Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Critical Choices
in HPLC
Video Introduction
Laura Bush
Column Selection
Tony Edge and
Dawn Watson
Gradient Methods
Dwight R. Stoll and
Scott Fletcher
Detectors
Scott Fletcher
CONTENTS
TOC
Welcome
The Fundamentals of
HPLC Detectors
INTRODUCTION
Welcome
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
Tony Edge and Dawn Watson
SPONSORED
Comparison of ReversedPhase Selectivity of SolidCore HPLC Columns
Click to
view PDF
SPONSORED
Optimizing Chromatographic
Results with Mobile-Phase
Preheating
Click to
view PDF
[solute]octanol
n-ionized
[solute]uwater
[1]
The log P value determines how soluble the compound is; larger positive
numbers indicate that the compound is more hydrophobic and less water soluble,
and negative numbers indicate that the compound is quite polar. In the case of
ionizable analytes the distribution coefficient (log D) provides a better estimate
of the analyte solubility as it takes into account all forms of the analyte molecule
(i.e., ionized and unionized, Equation 2). Log D is pH-dependent; hence, when it
is measured, the pH at which the measurement was carried out must be specified.
[solute]octanol
neutral
[solute]ionized
water
+ [solute]water
[2]
In general, the more carbon atoms present in a molecule, the greater the value
of log P, and in turn, the greater the retention under reversed-phase separation
conditions. The shape of the molecule can also affect analyte solubility, with
straight-chain molecules, in general, having larger log P values; hence greater
retention is seen for branched chain molecules. Furthermore, the greater the
saturation of the carbon-carbon bonds, the greater the log P value and hence,
a greater retention will be observed. In general, aliphatic compounds exhibit
greater retention than compounds with induced dipoles, which have greater
retention than compounds containing permanent dipoles, which have greater
retention than weak bases, weak acids, and strong acids. It should be noted at
this point that most molecules have many different functionalities, which can
make the exact interpretation quite tricky.
For a separation to occur the high performance liquid chromatography (HPLC)
column must be able to differentiate between similar molecules. As has already
been stated, this can be difficult to judge, because there may only be small
differences between molecules perhaps a difference of one carbon unit, or
perhaps two or three differences that could cancel each other out in terms of
the overall retention. It is necessary, therefore, to consider the analytes that will
be analyzed and how to maximize the differences in interactions between the
analytes and the stationary phase. The most predominant modes of interactions
when using a reversed-phased column are hydrophobic, dipoledipole, and
interactions.
There are other parameters to consider other than the chemistry between
the stationary phase and the analyte. For a separation to occur effectively,
the column has to have sufficient available surface area to load the sample. In
addition, the pH, temperature, and pressure can and do have an effect on the
selectivity of the separation mechanism and also on the robustness of the assay.
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
Column Specifications
Column manufacturers will, generally, provide information regarding the following
aspects of an HPLC column:
The nature of the solid support: This is the material to which the bonded
phase is attached, most commonly silica. Silica particles can be fully porous,
superficially porous, or nonporous. The silica particle type will have an effect on
the chromatography and can affect the efficiency of the separation mechanism.
Bonded phase: This is the chemistry of the moiety that is bonded to the silica
surface. Bonded phases are typically based on an alkyl or phenyl group, and it is
the interaction between the bonded phase and the analytes that primarily drives
the separation mechanism.
Particle size: Particle size is measured as the average diameter of the column
packing particles. Manufacturers will also report the distribution of the size
of the particles used to pack the column. In general, smaller particles and
tighter particle-size distributions will give sharper and, hence, more efficient
chromatography.
Particle shape (irregular and spherical): Irregularly shaped particles can be less
expensive, but they provide separations with poor efficiency because of the way
they pack into a column. It is much easier to pack a column with regularly shaped
particles than it is with irregularly shaped particles. Irregularly shaped particles
are also prone to shearing, which creates fines that can block columns, causing
both chromatographic and instrument-based problems, such as poor peak
shapes and increased back pressure.
Pore size: The majority of the stationary phase exists within the silica pore
structure; therefore, the analytes have to access the pores to interact with the
bulk of the bonded stationary phase. This means that the pore size needs to
be appropriate, because a big molecule will not fit into small pore. For small
molecules, the pore size should be about 150 or less. Larger molecules (>2000
Da) need bigger pores, of 300 . The larger the pores, the smaller the surface
area, which means that the analytes will have less bonded phase with which to
interact.
Surface area: Columns with high surface area may exhibit greater retention,
loading capacity, and resolution. However, low-surface-area columns have their
advantages. They equilibrate between runs more easily, which can be particularly
useful in gradient HPLC. Also, the reduced porosity results in better kinetics,
meaning that there is less dispersion in the column.
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
Si O Si
O
Si O H
C8 bonded phase
O
Si O Si
O
Si O H
C8 bonded phase
O
Si O Si
TMS group
O
Si O H
O
Si O Si
HO
Figure 1: Diagram showing various bonded phase groups, including the trimethylsilyl (TMS) group resulting from endcapping with trimethylchlorosilane.
Secondary Interactions
Silica is often referred to as type A or type B silica or type 1 and 2 silica. The
difference between the two types relates to the manufacturing process and
the resulting purity of the silica produced. Type 1 silica is manufactured by
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
polymerizing a metal silicate molecule, which results in high metal content in the
final silica that is produced. The metal atoms will tend to migrate to the surface,
where they are energetically favored. At the surface they affect the acidity
and, hence, the reactivity of the silica, increasing the strength of the secondary
interactions, which is very noticeable with basic compounds. Type 2 silica is
produced using an organosilicate monomer and therefore has less metal content;
this type of silica is less acidic and less reactive toward basic compounds. It is not
possible to say that one of these types of silica is better than another unless the
analytes are also discussed in the same context.
CH3
H3C Si CH3
O
O
Si
H3C Si CH3
OH
Si
O
O
Si
As well as type 1 and type 2 silicas there are also different forms of silanol groups
that exist at the surface. Different types of silanol species on the surface can
interact to different degrees. For example, acidic lone silanols will cause the
most peak tailing with basic analytes. A hydrated silanol will not induce much
interaction because it is lower in energy. Some examples of the different forms of
surface silica are shown in Figure 3.
