Instrument
Written By
: Eby Joy
Quality Assurance
Checked By : Hameed Najumudeen
Management Representative
Approved By : Deepak Rathore
General Manager
Sign :
Date
Sign :
Date :
Sign :
Date :
No: CI/01/RCDT1
Revision :0
Distributed to :LM,LAB
Original : QA File
Rev. 0 Issued Date:
27/01/12
Effective from :
30/01/12
Description
RCD testers are designed to simulate a range of fault currents, with restrictions on the
duration of the fault current, and to time the operation of the device. This will indicate
the ability of the RCD to interrupt a particular fault current within certain time limits to
ensure protection against fire, damage and electrocution.
It is important to understand that an RCD tester does not generate the current, but acts
as a resistor allowing current to flow from the live to earth, simulating a fault.
Purpose
To provide a written Calibration Procedure for RCD Testers to minimize the risk of error
and to maintain the accuracy limit.
Referance Standard
Electrical Test Calibrator Model: Transmille 3200HGA
Treacebility
NPL
Environmental Condition
Temperature: (22 5)C
Relative Humidity: (50 10)% RH
Procedure
1.
2.
3.
4.
Calibration Procedure
Instrument
Written By
Sign :
Date :
No: CI/01/RCDT1
Revision :0
Distributed to :LM,LAB
Original : QA File
Sign :
Date :
Sign :
Date :
: Eby Joy
Quality Assurance
13. If the instrument is not within the tolerance limit, make the adjustments if possible
and re-perform the test.
14. Prepare Documentation.
Acceptable Tolerance
Refer the appropriate manual or instrument data sheet.
Documentation
1.
2.
3.
4.
Document
Document
Document
Document
in
in
in
in
the
the
the
the
work sheet.
central computerized software.
certificate ( if calibration pass )
Faulty report ( if calibration fail )