General
Specifications
GS 01E20D01-01E
AXF
Magnetic Flowmeter
Integral Flowmeter
/Remote Flowtube
Integral Flowmeter
Remote Flowtube
FEATURES
User-oriented Functionality
Fluid Adhesion Level Diagnosis
By constantly monitoring the level of insulating
substance on the electrodes, it is possible to determine
when maintenance is required.
With the utilization of an optional replaceable electrode,
the electrodes can be easily removed from the flowmeter
and cleaned.
Flexible Electrical Connection Direction
The converter or the terminal box can be rotated
arbitrarily to change the directions of electrical
connection on the site.
Clear and Versatile Indications
The LCD indicator employs a large, backlit full dotmatrix, that can facilitate various displays.
One to three lines are available. When there is an alarm
condition, a full description of the countermeasure is
indicated.
Easy Setup Parameters
The most frequently used parameters are arranged in a
group at the top.
The infra-red switches enable the users to set
parameters without opening the cover.
P. 1
P. 2
P.12
P.16
P.18
P.22
P.23
P.23
P.24
P.35
P.40
P.56
P.57
GS 01E20D01-01E
Copyright June 2003(YK)
16th Edition Jan. 2015(KP)
<<Contents>> <<Index>>
STANDARD SPECIFICATIONS
Converter (Integral flowmeter)
The contents of (*1) and (*2) described in the converter
specifications are follows.
*1: Select two points from: one pulse output, one alarm
output, one status input, or two status outputs.
*2: For models without an indicator, the configuration tool
(Such as HHT (handheld terminal) or FieldMateTM
etc.) is necessary to set parameters.
Excitation Method:
Standard dual frequency excitation:
Size 2.5 to 400 mm (0.1 to 16 in.)
Enhanced dual frequency excitation:
Size 25 to 200 mm (1.0 to 8.0 in.)
(Optional code HF1 or HF2)
Input Signal (*1) :
One Status Input: Dry contact
Load Resistance: 200 or less (ON), 100 k or more
(OFF)
Output Signals :
One Current Output: 4 to 20 mA DC (load resistance:
750 maximum, including cable resistance)
One Pulse Output (*1):
Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)
Output rate: 0.0001 to 10,000 pps (pulse/second)
One Alarm Output (*1):
Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)
Two Status Outputs (*1):
Transistor contact output (open collector)
Contact capacity: 30 V DC (OFF), 200 mA (ON)
Communication Signals :
BRAIN or HART communication signal
(Superimposed on the 4 to 20 mA DC signal)
Distance from Power Line: 15 cm (6 in.) or more
(Parallel wiring should be avoided.)
BRAIN:
Communication Distance:
Up to 1.5 km (0.93 miles), when polyethylene insulated
PVC-sheathed cables (CEV cables) are used.
Communication distance varies depending on the type
of cable and wiring used.
Load Resistance:
250 to 450 (including cable resistance)
Load Capacitance: 0.22 F or less
Load Inductance: 3.3 mH or less
Input Impedance of Communicating Device:
10 k or more (at 2.4 kHz)
HART:
Load Resistance:
250 to 600 (including cable resistance)
Note: HART is a registered trademark of the HART
Communication Foundation.
Indicator (*2):
Full dot-matrix LCD (32132 pixels)
Lightning Protector:
The lightning protector is built into the current output and
pulse/alarm/status input and output terminals. When
optional code A is selected, the lightning protector is
built into the power terminals.
Protection:
General-purpose Use/Sanitary Type/TIIS Flameproof
type:
IP66, IP67
Explosion proof type except TIIS:
In case of explosion proof type except TIIS, refer to
description of "Enclosure" in "HAZARDOUS AREA
CLASSIFICATION".
Coating:
Case and Cover: Corrosion-resistant coating
Coating Color; Mint green coating (Munsell 5.6 BG
3.3/2.9 or its equivalent)
Converter Material:
Case and Cover : Aluminum alloy
Mounting/Shapes (Integral Flowmeter):
Electrical Connection:ANSI 1/2 NPT female
ISO M20 1.5 female
JIS G1/2 female
Direction of Electrical Connection: The direction can be
changed even after delivery.
Terminal Connection: M4 size screw terminal
Grounding:
Grounding resistance 100 or less
When optional code A is selected, grounding resistance 10 or less shall be applied.
* In case of explosion proof type except TIIS, follow the
domestic electrical requirements as regulated in each
country.
* In case of TIIS Flameproof type, refer to description of
HAZARDOUS AREA CLASSIFICATION.
Functions
How to Set Parameters (*2):
The indicators LCD and three infra-red switches
enable users to set parameters without opening the
case cover. Parameters can also be set with the
configuration tool (Such as HHT (handheld terminal)
or FieldMate, etc.). The language for the HHT is
English only.
Displayed Languages (*2):
Users can choose a language from among English,
Japanese, German, French, Italian, and Spanish.
Instantaneous Flow Rate/Totalized Value Display
Functions (for models with an indicator) (*2):
The full dot-matrix LCD enables user selections of
displays from one line to three lines for:
Instantaneous flow rate
Instantaneous flow rate (%)
Instantaneous flow rate (bar graph)
Current output value (mA)
Totalized forward-direction flow rate
Totalized reverse-direction flow rate
Totalized differential flow rate
Tag No.
Results of electrode adhesion diagnostics
Communication type
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Totalizer Display Function (*2):
The flow rate is counted one pulse at a time according to
the setting of totalization pulse weights. For forward and
reverse flow measurement functions, the totalized
values of the flow direction (forward or reverse) and the
flow direction are displayed on the indicator together
with the units. The difference of totalized values between
the forward and reverse flow rate can be displayed.
Totalization for the reverse flow rate is carried out only
when Forward and reverse flow measurement functions is selected.
Damping Time Constant (*2):
Time constant can be set from 0.1 second to 200.0
seconds (63% response). The default is 3 seconds.
Span Setting Function (*2):
Span flows can be set in units such as volume flow rate,
mass flow rate, time, or flow rate value. The velocity unit
can also be set.
Volume Flow Rate Unit: kcf, cf, mcf, Mgal (US), kgal (US),
gal (US), mgal (US), kbbl (US)*, bbl (US)*,
mbbl (US)*, bbl (US)*, Ml (megaliter), m3,
kl (kiloliter), l (liter), cm3
Mass Flow Rate Unit (Density must be set.):
klb (US), lb (US), t (ton), kg, g
Velocity Unit: ft, m (meter)
Time Unit: s (sec), min, h (hour), d (day)
* US oil or US Beer can be selected.
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Process Connection
Lining
PFA
Remote Flowtube
Integral Flowmeter
Wafer
EPDM
Rubber
Ceramics
(*1)
Generalpurpose Use
PFA
Flange
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0),
250 (10), 300 (12),
350 (14), 400 (16)
Union Joint
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
Polyurethane
100 (4.0), 125 (5.0),
Rubber
150 (6.0), 200 (8.0),
250 (10), 300 (12),
350 (14), 400 (16)
EPDM
Rubber
High Grade
Accuracy 0.2% of
Rate (*3)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Size of AXF Flowtubes: AXF Standard (Lay length code 1) (continued)
Unit: mm (in.)
Use
Process Connection
Lining
PFA
Wafer
Integral Flowmeter
15 (0.5), 25 (1.0),
32 (1.25), 40 (1.5)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0), 250 (10),
300 (12)
Union Joint
Clamp:
Tri-Clamp (*4),
DIN32676
ISO2852/SMS3016
Union:
Sanitary Type
DIN11851
ISO2853 (*5)
SMS1145 (*6)
Butt Weld:
DIN11850, ISO203
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
EPDM
Rubber
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
PFA
15 (0.5), 25 (1.0),
32 (1.25), 40 (1.5),
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0), 250 (10),
300 (12), 350 (14),
400 (16)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
EPDM
Rubber
PFA
PFA
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0), 65 (2.5),
80 (3.0), 100 (4.0),
125 (5.0), 150 (6.0),
200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
Polyurethane 65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
Rubber
150 (6.0), 200 (8.0),
250 (10), 300 (12),
350 (14), 400 (16)
Flange
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
Wafer
Explosion
proof Type
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
Polyurethane 65 (2.5), 80 (3.0),
Rubber
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0),
250 (10), 300 (12)
Submersible
Type
Flange
Remote Flowtube
High Grade
Accuracy 0.2% of
Rate (*3)
25(1.0),40(1.5),50(2.0),
25 (1.0), 40 (1.5),
80(3.0),100(4.0),
50 (2.0), 80 (3.0),
100 (4.0), 150 (6.0), 150(6.0),200(8.0)
200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0),
150 (6.0), 200 (8.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0)
25 (1.0), 32 (1.25),
40 (1.5), 50 (2.0),
65 (2.5), 80 (3.0),
100 (4.0), 125 (5.0)
PFA
T21.EPS
*1: AXF standard lay length dimensions for wafer type ceramics linings are the same as those for ADMAG ceramics linings.
*2: AXF standard lay length dimensions for union joint type ceramics linings are the same as those for ADMAG ceramics linings.
*3: Enhanced dual frequency excitation is not available for models with High grade accuracy.
*4: Not available with 32 mm (1.25 in.), 125 mm (5.0 in.)
*5: Not available with 125 mm (5.0 in.)
*6: Not available with 15 mm (0.5 in.), 125 mm (5.0 in.)
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Size of AXF Flowtubes: Replacement model for earlier ADMAG or ADMAG AE (Lay length code 2)
Unit: mm (in.)
Use
Process Connection
Lining
Wafer (*6)
Submersible
Type
PFA
80 (3.0), 100 (4.0),
150 (6.0), 200 (8.0)
25 (1.0), 40 (1.5),
50 (2.0), 80 (3.0), 100 (4.0),
150 (6.0), 200 (8.0)
25 (1.0), 40 (1.5),
Polyurethane 50 (2.0), 80 (3.0),
rubber
100 (4.0), 150 (6.0),
200 (8.0)
PFA
Flange (*7)
Explosion
proof Type
High Grade
Accuracy 0.2% of
Rate
Wafer (*6)
Flange (*7)
Integral Flowmeter
PFA
Generalpurpose use
Remote Flowtube
Wafer (*6)
PFA
Flange (*7)
PFA
T22.EPS
*6: ADMAG lay length dimensions for wafer type of 250 mm (10 in.), and 300 mm (12 in.) are the same as those for AXF Standard.
And, in case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in wafer type of 2.5 mm (0.1 in.)
to 15 mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
*7: ADMAG lay length dimensions for flange type of 15 mm (0.5 in.) to 100 mm (4.0 in.), or 300 mm (12 in.) to 400 mm (16 in.) are the same as those for AXF
Standard. However, in case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in flange type of 15
mm (0.5 in.) to 100 mm (4.0 in.), the lay length of AXF Standard are longer by approx. 4mm (0.16 in) than those of earlier ADMAG or ADMAG AE.
Protection:
General-Purpose Use/Sanitary Type/TIIS Flameproof
Type:
IP66, IP67
Explosion proof type except TIIS:
In case of explosion proof type except TIIS, refer to
description of Enclosure in HAZARDOUS AREA
CLASSIFICATION.
Submersible Type (only for Remote Flowtube):
IP68 (Conforms to continuous immersion under the
following test condition)
Test Condition:
50 m below the surface of the water, equivalent to
0.5 MPa hydraulic pressure, for one month.
Cable should be protected at customer site.
Coating:
General-Purpose Use/Explosion proof Type:
Size 2.5 to 125 mm (0.1 to 5.0 in.) (Wafer type),
Size 2.5 to 125 mm (0.1 to 5.0 in.) (Process connection
code B or D of flange type):
Housing: No coating (Stainless steel surface)
Flange (Flange type only) : No coating (Stainless
steel surface)
Terminal Box and Cover (Remote Flowtube):
Corrosion-resistant coating
Coating color; Mint green (Munsell 5.6 BG 3.3/2.9
or its equivalent)
Size 150 to 300 mm (6.0 to 12 in.) (Wafer type),
Size 150 to 400 mm (6.0 to 16 in.) (Process connection
code B of flange type),
Size 50 to 400 mm (2.0 to 16 in.) (Process connection
code C of flange type):
Housing, Flange (Flange type only), Terminal Box
and Cover (Remote Flowtube):
Corrosion-resistant coating
Coating color; Mint green (Munsell 5.6 BG 3.3/2.9
or its equivalent)
Sanitary Type:
Size 15 to 125 mm (0.5 to 5.0 in.):
Housing: No coating (Stainless steel surface)
Adapter : No coating (Stainless steel surface)
Terminal Box and Cover (Remote Flowtube):
Corrosion-resistant coating
Coating color; Mint green (Munsell 5.6 BG 3.3/2.9
or its equivalent)
Submersible Type: Non-tar epoxy coating (black)
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Flowtube Material:
MiniFlange
Part Name
Material
Housing
Flange
Wafer Type
PFA/Polyurethane
Rubber lining
Wafer Type
Ceramics lining [only
for 15 mm (0.5 in.)]
Stainless steel-JIS
SUS316L (AISI 316
SS/EN 1.4404 equivalent)
Sanitary Type
[only for 15 mm (0.5 in.)]
Wafer Type
PFA/Polyurethane
Rubber lining
Flange Type
PFA lining
Sanitary Type
[only for 15 mm (0.5 in.)]
Pipe
Part Name
Material
Housing
Flange
Size
25 mm (1.0 in.)
(Lay Length code 1)
Aluminum alloy
Sanitary Type
T03.EPS
Wafer Type
PFA/Polyurethane
Rubber/
Natural Soft Rubber/
EPDM Rubber lining
Pipe
Flange Type
PFA/Polyurethane
Rubber/
Natural Soft Rubber/
EPDM Rubber lining
Wafer Type
Ceramics lining
Sanitary Type
Size
80 mm (3.0 in.),
100 mm (4.0 in.)
Size
25 mm (1.0 in.)
Size
32 mm (1.25 in.) to
125 mm (5.0 in.)
Size
25 mm (1.0 in.)
(Lay Length code 1)
Size
25 mm (1.0 in.)
(Lay Length code 2)
Size
32 mm (1.25 in.) to
125 mm (5.0 in.)
Size
25 mm (1.0 in.)
Size
32 mm (1.25 in.) to
125 mm (5.0 in.)
Size
25 mm (1.0 in.) to
100 mm (4.0 in.)
Size
25 mm (1.0 in.)
Size
32 mm (1.25 in.) to
125 mm (5.0 in.)
Aluminum alloy
T04.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Size 150 mm (6.0 in.) to 400 mm (16 in.)
Part Name
Housing
Flange
Gasket:
Material
Use
Wafer Type
Carbon steel-JIS SS400 or
PFA/Polyurethane Rubber/Natural Soft Rubber/ SFVC 2A
EPDM Rubber lining
Lining
Wafer Type
Ceramics lining [available with
150 mm (6.0 in.), 200 mm (8.0 in.)]
Terminal Box (Remote Flowtube)
PFA/Polyurethane Rubber/
Natural Soft Rubber/
EPDM Rubber
grounding
ring
Ceramics
grounding
ring
Standard
ceramic fillers
(Valqua #7020)
grounding
ring
grounding
ring
Aluminum alloy
T05.EPS
*3:
Electrode:
Stainless steel-JIS SUS316L (AISI 316L SS/EN 1.4404
equivalent), Hastelloy*1 C276 equivalent, Titanium,
Tantalum, Platinum-Iridium, Tungsten Carbide,
Platinum-Alumina cermet(only for ceramics lining)
Optional code
Gasket within Flowtube Gasket within Flowtube
(GA, GC,
GD, or GF)
Gasket Material (within Flowtube)
Only when
selecting
the PFA lining/
ceramics lining
grounding
ring
Optional code
(BCF, BSF,
BCC, or
Flange of users
BSC)
pipe
*1:
*2:
T23-1.EPS
Use
Sanitary Type
Lining
PFA
Standard
GS 01E20D01-01E
Jan. 7, 2015-00
<<Contents>> <<Index>>
Joints:
Lining
nut
screw joint
(GUN, GUR)
Standard
Gasket within Flowtube
Materials for Union Joint
Process Connection Code
GUW: Union Joint (weld joint)
Process Connection Code
GUN, GUR: union joint
(screw joint)
Stainless steel
(JIS SUS316L
(ANSI 316L SS/EN
1.4404
equivalent))
Use
Sanitary Type
Adapter for clamp
connection
Adapter for union
connection
Standard
GS 01E20D01-01E
Jan. 7, 2015-00
10
<<Contents>> <<Index>>
Inner Diameter of Grounding Ring, Outer Diameter for Effective Sealing, Recommended Inner Diameter of Gasket;
Be sure to choose a gasket with an inner and outer diameter that does not protrude inside the piping.
If the inner diameter of the gasket is too large, or outer diameter of the gasket is too small, fluid leakage may result.
AXF Standard:
Unit : mm (in.)
PFA/ Polyurethane Rubber/ Natural Soft Rubber/ EPDM Rubber
Flange
Water
Recommended Inner
Diameter of Gasket
Size
Ceramics
Outer
Diameter
for Effective
Sealing
[B]
Flat Gasket
[C]
PTFE-sheathed
Non-asbestos
Gasket
[D]
Outer
Inner
Diameter of Diameter for
Effective
Grounding
Sealing
Ring
[B]
[A]
Recommended Inner
Diameter of Gasket
PTFEsheathed
Nonasbestos
Gasket
[D]
Flat Gasket
[C]
2.5 (0.1)
15 (0.59)
38 (1.50)
17 (0.67)
22 (0.87)
30 (1.18)
5 (0.2)
15 (0.59)
38 (1.50)
17 (0.67)
22 (0.87)
30 (1.18)
10 (0.4)
15 (0.59)
38 (1.50)
17 (0.67)
22 (0.87)
30 (1.18)
15 (0.5)
15 (0.59)
38 (1.50)
22 (0.87)
15 (0.59)
34 (1.34)
22 (0.87)
15 (0.59)
33 (1.30)
22 (0.87)
35 (1.38)
25 (1.0)
28 (1.10)
53 (2.09)
35 (1.38)
28 (1.10)
53 (2.09)
35 (1.38)
27 (1.06)
50 (1.97)
32 (1.25)
34 (1.34)
58 (2.28)
43 (1.69)
34 (1.34)
58 (2.28)
43 (1.69)
40 (1.5)
41 (1.61)
71 (2.80)
49 (1.93)
41 (1.61)
71 (2.80)
49 (1.93)
40 (1.57)
68 (2.68)
49 (1.93)
50 (2.0)
53 (2.09)
84 (3.31)
61 (2.40)
53 (2.09)
84 (3.31)
61 (2.40)
52 (2.05)
82 (3.23)
61 (2.40)
65 (2.5)
66 (2.60)
103 (4.06)
84 (3.31)
66 (2.60)
103 (4.06)
84 (3.31)
80 (3.0)
77 (3.03)
114 (4.49)
90 (3.54)
77 (3.03)
114 (4.49)
90 (3.54)
81 (3.19)
112 (4.41)
90 (3.54)
100 (4.0)
102 (4.02)
140 (5.51)
115 (4.53)
102 (4.02)
140 (5.51)
115 (4.53)
98 (3.86)
134 (5.28)
115 (4.53)
125 (5.0)
128 (5.04)
165 (6.50)
141 (5.55)
128 (5.04)
165 (6.50)
141 (5.55)
150 (6.0)
146.1 (5.75)
190 (7.48)
167 (6.57)
146.1 (5.75)
190 (7.48)
167 (6.57)
144 (5.67)
188 (7.40)
167 (6.57)
200 (8.0)
193.6 (7.62)
240 (9.45)
218 (8.58)
193.6 (7.62)
240 (9.45)
218 (8.58)
192 (7.56)
240 (9.45)
218 (8.58)
250 (10)
243.7 (9.59)
300 (11.81)
270 (10.63)
243 (9.57)
315 (12.40)
270 (10.63)
300 (12)
294.7 (11.60)
348 (13.70)
321 (12.64)
291.3 (11.47)
360 (14.17)
321 (12.64)
350 (14)
323.4 (12.73)
405 (15.94)
359 (14.13)
400 (16)
373.5 (14.70)
465 (18.31)
410 (16.14)
*1: The inner diameter of the process connection code: DD4, DJ, DJ2 is values in brackets [ ].
