Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
AND HOW IS IT
MEASURED?
A Technical Handbook For Industry
By Joe Novak
Contents
CHAPTER 1 INTRODUCTION
The Properties of Water ______________________________________________3
Ion Product Constant of Water _________________________________________3
BIBLIOGRAPHY _____________________________________________21
GLOSSARY ________________________________________________22
2
Chapter 1 INTRODUCTION
The Properties of Water
Water is the most common substance and the most important. In vapor, liquid, or solid form it covers more than seventy percent of the Earths surface and is a major
component of the atmosphere. Water is also an essential
requirement for all forms of life, most living things are
largely made up of water. Human beings, for example, are
about two-thirds water.
Water molecules are in continuous motion, even at low temperatures. Occasionally, when two water molecules collide, a
hydrogen ion is transferred from one molecule to the other
(Figure 1). The water molecule that loses the hydrogen ion
becomes a negatively charged hydroxide ion. The water
molecule that gains the hydrogen ion becomes a positively
charged hydronium ion. This process is commonly referred
to as the self-iionization of water.
Smaller ions generally move through a solution more rapidly than larger ones. The hydrogen ion [H+] and the
hydroxyl ion [OH-] are extremely mobile due to their geometry, as shown in Figure 4.
Polarization
When a DC voltage is applied to the electrodes of a conductivity sensor, positive (H+) ions and negative (OH-) ions
move to the positive and negative electrodes of the sensor
(Figure 5). The ions gain or lose electrons, and are converted
to hydrogen gas at the negative electrode and oxygen gas at
the positive electrode. These gases effectively form an insulation barrier at the electrodes, increasing the apparent
resistance that in turn decreases the apparent conductivity
of the solution. Using an AC voltage, at 60 Hertz or higher,
and increasing the cross sectional area of the electrodes, virtually eliminates the effects associated with polarization.
However, there are practical limits to increasing the electrode surface area. Electrode size cannot become too large
due to the physical constraints of sensor installation.
If all salts and free ions are removed from water, what
remains is referred to as theoretically pure water. Theoretically pure water has, at 25C, a specific conductance of
0.056 S/cm and a resistivity of 18.21 megohms cm. High
quality condensed steam and distilled or demineralized
water have specific conductances of 1.0 S/cm or less. This
value is caused by about 0.5 ppm of dissolved salts. Table 1
and Figure 6 (on the next page) show typical conductivity
ranges of various substances.
of 0.4 to 0.75 times the raw conductivity value. The variation is due to the type of dissolved solid(s) that are in the
sample. Table 2 can be used to convert between conductivity, resistivity, and TDS.
Conductivity/Concentration
Relationships
Different solutions with different conductivities, do not
always have directly proportional concentrations of salts or
solids in solution. In dilute solutions, an increase in conductivity is linear with an increase in concentration of the solution provided an electrolyte is dissolved in the solution and
there are no interactions between the solution and the dissolved electrolyte. When these two conditions are met, the
Resistivity
(ohms cm)
at 25C
Dissolved Solids
Parts/million
(ppm)
18,000,000
17,000,000
16,000,000
15,000,000
14,000,000
13,000,000
12,000,000
11,000,000
10,000,000
0.0277
0.0294
0.0313
0.0333
0.0357
0.0384
0.0417
0.0455
0.0500
9,000,000
8,000,000
7,000,000
6,000,000
5,000,000
4,000,000
3,000,000
2,000,000
1,000,000
0.0556
0.0625
0.0714
0.0833
0.100
0.125
0.167
0.250
0.500
Conductivity
(S/cm)
at 25C
Pure Water
1.110
1.250
1.430
1.670
2.000
2.500
3.330
5.000
10.00
Resistivity
(ohms cm)
at 25C
Dissolved Solids
Parts/million
(ppm)
900,000
800,000
700,000
600,000
500,000
400,000
300,000
200,000
100,000
0.556
0.625
0.714
0.833
1.000
1.250
1.670
2.500
5.000
pH versus Conductivity
Measurements
A frequently asked question when making water quality
measurements is whether conductivity or pH should be
measured.
pH is specific to hydrogen ions it will not respond to other
ions that may be present in solution. For example, if iron
sulfate were dissolved in the sulfuric acid solution discussed
above, the measurement of pH, the acid concentration,
would be unaffected. However, because conductivity
responds to all ions present, the measurement of conductivity will change with changes in iron sulfate content, while
the acid content remains constant. When determining
whether to measure pH or conductivity, always consider
solution concentrations and ion activity.