CH3
H3C Si CH3
OH
O
Si
O
O
Si
H3C Si CH3
OH
Si
O
O
H3C Si CH3
OH
Si
Si
O
Si
H3C Si CH3
OH
Si
OH
O
Si
O
O
Si
Si
O
O
Si
CH3
CH3
H3C Si CH3
H3C Si CH3
OH
O
Si
O
O
Si
OH
O
Si
CH3
O
O
Si
H3C Si CH3
OH
Si
OH
O
Si
OH
O
Si
O
O
Si
H3C Si CH3
OH
Si
OH
O
Si
Bridged (vicinal)
Vicinal hydrated
Lone acidic
Metal activated
Geminal
Figure 3: Silica surface silanol groups.
O
O
Si
Advancements in solid support are helping ensure faster and more efficient
HPLC. They include the following supports:
Coreshell: Coreshell particles have a solid silica core and a porous outer layer.
In comparison to traditional fully porous silica supports they produce faster and
more efficient chromatography. They also have a narrow size distribution, which
can contribute to increased chromatographic efficiency.
Monolithic silica rods: Monolithic silica rods allow for high-speed separation with
good resolution and shorter analysis time. These supports contain macropores
that are greater than 50 nm in diameter and mesopores that are 250 nm in
diameter. This structure allows separations to be performed at very low back
pressures and at high mobile-phase linear velocities, or with samples that are
viscous. Monolithic silica rods are also good for direct injection of dirty samples
of plasma or food extracts. Because of the increased flow rate, analysis time is
also reduced.
Fully porous silica (traditional silica): Fully porous silica has a high surface
area and excellent mechanical strength. It can be used as a support material
for normal-phase chromatography, and with surface modification it can be
used for reversed-phase chromatography. As previously stated, one of the
major drawbacks of silica is its susceptibility to hydrolysis at pH extremes. One
way manufacturers have overcome this problem is to use organosilica hybrids.
An organo group grafted into the silica layers makes them more resistant to
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
dissolution at high pH, and this characteristic will extend the column life and
applicability in applications that require the use of high pH.
Porous graphitic carbon: This is a unique chemistry phase. Porous graphitic
carbon is composed of flat sheets of hexagonally arranged carbon atoms;
consequently it has no surface silanols and therefore, unwanted interactions will
not occur. Porous graphitic carbon phases have total pH stability, meaning that
they can be used over the full pH range. This wide applicability of pH makes
them ideal for the analysis of compounds where extreme pH levels are required
to drive the separation. This capability is very good for the separation of strong
acids and bases where the neutral form of the molecule may be required to
increase retention, which requires extremes of pH. This phase is very versatile and
can be used in reversed-phase LC, normal-phase LC, and hydrophilic interaction
chromatography (HILIC), and for LCmass spectrometry (MS) applications.
100
log k
10
0.1
10
12
14
pH
Acetaminophen
Doxepin
Ibuprofen
Imipramine
Nortriptyline
p-Toluamide
Lidocaine
Figure 4: Plot showing the dependence of retention factor for various pharmaceutical compounds on pH. Mobile phase: 35% acetonitrile, 65% 20 mM buffer.
Nonpolar
Polar
N
C
OH
Si
Si
Si
Si
O O O
Alkyl
Dispersive
Phenyl
- interactions
Cyano
Electrostatic
/dipole
Silica
H-bonding
The pH of the mobile phase is an important parameter for the retention of acidic
and basic compounds. As one changes the pH (Figure 4), it is possible to change
the ionization state of acidic and basic molecules; this renders them more or less
polar, which in turn affects their retention time. For basic compounds at a low pH,
the base can accept a proton to become positively charged. As the pH increases,
the protons in the surrounding environment are removed until eventually all the
basic protons within the analyte are abstracted, leaving a neutral species. When
the molecule is charged, there is little retention, but as pH increases, the neutral
form of the molecule becomes apparent, and retention is increased.
The opposite situation occurs for acids, which are proton donors. At low pH, the
neutral form of the molecule exists and, hence, the molecule will exhibit greater
retention. As the pH is increased above the analyte pKa, any acidic protons will be
removed from the analyte to produce a negatively charged species that exhibits
less retention in comparison to its neutral counterpart.
A good rule of thumb for determining the extent of analyte ionization is the 2
pH rule. For acids, at 2 pH units above the analyte pKa the analyte will exist in
the ionized (negative) form. Conversely, for basic moieties, adjusting the pH 2
pH units below the pKa will produce the ionized (positive) species. Therefore, for
ionizable molecules, retention can be altered and controlled by changing the pH
of the mobile phase.
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
We sometimes make the assumption that there is only one mode of interaction
in chromatography when actually there are multiple modes of interactions that
can occur simultaneously within a column. It is important to understand where
those different modes of interactions come from and that on some occasions a
separation scientist may want a particular interaction to drive a separation and
on other occasions that interaction may be undesirable. Thus, it is not possible
to say that a particular column is good or bad without describing the type of
compounds that are being separated.
CN phase
7
4,5
Phenyl phase
2
3
So how do we go about selecting our column, given that there are no really
bad columns? To answer this we need to be able to fingerprint the retention
mechanisms of a column and better understand how they interact with the
molecules that we are trying to separate.
5
2
34
C8 phase
6
Time (min)
15
20
C(2.8)
10
A variety of modes of interaction potentially can exist between analytes and the
stationary phase:
Dispersive forces: These forces exist in all molecules and are the major retention
mechanism for alkyl phases. Retention is proportional to the hydrophobicity of
the molecule. This means that the more hydrophobic the molecule, the longer the
retention time.
H/10
C(7.0)/10
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
H/10
the compounds that are eluted first. Some compounds are not eluted at the same
retention time from the various stationary phases and a degree of orthogonality
appears among these different phases.
