Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
The actual Proof of Gross (and/or compensated) Volume is NOT done under the Meter Calibration.
Once the meter is calibrated, Proofs or Checks are done in Standard delivery mode.
For Initial Calibration or Re-Calibration if the Proof does not meet local criteria:
EMR CALIBRATION STEPS - Autocalibrate
1. Standard VOLUME MODE on Display (Gallons)
2. Break Seal
3. Remove or lower Left Rear Bolt. (7/16 hex)
4. Push MODE button to get to SETUP. Display will show LANGUAGE.
5. Arrow down to METER CALIBRATION
6. NEXT PROD CAL 1
7. NEXT, AUTOCALIBRATE
8. NEXT: Prompted with START DISPENSE
9. Push the START BUTTON.
10. Fill the Prover Pay attention to the prover volume and sight glass.
11. When the prover is filled: Push the FINISH button. Prompt: ENTER NTC VOLUME
12. Using the NEXT key to move the cursor in the Preset Field and the Arrow Keys to change the
values. Enter the Non-Compensated Prover Volume.
13. Push the ENTER key and a new calibration factor will be Calculated.
14. Record this K-Factor
15. Hit ENTER to get back to METER CALIBRATION
16. Insert the Left Rear Bolt.
17. MODE button to get to Volume(Gallons) Mode.
18. Clear the Ticket.
19. Run check under normal Volume/Delivery Mode.
20. If Temperature Compensated, the Delivered volume will have the thermometer.
21. After pushing FINISH to complete the transaction, Push the down arrow to display the NonCompensated Volume.
If your non-compensated checks in standard Volume/delivery mode are consistent, but out of
tolerance, a new K-factor can be calculated per the attached information and entered in Manual
Calibrate. Or, an Autocalibrate run can be made to calculate a new K-factor.
SPECIAL CALIBRATION ACCEPTANCE IF MORE THAN ONE PRODUCT PRODUCT ADDED AFTER
INITIAL CALIBRATION
If all the products were set up and assigned Cal 1 before the calibration was done, they would all get that
calibration properly assigned to that product.
If a new product is added after the calibration of that Cal 1, say 25, the new product unfortunately is not
automatically given the 25, but gets the 96 factory default in the background. Going to Manual calibrate and
getting to Encoder counts, NEXT, then ENTER to "recalculate" gets the CAL 1 of 25 assigned to the new
product.
23
Detailed Steps to manual calibrate to accept the calibration factor if product added after calibration:
18. Standard Volume Mode on Display
19. Break Seal
20. Remove or lower Left Rear Bolt. (7/16 hex)
21. Push MODE button to get to SETUP. Display will show LANGUAGE.
22. Arrow down to METER CALIBRATION
23. NEXT PROD CAL 1
24. NEXT AUTOCALIBRATE
25. Arrow up to MANUAL CALIBRATE
26. NEXT CALB RATE 1 (Encoder Counts K-factor in Preset field, Flow rate in Volume delivered field.)
27. NEXT FLOW RATE (DEFAULT IS 60 and okay to leave if single calibration)
28. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
29. ENCODER COUNTS and value in preset field
30. To ACCEPT the ENCODER COUNTS, NEXT SET COUNTS and Cursor Flashes.
31. Push ENTER. Recalculation Occurs
32. Then ENTER(s) to get back to METER CALIBRATION.
33. Insert the Left Rear Bolt.
34. MODE button to get to VOLUME (Gallons) Mode.
How to make corrections in calibration without re-calibrating
Example:
Calibration Factor: 95.966 pulses per Gallon
Register vol: 100.4
Prover volume: 100.0
(Prover-reg)/prover = (100.0-100.4)/100.0 = - .004
New factor= old factor /(1+diff) note sign is important
New factor = 95.966/(1-.004) = 95.966/0.996 = 96.351
If register more than prover, factor up. (less Gallons/pulse) (greater Pulses/gallon)
If register less than prover, factor down (more Gallons/pulse) (less pulses/gallon)
The adjustment is made in Manual Calibration with the C&C switch open.
24
Manual Calibrate: For entering known calibration values or normal values for specific meter if known.
30. Standard Volume Mode on Display
31. Break Seal
32. Remove or lower Left Rear Bolt. (7/16 hex)
33. Push MODE button to get to SETUP. Display will show LANGUAGE.
34. Arrow down to METER CALIBRATION
35. NEXT PROD CAL 1
36. NEXT AUTOCALIBRATE
37. Arrow up to MANUAL CALIBRATE
38. NEXT CALB RATE 1 (Encoder Counts K-factor in Preset field, Flow rate in Volume delivered field.)
39. NEXT FLOW RATE (DEFAULT IS 60)
40. To change FLOW RATE : NEXT to get flashing cursor and use arrow keys to change numbers, THEN
enter
41. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
42. ENCODER COUNTS and value in preset field
43. To change ENCODER COUNTS, NEXT SET COUNTS
44. Use the NEXT key to move the cursor and the arrow keys to change the numbers.
45. Once the number is changed, hit ENTER. CAL RATE 1
46. For another point Arrow down to CAL RATE 2, if not, ENTER back to Meter Calibration.
47. To change FLOW RATE : NEXT to get flashing cursor and use arrow keys to change numbers, THEN
enter YOU MUST ENTER A FLOW RATE for each additional point.
48. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
49. ENCODER COUNTS and value in preset field
50. To change ENCODER COUNTS, NEXT SET COUNTS
51. Use the NEXT key to move the cursor and the arrow keys to change the numbers.
52. Once the number is changed, hit ENTER. CAL RATE 2
53. Repeat for additional calibration points.
54. Then ENTER(s) to get back to METER CALIBRATION.
55. Insert the Left Rear Bolt.
56. MODE button to get to VOLUME (Gallons) Mode.
57. Clear the Ticket.
58. Run checks under normal Volume/Delivery Mode.
METER CALIBRATION
There are now 4 Calibrations Available. For multiple products,
If using Temperature Compensation, calibrate the Temperature Probe first as noted previously.
The system calibration is running a known Non-Compensated volume through the meter and having the
register electronics calculate the Calibration Factor. The calibration factor is Encoder Counts per Unit
Volume. During the Auto or Multi calibrate procedures, the register remembers the number of counts
during flow (meter turning). Once the Non-Temperature Compensated volume, Gross Volume, of the Prover
vessel is entered, the register calculates the Calibration factor.
The actual Proof of Gross and Compensated Volumes is NOT done under the Meter Calibration. Once the
meter is calibrated, close the C&C switch, and return to Volume Mode where the Proof run is made. Here
the register applies the Density or Coefficients to obtain the Compensated Volume. If the displayed volume is
Compensated, the thermometer icon to the right of the delivered volume is displayed. The Gross volume is
displayed by pushing an arrow button. These Compensated and Gross Volumes are compared to the Prover
Volume and W&M tables to verify the meter calibration.
25
AUTO CALIBRATE is the easiest option. If the system will normally be run at only one flow rate, then Auto
Calibrate at that flow rate will be fine. Follow the procedure noted in the Auto Calibrate section of the Setup
and Operation Manual.
IMPORTANT: Once the GROSS prover volume is entered and the ENTER button is pushed and the
Register calculates the calibration factor (the Descriptor Field changes and a new calibration factor shows),
RECORD THIS FACTOR for future reference. Then push ENTER UNTIL YOU RETURN TO METER
CALIBRATION. Returning to METER CALIBRATION saves the Calibration Factor.
MULTI CALIBRATE is used for systems that will be run at more than one flow rate. For large meters that will
be run at low flow rates, more than one calibration in the low flow rate range is recommended. It is not
necessary to return to Meter Calibration between rates, however once done, push ENTER to get back to
Meter Calibration.
If an E65 CALIBRATION ERROR occurs more points at closer flow rates will be needed. The error occurs
because one of the multipoint calibration values is different by more that 0.25% of the previous value. This is
built in the software to satisfy a Canadian W&M requirement. For example if calibration point 1 is 100.00 than
calibration point 2 must be between 99.75 and 100.25 counts. Push any key to clear the error code. Push
the NEXT key to begin calibrating a new flow rate point using either Autocalibrate or Multicalibrate. It is also
acceptable to calculate the counts and enter them manually using the Manual Calibrate Feature. If this
occurs, Single point autocalibrate points can be done and the curve entered manually. For example: at a
flow rate of 400 gpm: the calibration factor is 100.00. When the flow rate was 200 the E65 Calibration error
occurred. Either try to throttle the valve to get between the 200 and 400 flow rates, and re-do the second
point, or record the flow rate and K-factor for the 400. Stay at the 200 valve throttling and do a single point
AutoCalibrate. Record that K-factor and flow rate. Draw a straight-line curve between the two flow rates,
i.e. interpret points to obtain a K-factor and flowrate that does not violate the 0.25% difference.
MANUAL CALIBRATION is an option available as outlined in the Manual. It is normally used for entering
known calibration values or normal values for specific meter if known.
If for some reason the display head is replaced, or Restart is selected and the stored information in the
display head is lost, the Calibration Factor, if recorded, is entered under Manual Calibration.
Note: For Large Volume prover applications, a prover vessel metal expansion factor may need to be applied
per the local Weights and Measures regulations.
Recommend tenths resolution on Calibration even if whole units on Register for Delivery
To get the best accuracy on the calibration, Both register resolution and preset resolution (for k-factor)
should be temporarily set to Tenths. For calibration, the seal must be broken anyway, so the resolution
can be changed for the calibration and proof, then set back to whole units if that is the delivered resolution.
With C&C switch open, go to DISPLAY SYNTAX. Go to PRESET RES, PRESET VOLUME and move
decimal using NEXT, the ENTER back to PRESET RES to save. Do the same with REGISTER RES,
REGISTER VOL.
Remember To Go Back To Whole Units When Complete With Calibration And Proofs.
MULTICALIBRATE AND METER CAPABILITY WITH THE EMR
26
Sometimes meters are run at 4 to 1 or greater turndowns or the meter has some wear, making a single point
calibration inadequate. Multiple calibration points can be set up to compensate for many of these situations.
Technically the meter performance curve is not linear, but a linear model is easier to use for setting up a
multi-calibrate table and minimizing the prover runs.
The EMR can accommodate 8 points of calibration (or 7 segments). The multipoint calibration has a 0.25%
difference between calibration points limitation built into the software. The 0.25% difference is based on a
Measurement Canada requirement.
There will however be circumstances where the meter performance is beyond the multipoint capability.
The difference between the high/full flow and the low flow requirements must be about 1.70% or less
difference for the multipoint calibration to work. You will need to make a calculation to determine your
situation.
Examples:
High flow K-factor: 94.100
Low Flow K-Factor: 92.475
Segments maximum of 7
Percent per segment:
(((High k-factor Low k-factor)/low k-factor)/7 *100) - must be 0.25% or less
(((94.100-92.475)/ 92.475)/7*100) = 0.251% NO GO, greater than 0.25%
High flow K-factor: 94.100
Low Flow K-Factor: 93.399
Segments maximum of 7
Percent per segment:
(((High k-factor Low k-factor)/low k-factor)/7 *100) - must be 0.25% or less
(((94.100-93.399)/ 93.399)/7*100) = 0.107% - Good
Might want to try 4or 5 segments.