T25.EPS
Unit : mm (in.)
Size
Flange
Recommended Inner
Diameter of Gasket
Inner Diameter Outer Diameter
PTFEof Grounding
for Effective
sheathed
Ring
Sealing
NonFlat Gasket
[A]
[B]
asbestos
[C]
Gasket
[D]
Inner Diameter of
Grounding Ring
[A]
Outer
Diameter
for Effective
Sealing
[B]
Recommended Inner
Diameter of Gasket
PTFE-sheathed
Non-asbestos
Gasket
[D]
Flat Gasket
[C]
2.5 (0.1)
15 (0.59)
38 (1.50)
17 (0.67)
22 (0.87)
5 (0.2)
15 (0.59)
38 (1.50)
17 (0.67)
22 (0.87)
22 (0.87)
10 (0.4)
15 (0.59)
38 (1.50)
15 (0.5)
15 (0.59)
38 (1.50)
22 (0.87)
25 (1.0)
27 (1.06)
56 (2.20)
35 (1.38)
40 (1.5)
40 (1.57)
71 (2.80)
49 (1.93)
50 (2.0)
52 (2.05)
85 (3.35)
61 (2.40)
80 (3.0)
81 (3.19)
115 (4.53)
90 (3.54)
100 (4.0)
98 (3.86)
144 (5.67)
115 (4.53)
150 (6.0)
140.7 (5.54)
190 (7.48)
167 (6.57)
140.7 (5.54)
205 (8.07)
167 (6.57)
200 (8.0)
188.9 (7.44)
240 (9.45)
218 (8.58)
188.9 (7.44)
255 (10.04)
218 (8.58)
250 (10)
243 (9.57)
315 (12.40)
270 (10.63)
17 (0.67)
D *2
C A
AXF
Flowtube
Lining
Grounding Ring
Flat Gasket
PTFE-sheathed Non-asbestos Gasket
T16-3.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
11
<<Contents>> <<Index>>
Electrode Construction:
Non-replaceable Electrode Type
General-Purpose Use/Submersible Type/Explosion
proof Type:
PFA, Polyurethane Rubber lining:
External insertion type
Natural Soft Rubber, EPDM Rubber lining:
Internal insertion type
Ceramics lining: Integral type
Sanitary Type: Internal insertion type
Replaceable Electrode Type
Electrode parts can be put into unit to facilitate
replacement or mounting at customer site.
The optional dedicated tool (F9807SK) is required.
Combined Converter:
A remote flowtube can be combined with the AXFA11
Converter or the AXFA14 Converter. If a combined
converter is changed from AXFA11 to AXFA14 or vice
versa, a new meter factor must be adjusted by flow
calibrations.
In case that size 250 mm (10 in.) or larger is used in low
conductivity or high concentration slurries, please use
the AXFA11 Converter.
Maximum Cable Length:
Combination of AXF remote Flowtube and AXFA11:
up to 200 m (660 ft)
Combination of AXF remote Flowtube and AXFA14:
up to 100 m (330 ft)
Process
Connection
Wafer
GeneralPurpose Use
Flange
Available
Size
Lining
25 to 300 mm
(1.0 to 12 in.) PFA/
Polyurethane
25 to 400 mm Rubber
(1.0 to 16 in.)
Electrode
Material
JIS SUS316L
(AISI 316L
SS/EN
1.4404
equivalent)(*1)
T06.EPS
Process
Connection
GeneralFlange
Purpose Use
Available
Size
Lining
Electrode
Material
JIS SUS316L
(AISI 316L
SS/EN
1.4404
equivalent)(*1)
T07.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
12
<<Contents>> <<Index>>
ATEX:
(Integral Flowmeter)
Explosion proof for Class I, Division 1, Groups A, B, C
& D.
Dust-ignition proof for Class II/III, Division1, Groups E,
F & G.
Intrinsically safe (electrodes) for Class I, Division 1,
Groups A, B, C & D.
SEAL ALL CONDUITS WITHIN 18 INCHES
WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: NEMA 4X
Temperature Code: T6
Note: Temperature Code T5 to T3 included in the scope of
application and its approval.
Maximum Process
Temperature
Minimum Process
Temperature
T6
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+130C (+266F)
40C (40F)
T27-1_1.EPS
*AXF002C AXF400C
Applicable Standard:
EN 50014, EN 50018, EN 50019,
EN 50020, EN 50028, EN 50281-1-1
Certificate: KEMA 03ATEX2435
(Integral Flowmeter)
ATEX Flameproof Type
Group: II
Category: 2G
EEx dme[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Temperature Class:
Temperature
Class
T6
Maximum Process
Temperature
Minimum Process
Temperature
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+130C (+266F)
40C (40F)
T27-2.EPS
Maximum Process
Temperature
T75C (+167F)
+70C (+158F)
T85C (+185F)
+85C (+185F)
T100C (+212F)
+120C (+248F)
T110C (+230F)
+130C (+266F)
T29.EPS
Temperature
Code
Maximum Process
Temperature
Minimum Process
Temperature
T6
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+150C (+302F)
40C (40F)
T28-1_1.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
13
<<Contents>> <<Index>>
CSA:
(Remote Flowtube)
ATEX Flameproof Type
Group: II
Category: 2G
EEx dme[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Temperature Class:
Temperature
Class
Maximum Process
Temperature
Minimum Process
Temperature
T6
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+150C (+302F)
40C (40F)
T28-2.EPS
Maximum Process
Temperature
T75C (+167F)
+70C (+158F)
T85C (+185F)
+85C (+185F)
T100C (+212F)
+120C (+248F)
T115C (+239F)
+150C (+302F)
T30.EPS
*AXF002C AXF400C
Applicable Standard:
For CSA C22.2 Series;
C22.2 No 0, C22.2 No 0.4, C22.2 No 0.5,
C22.2 No 25, C22.2 No 30, C22.2 No 94,
C22.2 No 157, C22.2 No.61010-1-12,
C22.2 No.61010-2-030-12
For CSA E79 Series;
CAN/CSA-E79-0, CAN/CSA-E79-1,
CAN/CSA-E79-7, CAN/CSA-E79-11,
CAN/CSA-E79-18
Certificate: 1481213
Process Sealing Certification:
Dual Seal certified by CSA to the requirements of
ANSI/ISA 12.27.01.
No additional sealing required.
Primary seal failure annunciation;
Deterioration of the flowrate output at nonzero flow
point.
Unstable flowrate output at zero flow point.
(Integral Flowmeter)
For CSA C22. 2 Series
Explosion proof for Class I, Division 1, Groups A, B, C
& D.
Dust-ignition proof for Class II/III, Division 1, Groups E,
F & G.
Intrinsically safe (electrodes) for Class I, Division 1,
Groups A, B, C & D.
SEAL ALL CONDUITS WITHIN 50 cm OF THE
ENCLOSURE
WHEN INSTALLED IN DIV. 2, SEALS NOT REQUIRED
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: Type 4X
Temperature Code:
Temperature
Code
T6
Maximum Process
Temperature
Minimum Process
Temperature
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+130C (+266F)
40C (40F)
T27-1.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
14
<<Contents>> <<Index>>
IECEx:
Maximum Process
Temperature
Minimum Process
Temperature
T6
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+130C (+266F)
40C (40F)
*AXF002C AXF400C
Applicable Standard:
IEC60079-0, IEC60079-1, IEC60079-7,
IEC60079-11, IEC60079-18, IEC61241-0,
IEC61241-1, IEC60529 + Edition 2.1
Certificate: IECEx KEM 05.0018
(Integral Flowmeter)
IECEx Flameproof Type
Ex demb[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Temperature Class:
T27-1.EPS
Maximum Process
Temperature
Minimum Process
Temperature
Temperature
Class
Process Temperature
T6
T5
T4
T3
Ambient Temp.:
PFA Lining; 40C to +60C (40F to +140F)
Ceramics Lining; 15C to +60C (5F to +140F)
IECEx Type of Protection Dust
Ex tD A21 IP6x T95C, T105C, T120C, T130C
Electrode Circuit Um: 250 Vac/dc
Maximum power supply voltage: 250 Vac/130 Vdc
Excitation Circuit: 140V max
Enclosure: IP66, IP67
Maximum surface temperature:
Maximum Surface
Temperature
Process Temperature
T6
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T95C (+203F)
T4
+120C (+248F)
40C (40F)
T105C (+221F)
T3
+150C (+302F)
40C (40F)
T120C (+248F)
T130C (+266F)
T28-1.EPS
T27-4.EPS
Ambient Temp.:
PFA Lining; 40C to +60C (40F to +140F)
Ceramics Lining; 15C to +60C (5F to +140F)
T6
Maximum Process
Temperature
Minimum Process
Temperature
+70C (+158F)
40C (40F)
T5
+85C (+185F)
40C (40F)
T4
+120C (+248F)
40C (40F)
T3
+150C (+302F)
40C (40F)
T28-1.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
15
<<Contents>> <<Index>>
TIIS:
(Remote Flowtube)
IECEx Flameproof Type
Ex demb[ia] IIC T6...T3
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Temperature Class:
Temperature
Class
Process Temperature
T6
T5
T4
T3
Ambient Temp.:
PFA Lining; 40C to +60C (40F to +140F)
Ceramics Lining; 15C to +60C (5F to +140F)
IECEx Type of Protection Dust
Ex tD A21 IP6x T95C, T105C, T120C, T135C
Electrode Circuit Um: 250 Vac/dc
Excitation Circuit: 170V max
Enclosure: IP66, IP67
Maximum surface temperature:
Maximum Surface
Temperature
Process Temperature
T95C (+203F)
T105C (+221F)
T120C (+248F)
T135C (+275F)
Ambient Temp.:
PFA Lining; 40C to +60C (40F to +140F)
Ceramics Lining; 15C to +60C (5F to +140F)
(Integral Flowmeter)
Construction: Ex de[ia] IIC T4
: Converter ; Explosion proof
Flowtube ; Increased Safety and
Intrinsically Safety(ia)
Electrode ; Intrinsically Safety(ia)
Gas Group and Temperature Class: IIC T4
Ambient Temperature: 20 to 60C* (power supply code 1)
: 20 to 50C (power supply code 2)
Fluid Temperature: 120C max
Electrode Circuit: 250V AC/DC
Maximum power supply voltage: 250V AC/130V DC
Grounding: JIS Class C(grounding resistance 10 or
less) or JIS Class A(grounding resistance
10 or less)
*In case that ambient temperature exceeds 50C, use
heat-resistant cables with maximum allowable temperature of 70C or above.
(Remote Flowtube)
Construction: Ex de[ia] IIC T4
: Terminal box ; Explosion proof
Flowtube ; Increased Safety and
Intrinsically Safety(ia)
Electrode; Intrinsically Safety(ia)
Gas Group and Temperature Class: IIC T4
Ambient Temperature: 20 to 60C*
Fluid Temperature: 120C max
Electrode Circuit: 250V AC/DC
Grounding: JIS Class C(grounding resistance 10 or
less) or JIS Class A(grounding resistance
10 or less)
*In case that ambient temperature exceeds 50C, use
heat-resistant cables with maximum allowable temperature of 70C or above.
Note : In case of TIIS Flameproof type, a remote flowtube
is available for combined use with the AXFA14
converter only.
GS 01E20D01-01E
Jan. 7, 2015-00
16
<<Contents>> <<Index>>
STANDARD PERFORMANCE
Accuracy
Size
mm (in.)
T09.EPS
0.8
High Grade
Standard
Flow Velocity Accuracy Flow Velocity Accuracy
V m/s (ft/s) (Calibration V m/s (ft/s) (Calibration
code B)
code C)
V 0.3 (1)
0.5 mm/s
0.18% of Rate
0.2mm/s
0.2% of Rate
T08.EPS
0
1
Velocity[m/s]
10
F14.EPS
0
0
6
Velocity[m/s]
6
Velocity[m/s]
Pulse
0.6
Output
Accuracy 0.4
0.35
0.2
High
Standard
gradeAccuracy
Accuracy
Pulse
0.6
Output
Accuracy 0.4
0.35
0.2
0
1.0 mm/s
0.3 V 10 0.35% of
(1)
(33) Rate
1.0 mm/s
0.3 V 10
0.35% of Rate
(1.0)
(33)
1.0
Standard Accuracy
(Calibration code B)
PFA/Ceramics Lining:
2.5 (0.1)
to
15 (0.5)
V 0.3 (1.0)
25 (1.0)
to
400 (16)
Pulse Output:
Size
mm (in.)
Flow Velocity
V m/s (ft/s)
10
F15.EPS
10
F14.EPS
Jan. 7, 2015-00
17
<<Contents>> <<Index>>
Remote Flowtube (optional code FF1)
Between signal terminals and ground terminal : 500V
AC for 1 minute or 600V AC for 1 second
Between signal terminals and excitation current
terminal : 2000V AC for 1 minute or 2400V AC for
1 second.
Remote Flowtube (optional code WT1)
Between excitation current terminal and ground
terminal : 1000V AC for 1 minute
Remote Flowtube (optional code WT2)
Between excitation current terminal and ground
terminal : 1500V AC for 1 minute
Between signal terminals and excitation current
terminal : 1500V AC for 1 minute
CAUTION
*1: When performing the Insulation Resistance Test or
the Withstand Voltage Test, please obey the following
caution.
Following the relevant test, wait for more than 10
seconds after the power supply has been turned off
before removing the cover.
Remove all wires from terminals before testing.
When the power terminal has a lighting protector
(optional code A), remove the short bar at the ground
terminal.
After testing, be sure to discharge by using a resistance and return all wires and the short bar to its
correct position.
Screws must be tightened to a torque of 1.18 N-m or
more.
After closing the cover, the power supply can be
restored.
DN
PS
(mm) (*1) (MPa) (*1)
PS DN
(MPa mm)
CATEGORY(*2)
AXF002G/C
2.5
10
Article 3, (*3)
paragraph 3
AXF005G/C
20
Article 3, (*3)
paragraph 3
AXF010G/C
10
40
Article 3, (*3)
paragraph 3
AXF015G/W/C
15
60
Article 3, (*3)
paragraph 3
AXF025G/W/C
25
100
Article 3, (*3)
paragraph 3
AXF032G/W/C
32
128
II
AXF040G/W/C
40
160
II
AXF050G/W/C
50
200
II
AXF065G/W/C
65
130
II
AXF080G/W/C
80
160
II
AXF100G/W/C
100
200
II
AXF125G/W/C
125
250
II
AXF150G/W/C
150
300
II
AXF200G/W/C
200
400
III
AXF250G/W/C
250
500
III
AXF300G/W/C
300
600
III
AXF350G/W/C
350
350
II
AXF400G/W/C
400
400
III
T10-1.EPS
Sanitary Type:
MODEL
DN
(mm) (*1)
PS D
PS
CATEGORY (*2)
(MPa) (*1) (MPa mm)
AXF015H
15
15
Article 3, (*3)
paragraph 3
AXF025H
25
25
Article 3, (*3)
paragraph 3
AXF032H
32
32
AXF040H
40
40
AXF050H
50
50
AXF065H
65
65
AXF080H
80
80
AXF100H
100
100
AXF125H
125
125
II
T10-2.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
18
<<Contents>> <<Index>>
Cable length
As show in accuracy
description
100 (330)
Combined
with AXFA11
or AXFA14
converter
50 (170)
3 (10)
0 (0)
1
5
10
20
Liquid conductivity
50
(S/cm)
m (ft)
200 (660)
As show in accuracy
description
Combined with
AXFA11
converter
Cable length
100 (330)
25 (85)
0 (0)
1
Combined with
AXFA14
converter
50
3
10
20
(S/cm)
Liquid conductivity
m (ft)
200 (660)
Cable length
As show in accuracy
description
Combined with
AXFA11
converter
100 (330)
50 (170)
25 (85)
0 (0)
Combined with
AXFA14
converter
50
1
10
20
(S/cm)
Liquid conductivity
796 (2610)
F03.EPS
600 (1970)
446 (1460)
200 ( 660)
0
20.4
22
24
26
28.8
Usable range E (V)
Cable cross section area: 1.25 mm2
Cable cross section area: 2 mm2
F01.EPS
Fluid Conductivity:
Size 2.5 to 10 mm (0.1 to 0.4 in.): 5 S/cm or larger
Size 15 to 125 mm (0.5 to 5 in.): 1 S/cm or larger
Size 150 to 400 mm (6 to 16 in.): 3 S/cm or larger
Note: Fluids with large flow noise (pure water, fluid with
low conductivity and low viscosity such as alcohol)
cause the output fluctuation. Be careful that it affects
the flow rate measurement. It is recommended to
use the CA capacitance magnetic flowmeter.