H/10
C(7.0)/10
C(7.0)/10
We have talked about different modes of interactions, but how can we start to
quantify those modes? The Physical Quantitation Research Institute (PQRI) has
been trying to gain a better understanding of the different interactions that
molecules can have with the stationary phase. The radar plot shown in Figure
7 was generated for a Hypersil Beta Basic C18 column. This is the fingerprint or
characterization of this particular column. To get this information, it is necessary
to test individual columns under the same conditions, using identifiable test
probes throughout the testing regime.
Column Comparison
C(2.8)
Type A
C(2.8)
Type B
Both Type B
Figure 8: Column characterization plots for type A and type B columns (left)
and two type B columns. See Figure 7 for symbol identification.
11
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
12
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
Mass loading considerations: The amount of sample that can be loaded onto
a column is dependent on the column dimensions and stationary phase type.
Loading an excess of sample onto a column will result in poor peak shapes (broad
peaks, change in apex retention time, and fronting or tailing peaks) and will
ultimately decrease resolution.
Peak capacity: This parameter is important in modern HPLC and describes the
number of components that can be successfully separated with a given column
under gradient conditions. Peak capacity (P) is calculated using equation 3. The
peak capacity can be optimized by changing the gradient time as a function of
flow rate.
tg
P1+ w
[3]
Summary
It has been shown that numerous parameters pertaining to the stationary phase
and dimensions of an HPLC column should be considered to select the correct
column for a particular application.
13
COLUMN SELECTION
SELECTING COLUMN
STATIONARY PHASES
AND DIMENSIONS
Thermo Scientific
Accucore Vanquish Columns
1.5 m solid core particles for
unmatched resolution and throughput
2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are
the property of Thermo Fisher Scientific and its subsidiaries.
Factors to Consider
By Dwight R. Stoll and Scott Fletcher
SPONSORED
Tune Your Mixing Volume for
Gradient Generation
Click to
view PDF
SPONSORED
Eliminating Delays Caused by
Column Wash and Reconditioning in Gradient Methods
Click to
view PDF
15
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
initially at 20% acetonitrile in the mobile phase and then moving to 60% in a
linear gradient over 30 min. One of the key differences that results is that we have
improved resolution, for both the early- and late-eluted compounds. Also, when
we have analytes with very diverse chemistries, we have increased or improved
detection capabilities, because now the later-eluted compounds have much
narrower peak widths and therefore much higher peak heights.
We also have an increased ability to separate complex samples, mainly because
we can spread the peaks out better and because on average they have narrower
widths. This approach can translate to a shorter analysis time. Because the mobile
phase has the ability to elute strongly retained compounds at the end of the run,
column deterioration from the retention of those compounds is avoided.
1,2
1
0
25
50
75
Time (min)
10
15
20
25
30
Time (min)
16
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
in Figure 3, which is a plot of log of retention factor, k, versus the composition of
the mobile phase expressed as a ratio, .
100
As can be seen in the figure, for a rather small simple molecule like benzene,
the retention of that molecule is reduced as we increase the amount of organic
solvent in the mobile phase, but that change is rather slow compared to a
peptide like enkephalin, which has a much steeper slope. For a small protein like
lysozyme this dependence becomes very strong, and with a small change in the
concentration of organic solvent in the mobile phase, the compound is either very
highly retained or not retained at all. So this dependence of the retention of these
molecules on the mobile-phase composition is very important.
Leucine enkephalin
s = 11
Lysozyme
s = 40
Benzene
s = 2.7
As mentioned, one of the major benefits of gradient elution is the fact that
narrow peaks are obtained, where the peak width is nominally independent of the
retention time. So lets investigate this advantage in greater detail. A significant
factor is the focusing of the analyte band at the inlet of the column. Figure 4
includes plots of two analytes and shows how they are affected during a gradient
separation below the column diagram. The top one shows the distance that the
analytes travel in the column as a function of time and the bottom plot shows the
retention as a function of time.
10
0.14
0.18
0.22
0.26
0.30
0.34
0.38
0.42
These two plots provide different perspectives on how the analytes are behaving
inside the column. But the conclusion is that when the elution strength of the mobile
phase is low, the analytes come into the column and basically stick at the column inlet
they have very high retention and very low velocity. As the elution strength of the
mobile phase increases, the retention of those compounds goes down, as shown in
the lower graph in Figure 4, and at the same time, their velocity increases.
A secondary effect that contributes to the narrow peak width is that the mobilephase composition in the column close to the analyte band is weaker than the
solvent composition thats coming behind the band. Thus, the mobile phase that
follows the analyte through the column tends to have a slightly higher elution
strength, which tends to give the analyte molecules in the tail of the peak a higher
velocity, whereas the solutes on the leading edge of the peak have slightly higher
retention and lower velocity. These factors again compress the band somewhat
and also lead to narrow peak widths.
17
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Both high- and low-pressure pumping systems are used for gradient separations.
The first type, a high-pressure binary pumping system, is shown in Figure 5. In
the lower left and right parts of this figure are two independent pump heads.
One of them is pulling in solvent such as water from a bottle going through a
degasser and the other one is pulling in a second solvent, such as acetonitrile
or methanol. The solvent, or mobile phase, is then pumped out of these two
pump heads and mixed in a low-volume mixing chamber, where it goes through a
secondary mixture chamber and a pulse-dampening device to minimize pressure
fluctuations during the flow through the column.
20
30
40
50
60
70
80
90
100
Organic
modifier (%)
Distance (cm)
0 10
Start
End
14 min 22 min
Start
20
10
End 0
10
20
Time (min)
18
Its important to emphasize that the solvents are mixed under high-pressure
conditions. This pump design is typically characterized by a low internal mixing
volume, which is a very important factor with respect to gradient dwell volume,
which is the volume in the system from the point where the gradient is formed to
the top of the column. But on the other hand, they tend to be more complicated
designs and typically are more expensive to purchase.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Pulse damper
To autosampler
Low-volume
mixing chamber
This test can be done in different ways with various solvents used as solvents A
and B. One common way to conduct this test is to use pure water for A, and then
for B, to use water spiked with some compound that absorbs UV light, such as
acetone or benzyl alcohol.
