Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Model 577PC
Reflectometer
Users Manual
WARNING
This manual contains information and warnings, which must be followed to ensure proper and
safe operation of your Model 577PC Reflectometer. For accurate results and safe use of the
Model 577PC Reflectometer, correct operation and maintenance including periodic inspection
are required. You should read the instruction manual paying attention to all warnings and notes
prior to practical use and maintenance of your instrument.
Cautions regarding operation of the Model 577PC Reflectometer are indicated by the following
format throughout the instruction manual. Be sure to read and understand all of these cautions.
WARNING
NOTE
POINT
This document contains information for use by owners and operators of the Photovolt Instruments
Model 577PC Reflectometer. Information in this document is subject to change without notice. This
document is considered proprietary and no part may be reproduced or transmitted in any form or by any
means, electronic or mechanical, for any purpose, without the express prior written consent of Photovolt
Instruments. If copies are then permitted, they must contain this proprietary notice.
2012 Photovolt Instruments. All rights reserved.
Model 577PC
Reflectometer
Users Manual
(Revision A)
3. Instrument Set-up
10
11
7. Calibration Instructions
12
15
15
17
17
17
18
21
22
LIST OF FIGURES
Figure 1. Model 577PC Reflectometer system with YA Search Unit ........................................4
Figure 2. 577PC console shipping carton contents......................................................................5
Figure 3. Search units and related items .....................................................................................5
Figure 4. Power Adapter ..............................................................................................................6
Figure 5. Model 577PC Front Panel.............................................................................................7
Figure 6. Model 577PC Back Panel ...........................................................................................10
Figure 7. Example of a Two-Point Calibration............................................................................12
Figure 8. ZERO Calibration........................................................................................................13
Figure 9. Plaque.........................................................................................................................14
Figure 10. Format in Excel for Saved Data ................................................................................15
Figure 11. Types of Search Units...............................................................................................18
Figure 12. Search Unit Apertures...............................................................................................18
Figure 13. Tristimulus Filter Characteristics...............................................................................20
Figure 14. Applications by Search Unit and Color .....................................................................21
Figure 15. Lamp Assembly .........................................................................................................24
Figure 16. Lamp Installation........................................................................................................24
Figure 17. 577PC-S Stainless Steel Model................................................................................. D
APPENDICES
The new Photovolt 577PC reflectometer is similar to older models but, contains
state-of-the-art electronics and has many new features including an embedded PC
computer, Ports for a keyboard, a mouse, two USB devices, Ethernet, RS232 serial
port, and a color display.
Figure 1 shows a typical 577PC Reflectometer system with all its components.
The type YA sensor head (search unit) shown is the most common type. Other search
unit types may be supplied with black cavity standards and calibration plaques that
differ in appearance from the ones shown. Please refer to section 13 of this manual for
more information on the different search unit types available.
(England)
(Europe Germany)
Warranty Card
9957711P
0257720
4457004
0130168
The search unit, calibration plaque, black cavity, spare lamp, calibration certificate,
and color filter(s) are shipped in the smaller carton. There are a lot of possible options
so check the contents against the list of possibilities shown below.
Type YA search unit
0457711
Type WA search unit
0457720
Type TA search unit (with filters)
0457731
Type DA search unit
0457704
Type XA search unit
2457002
Color filters (for X,Y and D only) AMBER
BLUE
2457003
GREEN
2547001
Black cavity standard (not for D)
2657012
Black calibration plaque (for D only)
Brightness calibration Plaque (for W only) 2657030-2657049
Reflectance calibration plaque (X,Y,T)
2657050-2657069
Calibration plaque (for type D only)
Spare lamp for types YA, WA, TA, XA
1157707
Spare lamp for type DA only
Figure 3. Search units and related items
Connect the adapter output connector to the Power socket on the backside of
the reflectometer. Connect the power cord to the adapter and plug into your power
outlet. There is a small green indicator light on the top of the adapter that is illuminated
whenever it is connected to a functioning power outlet.
NOTE
Do not attempt to power the 577PC from any source
of power other than the Photovolt provided module.
Doing so will void the warranty and may damage the
instrument.
Difference Mode
While pressing the HOLD button pushing the CALIBRATE button puts the 577PC in the
difference mode.
When the difference mode is active all three color select lights are on. The reflectance
display shows the difference in percent between the current reading and the reading at the time
this mode was started.
The difference mode can be used to directly display variations across a large sample as
the search unit is moved about. It can also be used to show variations across a group of
samples or differences from a master sample. It can also directly display the darkening of filter
paper in the ASTM fuel stability test.
To exit the difference mode just repeat the HOLD plus CALIBRATE button procedure.
Unlike other modes the 577PC does not stay in the difference mode when power is turned off
and back on again.
