1. Scope
1.1 This test method covers the determination of 1,2dibromoethane (commonly referred to as ethylene dibromide
or EDB) and 1,2-dibromo-3-chloropropane (commonly referred to as DBCP) in water at a minimum detection level of
0.010 g/L by liquid-liquid extraction combined with gasliquid chromatography. This test method is applicable to the
analysis of drinking waters and groundwaters. It is not recommended for wastewaters, due to the potential for interferences
from high concentrations of other extractable organics. Similar
information can be found in EPA Method 504.
1.2 This test method was used successfully with reagent
water and groundwater. It is the users responsibility to ensure
the validity of this test method for waters of untested matrices.
1.3 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. For specific hazard
statements, see Sections 6 and 9.
2. Referenced Documents
2.1 ASTM Standards: 2
D 1066 Practice for Sampling Steam
D 1129 Terminology Relating to Water
D 1192 Specification for Equipment for Sampling Water
and Steam in Closed Conduits3
D 1193 Specification for Reagent Water
D 3370 Practices for Sampling Water from Closed Conduits
D 3856 Guide for Good Laboratory Practices in Laboratories Engaged in Sampling and Analysis of Water
D 4210 Practice for Interlaboratory Quality Control Procedures and a Discussion on Reporting Low-Level Data3
1
This test method is under the jurisdiction of ASTM Committee D19 on Water
and is the direct responsibility of Subcommittee D19.06 on Methods for Analysis for
Organic Substances in Water.
Current edition approved June 1, 2004. Published June 2004. Originally
approved in 1992. Last previous edition approved in 1998 as D 531698.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standards Document Summary page on
the ASTM website.
3
Withdrawn.
4
Available from U.S. Environmental Protection Agency, 26 W. Martin Luther
King Ave., Cincinnati, OH 45268.
Copyright ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
D 5316 98 (2004)
TABLE 1 Chromatographic Conditions for 1,2-dibromethane
(EDB) and 1,2-dibromo-3-chloropropane (DBCP)
Analyte
EDB
DBCP
Column A
9.5
17.3
Column C
4.1
12.8
6. Interferences
6.1 Impurities contained in the extracting solvent usually
account for the majority of the analytical problems. Solvent
blanks should be analyzed on each new bottle of solvent before
use. Indirect daily checks on the extracting solvent are obtained
by monitoring the water blanks. Whenever an interference is
noted in the water blank, the analyst should reanalyze the
extracting solvent. Low-level interferences generally can be
removed by distillation or column chromatography.
NOTE 1When a solvent is purified, stabilizers put into the solvent by
the manufacturer are removed, thus potentially making the solvent
hazardous. Also, when a solvent is purified, preservatives put into the
solvent by the manufacturer are removed, thus potentially making the
shelf-life short. However, it is generally more economical to obtain a new
source of solvent. Interference-free solvent is defined as a solvent
containing less than 0.1 g/L individual analyte interference. Protect
interference-free solvents by storing them in an area known to be free of
organochlorine solvents.
6.2 This liquid-liquid extraction technique efficiently extracts a wide boiling range of nonpolar organic compounds
and, in addition, extracts polar organic components of the
sample with varying efficiencies.
6.3 Current column technology suffers from the fact that
EDB at low concentrations may be masked by very high levels
of dibromochloromethane (DBCM), a common disinfection
by-product of chlorinated drinking waters.
7
J & W DB-1, 1.0 m film, available from J & W Scientific, or its equivalent,
has been found suitable for this purpose.
8
RtxVolatiles, 2.0 m film thickness. Restek part #10902, available from Restek
Corp., 110 Benner Circle, Bellefonte, PA 16823, or its equivalent has been found
suitable for this purpose.
9
These parameters were obtained by Restek Corporation during preliminary
attempts to improve the separation of EDB and DBCM.
5
An alternative column has been recommended by the Restek Corporation and
is described in 7.1.5 as Column C.