One good approach is to use a 50:50 mixture of methanol and water for these
tests. If you use pure water or a pure organic solvent, sometimes the test
19
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Proportioning valve
molecule will adsorb onto various instrument components. There are other
considerations, too. In the case of biological applications, for example, you
should use mobile phases that are similar to the mobile phases that actually are
going to be used in your application. And when your mobile phases consist of
highly aqueous solutions, benzyl alcohol may not be soluble enough; in such
cases, acetone, uracil, or thiourea would all be good alternatives.
Pulse damper
Ternary pumps
Quaternary pumps
Outlet valve
To autosampler
Inlet valve
500
0.125
150
0.250
0.550
100
75
100
50
50
25
0
0
20
40
60
Time (min)
80
100
20
%B
Absorbance (mAU)
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
RS = 6.97
volume act like the system with the higher dwell volume by deliberately
programming into the pumping system control an isocratic hold at the beginning
of the run to effectively mimic the high gradient delay volume.
RS = 1.63
Washout Volume
0
10
15
20
RS = 5.91
RS = 1.19
10
15
20
20
Absorbance (mAU)
-0
-20
-40
-60
-80
-100
-120
-140
-160
0.0
0.2
0.4
0.6
Time (min)
21
0.9
1.0
Figure 9: Graphical
display of washout
time, which is the delay
in time from when
the pumping system is
programmed to change
the solvent composition
relative to when the
composition actually
changes. Adapted
with permission from
reference (2).
So far we have discussed the characteristics of the gradient profile that we can
test by carrying out the composition steps and looking at what happens at
the detector. We also talked about the dwell volume, which is the delay of the
gradient actually arriving at the column. Lets now turn our attention to what
happens at the end of the gradient.
Typically a scouting type of gradient proceeds from 10 to 90% B during the run.
At the end of the gradient, we make a step change from 90% B back down to
10% B to equilibrate the system and column for the next injection of sample and
the next gradient elution. Chromatographers should be aware that there is also
a delay in that process caused by the washout volume in the system. Although a
step change is made from 90% down to 10%, it doesnt happen immediately.
This is exemplified in Figure 9, which shows the delay when using two solvents, A
and B, where B is spiked, in this case water spiked with acetone. If a step change
from 100% B to 0% B is made at time 0, we see that there is a slight delay and
then an exponential flush of the B solvent out of the system.
This delay is measured using an approach similar to that used to measure the
dwell volume, and for the purpose of discussion, we characterize this washout
volume by looking at the time it takes for the B solvent to be 97% flushed out
of the system. This washout volume becomes important in determining or
estimating how much time we should allow for reequilibration of the analytical
column because we want to make sure that the analytical column is prepared
for the next run by flushing the final mobile phase composition out and refilling
it with whatever solvent composition we are using at the start of the gradient
elution run.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Gradient slope
New system
Programmed
gradient
We can devise a way to systematically determine times that we should use for
these various factors when transferring a method from one system to another.
With respect to washout volume, we can look at the ratio of the washout volumes
on the two systems (see Figure 10). Equation 1 can be used to readjust our
expectations for how much time we need to allow for the last segment in the
gradient on the new system:
Gradient slope
Original system
Composition or response
80%
Gradient profile
Original system
Gradient profile
New system
Lets now turn our attention to optimizing essential gradient parameters and
in particular the benefits of running a scouting gradient. A scouting gradient
is probably the most important step in developing any method and makes it
possible to account for the wide polarity of analytes.
20%
Gradient dwell
New system
Time (min)
Figure 10: Plots showing how the washout volume can impact the transfer of a
method from one system to another.
Final %B
%B
tg
Purging
Conditioning
Initial isocratic
hold
Reequilibration
Initial %B
Time
22
New segment time = original segment time X (original system washout volume/
new system washout volume) [1]
When we dont know how many compounds or the types of compounds we are
looking for, we need to understand the range of analyte polarities during the
method development process (the essential gradient parameters are shown in
Figure 11) so that we can encompass and retain as many of those analytes as
possible. And to give ourselves the best chance of capturing these analytes, we
use a scouting gradient for the most nonpolar analytes that starts at 5% B and
goes up to 100% B (that is,100% organic mobile phase); this gradient elutes the
most highly retained, nonpolar (hydrophobic) analytes and also provides the best
chance of retaining the more polar, hydrophilic, analytes. The information that
we gather from this initial scouting gradient is helpful in determining whether a
gradient is needed or whether the method should be run isocratically.
Isocratic runs will provide the best resolving power for analytes of similar
polarties and the best indication of whether the analytes are interacting with the
stationary phase as much as possible. So a scouting gradient run may indicate
that an isocratic run is recommended or it might suggest the use of a gradient
run because of the differing polarity of analytes. However, it will be extremely
difficult to pick an isocratic mobile-phase composition that will retain the highly
polar analytes and not retard the more hydrophobic analytes so much that the
peaks broaden or remain bound onto the stationary phase. If the scouting run
is advising the use of an isocratic mobile phase it can also tell us what mobilephase composition to use and, if a gradient approach is suggested, it will indicate
whether we can actually increase our initial and final organic compositions or
perhaps decrease them to save time.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
1.0
2.0
1.0
10
3.0
4.0
2.0
5.0
3.0
Initial %B 5
Final %B 100
%B/min 1.9
Gradient time 50 min
15
6.0
7.0
4.0
Initial %B 20
Final %B 100
%B/min 1.9
Gradient time 40 min
8.0
5.0
Initial %B 40
Final %B 100
%B/min 2.0
Gradient time 30 min
where tf and ti are the final and initial retention times, respectively, and tG is the
total time during which the eluent composition is changing. If that difference is
25% or greater, then we typically recommend using a gradient, whereas if it is less
than 25%, an isocratic run is usually optimal. If the analytes are eluted significantly
below the 25% threshold of the gradient, we want to know what isocratic portion
to run. To identify that portion, there are a couple of further calculations that can
be used to better understand the average retention time that is, the retention
time in the middle of the peak elution window. We also need to calculate the rate
of change of the organic component of the mobile phase (the speed at which
the mobile-phase composition is changing every minute). For example, in the
method described previously, if we change from 95% aqueous down to 0% over
20 min, the rate is about 4.75%/min. This rate can be calculated by dividing the
difference between the initial and final %B by the time of the gradient. We can
then use these two values to carry out further optimization studies of the gradient
parameters. For the sake of clarity, these equations will not be described but
instead we will provide a general overview of the optimization procedure.
