:Introduction
Our project is to build ECG machine that take 3 point from the human
body using electrodes and enter this small voltage to amplification
.circuit in order to give voltage signal that can be introduce
So our project depend on amplification circuit ,this circuit contain
powerful amplifier which has high gain in order to be able to take signal
. that can be seen and measured
Our project its so sensitive because we have high voltage gain and
wire and electrodes connect to the human body ,all these component
give us
Noise that will be effect to our signal, so now we need very accurate
.filter that able to remove this noise from our signal without effect it
:Project Component
amplification circuit .1
shift circuit .2
PIC 18F4620 circuit contain ADC circuit .3
C# interface progame .4
:amplification circuit
Amplification circuit is the most important part of this project that
take small voltage from the human body .enter it to this circuit and
.give output as signal with amplitude from -5 volt to + 5 volt
In this circuit we have operational amplifier (AD624) which has a
very high CMRR (90dB) and high gain (1000), with power supply
.+9V and -9V
Instrumentation Amplifier:
We choose Analog Device AD624 instrumentation amplifier to amplify the
ECG voltage from electrodes, which is in the range of several mV.
The AD624 is set up with gain of 1000, and is supplied by +9 V and -9V
battery power. Some important features of the AD624 are listed below
Supply voltage
Gain
CMRR
Gain Bandwidth
Input offset
Her in this circuit we give the how the AD624 operational amplifier
connect and the value of resistance and capacitor value must
connect to it
:Filters circuit
next step after amplification is filter circuit in order to remove un
liked noise that will be affect to our signal .in this circuit we will use
.3 type of filter to remove noise
:high pass filter or (low frequency filter) .1
this filter is called is the low frequency filter, because signals
above the threshold are allowed to pass) is set at either 0.5 Hz or 1
Hz
frequency filter) .2
5
:notch filter .3
This filter is the most important filter .its work is important in ordewr
to remove 50Hz noise this noise come from the air and any
electronic machine around our circuit.this 50Hz can be come from
6
The circuit for notch filter that will be used and the value of
capacitor and resistor that used as shown in the circuit below
When we reach this point we must have pure ECG signal that can be
seen on the oscilloscope without any problem after we connect the
.circuit to two electrode on the left and right hand
:Step 2
7
All above procedure is the first step in our project .output the ECG
signal is the first step, the second step is to show the ECG signal on
.PC
:Shift circuit
As we said above ADC cant deal with negative value so we need shift
circuit to shift signal from-5volt to +5vlot to be signal from 0 __5
volt .in this circuit we will use 2 amplifier type (LM741 instrumental
amplifier ) with voltage offset -9 volt to +9 volt to be accurate . This
:shift circuit we use as shown below
this circuit give us the approperiat signal which have the signal with voltage
between 0 to 5 volt this signal can be enter to ADC and ADC will digitize it to
digital value to draw them on PC.
Interface On PC:
9
We will take the digital value from the serial port and draw this value
on the C# interface in our PC
This form use to draw the ECG signal from the input machine
10