At 5 segments ( 6 points) ,
0.107% *7/5 = 0.150% (< 0.25%)
Example
Each segment (94.100-93.399)/5 = 0.140
K-factor
Flow Rate
93.399
55
93.539
100
93.679
145
93.819
190
93.959
235
94.100
280
In the real world you get a close a practical in the segments and flow rates.
How To Determine The High And Low K-Factors In The Field And Minimize Prover Runs
27
For How to make corrections in calibration without re-calibrating see page above.
Temperature Probe Calibration
14. Disconnect the Temperature probe from the back of the display.
15. Unscrew the temperature probe from the pipe/thermowell.
16. Reconnect the temperature probe to the back of the display
17. Immerse the temperature probe in an ice-water bath 1quart for 5 minutes.
18. In C&C Mode, Arrow down to Temperature
19.
20.
21.
22.
23.
24.
25.
26.
NEXT, Fahrenheit
NEXT, CALIBRATE
NEXT, flashing cursor
Use the NEXT key to move cursor and the keypad or arrow keys to set the temperature to 32F.
ENTER, ENTER, ENTER back to Temperature.
Disconnect the Temperature probe from the back of the display.
Screw the temperature probe into the pipe/thermowell.
Reconnect the temperature probe to the back of the display
28
Push the Mode button so the small arrow is above Setup at the bottom of the screen. The display should say DEFAULTS.
Arrow (+) up to SYSTEM ADDRESS
NEXT , IB ADDRESS 1
Arrow down to PORT 2 ASSIGN,
NEXT, it will say PC SETUP
Arrow down to OBC
ENTER
ENTER back to SYSTEM ADDRESS
29
30
NO PRICE ENTRY
While trying to START a delivery, in the Currency Mode prior to selecting a price, the EMR will
display this message. Once the desired price is on the preset display, use the ENTER button to select
it.
SELECT PRODUCT
When more than one product is available, the user must choose one product prior to each delivery.
Press +/- button to display the desired product then press ENTER to select it.
Delivery Error Codes
The current delivery will automatically FINISH when a delivery error occurs.
E04 - ENCODER PULSE ERROR
This error is an indication that an illegal quadrature state has occurred. It is possible this error is
caused by a random event and another delivery may be started. Should this error repeat itself, the
source of the error must be fixed. Before replacing the encoder check the IB input voltage for dropout
or power surges. Some solenoid valves do not have diode suppression and may cause the supply
voltage to drop out. Add diodes (Veeder-Root P/N 846000-022) across the solenoid coils as noted in
the installation manual. An over speed condition (>1250 rpm) will also cause an E04 error.
E06 - TEMP PROBE TOO HOT
The measured temperature is greater than +160F (71.1C). This error can also be caused by poor
connections in a junction box when wire is added to the original temperature probe and a poor
connection is made. Soldering is recommenced - not just wire nuts.
E07 - TEMP PROBE TOO COLD
The measured temperature is less than -55F (-48.3C). When the fluid temperature is outside of the
EMR 's measurable range, a delivery can not continue or start. Go into the C&C Menu, under
TEMPERATURE, check what the EMR reads for the fluid's temperature. If the EMR temperature is
wrong remove or replace the thermoprobe and recalibrate.
E08 - TEMPERATURE PROBE FAILURE
The thermoprobe wires are either shorted together or are not connected to the TP terminals on the
Display Head. Check the thermoprobe connections on the back of the Display Head. If the
thermoprobe wiring is corroded, remove both leads and trim them. Cut back the cable insulation and
strip each wire by 1/4 of an inch (5 mm). Insert both wires into the terminal block and tighten. To test a
thermoprobe, connect each lead to an ohmmeter and measure the resistance. At 77F (25C), the
meter should read about 10 K ohms. If not, replace the probe (Veeder-Root P/N 846000-002).
Startup Error Codes
The following messages apply to self-testing during the power up sequence. Should any of these
messages appear, try cycling the power before replacing the Display Head Module.
E05 - NVRAM FAIL
Test of nonvolatile random access memory. The NVRAM is a piece of semiconductor memory
hardware used as a data storage device. To either write or read data into NVRAM the CPU (central
processing unit) places address and data information on the bus lines and then toggles the
appropriate control signals. The EMR3 uses NVRAM to store setup values, the current amount of
product dispensed and other historical data. Each value stored in NVRAM has a checksum that
validates the integrity for each piece of data. This error occurs when data being read does not agree
with its respective checksum.
10
The actual Proof of Gross (and/or compensated) Volume is NOT done under the Meter
Calibration. Once the meter is calibrated, Proofs or Checks are done in Standard delivery
mode.
For Initial Calibration or Re-Calibration if the Proof does not meet local criteria:
EMR AUTOCALIBRATE CALIBRATION STEPS
1. Standard VOLUME MODE on Display (Gallons)
2. Break Seal
3. Remove or lower Left Rear Bolt. (7/16 hex)
4. Push MODE button to get to SETUP. Display will show LANGUAGE.
5. Arrow down to METER CALIBRATION
6. NEXT PROD CAL 1
7. NEXT, AUTOCALIBRATE
8. NEXT: Prompted with START DISPENSE
9. Push the START BUTTON.
10. Fill the Prover Pay attention to the prover volume and sight glass.
11. When the prover is filled: Push the FINISH button. Prompt: ENTER NTC VOLUME
12. Using the NEXT key to move the cursor in the Preset Field and the Arrow Keys to change the
values. Enter the Non-Compensated Prover Volume.
13. Push the ENTER key and a new calibration factor will be Calculated.
14. Record this K-Factor
15. Hit ENTER to get back to METER CALIBRATION
16. Insert the Left Rear Bolt.
17. MODE button to get to Volume (Gallons) Mode.
18. Clear the Ticket.
19. Run check under normal Volume/Delivery Mode.
20. If Temperature Compensated, the Delivered volume will have the thermometer.
21. After pushing FINISH to complete the transaction, Push the down arrow to display the NonCompensated Volume.
If your non-compensated checks in standard Volume/delivery mode are consistent, but out of tolerance,
a new K-factor can be calculated per the attached information and entered in Manual Calibrate. Or, an
Autocalibrate run can be made to calculate a new K-factor.
D. Rajala Veeder-Root
4-2007
Detailed Steps to manual calibrate to accept the calibration factor if product added after calibration:
1.
2.
3.
4.
5.
6.
7.
8.
9.
D. Rajala Veeder-Root
4-2007
11. To change FLOW RATE : NEXT to get flashing cursor and use arrow keys to change numbers,
THEN enter
12. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
13. ENCODER COUNTS and value in preset field
14. To change ENCODER COUNTS, NEXT SET COUNTS
15. Use the NEXT key to move the cursor and the arrow keys to change the numbers.
16. Once the number is changed, hit ENTER. CAL RATE 1
17. For another point Arrow down to CAL RATE 2, if not, ENTER back to Meter Calibration.
18. To change FLOW RATE : NEXT to get flashing cursor and use arrow keys to change numbers,
THEN enter YOU MUST ENTER A FLOW RATE for each additional point.
19. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
20. ENCODER COUNTS and value in preset field
21. To change ENCODER COUNTS, NEXT SET COUNTS
22. Use the NEXT key to move the cursor and the arrow keys to change the numbers.
23. Once the number is changed, hit ENTER. CAL RATE 2
24. Repeat for additional calibration points.
25. Then ENTER(s) to get back to METER CALIBRATION.
26. Insert the Left Rear Bolt.
27. MODE button to get to VOLUME (Gallons) Mode.
28. Clear the Ticket.
29. Run checks under normal Volume/Delivery Mode.
METER CALIBRATION INFORMATION
There are now 4 Calibrations Available. For multiple products,
If using Temperature Compensation, calibrate the Temperature Probe first as noted previously.
The system calibration is running a known Non-Compensated volume through the meter and having the
register electronics calculate the Calibration Factor. The calibration factor is Encoder Counts per Unit
Volume. During the Auto or Multi calibrate procedures, the register remembers the number of counts
during flow (meter turning). Once the Non-Temperature Compensated volume, Gross Volume, of the
Prover vessel is entered, the register calculates the Calibration factor.
The actual Proof of Gross and Compensated Volumes is NOT done under the Meter Calibration.
Once the meter is calibrated, close the C&C switch, and return to Volume Mode where the Proof run is
made. Here the register applies the Density or Coefficients to obtain the Compensated Volume. If the
displayed volume is Compensated, the thermometer icon to the right of the delivered volume is
displayed. The Gross volume is displayed by pushing an arrow button. These Compensated and Gross
Volumes are compared to the Prover Volume and W&M tables to verify the meter calibration.
AUTO CALIBRATE is the easiest option. If the system will normally be run at only one flow rate, then
Auto Calibrate at that flow rate will be fine. Follow the procedure noted in the Auto Calibrate section of
the Setup and Operation Manual.
IMPORTANT: Once the GROSS prover volume is entered and the ENTER button is pushed and the
Register calculates the calibration factor (the Descriptor Field changes and a new calibration factor
shows), RECORD THIS FACTOR for future reference. Then push ENTER UNTIL YOU RETURN TO
METER CALIBRATION. Returning to METER CALIBRATION saves the Calibration Factor.
MANUAL CALIBRATION is an option available as outlined in the Manual. It is normally used for entering
known calibration values or normal values for specific meter if known.
D. Rajala Veeder-Root
4-2007
If for some reason the display head is replaced, or Restart is selected and the stored information in the
display head is lost, the Calibration Factor, if recorded, is entered under Manual Calibration.
Note: For Large Volume prover applications, a prover vessel metal expansion factor may need to be
applied per the local Weights and Measures regulations.
Recommend tenths resolution on Calibration even if whole units on Register for Delivery
To get the best accuracy on the calibration, Both register resolution and preset resolution (for k-factor)
should be temporarily set to Tenths. For calibration, the seal must be broken anyway, so the
resolution can be changed for the calibration and proof, then set back to whole units if that is the
delivered resolution.
With C&C switch open, go to DISPLAY SYNTAX. Go to PRESET RES, PRESET VOLUME and move
decimal using NEXT, the ENTER back to PRESET RES to save. Do the same with REGISTER RES,
REGISTER VOL.
Remember To Go Back To Whole Units When Complete With Calibration And Proofs.
MULTICALIBRATE AND METER CAPABILITY WITH THE EMR
Sometimes meters are run at 4 to 1 or greater turndowns or the meter has some wear, making a single
point calibration inadequate. Multiple calibration points can be set up to compensate for many of these
situations.
Technically the meter performance curve is not linear, but a linear model is easier to use
for setting up a multi-calibrate table and minimizing the prover runs.
The EMR can accommodate 8 points of calibration (or 7 segments). The multipoint calibration has a
0.25% difference between calibration points limitation built into the software. The 0.25% difference is
based on a Measurement Canada requirement.
