Chroma
iCheck Chroma is a test kit to measure
Vitamin A in oil, empowering you with instant
results to make confident decisions.
User Manual
Quality
Guarantee
Dear customer,
Congratulations on your acquisition of iCheck Chroma!
iCheck Chroma will be your reliable partner for the vitamin A analysis.
iCheck is a high-tech portable photometer with precise and reliable results.
iCheck is produced following strict rules of quality assurance according to ISO
9001:2008. This is accomplished by the use of high-grade components and
equipment as well as a stream-lined production process. This process includes
quality controls of each component and rigorous calibration of the device by
trained technicians.
Your iCheck Chroma comes with a 2-year warranty.
If you have any questions, please contact us by calling +49 (0)33 28 35 15 000
or sending an e-mail to support@bioanalyt.com.
www.bioanalyt.com
www.facebook.com/bioanalyt
www.linkedin.com/company/bioanalyt
Contents
Case Content
Device Description
Instructions
Menu Functions
6 11
12 13
Software Installation
14
Data Transfer
15
Technical Data
16 17
18
Measurement
19 21
General
22
USB Stick
23
Check your
Case Content
Grafik Case Content
H
USB
iCheck
Chroma
iCheck Chroma is a test kit to measure
vitamin A in oil, empowering you with instant
results to make confident decisions.
iCheck
Chroma
iCheck Chroma is a test kit to measure
vitamin A in oil, empowering you with instant
results to make confident decisions.
USB
User Manual
s now.
ood start.
User Manual
ou need
urement
ded.
evice Description
Device
Description
G
F
H
I
K
Measurement key
Escape key
Not shown:
Battery compartment on the bottom side.
Instructions
Chroma
1Insert the batteries
note:
The batteries can be recharged with the supplied
charger. It takes about 2-3 hours to fully charge an
empty battery. Charging is best performed within
the temperature range of +5 C to +45 C.
Sample 1
The iCheck has an energy-saving function. 10 minutes after the last keystroke the photometer will
switch o automatically.
Important: iCheck must be used with charged batteries at all times even when connected to the computer
via USB cable.
6
Remove the metal cap and insert the vial into the
measurement chamber of the iCheck.
Press the measurement key again. Wait until the
iCheck displays Prepare for measurement.
CHROMA
Insert Blank
Instructions
Chroma
5Prepare your sample
approx. 0.7 mL
0.5 mL
exactly
Now that your sample is ready, press the measurement key on the iCheck Chroma and follow the
instruction as indicated on the display.
The display will indicate Mix your sample at zero
and start a countdown: 3 > 2 > 1 > 0.
CHROMA
2x
Note:
The content of the vial will instantly turn blue, if the
sample contains vitamin A.
Instructions
Chroma
9Result display
iCheck Chroma will now analyze your sample. This
will take approx. 30 seconds.
Do not move the vial or the iCheck until the result
is displayed.
Sample result
The result is displayed in mg RE/kg. Multiply your
result by 3.33 to convert from mg RE/kg to IU/g.
CHROMA
Batch 1, Sample 1
(Range: 3 - 30 mg RE/kg)
8.55 mg RE/kg
10
10Data storage
11Disposal
11
Menu
Functions
Delete data
By pressing the enter key you enter the menu of
iCheck Chroma. Using the arrow keys you can scroll
through the dierent options of the menu and with
the enter key you can choose a function.
iCheck CHROMA
View Data
Delete Data
Transfer Data
Configuration
Transfer data
View data
Note:
Data menu function is only displayed after a minimum
of 1 measurement. Data Transfer menu function is
only displayed after a minimum of 2 measurements.
12
Configuration
To configure your iCheck you can select the following options:
Configuration
Display
Date & Time
OPC Test
Display
Set the brightness and contrast of the display.
Date & Time
Set the correct time and date of your time zone.
OPC Test
Use this function to check the calibration of your
device. To do this cover the measurement chamber with the metal cap and select OPC Test with
the enter key.
Control that the value displayed at (i.e. 45) is with
in the range indicated in the back of the device
(i.e. OPC: 40 - 70). OPC stands for Optical Performance Control.
When the value is within the range proceed with
your measurement.
When the value is outside the indicated range,
repeat the measurement. If the value remains
outside contact BioAnalyt Support at
support@bioanalyt.com for assistance.
Note:
The Calibration Data of your iCheck can be provided
on request. For this contact BioAnalyt Support at
support@bioanalyt.com and provide the serial number indicated on the back of your device.