Initially, we need to know the percentage of organic solvent in the isocratic
mobile phase. It can be determined by adding the initial %B to the amount
that the organic composition has increased by the time a peak is eluted, or by
the time the middle of that peak is eluted, if its an isocratic elution. If we then
multiply the average retention time by the rates of change of %B, the summation
of that plus the initial concentration tells us what mobile-phase composition the
pumps are pumping, which is a very useful parameter to know.
However, that composition is not what is passing through the column. We therefore
need to account for the delay or dwell volume. The way we do that is to convert
the dwell volume back to a time by dividing dwell volume by the flow rate, and
then multiplying that value by the rate of change in units of %B per minute. Then,
by subtracting the %B value obtained from the previous calculation from what the
pumps are pumping, we can determine what mobile-phase composition is passing
through the column at the time the analytes are detected. Because the analytes have
passed through the column and have been detected, we subtract 10%. Essentially,
we are calculating what mobile-phase composition is passing through the column
when the middle of that peak grouping is eluted, and then we take away 10%.
23
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
If we are optimizing the parameters for a gradient analysis, we repeat the same
calculation twice, but rather than using the average peak retention time, we use
the retention time of the first peak to be eluted, and then we calculate when the
last peak is eluted. When we use the initial peak retention time, we obtain the
initial %B, and when we use the final retention time, we obtain the final %B.
Initial %B 10
Final %B 100
%B/min 1.5
Gradient time 60 min
0
10
Initial %B 10
Final %B 60
%B/min 1.43
Gradient time 35 min
0
10
Initial %B 10
Final %B 40
%B/min 1.5
Gradient time 20 min
0
10
100% B
Steep
Shallow
We typically use the same range as with an isocratic separation, looking for a
retention factor somewhere between 2 and 10 with conventional HPLC systems.
However, for modern ultrahigh-pressure liquid chromatography (UHPLC) columns,
values of 0.55 are fairly typical.
100% B
tg = 20
tg = 5
100% B
0% B
100% B
tg = 40
tg = 10
S = 0.25MW0.5 [4]
0% B
0% B
10
10
20
30
40
24
So we take the square root of the molecular weight of the analyte, which really
drives its S value, and then we multiply it by 0.25. As a rule of thumb, if you work
on anything less than a 1000 Da in size, an S value of 5 is a very good starting
point.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Initial %B 10
Final %B 90
%B/min 1.333
Gradient time 60 min
Flow rate 0.5 mL/min
Column length 150 mm
Column i.d. 4.6 mm
Rs = 2.16
10
15
20
Initial %B 10
Final %B 90
%B/min 5.333
Gradient time 15 min
Flow rate 2.0 mL/min
Column length 150 mm
Column i.d. 4.6 mm
Rs = 1.99
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
Initial %B 10
Final %B 90
%B/min 13.333
Gradient time 6 min
Flow rate 5.0 mL/min
Column length 150 mm
Column i.d. 4.6 mm
Rs = 1.66
1.0
2.0
9.0
3.0
Figure 15: Chromatograms showing the effect of changing flow rate and
gradient time on selectivity and sensitivity.
Equation 3 can be rearranged to account for tG , which can be very useful if you
are actually trying to calculate what a gradient time should be. With a known flow
rate, an S value of 5, a of 0.95, and a column volume that has been calculated
using the standard column volume calculation, we can then use a k* value of 5
because we know what we are looking for. And for a standard 150 mm x 4.6 mm
i.d. column with a flow rate of 2 mL/min, we obtain a k* value of 5, which will
result in a tG of about 20 min.
Figure 14 emphasizes what can happen when the rate of change is too fast, or the
slope of the line is too steep. If the gradient time is too short, there is too much
compression of the analyte elution window. Alternatively, if we make the slope
too shallow, we are wasting time, as can be seen with the tG = 40 chromatogram
where there is a significant dead time in the separation.
When analyzing a multiple-component sample, you will find that analytes can be
affected to a different degree by changes in the gradient time. Its not always
the case that reducing the gradient time will improve resolution or increasing
the gradient time will improve resolution depending on the composition of
a sample, the optimal gradient time can be found somewhere in the middle,
which is contrary to the results obtained with isocratic separations. In gradient
separations, changing the gradient time can also change the selectivity, which in
turn changes the resolution. Arbitrarily changing the gradient time can affect the
separation of your samples both positively and negatively.
25
Where Vd is the dwell volume of the system. This rule of thumb is an incredibly
useful guide for estimating the reequilibration time that is required post-gradient.
An important parameter to remember is that a run time is not purely the gradient
time; it is a summation of the gradient time plus reequilibration time. It should always
be determined empirically. Although equation 5 provides a good estimate for the
required reequilibration time, you should always ensure that your analytes are not
affected by insufficient equilibration. Irreproducible retention times can be caused by
giving the column insufficient reequilibration time before the next injection.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
Method Transfer
Now we are going to discuss method transfer and translation in terms of flow
rate, length, and column internal diameter. Previously we talked about gradient
time and column flow rates. Changes in the flow rate can affect resolution and
selectivity. If you want to maintain selectivity, k* should remain the same for
the analytes and therefore resolution is maintained as much as possible. If the
flow rate is doubled, for example, the same k* value (sometimes referred to as
B value) can be maintained by halving the gradient time. If you want to maintain
selectivity, the equation must be balanced by making a proportional change to
the gradient time, as we did for the flow rate, and vice versa.
Figure 15 shows that as we go from a 60-min gradient in the top run, to 15 min
in the middle run, and down to 6 min with the bottom run, the resolution will be
affected. This order of magnitude reduction in run time can be accounted for
and selectivity can be maintained by ramping up the flow rate by an order of
magnitude. Yes, the efficiency has been lost, but selectivity is good and actually
the resolution will be quite adequate in most cases.