Turn on the master power rocker switch. When it is on it lights up in a green color. This is a
master power switch like those on the power supply on the back of your PC. It should be left on
all the time. The red power button on the front of the 577PC is used to control operation of the
instrument.
4.2.2.
While pressing the HOLD button pushing the ZERO button tells the 577PC to
transfer the internally stored reflectance readings to a USB flash memory stick. See
section 8 of this manual for a complete description of saving and transferring reflectance
readings.
4.3. Reflectance Display
The digital display shows the sample reflectance in percent from 0.00 to 99.9%.
This display is continuously updated unless the hold button is used. When the reading is
less than 10% the hundredths digit is shown. During abnormal conditions error
messages may be shown such as: Lo Hi or Err. The display is also used during
instrument calibration.
4.4. Color Select Buttons
The color select buttons select the reflectometer operation for each color:
AMBER, BLUE , or GREEN. Each color is calibrated individually and those settings are
selected with these buttons. This allows quick readings of a sample in each color
especially when using the type T search unit that has a color filter selection wheel. The
appropriate color must be used to match the color of light emitted by the search unit.
8
10
11
7. Calibration Instructions
The Model 577PC uses internal flash memory to store the current mode and the
calibration values for each of the tristimulus filters. The values for one calibration
plaque are retained when the instrument is powered down. Thus it is not necessary to
reenter the plaque calibration values before each recalibration if you use the same
calibration plaque. The 577PC should be allowed to warm up for at least 30 minutes
before calibration or use. To avoid the warm up period you may leave the instrument on
continuously. If left on continuously the 577 draws less than 25 watts of power and
lamp life will exceed 1.5 years. A detailed overview of the Model 577PCs calibration
procedure follows:
7.1. Filter Selection/Reflectance Display
When measuring samples the Amber, Blue, and Green buttons are used to
select appropriate calibration curve for the specific tristimulus filter in the search unit.
When one of these three buttons is depressed, the corresponding LED above it will
illuminate and indicate which filter calibration curve is being used.
12
POINT
The calibration curve of the 577PC will be different for each of the three filter
channels. Be sure to select the correct filter channel for the filter in the search
unit. Incorrect filter channel settings will result in erroneous reflectance
measurements.
When filters are changed in the search unit, the corresponding channel should be
selected via the Amber, Blue, or Green buttons. Filters and their corresponding
memory settings can be changed at will, once the Model 577PC has been calibrated.
Only those filter channels that are being used for samples need to be calibrated.
The normal display mode displays percent reflectance of the sample.. This is the
default display mode and is active when NOT in calibration mode (whenever the
CALIBRATE, or STD lights are extinguished).
A samples reflectance value is calculated based on the calibration curve saved in
memory for the currently selected filter channel (Amber, Blue, or Green). If the
calculation results in a reflectance value well beyond 100% the display will indicate an
overrange condition by displaying (Hi. ). (Reference Section 6.6)
7.2. Instrument Calibration
As stated earlier, a filter channel (Amber, Blue, or Green) should not be used for
reflectance measurements until it has been calibrated at least once. Calibration
requires setting both the zero offset (ZERO) and a standard plaque value (STD). These
values are set by using the CHANGE button as described in the following sections.
7.2.1.
To calibrate the Model 577PC, the desired filter channel must be selected via the
Amber, Blue, or Green buttons. Next, the CHANGE button is depressed causing the
CHANGE LED to illuminate. The instrument is now ready for a two-point calibration,
starting with the zero offset setting and followed by the standard plaque value setting.
7.2.2.
Figure 9. Plaque
light will Illuminate, then press the STD button.
The
STD light will illuminate as will all three filter color lights (Amber, Blue, and Green).
The display will show the assigned reflectance value for the previously entered
standard plaque.
Read the assigned plaque reflectance value for the selected color from the label on
the back of the plaque.
Enter the standard plaque value from the plaque label into the Model 577PC by using
the Amber, Blue and Green buttons to increment the tens, units, and tenths digits
respectively. The digits will individually increment between 0 and 9 when the buttons
are depressed. Standard plaque values of between 00.0 and 99.9 may be entered.
Once the display reads the correct standard plaque value, depress the STD button a
second time. The STD LED will extinguish, as will the CALIBRATE LED. Normal
reflectance display will resume and the display should be identical as the standard
plaque value ( 0.2%) after the standard plaque value is entered.
The reflectometer is now ready for sample measurements on the newly calibrated
color channel. For best results let the instrument warm up for at least 30 minutes before
calibration. During use calibration should be rechecked periodically by replacing the
standard plaque on the search unit. If the plaque value ( 0.5%) is displayed when the
standard plaque is replaced on the search unit, then the instrument is still in calibration.
Otherwise, calibration should be repeated.
POINT
Two-point calibration (zero offset and standard plaque entry) must be
done at least once for each filter and filter channel combination to be used.
