Caution: The presence of minute traces of organic matter on the equipment will
cause large errors in the test results. So clean all equipment thoroughly before
using.Erlenmeyer flask
-Small beaker
-Titration apparatus:
25 or 50 mL burette, graduated in 0.1 mL
burette support
100 mL graduated cylinder
rubber-tipped stirring rod, or magnetic stirrer and stir bar
white porcelain evaporating dish, 4.5 inches in diameter
-Reflux apparatus:
500 or 250 mL erlenmeyer flasks with ground-glass 24/40 neck
300-mm jacket Liebig, West, or equivalent condenser with 24/40 ground-glass
joint
hot plate with sufficient power to produce at least 1.4 W/cm2 of heating
surface
-Blender
-Pipets
-Glass beads
-Fume hood
Reagents
-Standard potassium dichromate solution, 0.250N or 0.025N
-Sulfuric acid reagent containing silver sulfate catalyst
-Standard ferrous ammonium sulfate titrant
-Ferroin indicator solution
-Mercuric sulfate crystals
-Sulfamic acid
-Concentrated sulfuric acid
-Distilled water
Laboratory Procedure
1. Standardize the ferrous ammonium sulfate.
a. Place 10.0 mL of 0.250 N potassium dichromate solution in a 500 mL erlenmeyer
flask.
b. Dilute the potassium dichromate solution to 100 mL with distilled water.
c. Add 30 mL of concentrated sulfuric acid. Add the acid slowly while mixing the
solution.
d. Cool the solution to room temperature.
e. Add 2 or 3 drops of ferroin indicator.
f. Titrate with ferrous ammonium sulfate until the solution changes from a bluegreen color to a reddish-brown color. Record the amount of titrant used in the Data
section.
g. Calculate the normality of the ferrous ammonium sulfate using the following
formula. Record the result in the Data section:
3. Prepare a blank.\
a. Pipet a volume of distilled water equal to that of the sample into a 500 mL
refluxing flask.
b. Add 1 gram of mercuric sulfate to the distilled water.
Caution: This step in the procedure can be extremely hazardous since a violent
reaction may occur. Perform this step and the reflux step in a laboratory fume
hood. Be sure to mix the solution thoroughly before applying heat in the next step
or local heating of the flask bottom may cause a blowout of the flask contents.
c. Cover the end of the condenser with a small beaker to prevent foreign material
from entering the reflux mixture.
d. Turn on the hot plate and reflux (boil) the mixture for two hours.
e. Cool the flask and condenser.
f. Wash down the inside of the condenser with distilled water and then remove the
condenser.
Alternative Procedure:
-
When your sample contains less than 50 mg/L COD, then you must change a
few parts of the usual procedure. First, in steps 1.a. and 2.f., dilute the
dichromate solution to a normality of 0.0250 N. (You will also have to change
the normality in the calculation in step 1.g.) Second, the ferrous ammonium
sulfate solution used as a titrant in steps 1.f. and 5.c. will also have to be
diluted to 0.025 N.