13
Software
Installation
Software installation
Software update
BioAnalyt Lab software can be updated by clicking on the Update window. For the program to
detect whether there are new updates available
from BioAnalyt computer must be connected to
the internet.
If your current version (e.g. 1.1.0) is dierent from
the newest version click on the Update key to
proceed with the software update.
Note:
BioAnalyt Lab software only works with Windows
operating system (XP and later versions).
14
Data
Transfer
Start BioAnalyt Lab program by double-clicking
the link on your desktop or by going to the Start
Menu >> Programs >> BioAnalyt GmbH >> BioAnalyt Lab.
Plug in your iCheck to your computer via USB
cable. A conguration window will appear after
you connect your device to the computer. Here
you can enter the information about your device.
The serial number of the device can be found on
the back side of the iCheck.
Now the information about iCheck is saved on
your computer and will be displayed the next time
you connect your iCheck to your computer. This
way, information about multiple iCheck devices
can be stored on your computer.
15
Technical
Data
iCheck Chroma is a state-of-the-art portable photometer for measuring vitamin A concentration in edible
oil. The iCheck is specifically developed and equipped
with innovative technology to enable technically
sophisticated measurement of absorption. iCheck
Chroma displays vitamin A concentration in mg RE/kg.
16
iCheck Chroma is equipped with the latest light emitting diode (LED) technology. Using LEDs as the light
source eliminates all known problems with halogen
lamps used in other readers. Modern LEDs emit
particularly stable light and have very low energy
consumption. Furthermore, each iCheck Chroma is
calibrated at the production site in Germany with
vitamin A standards according to quality management system. This eliminates the need for the user
to calibrate the device. Based on the half life of the
LED, approximately 10 million measurements can be
performed reliably on iCheck. iCheck Chroma reagent
vial is a 2-in-1 disposable reaction vial and cuvette.
iCheck Chroma reagent vials contain a certified
volume of reagents to determine vitamin A in the
sample. One iCheck Chroma test kit consists of 100
reagent vials. No additional chemicals or equipment
are required.
Quality assurance
iCheck and iCheck Chroma reagent vials are produced according to quality management system (DIN
EN ISO 9001:2008) certied by TV Nord in Germany.
We guarantee that each vial contains the exact mixture of reagents safely sealed and crimped to ensure
a shelf-life of 12 months (at 20 - 30 C), and that your
iCheck delivers reliable results for at least 2 years.
Technical data
Analyte:
Retinyl palmitate
Sample type:
Sample preparation:
Not required
Analysis method:
Photometric
Linear range:
3 30 mg retinol
equivalents (RE)/ kg
2 min
20 30C
Variation:
10%
Method comparison:
Staff qualification:
1 day training
Weight:
0.45 kg
Dimensions:
11 x 4 x 20 cm (W x H x L)
17
Frequently Asked
Questions
Power Supply
iCheck does not turn on.
You may use other AA/2100mAh/1.2V batteries. However you cannot recharge those in the supplied charger.
18
Measurement
Do I need to calibrate iCheck Chroma?
1. iCheck Chroma variation for repeated measurements of the same sample is below 10%. Therefore
10% deviation from expected value is acceptable.
2. The oil sample was not well mixed before taking up
with the syringe. Mix the oil sample well before taking up into the syringe.
3. No device warm-up was carried out.
4. Incorrect execution of the test:
It is very important to mix the vial at exactly 0.
Insert the vial immediately after mixing. Maximum
waiting time allowed between mixing and inserting
into iCheck Chroma is 2 seconds.
19
Frequently Asked
Questions
Measurement
5. Incorrect volume:
Make sure that exactly 500 L (0.5 mL) of sample is
injected into iCheck Chroma reagent vials.
6. Incorrect injection:
There has to be a distinct interface between oil sample and reagents in the vial before mixing.
7. The operator was not well trained. Contact
BioAnalyt to obtain training and iCheck certification.
What might interfere with the measurement
procedure?
20
1. It is recommended to measure in the range of 20 30 C (68 - 86 Fahrenheit). Do not use iCheck at temperatures above 40 C.
2. It is further recommended to store the iCheck and
the iCheck reagent vials at least two hours before
starting the measurement in the room in which the
measurement will be performed. This procedure
ensures that both, the vials and the device have the
same temperature.
3. Humidity does not aect iCheck measurements.
21
Frequently Asked
Questions
General
Which form of vitamin A can be measured?
iCheck Chroma measures retinyl palmitate, a common type of vitamin A added to refined edible oil.
22
USB Stick
Find the BioAnalyt Lab Software and further product
information on the USB stick.
23