There will however be circumstances where the meter performance is beyond the multipoint capability.
The difference between the high/full flow and the low flow requirements must be about 1.70% or less
difference for the multipoint calibration to work. You will need to make a calculation to determine your
situation.
Examples:
High flow K-factor: 94.100
Low Flow K-Factor: 92.475
Segments maximum of 7
Percent per segment:
(((High k-factor Low k-factor)/low k-factor)/7 *100) - must be 0.25% or less
(((94.100-92.475)/ 92.475)/7*100) = 0.251% NO GO, greater than 0.25%
High flow K-factor: 94.100
Low Flow K-Factor: 93.399
Segments maximum of 7
Percent per segment:
D. Rajala Veeder-Root
4-2007
How To Determine The High And Low K-Factors In The Field And Minimize Prover Runs
1. K-Factor at high flow rate
I suggest doing an AutoCalibrate at the high flow condition. Then go to Volume Mode (standard
delivery mode) and run a couple of checks for consistency and verification of the k-factor. During
the prover filling, push the Mode button to rate, observe and record the average flow rate. If the
results of the register when compared to the prover are consistent, but outside the accepted
tolerance, calculate a new K-factor manually, enter that k-factor using Manual Calibrate, and verify
with one or two prover runs.
2. K-Factor at Low flow rate
With the high flow rate K-factor, execute 2 or more runs at the low flow rate condition. During the
prover filling, push the Mode button to Rate, observe and record the average flow rate.
If the results are within tolerance, great. If consistent, but not within tolerance, calculate a new Kfactor for the low flow condition.
3. Manually, enter that low flow k-factor using Manual Calibrate, and verify with one or two prover runs.
4. Do the High-Low acceptability check outlined on page 1. If not okay, then the meter is incapable.
If Okay, calculate a K-factor/Flow Rate table. Use a number of segments that makes the calculation
easy.
5. Manually enter that k-factor/Flow Rate table using Manual Calibrate, and verify with one or two
prover runs.
For How to make corrections in calibration without re-calibrating see page 2 above.
D. Rajala Veeder-Root
4-2007
D. Rajala Veeder-Root
4-2007
Training Program
Updated: May 2007
125 Powder Forest Drive - Simsbury, CT 06070 USA - (860) 651-2766 www.veeder.com
TABLE OF CONTENTS
SECTION 1
Class Agenda
Introduction
Veeder Root
Gilbarco/Veeder Root
Danaher
SECTION 2
SECTION 3
SECTION 4
SECTION 5
SECTION 6
SECTION 7
SECTION 8
PC Interface/Set-up Program
Wireless Data Transfer
SECTION 9
Review - Certification
Written (open book)
Practical Hands On (open book)
SECTION 10
SECTION 1 Introduction
Veeder Root (Subsidiary of the Danaher Corporation):
Veeder-Root innovations have met the petroleum industry's growing needs for nearly a century.
Veeder-Root's beginnings can be traced to 1866, when the Root Company, a hinge manufacturer,
was founded in Bristol, CT. The following year, the company expanded to manufacture counting and
measuring devices for production, textile equipment and census taking.
Almost 30 years later in 1895, Curtis Veeder, the founder of Veeder Manufacturing Company,
invented a Cyclometer to record the miles traveled on a bicycle. He promoted the Cyclometer with
the slogan; its Nice to Know How Far You Go. The Cyclometer's success led to a full line of Veeder
counting devices.
In 1928, the Root and the Veeder Companies merged to form Veeder-Root, Incorporated of
Hartford, Connecticut, the largest manufacturer of counting and computing devices in the world.
While the majority of Veeder-Root's customers used the counters to control manufacturing
processes, in the early 1930s, one customer had the foresight to incorporate a Veeder-Root counter
into his gasoline pump. This innovation introduced Veeder-Root to the petroleum industry and
marked its beginning as a petroleum technology leader. These new mechanical computers allowed
gas station attendants to "filler up" without having to calculate cash sales as they pumped gas.
Through the 1940s, 1950s and 1960s, Veeder-Root counters went through many changes because
of wartime material shortages, growing competition and the addition of tenths of cents to the
counters. Through these decades of change, Veeder-Root mechanical computers could be found on
more than 90 percent of U.S. and Canadian gasoline pumps.
Around the same time, Veeder-Root engineers were developing an accurate meter register and
ticket system for gasoline delivery tracking. Their experiments with inventory control devices
sparked ideas that later surfaced in Veeder-Root's Tank Level Sensing (TLS) product line during the
late 1970s.
With the late 1970's gas crisis, gas station owners needed accurate inventory control and the ability
to detect product loss. During this time, Veeder-Root pioneered the automatic tank gauge and
released the first generation of TLS monitoring products, the TLS-100, in 1979.
Veeder-Root purchased Red Jacket Pumps in 2001 and moved the Manufacturing to Altoona.
Gilbarco/Veeder-Root
In 2002, Danaher (Veeder-Root Parent) purchased Gilbarco. Veeder Root and Gilbarco are aligned
under the same Senior management. Veeder-Root including Veeder-Root and Red Jacket
are a Profit Center.
Gasboy was purchased by Gilbarco/Veeder Root in 2003 and is incorporated under the umbrella.
Danaher Corporation:
Danaher is a leading U.S.-based manufacturer with a worldwide workforce of approximately 45,000.
The company generated revenues totaling 9.5 billion (US) in the 2006 fiscal year, almost half of
which was generated outside the U.S.A. With leading businesses in Professional Instrumentation,
Industrial Technologies, and Tools & Components, Danaher makes long-term investments in
established and internationally active industrial enterprises.
The origin of "Danaher" goes back to the root "Dana," a Celtic word dating from before 700 BC and
meaning "swift flowing". In the early 1980's, the vision of a manufacturing company, dedicated to
continuous improvement and customer satisfaction, was conceived during a fishing trip on the
Danaher, a tributary to the south fork of the Flat Head River in western Montana. The founders of
the company adopted the name for their new organization.
As Danaher has evolved, the elements of a swift flowing river have been retained. The company has
never strayed from the clarity of its initial vision. The flow of the business is ever changing but the
guiding principles--continuous improvement and customer satisfaction--remain constant. Over time,
the company has grown rapidly in size and success, achieving record levels again in 2004.
Danaher products help to make life healthier, safer and more enjoyable for people around the world.
Some of the products carry high-profile brand names, while others are hidden inside complex
systems. Danaher capitalized on favorable market conditions, achieving strong organic growth,
record sales, earnings, and cash flow in 2007.
Danaher has a proven system for achieving performance: the Danaher Business System (DBS). It
drives every aspect of the groups culture and performance. DBS is system of continuous
improvement and used to guide and measure everyday activities.
The Danaher Business segments include Professional Instrumentation, Industrial Technologies and
Tools & Components.
The Professional Instrumentation segment encompasses three strategic platforms: Electronic Test,
Environmental, and Medical Technologies.
The Industrial Technologies segment encompasses two strategic platforms: Motion and Product
Identification and three niche businesses; Aerospace and Defense, Power Quality, and Sensors and
Controls.
The Tools & Components segment encompasses one strategic platform, Mechanics Hand Tools,
and four focused niche businesses Jacobs Chuck, Delta Consolidated Industries, Hennessy
Industries, and Jacobs Vehicle Systems.
The EMR Electronic Meter Register System is independently approved by the UL and ATEX
organizations. These approvals apply to the EMR System when installed per the specifications
provided by Veeder-Root (See the Installation Manual for additional information). Systems not
installed per the Veeder-Root specifications may result in a violation of the safety approvals and or a
violation of the warranty for the EMR System (See page 4 of the Installation Manual for Safety
Warning).
Organizations or individuals associated with the installation of individual components of the EMR
System, such as but not limited to the 3-Way Differential Lockout Valve or the 2-Stage Petroleum
Preset Valve, must adhere to the practices applicable by law and by industry standard. For
example, if a flange needs to welded onto the piping for installation of the 2-Stage Petroleum Preset
Valve, this procedure should be conducted by an ASME (or equivalent) certified welder who is
qualified in such procedures and who is cognizant of the procedures for safe welding on the vessel
to which the valve will be installed.
The following organizations are internationally recognized agencies whose published practices
should be familiar to the organizations or individuals associated with EMR installations. This list
does not constitute the only organizations recognized globally. For each EMR installation, the
installer should insure compliance with all relevant codes and practices from all associated agencies
whose practices are applicable at the local level.
NEC
SAE
NFPA
API
NPGA
ASME
DOT
Positive Displacement:
Positive Displacement meters, as stated by E.L. Upp work on the bucket principle of a bucket
alternatively filled and emptied. To keep operation from being a strictly batch operation, most PD
meters have multiple buckets that are geared and valved together so that while some buckets are
filling, others are emptying. With proper timing and valving, there is an uninterrupted flow through
the meter. The driving force for this action comes from the flowing stream as a pressure drop.
PD meters are designed to measure liquids. Major types of meters by technology include;
Positioned-Vane Rotary, BiRotor, Sliding Vane, Oval Gear, Rotating Cresent, Oscillating Piston and
Rotating Paddle.
Turbine:
Turbine Meters are velocity measuring devices. Flow passes through a free turning rotor mounted
coaxially on the meter body centerline and exits the body. Fluid (or gas) imparts an angular velocity
to the angled rotor so that the rotation is proportional to the flow rate. With accurate measurement
of rotor speed from mechanical gearing or magnetic pickup and by knowing the hydraulic area that
the flow is passing through, volume at line conditions can be determined.
Readout Devices:
Secondary systems are a part of any measurement installation for reading primary element signals
and the variables necessary to correct flow from flowing to base conditions. These elements fall into
three main categories: mechanical, pneumatic and electronic. All have applications in flow
measurement. Veeder-Root Mechanical Meter Registers qualify as mechanical readout devices
and Veeder-Root Electronic Meter Registers (EMR) qualify as electronic readout devices.
Mechanical Register:
The Meter Register is the outstanding choice for recording and displaying delivery data in a wide
variety of industries. In the petroleum refining and distribution field, it is used from the well to
pipelines and storage centers, then to gasoline and fuel oil delivery trucks. In the chemical
processing field, it can be used for displaying totals during mixing and batching operations.
The Meter Register totalizes and displays high speed deliveries and transactions in large easy-toread figures. Ideal for increased efficiency in all kinds of liquid inventory and delivery operations.
Utilizing a geared drive available in a wide selection of ratios, the Meter Register records in
practically any unit of liquid measure - gallons, liters, barrels, etc. - as well as non-liquid units.
Special optional adaptors are available to mate the Meter Register with all standard meters and
pumps. It also easily combines with Veeder-Root Meter Preset and Printer to form a complete count,
control and print system.
Finished in corrosion resistant epoxy, the sturdy housing protects the Meter Register in demanding
environments and extremes in weather.
Electronic Meter Register (EMR):
The New Electronic Meter Register System from Veeder-Root is an alterative to the Mechanical
Register that has set the worldwide standard for quality, performance and value over the past four
decades. As the leader in Meter Registers, Veeder-Root has incorporated in the design of the
EMR, the most important features to meet the new demands of the market.