GS 01E20D01-01E
Jan. 7, 2015-00
19
<<Contents>> <<Index>>
Measurable Flow Rate Range:
SI Units (Size: mm, Flow rate: m3/h)
0 to 0.0018 m3/h
0 to 0.1767 m3/h
0 to 0.0071
0 to 0.7068
10
0 to 0.0283
0 to 2.8274
15
0 to 0.0637
0 to 6.361
25
0 to 0.1768
0 to 17.671
32
0 to 0.2896
0 to 28.952
40
0 to 0.4524
0 to 45.23
50
0 to 0.7069
0 to 70.68
65
0 to 1.1946
0 to 119.45
80
0 to 1.8096
0 to 180.95
100
0 to 2.8275
0 to 282.74
125
0 to 4.418
0 to 441.7
150
0 to 6.362
0 to 636.1
2 (290)
200
0 to 11.310
0 to 1,130.9
1 (145)
250
0 to 17.672
0 to 1,767.1
300
0 to 25.447
0 to 2,544.6
350
0 to 34.64
0 to 3,463
400
0 to 45.24
0 to 4,523
Size
(mm)
2.5
Pressure
MPa (psi)
4 (580)
0.1 (14.5)
*1
Temperature C (F)
T11.EPS
F18-1.EPS
0.1
0 to 0.0078 GPM
0 to 0.7780 GPM
0.2
0 to 0.0312
0 to 3.112
Pressure
MPa (psi)
0.4
0 to 0.1245
0 to 12.44
4 (580)
0.5
0 to 0.1946
0 to 19.45
1.0
0 to 0.7781
0 to 77.80
2 (290)
1.25
0 to 1.216
0 to 121.5
1 (145)
1.5
0 to 1.751
0 to 175.0
2.0
0 to 3.113
0 to 311.2
2.5
0 to 4.863
0 to 486.2
3.0
0 to 7.003
0 to 700.2
4.0
0 to 12.45
0 to 1,244
5.0
0 to 19.46
0 to 1,945
6.0
0 to 28.01
0 to 2,800
8.0
0 to 49.80
0 to 4,979
10
0 to 77.81
0 to 7,780
12
0 to 112.1
0 to 11,203
14
0 to 152.5
0 to 15,249
16
0 to 199.2
0 to 19,918
Size
(in.)
40
(104)
40 10 0
(40) (14) (32)
0.1 (14.5)
*1
40 10 0
( 40) (14) (32)
40
(104)
100 130
(212) (266)
Temperature C (F)
*1:
*2:
F18-2.EPS
T24.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
20
<<Contents>> <<Index>>
General-Purpose Use, Remote Flowtube (electrode
structure code 2: replaceable electrode)
25 to 200 mm (1.0 to 8.0 in.) (flange type, wafer type)
250, 300 mm (10, 12 in.) (flange type)
250, 300 mm (10, 12 in.) (wafer type)
350, 400 mm (14 to 16 in.) (flange type)
Pressure
MPa (psi)
4 (580)
3.25 (471)
2 (290)
2 (290)
1 (145)
0.1 (14.5)
10 0
(14) (32)
40
(104)
Temperature C (F)
0.1(14.5)
F18-3.EPS
Pressure
MPa (psi)
Temperature C (F)
*1:
F19-2.EPS
1 (145)
10 0
(14) (32)
40
(104)
100 130
(212) (266)
Temperature C (F)
F18-4.EPS
180
(356)
2 (290)
0.1 (14.5)
100 130
(212)(266)
10 0
(14) (32)
Pressure
MPa (psi)
4 (580)
2 (290)
1 (145)
0.75(109)
0.1(14.5)
1(145)
10 0
(14) (32)
40
(104)
Temperature C (F)
0.1(14.5)
130
(266)
10 0
(14) (32)
Temperature C (F)
160
(320)
F18-5.EPS
Ceramics Lining
General-Purpose Use and Explosion proof Type,
Remote Flowtube (electrode structure code 1: Nonreplaceable electrode)
2.5 to 50 mm (0.1 to 2.0 in.)
80 to 200 mm (3.0 to 8.0 in.)
Pressure
MPa (psi)
F20-1.EPS
Pressure
MPa (psi)
2 (290)
1 (145)
0.1(14.5)
4 (580)
10 0
(14) (32)
40
(104)
F20-2.EPS
Temperature C (F)
2 (290)
0.1 (14.5)
10 0
(14) (32)
100
(212)
Temperature C (F)
180
(356)
F19-1.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
21
<<Contents>> <<Index>>
Natural Soft Rubber Lining
100 (212)
Pressure
MPa (psi)
Decrease
60 (140)
4(580)
1(3)
5(15)
Flow Velocity m/s(ft/s)
Size 40 and 50 mm (1.5 and 2.0 in.)
TC(F)
100 (212)
90 (162)
2(290)
1(145)
*1
0.05(7.3)
10 0
(14) (32)
40
(104)
80
(175)
10(30)
Increase
Temperature C (F)
*1 : 0.04 MPa (5.7 psi) for sizes of 350 mm (14 in.) and 400 mm (16 in.)
F05-2.EPS
Decrease
30 (86)
Pressure
MPa (psi)
4(580)
50 (122)
1(3)
5(15)
Flow Velocity m/s(ft/s)
Size 80 to 200 mm (3.0 to 8.0 in.)
TC(F)
70 (158)
Decrease
1(3)
1(145)
*1
0.05(7.3)
10 0
(14) (32)
40
(104)
80
(175)
Temperature C (F)
*1 : 0.04 MPa (5.7 psi) for sizes of 350 mm (14 in.) and 400 mm (16 in.)
F05-3.EPS
Increase
50 (122)
30 (86)
2(290)
10(30)
5(15)
Flow Velocity m/s(ft/s)
10(30)
F21.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
22
<<Contents>> <<Index>>
5D or more
Tee
D: Flowtube Size
Expander
pipe
Reducer
pipe
2D
or more 0 is allowable. 0 is allowable. 10D or more 2D or more
90-degree bent
Various valves
(Correct)
(Correct)
(Incorrect)
Valve
F10.EPS
2D
5D or more 0 is allowable. 5D or more 0 is allowable.10D or more or more
F08.EPS
Mounting Orientation
If electrodes are perpendicular to the ground, air
bubbles near the top or precipitates at the bottom may
cause measurement errors.
Ensure that the terminal box of a remote flowtube and
converter of an integral flowmeter are mounted above
the piping to prevent water from entering them.
(Correct)
(Incorrect)
(Incorrect)
Air bubble
Electrode
Electrode
Precipitate
Water can
seep into the
terminal box.
F11.EPS
Mounting Orientation
Mounting Positions
Pipes must be fully filled with liquids.
It is essential that pipes remain fully filled at all times,
otherwise flow rate indications may be affected and
measurement errors may be caused.
Piping shall be designed so as to maintain the flowtube
filled with fluids.
Vertical mounting is effective in such cases as when
fluids tend to separate or solid matter may be precipitated. When employing vertical mounting, direct the
fluids from the bottom to the top to ensure that the
pipes remain fully filled.
(Correct)
(Incorrect)
(Correct)
h
h>0
(Incorrect)
h
h>0
F09.EPS
Mounting Positions
GS 01E20D01-01E
Jan. 7, 2015-00
23
<<Contents>> <<Index>>
ACCESSORIES
Centering device (wafer type only): 1 pc.
Hexagonal wrench: 2 pcs.
POWER
SUPPLY
DIO
DO
+ L/+ N/ -
Terminal configuration
Terminal wiring
Terminal
Symbols
Description
Functional grounding
N/L/+
I+
IDO+
DODIO+
DIO-
Power supply
Current output 4 to 20mA DC
Pulse output/Alarm output/
Status output
Alarm output/Status output
Status input
Protective grounding
(Outside of the terminal)
F41.EPS
Remote Flowtube
EX2
EX2
EX1
EX1
Terminal configuration
Terminal wiring
Terminal
Symbols
A
B
C
EX1
EX2
Description
Flow signal output
Excitation current input
Functional grounding
(Only for explosion proof type)
Protective grounding
(Outside of the terminal)
F42_2.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
24
<<Contents>> <<Index>>
Suffix Code
Description
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Applicable Model
Use
General-Purpose Use
Submersible Type
Explosion proof Type (*5)
Integral Flowmeter with 4 to 20mA DC Output and BRAIN Communication
Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
Integral Flowmeter with Digital communication
Communication
(FOUNDATION Fieldbus protocol) (*9)
-G Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*10)
-N Remote Flowtube for Combined Use with AXFA11
-P Remote Flowtube for Combined Use with AXFA14
G
W
C
-D
Converter
-E
Output Signal
-F
and
Power Supply
Lining (*8)
Electrode
Material (*8)
1
2
N
A
U
D
G
L
P
H
T
V
W
Electrode
Structure
Process Connection
(*3)(*11)
Lay Length
Electrical Connection (*6)
Calibration
Option
1 Non-replaceable
2 Replaceable
Indicator (*4)(*7)
N
S
L
P
H
T
V
-AA1
-AA2
-AD1
-AD2
-AD4
-AJ1
-AJ2
-AG1
None
JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Platinum-iridium
Hastelloy C276 Equivalent
Tantalum
Titanium
Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
Wafer (*1)
Wafer (*1)
Wafer (*2)
Wafer (*2)
Wafer (*1)(*2)
Wafer (*1)
Wafer (*1)
Wafer
1 Standard
-0 JIS G1/2 female
-2 ANSI 1/2 NPT female
-4 ISO M201.5 female
1 Integral Flowmeter with indicator (Horizontal)
2 Integral Flowmeter with indicator (Vertical)
N Integral Flowmeter without indicator /Remote Flowtube
B Standard
C High Grade (*12)
/ Optional code (See the Table of Optional Specifications)
Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
GS 01E20D01-01E
Jan. 7, 2015-00
25
<<Contents>> <<Index>>
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces.
1
2
N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
GS 01E20D01-01E
Jan. 7, 2015-00
26
<<Contents>> <<Index>>
Suffix Code
Description
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Applicable Model
G General-Purpose Use
C Explosion proof Type(*5)
-D
Converter
-E
Output Signal
-F
and
Communication
-G
Power Supply
Lining (*8)
1
2
N
C
Platinum-alumina Cermet
Non-replaceable
None
JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Platinum-iridium
Hastelloy C276 Equivalent
Tantalum
Titanium
ANSI Class 150
Wafer
ANSI Class 300
Wafer
Wafer (*1)
-AD1 DIN PN 10
Wafer (*1)
-AD2 DIN PN 16
Wafer (*1)
-AD4 DIN PN 40
Wafer
-AJ1 JIS 10K
Wafer
-AJ2 JIS 20K
Wafer
-AG1 JIS F12 (JIS75M)
-GUW Union Joint (Weld Joint) (*8)
-GUN Union Joint
(1/4NPT Male for 2.5 or 5 mm dia.: 3/8NPT Male for 10 mm dia.) (*8)
-GUR Union Joint (R1/4 Male for 2.5 or 5 mm dia.: R3/8 Male for 10mm dia.) (*8)
Lay Length (*3)
1 Standard
Electrical Connection (*6)
-0 JIS G1/2 female
-2 ANSI 1/2 NPT female
-4 ISO M201.5 female
1 Integral Flowmeter with indicator (Horizontal)
Indicator (*4)(*7)
2 Integral Flowmeter with indicator (Vertical)
N Integral Flowmeter without indicator /Remote Flowtube
Electrode Material (*8) E
Electrode Structure
1
N
Grounding Ring and
S
Grounding Electrode
L
Material (*8)
P
H
T
V
Process Connection
-AA1
(*2) (*11)
-AA2
Calibration
Option
B Standard
C High Grade (*12)
/ Optional code (See the Table of Optional Specifications)
GS 01E20D01-01E
Jan. 7, 2015-00
27
<<Contents>> <<Index>>
*1: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code : AD4)
because there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code : AD2) because there is no difference
in the dimensions of the mating faces.
*2: Mating dimensions are based on standards as follow:
1
2
N
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: AXF standard lay length dimension for ceramics linings are the same as
those for ADMAG ceramics linings.
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11.
Available only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*9: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*10: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*11: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*12: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
GS 01E20D01-01E
Jan. 7, 2015-00
28
<<Contents>> <<Index>>
Suffix Code
Description
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
G General-Purpose Use
W Submersible Type
C Explosion proof Type (*6)
Converter
-D Integral Flowmeter with 4 to 20 mA DC Output and BRAIN Communication
-E Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
Output Signal
-F Integral Flowmeter with Digital communication
and
(FOUNDATION Fieldbus protocol) (*10)
Communication
-G Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*11)
-N Remote Flowtube for Combined Use with AXFA11
-P Remote Flowtube for Combined Use with AXFA14
Applicable Model
Use
Power Supply
1
2
N
Lining (*9)
A
U
D
G
Electrode Material (*9) L
P
H
T
V
W
Electrode Structure
1 Non-replaceable
2 Replaceable
Process Connection
(*4)(*12)
Lay Length
Electrical Connection (*7)
Indicator (*5)(*8)
Calibration
Option
N
S
L
P
H
T
V
-BA1
-BA2
-BD1
-BD2
-BD4
-BJ1
-BJ2
-BG1
-CA1
-CA2
-CD1
-CD2
-CD4
-CJ1
-CJ2
-CG1
-DD4
-DJ1
-DJ2
1
-0
-2
-4
None
JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Platinum-iridium
Hastelloy C276 Equivalent
Tantalum
Titanium
Size 2.5 mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
Standard
JIS G1/2 female
ANSI 1/2 NPT female
ISO M201.5 female
Size 2.5mm (0.1 in.) to 200 mm (8.0 in.), PFA lining only
GS 01E20D01-01E
Jan. 7, 2015-00
29
<<Contents>> <<Index>>
*1: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: BA1, BA2, BD4, BJ1, and BJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection codes: BD4,CD4 and DD4) because
there is no difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection codes: BD2 , CD2) because there is no
difference in the dimensions of the mating faces.
*3: For a flange type of 2.5 to 10 mm (0.1 to 0.4 in)(Process connection codes: DJ1, DJ2, and DD4), prepare 10 mm (0.4 in.) diameter nominal flanges on the
process pipe side.
1
2
N
*4: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*5: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*6: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is
available only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or
G11. Available only for JIS G1/2 female electrical connections.
*7: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*8: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above).
Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
*13: In the case of calibration code C in remote type, please order a combination of flowtube and converter.
GS 01E20D01-01E
Jan. 7, 2015-00
30
<<Contents>> <<Index>>
Suffix Code
-D
Converter
-E
Output Signal
-F
and
Communication
-G
-N
-P
1
2
N
A
Power Supply
Lining (*4)
-JKB
-JSB
-KDB
-KKB
Lay Length
Description
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Integral Flowmeter/Remote Flowtube
Applicable Model
Sanitary Type
Integral Flowmeter with 4 to 20 mA DC Output and BRAIN Communication
Integral Flowmeter with 4 to 20 mA DC Output and HART Communication
Integral Flowmeter with Digital communication
(FOUNDATION Fieldbus protocol) (*5)
Integral Flowmeter with Digital communication
(PROFIBUS PA protocol) (*6)
Remote Flowtube for Combined use with AXFA11
Remote Flowtube for Combined use with AXFA14
Integral Flowmeter, 100 V to 240 V AC or 100 to 120 V DC
Integral Flowmeter, 24 V AC/DC
Remote Flowtube
Fluorocarbon PFA
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Non-replaceable
None
Tri-Clamp (3A), JIS SUS316L or SUSF316L (AISI 316L SS/EN1.4404
Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)(*1)
DIN32676 Clamp, JIS SUS316L or SUSF316L (AISI 316L SS/EN1.4404
Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
ISO2852/SMS3016 Clamp, JIS SUS316L or SUSF316L (AISI 316L SS/
EN1.4404 Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
DIN11851 Union, JIS SUS316L or SUSF316L (AISI 316L SS/EN1.4404
Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
ISO2853 Union, JIS SUS316L or SUSF316L (AISI 316L SS/EN1.4404
Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
SMS1145 Union, JIS SUS316L or SUSF316L (AISI 316L SS/EN1.4404
Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
Butt Weld for DIN 11850 Pipe Connection, JIS SUS316L or SUSF316L
(AISI 316L SS/EN1.4404 Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
Butt Weld for ISO 2037 Pipe Connection, JIS SUS316L or SUSF316L
(AISI 316L SS/EN1.4404 Equivalent or ANSI F316L SS/EN 1.4435 Equivalent)
1 Standard
-0 JIS G1/2 female
-2 ANSI 1/2 NPT female
-4 ISO M201.5 female
1 Integral Flowmeter with indicator (Horizontal)
2 Integral Flowmeter with indicator (Vertical)
N Integral Flowmeter without indicator /Remote Flowtube
Electrical Connection
Indicator (*3)
B Standard
C High Grade (*8)
/ Optional code (See the Table of Optional Specifications)
Calibration
Option
*1: For a tri-clamp type of size15 mm (0.5 in.)(Process connection code: HAB), prepare a 3/4 in. tri-clamp on the process pipe side.
*2: The detail dimensions of process connections (clamp/union/butt weld) are shown in the EXTERNAL DIMENSIONS section of the sanitary type.
In case of Butt Weld type, ferrules, sleeves, or alternative must be provided by the user. User need to weld these parts to the butt weld adapter.
*3: N shall be always selected for remote flowtubes.
In the case of an integral flowmeter, select from among the following figures.
1
*4:
*5:
*6:
*7:
*8:
T18.EPS
Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
Allowable fluid pressure should also be limited according to fluid temperature and pressure.
In the case of calibration code C in remote type, please order a combination of flowtube and converter.
GS 01E20D01-01E
Jan. 7, 2015-00
31
<<Contents>> <<Index>>
Suffix Code
Use
G
W
General-Purpose Use
Submersible Type
C
Converter
Output Signal
and
Communication
-D
-E
-F
-G
-N
-P
Power Supply
Lining (*9)
1
2
N
A
U
None
JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Platinum-iridium (*8)
Hastelloy C276 Equivalent
Tantalum (*8)
Titanium
ANSI Class 150
Wafer(*1)
ANSI Class 300
Wafer(*1)
DIN PN 10
Wafer(*2)
DIN PN 16
Wafer(*2)
DIN PN 40
Wafer(*1)(*2)
JIS 10K
Wafer(*1)
JIS 20K
Wafer(*1)
JIS F12 (JIS75M)
Wafer
N
S
L
P
H
T
V
Lay Length
Electrical Connection (*6)
Indicator (*4)(*7)
-AA1
-AA2
-AD1
-AD2
-AD4
-AJ1
-AJ2
-AG1
Process Connection
(*3)(*12)
Applicable Model
Calibration
Option
Description
AXF002
AXF005
AXF010
AXF015
AXF025
AXF040
AXF050
AXF080
AXF100
AXF150
AXF200
GS 01E20D01-01E
Jan. 7, 2015-00
32
<<Contents>> <<Index>>
*1: For a wafer type of 2.5 to 10 mm (0.1 to 0.4 in.), prepare 15 mm (0.5 in.) diameter nominal flanges on the process pipe side.
(Process connection codes: AA1, AA2, AD4, AJ1, and AJ2)
*2: Even when DIN PN10 or 16 is required for a model of size 2.5 to 50 mm (0.1 to 2.0 in.), select PN40 (Process connection code: AD4) because there is no
difference in the dimensions of the mating faces.
Even when DIN PN10 is required for a model of size 65 to 150 mm (2.5 to 6.0 in.), select PN16 (Process connection code: AD2) because there is no difference
in the dimensions of the mating faces.
