Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User Manual
For SPM and Software
004-1005-000 (Standard)
004-1005-100 (Cleanroom)
Ref.
DCR
Revision
Date
Approval
02A
5/2007
Pre-Release Draft
N/A
D.Kasai
1/2008
Initial Release
N/A
D.Kasai
11/3/08
New Updates
N/A
D. Paszkeicz
Notices: The information in this document is subject to change without notice. NO WARRANTY OF ANY KIND IS MADE WITH REGARD TO
THIS MATERIAL, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. No liability is assumed for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. This document contains proprietary information which is protected by copyright. No part of this
document may be photocopied, reproduced, or translated into another language without prior written consent.
Copyright: Copyright 2004 Veeco Instruments Inc. All rights reserved.
Trademark Acknowledgments: The following are registered trademarks of Veeco Instruments Inc. All other trademarks are the property of their
respective owners.
Product Names:
NanoScope
MultiMode
Dimension
BioScope
BioScope II
CP II
Atomic Force Profiler
Dektak
Innova
Caliber
(AFP )
Software Modes:
TappingMode
Tapping
TappingMode+
LiftMode
AutoTune
TurboScan
Fast HSG
PhaseImaging
DekMap 2
HyperScan
StepFinder
SoftScan
Hardware Designs:
TrakScan
StiffStage
Hardware Options:
TipX
Extender
TipView
Interleave
LookAhead
Quadrex
Software Options:
NanoScript
Navigator
FeatureFind
Miscellaneous:
NanoProbe
WARRANTY INFORMATION
This product is covered by the terms of the Veeco standard warranty as in effect on the date of shipment
and as reflected on Veeco's Order Acknowledgement and Quote. While a summary of the warranty
statement is provided below, please refer to the Order Acknowledgement or Quote for a complete
statement of the applicable warranty provisions. In addition, a copy of these warranty terms may be
obtained by contacting Veeco.
WARRANTY. Seller warrants to the original Buyer that new equipment will be free of defects in material
and workmanship for a period of one year commencing (x) on final acceptance or (y) 90 days from
shipping, whichever occurs first. This warranty covers the cost of parts and labor (including, where
applicable, field service labor and travel required to restore the equipment to normal operation).
Seller warrants to the original Buyer that replacement parts will be new or of equal functional quality and
warranted for the remaining portion of the original warranty or 90 days, whichever is longer.
Seller warrants to the original Buyer that software will perform in substantial compliance with the written
materials accompanying the software. Seller does not warrant uninterrupted or error-free operation.
Sellers obligation under these warranties is limited to repairing or replacing at Sellers option defective
non-expendable parts or software. These services will be performed, at Sellers option, at either Sellers
facility or Buyers business location. For repairs performed at Sellers facility, Buyer must contact Seller in
advance for authorization to return equipment and must follow Sellers shipping instructions. Freight
charges and shipments to Seller are Buyers responsibility. Seller will return the equipment to Buyer at
Sellers expense. All parts used in making warranty repairs will be new or of equal functional quality.
The warranty obligation of Seller shall not extend to defects that do not impair service or to provide
warranty service beyond normal business hours, Monday through Friday (excluding Seller holidays). No
claim will be allowed for any defect unless Seller shall have received notice of the defect within thirty days
following its discovery by Buyer. Also, no claim will be allowed for equipment damaged in shipment sold
under standard terms of F.O.B. factory. Within thirty days of Buyers receipt of equipment, Seller must
receive notice of any defect which Buyer could have discovered by prompt inspection. Products shall be
considered accepted 30 days following (a) installation, if Seller performs installation, or (b) shipment;
unless written notice of rejection is provided to Seller within such 30-day period.
Expendable items, including, but not limited to, lamps, pilot lights, filaments, fuses, mechanical pump
belts, V-belts, wafer transport belts, pump fluids, O-rings and seals ARE SPECIFICALLY EXCLUDED
FROM THE FOREGOING WARRANTIES AND ARE NOT WARRANTED. All used equipment is sold
AS IS, WHERE IS, WITHOUT ANY WARRANTY, EXPRESS OR IMPLIED.
Seller assumes no liability under the above warranties for equipment or system failures resulting from (1)
abuse, misuse, modification or mishandling; (2) damage due to forces external to the machine including,
but not limited to, acts of God, flooding, power surges, power failures, defective electrical work,
transportation, foreign equipment/attachments or Buyer-supplied replacement parts or utilities or services
such as gas; (3) improper operation or maintenance or (4) failure to perform preventive maintenance in
accordance with Sellers recommendations (including keeping an accurate log of preventive maintenance).
In addition, this warranty does not apply if any equipment or part has been modified without the written
permission of Seller or if any Seller serial number has been removed or defaced.
No one is authorized to extend or alter these warranties on Sellers behalf without the written authorization
of Seller.
THE ABOVE WARRANTIES ARE EXPRESSLY IN LIEU OF ANY OTHER EXPRESS OR IMPLIED
WARRANTIES (INCLUDING THE WARRANTY OF MERCHANTABILITY), AND OF ANY OTHER
OBLIGATION ON THE PART OF SELLER. SELLER DOES NOT WARRANT THAT ANY
EQUIPMENT OR SYSTEM CAN BE USED FOR ANY PARTICULAR PURPOSE OR WITH ANY
PARTICULAR PROCESS OTHER THAN THAT COVERED BY THE APPLICABLE PUBLISHED
SPECIFICATIONS.
NO CONSEQUENTIAL DAMAGES. LIMITATION OF LIABILITY. Seller shall not be liable for
consequential damages, for anticipated or lost profits, incidental, indirect, special or punitive damages, loss
of time, loss of use, or other losses, even if advised of the possibility of such damages, incurred by Buyer
or any third party in connection with the equipment or services provided by Seller. In no event will Sellers
THE ABOVE WARRANTIES ARE EXPRESSLY IN LIEU OF ANY OTHER EXPRESS OR IMPLIED
WARRANTIES (INCLUDING THE WARRANTY OF MERCHANTABILITY), AND OF ANY OTHER
OBLIGATION ON THE PART OF SELLER. SELLER DOES NOT WARRANT THAT ANY
EQUIPMENT OR SYSTEM CAN BE USED FOR ANY PARTICULAR PURPOSE OR WITH ANY
PARTICULAR PROCESS OTHER THAN THAT COVERED BY THE APPLICABLE PUBLISHED
SPECIFICATIONS.
NO CONSEQUENTIAL DAMAGES. LIMITATION OF LIABILITY. Seller shall not be liable for
consequential damages, for anticipated or lost profits, incidental, indirect, special or punitive damages, loss
of time, loss of use, or other losses, even if advised of the possibility of such damages, incurred by Buyer
or any third party in connection with the equipment or services provided by Seller. In no event will Sellers
liability in connection with the equipment or services provided by Seller exceed the amounts paid to Seller
by Buyer hereunder.
Service
Field service is available nationwide. Service and installations are performed by factory trained Veeco
service personnel. Contact the Veeco Metrology sales/service office for prompt service.
Veeco Instruments Inc.
112 Robin Hill Road
Santa Barbara CA 93117
Attn.: Service Center
Phone: (805) 967-2700
Fax: (805) 967-7717
www.veeco.com
Disclaimer:
Some images contained in this document may differ from installed equipment. The differences are usually
cosmetic only and still provide useful references for the accompanying text.
Table of Contents
Ch. 1 - Introduction 1
1.1 Overview of Manual ................................................................................................ 1
1.2 Safety ....................................................................................................................... 1
1.3 Microscope Specifications ....................................................................................... 2
1.3.1 Special Hardware Requirements.....................................................................2
1.3.2 Scanner Types.................................................................................................2
1.3.3 Scanning Techniques ......................................................................................2
1.4 SPM Fundamentals .................................................................................................. 4
1.4.1 Terminology....................................................................................................4
1.4.2 SPM Overview................................................................................................5
1.4.3 The Feedback Loop ........................................................................................6
1.4.4 Scan Size, Scan Rate, Feedback Parameters (gains) and Setpoint .................7
1.4.5 Main menu items ..........................................................................................11
1.4.6 Z Position Bar: Monitoring the Scanner's Position......................................14
1.4.7 Feedback Signal: Monitoring the Feedback Signal and Setpoint Values ....15
1.4.8 Setpoint: Setting the Reference Signal for the Feedback Loop ...................15
1.4.9 Setting the Gain of the Feedback Loop ........................................................16
1.4.10 Feedback Checkbox: Setting the Scanner's Z Position..............................17
Table of Contents
Ch. 3 - Safety 29
3.1 Operating Safety .................................................................................................... 29
3.1.1 Safety Symbols .............................................................................................29
3.1.2 Definitions: Warning, Caution, and Note .....................................................30
3.1.3 Summary of Warnings and Cautions ............................................................31
3.1.4 Grounding Innova .........................................................................................32
3.1.5 Setting the Line Voltage ...............................................................................32
3.2 Laser Safety ........................................................................................................... 33
3.3 Specifications and Performance............................................................................. 35
3.4 Innova User Documentation .................................................................................. 37
3.4.1 Innova Users Guide ....................................................................................37
3.4.2 SPMLab Display and Image Analysis ..........................................................37
Table of Contents
4.10.2 Aligning the Deflection Sensor...................................................................58
4.11 Troubleshooting Tips ........................................................................................... 59
4.12 Typical startup ..................................................................................................... 59
4.13 Deflection Sensor................................................................................................. 62
4.13.1 Alignment Knobs ........................................................................................62
4.13.2 Laser Indicators...........................................................................................64
4.14 Motor stage controls ............................................................................................ 66
4.15 Engage ................................................................................................................. 67
Ch. 6 - Menus 85
6.1 File Menu ............................................................................................................... 85
Table of Contents
6.2 Setup Menu ............................................................................................................ 86
6.3 Real Time Control Menu ....................................................................................... 87
6.4 Tools Menu ............................................................................................................ 88
6.5 Window Menu ....................................................................................................... 89
6.5.1 Control Pane Info Menu ...............................................................................90
6.6 Toolbuttons ............................................................................................................ 90
6.7 Other Controls........................................................................................................ 91
Table of Contents
8.2 Startup .................................................................................................................. 114
8.3 Cantilever Tuning - Manual Tuning Method...................................................... 115
8.4 Autotune............................................................................................................... 123
8.5 Engage ................................................................................................................. 127
8.6 Scanning Windows .............................................................................................. 130
Table of Contents
10.4 Engage ............................................................................................................... 153
10.5 Prepare to Ramp................................................................................................. 154
10.6 Ramp .................................................................................................................. 156
10.7 Sample Session with Probe Positioning............................................................. 158
Table of Contents
13.2 Startup ................................................................................................................ 187
13.3 Additional Instructions and Information............................................................ 188
13.4 Nanolithography - A Sample Session ................................................................ 188
Table of Contents
A.4 Run Synchronizer................................................................................................ 222
Ch. 1 - Introduction
Overview of Manual
Chapter 1
1.1
Introduction
Overview of Manual
This manual provides information specific to the Innova system. This manual describes installation
procedures and provides information on the operation of the SPMLab program. The information
includes description of several modes of operation and imaging as well as real time data analysis.
Some chapters are for reference only and describe modes and capabilities which are not included in
the basic Innova system. Applications modules are required to obtain these additional capabilities.
Innova uses a stationary probe and samples are scanned back and forth beneath the probe.
Typically, Innova samples are attached to round metal disks (pucks) which are magnetically
attached to the top of the scanner tube. The scanner moves the sample in the x,y and z direction and
the probe obtains information from the sample. The type of information will depend on the data
channels chosen.
PC
Innova
Dual Monitors
NanoDrive
Controller
Isolation table
1.2
Safety
The Innova system is designed with features intended to safeguard operators from injury, prevent
damage to the test samples and to be robust against damage from normal use. As with any
electrical, mechanical device, some hazard is inevitable. Chapter 3 provides important safety
information which should be read and understood by all users.
Rev. B
Ch. 1 - Introduction
Microscope Specifications
1.3
Microscope Specifications
Small Area
Large Area
Contact Mode Imaging - Measures height by sliding a probe tip across and in contact
with the sample in either air or liquid. LFM (Lateral Force Microscopy) characterizes
frictional effects. See Chapter 7.
Rev. B
Ch. 1 - Introduction
Microscope Specifications
These imaging techniques are described in this manual. Additional information may be available in
support notes (See Chapter 2). Contact Veeco if additional information is desired.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
1.4
SPM Fundamentals
1.4.1 Terminology
This section contains a brief list of terms and abbreviations to assist the reader. Other terms and
abbreviations are referred to in the Index at the end of this manual.