If the search unit to be used does not require multiple filters, calibration
need be done only on the filter channel required for use with the selected
search unit. The 577PC remembers the calibration information but it is still
good practice to check it fairly often.
14
16
10.
Computer Settings
The two situations where the connection of a display and keyboard are needed
are in setting the internal clock time and date and updating the 577PC software.
10.1. Setting Time and Date
When shipped from Photovolt the 577PC internal clock is set to universal (GMT)
time and date. To change this you will need to connect a keyboard (or USB keyboard)
and a display to the 577PC. You must start this procedure with the 577PC power turned
off. Push the red power button to turn on the 577 and immediately start repeatedly
pushing the DELete key on the keyboard. This will cause the BIOS setup page to
appear on the display. Use the arrow keys and PAGE UP and PAGE DOWN keys to set
the desired time and date and then save and exit by pushing the F10 key.
Once the BIOS setup is completed the computer will restart the reflectance
program. Unless you are sure you know what you are changing do not modify the other
BIOS settings as they may keep the reflectance program from operating properly.
10.2. Updating the 577PC Software
The 577PC software can be easily updated by the user. Normally the updated
program is supplied on a USB memory stick or sent by email.
When the update is sent by email you will receive a file with the name:
577_580_update.tar.gz. Copy this file to the root directory of a USB memory stick.
The file is about 125Kb. The USB stick that is used does not have to be empty and may
contains other files on them.
Connect a PC or USB keyboard, and the memory stick to the 577PC. Connecting
a monitor to the DISPLAY connector is desirable but not necessary. Wait a couple of
minutes for the new items to be recognized by the 577PC. Then push only the U key on
the keyboard, just the U key, nothing else. The reflectance display will go blank and the
new software will be installed. This only takes a few seconds. Wait a minute for it to
17
11.
Transmittance
Range
Transmittance
Peak (nm)
Amber
Green
Blue
540 660 nm
450 620 nm
410 490 nm
590
550
440
20
Industry
Biomedical
Burlap
Ceramics (Tile &
Porcelain)
Chemical
Cosmetics
Detergents
Distillate Fuels
Food &
Agricultural
Industrial
Products
Metal Finishing
Packaging
Paint & Varnish
Pharmaceutical
Plastics
Pulp & Paper
Textile
Type
Search
Unit
Application
Erythema, skin color,
frog skin, etc.
Grading
Color classification.
Opacity, or contrast ratio
Y/T
Varies
Y/T
Tristimulus
By user
Y/D
Green-Tristimulus
By user
Y/T
Green- Tristimulus
ASTM 6468
Tristimulus
Coffee Brewing
Institute, SCAA,
etc.
By user
Y/T
Y/T
Y/T/D
Green-Tristimulus
and others
Green-Tristimulus
Varies
Varies
Test Method or
Standard
Numerous by
users
By user
ASTM E97
ASTM C347
ASTM C346
ASTM C584
By user
Y/T
Filter
Y/T
Tristimulus
Y/T
Y/T
Y/T
Y/T
Y/T
Tristimulus
Tristimulus
Tristimulus
Green Tristimulus
By user
ASTM E97
ASTM D1455
ASTM D3052
By user
By user
Federal Test
Method No. 141;
ASTM D523
All types
Varies
Color
Y/T
Varies
By user
ASTM D2457
Brightness
Opacity
Color
Rapid and convenient
color matching of textiles
W
Y/T
Y/T
Built-in
Green-Tristimulus
Tristimulus
TAPPI Methods
ASTM
Tristimulus
By user
Y/T/D
21
The Photovolt model 577PC reflectometer system does not require any regular
preventative maintenance. Normally all that is needed is to keep the instrument
relatively clean and free of dust. In especially dusty environments it will probably be
necessary to use low pressure air to blow out dust that collects down inside the search
unit aperture and on the filters.
12.1. Care and Maintenance of Calibration Standards
The model 577PC reflectance measurements can only be as accurate as the
calibration standard plaque used for calibration. Normally the human eye can detect
only about a 5% change in reflectance and it is easily fooled by changes in color tint.
The 577PC reflectance resolution is 0.1%. Thats fifty times more sensitive than the eye.
Obviously any dirt or contamination you can see on the calibration plaque is going to
have a big detrimental affect on the accuracy of the readings.
12.1.1.
Environment:
Avoid exposing the calibration plaque to sources of heat, chemicals, dust, or oil.
Contamination of the surface of the plaque will result in decreased accuracy. Avoid
damaging the surface of the plaque. Scratches, chipping of the surface, or dulling of the
surface will result in incorrect readings and make cleaning difficult.
12.1.2.