1
2
N
*3: Mating dimensions are based on standards as follow:
ANSI:ASME B 16.5, DIN: DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*4: N shall be always selected for remote flowtubes
In the case of an integral flowmeter, select from among the figures at the right:
*5: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available
only for combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available
only for JIS G1/2 female electrical connections.
*6: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*7: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*8: In case of platinum-iridium (grounding ring code P) or tantalum (grounding ring code T) or None (grounding ring code N) in wafer type of 2.5 mm (0.1 in.) to 15
mm (0.5 in.), the lay lengths of Replacement model are the same as those for AXF Standard.
In this case, AXF Standard shall be selected.
*9: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is
also possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature
steam (150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*10:For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*11: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*12: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
GS 01E20D01-01E
Jan. 7, 2015-00
33
<<Contents>> <<Index>>
Suffix Code
Description
Size 150 mm (6.0 in.) Integral Flowmeter/Remote Flowtube
Size 200 mm (8.0 in.) Integral Flowmeter/Remote Flowtube
Size 250 mm (10 in.) Integral Flowmeter/Remote Flowtube
G General-Purpose Use
W Submersible Type
C Explosion proof Type (*4)
Converter
-D
Output Signal
-E
and
-F
Communication
-G
-N
-P
1
2
N
Lining (*7)
A
U
Electrode Material (*7) L
P
H
T
V
W
Power Supply
Electrode Structure
N
S
L
P
Lay Length
Electrical Connection (*5)
Indicator (*3)(*6)
Calibration
Option
None
JIS SUS316 (AISI 316 SS/EN 1.4401 Equivalent)
JIS SUS316L (AISI 316L SS/EN 1.4404 Equivalent)
Platinum-iridium
Process Connection
(*2)(*10)
1 Non-replaceable
2 Replaceable
Applicable Model
V Titanium
-CA1 ANSI Class 150
-CA2
-CD1
DIN PN 10
-CD2
DIN PN 16
-CJ1
JIS 10K
-CJ2
JIS 20K
-CG1
B Standard
/
GS 01E20D01-01E
Jan. 7, 2015-00
34
<<Contents>> <<Index>>
*1: Even when DIN PN10 is required for a 150 (6.0 in.)-mm model, select PN16 (Process connection code: CD2) because there is no difference in the dimensions of the
mating faces.
*2: Mating dimensions are based on standards as follow:
ANSI: ASME B 16.5, DIN:DIN 2501, JIS:JIS B 2220 and JIS G 3443-2
*3: N shall be always selected for remote flowtubes.
In the case of an integral, select from among the following figures:
N
1
2
*4: For explosion proof types, specify types of explosion proof certification using the optional codes. In case of TIIS flameproof type, the remote flowtube is available only for
combined use with the AXFA14. For the TIIS flameproof type with wiring using a flameproof packing adapter, select optional code G12 or G11. Available only for JIS G1/2
female electrical connections.
*5: For explosion proof types, JIS G1/2 female electrical connection is available only for TIIS flameproof type.
*6: In case of integral flowmeters of the TIIS flameproof type, select with indicator(code 1 or 2).
*7: Users must consider the characteristics of selected wetted parts material and influence of process fluids.
The use of inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant facilities. It is also
possible that the instrument itself can be damaged and that fragments from the instrument can contaminate the user's process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite, and high-temperature steam
(150C [302F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*8: For FOUNDATION Fieldbus protocol, refer to GS 01E20F02-01E
*9: For PROFIBUS PA protocol, refer to GS 01E20F12-01E.
*10: Allowable fluid pressure should also be limited according to fluid temperature and pressure.
GS 01E20D01-01E
Jan. 7, 2015-00
35
<<Contents>> <<Index>>
: Not available
Applicable Model
Item
N
AXF***G- P
D
E
AXF***C- F
G
N
AXF***C- P
N
AXF***W- P
D
E
AXF***H- F
G
N
AXF***H- P
Code
D
E
AXF***G- F
G
Specifications
Submersible
Sanitary
Explosion proof
General
Integral
Remote
Integral
Remote
Remote
Integral
Remote
Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube
DHC
Available for the submersible type and a model with optional code DHC. The
User-specified
Signal and Excitation cable length is limited up to 200 meters when combined with an AXFA11
converter, or 100 meters when combined with an AXFA14 converter.
Cable Length
Following L, specify the cable length in three digits as a multiple of 1 meter
(e.g., 001, 002, or 005) for a length up to 5 meters, or as a multiple of 5
meters (i.e., 005, 010, 015, or the like) for a length of 5 meters or more. If this
optional code is not selected, a 30-meter length cable is attached.
L***
Lightning Protector
DC Noise Cut Circuit The DC Noise Cut Circuit is built in. Available for 15 mm (0.5 in.) and larger
sizes, and for fluids with the conductivity of 50 S/cm or higher. Nullifies the
empty check and electrode adhesion diagnostic function.
ELC
C1
C2
C3
EM
MU
For District
Heating and
Cooling or
Condensation-proof
NAMUR NE43
Compliance (*9)
The output level is set to 0 mA during a CPU failure and is set 2.4 mA (-10 %)
or less during an alarm. Standard products are delivered with a setting 25 mA
during a CPU failure and 21.6 mA (110%) or more during an alarm.
Output signal
limits: 3.8 to
20.5 mA
T26-1.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
36
<<Contents>> <<Index>>
: Not available
Applicable Model
N
AXF***H- P
D
E
AXF***H- F
G
Waterproof Glands
N
AXF***W- P
electrical connections. Available only for JIS G1/2 female electric connections.
N
AXF***C- P
Waterproof glands for G3/4 conduits or flexible tubes are attached to the
Waterproof Glands
D
E
AXF***C- F
G
G3/4 Female
Code
N
AXF***G- P
Specifications
D
E
AXF***G- F
G
Item
Submersible
Sanitary
Explosion proof
General
Integral
Remote
Integral
Remote
Remote
Integral
Remote
Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube
EW
EG
EU
EP
PM
CM
SCT
RA
RB
RC
BCC
BCF
BSC
BSF
GA
GC
GD
GF
GH
Plastic Glands
Plastic glands are attached to the electrical connections. Available only for
JIS G1/2 female electric connections.
Mirror finishing on the PFA lining inside of the tube to the smoothness lining.
Lining
Mirror Finished
Ceramics
A pendant tag plate of JIS SUS304 (AISI 304 SS/EN 1.4301 equivalent) is
Plate
Direction change of
90 degrees rotated converter (or terminal box) to change the direction of
the electrical
connection (*1)
180 degrees rotated converter (or terminal box) to change the direction of
the electrical connection.
90 degrees rotated converter (or terminal box) to change the direction of
the electrical connection.
Gaskets (*2)
gaskets are
connections.
connections.
equivalent) ;
Gaskets: Chloroprene rubber (*3)
Bolts: JIS SUS304 (AISI 304 SS stainless steel
equivalent);
Nuts: JIS SUS403 (AISI 403SS stainless steel
equivalent) ;
Gaskets: PTFE-sheathed non-asbestos (*4)
Viton gaskets for use with a PFA or ceramics lining with PVC piping.
Allowable temperature and pressure are equivalent to Valqua #4010,
special fluororubber not mixed. Available for 2.5 mm (0.1 in.) to 200 mm
(8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Acid-resistant Viton gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
special fluororubber mixed (mixing #D2470). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant Viton gaskets for use with a PFA or ceramics lining with PVC
piping. Allowable temperature and pressure are equivalent to Valqua #4010,
special fluororubber mixed (mixing #RCD970). Available for 2.5 mm (0.1 in.) to 200
mm (8.0 in.) of PFA lining or 15 to 200 mm (0.5 to 8 in.) sizes of ceramics lining.
Alkali-resistant carbonized fluororesin gaskets for use with a ceramics lining with
metal piping. Allowable temperature and pressure are equivalent to Valqua #7026.
Silicon rubber gaskets for Sanitary Type, provided between the lining and
the adapter. For the condition of fluid temp. 120 to 160 C (248 to 320 F).
T26-2.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
37
<<Contents>> <<Index>>
: Not available
Applicable Model
Explosion proof
General
N
AXF***G- P
D
E
AXF***C- F
G
N
AXF***C- P
N
AXF***W- P
D
E
AXF***H- F
G
N
AXF***H- P
Oil-prohibited Use
Specifications
D
E
AXF***G- F
G
Item
Sanitary
Submersible
K1
K5
Oil-prohibited Use with Electrodes, linings, and grounding rings are assembled and packed with
Dehydrating Treatment polyethylene including desiccants after being cleaned with water and
acetone and dried with air. The label OIL & WATER FREE is affixed.
P1
P2
P7
X1
X2
Level 2: The Declaration and the Calibration Equipment List are issued.
L2
Level 3: The Declaration and the Primary Standard List are issued.
L3
L4
M01
E01
T01
WT1
WT2
PM1
PM2
PM3
PT
High Anti-corrosion
Coating
coating) in the same range as that for the standard coating. The color is
same as standard type. Salt/alkali/acid/weather-resistance.
Calibration Certificate
Hydrostatic Test
The test verifies the absence of leaks by applying the following water
pressures (which are determined under process connection conditions) to
linings for ten minutes. Test results are described in the Note column of a
test certificate (QIC).
Process Connection:
Water Pressure:
1.5 MPa
3.0 MPa
6.0 MPa
JIS F12
1.25 MPa
Withstand Voltage Test The test verifies the withstand voltage by applying the following conditions for
Certificate (*11) (*12)
Liquid Penetration Test Liquid penetration test certificate for the welded part of flanges or mini
Certificate (*16)
flanges.
Vent Hole
With a vent hole provided for permeable fluids (such as nitric acid,
hydrofluoric acid, or sodium hydroxide at high temperature).
Available only for a PFA lining flange type.
T26-3.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
38
<<Contents>> <<Index>>
: Not available
Applicable Model
and the Enhanced dual frequency excitation mode are inscribed on the
N
AXF***H- P
Although the meter factors for the Standard dual frequency excitation mode
D
E
AXF***H- F
G
N
AXF***W- P
Products are delivered with the Standard dual frequency excitation mode
(*6)
N
AXF***C- P
Frequency Excitation
D
E
AXF***C- F
G
Enhanced Dual
Code
N
AXF***G- P
Specifications
D
E
AXF***G- F
G
Item
Submersible
Explosion proof
Sanitary
General
Integral
Remote
Integral
Remote
Remote
Integral
Remote
Flowmeter Flowtube Flowmeter Flowtube Flowtube Flowmeter Flowtube
HF1
HF2
SC
FF1
KF2
CF1
SF2
(*8)
JF3
G12
(*9)
G11
name plate, the flow calibration of optional code HF1 is not performed in
Enhanced dual frequency excitation.Excitation mode select optional code
HF2 when an accurate flow measurement is required.
Available for 25 to 200 mm (1.0 to 8.0 in.) sizes.
Products are delivered with the Standard dual frequency excitation mode
and the Enhanced dual frequency excitation mode enabled.
The meter factor for the Enhanced dual frequency excitation obtained by
flow calibration is inscribed on the name plate and set into the combined
converter in addition to the meter factor for the Standard dual frequency
excitation.
Five-point Calibration
A flow test near 0, 25, 50, 75, and 100% of the user-specified span is
in User-specified
performed instead of the flow test of the standard 2m/s span and a test
Span
certificate (QIC) is submitted. Specify the span (100% flow span) whose
corresponding flow velocity lies between 0.5 to 10 m/s (0.8 to 10 m/s for
sizes 32, 65, and 125 mm) and that is less than the maximum line capacity.
Selectable range of flow rate span is showing below.
Size : mm
(in.)
2.5 (0.1)
5 (0.2)
FM Approval
0.2 (2.83)
10 (0.4)
15 (0.5)
0.32 (0.5) to
2.8 (4.40)
25 (1)
0.89 (0.5) to
11 (6.22)
32 (1.25)
40 (1.5)
50 (2)
3.54 (0.5) to
56 (7.92)
65 (2.5)
9.56 (0.8) to
80 (6.70)
80 (3)
100 (4)
125 (5)
150 (6)
200 (8)
250 (10)
300 (12)
350 (14)
400 (16)
FM Explosion proof
See HAZARDOUS AREA CLASSIFICATION
ATEX Certification
CSA Certification
IECEx
Certification
TIIS Certification
TIIS Flameproof
T26-4.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
39
<<Contents>> <<Index>>
Electrical
Integral Flowmeter Connection
Optional Code RB
Optional Code RC
Electrical
Connection
Indicator
Electrical
Connection
Remote Flowtube
Electrical
Connection
*2: When specifying the optional code BCC or BSC for a PFA or ceramics lining, it is advisable to specify the optional code GA, GC, or GD at the same time to prevent
potential leakage caused by the difference in elasticity between the flowtube and chloroprene gaskets. Refer to description of Gasket in the Wetted Part Material.
*3: Allowable temperature and pressure with the optional code BCC or BSC (only for Gaskets:Chloroprene rubber) are equivalent to Valqua #2010.
*4: Allowable temperature and pressure with the optional code BCF or BSF (only for Gaskets:PTFE-sheathed non-asbestos) are equivalent to Valqua #7030 (S).
*5: Special gaskets are inserted between the flowtube and the grounding ring or grounding electrode.
*6: Enhanced dual frequency excitation is not available for models with calibration code C (High Grade Accuracy).
*7: Select optional code G12 or G11 when TIIS Flameproof type with wiring using a flameproof packing adapter. Available only for JIS G1/2 female electric connection.
*8: The TIIS flameproof type is only available for AXF***C-P (remote flowtube for combined use with AXFA14).
*9: In the case of Fieldbus communication type, optional codes C1, C2, C3, EM and G11 are not available.
*10:In the case of size 32 mm, optional code GH is not available.
*11:Neither optional code WT1 nor WT2 is available for models with lining material "Natural Soft Rubber" or "EPDM Rubber" (Lining code: D or G).
*12:Either optional code WT1 or WT2 can be selected.
*13:In case of sanitary type or union joint type with ceramics lining, optional code PM1, PM2 or PM3 is not available as they have no grounding rings.
*14:In case of carbon steel flanges or mini flanges, the amount of Nickel, Chromium and Molybdenum in the carbon steel are not stipulated and optional code PM2 or PM3 is not
available. For material specifications, refer to description of "Flowtube Material".
*15:Only one code from PM1, PM2 or PM3 can be selected.
*16:In case of no welded portion, optional code PT is not available. For available sizes, refer to the following table.
Use
General-purpose Use
Submersible Type
Explosion proof Type
Sanitary Type
Size
Process Connection
Wafer
Flange
GS 01E20D01-01E
Jan. 7, 2015-00
40
<<Contents>> <<Index>>
EXTERNAL DIMENSIONS
AXF Standard, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube
51.5 51.5
(2.03) (2.03)
111(4.37)
Hr
70(2.76)
Hi
73(2.87) 49(1.93)
66*1
(2.6)
Remote Flowtube
Ground Terminal
(M4)
28(1.1)
128(5.04)
154(6.06)
AXF002
G
AXF005
W
AXF010
C
AXF015
197(7.76)*1
Ground Terminal
(M4)
86(3.38)
(1.89)
48
*4
D
E 1
F 2 A
G N
N
P
Integral Flowmeter
(d
3(0.12)
D
H1
63.5(2.50)
Size code
002
005
010
015
Size
2.5(0.1)
5(0.2)
10(0.4)
15(0.5)
Lining code
Model
Face-to-face
length
L*
81(3.19)
44(1.73)
Inner diameter of
Grounding ring d
15(0.59)
H1
144(5.67)
Hr
Remote Max. Height
flowtube
Weight kg (lb) *3
268(10.55)
Hi
306(12.03)
Height
L*2
72 (2.83)
58 (2.28)
4- 6.2(0.24)
P, T
+26(1.02) -2(0.08)
+0
None
2.4(5.3)
4.1(9.0)
F22.EPS
AXF Standard, AXF025-AXF125, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
Integral flowmeter Remote Flowtube Integral flowmeter Remote Flowtube
025
032
040
050
065
080
100
125
Size
25(1)
32(1.25)
40(1.5)
50(2)
65(2.5)
80(3)
100(4)
125(5)
A,U
A,U
D,G
A,U
D,G
A,U
D,G
A,U
D,G
A,U
D,G
*2
A,U
A,U
H1
Max. Height Hi
Height
Integral
Flowmeter
Weight kg (lb)
1.9(4.1)
3.6(7.8)
2.0(4.5)
3.7(8.2)
2.2(4.9)
3.9(8.7)
d)
128(5.04)
60(2.36) 70(2.76)
41(1.61) 53(2.09)
2.7(5.8)
4.4(9.6)
3.4(7.6)
4.1(9.1)
5.6(12.3) 9.3(20.4)
W*3
L*2
111(4.37)
Hr
Size code
Face-to-face
length
Ground Terminal
(M4)
28(1.1)
Lining code
51.5 51.5
(2.03) (2.03)
70
(2.76)
Model
197(7.76)*1
66*1
(2.6)
48(1.89)
1
2
Hr
D
A
E 1 U
F 2 D
G N G
N
P
Ground Terminal
(M4)
86(3.38)
154(6.06)
*5
Hi
73(2.87) 49(1.93)
AXF025
AXF032
AXF040
G
AXF050
W
AXF065
C
AXF080
AXF100
AXF125
S, L, H, V
P, T
+0
+26(1.02)
-2(0.08)
+30(1.18)
None
GA, GC, GD
+8(0.31)
(Special Gaskets)
25
32, 40, 50
65, 80
100
125
GS 01E20D01-01E
Jan. 7, 2015-00
41
<<Contents>> <<Index>>
AXF Standard, AXF150-AXF300, Wafer Type, PFA /Polyurethane Rubber /Natural Soft Rubber /EPDM
Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter
Remote Flowtube
154(6.06)
D
A
E 1 U
F 2 D
G N G
N
P
111(4.37)
(
d)
Hi
51.5 51.5
(2.03) (2.03)
(1.89)
128(5.04)
66*1
(2.6)
Ground Terminal
(M4)
28(1.1)
70(2.76)
1
2
73(2.87) 49(1.93)
48
Remote Flowtube
197(7.76)*1
Ground Terminal
(M4)
86(3.38)
*5
AXF150
G
AXF200
W
AXF250
C
AXF300
Integral Flowmeter
H1
Hr
* In case of size
250mm or 300mm,
position of grounding
ring is different
from the figure to
45-degree.