AFMAtomic force microscopy; atomic force microscope (see SPM).
biasElectrical potential applied to a tip or sample which causes electron flow from one to the
other.
calibrationMeasurement of known features to ensure accuracy of SPM images.
cantileverFlexible portion of probe extending from the substrate and to which the tip is attached.
cantilever tuneProcess of finding a cantilevers natural, resonant frequency by exciting the
cantilever through a range of frequencies until maximum amplitude (resonance) is obtained.
DSPDigital signal processor. Computer processor used to control SPM feedback loop.
drive amplitudeAmplitude of the signal used to oscillate a tip in TappingMode.
drive frequencyFrequency of the signal used to oscillate a tip in TappingMode.
engagementProcess of bringing a probe tip and sample together in a controlled manner such that
useful information about the surface is obtained without damaging either the tip or the sample.
errorDifference between feedback signal value and feedback setpoint.
false engagementCondition where probe is not racking the sample surface while in feedback. It
may be caused by long range interactions or air damping. Selecting a more positive setpoint
(contact mode) or smaller amplitude setpoint (TappingMode) may correct the problem
feedbackProcess of self-correction of the error signal.
fluid cellAccessory used for imaging materials in fluid.
integral gainThe multiplier of correction applied in response to the average error signal.
probeThe mechanical device (usually a integrated substrate, cantilever and tip) used for
imaging.
probe holderRemovable appliance for mounting SPM probes.
proportional gainThe multiplier of correction applied in response to the error signal in direct
proportion to the error.
PSPD - Position Sensitive Photo Detector. Produces output based on the position and intensity of a
laser spot. The Innova AFM utilizes a quad photodetector (QPD) form of pspd.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
QPD - Quad Photo Detector. A form of PSPD consisting of a four photodetector segment array.
The output of each segment varies with the total intensity of the laser spot (or portion of laser spot)
incident on the segment. The outputs of the four segments may be summed or differenced to
determine the position of the laser spot on the array. In TappingMode operation, the laser spot
oscillates across segment boundaries so the output of the QPD is an AC voltage.
RMS amplitudeRoot mean square (RMS) signal measured at the detector. (TappingMode only.)
SPMScanning probe microscopy; scanning probe microscope. A general term encompassing all
types of microscopy which utilize a scanned micro-sharpened probe and feedback circuitry to
image nanoscale phenomena. SPM modes of operation include AFM, ECSTM, EFM, MFM, STM,
and many others.
sensitivityAmount of movement produced by a scanner piezo by a specific input voltage.
setpointOperator-selected threshold used as the target of the feedback control loop.
spring constantAmount of force required to bend a cantilever a specific amount (typical units
are N/m).
tipholderSee probe holder
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
An atomic force microscope (AFM) operating in contact mode, touches the surface of a sample
with a sharp tip (~2 m long and often less than 5 nm radius), which is located at the free end of a
cantilever 100-200m long. The cantilever has a spring constant lower than the forces binding the
atoms of the sample. As the scanner moves the sample under the tip, the contact force causes the
cantilever to bend or deflect in response to changes in height.
A detection scheme measures the cantilever deflection as the sample moves relative to the tip. This
detection scheme consists of a laser reflected off the back of the cantilever and onto a position
sensitive photodetector (QPD). The measured cantilever deflections are used by the computer to
generate a map of surface height.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
1.4.4 Scan Size, Scan Rate, Feedback Parameters (gains) and Setpoint
Scan and feedback parameters are adjusted to obtain satisfactory imaging which depends on
obtaining a stable signal trace, free of noise or spurious signals. Iterative adjustment of some of the
parameters described in this section is usually required to produce a high quality image. All
parameters can be adjusted in real-time during a scan, without lifting the tip.
Scan Size
The scan size refers to the area of the sample which is being imaged. Since all scans are square,
scan width alone is specified to set the scan size. The scan size selected depends on the features of
interest and the size of the scanner.
For the standard Large Area scanner, the maximum scan size is >90 m. When using the 10 m
grating included with the Large Area scanner, a scan size of 30 m will display about three rows of
the grating. With a scan size of 10 m, only one row of the grating will be displayed.
A scan size may be chosen so that the features of interest are represented in reasonable proportions.
For example, using a 1 m calibration grating, a scan size of 5 to 10 m will display several rows of
the grating.
Generally, scan size should be selected to be greater than the lateral resolution multiplied by the
number of data points per scan line, i.e., the spacing between the pixels should be larger or
comparable in size to the area covered by each data point. Selecting a smaller scan size will result
in adjacent data points containing redundant information.
Scan Rate
The scan rate sets the frequency of the back and forth rastering of the sample beneath the probe.
Scan rate can be adjusted while acquiring an image. In general, as the scan size is increased, the
scanning velocity also increases (a longer line is scanned at the same scan rate, so the tip travels
faster over the sample). Large scan sizes usually are better imaged by decreasing the scan rate in
order to decrease the tip to sample scanning velocity.
Generally, slower scan rates give better resolution because the feedback system has time to
respond, while faster scan rates save time. If the scan rate is too fast, the feedback loop may not
respond appropriately to changes in height. As a result, the image may appear smeared (poor
surface tracking) or the tip may crash into protrusions on the surface, possibly damaging the tip
and/or sample. The scan rate setting interacts with the gain setting (see the following section). The
higher the gain setting, the faster the system can track the height and the faster the scan rate may be
set and still obtain good images.
The scan rate also depends on the type of image being taken. For images of height using a scan size
from 1 to 10 m and with feedback enabled, typical scan rates are from 1 to 4 Hz.
To determine which scan rate is best for a particular set of scan conditions, take several scans at
different rates and use a scan rate below the rate at which image quality begins to degrade.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
Deflection
Setpoint
Deflection Signal
Figure 1.4b Hardware components and signal pathways for contact mode.
The cantilever deflection is compared to the setpoint value in the setpoint scrollbox (the reference
value for the z feedback loop), and an error signal is generated. The error signal is sent to the
feedback electronics, which generates a feedback voltage. This feedback voltage controls the
scanner tube, causing it to extend or retract. The probe to sample spacing is controlled to maintain a
constant cantilever deflection which minimizes the error signal. The feedback signal can be used to
generate an image of sample height.
Gain
The magnitude of the error signal is often insufficient to cause the z scanner to respond properly.
To obtain better response, a gain is specified. The gain (of the feedback loop) is a multiplier
applied to the Error signal to generate the feedback signal to the scanner. This parameter must be
optimized for each set of scan conditions. Higher gain values mean that the feedback loop is more
responsive to changes in deflection of the cantilever. Proper gain settings are required for good
imaging.
If the gain is too high, the feedback signal will over react to small changes and the system will
oscillate. Feedback oscillations appear on the signal trace as fringes or ripples. If the gain is set too
low, the z feedback will not track surface height properly. When surface tracking is poor, surface
features can appear lopsided or the tip can damage features on the surface or be damaged by the
sample.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
Set gain by beginning with the default value. Gradually increase the value until the system begins
to show oscillation and reduce the gain until the oscillation ceases. An acceptable gain setting
usually will lie within a range of settings rather than at a single value.
Figure 1.4c shows the signal trace on the Profile Display in the scanning window. As the gain is
increased from the default value, the contours of the grating lines will be reestablished on the
Oscilloscope Display.
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
When scan conditions change (for instance, at different portions of the sample), the gain parameter
should be verified as still optimized. If oscillations appear, lower the gain until the system is just
below the oscillation point.
Setpoint
In contact imaging mode, setpoint represents the amount of cantilever bending, or deflection. High
setpoint produces higher contact force between the tip and sample in contact mode. When the z
feedback loop is enabled, the system operates to keep the amount of cantilever bending constant by
moving the sample up or down. In TappingMode, setpoint represents the amplitude of cantilever
oscillation. High setpoint produces lower contact force between the tip and sample and because the
oscillation is less affected by interaction with the sample. The high setpoint relates to a high RMS
voltage resulting from higher amplitude oscillation of the cantilever. The remainder of this
discussion deals with contact mode setpoint. Additional discussion of TappingMode setpoint is
contained in Chapter 8.
The optimal value of the setpoint parameter depends on a number of factors, the most influential is
the sample. If horizontal streaking occurs in an image, the setpoint is too high (the cantilever is
exerting more force against the sample surface), and dirt (unidentified particles, material, etc.) is
being dragged on the sample. If streaking remains after reducing the setpoint value, the sample is
probably too soft to examine using contact mode -- or a softer cantilever may be needed. Setpoint
can be adjusted during a scan so that its effect may be observed in real time. When optimizing the
setpoint, it is useful to compare the forward and reverse traces on the oscilloscope display, as
described in Chapter 7.
If the setpoint is too low, the probe will not be able to track the sample height. Check the signal
trace on the profile display in the scanning window to see if it realistically represents the height.
When imaging the calibration grating with a setpoint value that is too low, the shape of the signal
trace will appear to flatten out, indicating that the probe is unable to follow the height.
10
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
X and Y Offsets
X offset; Y offsetThese controls allow selection of the center of the area to be scanned.
Scan Angle
Scan angleCombines X-axis and Y-axis drive voltages, enabling the piezo to scan the sample at
variable X-Y angles.
Scan Rate
Scan rateSets the number of lines scanned per second.
Leveling
Leveling is used when the average surface of the sample is not parallel to the scanning plane. The
leveling options are:
None
1DAC - subtracts the average from the signal for each line
1D line fit - subtracts a 1st order least mean square fit for each line
1D bow removal - subtracts a 2nd order least mean square fit for each line
2D AC - subtracts the average for all the date acquired at the time
2D plane - plane tilt removal for all data acquired at the time
Rev. B
The high gain range is ~0 to 400 volts and the low gain range is ~0 to 100 volts. If switching
from high gain to low gain mode, avoid damage to the probe tip and sample in by raising the Z
stage a short distance using the z direction pad to ensure that the probe tip clears the sample
surface.
11
Ch. 1 - Introduction
SPM Fundamentals
Both high- and low-gain modes are available on Innova. High gain mode accommodates most
applications. Low gain mode produces the best noise performance for ultra flat samples a atomic
resolution with the large area scanner. It is not usually necessary to select the low gain mode when
using the small area scanner.
High gain mode applies the full voltage range to the scanner to produce xy and z motion and is
most often used for scan sizes in the micron range. The maximum scan size in high gain mode is
limited by the available range of scanner motion. The maximum xy range of a Large Area scanner
is nominally 90 m. The maximum z range of a Large Area scanner is nominally 7.5 m.
Low gain mode uses only a portion of the scanner xy and z range and is generally used for smaller
scanson the order of tens to hundreds of nanometers. The range of xy motion is reduced to ~1/4
of its full range, and the range of z motion is reduced to ~1/3 of its full range. For example, the xy
range of a Large Area scanner is reduced to ~25 m, and the z range is reduced to ~2.5 m.
This section explains why low gain mode is important for obtaining improved lateral resolution
with smaller scan sizes and gives step-by-step instructions for switching from high gain to low gain
mode.
the scan size divided by the number of data points per scan line
If a 10 m image is taken with 256 x 256 data points, then one data point is taken every 10 m/256,
or 39 nm, which represents the limit of the lateral resolution. Lateral resolution will be better by a
factor of 2 for a 10 m image taken with 512 x 512 data points, however, it will take twice as long
because there are twice as many lines of data.
In order to improve the lateral resolution as limited by the scan size without increasing the time to
take an image, use a smaller scan size. For example, for a 256 x 256 image and a scan size of 50 nm
(0.05 m), the lateral resolution limit improves to 50 nm/256, or 0.195 nm.
The lateral resolution is no better than the largest limiting factor. For small scan sizes (below 5
m), the largest limiting factor is not likely to be the scan size divided by the number of data points
per scan line. In general, do not select a scan size that is smaller than the lateral resolution (as
limited by any of the factors described here) multiplied by the number of data points per scan line.
Selecting a smaller scan size will result in adjacent data points containing redundant information.
12
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
In most cases, the primary limiting factor to the lateral resolution is the interaction area between the
tip and the sample, or the effective tip radius. The interaction area is affected by type of imaging
being performed, the characteristics of the surface being imaged (height, chemistry, surface fluids,
field effects). The sharpness and geometry of the tip also influence the interaction area and lateral
resolution.
In STM mode, the exponential relationship between tunneling current and tip-to-sample spacing
isolates the interaction between the tip and the sample to atoms at the very end of the tip. Thus,
even a very blunt tip with a radius on the order of 100 nm can be used in STM mode to achieve
atomic resolution when the tip has a single atom that protrudes more than its neighbors. This same
tip, however, may not be able to resolve features that are wide if those features are also very deep
(high aspect ratio features).
For other modes, the lateral resolution as limited by the effective tip radius is on the order of
nanometers to tens of nanometers. Factors such as tip wear and deformation increase the interaction
area for contact mode operation. The response of the measured signal to changes in tip-to-sample
spacing affects the lateral resolution for tapping (lift) modes. The way to determine the smallest
features that can be imaged using a particular tip in a particular operating mode is to optimize all of
the other factors that limit the lateral resolution and then experiment imaging small features on a
sample.