Example: Should pH or conductivity be used to measure the
concentration of sulfuric acid? This depends on the parameters of the process. If the concentration of the acid is less
than 0.5%, pH or conductivity measurement could be used.
If the acid concentration is greater, conductivity measurement should be used. The limitation for using pH measurement is that the hydrogen sensitive measuring electrode
becomes saturated at high concentrations of hydrogen ions.
Conductivity sensors are not hampered by high concentrations of hydrogen ions.
Effects of Temperature on
Solution Conductivity
In a solid, electrons move almost instantly upon the application of a voltage potential. In a solution, the electron
charge is carried by ions. Therefore, the applied potential,
species and concentration of the ion, and temperature of the
solution determine the speed at which the charge is transferred.
The resistance of a solid conductor has a positive temperature coefficient. Resistance increases with increasing temperature. Solid conductors vary chemically and possess
different positive temperature coefficients.
In solutions, ion mobility increases with increasing temperature. This increase in ion mobility means that ions will
carry more electrons per unit of time, resulting in a decrease
in solution resistance and a corresponding increase of conductivity which is a negative temperature coefficient.
The mobility of the dissociated ions depends on the temperature of the solution, and is not always linear because of the
contaminants in solution.
Temperature Compensation
Electrical conduction in aqueous solution is related to the
temperature, solution concentration, and the specific conductivity of the ions in the solution (Figure 8). Temperature
compensation corrects the conductivity measured at the
sample temperature to what that value would be if the sample was at a reference temperature (normally 25C).
Solution conductivity varies with ion concentration as well
as temperature and must be related to conductivity measurements at a reference temperature. The temperature coefficient (TC or temperature compensation factor) of
conductivity solutions in water is typically positive, linear,
and ranges between 2% and 7% per C, depending on the
nature and concentration of the conducting ions. Water
solutions typically have a compensation factor of 2% per C.
An analyzer or meter that is properly temperature compensated will read conductivity corrected to the reference temperature and the temperature coefficient curve, regardless of
sample temperature.
For example, water with a TC of 2% per C (0.02) and
500 S/cm at 25C, would actually read 510 S/cm at 26C,
an increase of 10 S/cm, if the measurement was not properly temperature compensated.
Information on specific temperature coefficients for most
chemicals can be found in published reference books and
periodicals.
The overall accuracy of a conductivity system is governed
to a large extent by the accuracy of its temperature compensator rather than by any other single factor and, many
times, by all other factors combined.
Non-linear Temperature
Compensation
In high purity water (1-18.3 megohms cm), temperature
compensation is non-linear. The conductivity versus temperature is not as easily defined. Special resistivity instruments are used to monitor high purity water and have
pre-programmed electronics for handling the conductivity
versus temperature equation.
Resistance Measurement of
Conductivity
Measuring the resistance of the solution and calculating its
reciprocal determine conductance. The Whetstone Bridge
(Figure 9), due to its inherent accuracy, is normally used to
make this resistance measurement. The conductivity sensor
forms one section of the bridge (R1 and R2), and from the
balancing arm (R3 and Rs), the resistance of the solution
being measured can be calculated.
A practical means of providing temperature compensation is
to introduce into the bridge circuit a resistive element (Rt)
that will change with temperature at the same rate as the
solution being measured. This temperature compensator arm
of the Whetstone Bridge can be:
Automatican RTD or thermistor in good thermal contact with the sample being measured.
Regardless of the type of device used to accomplish the temperature measurement, accurate compensation for temperature changes requires the temperature coefficient
compensator to match that of the solution being measured,
and that the temperature compensator be in direct contact
with the solution.