Size code
150
200
250
300
Size
150(6)
200(8)
250(10)
300(12)
A,U
D,G
A,U
D,G
A,U
D,G
A,U
D,G
200(7.87)
250(9.84)
Model
Lining code
Face-to-face
length
*2
300(11.81) 350(13.78)
Hr
Remote Max. Height
Flowtube
Weight kg (lb)*4
Height
W*3
Option
Code
16.2(35.7) 23.8(52.4)
S, L, H, V
P, T
+0
-2(0.08)
14.5(32.0) 22.1(48.7)
P, T
+32(1.26) -2(0.08)
+0
None
39.0(86.0) 48.3(106.5)
Nominal size
150
200
250
300
+49(1.93)
+50(1.97)
+49(1.93)
+53(2.09)
40.7(89.7) 50.0(110.2)
F24.EPS
Remote Flowtube
*2
154(6.06)
Remote
flowtube
Integral
flowmeter
002
005
010
002
005
010
Size
2.5
(0.1)
5
(0.2)
10
(0.4)
2.5
(0.1)
5
(0.2)
10
(0.4)
Lining code
Remote
flowtube
Face-to-face
length
140(5.51)
130(5.12)
Height
H1
144(5.67)
144(5.67)
Max. Height
Hr
268(10.55)
268(10.55)
2.3(5.1)
2.3(5.1)
306(12.03)
306(12.03)
4(8.8)
4(8.8)
Weight kg (lb)
Hi
Integral Max. Height
flowmeter
Weight kg (lb)
128(5.04)
111(4.37)
3(0.12)
72
L
4- 6.2(0.24)
(2.83)
58 (2.28)
Size code
Model
28(1.1)
70(2.76)
H1
63.5(2.50)
51.5 51.5
(2.03) (2.03)
Hr
Process connection
66*1
(2.6)
48(1.89)
W
G U N 1
R
73(2.87) 49(1.93)
D
E 1
F 2 C E 1 N
G N
N
P
Hi
AXF002
G
AXF005
C
AXF010
(M4)
197(7.76)*1
Ground Terminal
(M4)
86(3.38)
Weld joint
Union joint
D
A
4(0.16)
Size
A
2.5 22(0.87)
(0.1) 22(0.87)
5
22(0.87)
(0.2) 22(0.87)
10 25(0.98)
(0.4) 25(0.98)
B C
4(0.16)
11.5 (0.45)
30(1.18)
B
8(0.31)
8(0.31)
8(0.31)
8(0.31)
10(0.39)
10(0.39)
C
18.5(0.73)
18.5(0.73)
18.5(0.73)
18.5(0.73)
22.5(0.89)
22.5(0.89)
C D
A B
D
R1/4
NPT1/4
R1/4
NPT1/4
R3/8
NPT3/8
Size
2.5(0.1)
5(0.2)
10(0.4)
35(1.38)
A
22(0.87)
22(0.87)
25(0.98)
10 (0.39)
B
C
D
8(0.31) 14.3(0.56) 18.5(0.73)
8(0.31) 14.3(0.56) 18.5(0.73)
10(0.39) 17.8(0.70) 22.5(0.89)
F25.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
42
<<Contents>> <<Index>>
(d
H1
63.5(2.50)
3(0.12)
D
L*2
Model
Size code
015
Size
15(0.5)
Lining code
Remote
flowtube
L*
85(3.35)
Outside dia.
44(1.73)
4- 6.2(0.24)
15(0.59)
Height
H1
144(5.67)
Remote
flowtube
Max. Height
Hr
268(10.55)
Integral
flowmeter
Max. Height
Hi
306(12.03)
Weight kg (lb)
72 (2.83)
58 (2.28)
Face-to-face
length
111(4.37)
70(2.76)
D
E 1
F 2 C E 1
G N
N
P
Ground Terminal
(M4)
28(1.1)
51.5 51.5
(2.03) (2.03)
66*1
(2.6)
Hr
G
AXF015
C
Hi
73(2.87) 49(1.93)
*3
197(7.76)*1
128(5.04)
Ground Terminal
(M4)
86(3.38)
(1.89)
48
154(6.06)
+0
P, T
+22(087)
-6(0.24)
2.3(5.1)
Weight kg (lb)
4(8.8)
F26.EPS
154(6.06)
*3
Size code
025
040
050
080
100
Size
25(1)
40(1.5)
50(2)
80(3)
100(4)
Lining code
*2
99(3.90)
129(5.08) 155(6.10)
27(1.06)
40(1.57)
52(2.05)
81(3.19)
Hr
Remote Max. Height
Flowtube
Weight kg (lb)
Hi
Integral Max. Height
Flowmeter
Weight kg (lb)
3.2(7.0)
4.9(10.8)
d)
(
4.1(9.0)
+0
P, T
+22(0.87) -6(0.24)
98(3.86)
2.3(5.1)
111(4.37)
L*2
L(Face-to-face length)
H1
Height
28(1.1)
H1
Face-to-face
length
Remote
Flowtube Outside dia.
51.5 51.5
(2.6) (2.03) (2.03)
Model
197(7.76)*1
66*1
70(2.76)
Hr
Ground Terminal
(M4)
128(5.04)
48(1.89)
D
E 1
F 2 C E 1
G N
N
P
Hi
73(2.87) 49(1.93)
AXF025
AXF040
G
AXF050
C
AXF080
AXF100
6.8(15.0) 9.6(21.1)
F27.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
43
<<Contents>> <<Index>>
*3
Remote Flowtube
Ground Terminal
(M4)
86 (3.38)
154(6.06)
197(7.76)*1
66*1 51.5 51.5
(2.6) (2.03) (2.03)
111(4.37)
Hr
Hi
28(1.1)
(1.89)
70(2.76)
73 49
(2.87) (1.93)
48
128(5.04)
AXF150 G
AXF200 C
Integral Flowmeter
D
E 1
F 2 C E 1
G N
N
P
H1
(d
Model
Size code
150
200
Size
150(6)
200(8)
232(9.13)
302(11.89)
214(8.43)
264(10.39)
Lining code
Face-to-face
length
Remote
Flowtube Outside dia.
Inner diameter of
Integral
Flowmeter Grounding ring
Height
L*
144(5.67)
192(7.56)
254(10.00)
304(11.97)
Hr
378(14.88)
428(16.85)
20.2(44.5)
33.5(73.9)
416(16.36)
466(18.33)
21.9(48.3)
35.2(77.6)
Hi
L*2
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code,
add the following value to L (face-to-face length).
Grounding Ring Code S, L, H, V
H1
L(Face-to-face length)
+0
P, T
+30(1.18) -6(0.24)
F28.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
44
<<Contents>> <<Index>>
111(4.37)
(1.89)
48
t *2
N-h
Hr
28(1.1)
128(5.04)
Ground Terminal
(M4)
86(3.38)
154(6.06)
Remote Flowtube
Ground Terminal
(M4)
197(7.76)*1
70(2.76)
AXF002
G
AXF005
W
AXF010
C
AXF015
Integral Flowmeter
(H2)
D
E 1
F 2 A
G N
N
P
1
2
1
2
1
4
1
2
4
J
J
A
A
D
J
J
D
Hi
73(2.87) 49(1.93)
B
B
B
B
B
D
D
D
*4
Remote Flowtube
60.5
(2.38)
H1
D
C
(d)
Process Connection
Size code
Model
BJ1(JIS10K)
BJ2(JIS20K)
BA1(ANSI Class 150) BA2(ANSI Class 300)
BD4(DIN PN40)
DJ1(JIS10K)
DJ2(JIS20K) DD4(DIN PN40)
002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 015 002 005 010 002 005 010 002 005 010
5
10 15 2.5
5
10 15 2.5
5
10 15 2.5
2.5
5
10 15 2.5
5
10 15 2.5
5
10 2.5
5
10 2.5
5
10
(0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.5) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4) (0.1) (0.2) (0.4)
Size
Lining code
A
2
L*2
t *2
d
200
(7.87)
150(5.91)
A
200
(7.87)
150(5.91)
95(3.74)
95(3.74)
150(5.91)
A
200
(7.87)
150(5.91)
A
200
(7.87)
150(5.91)
A
200
(7.87)
150(5.91)
95(3.74)
95.3(3.75)
88.9(3.50)
90(3.54)
21 to 25
17 to 22
20
18 to 22
20 to 25
15.2
18.2
18
(0.71) (0.67 to 0.87) (0.60) (0.79 to 0.98) (0.72) (0.83 to 0.98) (0.79) (0.71 to 0.87)
20 to 24
18 to 22
16
(0.71 to 0.87) (0.63) (0.79 to 0.94)
150(5.91)
90(3.54)
90(3.54)
20 to 24
(0.79 to 0.94)
21 to 25
(0.83 to 0.98)
15(0.59)
15(0.59)
15(0.59)
15(0.59)
15(0.59)
12(0.47)
12(0.47)
12(0.47)
70(2.76)
70(2.76)
60.5(2.38)
66.5(2.62)
65(2.56)
65(2.56)
65(2.56)
60(2.36)
Integral
Hole dia.
h
Flowmeter
Number of holes N
45
45
45
45
45
45
45
45
15(0.59)
15(0.59)
15.7(0.62)
15.7(0.62)
14(0.55)
15(0.59)
15(0.59)
14(0.55)
Height
H1
Height
H2
150(5.91)
141(5.54)
141(5.54)
141(5.54)
141(5.54)
141(5.54)
141(5.54)
141(5.54)
141(5.54)
80(3.15)
80(3.15)
80(3.15)
80(3.15)
80(3.15)
80(3.15)
80(3.15)
80(3.15)
265(10.43)
265(10.43)
265(10.43)
265(10.43)
265(10.43)
265(10.43)
265(10.43)
265(10.43)
3.5
(7.7)
3.4(7.5)
5.2
(11.5)
5.1(11.2)
3.7
(8.2)
3.6(7.9)
3.2(7.1)
303(11.91)
303(11.91)
5.4
(11.9)
5.3(11.7)
3.3
(7.3)
4.9(10.8)
3.6(7.9)
3.7
(8.2)
5.0
(11.0)
5.3(11.7)
3.8(8.4)
3.9
(8.6)
303(11.91)
303(11.91)
303(11.91)
5.4
(11.9)
5.5(12.1)
5.6
(12.4)
3.3(7.3)
3.4(7.5)
3.6(7.9)
303(11.91)
303(11.91)
303(11.91)
5.0(11.0)
5.1(11.2)
5.3(11.7)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness
of flange).
L
Grounding Ring Code
Option
Code
None
S, L, H, V
+0
+0
P, T
t
N
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
GS 01E20D01-01E
F29.EPS
Jan. 7, 2015-00
45
<<Contents>> <<Index>>
AXF Standard, AXF025-AXF050, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter
51.5
28(1.1)
(2.03) (2.03)
128(5.04)
111(4.37)
t *2
N-h
)
(d
H1
51.5
66*1
(2.6)
(1.89)
70(2.76)
1
2
48
Hr
AXF025
G
AXF032
W
AXF040
C
AXF050
Ground Terminal
(M4)
197(7.76)*1
(H2)
D
A
E 1
U
2
F
D
G N
G
N
P
154(6.06)
A
J
D 4
A
J
D 4
B
B
B
C
C
C
Ground Terminal
(M4)
86(3.38)
Hi
73(2.87) 49(1.93)
*4
Remote Flowtube
D
C
L*2
Process Connection
Model
025
Size
25
(1)
32
(1.25)
Lining code
2
BJ1/CJ1
(JIS10K)
BJ1(JIS10K)
032
040
Size code
40
(1.5)
A,U
A,U
A,U
200
200
200
28
(1.10)
90
(3.54)
45
19
(0.75)
4
120
(4.74)
58
(2.28)
244
(9.62)
4.4
(9.8)
282
(11.09)
6.1
(13.5)
34
(1.34)
100
(3.94)
45
19
(0.75)
4
129
(5.08)
61
(2.40)
253
(9.96)
5.3
(11.7)
291
(11.46)
7.0
(15.5)
41
(1.61)
105
(4.13)
45
19
(0.75)
4
138
(5.43)
68
(2.67)
262
(10.31)
5.7
(12.6)
299
(11.79)
7.4
(16.4)
Integral
Hole dia.
Flowmeter Number of holes
h
N
Remote
C
Flowtube Pitch circle dia.
Bolt hole interval
Height
H1
Height
H2
BJ2(JIS20K)
032
040
050
025
50
(2)
A,U
D,G
200
(7.87)
155
(6.10)
20
(0.79)
53
(2.09)
120
(4.72)
45
19
(0.75)
4
157
(6.16)
79
(3.11)
281
(11.04)
6.8
(14.9)
318
(12.52)
8.5
(18.6)
25
(1)
32
(1.25)
40
(1.5)
A,U
A,U
A,U
200
(7.87)
125
(4.92)
20
(0.79)
28
(1.10)
90
(3.54)
45
19
(0.75)
4
120
(4.74)
58
(2.28)
244
(9.62)
4.8
(10.5)
282
(11.09)
6.5
(14.3)
200
(7.87)
135
(5.31)
22
(0.87)
34
(1.34)
100
(3.94)
45
19
(0.75)
4
129
(5.08)
61
(2.40)
253
(9.96)
5.7
(12.6)
291
(11.46)
7.4
(16.4)
200
(7.87)
140
(5.51)
22
(0.87)
41
(1.61)
105
(4.13)
45
19
(0.75)
4
138
(5.43)
68
(2.67)
262
(10.31)
6.2
(13.6)
299
(11.79)
7.9
(17.4)
BJ2/CJ2
(JIS20K)
BA1/CA1
050
025
50
(2)
A,U
D,G
200
(7.87)
155
(6.10)
22
(0.87)
53
(2.09)
120
(4.72)
22.5
19
(0.75)
8
157
(6.16)
79
(3.11)
281
(11.04)
7.0
(15.4)
318
(12.52)
8.7
(19.1)
25
(1)
040
050
025
32
40
(1.25) (1.5)
50
(2)
25
(1)
032
A,U
A,U
A,U
200
(7.87)
108.0
(4.25)
18.2
(0.72)
28
(1.10)
79.2
(3.12)
45
15.7
(0.62)
4
112
(4.40)
58
(2.28)
236
(9.28)
3.9
(8.5)
273
(10.76)
5.6
(12.2)
200
(7.87)
117.3
(4.62)
19.7
(0.78)
34
(1.34)
88.9
(3.50)
45
15.7
(0.62)
4
120
(4.72)
61
(2.40)
244
(9.61)
4.5
(9.9)
282
(11.10)
6.2
(13.6)
200
(7.87)
127.0
(5.00)
21.5
(0.85)
41
(1.61)
98.6
(3.88)
45
15.7
(0.62)
4
131
(5.17)
68
(2.67)
255
(10.05)
5.4
(11.9)
293
(11.53)
7.1
(15.7)
A,U
D,G
200
(7.87)
152.4
(6.00)
23.1
(0.91)
53
(2.09)
120.7
(4.75)
45
19.1
(0.75)
4
155
(6.11)
79
(3.11)
279
(10.99)
7.4
(16.4)
317
(12.47)
9.1
(20.1)
BA2/CA2
(ANSI Class 300)
040
050
32
40
(1.25) (1.5)
50
(2)
032
A,U
A,U
A,U
200
(7.87)
124.0
(4.88)
21.5
(0.85)
28
(1.10)
88.9
(3.50)
45
19.1
(0.75)
4
120
(4.72)
58
(2.28)
244
(9.60)
5.0
(11.0)
281
(11.07)
6.7
(14.7)
200
(7.87)
133.4
(5.25)
23.1
(0.91)
34
(1.34)
98.6
(3.88)
45
19.1
(0.75)
4
128
(5.04)
61
(2.40)
252
(9.92)
5.8
(12.9)
290
(11.42)
7.5
(16.6)
200
(7.87)
155.4
(6.12)
24.6
(0.97)
41
(1.61)
114.3
(4.50)
45
22.4
(0.88)
4
146
(5.73)
68
(2.67)
270
(10.61)
7.8
(17.1)
307
(12.09)
9.5
(20.8)
A,U
D,G
200
(7.87)
165.1
(6.50)
26.4
(1.04)
53
(2.09)
127.0
(5.00)
22.5
19.1
(0.75)
8
162
(6.36)
79
(3.11)
286
(11.24)
9.0
(19.8)
323
(12.72)
10.7
(23.6)
BD4(DIN PN40)
025
25
(1)
A,U
032 040
32
40
(1.25) (1.5)
A,U
A,U
200
200
200
(7.87) (7.87) (7.87)
150
140
115
(4.53) (5.51) (5.91)
22
22
22
(0.87) (0.87) (0.87)
41
34
28
(1.10) (1.34) (1.61)
110
100
85
(3.35) (3.94) (4.33)
45
45
45
18
18
14
(0.55) (0.71) (0.71)
4
4
4
143
131
115
(4.54) (5.16) (5.63)
68
61
58
(2.28) (2.40) (2.67)
267
255
239
(9.42) (10.04) (10.51)
6.9
6.1
4.7
(10.4) (13.4) (15.2)
304
293
277
(10.90) (11.54) (11.98)
8.6
7.8
6.4
(14.1) (17.2) (19.0)
BD4/CD4
(DIN PN40)
050
50
(2)
A,U
D,G
200
(7.87)
165
(6.50)
24
(0.94)
53
(2.09)
125
(4.92)
45
18
(0.71)
4
162
(6.36)
79
(3.11)
286
(11.24)
8.7
(19.2)
323
(12.72)
10.4
(22.9)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness
of flange).
L
Grounding Ring Code
Option
Code
None
S, L, H, V
+0
+0
P, T
t
N
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F30.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
46
<<Contents>> <<Index>>
AXF Standard, AXF065-AXF125, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter
Remote Flowtube
Integral Flowmeter
Remote Flowtube
Ground Terminal
(M4)
Ground Terminal
(M4)
154(6.06)
2
1
Model
BJ2/CJ2(JIS20K)
Size code
065
080
100
125
065
080
Size
Lining code
65
(2.5)
A,U
D,G
200
(7.87)
175
(6.89)
22
(0.87)
66
(2.60)
140
(5.51)
45
19
(0.75)
4
176
(6.93)
89
(3.50)
300
(11.81)
9.0
(19.8)
338
(13.31)
10.7
(23.5)
80
(3)
A,U
D,G
200
(7.87)
185
(7.28)
22
(0.87)
77
(3.03)
150
(5.91)
22.5
19
(0.75)
8
187
(7.36)
95
(3.74)
311
(12.24)
9.6
(21.2)
349
(13.74)
11.3
(25.0)
100
(4)
A,U
D,G
250
(9.84)
210
(8.27)
22
(0.87)
102
(4.02)
175
(6.89)
22.5
19
(0.75)
8
211
(8.30)
106
(4.16)
335
(13.18)
12.4
(27.3)
372
(14.65)
14.1
(31.0)
125
(5)
A,U
D,G
250
(9.84)
250
(9.84)
24
(0.94)
128
(5.04)
210
(8.27)
22.5
23
(0.91)
8
245
(9.65)
120
(4.73)
369
(14.53)
17.4
(38.3)
407
(16.01)
19.1
(42.1)
65
(2.5)
A,U
D,G
200
(7.87)
175
(6.89)
24
(0.94)
66
(2.60)
140
(5.51)
22.5
19
(0.75)
8
176
(6.93)
89
(3.50)
300
(11.81)
9.3
(20.5)
338
(13.31)
11.0
(24.3)
80
(3)
A,U
D,G
200
(7.87)
200
(7.87)
26
(1.02)
77
(3.03)
160
(6.30)
22.5
23
(0.91)
8
195
(7.68)
95
(3.74)
319
(12.56)
12.4
(27.3)
357
(14.06)
14.1
(31.0)
Thickness
Inner diameter of
Remote Grounding ring
t*
d
BJ1/CJ1(JIS10K)
h
Number of holes N
Hole dia.