Assuming a small scan size and good tip conditions, the factor next most likely to limit the lateral
resolution is the resolution of the x-y detector, which is on the order of 1 nm. This limit applies only
if Closed Loop is on. Thus, the highest lateral resolution for small scan sizes is obtained with
Closed Loop off.
Note:
Finally, the digitized step size of the scanner limits the lateral resolution. The voltage applied to the
scanner is digitized, and the number of possible voltage values depends on the number of bits of the
digital-to-analog converter (the DAC) used to send the voltage signal to the scanner. Innova uses
20-bit DACs for sending the voltage signal to the scanner, so the voltage can be expressed as a
20-bit number, which has 220 possible values. The total range of motion of the scanner can
therefore be divided into 220 digitized steps. In low gain mode with a Large Area scanner, the
minimum step size of the scanner is 25 m/220 steps = ~0.025 nm/step. For a Large Area scanner
and a 20-bit digital-to-analog converter, the resolution as limited by the step size of the scanner is
always 0.025 nm -- substantially smaller than the length of a typical chemical bond.
The feedback controls and displays used to optimize operation of the Z feedback loop are listed in
the following table and are described in the later sections
Rev. B
13
Ch. 1 - Introduction
SPM Fundamentals
.
Control
Function
Z Position Bar
Feedback signal
Setpoint
PID Gains
Feedback Box
Cartoon Window
14
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
The bottom end of the Z Position bar represents the scanner's position when it is fully retracted.
The top end of the Z Piezo bar represents the scanner's position when it is fully extended. The
scanner tube retracts when the probe tip encounters peaks in the surface and extends when the tip
encounters valleys in the sample surface.
1.4.8 Setpoint: Setting the Reference Signal for the Feedback Loop
The setpoint scrollbox is used for specifying the setpoint, which is the reference signal for the
feedback loop that is maintained during an auto engage and a scan. For Contact mode, the setpoint
represents cantilever deflection, or force between the tip and the sample. For TappingMode, MFM
and EFM, the setpoint controls the amplitude of cantilever vibration. For STM, the setpoint
represents tunneling current. For LFM, the setpoint value also has an effect on the lateral force due
to friction. Increasing the setpoint value increases the force between the probe tip and sample. The
allowed range of values of the setpoint depend on the operating mode and the system calibration
parameter values.
In general, the setpoint value is increased (i.e., the value becomes more positive) the distance
between the probe and the sample decreases. In contact mode, increasing the setpoint value
decreases the tip-to-sample spacing in order to achieve a greater cantilever deflection, or vertical
force, between the probe and the sample. For STM, increasing the setpoint value, brings the probe
and the sample surface closer together to produce a higher tunneling current. During a scan, a
signal from the feedback loop is sent to the scanner, causing the scanner to retract or extend so that
Rev. B
15
Ch. 1 - Introduction
SPM Fundamentals
the feedback signal matches the setpoint value. For TappingMode, increasing the setpoint value
results in a higher amplitude and a lower amount of force applied to the sample.
To enter a setpoint value: Enter a new value in the setpoint scrollbox and then press the [Enter]
key. Or, use the scrollbox arrows to scroll through a range of values. The increment of the scroll
box can be changed by double clicking inside the scroll box.
Do not lower the gain to zero. When the gain is set to zero, the feedback loop is disabled, and
the system will not track changes in surface features. If the feedback is completely disabled,
the tip can be damaged if the sample surface is very rough. For STM, some finite feedback
response is needed to prevent the tip from crashing into the sample.
To adjust gains: Enter a value in the Gain scrollbox and then press the [Enter] key. Or, use the
scrollbox arrows to scroll through a range of values.
16
Rev. B
Ch. 1 - Introduction
SPM Fundamentals
Rev. B
Be careful when using the Z(m) scrollbox to manually extend the scanner. Extending the
scanner too far will cause the probe to crash into the sample surface. A probe crash can damage
the probe, the scanner, and the sample.
17
Ch. 1 - Introduction
SPM Fundamentals
18
Rev. B
Chapter 2
2.1
Advanced Imaging
Advanced Imaging
The standard configuration of Innova has the ability to produce nanoscale images using Contact
mode, TappingMode and STM (Scanning Tunneling Microscopy) mode for many sample types in a
typical laboratory environment. Lateral force microscopy and phase imaging are included in the
standard configuration and provide contrast mechanisms based upon material properties. These
capabilities give access to information beyond simple topography. Optional enhancements for the
Innova system provide additional capabilities including the capability of probing additional
material properties as well as enabling operation in a fluid and/or temperature controlled
environment to perform imaging of samples under relevant conditions (viz. biological samples,
electrochemical reactions, etc.).
This section provides brief descriptions of several advanced imaging modes and accessories,
however, development continues to produce additional methods and improve existing ones. To
order any of the options or to obtain information on these items or more recently developed
methods, contact your Veeco representative.
Rev. B
19
Advanced Imaging
Advanced Methods:
2.2
Advanced Methods:
BNC cable to connect the Thermal Control Unit to the NanoDrive Controller.
BNC-type cable to connect the Thermal Control Unit to the thermal probe
20
Rev. B
Advanced Imaging
Advanced Methods:
Amplifier mount
C-AFM cable
Rev. B
21
Advanced Imaging
Advanced Methods:
MicroCell
1ml syringe
Plastic tubing
Spare glass
22
Rev. B
Advanced Imaging
Advanced Methods:
TappingMode cartridge
Rev. B
23
Advanced Imaging
Advanced Methods:
24
Rev. B
Advanced Imaging
Advanced Methods:
Also needed to perform FMM (not included in package) are probes suitable for FMM
such as FESP.
LCSTM Cartridge
Rev. B
25
Advanced Imaging
Advanced Methods:
2.2.8
Description
SCM characterizes the capacitance properties of the sample and provides a means for 2D dopant
profiling. The images are useful in analyzing doped semiconductor materials where the dopant
distributions are not readily determined by other means. The SCM option offered with Innova
provides a fully integrated hardware and software solution which maximizes setup efficiency. The
option includes closed-loop SCM Imaging to ensure a constant depletion volume.
The Innova SCM package includes:
SCM sensor
SCM cartridge
SCM Cable
26
Rev. B
Advanced Imaging
Advanced Methods:
2.2.10 Electrochemistry
Description
The Innova Electrochemistry option permits in situ Contact mode or STM studies of surfaces in a
controlled electrochemical environment. The electrochemistry option combines an
electrochemistry cell and probe cartridge designed for Innova and uses the VersaSTAT-3
potentiostat from Ametek/Princeton Applied Research.
Scans and electrochemical measurements can be taken simultaneously. An electrochemical
measurement can be interrupted or altered without affecting the progress of the scan, and a scan can
be interrupted or altered without affecting the electrochemical measurements. The V3 Studio
electrochemical software provided with the potentiostat displays the data in real-time in a window
separate from the SPMLab screen. This window can be hidden in the background while still
controlling the potentiostat.
The following components are included in the Electrochemistry option:
VersaSTAT3-200
Rev. B
27
Advanced Imaging
Accessories and documents
drives the magnitude and sign of the applied DC bias. The implementation of surface potentiol
microscopy on Innova is exceptionally well integrated in hardware and software, yet exceptionally
flexible. Dual-frequency, single-pass potential measurements are enabled by the 2 full lock-in
amplifiers integrated in the control electronics. This approach minimizes topographic artifacts by
matching the electrostatic potential while in feedback. Additionally, Veecos patented Lift Mode
can be employed in a dual-pas, single-frequency approach to maximize the signal-to-noise ratio of
the potential measurement.
2.3
28
Rev. B
Ch. 3 - Safety
Operating Safety
Chapter 3
3.1
Safety
Operating Safety
This section includes important information about the Innova system. It describes procedures
related to the operating safety of Innova and must be read and understood before operating the
Innova system.
WARNING:
FUNCTION
Rev. B
29
Ch. 3 - Safety
Operating Safety
FUNCTION
Power off
DEFINITION
WARNING
CAUTION
Note
It is important to read all warnings, cautions, and notes. Warnings, cautions, and notes include
information that are important to the operating safety of Innova system.
30
Rev. B
Ch. 3 - Safety
Operating Safety
Innova must be properly grounded before applying power. The mains power
cord must be inserted into an outlet with a protective earth ground contact.
WARNING:
The line voltage selection must be checked before applying power to Innova
system components. The line voltage selector switch is on the front panel of the
NanoDrive beneath the front cover and protective clear shield. The line voltage
selector switches can be set to the following voltages: 100 V, 120 V, 220 V, and
240 V. See section 3.1.5.
WARNING:
Do not open Innova. The NanoDrive and the base unit use hazardous voltages
and therefore present serious electric shock hazards.
WARNING:
Periodically inspect the cables of the Innova system to ensure they are not
frayed, loose, or damaged. Cables that are frayed, loose, or damaged must be
reported immediately to a local Veeco service representative. Do not operate
Innova when wires are frayed, loose, or damaged.
WARNING:
CAUTION:
All Innova system components must be handled with care. System components contain
delicate electromechanical components that can be damaged easily by improper handling.
CAUTION:
The power to the NanoDrive must be turned OFF before removing or installing the scanner.
CAUTION:
When removing and installing the scanner, personnel and equipment must be grounded to
ensure that the scanner is not damaged. The scanner is sensitive to electrostatic discharge.
CAUTION:
The four screws that connect the scanner to the base unit must be securely fastened to ensure
proper grounding. When the four screws are securely fastened vibrations are reduced and
maximum instrument performance is ensured.
CAUTION:
To preserve safety and EMC compliance, Innova must be used with the EMI filter supplied
with the Innova system.
CAUTION:
Rev. B
31
Ch. 3 - Safety
Operating Safety
Ground
Connection
32
Rev. B
Ch. 3 - Safety
Laser Safety
4. Remove the clear shield cover protecting the line voltage selector switches.
Power On/Off
Selector switches (6)
3.2
Laser Safety
Innova contains a diode laser powered by a low voltage supply with a maximum output of 0.2 mW
CW in the wavelength range 600 to 700 nm. Diode laser power up to 0.2 mW at ~670 nm could be
accessible in the interior. Innova should always be operated with the probe head properly installed.
Rev. B
33
Ch. 3 - Safety
Laser Safety
WARNING:
Figure 3.2a shows the warning labels located on the Innova probe head. Strict observance of these
warning labels is required.
34
Rev. B
Ch. 3 - Safety
Specifications and Performance
Figure 3.2b shows the location of the laser safety compliance label on the rear panel of the
NanoDrive electronics module.
Figure 3.2b NanoDrive Rear Panel, Showing Location of Laser safety Compliance Label.
3.3
Probe Head
Measurement Performance
Standard
Scanner
Scan range
Rev. B
35
Ch. 3 - Safety
Specifications and Performance
Optional
Scanner
Scan range
Microscope Stage
Translation range
Sample size
Tip-sample approach
Optical microscope
6 mm x 6 mm.
~45 mm x ~45 mm x 18 mm (thick).
Automatic with 3 independent stepper motors.
Optional on-axis microscope with color video
monitor for probe tip and sample view.
5:1 zoom, ~250 m field of view.
Standard 10X objective.
System power
Innova
NanoDrive
15 in. (380 mm) x 14 in. (355 mm) x 14 (355 mm); 26 lb (12 kg).
14 in. (585 mm) x 23 in. (191 mm) x
23 in. (585 mm); 50 lb (23 kg).
17 in. (432 mm) x 7.5 in. (191 mm) x
17.5 in. (445 mm); 27 lb (12 kg).
Computer
Operating Environment
Temperature
Humidity
Cleaning Agents
Base unit
Probe head
AEM and computer
WARNING:
CAUTION:
36
Isopropyl alcohol.
Isopropyl alcohol.
Isopropyl alcohol.
To avoid risk of electric shock, do not clean Innova system components when
power is applied.
Do not use acetone to clean Innova system components. Acetone may damage important
safety warning labels.
Rev. B
Ch. 3 - Safety
Innova User Documentation
3.4
Rev. B
37
Ch. 3 - Safety
Innova User Documentation
38
Rev. B
Chapter 4
4.1
Overview:
This chapter describes how to prepare the Innova instrument to take an image in Contact mode,
which is used for imaging the height of a sample. The sample in these tutorials is the 10 m
calibration grating provided with the Large Area scanner. The sample has been selected because its
features are relatively easy to identify.
Further details and background information related to many of the procedures in this chapter are
provided in Chapter 7 and Chapter 8.
This chapter assumes familiarity with Chapter 1. Before following the procedures in this chapter,
also ensure that:
The computer components and control electronics, have been installed and set up
properly.