10
Non-contacting (Electrodeless)
Style Sensors
Conductivity measurements are often made in solutions
which tend to coat, foul, or damage the surface of conventional contacting style conductivity sensors. When measuring solutions over 10,000 S/cm with electrode type sensors,
large cell constants must be used. These sensors have small
electrode surface areas and, consequently, are very susceptible to fouling and polarization which introduce measurement errors.
Non-contacting or electrodeless conductivity sensors do
not have electrodes in direct contact with the solution being
measured (Figure 12). They are immune to the normal problems associated with contacting style conductivity sensors.
Electrodeless style conductivity sensors were specifically
developed to solve these problems.
K = 5/50 = 0.1
The ratio Rc / Rs is defined as the cell constant (K) for the
measuring sensor. It can be used to convert readings made
with the previously uncalibrated sensor to specific conductance using the following equations:
Operation
The quotient obtained by dividing 1,000,000 by the resistance value yields a S/cm value. Example: 100 ohms =
1,000,000/100 or 10,000 S/cm. Use the same procedure to
convert S/cm to ohms. Example: 200 S/cm =
1,000,000/200 or 5,000 ohms.
Electrode Materials
Electrode materials affect the conductive properties of contacting style conductivity sensors. In the past, platinum and
gold, which exhibited high performance characteristics,
were used as electrode materials. When aggressive solutions
(concentrated acids and caustics) were measured, noble metals that resisted oxidation were used to eliminate dissolving
of the electrode. Today, most electrodes are made of titanium, graphite, or stainless steel. These materials have good
performance characteristics in addition to exhibiting good
corrosive resistance in most applications.
11
Sensor Installation
Best performance comes from a conductivity system that
has its sensor mounted properly. Conductivity sensors are
usually installed in-line (flow-thru) or immersed (submersion) in an open vessel.
In flow through applications the following should be
adhered too. For contacting conductivity sensors, flow
should contact the tip end of the sensor and completely
surround the sensor before exiting. Electrodeless conductivity sensors must have a flow path that flows through the
center of the sensor and completely surrounds the sensor
12
8
7
5
1
2
3
4
Insertion mounting
Insertion mounting
Non-metallic sensor, insertion mounting
Boiler wall insertion mouting
5
6
7
8
13
2.
3.
4.
Analyzers
Until the advent of the microprocessor-based conductivity
analyzer, many of the calculations (algorithms) necessary
for accurate conductivity/resistivity measurement could not
be effectively implemented. The flexibility of the microprocessor enables the inclusion of multiple concentration
ranges, linear and non-linear ranges, and for measuring systems using two sensors, dedicated math functions for TDS
14
(Total Dissolved Solids), % passage or % rejection, and difference, all within the same analyzer.
Also, separate alarms can be dedicated to a data point, set
to a math function, or configured for diagnostic and or temperature alarms. Typically, resolution can be read to 3-1/2
digits after the decimal point. Other microprocessor-based
analyzer attributes include system diagnostics for the sensor
and analyzer.
It is interesting to note that analog analyzers still have merit
and can be advantageous. For specific applications only
requiring moderate accuracy, the generally lower-priced
analog instruments may be the best choice.
Analyzer Installation
The analyzer should be installed according to the manufacturers specifications. Pay particular attention to grounding.
Ground loops commonly cause havoc with conductivity
readings.
Calibration
Some installations may require the sensor to be located further from the analyzer than the standard sensor cable
length, typically 20 ft. (6 m). In these cases, use an interconnect or extension cable recommended by the sensor
manufacturer and a suitable junction box to protect the
splicing between the sensor and interconnect cables. Use of
other cables may introduce errors due to cable capacitance
effect, noise intrusion, or line loss. In general, low conductivity ranges necessitate shorter cable runs, while higher
ranges enable longer runs. Be sure to follow good wiring
practices and use as few termination points as possible. If
cabling must be spliced, a soldered splice is better than a
crimped splice. Make sure to protect any splices from moisture. Sealing the splice with a non-conductive type RTV or
silicone sealant is highly recommended.