Height
H1
Height
H2
N-h
D
t *2
C
(d)
L*2
111(4.37)
128(5.04)
70(2.76)
28(1.1)
(H2)
Hr
1
2
(3.38)
(1.89)
H1
AXF065
G
AXF080
W
AXF100
C
AXF125
A
D
J
G
A
D
J
G
73(2.87) 49(1.93)
D
A
E 1
U
F 2
D
G N
G
N
P
Hi
B
B
B
B
C
C
C
C
*4
86
48
197(7.76)*1
66*1 51.5 51.5
(2.6) (2.03) (2.03)
100
125
BG1/CG1(JIS F12)
080
100
125
80
(4)
(5)
(3)
A,U
A,U A,U
D,G
D,G D,G
250
200
250
(9.84) (9.84) (7.87)
225
211
270
(8.86) (10.63) (8.31)
28
22
30
(1.10) (1.18) (0.87)
102
77
128
(4.02) (5.04) (3.03)
185
168
225
(7.28) (8.86) (6.61)
22.5 22.5
45
23
19
25
(0.91) (0.98) (0.75)
8
4
8
218
200
255
(8.59) (10.04) (7.87)
106
95
120
(4.16) (4.73) (3.74)
342
324
379
(13.47) (14.92) (12.76)
16.9 24.7 12.2
(37.3) (54.5) (26.9)
380
417
362
(14.95) (16.40) (14.25)
18.6 26.4 13.9
(41.0) (58.3) (30.7)
BD2/CD2(DIN PN16)
100
125
065
080
100
125
065
065
080
100
125
100
(4)
A,U
D,G
250
(9.84)
238
(9.37)
22
(0.87)
102
(4.02)
195
(7.68)
45
19
(0.75)
4
225
(8.85)
106
(4.16)
349
(13.73)
15.5
(34.2)
386
(15.21)
17.2
(37.8)
125
(5)
A,U
D,G
250
(9.84)
263
(10.35)
24
(0.94)
128
(5.04)
220
(8.66)
30
19
(0.75)
6
252
(9.90)
120
(4.73)
376
(14.79)
19.5
(43.1)
413
(16.26)
21.2
(46.8)
65
(2.5)
A,U
D,G
200
(7.87)
177.8
(7.00)
26.4
(1.04)
66
(2.60)
139.7
(5.50)
80
(3)
A,U
D,G
200
(7.87)
190.5
(7.50)
27.9
(1.10)
77
(3.03)
152.4
(6.00)
100
(4)
A,U
D,G
250
(9.84)
228.6
(9.00)
27.9
(1.10)
102
(4.02)
190.5
(7.50)
125
(5)
A,U
D,G
250
(9.84)
254.0
(10.00)
27.9
(1.10)
128
(5.04)
215.9
(8.50)
65
(2.5)
A,U
D,G
200
(7.87)
190.5
(7.50)
29.4
(1.16)
66
(2.60)
149.4
(5.88)
080
80
100
125
(3)
(4)
(5)
A,U
A,U
A,U
D,G
D,G
D,G
200
250
250
(7.87) (9.84) (9.84)
209.6 254.0 279.4
(8.25) (10.00) (11.00)
32.4
35.8
39.1
(1.28) (1.41) (1.54)
77
102
128
(3.03) (4.02) (5.04)
168.1 200.2 235.0
(6.62) (7.88) (9.25)
100
65
(2.5)
A,U
D,G
200
(7.87)
185
(7.28)
22
(0.87)
66
(2.60)
145
(5.71)
80
(3)
A,U
D,G
200
(7.87)
200
(7.87)
24
(0.94)
77
(3.03)
160
(6.30)
100
(4)
A,U
D,G
250
(9.84)
220
(8.66)
24
(0.94)
102
(4.02)
180
(7.09)
125
(5)
A,U
D,G
250
(9.84)
250
(9.84)
26
(1.02)
128
(5.04)
210
(8.27)
45
19.1
(0.75)
45
19.1
(0.75)
22.5
19.1
(0.75)
22.5
22.4
(0.88)
22.5
22.4
(0.88)
22.5
22.4
(0.88)
45
18
(0.71)
22.5
22.4
(0.88)
125
22.5
18
(0.71)
4
4
8
8
8
8
8
8
8
4
177
190
220
184
200
233
195
247
260
181
(6.97) (7.48) (8.66) (9.72) (7.24) (7.87) (9.16) (10.22) (7.13) (7.68)
89
95
106
89
95
106
95
120
120
89
(3.50) (3.74) (4.16) (4.72) (3.50) (3.74) (4.16) (4.72) (3.50) (3.74)
301
314
344
308
324
357
319
371
384
305
(11.85) (12.36) (13.54) (14.61) (12.13) (12.76) (14.04) (15.11) (12.01) (12.56)
10.8
12.9
17.7
12.6
16.6
26.8
11.9
20.8
9.8
34.9
(23.7) (28.5) (39.1) (45.9) (27.7) (36.6) (59.1) (76.9) (21.6) (26.2)
339
352
382
346
362
357
394
409
343
421
(13.35) (13.86) (15.02) (16.08) (13.62) (14.25) (15.52) (16.58) (13.50) (14.06)
12.5
14.6
19.4
14.3
18.3
13.6
28.5
22.5
11.5
36.6
(27.5) (32.2) (42.8) (49.6) (31.4) (40.4) (62.8) (80.7) (25.4) (29.9)
22.5
22.4
(0.88)
22.5
22.5
18
18
(0.71) (0.71)
8
8
245
216
(8.49) (9.65)
120
106
(4.16) (4.72)
369
340
(13.37) (14.53)
19.3
14.5
(32.0) (42.5)
407
377
(14.85) (16.00)
21.0
16.2
(35.7) (46.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness of
flange).
L
Grounding Ring Code
Option
Code
None
S, L, H, V
+0
+0
P, T
t
N
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F31.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
47
<<Contents>> <<Index>>
AXF Standard, AXF150, AXF200, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
B
B
D
A
B
E 1
G
AXF150
U
1
B
F
W
2
AXF200
D
2
C
G N
C
G
C
N
C
P
*4: D, E, F, G; Integral Flowmeter, C
*4
N, P;
A
D
J
G 1
1
A
D
J
G 1
Remote Flowtube
Integral Flowmeter
Integral Flowmeter
Remote Flowtube
Ground Terminal
(M4)
86 (3.38)
154(6.06)
48
111(4.37)
128(5.04)
t *2
N-h
(H2)
H1
Hr
Hi
70
(2.76)
28(1.1)
(1.89)
49 (1.93)
73
(2.87)
Remote Flowtube
Ground Terminal
(M4)
197(7.76)*1
d)
L*2
Process Connection BJ1/CJ1(JIS10K) BJ2/CJ2(JIS20K) BG1/CG1(JIS F12) BA1/CA1(ANSI Class 150) BA2/CA2(ANSI Class 300) BD1/CD1(DIN PN10) BD2/CD2(DIN PN16)
Size code
Model
Size
Lining code
Face-to-face
length
0 2
L -3 *
Outside dia.
D
2
t*
d
Thickness
Inner diameter of
h
N
Height
H1
Height
H2
Hr
Remote Max. Height
flowtube
Weight kg (lb) *3
Hi
Integral Max. Height
flowmeter
Weight kg (lb)
150
200
150
200
(6)
(8)
A,U
A,U
D,G
D,G
300
350
(11.81) (13.78)
280
330
(11.02) (12.99)
27
27
(1.06) (1.06)
146.1 193.6
(5.75) (7.62)
240
290
(9.45) (11.42)
22.5
15
23
23
(0.91) (0.91)
8
12
281
331
(11.06) (13.03)
141
166
(5.55) (6.54)
405
455
(15.94) (17.91)
27.8 37.3
(61.3) (82.2)
443
493
(17.42) (19.39)
29.5 39.0
(65.0) (86.0)
150
150
(6)
A,U
D,G
300
(11.81)
305
(12.01)
33
(1.30)
146.1
(5.75)
260
(10.24)
15
25
(0.98)
12
294
(11.56)
141
(5.55)
418
(16.44)
37.1
(81.8)
456
(17.95)
38.8
(85.5)
200
200
(8)
A,U
D,G
350
(13.78)
350
(13.78)
35
(1.38)
193.6
(7.62)
305
(12.01)
15
25
(0.98)
12
341
(13.43)
166
(6.54)
465
(18.31)
51.9
(114.4)
503
(19.80)
53.6
(118.2)
150
200
150
200
(6)
(8)
A,U
A,U
D,G
D,G
300
350
(11.81) (13.78)
290
342
(11.42) (13.46)
27
29
(1.06) (1.14)
146.1 193.6
(5.75) (7.62)
247
299
(9.72) (11.77)
30
22.5
19
19
(0.75) (0.75)
6
8
286
337
(11.26) (13.27)
141
166
(5.55) (6.54)
410
461
(16.14) (18.15)
29.9 43.2
(65.9) (95.3)
448
499
(17.64) (19.65)
31.6
44.9
(69.7) (99.0)
150
150
(6)
A,U
D,G
300
(11.81)
279.4
(11.00)
30.4
(1.20)
146.1
(5.75)
241.3
(9.50)
22.5
22.4
(0.88)
8
281
(11.05)
141
(5.55)
405
(15.93)
30.9
(68.0)
443
(17.43)
32.6
(71.8)
200
200
(8)
A,U
D,G
350
(13.78)
342.9
(13.50)
33.4
(1.31)
193.6
(7.62)
298.5
(11.75)
22.5
22.4
(0.88)
8
337
(13.29)
166
(6.54)
461
(18.17)
49.2
(108.4)
499
(19.66)
50.9
(112.2)
150
150
(6)
A,U
D,G
300
(11.81)
317.5
(12.50)
43.5
(1.71)
146.1
(5.75)
269.7
(10.62)
15
22.4
(0.88)
12
300
(11.80)
141
(5.55)
424
(16.68)
52.5
(115.7)
462
(18.18)
54.2
(119.5)
200
200
(8)
A,U
D,G
350
(13.78)
381.0
(15.00)
46.1
(1.81)
193.6
(7.62)
330.2
(13.00)
15
25.4
(1.00)
12
357
(14.04)
166
(6.54)
481
(18.92)
78.8
(173.7)
519
(20.41)
80.5
(177.5)
200
200
(8)
A,U
D,G
350
(13.78)
340
(13.39)
29
(1.14)
193.6
(7.62)
295
(11.61)
22.5
22
(0.87)
8
336
(13.23)
166
(6.54)
460
(18.11)
42.5
(93.7)
498
(19.61)
44.2
(97.5)
150
150
(6)
A,U
D,G
300
(11.81)
285
(11.22)
27
(1.06)
146.1
(5.75)
240
(9.45)
22.5
22
(0.87)
8
284
(11.16)
141
(5.55)
408
(16.04)
28.7
(63.2)
446
(17.54)
30.4
(66.9)
200
200
(8)
A,U
D,G
350
(13.78)
340
(13.39)
29
(1.14)
193.6
(7.62)
295
(11.61)
15
22
(0.87)
12
336
(13.23)
166
(6.54)
460
(18.11)
41.9
(92.5)
498
(19.61)
43.6
(96.2)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length)
and t (thickness of flange).
L
Grounding Ring Code
Option
Code
None
S, L, H, V
+0
+0
P, T
t
N
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
GS 01E20D01-01E
F32.EPS
Jan. 7, 2015-00
48
<<Contents>> <<Index>>
AXF Standard, AXF250-AXF400, JIS/ANSI/DIN Flange Type, PFA /Polyurethane Rubber /Natural Soft
Rubber /EPDM Rubber Lining
*4
AXF250
G
AXF300
W
AXF350
C
AXF400
D
A
E 1
U
F 2
D
G N
G
N
P
1
2
B
B
B
B
C
C
C
C
A
D
J
G 1
1
A
D
J
G 1
Integral Flowmeter
Integral Flowmeter
Ground Termonal
(M4)
Remote Flowtube
154(6.06)
86
Ground Terminal
(M4)
197(7.76)*1
(3.38)
(2.76)
28(1.1)
70
(4.37)
111
128(5.04)
Eye Bolt
N- h
D
C
H3
Hr
H1
Hi
(H2)
73
49 (1.93)
48(1.89)
(2.87)
Remote Flowtube
(d
)
t*2
L*2
250
Size code
BJ1/CJ1(JIS10K)
300
350
400
250
300
BG1/CG1(JIS F12)
250
300
350
400
300
350
400
BA2/CA2
(ANSI Class 300)
250
300
BD1/CD1(DIN PN10)
250
300
350
400
BD2/CD2
(DIN PN16)
250
300
300
250
300
350
400
250
300
350
250
350
400
250
300
250
300
350
400
250
300
400
250
300
(12) (14)
(16) (10)
(12) (10) (12) (14)
(10) (12) (14) (16) (10) (12) (10) (12) (14) (16) (10)
(16) (10)
(12)
A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U A,U
Lining code
D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G D,G
500
450
500
550
600
450
500
550
450
550
600
450
500
450
500
550
600
450
500
600
450
500
Face-to-face
0 2
L -5 * (17.72) (19.69) (21.65) (23.62) (17.72) (19.69) (17.72) (19.69) (21.65) (23.62) (17.72) (19.69) (21.65) (23.62) (17.72) (19.69) (17.72) (19.69) (21.65) (23.62) (17.72) (19.69)
length
445
505
400
490
560
430
480
410
464
530
582 406.4 482.6 533.4 596.9 444.5 520.7 395
445
565
405
460
Outside dia.
D (15.75) (17.52) (19.29) (22.05) (16.93) (18.90) (16.14) (18.27) (20.87) (22.91) (16.00) (19.00) (21.00) (23.50) (17.50) (20.50) (15.55) (17.52) (19.88) (22.24) (15.94) (18.11)
34
38.2 39.8 45.1 46.6 55.7 58.8
36
32
36
38
42
44
32
34
36
36
2
34
34
36
34
36
Thickness
t * (1.26) (1.34) (1.42) (1.50) (1.65) (1.73) (1.26) (1.34) (1.42) (1.42) (1.50) (1.57) (1.78) (1.83) (2.19) (2.31) (1.34) (1.34) (1.42) (1.42) (1.34) (1.42)
243 291.3 323.4 373.5 243 291.3 243 291.3 323.4 373.5 243 291.3 323.4 373.5 243 291.3 243 291.3 323.4 373.5 243 291.3
Inner diameter of
d
(9.57) (11.47) (12.73) (14.70) (9.57) (11.47) (9.57) (11.47) (12.73) (14.70) (9.57) (11.47) (12.73) (14.70) (9.56) (11.47) (9.57) (11.47) (12.73) (14.70) (9.57) (11.47)
Remote Grounding ring
400
460
355
445
510
380
430
360
414
472
524 362.0 431.8 476.3 539.8 387.4 450.9 350 400
515
355
410
flowtube Pitch circle dia.
C (13.98) (15.75) (17.52) (20.08) (14.96) (16.93) (14.17) (16.30) (18.58) (20.63) (14.25) (17.00) (18.75) (21.25) (15.25) (17.75) (13.78) (15.75) (18.11) (20.28) (13.98) (16.14)
15
15
15
11.25 11.25 11.25
11.25 11.25
15 11.25 11.25 11.25 15 11.25 22.5
18
18
15
15
15
15
15
Bolt hole interval
Size
Integral
flowmeter
Remote
flowtube
Hole dia.
Number of holes
Height
H1
Height
H2
Height
H3
Max. Height
Hr
3
Weight kg (lb)*
Max. Height
Hi
Integral
flowmeter
Weight kg (lb)
25
25.4 25.4 28.4 28.4 28.4 31.8
22
25
25
27
27
27
23
23
25
25
22
22
26
26
26
(0.98) (0.98) (0.98) (1.06) (1.06) (1.06) (0.91) (0.91) (0.98) (0.98) (1.00) (1.00) (1.12) (1.12) (1.12) (1.25) (0.87) (0.87) (0.87) (1.02) (1.02) (1.02)
16
12
12
12
16
16
16
16
12
16
16
12
16
8
10
10
12
12
12
16
12
12
447
403
466
512
572
422
485
498
400
491
553
415
464
405
456
511
564
397
447
556
402
454
(15.75) (17.60) (19.33) (21.77) (16.34) (18.27) (15.94) (17.95) (20.12) (22.20) (15.87) (18.35) (20.16) (22.52) (16.61) (19.09) (15.63) (17.60) (19.61) (21.89) (15.83) (17.87)
221
197
221
243
270
197
221
243
197
243
270
197
221
197
221
243
270
197
221
270
197
221
(7.76) (8.70) (9.57) (10.63) (7.76) (8.70) (7.76) (8.70) (9.57) (10.63) (7.76) (8.70) (9.57) (10.63) (7.76) (8.70) (7.76) (8.70) (9.57) (10.63) (7.76) (8.70)
499
460
537
596
660
499
575
568
454
553
623
484
534
464
518
593
645
449
499
628
459
514
(17.87) (19.65) (21.77) (24.53) (19.06) (21.02) (18.27) (20.39) (23.35) (25.39) (18.11) (21.14) (23.46) (25.98) (19.65) (22.64) (17.68) (19.65) (22.36) (24.72) (18.07) (20.24)
571
527
590
636
696
546
609
622
524
615
677
539
588
529
580
635
688
521
571
680
526
578
(20.63) (22.48) (24.21) (26.65) (21.22) (23.15) (20.83) (22.83) (25.00) (27.09) (20.75) (23.23) (25.04) (27.40) (21.50) (23.98) (20.51) (22.48) (24.49) (26.77) (20.71) (22.76)
70.0 78.0 107.0 135.0 98.5 114.5 73.4 85.0 121.2 137.0 83.4 104.8 151.8 185.3 133.0 176.7 73.0 79.4 112.5 129.7 74.8 87.9
(154.3) (172.0) (235.9) (297.6) (217.2) (252.4) (161.7) (187.4) (267.2) (301.9) (183.8) (231.0) (334.7) (408.5) (293.1) (389.5) (161.0) (174.9) (248.0) (285.9) (164.9) (193.8)
609 653
565
628
674
734
584
647
660
562
715
577
626
567
618
673
726
559
609
718
564
616
(22.13) (23.98) (25.71) (28.15) (22.72) (24.65) (22.32) (24.33) (26.50) (28.58) (22.24) (24.72) (26.54) (28.90) (22.99) (25.47) (22.01) (23.98) (25.98) (28.27) (22.20) (24.25)
71.7 79.7 108.7 136.7 100.2 116.2 75.1 86.7 122.9 138.7 85.1 106.5 153.5 187.0 134.7 178.4 74.7 81.1 114.2 131.4 76.5 89.6
(158.1) (175.7) (239.6) (301.4) (220.9) (256.2) (165.5) (191.2) (270.9) (305.7) (187.6) (234.8) (338.4) (412.3) (296.9) (393.3) (164.7) (178.7) (251.7) (289.6) (168.7) (197.6)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length)
and t (thickness of flange).