The cables are properly connected between the system components and the power cords
are plugged in.
WARNING:
4.2
Facilities
The basic requirements for facilities, including utilities and floor space requirements are listed in
this section. Innova is designed to operate acceptably in what would be considered a typical
laboratory environment in terms of temperature and humidity. A user supplied vibration isolation
table (VT-GB or equivalent) is required for Innova and critical applications may require additional
protection against mechanical or acoustic vibration (including air buffeting).
Rev. B
39
Vibration: Vibration of the AFM mounting surface should not exceed VC-D in any
direction.
100
1959-940-A
Workshop (ISO)
90
32,000
Office (ISO)
80
16,000
8,000
4,000
70
BBN Criterion A
2,000
BBN Criterion B
60
1,000
BBN Criterion C
50
500
BBN Criterion D
250
BBN Criterion E
125
40
6.3
4
5
16
10
8
12.5
40
25
20
31.5
63
50
100
80
160
125
40
Rev. B
Innova
Rev. B
41
4.3
Cable Connections
The cable connections described in this section should have been made by the authorized Veeco
representative who initially installed the system. Nevertheless, it is prudent to perform a check by
noting the following steps, which will also serve as familiarization with the system.
42
Rev. B
c. Connect the computer monitor cables to the computer monitor connectors on the back
panel of the computer unit.
3. Connect the Veeco supplied USB cable between the NanoDrive Controller and the back
panel of the PC.
4. Connect the 68 pin Low Voltage cable between the low voltage connector on the back panel
of the Innova microscope and the low voltage connector on the IO-I board of the NanoDrive
Controller. This cable carries low voltages signals (e.g. the setpoint) to the Innova
instrument.
5. Connect the 13 pin High Voltage cable between the high voltage connector on the back panel
of the Innova microscope and the high voltage connector on the IO-HV board of the
NanoDrive Controller. This cable carries high voltage signals that are applied to the
piezoelectric scanner.
6. Connect the 50-pin motor control cable between the motor control connector on the back
panel of the Innova instrument and the NanoDrive controller. The motor control cable carries
low-voltage signals to the stage.
7. Connect the video cable from the video BNC connection at the back panel of the Innova
microscope to the frame grabber card connector at the back of the PC.
Rev. B
43
4.4
CAUTION:
The scanner is extremely fragile. The piezoelectric ceramic tube is easily broken/damaged by
mechanical shock. Use extreme care when handling the scanner. Do not apply any pressure to
the sample holder on the top surface of the scanner since this may result in scanner tube
damage/breakage
CAUTION:
Ensure the head is not installed and the optics are moved out of the way before removing or
installing the scanner.
CAUTION:
The power to the NanoDrive must be turned OFF before removing or installing the scanner.
Sample holder above
Scanner tube
44
Rev. B
Scanner Mounting
Screws (4)
Do not allow the top of the scannerespecially the sample holderto come into contact with
another surface. If the scanner contacts another surface, the scanner tube may break.
9. After the electrical connector is disengaged, lift the scanner and tilt it to remove it from the
Innova stage.
10. Store the scanner on its side in the box originally supplied.
To install a scanner:
1. Insert the four hex-head screws into the scanner (as required).
2. Orient the scanner to align the scanner connector with the stage connector and insert the
scanner into the opening on the Innova stage. There is only one correct fit. The scanner
should fit snugly.
CAUTION:
Rev. B
The four screws that connect the scanner to the Innova base unit must be securely fastened to
ensure proper grounding. When the four screws are securely fastened, vibration is reduced and
imaging is improved.
45
3. Tighten the four hex-head screws with a 3/32" allen wrench until the scanner is firmly
seated.
Note:
4.5
Starting Innova
1. Turn the computer and monitors on. The computer on/off switch is located on the front panel
of the computer unit. The computer monitor on/off button is located on the front of the
monitor below the screen. Windows starts automatically and displays the desktop.
2. Turn on the NanoDrive. The on/off rocker switch is located on the lower left portion of the
front panel of the controller.
3. From the Windows START button, select All PROGRAMSVEECOSPMLAB XP
SPMLAB or double click the desktop icon for SPMLab.
Figure 4.5a Removing the Probe cartridge from the Probe Head.
46
Rev. B
4. Set the probe cartridge on a flat surface with the cantilever facing up.
5. Disconnect the electrical connector from the stage before removing the head.
Connector
Ensure that the head is properly seated on the three point mount.
2. Insert the electrical connector from the probe head into the connector on the stage, as shown
in Figure 4.5b.
4.6
Loading a Sample
1. Move the Innova Optics objective lens away from the probe head by rotating the swing arm.
Rev. B
47
CAUTION:
The objective lens must be out of the way before loading a sample. Otherwise, raising the
probe head will hit the objective lens, damaging both the probe head and the lens.
Lower probe
Simplified
48
Do not let the sample mounting disk snap down hard on the magnetic sample stub. The
piezoelectric scanner (mounted underneath the sample holder) is a ceramic material that can
easily break under mechanical shock.
Rev. B
5. Slide the mounting disk gently onto the sample holder (the small round stub attached to the
top of the scanner) as shown in Figure 4.6c.
6. Position the mounting disk so that the sample is centered on the sample holder. The sample
holder magnet holds the sample mounting disk securely in place.
Figure 4.6c Sliding the Sample Mounting Disk Onto the Sample Holder.
7. Use the down arrow in the Motor Stage window, refer to Figure 4.6b to lower the probe head
to within several millimeters of the sample surface.
8. Move the Innova Optics objective lens into position over the sample.
4.7
Rev. B
49
Chip Carrier
Spring clip
The spring tool should be kept in a secure place when not in use. If the tool is
lost, a charge will be made for a replacement.
1. Place the probe cartridge on a flat surface with the handling prongs facing to one side and the
spring clip facing up.
50
Rev. B
CAUTION:
Hold the chip carrier by the ceramic plate. Do not touch the microfabricated cantilever chip.
The cantilever chip is extremely fragile and can be damaged or broken easily.
Rev. B
51
8. Slide the chip carrier under the lip of the spring clip, as shown in Figure 4.7e, until the balls
engage the slots.
52
Rev. B
4.8
CAUTION:
Use only approved probe cartridges in the Innova head. An incorrect probe cartridge may
damage the instrument.
4.9
A stop inside the swing arm limits the arm rotation. Rotate the swing arm slowly to avoid
striking the probe head in the event that it or the head is not properly adjusted.
1. Rotate the Innova Optics swing arm slowly until the objective lens fits between the two arms
of the probe head.
2. The lens should be positioned directly over the cantilever and sample.
Rev. B
53
3. Start WinTV32 by Start > All Programs > Hauppauge WinTV > WinTV32.
54
Rev. B
5. Zoom out to the widest field of view by moving the camera slider control fully to the left and
adjust the light intensity by sliding the light intensity control to obtain the best possible
image.
Light intensity
control slider
Camera slider
Focus knobs
Coarse
Fine
Coarse focus moves the lens up and down. To avoid damaging the probe head and the lens, do
not drive the objective lens down into the probe head.
6. Adjust the coarse and fine focus adjustments to focus on the sample, refer to Figure 4.9c.
Rev. B
55
7. Monitor the focus adjustment by looking at the WinTV display. The cantilever chip with
triangular-shaped cantilevers should be seen.
56
Rev. B
Rev. B
57
Laser
Head
Head
Figure 4.10b Probe Head Alignment Controls and Laser Indicators.
1. Focus on the cantilever with the Innova Optics.
2. Adjust the laser spot to the end of the cantilever.
3. Center the laser spot on the end of the cantilever.
4. Adjust the QPD (Detector) alignment knobs to align the laser spot while watching the laser
position indicators on the front of the probe head (see Figure 4.10b). The central green LED
should be brightly illuminated and none of the red LEDs illuminated. It is normal to make
several adjustments, alternating between up/down and left/right to obtain proper alignment.
Near the correct position, the red lights are very sensitive, and some practice is needed to
make the adjustments so that all the red lights are off.
5 LEDs (center green)
58
Rev. B
Aligning the QPD may be especially difficult if the probe is defective. Residual strains
may cause a cantilever to bow and make alignment difficult. It may be advisable to
replace the probe.
If it is not possible to position the laser properly on the detector, insure the probe is
properly mounted or attempt aligning to the second cantilever.
The laser power is not on. -- Ensure that the head is properly connected to the stage and
check that the LASER ON/OFF toolbutton in the main window is clicked ON.
The laser spot doesnt come into view in the optics. -- It is important to ensure that the
laser spot is in the vicinity of the cantilever. Move the optics out of the way and adjust
the laser positioning controls until the spot can be seen near the cantilever.
It isnt clear what direction the steering knobs move the laser spot. -- Observe the
position of the spot on the sample surface and experiment with the knobs.
The brightness of the laser spot on the cantilever in the optical view is maximized, but
the laser intensity indicator does not illuminate. -- Do not try to maximize the brightness
of the laser spot in the optical view. The goal is to maximize the reflected light to the
QPD. When the spot is positioned so that the QPD is receiving the maximum reflected
light, the laser spot on the cantilever may not appear bright in the optical view.
The cartridge is not properly inserted in the probe head. -- Wiggle the cartridge in and
out to ensure that all three contact zones are engaged by the three balls on the cartridge.
The cartridge should feel firmly mounted when it is correctly positioned over the
contact zones.
The chip carrier is not properly inserted in the cartridge mount. -- Remove the cartridge
from the head and wiggle the chip carrier from side-to-side to ensure that all three balls
on the cartridge mount are engaged by the three slots on the chip carrier. The chip
carrier should click into place.
The cantilevers are broken. -- Examine the probe either visually or using the optical
microscope view to determine if the cantilever(s) are broken. Replace the probe as
required.
Rev. B
59
2. To start the SPMLab software, double-click the SPMLab startup icon on the computer
desktop. Select Yes when asked to load DSP Code, see figure Figure 4.12a. The SPMLab
software window appears and may span one or two monitor displays. This large window will
contain all the areas and panels needed to control the microscope and analyze results.
60
Rev. B
6. Select Yes when asked to Would you like to turn the HV on.
7. The main SPMLab window, Figure 4.12c displays the toolbuttons for all functions and
feedback controls in the left portion of the window.
Function toolbuttons
Engage Indicator
Feedback controls
Tip bias controls
Optical Microscope light and intensity
Camera zoom
Laser on/off
Individual sections may be turned on/off by right clicking on the left toolbar
which will open a selection window for the sections to be displayed.
Rev. B
61
62
Rev. B
The angle of each positioning mirror can be changed with two sets of alignment knobs shown in
Figure 4.13b.
Detector
(QPD)
Cantilever
(Laser)
Head
Head
Rev. B
63
(red)
(green)
(red)
(red)
(red)
Figure 4.13c Laser Position and Intensity Indicators.
64
Rev. B
The Laser Alignment window in the SPMLab program provides a visual representation of the laser
spot on the QPD. Open the Laser Alignment window either by clicking the
using the menu as shown in Figure 4.13d.
toolbutton or
Rev. B
65
Select to enable
distance scroll
66
Frequent or repeated use of Stage Reset may cause a safety shutdown due to controller
overheating. Stage motors will be disabled (several minutes) until the controller temperature is
restored to a safe range.
Rev. B
4.15 Engage
Approaching the sample is a two-part process. First, the motor stage controls (as described in
section 4.14) are used to perform a manual approach to move the probe close to the sample surface.
Then, auto engage is used to bring the probe to the distance required for imaging. When an auto
engage has succeeded, the probe is said to be in feedback at the surface. The greater the distance
between probe tip and sample, the longer the auto engage process will require.
The Engage dialog box is shown in Figure 4.15a in two configurations. The configuration can be
changed between Simplify and Extend by clicking the toolbutton in the lower right of the
window.
Stop engage/
Disengage
Autoengage
Simplify Window
Extend Window
Figure 4.15a The Engage Panel
Rev. B
The engage process involves adjusting the sewing voltage, the motor step (0.4 m min.)
and the delay time (delay between motor step and sewing range).
The settings for these parameters are specified in the Extend window, however, the
default values will generally produce good results.
The Z Center checkbox causes the piezo range to be set at midrange after an
autoengage has been completed and the probe is in feedback.
The piezo/tip interaction distance can be adjusted manually using the 2 step buttons (up
and down). A right mouse click opens a window where the desired Z adjustment can be
entered.
67
Clicking the down arrow in either window initiates an automatic approach/engage using
the settings which are in the Extend window.
To autoengage, stop the engage process or disengage, click the appropriate toolbutton as
indicated in Figure 4.15a. Right clicking on the disengage tool
where the retreat step distance may be entered.
68
opens a window
Rev. B
Chapter 5
5.1
Realtime Controls
5.2
Area Scanning
The area scan window allows setting of:
Rev. B
69
Scan rate: sets the number of lines scanned per second in the x direction. (min. = 0.1 Hz,
max = 100.0 Hz).