Conductivity sensors and their associated temperature compensators require only a one-point calibration. It may be
inconvenient or impossible for the operator to calibrate the
conductivity sensor to the analyzer using known solutions
or temperature baths. This is especially true when operating
at conductivity or temperature points that are at the
extremes of their ranges, or that would require unstable or
hazardous solutions.
Calibration of Conductivity
Systems
The old adage of an analyzer is only as good as its calibration holds true for conductivity analyzers, and is especially
critical when measuring high purity water (HPW). Wet calibration to known standards is the accepted method for calibrating most conductivity measurement systems. However,
it is extremely difficult, if not impossible, to wet calibrate
conductivity measurement systems used for HPW measurement. The difficulties in preparing qualified samples of HPW
standards without error are monumental.
Qualified HPW samples are unstable and costly to make.
They are, however, available in small quantities as 5 S/cm,
and 15 S/cm standards from NIST (National Institute of
Standards and Testing). Cost as of this writing is $360 for a
500 ml sample. These standards are only good for one time
use, and only under laboratory conditions. Therefore this
may not be the best choice for calibrating a system used to
measure HPW conductivity.
A better method that some analyzers offer is a dry calibration method in which samples are not required. This calibration is also known as the true cell constant (K)
method. To perform this method, the true cell K not just
the nominal cell K must be known. The analyzer then uses
the entered true cell K in its measuring algorithm to calculate the conductivity value:
Conductivity =
106 x Kcell
Rcell [1 + S (C - 25)]
Where:
Conductivity is in S/cm at 25C
Kcell = true cell constant
Rcell = resistance in ohms
S = the slope of the solution with temperature.
C = temperature in degrees C.
For these reasons, a dry calibration makes calibrating easier and more accurate. The conductivity sensor is precisely
measured at the factory for both conductivity slope and
temperature offset. The sensor is then labeled with the true
cell K and temperature T factor for entry into the analyzer. The assumption with dry calibration is that the characteristics of the sensor have not changed; the electrodes are
clean and not deformed or eroded away. If the physical
characteristics of the sensor have not changed from its manufactured state, dry calibration is valid. This makes the dry
calibration method viable for any conductivity range.
Maintenance Schedule/Cleaning
Conductivity sensors, like any other sensor installed in a
process, will eventually succumb to contamination from the
process. Regular cleaning of the conductivity sensor will
assure long, reliable, and accurate service.
Conductivity sensors are typically made of durable, corrosion-resistant materials. Other than cleaning to remove contaminants, no other maintenance should be required.
Careful examination of the electrodes should be performed
to make sure they have not been chemically attacked,
15
eroded, or physically altered. A slow degradation of the sensor can be hard to spot until its too late.
Cleaning should be done with a combination of soaks and
rinses in a solution of water and detergent. This usually
removes most contaminants. Persistent contaminants may
require soaking in weak acid or caustic solutions for brief
periods of time, followed by thorough rinsing with clean
water to remove and neutralize any residual acid or caustic.
NOTE: Always follow all safety procedures when using
chemicals to clean conductivity sensors.