190
170
150
130
110
Optimal
range
90
70
50
20
40
60
80
100
120
140
160
tg/t0
Figure 17: Plot of peak capacity against the ratio of gradient time (tG) and the
unretained peak time (t0) showing the optimal range. Adapted with permission
from reference (3).
26
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
(95)
Peak Capacity
1
(5)
10
15
20
25
(95)
Peak capacity is a term that has gained favor in recent years, predominantly
because of the power of modern UHPLC systems, which can resolve a greater
number of peaks in a gradient separation. Peak capacity is defined as the ratio
of the gradient time and the average peak width of the first and last eluted peak,
added to 1, which gives us the theoretical number of peaks that can be resolved.
It is our experience that the practical empirical number of peaks that can be
resolved is an order of magnitude lower than the theoretical number. However, it
is a good way of understanding the efficiency of a separation.
The gradient length for optimum peak capacity should be neither too short nor
too long. Figure 17 is a plot of peak capacity against the ratio of gradient time (tG)
and the unretained peak time (t0), often known as the holdup time. The optimal
range is the highlighted blue zone, where the peak capacity is highest. Very long
gradients provide little increase in peak capacity.
(5)
10
15
20
25
(88)
(51)
(33)
10
27
15
There is no question that the gradient profile can affect certain peaks, as
exemplified by the two critical peak pairs shown in Figure 18. There is almost
baseline resolution between the peak pairing 1 and only very poor resolution of
peak pair 2. The segmented gradient used for this separation allows control over
early and later portions of the gradient, but there are no really hard and fast rules
for when to implement the segment change.
So what happens when we slow the gradient down? Figure 19 shows the initial
gradient at the top and the gradient slowed down on the bottom. In this
example, the critical peak pair 2 is resolved by the slower gradient, but peak pair
1 is still fairly problematic. A much better approach is to incorporate an isocratic
hold and isocratic segments within the gradient.
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
(95)
A good place to start is 10% less than where each critical peak pair is eluted and
hold for two to three column volumes. If that hold time is not long enough, hold
for slightly longer. If the mobile phase is too strong, try using a lower % B. This
approach is a little more complex than using a traditional linear gradient from
5% to 95% or 100% B, but it is not that complex; using the calculation described
earlier it is very easy and straightforward to implement.
(5)
(52)
(95)
(40)
(40)
(5)
0
10
20
30
28
GRADIENT METHODS
GRADIENT HPLC:
GRADIENT METHODS
GRADIENT HPLC:
Factors to Consider
29
Versatility unleashed
Experience our innovative and versatile LC solutions for more productivity with less effort. Our philosophy of UHPLC-bydesign creates a culture dedicated to developing unique technologies and workflows to simplify your daily lab work. Ultrahighefficient columns, sample-specific chemistries and detection techniques let you see more analytes than ever before.
Combining intuitive automation tools and our world-class Thermo Scientific Dionex Chromeleon software increases
your productivity beyond traditional concepts, enabling you to tackle the most challenging analytical demands with ease.
HPLC Detectors
By Scott Fletcher
SPONSORED
Charged Aerosol Detection
(CAD) Bibliography
Click to
view PDF
SPONSORED
Electrochemical Detection
(ECD) Bibliography
Click to
view PDF
Lets start by considering the properties of the ideal detector for high performance liquid chromatography
(HPLC). Ideally, we would like to detect the presence of everything in a sample, independent of anything
else thats going on in the background of either the mobile or stationary phase. For example, we might
have a situation where we would like to detect as many of the analytes in our separation as we possibly can.
Alternatively, in a slightly different scenario, we might need more-selective detection, when we want to
measure only the solutes of interest, and make invisible the presence of matrix components that we are not
interested in measuring.
Obviously, we would like the detector to be stable, and for its performance not to vary with changes
in temperature or mobile phase. In a perfect world, we would also like to be able to detect very low
concentrations of analytes. We also want our detector to have certain physical properties that will not
negatively affect the separation procedure. For example, we dont want the detector cell to increase the
volume, because this will cause dispersion of our chromatographic peaks and thus will not only make it more
difficult to maintain the quality of the separation, but also to ensure sensitivity and detection capability.
On the other hand, we also would like to be able to detect the narrow peaks that are associated with
increasingly high performance forms of chromatography, such as ultrahigh-pressure LC (UHPLC), where the
31
DETECTORS
T H E F U N DA M E N TA L S O F
HPLC Detectors
peak volumes may be extremely small. If the detector response time is too slow, it
may miss very sharp peaks that arise between the detector observation periods.
And finally, we would like the detector to be robust and easy to optimize.
32
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
Detection method
Additionally, if we try to detect above the linear range of the detector, then we
overestimate the quantity of any impurities, because we are not counting the area
of the main peak proportionately compared to the increase in the height of the
impurities.
Selectivity
Sensitivity
Refractive index
Low
15 g
Conductivity
Low
1050 ng
Medium
0.51.0 ng
Electrochemical
High
50500 pg
Fluorescence
High
10100 pg
Low
0.11.0 ng
Charged aerosol
Low
0.11.0 ng
UVvis
Inlet capillary
Mobile-phase
flow from
column
Chromatogram
Flow cell
window
Collimated light
from UVvis source
Detector
diode
Mobile-phase flow to
waste, second detector,
or fraction collector
33
Outlet capillary
When detector signal is plotted against concentration, the slope is typically used
to determine the sensitivity of the method and the intercept indicates the degree
of error within the method, which is a direct result of the background response.
However, this is an area of much debate when we start talking about what
constitutes the limit of detection and the minimal detectible amount against the
signal-to-noise ratio.
Table I shows the typical selectivity and sensitivity of seven commonly employed
detectors. As can be seen, the most selective detection methods typically are
the most sensitive. When we require that a detector be more selective, we are
effectively demanding an increase in the specificity of detection parameters,
and its very unlikely that all of these criteria would be met by anything in the
general background noise. In fluorescence, for example, you just dont set the
wavelength at which your compound absorbs; you also effectively couple that
with the emission wavelength. And the chances are extremely unlikely that any
given interfering molecule will have the same set of coupled conditions as the
analyte. Similarly, with electrochemical detectors, you can set the parameters
of the detector to observe only the electrochemical effect of the molecule of
interest, which will often be in a range that other background contaminants are
not responsive to.