Nominal Size: 250 mm to 300 mm
L
t
L
Grounding Ring Code
Option Code is None
S, L, H, V
+0
+0
t
N
-6(0.24) -3(0.12)
S, L, H, V
+0
+0
t
N
-10(0.39) -5(0.20)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
All Rights Reserved. Copyright 2003, Yokogawa Electric Corporation
GS 01E20D01-01E
F33.EPS
Jan. 7, 2015-00
49
<<Contents>> <<Index>>
154(6.06)
015
025
032
040
050
065
080
100
125
Size
15
(0.5)
25
(1)
32
(1.3)
40
(1.5)
50
(2)
65
(2.6)
80
(3)
100
(4)
125
(5)
HKB
Height
H1
Max. Height
Hr
2
Weight kg (lb)*
Max. Height
Hi
Weight kg (lb)
246
(9.70)
155
(6.10)
97.4
(3.83)
100
(3.94)
97.6
(3.84)
183
(7.20)
307
(12.09)
8.1
(17.9)
345
(13.56)
9.8
(21.6)
316
(12.46)
183
(7.20)
HAB (TriClamp)
25
70
(2.76)
22.2
B
(0.87)
25.4
C
(1.00)
22.1
d
(0.87)
43.6
E
(1.72)
50.4
25
F
(0.98) (1.98)
50
50
G
(1.97) (1.97)
F9811 F9811
Parts No. HU
HV
125
(4.92)
135.7
(5.34)
212
(8.35)
336
(13.23)
12.1
(26.7)
374
(14.70)
13.8
(30.4)
Integral
Inner dia. d HDB
Flowmeter
A
216
(8.52)
129
(5.08)
72.9
(2.87)
81
(3.19)
72.9
(2.87)
157
(6.18)
281
(11.06)
5.7
(12.6)
319
(12.54)
7.4
(16.3)
HAB
A
196
(7.73)
117
(4.61)
60.2
(2.37)
66
(2.60)
60.3
(2.37)
147
(5.79)
271
(10.67)
4.8
(10.6)
309
(12.15)
6.5
(14.3)
176
166
166
(6.55) (6.55) (6.94)
99
86
73
(2.87) (3.39) (3.90)
47.5
34.8
(1.37) (1.87)
50
38
32
(1.26) (1.50) (1.97)
48.6
35.6
31.3
(1.23) (1.40) (1.91)
129
111
98
(3.86) (4.37) (5.08)
253
235
222
(8.74) (9.25) (9.96)
3.6
2.9
2.6
(5.7) (6.4) (7.9)
291
273
260
(10.22) (10.73) (11.44)
5.3
4.6
4.3
(9.5) (10.1) (11.7)
Outside dia.
Remote
Flowtube
A
166
(6.55)
73
(2.87)
22.1
(0.87)
26
(1.02)
22.6
(0.89)
98
(3.86)
222
(8.74)
2.5
(5.5)
260
(10.22)
4.2
(9.3)
A
166
(6.55)
73
(2.87)
15.7
(0.62)
16
(0.63)
15.2
(0.60)
98
(3.86)
222
(8.74)
2.7
(6.0)
260
(10.22)
4.4
(9.7)
(d)
Face-to-face
length
70
(2.76)
16
(0.63)
19.2
(0.76)
15.7
(0.62)
128(5.04)
Size code
Lining code
111(4.37)
H1
Model
15
51.5 51.5
(2.03) (2.03)
Hr
Process Connection
Process
Connection
Nominal
Size
Ground Terminal
(M4)
28(1.1)
Integral
Flowmeter
(2.6)
70(2.76)
H A B
H D B 1
H K B
Remote
Flowtube
66*1
48
(1.89)
49
73
(2.87) (1.93)
D
E 1
F 2 A L 1 N
G N
N
P
Ground Terminal
(M4)
86(3.38)
*3
Hi
AXF015
AXF025
AXF032
AXF040
AXF050 H
AXF065
AXF080
AXF100
AXF125
Remote Flowtube
40
50
65
80
100
15
25
32
40
50
65
80
100
125
15
25
32
40
50
65
80
100
125
83
(3.27)
34.6
(1.36)
38.1
(1.50)
34.8
(1.37)
43.6
(1.72)
50.4
(1.98)
50
(1.97)
F9811
HX
96
(3.78)
47.6
(1.87)
50.8
(2.00)
47.5
(1.87)
56.3
(2.22)
64
(2.52)
50
(1.97)
F9811
HY
114
(4.49)
59.5
(2.34)
63.5
(2.50)
60.2
(2.37)
70.6
(2.78)
77.4
(3.05)
50
(1.97)
F9811
HZ
126
(4.96)
72.3
(2.85)
76.2
(3.00)
72.9
(2.87)
83.3
(3.28)
91
(3.58)
50
(1.97)
F9811
JA
152
(5.98)
97
(3.82)
101.6
(4.00)
97.4
(3.83)
110.3
(4.34)
118.9
(4.68)
50
(1.97)
F9811
JB
70
(2.76)
16
(0.63)
20
(0.79)
16
(0.63)
27.5
(1.08)
34
(1.34)
50
(1.97)
F9811
JD
70
(2.76)
22.2
(0.87)
30
(1.18)
26
(1.02)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JE
70
(2.76)
29.4
(1.16)
36
(1.42)
32
(1.26)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JF
83
(3.27)
34.6
(1.36)
42
(1.65)
38
(1.50)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JG
96
(3.78)
47.6
(1.87)
54
(2.13)
50
(1.97)
56.5
(2.22)
64
(2.52)
50
(1.97)
F9811
JH
114
(4.49)
59.5
(2.34)
70
(2.76)
66
(2.60)
83.5
(3.29)
91
(3.58)
50
(1.97)
F9811
JJ
126
(4.96)
72.3
(2.85)
85
(3.35)
81
(3.19)
97
(3.82)
106
(4.17)
50
(1.97)
F9811
JK
152
(5.98)
97
(3.82)
104
(4.09)
100
(3.94)
110
(4.33)
119
(4.69)
50
(1.97)
F9811
JL
180
(7.09)
123
(4.84)
129
(5.08)
125
(4.92)
146
(5.75)
155
(6.10)
60
(2.36)
F9811
JM
70
(2.76)
16
(0.63)
18
(0.71)
15.2
(0.60)
27.5
(1.08)
34
(1.34)
50
(1.97)
F9811
JN
70
(2.76)
22.2
(0.87)
25.6
(1.01)
22.6
(0.89)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JP
70
(2.76)
29.4
(1.16)
34.3
(1.35)
31.3
(1.23)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JQ
83
(3.27)
34.6
(1.36)
38.6
(1.52)
35.6
(1.40)
43.5
(1.71)
50.5
(1.99)
50
(1.97)
F9811
JR
96
(3.78)
47.6
(1.87)
51.6
(2.03)
48.6
(1.91)
56.5
(2.22)
64
(2.52)
50
(1.97)
F9811
JS
114
(4.49)
59.5
(2.34)
64.1
(2.52)
60.3
(2.37)
70.5
(2.78)
77.5
(3.05)
50
(1.97)
F9811
JT
126
(4.96)
72.3
(2.85)
76.7
(3.02)
72.9
(2.87)
83.5
(3.29)
91
(3.58)
50
(1.97)
F9811
JU
152
(5.98)
97
(3.82)
102.5
(4.04)
97.6
(3.84)
110
(4.33)
119
(4.69)
50
(1.97)
F9811
JV
180
(7.09)
123
(4.84)
141.2
(5.56)
135.7
(5.34)
146
(5.75)
155
(6.10)
60
(2.36)
F9811
JW
F34.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
50
<<Contents>> <<Index>>
197(7.76)*1
66*1 51.5 51.5
(2.6) (2.03) (2.03)
28(1.1)
111(4.37)
70
(2.76)
Hr
Hi
J K B
J D B 1
J S B
73
49
(2.87) (1.93)
D
E 1
F 2 A L 1 N
G N
N
P
Ground Terminal
(M4)
86(3.38)
48
(1.89)
154(6.06)
*3
128(5.04)
AXF015
AXF025
AXF032
AXF040
AXF050 H
AXF065
AXF080
AXF100
AXF125
Remote Flowtube
H1
D
(d)
Adapter for union
connection
015
025
032
040
050
065
080
100
125
Size
15
(0.5)
25
(1)
32
(1.3)
40
(1.5)
50
(2)
65
(2.6)
80
(3)
100
(4)
125
(5)
Lining code
166
166
166
166
176
196
216
246
166
(6.55)
166
(6.55)
73
(2.87)
31.3
(1.23)
32
(1.26)
29.6
(1.17)
98
(3.86)
222
(8.74)
2.7
(6.0)
260
(10.24)
4.4
(9.7)
166
(6.55)
166
(6.55)
86
(3.39)
35.6
(1.40)
38
(1.50)
35.5
(1.40)
111
(4.37)
235
(9.25)
3
(6.6)
273
(10.73)
4.7
(10.4)
176
(6.94)
176
(6.94)
99
(3.90)
48.6
(1.91)
50
(1.97)
48.5
(1.91)
129
(5.08)
253
(9.96)
3.8
(8.4)
291
(11.44)
5.5
(12.1)
196
(7.73)
196
(7.73)
117
(4.61)
60.3
(2.37)
66
(2.60)
60.5
(2.38)
147
(5.79)
271
(10.67)
4.9
(10.8)
309
(12.17)
6.6
(14.6)
236
(9.31)
216
(8.52)
129
(5.08)
72.9
(2.87)
81
(3.19)
72.9
(2.87)
157
(6.18)
281
(11.06)
5.9
(13.0)
319
(12.54)
7.6
(16.8)
266
(10.49)
276
(10.88)
155
(6.10)
97.6
(3.84)
100
(3.94)
97.6
(3.84)
183
(7.20)
307
(12.09)
8.2
(18.1)
345
(13.56)
9.9
(21.8)
326
(12.85)
Integral
Flowmeter
JKB
Inside dia. d JDB
JSB
Remote
Flowtube
Integral
Flowmeter
Height
H1
Max. Height
Hr
2
Weight kg (lb)*
Max. Height
Hi
Weight kg (lb)
70
(2.76)
16
B
(0.63)
18
C
(0.71)
15.2
d
(0.60)
21.2
E
(0.83)
13.5
F
(0.53)
3
G
(0.12)
50
H
(1.97)
F9811
Parts No. LA
A
E
F
125
(4.92)
212
(8.35)
336
(13.23)
13
(28.7)
374
(14.70)
14.7
(32.4)
JSB:
3
F
H
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the
figure.
*2: When option code DHC is selected, waterproof glands and a 30m long cable are
attached. Add 9.5kg (20.9lb) to the weight in the table.
Process
Connection
Nominal
Size
183
(7.20)
d
E
JSB
Remote
Flowtube Outside dia.
JDB:
JDB
166
166
(6.55) (6.55)
166
(6.55)
73
73
(2.87) (2.87)
22.6
15.2
(0.60) (0.89)
26
16
(0.63) (1.02)
22.5
(0.89)
98
98
(3.86) (3.86)
222
222
(8.74) (8.74)
2.6
2.6
(5.7) (5.7)
260
260
(10.24) (10.24)
4.3
4.3
(9.5) (9.5)
Model
Size code
Face-to-face
L
length
Process Connection
JKB:
25
32
40
50
65
80
100
15
25
32
40
50
65
80
100
125
25
32
40
50
65
80
100
70
(2.76)
22.2
(0.87)
25.6
(1.01)
22.6
(0.89)
29.2
(1.15)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LB
70
(2.76)
29.4
(1.16)
34.3
(1.35)
31.3
(1.23)
38.2
(1.50)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LC
83
(3.27)
34.6
(1.36)
38.6
(1.52)
35.6
(1.40)
42.7
(1.68)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LD
96
(3.78)
47.6
(1.87)
51.6
(2.03)
48.6
(1.91)
56.2
(2.21)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LE
114
(4.49)
59.5
(2.34)
64.1
(2.52)
60.3
(2.37)
69.9
(2.75)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LF
126
(4.96)
72.3
(2.85)
76.7
(3.02)
72.9
(2.87)
82.6
(3.25)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LG
152
(5.98)
97
(3.82)
102.5
(4.04)
97.6
(3.84)
108.7
(4.28)
13.5
(0.53)
3
(0.12)
50
(1.97)
F9811
LH
70
(2.76)
16
(0.63)
20
(0.79)
16
(0.63)
18
(0.71)
25.8
(1.02)
RD34
1/8
50
(1.97)
F9811
KR
70
(2.76)
22.2
(0.87)
30
(1.18)
26
(1.02)
30
(1.18)
39.8
(1.57)
RD52
1/6
50
(1.97)
F9811
KS
70
(2.76)
29.4
(1.16)
36
(1.42)
32
(1.26)
36
(1.42)
45.8
(1.80)
RD58
1/6
50
(1.97)
F9811
KT
83
(3.27)
34.6
(1.36)
42
(1.65)
38
(1.50)
42
(1.65)
51.8
(2.04)
RD65
1/6
50
(1.97)
F9811
KU
96
(3.78)
47.6
(1.87)
54
(2.13)
50
(1.97)
54
(2.13)
63.8
(2.51)
RD78
1/6
50
(1.97)
F9811
KV
114
(4.49)
59.5
(2.34)
70
(2.76)
66
(2.60)
71
(2.80)
80.8
(3.18)
RD95
1/6
50
(1.97)
F9811
KW
126
(4.96)
72.3
(2.85)
85
(3.35)
81
(3.19)
85
(3.35)
94.8
(3.73)
RD110
1/4
60
(2.36)
F9811
KX
152
(5.98)
97
(3.82)
104
(4.09)
100
(3.94)
104
(4.09)
113.8
(4.48)
RD130
1/4
60
(2.36)
F9811
KY
180
(7.09)
123
(4.84)
129
(5.08)
125
(4.92)
130
(5.12)
141.8
(5.58)
RD160
1/4
65
(2.56)
F9811
KZ
70
(2.76)
22.2
(0.87)
25.4
(1.00)
22.5
(0.89)
32
(1.26)
11
(0.43)
RD40
1/6
50
(1.97)
F9811
LK
70
(2.76)
29.4
(1.16)
32
(1.26)
29.6
(1.17)
40
(1.57)
13
(0.51)
RD48
1/6
50
(1.97)
F9811
LL
83
(3.27)
34.6
(1.36)
38.1
(1.50)
35.5
(1.40)
48
(1.89)
15
(0.59)
RD60
1/6
50
(1.97)
F9811
LM
96
(3.78)
47.6
(1.87)
51
(2.01)
48.5
(1.91)
61
(2.40)
15
(0.59)
RD70
1/6
50
(1.97)
F9811
LN
114
(4.49)
59.5
(2.34)
63.5
(2.50)
60.5
(2.38)
73.5
(2.89)
19
(0.75)
RD85
1/6
50
(1.97)
F9811
LP
126
(4.96)
72.3
(2.85)
76.2
(3.00)
72.9
(2.87)
86
(3.39)
19
(0.75)
RD98
1/6
50
(1.97)
F9811
LQ
152
(5.98)
97
(3.82)
102.5
(4.04)
97.6
(3.84)
120
(4.72)
30
(1.18)
RD132
1/6
65
(2.56)
F9811
LR
F35.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
51
<<Contents>> <<Index>>
70(2.76)
032
040
050
065
080
100
125
Size
25
(1)
32
(1.3)
40
(1.5)
50
(2)
65
(2.6)
80
(3)
100
(4)
125
(5)
Lining code
L
Outside dia.