6
J & W Durawax DX-3, 0.25 m, available from J & W Scientific, 91 Blue
Ravine Rd., Folsom, CA 95630, or its equivalent, has been found suitable for this
purpose.
D 5316 98 (2004)
8.10.4 Store stock standard solutions in 15-mL bottles
equipped with PTFE-lined screw caps. Methanol solutions
prepared from liquid analytes are stable for at least four weeks
when stored at 4C.
8.11 Standard Solutions, Primary DilutionUse stock standard solutions to prepare primary dilution standard solutions
that contain both analytes in methanol. The primary dilution
standards should be prepared at concentrations that can be
easily diluted to prepare aqueous calibration standards (see
12.1.1) that will bracket the working concentration range. Store
the primary dilution standard solutions with minimal headspace, and check frequently for signs of deterioration or
evaporation, especially just before preparing calibration standards. The storage time described for stock standard solutions
also applies to primary dilution standard solutions.
9. Hazards
9.1 The toxicity and carcinogenicity of chemicals used in
this test method have not been precisely defined; each chemical
should be treated as a potential health hazard, and exposure to
these chemicals should be minimized. Each laboratory is
responsible for maintaining awareness of OSHA regulations
regarding safe handling of chemicals used in this test method.
Additional references to laboratory safety need to be made
available to the analyst.
9.2 EDB and DBCP have been tentatively classified as
known or suspected human or mammalian carcinogens. Pure
standard materials and stock standard solutions of these compounds should be handled in a hood or glovebox. A NIOSH/
MESA approved toxic gas respirator should be worn when the
analyst handles high concentrations of these toxic compounds.
NOTE 2When a solvent is purified, stabilizers put into the solvent by
the manufacturer are removed, thus potentially making the solvent
hazardous.
10
Reagent Chemicals, American Chemical Society Specifications, American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Standards for Laboratory Chemicals, BDH Limited, Poole, Dorset, UK, and the United States Pharmacopeia.
D 5316 98 (2004)
TABLE 2 Bias and Precision at 2.0 g/L over a Four-Week Study
Period
Analyte
EDB
DBCP
Matrix A
RW-A
GW
GW-A
TW
TW-A
RW-A
GW
GW-A
TW
TW-A
Concentration
(g/L)
Average Bias
(%)
Relative
Standard
Deviation (%)
16
15
16
16
16
16
16
16
16
16
+4
+1
4
7
7
+5
+5
+1
5
6
4.7
2.5
4.7
6.3
6.1
8.2
6.2
8.4
10.1
6.9
A
Matrix Identities: RW-A = Reagent water at pH 2, GW = Groundwater, ambient
pH, GW-A = Groundwater at pH 2, TW = Tap water, ambient pH, TW-A = Tap
water at pH 2.
10.2.2 The addition of sodium thiosulfate as a dechlorinating agent or acidification, or both, to pH 2 with HCl (1 + 1),
common preservative procedures for purgeable compounds,
has been shown to have no effect on EDB or DBCP (see Table
2). Nonetheless, sodium thiosulfate must be added to avoid the
possibility of reactions that may occur between residual
chlorine and indeterminate contaminants present in some
solvents, yielding compounds which may subsequently interfere with the analysis. The presence of sodium thiosulfate will
arrest the formation of DBCM (see 6.3). Also, samples should
be acidified to avoid the possibility of microbial degradation
that may periodically affect these analytes contained in other
groundwater matrices.
10.3 Sample Storage:
10.3.1 Store samples and field reagent blanks together at
4C until analysis. The sample storage area must be free of
organic solvent vapors.
10.3.2 Analyze all samples within 28 days of collection.
13. Procedure
13.1 Sample Preparation:
13.1.1 Remove samples and standards from storage and
allow them to reach room temperature.
13.1.2 For samples and field reagent blanks, contained in
40-mL bottles, remove container cap. Discard a 5-mL volume
using a 5-mL transfer pipet or 10-mL graduated cylinder.