The Register Head (display) and the Interconnect Box (I.B.) are the building blocks of the system.
These along with a series of accessories allow the owner to configure a system that meets their
needs on the vehicle or in the terminal. Two Registers may be connected to a single I.B. and up to
16 Interconnect Boxes can be networked together with I.B. NET.
Note: The performance, range, repeatability and tolerances of each
type of flowmeter is established by the meter manufacturer.
Glossary:
The following glossary provides information on metering terms that may be helpful to an installer of
the EMR. EMR installers should possess the appropriate manuals for the flowmeter to which the
EMR will be mounted.
Acceptance Test- A test of the measuring device, under carefully controlled conditions with expert
operators and a certified inspector, to determine that the device will comply with the requirements
for approval, for sale in a state requiring such approval. Typically these are the requirements found
in NIST Handbook 44. In many (not all) situations, a national approval will result in automatic
acceptance at the state level.
Accumulative- A term applied to indicating elements or recording elements to describe the fact that
they are not resettable to a zero indication in normal use.
Accuracy- Freedom from error, usually expressed in percentage.
Accuracy Curve- A graph or plotted curve showing the performance characteristics of a measuring
device. The information plotted is the amount of error at the rates of flow between the minimum and
maximum rated capacity of the device.
Adjuster- A ratio changing device which is used to obtain agreement between the indicated and/or
recorded volume and the actual volume measured. The ratio may be either continuously variable or
variable in increments. Typical with LC, Smith, Brooks and other PD meters.
Air Eliminator (see also Vapor Eliminator)- A device installed upstream of the measuring device
to avoid measuring air or vapor. Air or vapor measured with liquid will result in registering a volume
larger than the liquid volume.
Air Separator- A device of the air eliminator type designed to include the capacity of separating air
vapor intimately mixed with the liquid. It may include one or more air eliminator mechanisms.
Air (Vapor) Control System- An arrangement of various elements intended to assist the air
eliminator - or air separator - by stopping or reducing flow of liquid when air or vapor is detected.
Ambient Temperature- Literally, this is the temperature of the surroundings. It is usually used to
denote the temperature of the atmosphere in a given location at a specific time.
Back Pressure Valve- A device intended to maintain a desired minimum upstream pressure.
Bubble Point- The conditions of temperature and pressure under which a liquid will begin to form
vapor.
Calibration- The process or procedure of setting or bringing a meter or a prover into agreement
with an established standard.
C&C Calibration and Configuration - EMR mode that is Weights and Measured sealed where
key system set up is done including product and calibration.
3
Change Gears- A gear system located between the meter and the indicating and recording
elements usually used to provide registration by the counter in the desired units of measurement.
Also used to bring the registration closer to true value. These gears may be located in the meter or
be part of a subassembly between the meter and the indicating elements. Typical with Neptune
meters and registers.
Check Valve- A device designed to prevent a reversal of liquid flow.
Clingage- The liquid film that adheres to the inside surface of a container after it has been emptied.
Coefficient of Expansion- A number usually expressed as a decimal that indicates the change of
volume per unit of volume per degree of temperature change.
Counter (Register)- A device which indicates a quantity related to the volume measured by the
meter. Can be mechanical or electronic.
Delivery, Over - Delivery, Under- The terms, over-delivery and under-delivery, are commonly
used to indicate the kind of measuring error that is experienced. Over-delivery indicates that the
counter is registering a greater volume than is delivered by the meter, and under-delivery indicates
that the meter is delivering less than is being registered on the counter.
Density, Relative (Specific Gravity)- The relative density t1/t2 of a solid or liquid substance is
defined as the ratio of the mass of a given volume of the substance at a temperature (t1) to the
mass of an equal volume of pure water at a temperature (t2) - (this term replaces the normal term
specific gravity).
Differential Pressure Valve- A device designed to maintain the total pressure on a system. In this
system, the desired pressure must be higher than the vapor pressure of the liquid when being
metered.
Drainage Time- The drainage time for test measures and provers shall be 10 seconds for a
capacity of 10 gallons or less, and 30 seconds for capacities exceeding 10 gallons. Necessary to
achieve uniform clingage film.
Error- The difference between the indicated value and the true value.
Flow Limiting Device (Flow Control)- A device installed in the system to prevent the rate of flow
through the meter from exceeding the desired maximum flow rate.
Flow Meter- Commonly used to describe a liquid measuring device.
Flow Range- The minimum and maximum flow rates established by the manufacturer to provide
the maximum meter performance and accuracy with long life. If exceeded, accuracy or meter life
may be adversely affected.
Flow Rate- The rate of flow of liquid expressed in volume or mass units per unit of time.
Gravity, API- A measure of the specific gravity of a liquid hydrocarbon as indicated by a
hydrometer having a scale graduated in degrees API.
Laminar Flow- Liquid flow in which the liquid elements travel along relatively straight parallel paths.
Low Vapor Pressure Liquid- A liquid which at atmospheric pressure and temperature can be
maintained in an open system without boiling.
Master Meter- A meter used as a reference for testing other working meters.
Meniscus- A curved liquid surface at the end of a liquid column.
Positive Displacement Meter- A meter in which the primary sensing element separates the liquid
into measured segments of known volume providing a shaft motion or other signal that can be used
for indicating the volume delivered.
Preset Device- An indicating device that can be preset for a desired quantity which will stop the
flow automatically, or provide a signal when the preselected quantity of liquid has been metered or
delivered.
Pressure Loss (Pressure Drop) (Loss of Head) (Differential Pressure) The difference in pressure between the inlet and outlet of a device while operating.
Prover (Proving Tank) - Volumetric Type- A closed or open vessel designed especially for
accurate determination of the volume of a liquid delivered into or out of it during a meter run. The
volume of liquid is either observed from the liquid level in a gauge proving glass or is known from
previous calibration of a fixed-volume vessel.
Recording Element- A device designed to register and print the volume metered on a ticket, slip
tape, or card.
Reference Standard- A volumetric measure that has been verified to contain or to deliver a
known volume, which is used for establishing the volume of field and working standards.
Register- A device which indicates the quantity measured by the meter.
Registration, Over - Registration, Under- The terms, over-registration, are commonly used to
indicate the kind of measuring error that is experienced. Over-registration indicates the counter is
registering a greater volume than is delivered by the meter, and under-registration indicates that the
meter is delivering more than is registered in the counter.
Repeatability- A measure of the deviation of a series of test results from their mean value, all
determinations being carried out under identical conditions.
Reynolds Number- A number characteristic of the flow of fluid in a pipe or past an obstruction.
5
Slit Compartment Test- A test applied to a truck mounted metering system which simulates the
conditions encountered in actual operation when one truck compartment is emptied during a
delivery.
Strainer (Filter)- A device equipped with a porous medium or wire mesh element to prevent
particles of foreign material from passing through a meter.
Swirl Plate- A plate or baffle installed in or near the outlet connection of a storage container to
reduce or prevent formation of a vortex.
Tolerance- The amount of error that will be allowed. It is a plus or minus value.
Torque Load- The power required of the meter for driving accessory equipment such as counters,
printers, etc.
Turbulent Flow- Flow in a pipeline is turbulent when the liquid elements no longer travel along
straight parallel paths. This usually occurs when the Reynolds Number is well above 2,000.
Vapor Eliminator- A device installed upstream of the measuring device to avoid measuring air
and/or vapor.
Vortex- The swirling motion of liquid often encountered as it enters the outlet opening of a
container. The tendency of a vortex to form depends upon many details of the container
construction. Since the swirling liquid can form around a central column of air or vapor, the flow
area is reduced. This restricts the rate of flow and often results in carrying considerable quantities
of air or vapor with the liquid.
Connections:
A good Ground to the Truck chassis or Earth ground to the Interconnect Box is important.
On many newer vehicles, the V- of the battery may not be common with the chassis
ground or the defined ground in the vehicle electrical distribution box. The Input Power
Ground at the top of the EMR Interconnect Box terminal strip should come from the V-
terminal of the battery or the location where the V- battery terminal is directly connect to
the vehicle chassis. If another ground point is used, there may be insufficient power for
the display to operate properly.
Pay close attention to the connections on the terminal strip on the back of the display
head.
Caution: Do not power up the system until all connections have been verified. The IB is
protected by a fuse for incorrect power into the box, however the Intrinsically Safe side of the
IB that powers and communicates with the head does not have this type of replaceable
protection due to Agency Approval constraints.
Diode Suppression required for 2-stage valve. 2 kits per valve. If purchasing valve from
Veeder-Root, diode suppression comes with the valve.
2-stage solenoid valve kit has cable, but no electrical junction boxes
LPG valve does not require diode suppression, kit has cable, but no hoses and fittings
Maximum Distances between Components:
Installation Manual, page 4, EMR Control Drawing (for truck installations). This
configuration will apply to the majority of the EMR systems that are installed. ( Terminal on
Page 36)
Specific Installation Kits for connecting to Meter - Pages 6-19
Neptune Kits base on if flow meter CURRENTLY has Temperature Compensation or Not.
Small groove pins are in the kits that can get lost
Prepare your workspace:
2
Working with a group no greater than two persons to one EMR Head, place your components on
the table in front of you. Place the tools and the kits that the instructor has provided to one side and
align the components with the Display Head in front of you with the Interconnect box next to it. For
the training, your instructor will provide power supplies which will provide the 12 Volts DC necessary
to operate the system. Place the printer behind the Display Head and Interconnect Box at this time.
DISPLAY HEAD
Optional Keypad First Before Other Items
Pages 21-22
The optional keypad greatly decreases the time necessary for data entry during setup and it
is recommended that all certified installers have an Optional Keypad with longer wires for
such use in the field for set-ups where significant label changes will be made.
Installation of Optional pulse generator for Demo
DO NOT CONNECT WIRES FROM INSTALLED PULSE GENERAOR TO DISPLAY a
demonstration pulse generator will be installed.
Display Head on stand
Put shaft into housing.
Invert the display head and put stand on with two bolts provided
Invert upright.
Connect spring between shaft and pulse generator.
Wire Harness and Temperature Probe
Pages 22-24 of the Installation Manual. Do not tie wrap any cables at this time.
Use the special cord grip top and put wire and Temp Probe together(non-standard).
Review and Tie Wrap wires
Validate that all assemblies are installed correctly and all wire terminations are correct.
Route and tie wrap cables as shown on page 22, or as directed by Instructor.
Special Encoder for Demonstration
Install the Extra Pulse Generator/Encoder per your Instructors directions.
INTERCONNECT BOX
If all four cover screws of the Interconnection Box are installed, remove the right two and leave two
(creating a hinge). Refer to Figure 19 on page 26.
Knockouts
Remove specific knockouts per instructor, and install the box part of the cord grips.