But for a nonspecific, nonselective detector, such as an RI detector, noise,
temperature, and environmental changes may affect its performance, so it is quite
difficult to measure very small changes in concentration. Additionally, with some
detectors, particularly with low-selectivity detectors such as RI, its very difficult to
eliminate all the background effects that affect detection capability.
UVvis Detection
Lets now turn our attention to UVvisible, or UVvis, detection by first explaining
what happens in the flow cell. Figure 1 is a diagram of a generic UVvis flow cell
showing the liquid flow from the chromatograph arriving at the cell and passing
through the collimated light of the UVvisible source, which is in line with the
detector. We can use this principle to measure the difference between what is
going into the cell at the front end and what is passing through the cell and being
detected at the back end. This difference in the transmission of light can be
converted into an absorbance signal, which is shown here as the chromatogram.
This peak will be proportional to the concentration, so the more analyte
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
molecules of a substance that pass through the cell, the more light is absorbed,
and therefore the less that comes out at the back end, which results in a larger
peak appearing in the chromatogram.
Chromophores
UV chromophores give the molecule its UV activity. This activity is typically
electronic in nature, so the more mobile the electrons in the conjugated
34
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
system are, the easier it is to see good UV activity. Additionally, more highly
conjugated molecules will tend to absorb higher wavelengths, which translate
to lower energies of UV radiation. A general rule of thumb is that some solvents,
particularly acetonitrile, are transparent to UV light at 190 nm. With methanol
and some other common solvents, it is difficult to detect them below 220 nm. So
broadly speaking, to avoid seeing any significant effect from the background, we
should work above the 210220 nm range, particularly when running gradients
where a changing composition in the background of the solvent could lead to a
sizeable baseline drift.
Achromatic
lens
Vis lamp
UV lamp
Detector Optical
flow cell slit
254 nm
Holmium
filter
E1 / E2
240 nm
Grating
254 +
380 nm
Diode
array
240 +
320 nm
240
320 nm
320 nm
320 nm
Diode-Array Detection
Lets now take a look at diode-array detection (DAD). With these detectors, you
are looking at all wavelengths that are passing through the flow cell, instead of
just one wavelength as occurs with a UVvis detector. There is no wavelength
separation before the detection process. The detector determines which
wavelengths are missing from the original input light source (in other words,
which wavelengths were absorbed by the sample), after absorption has taken
place. So with diode-array detectors, you dont just get an absorption signal from
your solute at a specific wavelength; you actually get real-time spectra from the
molecule. These principles are presented schematically in Figure 2, which shows
35
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
Absorbance (mAU)
Anisic acid
Optimum: Slit: 8 mm (16)
Signal: 255/30 Ref. 340/100
Analytical
wavelength
50
that DAD can be used for detection at single or multiple wavelengths, where
spectra can be dynamically obtained and stored for peak purity analysis, library
searching, or extraction of signals.
40
Reference bandwidth
100 nm
30
20
10
0
Reference wavelength
(290 nm + 50 nm)
340 nm
30 nm
Bandwidth at
50% peak height
220
240
260
280
300
320
340
360
380
400
Wavelength (nm)
Figure 3: Spectrum of an analyte molecule (anisic acid) showing how a diode-array detector can be used monitor both the analytical wavelength and a reference
wavelength at the same time.
Mirror
Emission
monochromator
Lens
Lens
Photomultiplier
Excitation
monochromator
Flow cell
36
Its worth spending some time to understand how the response rate is optimized
for a diode-array detector. Basically, the faster you make the response time,
the faster the ability to respond to whatever species is coming through the cell,
and the more likelihood of increasing peak sensitivity. However, as the response
time goes down, the noise also goes up, so the overall sensitivity that results
from using a higher response factor may not be any better than using a lower
response factor, and may even be worse in some cases. Thus, to get the best
signal-to-noise ratio, these parameters have to be optimized based on the
chromatographic separation conditions and the detection capability required.
Generally speaking, on modern UPHLC instruments where you are using very
efficient chromatography and getting peaks that are 23 s in width, you rarely
get any better response frequency than 40 measurements per second, which
means you dont have to use anything faster than a response coefficient of 40 Hz.
Modern detectors go up to 240 Hz, but as soon you go higher than 40 Hz, you
can start to run into problems with noise.
Photodiode
DETECTORS
T H E F U N D A M E N TA L S O F
Internal
conversion
40
35
30
25
20
15
10
Excitation
spectrum
5
0
300
350
400
450
500
550
600
Wavelength (nm)
Ground state So
Purge valve 2
Purge valve 1
Waste
Fluorescence
250 nm
Excitation
S1
250
you dont want to look at all spectral information. For this purpose the most
important settings on a DAD are the detection wavelength and the bandwidth.
For example, you can choose a detection wavelength such as 250 nm and set the
bandwidth to 7080 nm. In this way, you will actually be detecting everything
that absorbs light at wavelengths ranging from 210 to 290 nm. This can be
problematic with quantitation in a mixture, but it gives you the best chance of
detecting any unknown components in the sample.
S2
Emission
spectrum
200 nm
Norm.
HPLC Detectors
Fluorescence Detection
A schematic of a fluorescence detector is shown in Figure 4. The radiation
source is typically a xenon arc flash lamp, which flashes every 3 s, producing a
continuous spectrum of light from 200 nm to 900 nm. Radiation from the lamp
is focused by the first lens, then reflected by the mirror onto the excitation
monochromator grating, which disperses and reflects the emitted radiation. The
light is then split in the flow cell to allow light to reach both the reference diode
and photomultiplier tube. Before the light reaches the emission monochromator,
a cutoff filter removes light below a certain wavelength to reduce noise from
first-order scatter and second-order stray light. The emission monochromator
determines the wavelength range of light reaching the photomultiplier tube
where the incident photons hit the photocathode and generate electrons, thus
multiplying the signal.