16
Applications
Industry
Applications
Chemical Production
Concentrations*
Purity
Leak Detection* (Heat Exchange)
Cleaning*
Scrubbers*
Waste Streams
Cooling Towers
Mining
Leaching
Concentration*
Waste Streams
Leak Detection* (Heat Exchanges)
Semiconductors
Reverse Osmosis
Deionization*
Electro Deionization
Concentration*
Waste Streams
Rinsing*
Food Processing
Tomato Paste
Pickling
Waste Streams
Chemical Peeling*
Papermaking
Concentration
Black Liquor*
Bleaching
Brown Stock*
Waste Streams
Cooling Towers*
Printing
Fountain Control*
Water Production
Reverse Osmosis
Demineralization*
Deionization*
Electro Deionization
Desalination*
Nano Filtration
Distilled Water*
Petroleum Drilling
Mud Control
Concentration*
Waste Streams
Pharmaceuticals
Reverse Osmosis
Deionized Water*
Concentration*
Waste Streams
Power Generation
Textiles
Steam Production*
Condensate Return*
Boiler Blow-down*
Cooling Towers*
Leak Detection* (Heat Exchangers)
Demineralizers*
Reverse Osmosis
Cleaning*
Concentration*
Waste Streams
Desalination*
Dye Vats*
Bleaching*
Washing*
Rinsing*
Waste Streams
Cooling Towers*
Concentration*
Cooling Towers*
Cleaning*
Waste Streams
Pickling Baths*
Metals Finishing
Plating Rinsing*
Concentration
Pickling Bath
Cleaning*
Waste Streams
Cleaning*
Rinsing*
Concentration*
17
Demineralizers/Deionizers
Two-bed Demineralizers
In a two-bed demineralizer, synthetic resins adsorb cations
and anions in the water as they pass through the cation and
anion beds. A conductivity sensor located in the discharge
of the cation bed will signal resin depletion by a drop in
conductivity. A conductivity sensor located in the discharge
of the anion bed will signal the depletion of resin in either
bed by a rise in conductivity. In large demineralizers, additional conductivity measuring systems are commonly used
to monitor the strength of the regenerate acid and caustic,
and check the rinse before the demineralizer is put back in
service. A properly operating deionizer train will remove
between 98-99.9% of all ionic impurities. The final product
water will have a conductivity of 0.056 S/cm which is considered theoretically pure.
Unlike aqueous solutions of higher conductivities (more
than 25 S/cm), high purity water (HPW) requires special
considerations for accurate conductivity measurements.
Sensors must be constructed of materials that do not readily dissolve in HPW, which would add ions. The sensor must
not have voids where air can become trapped, causing
unstable readings. It must have a very fast acting and accurate ( 0.1C) temperature compensation device with its
true cell K and T factor must be known. The analyzer
must be microprocessor-based and include algorithms specific to HPW measurement.
Mixed-bed Demineralizers
In a mixed-bed demineralizer, the anion and cation resins
are mixed. This has the effect of a number of two-bed demineralizers in series and produces higher quality product
water. A conductivity sensor is used to monitor the discharge. A rise in conductivity signals depletion of either
resin, similar to the two-bed system. Additional conductivity measuring systems can be used to monitor the acid and
caustic regenerate solutions as well as the rinse-down.
18
Distilled Water
The purity of distilled water is determined by conductivity.
Conductivity is measured at the discharge line of the still. If
there is a storage vessel for this water, this too can be
monitored.
Chemical Peeling
In canning processes, many fruits and vegetables such as
potatoes, onions, peaches, etc., are placed on a conveyor
and run through a hot caustic soda bath to loosen their
skins. The strength of the caustic solution is monitored by
placing a conductivity sensor between the discharge of the
bath and inlet of the circulating pump used to keep the concentration equal throughout the bath. The bath eventually
becomes contaminated with dirt, peel, and other foreign
material, and its viscosity increases. A change in conductivity signals when the bath must be cleaned.
Boiler Operation
Boiler Blowdown
Dissolved solids concentrate in boiler water where steaming
occurs. Depending on the boiler design, proper boiler operation requires that the solids concentration be maintained
below some limit. This is accomplished by blowing down
(draining) a portion of the boiler water and maintaining the
water level by adding feedwater with a low solids content.
Blowdown is normally measured as TDS (Total Dissolved
Solids).
Condensate
Monitoring returned condensate is one of the largest uses of
conductivity measuring systems. Where steam is manufactured, the return condensate may be contaminated with
acid, caustic, plating solution, brine, etc. These contaminates
are easily detectable by conductivity analyzers. In other
processes, the condensate may contain materials such as
sugar syrup, contaminated oils, etc. which are non-ionic but
which contain impurities that are ionic. Conductivity measurement is a practical way to detect these impurities.