Scan range: size of the scan along one side of a square (min. = 0 m, max = 5.0 m for
5.0 scanner and 90 m for 90 m scanner).
Rotation: combines x and y drive voltages such that the piezo scans the sample at
variable x y angles (scanning angle 0 to 360).
X and Y offset: these controls allows entering an offset of x and y from the center of the
full xy scan range.
Automatic
Scaling
Power
Spectrum
Tool
Divide by 2
Multiply by 2
Figure 5.2a Profile and Power Spectrum
The profile and power spectrum display shows typical parameters for an area scan of a 10.0 m
grid. The profile shown on the figure depicts a height vs. scan range plot for the current line being
scanned. The vertical axis can be scaled automatically by selecting the A button or manually set by
highlighting either the X2 or divide /2 button. The power spectrum which describes how the power
of the signal is distributed with frequency can be displayed by highlighting the power spectrum
button. The scales for the power spectrum display can be adjusted with the tools on the right hand
side of the screen (log, auto, x2 or divided by 2).
70
Rev. B
5.3
Line Scanning
Any line scan can be obtained from a scanned image. Line scans can be obtained as follows:
1. Highlight the scan area icon on top of the scanning window.
2. Drag the image from the image window into the scan area window in the scanning window.
Figure 5.3a
Rev. B
71
Drawn line
72
Rev. B
4. Level the Line Scan signal by selecting a leveling option (AC only, 1D 1st order or 1D 2nd
order) And press start in the main scanning window.
Leveling
Options
Start scan
Figure 5.3c
5.4
Scan Leveled
Rev. B
73
Zoom Toolbutton
Toolbuttons
74
Rev. B
3. Right click in the center of the zoom square to display the zoomed image of the selected area.
Rev. B
75
7. After selecting the Distance Measurement toolbutton, click and drag a measurement line on
the scanned image and the distance will be displayed as shown in Figure 5.4c.
Figure 5.4c Distance Measurement Line Drawn with Click and Drag
8. The Angle Measurement tool
image.
76
Rev. B
9. After selecting the Angle Measurement tool, click and drag a line for one side of the angle,
then right click and drag the other side of the angle to display the angle as shown in Figure
5.4d.
11. After selecting the Cross Section toolbutton, click and drag to define a line on the image for
cross sectioning.
12. When the line has been drawn, the Cross Section window will open and display the sample
height along the line.
Rev. B
77
13. Clicking on the cross section plot at points of interest (see the blue triangular markers in
Figure 5.4e) will display information about the markers and their relative positions.
5.5
Probe Positioning
1. Select the probe position icon Move the cursor to the desired probe position. The
coordinates will appear on the left hand side of the scanning window.
Probe position
Figure 5.5a
Probe Position
78
Rev. B
5.6
LiftMode Settings
The scanning window contains real time controls for the LiftMode. The LiftMode is used in MFM
and EFM and will be described in detail in Chapter 11 and Chapter 12.
1.
Rev. B
79
3. Figure 5.6b shows the LiftMode window for EFM. The parameters in the LiftMode section
are the same as used with MFM. Select the Double Bias tab, select Enabled and choose to
apply a voltage bias to either the tip or the sample and enter the appropriate bias voltages.
80
Rev. B
5.7
5.8
Rev. B
Resolution (number of data points, the minimum is 16 pixels, maximum 1024 pixels).
Thresholds
81
5.9
IV Curves
1. The signal current to voltage curve (bias line on tip or sample) is generated in the Point
Spectroscopy window. The following inputs may be made but the default values will usually
produce good results:
Resolution (number of data points, the minimum is 16 pixels, maximum 1024 pixels)
Z Start/End range
2. The scaling for the graphs displayed on the right are adjusted by selecting full, auto or
manual on the left and right signal scales.
82
Rev. B
Rev. B
83
2. Open the Probe Positioning window by clicking the Probe position toolbutton
to add
4. The selected probe positions will be shown in the image and listed by coordinates in the
probe location window.
Coordinates of selected points
84
Rev. B
Ch. 6 - Menus
File Menu
Chapter 6
Menus
The menu bar contains a list of menus. Menus contain menu items that provide
access to instrument controls. The menus and menu items of SPMlab are depicted
in the figures below.
6.1
File Menu
Menu item
Rev. B
Function
Save data
Exit
85
Ch. 6 - Menus
Setup Menu
6.2
Setup Menu
Menu Item
86
Function
Microscope
Laser
Tapping
Stage motors
Opens the stage motors window and allows moving the head
assembly
Closed Loop
Engage
Recalibration
Rev. B
Ch. 6 - Menus
Real Time Control Menu
6.3
Menu Item
Rev. B
Function
Scan Control
Opens the scanning window. Specifications include: the imaging channels, input scan parameters, activate LiftMode, activate
bias voltages for tip bias and sample bias, overscanning, closed
loop on/of, high/low gain HV and X/Y axis moves and control
scanning conditions.
Point
Spectroscopy
Signal
Tracing
Probe
positioning
Multimeter
Oscilloscope
87
Ch. 6 - Menus
Tools Menu
6.4
Tools Menu
Menu Item
88
Function
Image
Analysis
NanoPlot
Rev. B
Ch. 6 - Menus
Window Menu
6.5
Window Menu
Menu Item
Rev. B
Function
Cascade
Tile
Horizontally
Tile
Vertically
89
Ch. 6 - Menus
Toolbuttons
Menu Item
6.6
Function
About
System info
Toolbuttons
90
Rev. B
Ch. 6 - Menus
Other Controls
6.7
Other Controls
Engage indicator
Feedback signal
Feedback on/off
Piezo position indicator
Gains
{
Bias voltage controls
Rev. B
91
Ch. 6 - Menus
Other Controls
Right clicking on a section of the Main Window controls will open a window which permits
selection of which control sections will be displayed or hidden.
parameter window.
92
Rev. B
Chapter 7
7.1
Overview
This chapter assumes setup of Innova as described in Chapter 4. Further details and background
information related to many of the procedures in this chapter are provided in other chapters.
The Innova system is comprised of the piezo scanner, Optics and the AFM detection system and the
NanoDrive controller. The scanner houses the piezoelectric transducer. The piezo element
physically moves the sample in the X, Y and Z direction. The force detection system consists of a
laser which generates a spot of light that is reflected off of a microfabricated cantilever onto a
mirror and finally into a photodetector. The position of the spot is determined by circuitry which
generates a voltage from the difference between the photodiode segments. The circuit outputs a
voltage ranging from +10V to -10V depending upon the position of the spot on the photodiodes.
The Innova system maintains the tip at the end of the cantilever in contact with the sample surface.
The sample is scanned under the tip in X and Y. Features on the sample surface deflect the
cantilever, which changes the position of the laser spot on the photodiodes. This position change is
read by the feedback loop. The feedback loop moves the sample in Z to restore the spot to its
original position.
WARNING:
7.2
Startup
Rev. B
93
94
Rev. B
MICROSCOPE: Innova
Mode: Contact
SELECT SCANNER: Select the menu item for the scanner that is currently installed
(Large or Small Area).
6. Click
Rev. B
95
7.3
in = On
out = Off
Figure 7.3a Laser On/Off Toolbutton
96
Rev. B
Do not lower the probe head too far. If the probe tip hits the sample surface, both the probe tip
and the sample will be damaged.
7. Open the Motor Stage window by clicking on the Motor Stage toolbutton
and use the
motor stage buttons to lower the probe until the cantilever begins to come into focus in the
optical view. At this point, the cantilever will be close to, but not touching, the sample
surface. To avoid damage to the tip (a crash), use Slow when the tip is near the sample.
Rev. B
97
8. Choose a setpoint for the deflection feedback. The setpoint must be a larger positive value
than the force deflection. If the laser has been aligned in the center, corresponding to a force
deflection of approx. zero, a setpoint of +2.0 V may be appropriate. See Figure 7.3d.
9. Perform Auto engage as described in the following section
7.3.2 Engaging
CAUTION:
Once the tip and sample are in contact, raise the probe head before turning off the power to the
probe head. When the system is engaged, the Scan window (see Figure 7.3d) will appear.
A successful engage process will bring the probe into contact with the sample surface.
1. Ensure the auto engage mode is activated and click the Engage toolbutton
2. During the engage process, the Z position bar (which represents Z piezo movement) will be
sewing and the cantilever icon will indicate Engaging.
3. Use the default P, I and D gains in the feedback control window (see Figure 7.3d) these
values may be optimized at a later time as described in Adjusting Feedback Parameters:
Section 7.5.3.
98
Rev. B
4. The engage process is completed when the feedback signal bar shows the setpoint value and
the Z position piezo bar is near mid range. The cantilever icon will indicate Engaged and
the scanning window will be open.
Improperly set Size and Gain parameters could result in damage to the tip or sample.
Rev. B
99
7.4
100
Rev. B
Rev. B
101
7.5
SCAN RATE
1. Position the cursor over the number in the Rate textbox and double-click to highlight the old
text. Alternatively, click and drag the cursor to select the number.
2. Type in a new scan rate in Hertz (= scan lines per second), and then press the [Enter] key.
3. A new scan will start automatically.
The scan rate can be adjusted while an image is being acquired. Image quality can be
monitored as the scan rate is adjusted. When the optimum scan rate is determined, take a
new image using that scan rate.
102
Rev. B
LEVELING
A tilted signal trace on the profile display indicates that the sample is tilted. The following
steps describe how to adjust for this tilt.
1. Select the image window containing the image to be leveled.
2. Click the scroll box and select the leveling type
3. The available leveling types are:
None
4. Stop when oscillations begin appearing superimposed on the signal trace of the grating in the
Oscilloscope Display.
5. Lower the gain incrementally by clicking the scroll down tool
until the oscillations
disappear. The goal is to be close to, but not beyond, the point at which oscillation appears.
The proportional gain value is now optimized.
6. Repeat this procedure for integral gain. Usually integral gain changes produce the greatest
effect.
7. Derivative gain setting may be experimented with to attempt to improve the trace.
Rev. B
103
SETPOINT
1. Position the cursor over the number in the setpoint scrollbox, and double-click to highlight the
old value. Alternatively, click and drag the cursor to select the number.
2. Type in a new setpoint value, and then press the [Enter] key. Alternatively, use the
scrollbox arrows to scroll through the range of setpoint values.
3. Click the start scan toolbutton to begin a new scan.
4. Adjust the setpoint while a scanning to see the immediate effect.
Continuous
Scan
Capture tool
Continuous
Capture
Scan Area
Button
Resolution
Scrollbox
104
Rev. B
CAPTURING IMAGES
1. All images are saved to the computer hard drive as long as the capture tool (see Figure 7.5a) is
enabled.
2. To perform continuous scans, enable the Continuous Scan toolbutton (Figure 7.5a).
3. To save all images continuously, enable the Capture Tool and Continuous Capture tool
(Figure 7.5a).
4. Alternatively, a single frame scan can be saved after it is complete by the menu sequence:
File > Save.
7.6
LFM Imaging
Lateral Force Microscopy (LFM) is an extension of Contact mode that can provide additional
information about the sample surface by measuring the torsional deformation of the cantilever
(represented by the LFM signal) as well as the vertical bending (represented by the deflection
signal).
After completing performing Contact mode imaging, LFM images may be taken without
interrupting the imaging session. Since the LFM signal is always available during Contact mode
operation, monitor it by the following:
Rev. B
105
WARNING:
Observe all Warning and Caution statements in Chapters 1-4 when using the
Innova system.
This section assumes familiarity with taking Contact mode images with Innova and that the
system is installed and set up.
, to open the
5. Configure the system software by making the following selections in the SPMLab System
Configuration dialog box.
MICROSCOPE: Innova
MODE: Contact
SELECT SCANNER: Select the file that has the scanner calibration values for the scanner
that is currently installed.
6. Click
106
Apply
Rev. B
Rev. B
107
108
Rev. B
Figure 7.6e Cantilever Response to Change In Height with Corresponding Topography and LFM Signal Traces.
Figure 7.6ec shows the LFM signal trace that would result from the height of Figure 7.6ea. As the
figure shows, the LFM data reflects only the lateral components of bending (e.g., bending to the
right at the rise in height produces a positive signal and bending to the left at the drop in height
produces a negative signal).
Figure 7.6f illustrates how a cantilever responds to changes in frictional coefficients and how that
response correlates with the resulting LFM and height data.
Figure 7.6fa shows a change in frictional coefficient that causes the cantilever to bend to the right
for a scan that is taken from left to right. If the scan is taken from right to left, as illustrated in
Figure 7.6fb, the cantilever bends to the left as it passes over the change in frictional coefficient. A
change in height causes the same type of cantilever bending illustrated in Figure 7.6e.