Wire Display Heads and Printer Cable per Figure 20, Page 28
Use the ground wire on the display head as and anchor for strain relief
V+
GND
In this example:
TX IB to 9, RX IB to 12,
Idle state is 1 and 4 to DataLink modem cable: 1 to TX of Black wire of Serial cable to
DataLink Modem. 4 to RX or Green wire of Serial Cable to DataLink modem.,
Terminal Strip
V+ (13)
RX (12)
RS232 Serial
Output
TX (9)
RX (8)
TX (5)
GND
2PDT
Relay/Block
14
12
COM (14)
Display
Head 1
Relay 2
11
13
Datalink
Modem
10
RX (4)
TX (1)
GND
NO
COM (14)
Display
Head 2
Relay 2
NO
DESCRIPTION:
The 2-Stage, 2-inch valve Kit contains: Two: 0846000-022 Diode Suppression Kits with the
valve. These need to be installed per the diagram in these instructions or the EMR
Installation Guide, Manual 577013-758.
VALVE SPECIFICATIONS:
DIMENSIONS:
SINGLE FLOW IN IB
(RELAY 2)
FAST FLOW IN IB
(RELAY 1)
FLOW
Display Head
Plate in IB
HEAD 1
HEAD 2
Diagram in Installation
Manual
Set-Up Manual
RELAY CONTROL
SET ADVANCE STOP (2nd Stage Knockoff)
---------------------------RELAY CONTROL
SET SLOW FLOW
(1st Stage Knockoff)
RELAY CONTROL
SET ADVANCE STOP (2nd Stage Knockoff)
---------------------------RELAY CONTROL
SET SLOW FLOW
(1st Stage Knockoff)
Connect the red wires from each solenoid in the junction box with the diode suppressor as shown.
Connect the wires in the Interconnect Box per the diagram, providing power to the solenoid from the
V+ input power from the vehicle battery or accessory panel. Connect the wire from the solenoid to
the appropriate DH Relay per the previous page information. Reference: EMR Installation Guide
Manual 577013-758.
WARRANTY
This valve is under Warranty from the Manufacturer. Contact Veeder-Root for the Manufacturer
information.
VEEDER-ROOT
th
P.O. Box 1673, Altoona PA 16603-1673, 6 Avenue at Burns Crossing (16602), 814-695-4476, fax 814-695-7605
252135 Rev. A
10
Enable printer
Power down
Disconnect head 1
Connect head 2
Enable Printer
Power down,
Power up
Delivery Demonstration
Display and Standard Product Features Explained and Demonstrated
Simple Pump and Print Showing Temp Comp
Volume Preset
Currency
Currency preset
Multiple Deliveries
STUDENTS MAKE DELIVERIES
Make Pump and print, preset, and currency deliveries on the EMR System using the Operator Card
as a guide. You will learn the basic functionality of the system. Pages 46-50 are the details
Following learning deliveries, you will learn how to setup, configure, and calibrate.
VEEDER-ROOT
Basic Operating
Instructions
TO MAKE A DELIVERY
1.
TICKET INTO PRINTER: Press the RELEASE key on the Epson Printer, insert a ticket until the Paper Out Red
light goes out, and press the FORWARD key to engage the ticket. (The Release light goes out)
2.
Push FINISH to CLEAR THE DISPLAY of the PREVIOUS DELIVERY amount. Zeros appear in the Register
field on the display.
3.
DELIVERY: Push the START button on the Display. The Nozzle Symbol
will appear on the display in the upper left hand corner.
4.
FILL THE TANK: Pull the delivery hose and fill the customers tank.
5.
AFTER FUEL DISPENSE: When the delivery is complete, push the FINISH button. The ticket will print
automatically. The printer symbol will flash to indicate the ticket has printed.
6.
REMOVE TICKET: Push the RELEASE button on the printer to remove the delivery ticket.
Note: Once the filling begins, if flow is stopped for 3 minutes, the delivery will automatically terminate and a delivery
ticket will automatically be printed.
DUPLICATE TICKET
1.
TICKET INTO PRINTER: Press the RELEASE key on the Epson Printer, insert a ticket until the Paper Out red
light goes out, and press the FORWARD key to engage the ticket. (The Release light goes out)
2.
With the delivered amount showing to obtain a duplicate ticket. HOLD DOWN THE FINISH BUTTON FOR 3
SECONDS AND RELEASE.
MULTIPLE DELIVERIES
1.
To make a MULTIPLE TANK DELIVERY UNDER ONE SUMMARY TICKET, hold the START button in until the
Multiple Tank Symbol shows with the nozzle symbol.
The 3 minute time out is disabled under Multiple Delivery.
2.
3.
For an electronic record of each individual tank fill under the same ticket, push START for each tank.
The Multiple Delivery transaction is completed with FINISH, just like a normal delivery. The ticket will be a
SUMMARY, labeled MULTIPLE DELIVERY. Individual transactions under the Summary ticket are stored
electronically for printing via VIEW RECORDS.
PRESET DELIVERY
1. To make a preset delivery, push the PRESET BUTTON PRESET GALLONS shows.
2. Push the NEXT button, CHANGE PRESET shows.
3. Use the arrow keys to obtain a preprogrammed preset or use the NEXT button to move the cursor to a number
position and use the arrow keys to put a digit in the appropriate location in the preset field.
4. When the number you want is in the Preset Field, push ENTER to accept the value.
5. Begin delivering product. The preset window counts down as the delivery window counts up.
6. The system will slow down with some amount to go and then stop when the preset amount is delivered.
COMMON PROBLEMS AND SOLUTIONS
Cannot start next delivery: Push START, BUT NOTHING HAPPENS
Push finish to clear the display of the previous delivery (zeros appear ) then press start.
The printer symbol cannot be flashing. If flashing, remove the previous ticket from the printer and insert new
ticket.
The printer will not print::
Check to see if ticket is inserted properly and the paper out and release lights are out.
Make sure the printer power is on. Check on/off switch on side of printer.
Veeder-Root
Form Number
Rev. A
Push FINISH to CLEAR THE DISPLAY of the PREVIOUS DELIVERY amount. Zeros appear in the
Register field on the display.
5. Push the NEXT key. You will see a cursor at the beginning of the Tail number field in the register
under the N.
6. Push NEXT again, to move the cursor to the right of the N.
7. Using the Keypad, Type the TAIL NUMBER, the Keypad works just like a Cell Phone Keypad.
DO NOT push Enter until the number is correct.
If you make a mistake, you can move the cursor back around using the NEXT key on the
register. The clear key is the lower left of the keypad.
8. Once you have the TAIL NUMBER , push ENTER, either on the Keypad or, or the Register.
9. FILL THE AIRCRAFT: Connect Bonding/Grounding Cable, Pull the delivery hose and fill the
customers aircraft per Refueler instructions.
10. AFTER FUEL DISPENSE: When the delivery is complete, push the FINISH button.
Push FINISH to CLEAR THE DISPLAY of the PREVIOUS DELIVERY amount. Zeros appear in the
Register field on the display.
The Nozzle Symbol will appear on the display in the upper left hand
corner.
5.
6.
7.
8.
Once you have the TAIL NUMBER , push ENTER, either on the Keypad or, or the Register.
9.
FILL THE AIRCRAFT: Connect Bonding/Grounding Cable, Pull the delivery hose and fill the customers aircraft
per Refueler instructions.
10. AFTER FUEL DISPENSE: When the delivery is complete, push the FINISH button. The ticket will print
automatically. The printer symbol will flash to indicate the ticket has printed.
11. REMOVE TICKET: Push the RELEASE button on the printer to remove the delivery ticket.
1. Push FINISH to CLEAR THE DISPLAY of the PREVIOUS DELIVERY amount. Zero appears in the
Register field on the display.
2. DELIVERY: Push the START button on the Display.
3. The Nozzle Symbol will appear on the display in the upper left
hand corner.
4. You will get prompted to enter a TAIL NUMBER.
5. Push the NEXT key. You will see a cursor at the beginning of the Tail number field in the register. The
N may be used.
6. If using the N, Push NEXT again, to move the cursor to the right of the N.
7. Using the Keypad, Type the TAIL NUMBER, the Keypad works just like a Cell Phone Keypad.
DO NOT push Enter until the number is correct.
If you make a mistake, you can move the cursor back around using the NEXT key on the
register, or CLEAR. The clear key is the lower left of the keypad.
8. Once you have the TAIL NUMBER , push ENTER, either on the Keypad or, or the Register.
9. You will get a prompt for Selecting the Product:
OR
5.
6.
7.
8.
9.
Power down
Disconnect Head 2 PWR in the IB
Power up the system. It will recognize Head 1.
Check to make sure System Address IB Address 1, Head 1.
a. Push the Mode button so the small arrow is above Setup at the bottom of the screen.
The display should say DEFAULTS.
b. Arrow (+) up to SYSTEM ADDRESS
c. NEXT , IB ADDRESS 1 ( IF NOT 1, NEXT to SET IN ADDRESS, arrow to 1 and
ENTER)
d. Arrow down to HEAD ADDRESS, it should say 1. ( IF NOT 1, NEXT to SET HEAD
ADDRESS, arrow to 1 and ENTER)
e. ENTER to SYSTEM ADDRESS
Power down.
Disconnect Head 1 PWR in the IB.
Connect Head 2 PWR.
Power up the system.
At Head 2, Go to Setup, System Address, Make sure IB 1, then Head Address, and set to
Head 1.
a. Push the Mode button so the small arrow is above Setup at the bottom of the screen.
The display should say DEFAULTS.
b. Arrow (+) up to SYSTEM ADDRESS
c. NEXT , IB ADDRESS 1 ( IF NOT 1, NEXT to SET IN ADDRESS, arrow to 1 and
ENTER)
d. Arrow down to HEAD ADDRESS, it should say 1. ( IF NOT 1, NEXT to SET HEAD
ADDRESS, arrow to 1 and ENTER)
e. ENTER to SYSTEM ADDRESS
10
SETUP
Go to the Setup and Operation Manual. This Manual No. is 577013-766.
First Thing: INCREASE TIME DELAYS TO GET SCREEN AND BACKLIGHT TO STAY ON
LONGER page 28.
Next, etc.
Second thing: Turn On printer
11
For Product 1, NEXT, then go to PRODUCT TYPE FIRST, then NEXT. Select the Product
from the menu and push ENTER. This selects the Product and for Temperature
Compensation the associated tables. The PRODUCT NAME is automatically assigned after
selecting the PRODUCT TYPE. If you want to customize the PRODUCT NAME, that field is
editable. The PRODUCT TYPE will not change with a name change.
TEMPERATURE COMPENSATION
Defaults to Non-Compensated Arrow to TC COMPENSATED and ENTER for
Temperature Compensated Product.
The DENSITY or COEFFICIENT tables are displayed under that feature. Changes can be
made, but the default tables for the pre-loaded Product Types are Typically the Weights and
Measures approved ones.
CAL NUMBER allows 4 different calibrations. Any of these calibrations can be assigned to
any of the 8 products. This allows for different names to be used for the same product
Type. i.e. Clear Diesel and Dyed Diesel are both the same product Type and will have the
same Cal Number, but be different Products with different names.
Two REPORT FORMATS are available.