When electricity is generated by steam, a water-cooled condenser is usually used after the turbine that reduces pressure
and temperature to insure a liquid sample for making a con-
Steam
Steam can be checked for carryover of boiler water solids by
using proper sample conditioning equipment and conductivity measuring systems. Where steam is employed for its
heat content or where it is used to drive a reciprocating
steam engine, sensitivities below 1 ppm TDS are not generally required. In this case, a proper sampling system consists
of a suitably sized cooling coil and a throttling valve to drop
the temperature and pressure. The conductivity of the sample is then monitored.
Where steam is used to operate a turbine, much more stringent requirements are placed on the steam generator. Small
amounts of solids will foul a turbine and reduce its efficiency. The use of a sample cooler to reduce steam temperature and pressure and provide a condensed sample is a
common practice at solids levels above 1 ppm TDS in the
steam. However, if turbine operation requires steam purity
somewhat better than this for prolonged service, more elaborate sample conditioning is necessary.
The Larson-Lane Steam Analyzer (U.S. Patent 2,832,673) is
a combination thermal and chemical analyzer which substantially increases the sensitivity of the conductivity measurement. Not only dissolved solids, but also volatile
constituents such as carbon dioxide, ammonia and amines,
contribute to the conductivity of the condensed steam.
Degassing of the sample is a requisite for high sensitivity
and accuracy in measuring conductivity as a means of
measuring solids in steam. Mechanical degassing is effective
in stripping carbon dioxide. The more soluble ammonia and
amines are not appreciably removed. However, by preceding
a reboiling stage with ion exchange through a cation resin
bed, ammonia and amines are completely removed, and the
Cooling Towers
With growing emphasis on the conservation of water, cooling towers are prominent as a means of dissipating large
heat loads from chemical production, air conditioning systems, and power generation. Because evaporative losses in
the cooling tower concentrate the solids in the circulating
water, conductivity measurement is an effective way to control Blowdown and make-up water.
Rinsing
Rinsing of ionic solutions from a variety of materials
accounts for a large use of conductivity instrumentation.
Products like milk of magnesia, salt cake, titanium dioxide,
paint pigments, and latex curd fall into this category.
Measurement and control of the conductivity of the effluent
in the washing of these materials helps to insure consistent
high quality.
Adequate rinsing of metal and plastic in the electroplating
industry has special significance. Not only is adequate rinsing of corrosive cleaning and plating solutions from parts
important in maintaining their finish quality, but also
equally important is the reduction in the volume of water
being used. Continuous blow-down in this process effectively promotes adequate rinsing and consistent plating
quality while minimizing water volume and waste volume
that must be treated.
Semiconductors
Ultrapure water is an excellent solvent, leaving no residue.
It is used throughout the semiconductor manufacturing
process for rinsing wafers and other substrates. Monitoring
the quality of the deionized water used in these processes is
a critical part of quality control programs for this industry.
Metals
Cleaning with mixtures of caustic, polyphosphates, and other
compounds is used throughout the metal fabricating industry, starting with the mill and continuing through to the finishing shop. Changes in the cleaning bath concentration are
caused by drag-in and drag-out as well as by dilution.
Conductivity effectively monitors the concentration and
directly controls additions of make-up water or chemicals.
19
Acid Manufacture
Desalinization
Conductivity measuring systems are used in the manufacture of many primary industrial chemicals including sulfuric, nitric, hydrochloric, phosphoric, and hydrofluoric acids.
Many of these measurements are made at high acid concentrations, where conductivity will actually decrease with an
increasing acid concentration.
Scrubbing Towers
The composition and gradual depletion of the chemical contents of many industrial scrubbing towers are effectively
monitored by the measurement of conductivity. For example, a 10% caustic soda solution is commonly used to scrub
acid vapors from process gases and exhaust air. As hydroxyl
ions in the caustic soda are replaced by chloride, cyanide,
bicarbonate, etc., the conductivity of the solution decreases.
This decrease is proportional to the degree of depletion of
the original charge of caustic soda.