Figure 7.6fc shows a topography signal trace resulting from the surface of Figure 7.6fa: the data
only reflect the change in sample height. Figure 7.6fd shows the LFM signal trace that would result
from a scan taken from left to right. Figure 7.6fe shows the LFM signal trace that would result from
a scan taken from right to left. The sign of the LFM signal flips for the change in friction, but not
Rev. B
109
for the change in height. Changes in height appear on an LFM image as adjacent dark/bright
regions. By identifying these adjacent dark/bright regions, and by viewing data from two scan
directions, a user looking at an LFM image can distinguish between contrast changes due to
changes in frictional coefficient and those due to changes in height. Image subtraction, available in
the SPMLab analysis software, can be used to reduce topographic effects in LFM images.
Figure 7.6f Cantilever Response to Change in Frictional Coefficient and Height, and Corresponding Contact
Mode (AFM) and LFM Signal Traces.
Side-by-side height and LFM data therefore provide complementary information. By monitoring
the LFM signal, the contribution of lateral cantilever bending to a Contact mode image can be
inferred. Conversely, having the Contact mode information available confirms that contrast
changes on an LFM image that are due to changes in height, rather than frictional coefficient.
When scanning, the fast scan direction to be selected to be x or y (horizontal or vertical). As an
image is being taken, the scanner rasters back and forth over each scan line in the fast scan
direction, and then advances to the next line in the slow scan direction. To maximize LFM signals,
choose the scan angle so that the fast scan direction is perpendicular to the cantilever axis (scan
angle 0 or 180 degrees. Once a fast scan direction has been selected, forward or reverse sweep data
can be selected for viewing. For example, if fast scanning is x scan, data collected from the
110
Rev. B
right-to-left sweep of the scanner, or from the left-to-right sweep of the scanner, or from both
sweep directions may be selected for viewing.
Fast Scan Direction
Rev. B
111
112
Rev. B
Chapter 8
8.1
TappingMode Imaging
Overview
This chapter describes an example of a TappingMode imaging session. Briefly, TappingMode
imaging is performed by monitoring the oscillation amplitude and phase of an oscillating cantilever
probe. The probe to sample distance is set so that the probe tip lightly contacts (taps) the sample
surface during the cantilever oscillation. As the probe interacts with the surface of a sample, the
oscillation amplitude or phase relation to the exciting signal changes in response to the interaction.
The interaction provides the basis for producing an image of the sample.Throughout this chapter,
many of the settings used are default values of the program. Experienced users may wish to
override the defaults based on their knowledge of the sample and imaging requirements. In
general, the default settings are acceptable starting points for most applications.
One advantage of TappingMode AFM is reduction of frictional forces which exert torque on the
cantilever. Unlike traditional Contact mode which maintains a constant cantilever deflection, the
TappingMode feedback loop keeps the cantilever oscillating at a constant amplitude. The tip on the
cantilever is oscillated at a frequency near or at the cantilever resonance. A laser beam is reflected
off a microfabricated cantilever, onto a mirror, then onto a photodiode. As the cantilever vibrates,
the laser spot oscillates across the photodiode and produces an AC voltage. The signal from the
photodiode is rectified, then lowpass filtered into a DC voltage (measuring the RMS Amplitude).
The RMS amplitude is proportional to the amount of cantilever motion.
The feedback system compares the RMS amplitude to the setpoint voltage. The two voltages are
kept equal by controlling the amplitude of cantilever movement. The sample surface is in close
proximity to the cantilever such that the tip touches the surface only at the lowest point of its
oscillation. The RMS voltage is reduced to the setpoint voltage by the feedback loop moving the
sample nearer to the tip. The sample reduces the cantilever movement until the desired RMS
voltage is reached. The oscillation amplitude of the cantilever is held constant by the piezo moving
the sample in Z as it is simultaneously translated in X and Y.
Engagement in TappingMode AFM requires that the setpoint voltage be smaller than the RMS
voltage when the probe to sample distance is large (far from engaged). The tip to sample spacing is
reduced until the RMS amplitude is at the setpoint.
Rev. B
113
8.2
Startup
1. Set up with the appropriate probe and cartridge for TappingMode operation. Refer to Chapter
4.
2. With the system properly installed and turned on, launch SPMLab by double clicking the
desktop icon and select Yes when asked to load the DSP code.
114
Rev. B
5. Laser alignment is described in Aligning the Deflection Sensor: Section 4.10 and Deflection
Sensor: Section 4.13 in this manual. Ensure the alignment is proper before proceeding with
this exercise.
6. Click on the Laser Alignment toolbutton
to open the Laser Alignment window and
check the alignment of the deflection sensor by looking at the laser position and intensity
indicators on the probe head.
Note:
7. A representation of the laser spot will be displayed. The spot should be centered on the cross
hairs using mechanical controls on the SPM.
8.3
Rev. B
115
2. Set the Range setting to 0 - 600 kHz or select an appropriate range for the probe in use (the
resonance of TESP probes is near 300 kHz so 600kHz is appropriate) and click the
Start/Stop Frequency Sweep
toolbutton.
116
Rev. B
Rev. B
117
Note:
The zoom box must be drawn from Northwest (10 oclock) to Southeast (4
oclock) in order to zoom. If drawn in other directions, the result is an
unzoom to the original plot.
118
Rev. B
5. When the mouse button is released, the zoom box expands and the zoomed plot is shown.
Rev. B
119
7. Continue to zoom until the peak response is displayed over some width on the plot.
120
Rev. B
9. Position the mouse cursor at the peak of the amplitude curve and click to select the peak
frequency.
Rev. B
121
122
Phase adjustment is not required for basic tapping mode imaging because the
feedback loop utilizes amplitude, not phase. Phase adjustment is, however,
required to obtain the desired contrast in Phase Imaging, that is, when
selecting the Phase as a channel to be recorded.
Rev. B
8.4
Autotune
Autotune automates much of the cantilever tuning process. It automatically selects the cantilever
tune frequency based on the maximum response within a specified frequency sweep range. If
selected, autotune also adjusts drive amplitude and phase. The results of autotune, based on user
specifications usually produces good results and is easier than the manual tune process previously
described (see 8.3). The procedure for autotuning is:
Note:
Rev. B
Throughout this procedure for autotune, accepting the default values in any
window will generally provide good results. The example screens are for
illustrative purposes.
123
1. Select the frequency range for the autotune sweep, refer to Figure 8.4a (the maximum
response will be found within this range).
124
Rev. B
2. Drive amplitude for cantilever oscillation can be set in either of two ways. The Drive
Amplitude value may be scrolled up/down as shown in Figure 8.4b
Rev. B
125
Figure 8.4c. When a value is entered for the Target Tapping Signal and autotune is
performed, any value in the Drive Amplitude box will be overridden.
126
Rev. B
4. Autotuning is performed by clicking the Scan toolbutton and will produce a display similar
to that of Figure 8.4d.
8.5
Engage
Engage is the process of moving the sample and tip closer to each other until a desired degree of
interaction occurs. The Setpoint defines whether or not the SPM is engaged/operating in
TappingMode. When the cantilever is oscillating freely, a characteristic ac voltage results. As the
cantilever nears a sample and begins to interact with it, the oscillation amplitude (and the
corresponding ac voltage) is reduced. In this illustrated example, the free oscillation corresponds to
5.75 volts and 3.5 volts defines the value at which the tip is engaged with the sample and which
feedback will attempt to maintain as the sample height changes. A setpoint of greater than 5.75
volts would imply that the tip is held far enough away from the sample that there is no interaction
(therefore no imaging). A setpoint of 0 volts would occur when the cantilever is not oscillating
because it is in solid contact with the sample. Lower setpoints imply higher contact forces between
tip and sample with consequently higher probability of damage to the tip or sample. The suggested
setpoint value of 60-70% of the free amplitude is a rough starting value and depends upon the
sample.
Rev. B
127
toolbutton
128
Rev. B
Rev. B
129
8.6
Scanning Windows
1. The channel windows display images based on different criteria and many channels are
available. The available channels can be viewed by opening the Channel window.
130
Rev. B
3. For ease in viewing, drag the Scanning window onto the second monitor screen.
Rev. B
131
132
Rev. B
Start/Stop
Scanning
Height selected
Rev. B
133
drop down menus and 1D Line Fit leveling specified, which is a common choice, especially
for height data.
134
Rev. B
Chapter 9
STM Imaging
9.1 Overview
This chapter describes STM imaging and includes information on STM tips used to take STM
images. The instructions in this chapter assume familiarity with setting up the system and taking an
image.
WARNING:
Follow all Warning and Caution statements in Users Guide when using the
Innova system.
9.2
Rev. B
135
STM tips are very sharp. Be careful when handling an STM cartridge with a tip
loaded. Avoid leaving the STM cartridge on a table or other work surface with
the exposed tip pointing up. Store the STM cartridge and tip in a container with a
lid.
Cut the wire by using the wire cutters to pull and twist the end of the wire while snipping.
136
Rev. B
The resulting tip may not appear as sharp as it is. Tungsten STM tips oxidize fairly quickly and
should be discarded after 1 to 2 days. Platinum iridium STM tips, on the other hand, do not readily
oxidize and may be kept and used for a much longer time before degradation affects image quality.
The overall shape of tips made using wire cutters is not well-defined. STM images taken using
these relatively blunt tips can show multiple tip imaging effects. Sharper, higher aspect ratio
tungsten tips can be made using a tip etcher.
Front
Back
Rev. B
137
On the top side of the cartridge, a round metal contact pad delivers the tunneling current signal to
the control electronics.
CAUTION:
The tip is tilted relative to the sample instead of pointing straight down. The tilt allows seeing the
tip using the on-axis optical view.
The STM cartridge is installed in the probe head in the same way as the AFM cartridge, with the tip
pointing down. The instructions below explain how to insert STM tips into the STM cartridge.
WARNING:
STM tips are very sharp. Be careful when handling an STM cartridge with a tip
loaded. Avoid leaving the STM cartridge on a table or other work surface with
the exposed tip pointing up. Store the STM cartridge and tip in a container with a
lid.
Use a tip that has been made using wire cutters or an etched tip. Insert the back end of the tip first.
Note:
It is not necessary to be able to see the tip using the on-axis optical view.
However, to see the tip in the on-axis view, have the sharp end of the tip
extend from the STM cartridge approx. as far as the AFM cantilever chip on
the AFM cartridge. If the tip is too long or too short, locating the tip using the
on-axis optical view may be difficult.
138
Rev. B
5. Check that the tip length is about right by comparing the STM cartridge with an AFM
cartridge that has a cantilever chip loaded. The STM tip should extend out about as far as the
AFM cantilever chip does on the AFM cartridge. It is usually easier to see the tip in the
on-axis optical view if the vertical distance between the tip and the underside of the cartridge
is between 2 and 3 mm.
Be careful to cut the back end of the wire short enough to avoid scratching the objective lens.
Rev. B
139
9.3
9.3.1 Startup
1. Install a small area scanner as described in Chapter 4. STM studies usually are concerned
with atomic resolution. Although the large area scanner can also achieve atomic resolution,
the small area scanner is recommended.
2. Click on the Microscope Set-up tool
scanner is specified.
140
Rev. B
3. Select Scanning Window and inspect Conditions settings to insure that HV gains are set
appropriately. High is recommended for the Small area scanner; Low is strongly
recommended for the large area scanner.
Rev. B
141
9.3.2 Troubleshooting
Check the electrical resistance of the path between the STM tip and the probe head using a
multimeter on the Ohms setting.
1. Place one probe of the multimeter on the right brass screw of the two brass screws on the
probe head which are positioned directly above the probe cartridge, which connects to the
metal contact pad of the STM cartridge.
2. Place the other probe of the multimeter on the metal cylinder of the STM cartridge that holds
the STM tip wire.
The resistance between the probe head and the tip should be small for STM operation, typically
~0.2 .
If the STM cartridge is too loose or too tight, the conducting path may be disrupted. Try adjusting
the position of the probe cartridge to establish good contact. If this does not help, call Veeco
Customer Support for help in diagnosing the problem.
142
Rev. B
Setpoint is 1.0000 nA
Rev. B
143
To bring the tip close to the sample using the optical view. The following steps refer toFigure 9.3e:
9.3.4 Engaging
If desired, open a multimeter window and set it to display AL3-STM signal. This display allows
viewing the STM current signal in raw (volts) units.
144
Rev. B
Note:
It is important to select Fine in the engage dialog window because the STM
current generally becomes immeasurably small within a few nm of the
surface. For this reason, tip sewing is needed throughout the engage process.
Click the Engage tool to begin an auto approach. In STM mode, this initiates the following
sequence of steps:
The scanner extends, moving the sample toward the tip with feedback enabled.