User Defined Product: A product not available in the tables can be set-up under User
Defined. For Temperature Compensation of the product, Density or Coefficient information
will need to be loaded under that menu item. At least two densities at their respective
temperatures are needed.
Remember to Always ENTER back to the Main Menu to save changes in C&C mode after making
your selections.
12
METER CALIBRATION
The actual Proof of Gross (and/or compensated) Volume is NOT done under the Meter Calibration.
Once the meter is calibrated, Proofs or Checks are done in Standard delivery mode.
For Initial Calibration or Re-Calibration if the Proof does not meet local criteria:
EMR AUTOCALIBRATE CALIBRATION STEPS
1. Standard VOLUME MODE on Display (Gallons)
2. Break Seal
3. Remove or lower Left Rear Bolt. (7/16 hex)
4. Push MODE button to get to SETUP. Display will show LANGUAGE.
5. Arrow down to METER CALIBRATION
6. NEXT PROD CAL 1
7. NEXT, AUTOCALIBRATE
8. NEXT: Prompted with START DISPENSE
9. Push the START BUTTON.
10. Fill the Prover Pay attention to the prover volume and sight glass.
11. When the prover is filled: Push the FINISH button. Prompt: ENTER NTC VOLUME
12. Using the NEXT key to move the cursor in the Preset Field and the Arrow Keys to change the values.
Enter the Non-Compensated Prover Volume.
13. Push the ENTER key and a new calibration factor will be Calculated.
14. Record this K-Factor
15. Hit ENTER to get back to METER CALIBRATION
16. Insert the Left Rear Bolt.
17. MODE button to get to Volume (Gallons) Mode.
18. Clear the Ticket.
19. Run check under normal Volume/Delivery Mode.
20. If Temperature Compensated, the Delivered volume will have the thermometer.
21. After pushing FINISH to complete the transaction, Push the down arrow to display the NonCompensated Volume.
If your non-compensated checks in standard Volume/delivery mode are consistent, but out of tolerance, a
new K-factor can be calculated per the attached information and entered in Manual Calibrate. Or, an
Autocalibrate run can be made to calculate a new K-factor.
SPECIAL CALIBRATION ACCEPTANCE IF MORE THAN ONE PRODUCT PRODUCT ADDED AFTER
INITIAL CALIBRATION
If all the products were set up and assigned Cal 1 before the calibration was done, they would all get that
calibration properly assigned to that product.
If a new product is added after the calibration of that Cal 1, say 25, the new product unfortunately is not
automatically given the 25, but gets the 96 factory default in the background. Going to Manual calibrate and
getting to Encoder counts, NEXT, then ENTER to "recalculate" gets the CAL 1 of 25 assigned to the new
product.
13
Detailed Steps to manual calibrate to accept the calibration factor if product added after calibration:
1. Standard Volume Mode on Display
2. Break Seal
3. Remove or lower Left Rear Bolt. (7/16 hex)
4. Push MODE button to get to SETUP. Display will show LANGUAGE.
5. Arrow down to METER CALIBRATION
6. NEXT PROD CAL 1
7. NEXT AUTOCALIBRATE
8. Arrow up to MANUAL CALIBRATE
9. NEXT CALB RATE 1 (Encoder Counts K-factor in Preset field, Flow rate in Volume delivered field.)
10. NEXT FLOW RATE (DEFAULT IS 60 and okay to leave if single calibration)
11. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
12. ENCODER COUNTS and value in preset field
13. To ACCEPT the ENCODER COUNTS, NEXT SET COUNTS and Cursor Flashes.
14. Push ENTER. Recalculation Occurs
15. Then ENTER(s) to get back to METER CALIBRATION.
16. Insert the Left Rear Bolt.
17. MODE button to get to VOLUME (Gallons) Mode.
How to make corrections in calibration without re-calibrating
Example:
Calibration Factor: 95.966 pulses per Gallon
Register vol: 100.4
Prover volume: 100.0
(Prover-reg)/prover = (100.0-100.4)/100.0 = - .004
New factor= old factor /(1+diff) note sign is important
New factor = 95.966/(1-.004) = 95.966/0.996 = 96.351
If register more than prover, factor up. (less Gallons/pulse) (greater Pulses/gallon)
If register less than prover, factor down (more Gallons/pulse) (less pulses/gallon)
The adjustment is made in Manual Calibration with the C&C switch open.
Manual Calibrate: For entering known calibration values or normal values for specific meter if known.
1. Standard Volume Mode on Display
2. Break Seal
3. Remove or lower Left Rear Bolt. (7/16 hex)
4. Push MODE button to get to SETUP. Display will show LANGUAGE.
5. Arrow down to METER CALIBRATION
6. NEXT PROD CAL 1
7. NEXT AUTOCALIBRATE
8. Arrow up to MANUAL CALIBRATE
9. NEXT CALB RATE 1 (Encoder Counts K-factor in Preset field, Flow rate in Volume delivered field.)
10. NEXT FLOW RATE (DEFAULT IS 60)
11. To change FLOW RATE : NEXT to get flashing cursor and use arrow keys to change numbers, THEN
enter
12. To get to ENCODER COUNTS from the FLOW RATE screen, Arrow UP
13. ENCODER COUNTS and value in preset field
14
15
Note: For Large Volume prover applications, a prover vessel metal expansion factor may need to be applied
per the local Weights and Measures regulations.
Recommend tenths resolution on Calibration even if whole units on Register for Delivery
To get the best accuracy on the calibration, Both register resolution and preset resolution (for k-factor)
should be temporarily set to Tenths. For calibration, the seal must be broken anyway, so the resolution
can be changed for the calibration and proof, then set back to whole units if that is the delivered resolution.
With C&C switch open, go to DISPLAY SYNTAX. Go to PRESET RES, PRESET VOLUME and move
decimal using NEXT, the ENTER back to PRESET RES to save. Do the same with REGISTER RES,
REGISTER VOL.
Remember To Go Back To Whole Units When Complete With Calibration And Proofs.
MULTICALIBRATE AND METER CAPABILITY WITH THE EMR
Sometimes meters are run at 4 to 1 or greater turndowns or the meter has some wear, making a single point
calibration inadequate. Multiple calibration points can be set up to compensate for many of these situations.
Technically the meter performance curve is not linear, but a linear model is easier to use for setting up a
multi-calibrate table and minimizing the prover runs.
The EMR can accommodate 8 points of calibration (or 7 segments). The multipoint calibration has a 0.25%
difference between calibration points limitation built into the software. The 0.25% difference is based on a
Measurement Canada requirement.
There will however be circumstances where the meter performance is beyond the multipoint capability.
The difference between the high/full flow and the low flow requirements must be about 1.70% or less
difference for the multipoint calibration to work. You will need to make a calculation to determine your
situation.
Examples:
High flow K-factor: 94.100
Low Flow K-Factor: 92.475
Segments maximum of 7
Percent per segment:
(((High k-factor Low k-factor)/low k-factor)/7 *100) - must be 0.25% or less
(((94.100-92.475)/ 92.475)/7*100) = 0.251% NO GO, greater than 0.25%
High flow K-factor: 94.100
Low Flow K-Factor: 93.399
Segments maximum of 7
Percent per segment:
(((High k-factor Low k-factor)/low k-factor)/7 *100) - must be 0.25% or less
(((94.100-93.399)/ 93.399)/7*100) = 0.107% - Good
16
How To Determine The High And Low K-Factors In The Field And Minimize Prover Runs
1. K-Factor at high flow rate
I suggest doing an AutoCalibrate at the high flow condition. Then go to Volume Mode (standard delivery
mode) and run a couple of checks for consistency and verification of the k-factor. During the prover
filling, push the Mode button to rate, observe and record the average flow rate. If the results of the
register when compared to the prover are consistent, but outside the accepted tolerance, calculate a new
K-factor manually, enter that k-factor using Manual Calibrate, and verify with one or two prover runs.
2. K-Factor at Low flow rate
With the high flow rate K-factor, execute 2 or more runs at the low flow rate condition. During the prover
filling, push the Mode button to Rate, observe and record the average flow rate.
If the results are within tolerance, great. If consistent, but not within tolerance, calculate a new K-factor
for the low flow condition.
3. Manually, enter that low flow k-factor using Manual Calibrate, and verify with one or two prover runs.
4. Do the High-Low acceptability check outlined on page 1. If not okay, then the meter is incapable. If
Okay, calculate a K-factor/Flow Rate table. Use a number of segments that makes the calculation easy.
5. Manually enter that k-factor/Flow Rate table using Manual Calibrate, and verify with one or two prover
runs.
For How to make corrections in calibration without re-calibrating see page 2 above.
17
CONFIGURE INPUT/OUTPUT
Emergency stop; A big red button, V-R sells one. Connections on Display head. Interrupts
transaction and closes relays, but does complete transaction. Can open ES and complete
delivery.
Remote Start/Stop: Connections in IB. Function same as START and FINISH buttons.
Momentary contact only.
Pulse Output ( before January 2006, Software E01 and before) :
The pulses per unit may need to be adjusted to account for potential lag in measurement and
rounding. For example, at 10 pulses per unit, the last 1/10 pulse is not sent until the volume
measured is equal to 1/10. Because the EMR rounds off, there is no agreement of indication.
At 20 pulses per unit, a pulse is sent at every tenth and every half a tenth, therefore the EMR
will agree with the other indicating device.
Remote Display: If enabled, setup under Initial Remote Display on page 26.
Interlock:
1. Printer Interlock: Not associated with Epson Printer. A switch closure between terminals 15
and 16 on the back of the display is required to start. Can be used as a permissive switch.
2. Preset Interlock: When enabled, requires a preset and presets enabled in Defaults. The
switch part of this is the Printer Interlock, terminals 15 and 16. If the Switch is open, the
preset is required.
3. Transfer Interlock: Connect the transfer switch between termnials 15 and 16. With the switch
closed, The label Transfer Ticket is printed on the Delivery Ticket. UK application.
4. Disable Interlock: Clears the interlocks.
18
19
DELIVERY OPTIONS
PRICE CHANGING: See manual description. Can now lock a price that cannot be changed.
Useful for set price deliveries or for Density for Mass Deliveries.
TEMPERATURE COMPENSATION: Density or Coefficient, Default is Coefficient.
DELIVERY TIMEOUT: Now is variable. Pre-flow is if no flow is seen. Post-flow is idle flow
after flow seen.
TANK ID: The individual transaction can be tagged with a Customer or Tank identification
number. This number is entered by the operator through the standard display or optional
keypad. This feature is enabled under C&C in DELIVERY OPTIONS. If Enabled, an ID must
be entered (or accept default).
E04 PULSE ENCODER ERROR: Used in special cases for French W&M.
GROSS VOLUME: BEGINNING WITH E02 SOFTWARE/FIRMWARE, THE GROSS
VOLUME CAN BE PRINTED AS AN OPTION. Enable it here.
is always
Once you have all the lines on the ticket you want, then go to the next line, push NEXT, and
then place END OF REPORT in that line. Anything between line 1 Advance and End of
Report is printed on the ticket.
Hit ENTER all the way back to the REPORTS FORMAT line to save your new ticket format.