The most important parameters to optimize in a fluorescence detector are the
excitation and emission wavelengths. The excitation wavelength can be taken
from the excitation spectrum obtained on a spectrofluorimeter. The optimum
emission wavelength is dependent on the particular instrument and compound.
Fluorescence detectors can be extremely sensitive, but they detect only
37
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
Eluent
only
Lamp
Photomultiplier
Eluent +
sample
Photomultiplier
Eluent
only
Lamp
As mentioned above, the sensitivity of any detector is not only related to the
intensity of the peak height but also the intensity of the signal noise. Very often
the noise drives down sensitivity and ultimately impacts the detection limit.
Figure 5 exemplifies this for a fluorescence detector. Here is a great example
using a second-order filter. We have a specific excitation wavelength. It can be
seen from the electronic transitions that photons travel from the ground state
to the excited state, and then relax back down to the ground state. This occurs
at approximately 450 nm, where we actually measure the signal. So it is actually
the emission spectrum and not the excitation response that gives us the secondorder separation of the peak from the interference and the background signal. In
this example, it can be seen that the excitation wavelength is within the UV range,
while the emission spectrum is much broader, less defined and usually far more
practical to measure.
Column effluent
Nebulizer
Nebulizer gas
(air or nitrogen)
Nebulizer
chamber
Drift tube
(heated-zone
evaporation stage)
Analyte
Photomultiplier tube
or photodiode
Light source
Light-scattering
cell
Single output
Amplifier
38
The main advantage of fluorescence detectors is that not only do you achieve
good selectivity (because only a small handful of molecules fluoresce), but
you also get high sensitivity, which means that only small sample volumes are
required. But of course, the selectivity of these detectors can actually be a
disadvantage, because of the fact that not many compounds naturally fluoresce.
In addition, this type of detector can be affected by temperature because of
the energy required and the additional collisions that take place, and because
were looking at excitation and relaxation. And both the excitation and emission
wavelengths have to be optimized; you cannot just label the excitation and
emission wavelengths to be used, as is typically done with a UV detector. Also,
these settings tend be very detector-specific; with fluorescence detection,
both the excitation and emission wavelengths have to be set on every different
instrument.
Refractive-Index Detection
Figure 6 shows a schematic that explains how an RI detector works. We see that
there are two cells. On the right hand side, we can see the light path passing
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
through two cells. We have a reference and a sample cell. Before the analysis,
both cells are flushed with the mobile phase. When the injection is made, the
valve is rotated and column effluent then passes through the sample cell, with
the reference cell being filled with just the mobile phase. This technique relies
on comparing the degree of bending or refracting the light between the mobile
phase and the mobile phase containing the sample. So when only pure mobile
phase is coming from the column, that light is perfectly balanced and there is no
signal. As soon as anything different is eluted from the column and into the flow
cell, the degree by which the light is bent changes; the change in refractive index
can be caused by a sample compound or just by a change in the mobile phase.
This process is shown in Figure 7.
Electrometer
Charge is drawn off
and measured by
a sensitive electrometer
Nebulizer and
impactor
HPLC
column
eluent
Signal out
Signal is directly
proportional to
quantity of analyte
in sample
Drying
tube
Ion trap
Negatively charged
ion trap removes
high-mobility particles
Gas
inlet
Collector
Analyte particles
transfer their charge
Large
droplets
to waste
Positive charged
transferred to
analyte particles
by charged opposing
secondary gas steam
39
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
of the refractive-index detector. In addition, it can be used for analytes that dont
have any chromophoric properties, and unlike an RI detector, it can be used for
gradient separations. Its biggest drawback, however, is the fact that you cant use
it for volatile samples because they will be lost via evaporation in amongst the
mobile phase.
Additionally, the mobile phase must be volatile for this technique to work,
although this is not a huge drawback. Another challenge with these detectors is
that the signal does not respond linearly to the concentration.
Working
electrode
Reference
electrode
Electrochemical Detection
Counter
electrode
Figure 10: Schematic of an electrochemical detector.
40
DETECTORS
T H E F U N D A M E N TA L S O F
DETECTORS
T H E F U N D A M E N TA L S O F
HPLC Detectors
Typical samples that are applicable to this type of detection include phenol,
hydroxybenzene, catechol, dihydroxybenzene and similar types of aromatic
functional groups. Other sample matrices that lend themselves to amperometric
detection are catecholamine, dopamine, and epinephrine.
Methoxy
Hydroxy
Nucleobase
Amine
Phenol
Catechol
Nucleoside
Quinol
Quinone
Glycoside
Thiol
Carbohydrate
41
Charged aerosol detection is a revolutionary technology that will change the way you view
Charged aerosol detection has the flexibility to be used for a broad range of analytes in
every sample. This technique delivers consistent analyte response independent of chemical
many different matrices, opening new opportunities for broad discovery and enhanced
routine analysis.
levels.
An analytes response to charged aerosol detection does not depend on optical properties,
Specialty chemicals
chemical derivatization are not essential for detection. Charged aerosol detection is a
mass-sensitive technique that measures any non-volatile and many semi-volatile analytes.
Variance in inter-analyte relative response is minimal, whether analyzing small molecules
The new Thermo Scientific Dionex Corona Veo detector is designed to integrate into
any HPLC/UHPLC system. When combined with a UV, diode array, or mass spectrometer, it
25
Citric acid
provides an orthogonal and complementary detection solution, making it the ideal detector
Theophylline
Phenylalanine
Diclofenac
Progesterone
Naproxen
pA
Propranolol
Charged aerosol
The Corona Veo charged aerosol detector delivers sensitive, universal response through
a simple yet flexible design, perfectly matched for applications with capillary, microbore,
-2
600
Naproxen
Propranolol
mAU
UV
Diclofenac
Phenylalanine
Progesterone
Citric acid
-50
0
Minutes
10
12
14
16
18
Six pharmaceutical agents with an excipient (citric acid) were fully resolved using gradient reversedphase HPLC and their responses measured first by UV detection and then by charged aerosol detection.
As can be seen, UV detection significantly underestimates the levels of most analytes.
@2014 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and
its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries.
Please consult your local sales representative for details.