The system monitors the tunneling current, checking for the setpoint value as the
scanner extends.
If the setpoint value is not obtained with the scanner fully extended, then the system
retracts the scanner and lowers the probe head one step (of the stepper motor) toward the
sample.
The above steps are repeated until the tunneling current matches the setpoint value.
The engage sequence is a slow process. The reflection of the tip may flicker in the optical view with
each step. The Current signal in the multimeter window display should approach the setpoint value,
and then stop.
When the approach stops, the tip will be within 1 nm of the surface but will not actually be in
contact. After a successful engage, the green Piezo bar should show that the scanner has stopped
moving and is extended to about half of its full range.
Rev. B
Note:
The tip should never touch the surface in STM mode. If the tip makes contact,
both the tip and the sample will be damaged. Contact between the tip and the
sample, referred to as a tip crash, is indicated by an increased current reading
on the multimeter display. If the tip crashes, it must be changed before
proceeding.
Note:
145
When the engage has been successful, the display should appear similar to:
146
Rev. B
Since the tunneling current varies exponentially with tip-to-sample spacing, the setpoint current
basically controls the tip-to-sample spacing during a scan. Raising the setpoint value brings the tip
closer to the sample, while lowering the setpoint value moves the tip farther from the surface.
Ideally, the tip should never come into contact with the surface during an STM scan, as both the tip
and the sample surface will be damaged; e.g., the tip can leave a small pinhole or dent in the surface
if it is driven into the surface.
Typically, the setpoint is less than 5 nanoamps for STM.
The tip-to-sample bias during a scan is set using the sample bias and tip bias parameters. If the
sample is biased negative relative to the tip, then the STM image will represent tunneling from
filled electronic states on the sample surface. If the sample is biased positive relative to the tip, then
the STM image will represent tunneling into empty electronic states on the sample surface. The tip
and sample biases are given in volts.
The bias settings have an indirect effect on the tip-to-sample spacing. In constant-current mode
with feedback optimized, the system attempts to maintain a constant tunneling current by varying
the tip-to-sample spacing. If the tip-to-sample bias is increased, the tunneling current also
increases. The system therefore pulls the sample away from the tip to maintain constant tunneling
current.
Typically, the bias between the tip and the sample should be lower than about 2 V in air. If the bias
is larger than about 2 V, the STM will not be operating in the tunneling regime. The optimal bias to
use depends on whether the sample is conducting or semiconducting. With conducting samples, a
lower bias setting may be used. With semiconducting samples, use higher bias settings, but less
than 2 volts. A typical bias range for STM is from about 0.1 to 2 V.
Figure 9.3h shows typical settings for an STM scan using the small area scanner on HOPG.
Figure 9.3h Typical Scan Settings for Atomic Resolution Imaging Using STM or Contact Mode AFM
Rev. B
147
Figure 9.3i STM Image Showing Atomic Resolution on HOPG. 4nm x 4nm Image Size, 100mV Bias, 1nA
Current Setpoint.
9.4
148
Rev. B
10.2 Startup
1. Double click the SPMLab icon to launch the program.
2. If NanoDrive is not turned on, the following message will appear.
Rev. B
149
4. The system configuration panel will appear. The panel will typically display the previous
setting and should be changed to Contact mode and the Apply toolbutton clicked.
150
Rev. B
6. Start WinTV32 by Start > All Programs > Hauppauge WinTV > WinTV32.
Rev. B
151
Laser
Positioning
Head Position
Left/Right
Head Position
Front/Back
Figure 10.3a Adjustments on Innova Head
152
Rev. B
a
c
b
10.4 Engage
1. Use the focusing adjustment on Innova to focus onto the surface of the sample, then use the
Motors Stage controls to lower the cantilever until it begins to come into focus.
Rev. B
153
3. Select Z Adjustment in the engage window and click on the Engage toolbutton
Engage
toolbutton
154
Rev. B
2. Drag the Scanning window and reposition the WinTV window to suit personal preference.
To display the image windows simultaneously, use Windows > Tile Horizontally.
Rev. B
155
Note:
If values outside the permissible/possible range are entered for Z start and Z
end, the program will correct the values to acceptable ones.
10.6 Ramp
1. Begin ramping by clicking the Start/Stop Scanning toolbutton
156
Rev. B
2. To obtain the standard trace in the figure, select the Z Inverse option
3. The response of piezo positioning devices is inherently non-linear. The non-linear response
can be characterized and compensated for in hardware and software. This scanner
linearization feature is activated by enabling the Z piezo linearizer (= closed-loop Z mode).
Rev. B
157
4. Since the probe has usually been brought into feedback before ramping, the Z-position
maintained during feedback is usually a useful reference point. The relative Z-position
refers the Z-start and Z-end positioning to the feedback point instead of referring it to zero
volts of the Z-DAC drive voltage
3. Open the Probe Positioning window and click on the scanned image in the Scanning
window at the locations to measure force distance curves. As the locations are selected, the
coordinates of the locations will appear in the Probe Positioning window.
Location Coordinates
Selected locations
158
Rev. B
4. Open the Point Spectroscopy window and enable Merged to Probe Position.
Rev. B
159
6. When the force distance curves have been produced, save the data by clicking the Save
toolbutton in the Probe Positioning window. This will open a typical windows to specify
the file name and location for the saved data.
7. The saved data file is a text file (with the default file extension: .pos) which may be used
for analysis.
8. Use the Table toolbutton
to display the force curve data as a table. Clicking on a
column heading selects a force curve for display when the probe positioning tool has been
used to acquire a series of curves which results in a table with multiple columns.
160
Rev. B
Rev. B
MESP probes
161
11.3 Startup
1. Launch the SPMLab program by double clicking the program icon. The System Status
window will appear.
162
Rev. B
3. When the appropriate selections have been made, click Apply and the Turn HV on window
will appear.
Some procedure details have been omitted and are described in the MFM
support note.
6. The Laser Alignment display provides a representation of the laser spot location on the QPD
(quad photo detector). Make adjustments on Innova as required to center the spot.
Rev. B
163
Cantilever
Tuning
164
Rev. B
Start/Stop
Frequency
Sweep
Pan/Zoom option
165
4. Continue to draw selection boxes until the peak exhibits some width and click on the
Start/Stop Frequency Sweep toolbutton to produce a new scan within the range of interest.
5. Select the Set Frequency mode and move the cursor to the peak of the amplitude curve.
This will display the value of the amplitude as well as the frequency indicated by the cursor.
In this example, the amplitude is 6.695 volts at a frequency of 67.635 kHz.
Set Frequency
Adjust phase
166
Rev. B
Start/stop
scan
Rev. B
167
11.5 Engage
1. Using the adjustments on Innova, adjust the focus so that the WinTV2000 image shows the
sample surface to be in focus.
Note:
Most surfaces being scanned in MFM are highly polished and it may be
difficult to focus on the surface. Usually slight scratches, blemishes or
contaminants exist to assist focusing.
2. Use the Motor stage controls to lower the probe until it begins to come into focus and select
the engage toolbutton.
Engage toolbutton
168
Rev. B
3. In the Engage window, set the engage mode to Auto, select Z center and click the engage
toolbutton.
Engage
Auto
Z Center
Rev. B
169
Note:
The start height parameter is usually not needed. It allows a higher initial lift
height when needed to overcome adhesion on sticky samples.
LiftMode
Enabled
Start/Stop Scan
Select
Channels
170
Rev. B
8. Tile channel windows and click Start/Stop Scan toolbutton to obtain MFM images.
Rev. B
171
172
Rev. B
Rev. B
173
EFM sample
SCM-PIT probes
Probes: SCM-PIT
Carrier with
Attached Cable
Figure 12.2a Carrier Connection for EFM Plugs Into Black Socket at Front of Microscope Base
12.3 Startup
This section describes starting the program and making initial adjustments to Innova.
1. Launch the SPMLab program by double clicking the program icon. The System Status
window will appear.
174
Rev. B
2. Load the DSP code by clicking the Yes toolbutton and select the correct microscope
configuration (Tapping).
Rev. B
175
5. The Laser Alignment display provides a representation of the laser spot location on the QPD
(quad photo detector). Make adjustments on Innova as required to center the spot.
Cantilever
Tune
176
Rev. B
Start/Stop
Frequency
Sweep
Rev. B
177
Note:
Pan/Zoom Option
Figure 12.4c Pan/Zoom on Low Frequency Peak
178
Rev. B
3. Continue to draw selection boxes until the peak exhibits some width and click on the
Start/Stop Frequency Sweep toolbutton to produce a new scan within the range of interest.
The new scan will appear approximately as shown
Rev. B
179
4. Select the Set Frequency mode and move the cursor to the peak of the amplitude curve.
This will display the value of the amplitude as well as the frequency indicated by the cursor.
In this example, the amplitude is 6.695 volts at a frequency of 67.635 kHz.
Set Frequency
180
Rev. B
Adjust the phase setting and re-scan as required until the phase voltage at the frequency for peak
amplitude is approximately 0 volts. In this example, the phase was changed from 190 degrees (see
figure Figure 12.4e) to 310 degrees.
Adjust phase
12.5 Engage
1. Using the adjustments on Innova, adjust the focus so that the WinTV2000 image shows the
sample surface to be in focus.
Note:
Rev. B
Most surfaces being scanned in EFM will be highly polished and it may be
difficult to focus on the surface however, there will usually be slight scratches,
blemishes or contaminants to assist focusing.
181
2. Use the Motor stage controls to lower the probe until it begins to come into focus and select
the engage toolbutton.
Engage toolbutton
Auto
Z Center
182
Rev. B
4. The channel screens will appear automatically during the engage process and WinTV will
show the sample surface in focus and the probe in sharper focus than before engage.
LiftMode,
Enabled
Rev. B
183
8. Click the menu selection Channels in the Scanning window and select appropriate channels
for viewing EFM. Select Tapping Phase - Backward (LiftMode) to obtain EFM data
Begin scanning and modify settings (range, rate, etc.) as required to produce desired images.
10. Tile channel windows and click Start/Stop Scan toolbutton to obtain EFM images
184
Rev. B
Rev. B
185
186
Rev. B
Ch. 13 - Nanolithography
Overview:
Chapter 13 Nanolithography
13.1 Overview:
NanoPlot is an optional application which may be purchased separately and embedded within
SPMLab to facilitate nanolithography applications. NanoPlot provides tools to produce line
drawings and geometries and utilities to permit importing of graphics produced using other
graphics generating software.
Nanolithography utilizes the precise positioning controls of the SPM to produce graphic images at
the nanoscale. The images which can be produced range from simple geometries and line figures to
false grey scale two dimensional graphics. Images may be produced by engraving the sample with
the point of the SPM probe (scratching) or by applying an electrical potential to the probe and
utilizing an electrochemical process (anodic oxidation) with the sample.
13.2 Startup
With SPMLab running,
1. From the menu bar, select: Tools > NanoPlot
Rev. B
187
Ch. 13 - Nanolithography
Additional Instructions and Information
188
Rev. B
Ch. 13 - Nanolithography
Nanolithography - A Sample Session
1. Load a conductive tip with proper electrical contacts to permit applying an electrical bias.
Connection
Rev. B
189
Ch. 13 - Nanolithography
Nanolithography - A Sample Session
6. Menu Tools > Setup in NanoPlot to enter SetPoint and Tip Bias.
190
Rev. B
Ch. 13 - Nanolithography
Nanolithography - A Sample Session
Note:
Rev. B
191
Ch. 13 - Nanolithography
Nanolithography - A Sample Session
192
Rev. B
Ch. 14 - Calibration
Overview
Chapter 14 Calibration
14.1 Overview
This chapter describes how the scanner of the Innova instrument works and how to calibrate the
scanner to maintain its optimal performance.
The scanner is a crucial component of the Innova system. The precision of the scanner motion is
largely responsible for the quality and reliability of data. Understanding both the scanner's role in
producing images as well as how to calibrate the scanner is therefore an important part of operating
the instrument.
Typically, a Large Area scanner is provided with the Innova system. An optional small area
scanner, which is useful for high resolution applications, is also available for use with Innova. This
chapter focuses on the most common case of calibrating the large area scanner in high-gain,
closed-loop mode. Analogous procedures apply for low gain mode, for open-loop operation and
for the (open-loop) small area scanner.
Open loop scans may exhibit more nonlinearity and hysteresis.
Rev. B
193
Ch. 14 - Calibration
Test the X and Y Detector:
194
Rev. B
Ch. 14 - Calibration
Calibrating X and Y Measurements:
Figure 14.4a Horizontal and Vertical Reference Lines Dragged Onto Image
Rev. B
195
Ch. 14 - Calibration
Calibrating X and Y Measurements:
It is a good idea to rescan the sample at this point to make sure that the correction works properly.