For the actual location of the printing up-and-down the ticket, the Ticket Advance under the
non-C&C, Set-up item PRINTER OPTION. This is where you enable the printer, usually
select Slip Printer, Select Advance, and then enter the amount the ticket moves before it
begins to print.
Note:
You can edit the actual printer labels under the LANGUAGE category before setting up
the ticket.
For multiple tickets of the same type, we have available a small program that can
download the ticket format via RS-232.
If there is no data available, then the line will not print. i.e. if no price, then no pricerelated information will be printed on the ticket.
RESTART
NOTE: DO NOT RESTART UNLESS YOU WANT TO REBOOT TO FACTORY SETTINGS.
THIS MEANS YOU WILL LOSE CALIBRATION AND ALL SETUP INFORMATION
SECURITY CODES
Turned on under C&C page 44. Note capability under C&C to reset user codes to default.
Prompted on page 13, Setup on Page 17.
The security code system can be used to prevent access to changes in the set-up, or allow
only certain features to be changeable, for example allowing price changes if the Currency
Feature has been purchased. Up to 3 users with different security codes can be set-up.
The use of security codes is enabled under C&C, SECURITY CODE. The actual user
security codes are entered under Setup SECURITY after the Security Code feature is
enabled.
User 1 is always the Supervisor. The Supervisor has a passcode, can change all three
passcodes, and has default yes access to all Set-Up changes, but controls access to setup changes of Users 2 and 3. (C&C is still controlled by Sealing of the Head). All the User
Accesses to Setup changes are yes unless changed.
Logging in:
When Security Codes are enabled, the user is prompted when going to Set-Up for their
Security Code.
21
At the ENTER SECURITY CODE PROMPT, Select the USER. Next gets to LOGGED OUT.
Once the Security Code is entered and ENTER is pushed, LOGGED IN, appears. If
LOGGED IN appears for any user, then that user is Logged in. The specific USER is logged
in until LOGGED OUT is selected under that user and ENTER pushed. At that time
LOGGED OUT will be the only situation that appears under that user.
Security Notes:
Upon Powering down, the Security Code Status for all users is returned to LOGGED
OUT.
User 1, Supervisor, must log out or the capability of Changes will remain active if power is
not cycled.
Only one user can be logged in at a time. If another User logs in, other users are
automatically logged out.
Important Note: Security Codes do not prevent Dispensing, they only control Setup options.
22
Examples:
Example 1
$
$
%
$
$
%
$
$
$
$
$
$
100.0
1.1000
110.0000
6.0000
0.0660
6.6000
5.0000
-0.0583
-5.8300
1.1077
-5.50
6.27
110.77
When PRICE CODE 1 is selected under CURRENCY, the FINAL PRICE displayed in the PRESET
display location is 1.1077.
24
100.0
2.0000
200.0000
5.0000
0.1000
10.0000
210.0000 REGULAR PAY PRICE
3.0000
-0.0630
-6.3000
$
2.037
$
-6.00
$
9.70
$
203.70 EARLY PAY PRICE
$
$
%
$
$
$
%
$
$
Under PRICING, select PRICE CODE 2. Change the PRICE to 2.00. and Assign TAX/DIS 1 to this PRICE CODE 2.
Line Assignments:
Line 0 is Subtotal, Volume x base price
Line 1 is the Tax/unit
1. Under PRICING, select PRICE CODE 2. Change the PRICE to 2.00, and Assign TAX/DIS 2 to this PRICE
CODE 2.
2. Under TAX/DISCOUNT, select T/D CODE 2, T/D LINE 1 and CHNG T/D VALUE to TAX PERCENT 5.0000 and
ENTER. CHNG T/D ASSIGN to 0.
3. Go to T/D LINE 2 and CHNG T/D VALUE to T/D SUBTOTAL and ENTER. CHNG T/D ASSIGN TO both LINES
0,1. And ENTER back to main menu. This creates an Intermediate subtotal after the 5% tax is added.
4. (Note: This Printer Label can be changed to REGULAR PAY PRICE to identify to the customer what this
means.)
5. Go to T/D LINE 3 and CHNG T/D VALUE to DISCOUNT PERCENT 3.0000 and ENTER. CHNG T/D ASSIGN
TO both LINES 0,1. ENTER, ENTER. (Note Line 2 is the intermediate subtotal, it does not take assignments)
25
$
$
%
$
$
$
$
$
%
$
$
$
$
$
$
100.0
2.0000
200.0000
5.0000
0.1000
10.0000
210.0000 REGULAR PAY PRICE
0.1000 0.10 DIS PER GALLON
-10.0000
5.0000 Remove tax on the 0.10
Discount.
0.0050
0.5000
2.050
-9.50
10.00
200.50 EARLY PAY PRICE
Under PRICING, select PRICE CODE 2. Change the PRICE to 2.00. and Assign TAX/DIS 1 to this PRICE CODE 2.
Line Assignments:
Line 0 is Subtotal, Volume x base price
Line 1 is the Tax/unit
1. Under PRICING, select PRICE CODE 3. Change the PRICE to 2.00, and Assign TAX/DIS 3 to this PRICE
CODE 2.
2. Under TAX/DISCOUNT, select T/D CODE 3, T/D LINE 1 and CHNG T/D VALUE to TAX PERCENT 5.0000 and
ENTER. CHNG T/D ASSIGN to 0.
3. Go to T/D LINE 2 and CHNG T/D VALUE to T/D SUBTOTAL and ENTER. CHNG T/D ASSIGN TO both LINES
0,1. And ENTER back to main menu. This creates an Intermediate subtotal after the 5% tax is added.
4. (Note: This Printer Label can be changed to REGULAR PAY PRICE to identify to the customer what this
means.)
5. Go to T/D LINE 3 and CHNG T/D VALUE to DIS PER VOLUME 0.10 and ENTER. CHNG T/D ASSIGN TO
both LINES 0,1. ENTER, ENTER. (Note Line 2 is the intermediate subtotal, it does not take assignments)
6. Got to T/D LINE 4 and CHNG T/D VALUE TO DIS PERCENT of 5 and ENTER. CHNG T/D ASSIGN TO LINES
0,1,3 . And ENTER back to main menu.
26
Example 4:
$
$
%
$
$
$
$
$
$
$
$
100.0
1.1000
110.0000
6.0000
0.0660
6.600
15.0000
15.0000
1.1660
15.00
6.60
131.60
Line Assignments:
Line 0 is Subtotal, Volume x base price
Line 1 is the Tax/unit
Under PRICING, select PRICE CODE 3. Change the PRICE to 1.10. and Assign TAX/DIS CODE 2
to this PRICE CODE 3. TAX/DIS CODE 2 has not been set up yet.
Under TAX/DISCOUNT, select T/D CODE 2, T/D LINE 1 and CHNG T/D VALUE to TAX PERCENT
6.0000 and ENTER. CHNG T/D ASSIGN to 0.
Go to T/D LINE 2 and CHNG T/D VALUE to SURCHARGE, NEXT, and enter 15.0000 as the
surcharge. Because this is a Delivery Surcharge, it is not assigned to a specific line.
Note that a Delivery DISCOUNT SURCHARGE works the same way, but the Register assigns the
DISCOUNT a minus value.
27
Example 5
Base Price
Excise Tax
Road Tax
Local Tax
Sales Tax
VAT
$0.32/litre
$0.04/litre
$0.162/ litre
$0.015/ litre
7% on the above
7.5% on all the above
Under PRICING,
Select a Distinct Price Code, say 8, and change the price to 0.32.
Assign a unique Tax/discount Code, say 8.
Under TAX/DISCOUNT
Select T/D Code 8
Select T/D Line 1, CHNG T/D VALUE to TAX PER VOLUME, and set to 0.04
CHNG T/D ASSIGN to 0. 0 is the Delivery Subtotal
Select T/D Line 2, CHNG T/D VALUE to TAX PER VOLUME, and set to 0.162
CHNG T/D ASSIGN to 0. 0 is the Delivery Subtotal
Select T/D Line 3, CHNG T/D VALUE to TAX PER VOLUME, and set to 0.015
CHNG T/D ASSIGN to 0. 0 is the Delivery Subtotal
Select T/D Line 4, CHNG T/D VALUE to TAX PERCENT, and set to 7.000
CHNG T/D ASSIGN to 0,1,2,3. This captures all the items above
Select T/D Line 5, CHNG T/D VALUE to TAX PERCENT, and set to 7.500
CHNG T/D ASSIGN to 0,1,2,3,4. This captures all the items above
Net Volume Finish
PRICE/LITRE
SUBTOTAL
UNIT TAX T/D 1
AMOUNT
UNIT TAX T/D 2
AMOUNT
UNIT TAX T/D 3
AMOUNT
% TAX T/D 4
TAX/UNIT
AMOUNT SUB+T/D 1&2&3
% TAX T/D 5
TAX/UNIT
AMOUNT SUB+T/D 1&2&3&4
FINAL PRICE PER LITRE
TOTAL NON-TAX LINES
TAX THIS DELIVERY
GRAND TOTAL
(Same as ([{(0.32+.04+.162+.015 ) x 100)}] X
1.07) x 1.075
$
$
$
$
$
$
$
$
%
$
$
%
$
$
$
$
$
$
100.0
0.3200
32.0000
0.0400
4.0000
0.1620
16.2000
0.0150
1.5000
7.0000
0.0376
3.7600
7.5000
0.0431
4.3100
0.6177
0.00
29.77
61.77
Turn to page 15: DEFAULTS, START MODES, PRESETS and RELAY CONTROL.
Page 18: ENABLE PRINTER. Here is where the location on the ticket is set using ADVANCE.
Page 20: VIEW RECORDS and RESTORE RECORDS. VIEW RECORDS IS USED TO FIND AND
PRINT LOST TICKETS. The last 50 are stored in the head.
SHIFT REPORT pages 21 and 22. How to start and end a shift. Entering data. A sample shift
report can be found on page 23, Figure 21. Do a shift report.
Turn to page 23 & 24. Review of the DATE FORMAT and TIME/DATE MODE.
Turn to page 24. Pages 24 through 27 concern the assignments and addresses of the devices that
make up an EMR System. A single meter (single Display Head), single Interconnection Box,
single printer system would not need any inputs (typically) for these assignments and addresses.
Any system with a multiple of these devices will require data input.
Networking/Connecting IBs Together
Multiple IBs with their heads can be networked together and talk to one printer or get
commands from one external device. The IBs are connected via the RS-485 terminals A and
B labeled IB Network in each IB. Each IB is given a unique address as noted under
SYSTEM ADDRESS in the Setup and Operation instructions. The Printer Address, i.e. which
IB and port the printer is connected and assigned is defined in the Printer Address section of
SYSTEM ADDRESS. This is where on the Network each head needs to be directed via the
Printer Address for printing. Before Connecting the IBs together, each individual
system needs to be set-up independently and working properly. For two head
operation, follow the set-up for two heads information in the Getting Started Section.
Power down the systems before making connections or changes.