Handling:
Avoid touching the reading surface of the calibration plaque. Fingerprints, dirt, oils,
or any other surface contamination will result in incorrect readings. Handle the plaque
only by the anodized plaque holder. When placing the plaque on a table or other flat
surface, keep the reading side facing up, or place it face down on a clean, soft, dirt-free
surface such as a lint-free cloth.
12.1.3.
Operating Conditions:
It is suggested that the 577PC, search unit and calibration plaque be allowed to
equilibrate to room temperature with the power turned on for a period of one hour before
conducting critical calibrations.
12.1.4.
Remove dust by gently brushing the surface of the plaque with a soft camel-hair
brush or blow dust off of the surface with clean, dry, oil free, filtered compressed air. A
pressurized canister of air for cleaning photographic lenses is suggested for this
purpose.
12.1.5.
Severe Cleaning:
Recalibration Period:
Only the inside surface of the black cavity is used so as long as it is clean and
undamaged it should work properly. The black cavity does not need to be calibrated.
23
LAMP
BRACKET
LOCKWASHER
NUT
Figure 15. Lamp Assembly
INSTALL LAMP INTO BRACKET AS SHOWN. TIGHTEN NUT FINGERTIGHT ONLY.
LAMP WITH BRACKET. MOUNT AS LOW
AS POSSIBLE, LEAVE SCREWS LOOSE.
MOUNTING SCREWS 4 PLACES
WIRE
CONTACT
BLOCK
Move the lamp bracket up/down or tilt side to side until the sample aperture on the end of the search
unit is evenly illuminated. The easiest way to check the illumination is to hold a piece of paper against
the aperture and check for a evenly lit circle.
24
Some states do not allow disclaimer of express or implied warranties in certain transactions;
therefore, this statement may not apply to you.
This manual could include technical inaccuracies or typographical errors. Changes may be
periodically made to the information herein; these changes will be incorporated into new editions
of the manual. Photovolt Instruments may make modifications and/or improvements to the
products described in this manual at any time.
It is possible that this manual contains information about or makes reference to products or
services that are not released in your country. Such information or references must not be
construed to mean that Photovolt Instruments intends to announce such products or services in
your country. Contact your local Photovolt Instruments representative for further information.
The foregoing warranties are in lieu of all other warranties, expressed or implied, including but
not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Issued: Feb 2012
Revision A
26
P/N 9957710P
25
7. Problem: Power light is on but the display is blank and LED lights are OFF
a. Internal computer is still booting up. Wait for 30 seconds.
If additional assistance is required call Photovolt Technical Service at 1-800-222-5711.
3. Standard Plaque: A small calibration plate with a surface of known reflectance.
4. Photodetector: A device that produces an electrical current proportional to the light that
shines on its surface.
5. Reflectance Photocell: The Photodetector that measures the light reflected from the sample
surface.
6. Gain: The amount of amplification applied to the photosensor output signal before it is
converted to digital information for the computer.
7. Filter Channel: One of the three electronic channels (Amber, Blue, Green) of the Model
577PC. Each channel has its own linearity settings and memory locations, which allow it to
function independently of the other two.
8. Calibration Memory: Each filter channel has three independent memory locations associated
with it. These locations are:
Zero Offset Memory: The memory locations that store the current produced by the Model
577PC when a Black cavity standard is placed on the attached search unit.
Calibration Reflectance Memory: The memory locations that store the current produced
by the Model 577PC when a standard plaque is placed on the attached search unit.
Standard Plaque Value Memory: The memory locations that store the standard plaque
reflectance values entered during calibration.
9. 4-digit LED Display: The group of four display digits that display the percent reflectance.
10. LED: Light Emitting Diode - The small, red, rectangular indicators above the HOLD, Amber,
Blue, Green, STD, and CALIBRATE buttons on the front panel.
Yellowness Index
YI = 100 (A - B) / G
YI = 128X - 106Z / Y
YI = 100 (1 - 0.847 Z / Y)
ASTM E 313
ASTM D 1925
ASTM E 313
Appendix D - 577PC-S
The Photovolt Model 577A reflectometer is available with the SS stainless steel
case option. This model has a shallow case with all controls and common connections
on the front side. This allows mounting on a wall or shelf above and away from the work
area. The case and search unit are made from stainless steel and sealed to a NEMA-4
splash proof rating. An extra long search unit coiled cable is included so samples up to
ten feet away can be easily measured.
The operation of the 577PC-S is identical to the standard instrument and all
previous sections of this manual apply to this model equally.
This document contains information for use by owners and operators of the Photovolt Instruments
Model 577PC Reflectometer. Information in this document is subject to change without notice. This
document is considered proprietary and no part may be reproduced or transmitted in any form or by any
means, electronic or mechanical, for any purpose, without the express prior written consent of Photovolt
Instruments. If copies are then permitted, they must contain this proprietary notice.
2012 Photovolt Instruments. All rights reserved.
Photovolt Instruments
P/N: 9957710P Rev A