20
Reverse Osmosis
During reverse osmosis (RO), water is drawn first through a
filter unit to remove suspended solids, and then sent on to
the RO unit. In the RO unit, pressure is applied to water on
one side of a semi-permeable membrane, forcing the water
to diffuse through the membrane. Minerals, salts and colloids are rejected, while purified water passes through. Total
dissolved solids (TDS) removal can be 90% or higher with
conductivity readings below 5 S/cm.
The RO process functions well with waters that have high
dissolved solids. In fact, RO units are often used to remove
the majority of dissolved solids from water destined for an
ion exchange system. The product water from the reverse
osmosis unit is then passed through a two-bed or mixed-bed
deionization unit to remove the remaining ions that are
present. This step is sometimes referred to as polishing the
water.
BIBLIOGRAPHY
1.
Beckman, 1958, Principles of Industrial Solution Conductivity Measurements, Beckman Instruments, Inc.
2.
Leeds and Northrup, 1988, On Line Conductivity and Resistivity Measurements, Leeds and Northrup
3.
4.
5.
Thornton and Associates, 1998, Resistivity Measurement Techniques, Thornton and Associates
6.
John C. Russ, 1996, Materials Science and Engineering, North Carolina State University, Raleigh NC
21
GLOSSARY
Anion: A negatively charged ion in an aqueous solution.
Blowdown: In boiler operation, the purge of a small portion
of concentrated boiler water from the system to maintain a
desired level of dissolved and suspended solids in the system.
Cation: A positively charged ion in an aqueous solution.
Concentrate: In crossflow filtration, the portion of a feed
stream which does not permeate the media, but retains the
ions, organics, and suspended particles which are rejected
by the media.
Concentration: The amount of material contained in a unit
volume of fluid; the process of increasing the dissolved
material per unit volume.
Condensate: Water obtained through evaporation and subsequent condensation. Normally the water resulting from condensing plant steam originally generated in a boiler. (Water
condensed in a water still operation is usually called distillate.)
Contaminate: A source of contamination, an impurity. Any
substance in water which is not H2O.
Deionization (DI): The process of removing ionized salts
(theoretically up to 100%) from water through specially
manufactured ion exchange resins. Deionization typically
does not remove organics, viruses or bacteria, except
through accidental trapping in the resin, or when using
specially made strong base anion resins which remove gram
negative bacteria.
Demineralization: The process of removing minerals from
water, usually through deionization, reverse osmosis, or distillation.
Dissolved Solids: The residual material remaining after filtering the suspended material from water and evaporating
the solution to a dry state.
pH: An expression of hydrogen ion concentration, specifically the negative logarithm of the hydrogen ion concentration. The pH measurement range is 0-14, with 7 as neutral,
0-7 as acid, and 7-14 as alkaline (base).
ppb: An acronym for parts per billion, commonly considered equivalent to micrograms per liter (g/L).
22
ppm: An acronym for parts per million, commonly considered equivalent to milligrams per liter (mg/L).
ppt: An acronym for parts per trillion, commonly considered
equivalent to nanograms per liter (ng/L).
Precipitate: An insoluble product of a chemical reaction of
soluble compounds in water.
Reagent-ggrade Water (ASTM): Water that meets the standards for reagent use promulgated by American Society of
Testing and Materials. Four grades, RI through RIV, have
been established for specifically intended uses.
Semi-p
permeable Membrane: A membrane which allows a
solvent such as water to pass through, while rejecting certain dissolved or colloidal substances.
Sepralator: In crossflow membrane systems, a modular and
replaceable spiral-wound membrane cartridge or element.
Solutes: Matter dissolved in a solvent.
Suspended Solids (SS): Solid organic and inorganic particles
in a solution that are held in suspension.
TDS: An acronym for Total Dissolved Solids.
Two-bbed Tank: A pairing of cation and anion exchange
tanks, typically operating in series. Best used for the deionization of relatively high volumes of water. Capable of
product water resistivity of up to 1 megohmcm.
Ultrapure Water: Highly treated water of high resistivity
with no organics; usually used in the semiconductor and
pharmaceutical industries.
23