Turn to page 28. TIME DELAYS and VERSION NUMBER.
Version Number is the official W&M version number, not the version that shows at end of the
display startup.
29
PRACTICE
Leave SETUP MODE and put your EMR in VOLUME MODE.
IF YOU CANNOT GET TO VOLUME MODE, THEN YOU WILL NEED TO CLOSE THE C&C
SWITCH (PUT THE BOLT IN)
SHIFT REPORT
Return to Setup and begin a Shift (Time, Odometer, and Load).
1.
2.
3.
4.
5.
6.
7.
Return to Volume.
Go to Currency.
Select a Price Code with a value.
Make a sample delivery.
Make 2 more Sample Deliveries using Volume and Currency preset.
Go to setup and end the shift with an odometer reading greater than the one you started with.
Print the shift report.
Temperature Calibration
Product Set-up (FUEL SOURCE-PRODUCT DESCRIPTION)
Calibration
Report/Ticket Format.
Security Codes
Currency: Example 1
Congratulations! You completed the Setup and Operation sections of the course.
30
Activity
FOR TWO HEAD SYSTEMS
DISPLAY SYNTAX
TEMPERATURE
If Temperature Compensation,
Temperature Compensation by
Density? Or Coefficient? (C&C
DELIVERY OPTIONS)
C&C TEMPERATURE
Default is Fahrenheit, if Celcius, need
to change
Comment
Temperature Calibration
31
Order
7
8
9
Activity
METER ID (C&C FUEL SOUCE)
INPUT TYPE AND METER DIRECTION (C&C
FUEL SOUCE)
PRODUCT SETUP
(Product Description under C&C FUEL
SOURCE)
10
11
CONFIGURE I/O
Emergency Stop
Remote Start/Stop
Pulse Output and Setting Pulses (before
1-06)
Remote Display and What Master
Printer Interlock
Preset Interlock
Transfer Interlock
12
13
OEM MESSAGE
DELIVERY OPTIONS
14
15
16
SECURITY
17
18
19
RELAY CONTROL
Comment
Text for Meter ID such as Truck Number
Defaults to Quadrature, which is the Pulse Generator
Supplied by V-R. Here is where the meter direction can be
changed if counting backwards. CW is default.
formats.
Select PRODUCT TYPE first. This loads the
proper Temp Comp Table/factor, and assigns the
default name to the product. Can change text
name under Name. Select Temperature
Compensation here, and Which of 1 Calibrations.
32
21
Activity
TANK ID
22
SHIFT Report
23
Order
20
24
25
26
27
28
29
30
31
RESTORE RECORDS
RESTART (C&C)
Comment
If enabled, the operator is required to enter an
identification for each delivery.
FOR DATALINK ENABLE AND CHANGE PROMPT
TO TAIL NUMBER, CUSTOMER NO OR ACCOUNT
NO AND DEFAULT BLANK OR WHAT WANTED
Can view and print up to the last 50 transactions. Very
helpful for lost tickets.
This is where the Shift Report capability can be
accessed. Between the Accept Start and Accept
End times you get a summary of the transactions. You
can also enter Load and Odometer information. Load
information is adjusted by the transactions.
If you want a special Shift Report, then set up in the
same way as Delivery Report/Ticket.
If want to be other than default
Set date and time.
Defaults okay for typical truck, two heads noted above.
For terminal with more than one IB will need to
specifically address.
Not available if Display does not have Currency.
Not available if Display does not have Currency.
This is a read-only system. These numbers are the
official W&M software numbers.
Only used in Special Cases
NORMALLY DO NOT USE WILL REBOOT TO
FACTORY DEFAULTS AND SETUP INFORMATION
WILL BE LOST.
33
34
Language
o Setup Labels: Tail Number Prompt and default value for tail number for
DataLink.
o Printer labels if need to change or create Text and PC not available.
Display Syntax
o Whole gallons for Jet, but calibration in tenths first.
Configure I/O
o Pulse output for Pre -19 Jan 2006 systems.
Delivery Options
o Tail Number Enable Prompt
o Delivery Timeout Disable
Meter Calibration
5.
6.
7.
8.
9.
Power down
Disconnect Head 2 PWR in the IB
Power up the system. It will recognize Head 1.
Check to make sure System Address IB Address 1, Head 1.
a. Push the Mode button so the small arrow is above Setup at the bottom of the screen.
The display should say DEFAULTS.
b. Arrow (+) up to SYSTEM ADDRESS
c. NEXT , IB ADDRESS 1 ( IF NOT 1, NEXT to SET IN ADDRESS, arrow to 1 and
ENTER)
d. Arrow down to HEAD ADDRESS, it should say 1. ( IF NOT 1, NEXT to SET HEAD
ADDRESS, arrow to 1 and ENTER)
e. ENTER to SYSTEM ADDRESS
Power down.
Disconnect Head 1 PWR in the IB.
Connect Head 2 PWR.
Power up the system.
At Head 2, Go to Setup, System Address, Make sure IB 1, then Head Address, and set to
Head 1.
a. Push the Mode button so the small arrow is above Setup at the bottom of the screen.
The display should say DEFAULTS.
b. Arrow (+) up to SYSTEM ADDRESS
c. NEXT , IB ADDRESS 1 ( IF NOT 1, NEXT to SET IN ADDRESS, arrow to 1 and
ENTER)
d. Arrow down to HEAD ADDRESS, it should say 1. ( IF NOT 1, NEXT to SET HEAD
ADDRESS, arrow to 1 and ENTER)
e. ENTER to SYSTEM ADDRESS
12. Push the mode button to Volume and you should get an E50 IB Comm Failure. This is normal
in this case because the system sees Head 2, but no Head 1.
13. Power down and connect head 1 PWR in the IB.
14. Power Up and there should be no IB Comm Failure due to Software/address conflicts.
Time Delays
To have the screen stay lit longer:
1. Push the Mode button so the small arrow is above Setup at the bottom of the screen. The
display should say DEFAULTS.
2. Arrow (+) up to TIME DELAYS
3. NEXT, BACKLIGHT 30,
4. NEXT, Cursor flashing before 30
5. Arrow down to a 9 and ENTER
6. Arrow down to BLANK SCREEN 30,
7. NEXT, Cursor flashing before 30
8. Arrow down to a 9 and ENTER
Time and Date Set Head 1, Only done through the screen - T&D in IB Box
1. If not already in SETUP MODE, Push the Mode button so the small arrow is above Setup at
the bottom of the screen. The display should say DEFAULTS.
2. Arrow up (+) to SEE the Time and Date.
3. To change, use the NEXT key to move the flashing cursor under the item to change and use
the arrow keys to make changes.
4. When done ENTER to accept the changes.
Printer Enable if Printing
1. If not already in SETUP MODE, Push the Mode button so the small arrow is above Setup at
the bottom of the screen. The display should say DEFAULTS.
2. Arrow (+) up to PRINTER OPTIONS
3. NEXT, DISABLE, Arrow down to ENABLE to enable the printer.
4. If a Slip Printer:
a. ENTER, ENTER back to PRINTER OPTIONS
5. If Slip Printer, MAKE SURE DIP SWITCH 3 ON THE BOTTOM OF THE PRINTER IS ON.
6. If roll printer, after ENABLE, NEXT, Slip Printer, Arrow down to ROLL PRINTER,
7. ENTER, ENTER, ENTER back to PRINTER OPTIONS.
8. If Roll printer, DIP SWITCHES UNDER COVER ON BOTTOM OF PRINTER - MAKE SURE
DIP SWITCH 1 IS 3, AND DIP SWITCH 2 IS 1
9. If Roll printer, PORT 1 ASSIGN UNDER SYSTEM ADDRESS NEEDS TO BE CHANGED
TO ROLL PRINTER. (Default is Slip Printer)
C&C MODE
To get into C&C MODE, it is cleanest to start in VOLUME MODE (Gallons on Screen)
1. Drop the Left Rear Bolt on the Register housing to open the C&C switch.
2. Push the Mode button so the small arrow is above Setup at the bottom of the screen. The
display should say LANGUAGE. LANGUAGE is the first item in the C&C menu.
3. If you are in SETUP mode already and drop the bolt to open the C&C switch, YOU
MUST PUSH THE MODE BUTTON AGAIN to get into C&C mode
4. If you are in the SETUP Menu, C&C is just above DEFAULTS, Push next to get to
LANGUAGE.
5. You will know that you are in the C&C mode if you push the MODE button and you cannot
get to Volume (Gallons).
LANGUAGE
Setup Labels: Tail Number Prompt and default value for tail number for DataLink.
1. NEXT, English, NEXT, LABEL CHANGES,
2. NEXT, PRINTER LABELS.
3. Arrow down to SETUP LABELS
4. NEXT, TANK ID
5. NEXT for cursor flashing under the T. Use the Cursor to move in the field and the Keypad
for entering TAIL NO , or TAIL NUMBER, or ENTER TAIL NO. in this field as the prompt
for the operator to enter Tail number in the Transaction.
6.
7.
8.
9.
Note that the keypad works like a cell phone, and automatically indexes over if the next
letter or number is on a different key. If same key, NEXT to move cursor.
The lower left key on the keypad is CLEAR.
ENTER when done with editing this prompt.
Arrow down to TANK 1. This is the Default value for Tail Number. If you want to start
with blanks, then use the lower left clear key on the keypad to clear this default.
Note: If you need an N leading the tail number, you can put an N in the first character
as a default. This may or may not be value added. If the N is there, the operator pushes
NEXT, Again at the prompt to enter the numbers of the tail number.
ENTER to accept the changes
ENTER ,ENTER, ENTER back to LANGUAGE.
NOTE: If you Enter to many times to DEFAULT, then arrow up to C&C and NEXT to
LANGUAGE to get back in the C&C menu.
Printer Labels
If you want or need to change printer labels, these can be edited one character at a time
through the display screen, but the PC Setup Program approach for these labels and setting
up a ticket is significantly easier. Particularly because the editing of the labels and the printer
ticket setup must be done for each register display.
Pulse Output For software versions Pre -19 Jan 2006 systems.
1. In C&C menu Arrow down to CONFIGURE I/O
2. NEXT, Emergency Stop.
3. Arrow down to PULSE OUTPUT
4. NEXT to SET PULSES/VOL and the flashing cursor.
5. NEXT key and keypad or arrow keys to set the pulses/volume.
6. ENTER several times back to CONFIGURE I/O
DELIVERY OPTIONS
Defeat Delivery Timeout - Normally defeated for Aviation.
1.
2.
3.
4.
5.
6.
Label on Ticket
ADVANCE
ABC AVIATION
BEST AIRPORT
1234 JET BLVD
ANYWHERE, ST 12345
800-123-456
ARINC 123.4 UNICOM 134.5
BLANK LINE
TAIL NUMBER ( Edit TANK ID printer label)
START
END
TICKET NUMBER
METER ID
PROD DRSCPT
TOTALIZER START
TOTALIZER END
VOLUME FINISH (Edit GROSS VOLUME FINISH printer label)
END OF REPORT
Set up ticket
1.
2.
3.
4.
5.