Inner dia. d
KDB
Integral
Flowmeter
Height
H1
Max. Height
Hr
2
Weight kg (lb)*
Max. Height
Process
Connection
Nominal
Size
Hi
Weight kg (lb)
126
(4.98)
73
(2.87)
22.6
(0.89)
26
(1.02)
98
(3.86)
222
(8.74)
2.3
(5.1)
260
(10.24)
4
(8.8)
126
(4.98)
73
(2.87)
31.3
(1.23)
32
(1.26)
98
(3.86)
222
(8.74)
2.5
(5.5)
260
(10.24)
4.2
(9.3)
126
(4.98)
86
(3.39)
35.6
(1.40)
38
(1.50)
111
(4.37)
235
(9.25)
2.8
(6.2)
273
(10.73)
4.5
(9.9)
136
(5.37)
99
(3.90)
48.6
(1.91)
50
(1.97)
129
(5.08)
253
(9.96)
3.4
(7.5)
291
(11.44)
5.1
(11.2)
156
(6.16)
117
(4.61)
60.3
(2.37)
66
(2.60)
147
(5.79)
271
(10.67)
4.5
(9.9)
309
(12.17)
6.2
(13.7)
176
(6.94)
129
(5.08)
72.9
(2.87)
81
(3.19)
157
(6.18)
281
(11.06)
5.3
(11.7)
319
(12.54)
7
(15.4)
206
(8.13)
155
(6.10)
97.6
(3.84)
100
(3.94)
183
(7.20)
307
(12.09)
7.1
(15.7)
345
(13.56)
8.8
(19.4)
276
(10.88)
183
(7.20)
135.7
(5.34)
125
(4.92)
212
(8.35)
336
(13.23)
11
(24.3)
374
(14.70)
12.7
(28.0)
70
(2.76)
16
B
(0.63)
18
C
(0.71)
15.2
d
(0.60)
30
E
(1.18)
F9811
Parts No. NN
A
KKB
A
126
(4.98)
73
(2.87)
15.2
(0.60)
16
(0.63)
98
(3.86)
222
(8.74)
2.6
(5.7)
260
(10.24)
4.3
(9.5)
025
15
(0.5)
015
Size code
Face-to-face
length
KKB / KDB
Model
(d)
Process Connection
Integral
Flowmeter
28(1.1)
Remote
Flowtube
51.5 51.5
(2.03) (2.03)
111(4.37)
H1
Remote
Flowtube
66*1
(2.6)
Hr
Hi
K K B
1
K D B
73(2.87) 49(1.93)
48
(1.89)
D
E 1
F 2 A L 1 N
G N
N
P
Remote Flowtube
Ground Terminal
(M4)
197(7.76)*1
Ground Terminal
(M4)
86(3.38)
154(6.06)
*3
Integral Flowmeter
128(5.04)
AXF015
AXF025
AXF032
AXF040
AXF050 H
AXF065
AXF080
AXF100
AXF125
Remote Flowtube
25
32
40
50
65
80
100
125
15
25
32
40
50
65
80
100
125
70
(2.76)
22.2
(0.87)
25.6
(1.01)
22.6
(0.89)
30
(1.18)
F9811
NP
70
(2.76)
29.4
(1.16)
34.3
(1.35)
31.3
(1.23)
30
(1.18)
F9811
NQ
83
(3.27)
34.6
(1.36)
38.6
(1.52)
35.6
(1.40)
30
(1.18)
F9811
NR
96
(3.78)
47.6
(1.87)
51.6
(2.03)
48.6
(1.91)
30
(1.18)
F9811
NS
114
(4.49)
59.5
(2.34)
64.1
(2.52)
60.3
(2.37)
30
(1.18)
F9811
NT
126
(4.96)
72.3
(2.85)
76.7
(3.02)
72.9
(2.87)
30
(1.18)
F9811
NU
152
(5.98)
97
(3.82)
102.5
(4.04)
97.6
(3.84)
30
(1.18)
F9811
NV
180
(7.09)
123
(4.84)
141.2
(5.56)
135.7
(5.34)
40
(1.57)
F9811
NW
70
(2.76)
16
(0.63)
20
(0.79)
16
(0.63)
30
(1.18)
F9811
ND
70
(2.76)
22.2
(0.87)
30
(1.18)
26
(1.02)
30
(1.18)
F9811
NE
70
(2.76)
29.4
(1.16)
36
(1.42)
32
(1.26)
30
(1.18)
F9811
NF
83
(3.27)
34.6
(1.36)
42
(1.65)
38
(1.50)
30
(1.18)
F9811
NG
96
(3.78)
47.6
(1.87)
54
(2.13)
50
(1.97)
30
(1.18)
F9811
NH
114
(4.49)
59.5
(2.34)
70
(2.76)
66
(2.60)
30
(1.18)
F9811
NJ
126
(4.96)
72.3
(2.85)
85
(3.35)
81
(3.19)
30
(1.18)
F9811
NK
152
(5.98)
97
(3.82)
104
(4.09)
100
(3.94)
30
(1.18)
F9811
NL
180
(7.09)
123
(4.84)
129
(5.08)
125
(4.92)
40
(1.57)
F9811
NM
GS 01E20D01-01E
F36.EPS
Jan. 7, 2015-00
52
<<Contents>> <<Index>>
Replacement model for Earlier ADMAG or ADMAG AE, AXF002-AXF015, Wafer Type, PFA Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube
51.5 51.5
(2.03) (2.03)
66*1
(2.6)
111(4.37)
70(2.76)
Ground Terminal
(M4)
28(1.1)
Hr
63.5
(2.50)
3(0.12)
H1
(d
D
Hi
73(2.87) 49(1.93)
AXF002
G
AXF005
W
AXF010
C
AXF015
Remote Flowtube
197(7.76)*1
128(5.04)
Ground Terminal
(M4)
86(3.38)
(1.89)
48
154(6.06)
*4
D
E 1
F 2 A
G N
N
P
Integral Flowmeter
L*2
4- 6.2(0.24)
72
58
(2.83)
(2.28)
Model
Size code
002
005
010
015
Size
2.5(0.1)
5(0.2)
10(0.4)
15(0.5)
Lining code
Face-to-face
length
Remote
Flowtube Outside dia.
*2
85(3.35)
44(1.73)
15(0.59)
H1
144(5.67)
Hr
Remote Max. Height
Flowtube
3
Weight kg (lb)*
268(10.55)
Hi
306(12.03)
Height
Integral
Flowmeter
Max. Height
None
P, T
+22(0.87) -6(0.24)
+0
2.4(5.3)
Weight kg (lb)
F37.EPS
4.1(9.0)
Replacement model for Earlier ADMAG or ADMAG AE, AXF025-AXF100, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter Remote Flowtube
Size code
025
040
050
080
100
Size
25(1)
40(1.5)
50(2)
80(3)
100(4)
A,U
A,U
A,U
A,U
A,U
Face-to-face
length
*2
Remote
Flowtube Outside dia.
Inner diameter of
Grounding ring
Integral
Flowmeter
Height
H1
Remote Flowtube
Ground Terminal
(M4)
28(1.1)
128(5.04)
111(4.37)
H1
Lining code
197(7.76)*1
51.5 51.5
(2.03) (2.03)
70
(2.76)
Hr
Model
66*1
(2.6)
48(1.89)
Hi
73(2.87) 49(1.93)
D
E 1
A
F 2
U
G N
N
P
Ground Terminal
(M4)
86(3.38)
(
d)
154(6.06)
*4
AXF025
AXF040 G
AXF050 W
AXF080 C
AXF100
Integral Flowmeter
L*2
None
+0
P, T
+22(0.87) -6(0.24)
Max. Height
Weight kg (lb)
GS 01E20D01-01E
Jan. 7, 2015-00
53
<<Contents>> <<Index>>
Replacement model for Earlier ADMAG or ADMAG AE, AXF150, AXF200, Wafer Type,
PFA /Polyurethane Rubber Lining
Unit : mm (approx. inch)
Integral Flowmeter
Remote Flowtube
154(6.06)
86(3.38)
66*1
51.5 51.5
(2.6)
(2.03) (2.03)
28(1.1)
111(4.37)
128(5.04)
d)
(
70(2.76)
Hr
73(2.87) 49(1.93)
48(1.89)
D
E 1
A
F 2
U
G N
N
P
Hi
G
AXF150
W
AXF200
C
Remote Flowtube
Ground Terminal
(M4)
197(7.76)*1
Ground Terminal
(M4)
*4
Integral Flowmeter
H1
Model
Size code
150
200
Size
150(6)
200(8)
A,U
A,U
230(9.06)
300(11.81)
202(7.95)
252(9.92)
140.7(5.54)
188.9(7.44)
Lining code
Remote
flowtube
Face-to-face
length
L*
Outside dia. D
Inner diameter of
Integral
Grounding ring d
flowmeter
H1
Height
Remote
flowtube
Integral
flowmeter
L*2
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the
value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add
the following value to L (face-to-face length).
Grounding Ring Code S, L, H, V
243(9.57)
293(11.54)
Max. Height Hr
367(14.45)
417(16.42)
Weight kg (lb)*3
17.9(39.5)
26.8(59.1)
Max. Height Hi
405(15.93)
455(17.89)
Weight kg (lb)
19.6(43.2)
28.5(62.8)
Option
Code
None
+0
P, T
+28(1.1)
-6(0.24)
*3: When submersible type or option code DHC is selected, waterproof glands
and a 30m long cable are attached. Add 9.5kg(20.9lb) to the weight in the
table.
F39.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
54
<<Contents>> <<Index>>
Replacement model for Earlier ADMAG or ADMAG AE, AXF150-AXF250, JIS/ANSI/DIN Flange Type,
PFA /Polyurethane Rubber Lining
*4
AXF150 G
AXF200 W
AXF250 C
D
E 1
A
F 2
U
G N
N
P
1
2
A
D
2
J
G 1
C
C
C
C
Integral Flowmeter
Integral Flowmeter
Ground Terminal
(M4)
197(7.76)*1
Remote Flowtube
154(6.06)
86
Ground Terminal
(M4)
(3.38)
28(1.1)
70
(2.76)
Eye Bolt
D
Hr
(H2)
C
)
(d
H1
H3
Hi
111(4.37)
N- h
Process Connection
t *2
Model
Remote Flowtube
128(5.04)
73 49 (1.93)
(2.87)
48(1.89)
L*2
CJ1(JIS10K)
CG1(JIS F12)
CJ2(JIS20K)
CD2(DIN PN16)
Size code
150
200
250
150
200
250
150
200
250
150
200
250
150
200
250
200
250
150
200
250
Size
150
(6)
200
(8)
250
(10)
150
(6)
200
(8)
250
(10)
150
(6)
200
(8)
250
(10)
150
(6)
200
(8)
250
(10)
150
(6)
200
(8)
250
(10)
200
(8)
250
(10)
150
(6)
200
(8)
250
(10)
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
A,U
270
(10.63)
280
(11.02)
31
(1.22)
140.7
(5.54)
240
(9.45)
340
(13.39)
330
(12.99)
31
(1.22)
188.9
(7.44)
290
(11.42)
430
(16.93)
400
(15.75)
33
(1.30)
243
(9.57)
355
(13.98)
270
(10.63)
305
(12.01)
37
(1.46)
140.7
(5.54)
260
(10.24)
340
(13.39)
350
(13.78)
39
(1.54)
188.9
(7.44)
305
(12.01)
430
(16.93)
430
(16.93)
43
(1.69)
243
(9.57)
380
(14.96)
270
(10.63)
290
(11.42)
31
(1.22)
140.7
(5.54)
247
(9.72)
340
(13.39)
342
(13.46)
33
(1.30)
188.9
(7.44)
299
(11.77)
430
(16.93)
410
(16.14)
33
(1.30)
243
(9.57)
360
(14.17)
270
(10.63)
279.4
(11.00)
34.4
(1.35)
140.7
(5.54)
241.3
(9.50)
340
(13.39)
342.9
(13.50)
37.4
(1.47)
188.9
(7.44)
298.5
(11.75)
430
(16.93)
406.4
(16.00)
39.2
(1.54)
243
(9.57)
362.0
(14.25)
270
(10.63)
317.5
(12.50)
47.5
(1.87)
140.7
(5.54)
269.7
(10.62)
340
(13.39)
381.0
(15.00)
50.1
(1.97)
188.9
(7.44)
330.2
(13.00)
430
(16.93)
444.5
(17.50)
56.7
(2.23)
243
(9.57)
387.4
(15.25)
340
(13.39)
340
(13.39)
33
(1.30)
188.9
(7.44)
295
(11.61)
430
(16.93)
395
(15.55)
35
(1.38)
243
(9.57)
350
(13.78)
270
(10.63)
285
(11.22)
31
(1.22)
140.7
(5.54)
240
(9.45)
340
(13.39)
340
(13.39)
33
(1.30)
188.9
(7.44)
295
(11.61)
430
(16.93)
405
(15.94)
35
(1.38)
243
(9.57)
355
(13.98)
15
15
15
15
15
30
22.5
22.5
22.5
22.5
15
15
15
11.25
22.5
15
22.5
15
15
Lining code
Face-to-face
length
L*2
Outside dia.
Thickness
Inner diameter of
Remote Grounding ring
t *2
d
flowtube
Pitch circle dia. C
Integral
flowmeter Bolt hole interval
Hole dia.
Remote
flowtube
22.5
23
(0.91)
Number of holes
Height
H1
Height
H2
Height
H3
Max. Height
Hr
Weight kg (lb)*3
Hi
Max. Height
Integral
flowmeter
Weight kg (lb)
23
25
25
25
27
19
19
23
22.4
(0.91) (0.98) (0.98) (0.98) (1.06) (0.75) (0.75) (0.91) (0.88)
12
284
(11.18)
141
(5.55)
325
(12.80)
408
(16.06)
29
(63.9)
446
(17.56)
30.7
(67.7)
12
12
334
400
297
(13.15) (15.75) (11.69)
166
197
141
(6.54) (7.76) (5.55)
375
454
350
(14.76) (17.87) (13.78)
458
524
421
(18.03) (20.63) (16.57)
39
64
38.3
(86.0) (141.1) (84.4)
496
562
459
(19.53) (22.13) (18.07)
40.7
65.7
40.0
(89.7) (144.8) (88.2)
12
12
344
415
289
340
(13.54) (16.34) (11.38) (13.39)
166
197
141
166
(6.54) (7.76) (5.55) (6.54)
395
484
335
387
(15.55) (19.06) (13.19) (15.24)
468
539
413
464
(18.43) (21.22) (16.26) (18.27)
53.6
92.5 31.1
44.6
(118.2) (203.9) (68.6) (98.3)
506
577
451
502
(19.92) (22.72) (17.76) (19.76)
55.3
94.2
32.8
46.3
(121.9) (207.7) (72.3) (102.1)
22.4
25.4
22.4
25.4
28.4
(0.88) (1.00) (0.88) (1.00) (1.12)
22
22
(0.87) (0.87)
22
(0.87)
22
26
(0.87) (1.02)
12
12
12
16
12
12
12
405
(15.94)
197
(7.76)
464
(18.27)
529
(20.83)
67.4
(148.6)
567
(22.32)
69.1
(152.3)
284
(11.18)
141
(5.55)
324
(12.76)
408
(16.06)
32.1
(70.7)
446
(17.56)
33.8
(74.4)
340
(13.39)
166
(6.54)
387
(15.24)
464
(18.27)
50.9
(112.2)
502
(19.76)
52.6
(115.9)
403
(15.87)
197
(7.76)
460
(18.11)
527
(20.75)
77.4
(170.6)
565
(22.24)
79.1
(174.4)
303
(11.93)
141
(5.55)
362
(14.25)
427
(16.81)
53.7
(118.4)
465
(18.31)
55.4
(122.1)
360
(14.17)
166
(6.54)
426
(16.77)
484
(19.06)
80.5
(177.5)
522
(20.55)
82.2
(181.2)
422
(16.61)
197
(7.76)
499
(19.65)
546
(21.50)
127.0
(279.9)
584
(22.99)
128.7
(283.6)
339
(13.35)
166
(6.54)
385
(15.16)
463
(18.23)
44.2
(97.5)
501
(19.72)
45.9
(101.2)
397
(15.63)
197
(7.76)
449
(17.68)
521
(20.51)
67.0
(147.7)
559
(22.01)
68.7
(151.5)
287
(11.30)
141
(5.55)
330
(12.99)
411
(16.18)
29.9
(65.8)
449
(17.68)
31.6
(69.6)
339
(13.35)
166
(6.54)
385
(15.16)
463
(18.23)
43.6
(96.2)
501
(19.72)
45.3
(100.0)
402
(15.83)
197
(7.76)
459
(18.07)
526
(20.71)
68.8
(151.7)
564
(22.20)
70.5
(155.5)
*1: When indicator code N is selected, subtract 12 mm (0.47 inch) from the value in the figure.
In case of explosion proof type with indicator, add 5 mm (0.2 inch) to it.
*2: Depending on the selection of grounding ring code and optional code, add the following value to L (face-to-face length) and t (thickness of
flange).
Nominal Size: 250 mm
L
t
None
S, L, H, V
+0
+0
P, T
S, L, H, V
+0
+0
t
N
-6(0.24) -3(0.12)
*3: When submersible type or option code DHC is selected, waterproof glands and a 30m long cable are attached.
Add 9.5kg(20.9lb) to the weight in the table.
F40.EPS
GS 01E20D01-01E
Jan. 7, 2015-00
55
<<Contents>> <<Index>>
Flameproof Packing Adapter for TIIS Flameproof Type (Optional code G12 or G11)
T1
Unit : mm
(Approx. inch)
18(0.71)
O-Ring
Washer
Union nut
Packing gland
*Packing
(Choose from the table below
depend on cable outside diameter)
Clamp ring
Clamp nut
T2
O-Ring
B.coupling
Cable(user's scope)
Nominal diameter
T1
T2
G 1/2
G 1/2
Packing diameter
Dimension
C
35
39
94.5
(1.38)
(1.54)
(3.72)
F43.EPS
Identification
mark
10.0(0.39)
20.0
16
12.0(0.47)
(0.79)
16 10-12
8-10
Weight
kg
(lb)
0.26
Parts No.
G9601AM*
(0.57)
*: G 11: 1 unit
G 12: 2 units
T31.EPS
Unless otherwise specified, difference in the dimensions are refer to the following table.
General tolerance in the dimensional outline drawing.
Category of basic dimension
Above
Equal or below
3 (0.12)
6 (0.24)
10 (0.39)
18 (0.71)
30 (1.18)
50 (1.97)
80 (3.15)
120 (4.72)
180 (7.09)
250 (9.84)
315 (12.40)
400 (15.75)
3 (0.12)
6 (0.24)
10 (0.39)
18 (0.71)
30 (1.18)
50 (1.97)
80 (3.15)
120 (4.72)
180 (7.09)
250 (9.84)
315 (12.40)
400 (15.75)
500 (19.69)
Tolerance
0.7 (0.03)
0.9 (0.04)
1.1 (0.04)
1.35 (0.05)
1.65 (0.06)
1.95 (0.08)
2.3 (0.09)
2.7 (0.11)
3.15 (0.12)
3.6 (0.14)
4.05 (0.16)
4.45 (0.18)
4.85 (0.19)
Unit : mm (approx.inch)
630 (24.80)
800 (31.50)
1000 (39.37)
1250 (49.21)
1600 (62.99)
2000 (78.74)
2500 (98.43)
3150 (124.02)
Tolerance
5.5 (0.22)
6.25 (0.25)
7.0 (0.28)
8.25 (0.32)
9.75 (0.38)
11.5 (0.45)
14.0 (0.55)
16.5 (0.65)
Remarks: The numeric is based on criteria of tolerance class IT18 in JIS B 0401.
GS 01E20D01-01E
Jan. 7, 2015-00
56
<<Contents>> <<Index>>
80
10
0
12
5
15
0
40
50
65
15
10
10.0
25
32
Size: mm
2.
5
SI Units
5.0
Flow velocity
(m/s)
1.0
0.5
0.3
0.1
0.001
0.01
1,000
100
10
0.1
10,000
10
12
14
16
5
6
2.5
1.2
5
1.5
0.5
0.4
33
Size: inch
0.2
0.1
English Units
Flow velocity
(ft/s)
10
5.0
2.0
1.0
0.33
0.01
0.1
1.0
100
10
1,000
10,000
100,000
F44.EPS
Flowrate (GPM)
GS 01E20D01-01E
Jan. 7, 2015-00
57
<<Contents>> <<Index>>
ORDERING INFORMATION
Note 1: When ordering, the span flow rate, unit, output
pulse weight, and totalizer display pulse weight can
be specified. These parameters will then be set
before shipment.
In the case of remote type, please specify combination information of flowtube and converter.
These parameters will be set in the combined
converter.
When ordering of a flowtube alone or a converter
alone, these parameters cannot specify.
Custom configuration request is necessary for
setting these parameters out of regular setting
range.
Note 2: Some options, if ordered, require the relevant
specifications to be input when ordering.
RELATED INSTRUMENTS
Calibrator for Magnetic Flowmeter (AM012):
GS 1E6K2-E
BT200 BRAIN TERMINAL:
GS 01C00A11-00EN
GS 01R01A01-01E
REFERENCE
ADMAG, AXF, ADMAG AXF, BRAIN TERMINAL and
FieldMate are registered trademarks of Yokogawa Electric
Corporation.
Company names and product names used in this material
are registered trademarks or trademarks of their respective
owners.
GS 01E20D01-01E
Jan. 7, 2015-00