaseno K, Naito
a dull potential during SR (dull/spiky pattern), and the se- S, Nogami A, Oshima S, Taniguchi K: Idiopathic ventricular arrhyth-
quence was reversed during VT/PVCs (spiky/dull pattern); mias arising from the pulmonary artery: Prevalence, characteristics, and
(2) the spiky potential became diminished leaving a dull po- topography of the arrhythmia origin. Heart Rhythm 2008;5:419-426.
tential after a successful ablation above the pulmonary valve 4. Whitwam W, Rule S, Narayan SM: Noncontact mapping of small pul-
monary artery potentials preceding ectopy from the right ventricular
(Fig. 2); and (3) when the successful ablation sites were lo- outflow tract. Heart Rhythm 2007;4:959-963.
cated at more distant sites from the pulmonary valve, the 5. Srivathsan KS, Bunch TJ, Asirvatham SJ, Edwards WD, Friedman PA,
2 components were more separated (Table 3, Figs. 2 and Munger TM, Hammill SC, Cha YM, Brady PA, Jahangir A, Bradley
3). These findings indicate that a spiky potential recorded DJ, Rea RJ, Packer DL, Shen WK: Mechanisms and utility of dis-
above the pulmonary valve reflects the near-field activation crete great arterial potentials in the ablation of outflow tract ventricular
arrhythmias. Circ Arrhythmia Electrophysiol 2008;1:30-38.
of the myocardial extension into the PA, and that the dull 6. Verma A, Kilicaslan F, Schweikert RA, Tomassoni G, Rossillo A,
potential reflects the far-field ventricular activation. Several Marrouche NF, Ozduran V, Wazni OM, Elayi SC, Saenz LC, Minor S,
previous studies also demonstrated that tiny multicompo- Cummings JE, Burkhardt JD, Hao S, Beheiry S, Tchou PJ, Natale A:
nent or discrete potentials recorded above the pulmonary Short- and long-term success of substrate-based mapping and ablation
of ventricular tachycardia in arrhythmogenic right ventricular dysplasia.
valve have consisted of PA potentials and far-field RV po- Circulation 2005;111:3209-3216.
tentials.1,3-5 They mentioned that mapping and direct ab- 7. Corrado D, Basso C, Leoni L, Tokajuk B, Bauce B, Frigo G, Tarantini
lation of the PA potential were useful to eliminate the PA G, Napodano M, Turrini P, Ramondo A, Daliento L, Nava A, Buja G,
arrhythmias. Iliceto S, Thiene G: Three-dimensional electroanatomic voltage map-
ping increases accuracy of diagnosing arrhythmogenic right ventricular
Distribution of the Successful Ablation Sites of the cardiomyopathy/dysplasia. Circulation 2005;111:3042-3050.
8. Corrado D, Basso C, Leoni L, Tokajuk B, Turrini P, Bauce B, Migliore
RVOT-VT/PVCs and PA-VT/PVCs F, Pavei A, Tarantini G, Napodano M, Ramondo A, Buja G, Iliceto
S, Thiene G: Three-dimensional electroanatomical voltage mapping
In the present study, most of the successful ablation sites and histologic evaluation of myocardial substrate in right ventricular
of the RVOT-VT/PVCs were located within 6 mm below the outflow tract tachycardia. J Am Coll Cardiol 2008;51:731-739.
pulmonary valve (73%), and most of the successful ablation 9. Boulos M, Lashevsky I, Reisner S, Gepstein L: Electroanatomic map-
sites of the PA-VT/PVCs were located within 6 mm above the ping of arrhythmogenic right ventricular dysplasia. J Am Coll Cardiol
2001;38:2020-2027.
pulmonary valve (62.5%). We speculated that this distribu- 10. Boulos M, Lashevsky I, Gepstein L: Usefulness of electroanatomical
tion pattern of the successful ablation sites suggested that the mapping to differentiate between right ventricular outflow tract tachy-
possible origin of idiopathic right-sided outflow tract arrhyth- cardia and arrhythmogenic right ventricular dysplasia. Am J Cardiol
mias, including both RVOT arrhythmias and PA arrhyth- 2005;95:935-940.
mias, often arise from active tissue around the pulmonary 11. Cassidy DM, Vassallo JA, Marchlinski FE, Buxton AE, Untereker WJ,
Josephson ME: Endocardial mapping in humans in sinus rhythm with
valve. normal left ventricles: Activation patterns and characteristics of elec-
trograms. Circulation 1984;70:37-42.
Study Limitations 12. Kamakura S, Shimizu W, Matsuo K, Taguchi A, Suyama K, Kurita
T, Aihara N, Ohe T, Shimomura K: Localization of optimal ablation
The study population in the present study was small. Fur- site of idiopathic ventricular tachycardia from right and left ventricular
ther evaluation is needed in a larger study population. In outflow tract by body surface ECG. Circulation 1998;98:1525-1533.
addition, all patients in the present study were Japanese. We 13. Lehman BB, Stein KM, Markowitz SM, Mittal S, Iwai S: Ventricular
could not analyze the relationship between the prevalence of tachycardia in patients with structurally normal hearts. In: Zipes DP,
Jalife J, eds.: Cardiac Electrophysiology: From Cell to Bedside, 4th ed.
PA-VT/PVCs and genetic/ethnic differences. However, all Philadelphia, PA: Saunders, 2004, pp. 668-682.
patients in the present study had no familial history of car- 14. Movsowitz C, Schwartzman D, Callans DJ, Preminger M, Zado E,
diac sudden death, inherited cardiac diseases, or ventricular Gottlieb CD, Marchlinski FE: Idiopathic right ventricular outflow tract
arrhythmias. tachycardia: Narrowing the anatomic location for successful ablation.
Am Heart J 1996;131:930-936.
15. Josephson ME: Catheter and Surgical Ablation in the Therapy of Ar-
Conclusions rhythmias. Clinical Cardiac Electrophysiology. Techniques and Inter-
pretations, 3rd ed. Philadelphia, PA: Lippincott & Wilkins, 2002.
PA arrhythmias may be more common than previ- 16. Moorman AF, Christoffels VM: Cardiac chamber formation: Develop-
ously recognized. Careful mapping and interpretation of ment, genes, and evolution. Physiol Rev 2003;83:1223-1267.
17. Hasdemir C, Aktas S, Govsa F, Aktas EO, Kocak A, Bozkaya YT,
low voltage and multicomponent potentials recorded above Demirbas MI: Demonstration of ventricular myocardial extensions into
the pulmonary valve are important so as not to over- the pulmonary artery and aorta beyond the ventriculo-arterial junction.
look any PA-VT/PVCs. It will contribute to the in- Pacing Clin Electrophysiol 2007;30:534-539.

Potrebbero piacerti anche