Replace the container cap and weigh the container with
contents to the nearest 0.1 g and record this weight for
subsequent sample volume determination (13.3).
13.1.3 For calibration standards, laboratory fortified blanks,
and laboratory reagent blanks, measure a 35-mL volume using
a 50-mL graduated cylinder and transfer it to a 40-mL sample
container.
13.2 Microextraction and Analysis:
13.2.1 Remove the container cap and add 6 g NaCl (see 8.8)
to the sample.
13.2.2 Recap the sample container and dissolve the NaCl by
shaking by hand for about 20 s.
13.2.3 Remove the cap and, using a transfer pipet, add 2.0
mL of hexane. Recap and shake vigorously by hand for 1 min.
Allow the water and hexane phases to separate. (If stored at
this state, keep the container upside down.)
13.2.4 Remove the cap and carefully transfer 0.5 mL of the
hexane layer into an auto-injector using a disposable glass
pipet.
13.2.5 Transfer the remaining hexane phase, being careful
not to include any of the water phase, into a second autoinjector vial. Reserve this second vial at 4C for a reanalysis if
necessary.
13.2.6 Transfer the first sample vial to an auto-injector set
up to inject 2.0 L portions into the gas chromatograph for
D 5316 98 (2004)
TABLE 3 Single Laboratory Bias and Precision for EDB and
DBCP in Tap Water
Analyte
EDB
DBCP
Number of
Samples
Concentration
(g/L)
Average Bias
(%)
Relative
Standard
Deviation (%)
7
7
7
7
7
7
0.03
0.24
50.0
0.03
0.24
50.0
+ 14
2
5
10
+2
6
9.5
11.8
4.7
11.4
8.3
4.8
Water Type
Applicable Conc. Range
Reagent Water:
Single-Analyst Precision
Overall Precision
Recovery
Groundwater:
Single-Analyst Precision
Overall Precision
Recovery
A
SR = 0.041X + 0.004
S = 0.075X + 0.008
X = 1.072C 0.006
SR = 0.065X + 0.000
S = 0.143X 0.000
X = 0.987C 0.000
SR = 0.046X + 0.002
S = 0.102X + 0.006
X = 1.077C 0.001
SR = 0.076X 0.000
S = 0.160X + 0.006
X = 0.972C + 0.007
X = Mean recovery
C = True value for the concentration
(1)
1,2-dibromo3-chloropropane
14. Calculation
14.1 Identify EDB and DBCP in the sample chromatogram
by comparing the retention time of the suspect peak to the
retention times generated by the calibration standards and the
laboratory control standard.
14.2 Use single point calibrations (12.1.4) or use the calibration curve or calibration factor (12.1.3) to directly calculate
the uncorrected concentration (Ci) of each analyte in the
sample (for example, calibration factor 3 response).
14.3 Calculate the sample volume (Vs) as equal to the net
sample weight:
Vs 5 gross weight ~13.1.2! 2 bottle tare ~13.3.3!
1,2-dibromoethane
(2)
D 5316 98 (2004)
TABLE 5 Ranges for Quality Control Sample
Spike
Concentration,
g/L
0.240 (EDB)
0.240 (DBCP)
Proficiency
Acceptable
Standard
Deviation, max
0.0276 g/L
0.0312 g/L
Demonstration
QC Check
Acceptance Range
for Mean Recovery
Acceptance Range
for QC Check
Recovery Spike
0.240 (EDB)
0.240 (DBCP)
Acceptance Range
for Recovery Spike
0.149 0.331 g/L
0.107 0.373 g/L
18. Keywords
18.1 DBCP;1,2-dibromo-3-chloropropane;1,2-dibromoethane; EDB; ethylene dibromide; gas chromatography;
microextraction
APPENDIX
(Nonmandatory Information)
X1. Reference Statistics
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.