If further correction is necessary, continue the angle / crosstalk correction process. Collect a
suitable scan of the standard grid and drag it into the Scanning control window. Choose Setup /
Recalibration / XY Recalibration and activate the Distance Measurement dialog by clicking the
button with the line segment icon on the right side of the Scanning window. The label XY
Recalibration should appear on the left side of the window under the parameters input section.
Click on the downward pointing red/blue arrows. To perform the X axis calibration make sure that
the X radio button is selected and then draw a horizontal line across a set of features (known
distance) in the image. The software will show the measured distance in the display box above the
arrows. Type in the correct (nominal) distance in the Distance dialog box. Example: If your line
covers 8 pitches, type in 80, if it covers 9 pitches type in 90. Repeat this procedure to calibrate
the Y axis, by selecting the Y radio button and drawing the line vertically. Once completed, rescan
the standard sample image and check the results.
196
Rev. B
Ch. 14 - Calibration
Calibrating Z Measurement:
Rev. B
197
Ch. 14 - Calibration
Calibrating Z Measurement:
In the example below we can see a Z Lin value of 104.2mV per 200nm. This results in a factor of
1.92nm/mV.
198
Rev. B
15.2 Set-up
The thermal tune procedure is the same whether the experiment will be performed in contact mode
or tapping mode. It should be noted that tapping mode imaging in liquid typically uses a probe
which is also used for contact imaging in air and liquid. The probes commonly used for tapping
mode imaging have high resonant frequencies in air but liquid damps the oscillation. The common
choice of probe for tapping mode imaging in liquid is the short, narrow-legged DNP or DNPs. Even
softer cantilevers can be used for contact mode but typically these do not perform well for tapping
mode imaging in liquid.
1. Mount the specimen to be imaged. If the imaging is to be performed in liquid, the thermal
tune procedure is performed with the specimen in the liquid.
2. Set-up Innova to perform contact mode imaging as described in Chapter 7 up to engaging.
Rev. B
199
200
Rev. B
3. For later reference, click on the drop down window to see what the display options are. Note
that only two y-axis options are available: V or m.
Options
Rev. B
201
Point 1
Point 2
202
Rev. B
7. Click the Calibrate toolbutton and a Deflection Sensitivity Calibration message window will
appear.
Rev. B
203
3. The Thermal Tune window will open. Notice that the sensitivity value 0.0195, which was
determined in the previous section, is shown in this window. The spring constant value
which is displayed is a remnant from some earlier analysis and is not valid. A correct spring
constant will be determined later in this procedure.
204
Rev. B
4. Clicking the Start/Stop toolbutton will begin to perform/average spectrum analyses and
display the number of analyses performed in window at the bottom of the screen until the
toolbutton is again clicked. The example was stopped at 209 iterations.
Rev. B
205
5. After the acquisition has been stopped, select the measurement tool
corner of the window.
Zoom window
206
Rev. B
7. Release the mouse button and the selected zoom area will be displayed. As the cursor is
moved on the response curve the values of the curve are displayed at the bottom of the
screen.
Rev. B
207
Note:
If desired/needed, the curve fit may be removed and redrawn by clicking on the
Fit toolbutton.
Note:
9. Click the Calibrate toolbutton and the spring constant value will be updated
208
Rev. B
Spectroscopy as shown in the revised window. This option allows force curves to be
displayed and saved in force units.
Rev. B
209
210
Rev. B
2. If the data extends out of the screen range, the input gain is set too high and must be adjusted
to produce an acceptable level as illustrated in Figure 15.4c
Adjust
as needed
Rev. B
211
reflected peak will occur at different locations. Peaks truly located below the Nyquist frequency
will appear at the same frequency (their true frequency) regardless of the sampling ratem.
212
Rev. B
Appendix A Synchronizer
Notice:
This section describes specialized applications for advanced users. Operation of the Innova system
does not require familiarity with the contents of this section except for those specialized
applications. In general, the majority of users may skip this section unless, and until, the
Synchronizer capability is known to be required.
A.1
Introduction:
Veeco Innova Synchronizer is an option which provides the capability to synchronize the image
scanning operations of an Innova AFM with the operation of another device. The synchronization
is performed by handshaking signals between the two instruments. The Nanodrive controller can
output handshaking signals with designated voltage levels and durations. It can also accept input
handshaking signals with designated voltage levels.
The Innova synchronizer requires no additional hardware. The handshaking signals are fed in and
out via the interface board installed in the standard configuration Nanodrive controller. The
Synchronizer option requires SPMLab Version 7.11or higher. Contact Veeco to purchase the
Synchronizer option functionality and refer to option part number INSYNC.
Rev. B
213
Appendix A: Synchronizer
Set-up
A.2
Set-up
A.2.1
OUT2
Figure A.2a
214
BNC Connections
Rev. B
Appendix A: Synchronizer
Set-up
Figure A.2b
3. In the Synchronizer window, specify the pulse required by the synchronizing device.
Check boxes:
Active: checked.
Handshake: unchecked.
Test: checked. (Changes the Apply button to Run for testing synchronization
handshaking. When imaging, Test should be unchecked.)
Figure A.2c
Rev. B
215
Appendix A: Synchronizer
Set-up
Figure A.2d
CAUTION:
Ensure the output voltage levels are compatible with the receiving device to prevent
electronics damage.
4. When the Run button is clicked, the NanoDrive controller outputs a single pulse. During the
output, the Synchronizer window becomes temporarily busy, depending on the pulse width,
then returns to the idle state.
216
Rev. B
Appendix A: Synchronizer
Set-up
A.2.2
IN2
Figure A.2e
Rev. B
217
Appendix A: Synchronizer
Set-up
Active: checked.
Handshake: checked.
Tuning: checked.
LED indicator
Figure A.2f
218
Rev. B
Appendix A: Synchronizer
Set-up
3.
High level: the upper level of the input handshaking pulse. Maximum: 10V.
Low level: the lower level of the input handshaking pulse. Minimum: -10V.
Rising Edge: When checked, the rising edge is used for handshaking, otherwise, the
falling edge is used.
Handshake
Output Signal
Handshake
Input Signal
Figure A.2g
4. Click Run. The NanoDrive controller should immediately output a voltage specified by
high level in the Pulse Out section, then wait for the handshake signal to lower the output
to the low level. The small LED (see Figure A.2f) indicates the status of the handshake
signal. It has three states:
Rev. B
219
Appendix A: Synchronizer
Configure Software
A.3
Configure Software
1. Specify the handshaking signals as described in the preceding sections.
2. Un-check the Tuning selection and click Apply to activate the synchronizer.
Figure A.3a
3. Open the Scanning Conditions window using the menu selection Conditions.
Figure A.3b
220
Rev. B
Appendix A: Synchronizer
Configure Software
Figure A.3c
Backward line:
Beginning: When checked, the Nanodrive controller outputs a specified handshaking output signal
when it is ready to begin a scan line in the backward direction. If Handshake is enabled (see the
checkbox in Figure A.3a), Nanodrive will wait for a handshaking input signal to start the scan.
End: When checked, the Nanodrive controller outputs a specified handshaking output signal when
it finishes scanning a line in the backward direction.
Rev. B
221
Appendix A: Synchronizer
Run Synchronizer
Pixels: When checked, the Nanodrive controller outputs a specified handshaking output signal
when it is ready to move to the next pixel in the backward direction. If Handshake is enabled (see
the checkbox in Figure A.3a), Nanodrive will wait for a handshaking input signal to start the scan.
Note:
Since the end of a forward line is usually also the beginning of the next
backward line, Nanodrive will output two handshaking pulses when both
checked. The same is true for the end of backward line and the beginning of
forward line.
A.4
Run Synchronizer
1. In the Scanning control window, click the Start/Stop scanning button to start scanning and
synchronization will be performed.
Figure A.4a
222
Rev. B
Note:
This section describes specialized applications for advanced users. Operation of the Innova system
does not require familiarity with the contents of this section except for those specialized
applications. Even advanced modes such as Scanning Capacitance Microscopy do not usually
require Open Hardware features. In general, the majority of users may skip this section unless, and
until, the Open Hardware capability is known to be required.
B.1
Introduction
The included Open Hardware feature provides wide flexibility in the configuration of the Innova system.
The Open Hardware option provides access to the configurable electronics including: DACs, ADCs,
Multiplexers, lock-in amplifiers, signal generators, and others. With Open Hardware access, the various
accessible items may be configured as needed for special requirements of imaging or testing.
It is important to understand several aspects of the Open Hardware feature:
The default SPMLab values will appear in the various control windows until Get
Parameters is executed.
There are no safeguards. Reversing feedback or other actions can cause the system to behave
erratically and may result in damage to the probe and/or sample.
None of the changes made using Open Hardware controls will be saved. By closing and reopening SPMLab, all normal default conditions will be restored.
Rev. B
223
Open Hardware
Software Setup
B.2
Software Setup
1. Menu select: Tools > Open HW Access.
224
Rev. B
Open Hardware
Software Setup
3. The Open hardware tool bar will display on the right side of the SPMLab main window.
Rev. G
225
Open Hardware
Open Hardware System Diagram
B.3
B.3.1
Figure B.3a displays the main components of the Innova system. The three main components are the SPM
Instrument, the NanoDrive Controller and the PC. During normal operation, the DSP inside the NanoDrive
controller controls the Instrument, receives commands from the PC and sends data back to the PC.
USB
DSP Board
(DSP & FPGA)
SPM
Instrument
Data Acquisition
Board
Interface
Board
(IO-I)
Lock-in
Board
HV Board
(IO-HV)
Optional Boards:
IO-X signal access
Digital Bus
(IO-MOD+)
Analog Bus
(Innova)
226
Rev. B
Open Hardware
Open Hardware System Diagram
The sub-functions of the NanoDrive controller are split over a number of modular boards:
1. Data Acquisition board:
This board handles most of the Analog to Digital conversions (ADC) and a number of the
Digital to Analog conversions (DAC). It interacts with the other boards through the
digital and analog bus.
This board interacts directly with the Innova SPM instrument and handles all of the low
voltage signals coming and going to the instrument.
This is a two channel lock-in board that is, as an example, used during Tapping mode to
generate the Tapping drive and determine the Tapping amplitude and phase. It normally
interacts with the other boards through the analog and digital bus, but has powerful input
and output capabilities through the BNCs mounted on its front panel.
This board generates all the high voltages needed for the scan motion.
5. Optional boards:
Rev. G
Additional boards, such as the IO-X Signal Access board, can be placed in the NanoDrive
controller and interact with the other boards through the analog and digital bus.
227
Open Hardware
Open Hardware System Diagram
B.3.2
The Open Hardware feature in the software allows for direct interaction with the different boards in the
NanoDrive controller, outside of the normal SPMLab software routine. The diagram in Figure B.3b gives an
overview of the Open Hardware functions and where they interact with the NanoDrive controller.
Main clock
XYZ Feedback selection and
settings
Data AcquisitionBoard
x
x
x
x
Interface Board(IO-I)
x
x
x
x
Lock-in Board
x
HV Board
x
x
x
x
Digital Bus
Analog Bus
Optional Boards:
IO-X signal access
228
Rev. B
Open Hardware
Input/Output Signal Access
B.4
B.5
B.5.1
Feedback Control
Rev. G
toolbutton.
229
Open Hardware
Open Hardware Functions
DSP Clock: Main FB period in sec. This changes the overall speed of the controller. At
values below 10.0 sec. (= 100kHz), errors may occur on some applications
X/Y/Z Channel: For each of the Z, Y and X Channel sections, the PID Coefficients are
different from the feedback values in the SPMLab main window GUI. The values in this
window are the actual values applied to the hardware feedback loops while the values in the
GUI are scaled for convenience.
B.5.2
On: When unchecked, the feedback loop is inactive. When checked, the feedback
operation is active. The value is the actual value of the target under feedback control
(feedback input signal).
ADC: The ADC used to measure the input signal that is under feedback control.
Get Parameters Button: When clicked all the parameters in the window are updated to
display the current status of all the parameters in the window. To change any parameter or
setting, change the parameter and <Enter> from keyboard.
Multiplexer Control
toolbutton.
230
Rev. B
Open Hardware
Open Hardware Functions
This dialog permits setting the multiplexers present on the data acquisition board. There are two types of
multiplexers. The first type has 16 inputs and the output feeds an ADC channel. The second is a 64-to-1
multiplexer that can be used to access more signal lines and is primarily used in the Oscilloscope and
Multimeter functions.
The 16-to-1 ADC Multiplexers
MUX
Gain
ADC
0..6
2 kHz
9
Offset
Offset
DAC
10 kHz
LPF
50 kHz
Line
AL0..AL15
Each ADC channel has one 16-to-1 multiplexer. Since there are 8 ADC channels there are also 8 such
multiplexers. Seven of them, those that are associated with ADCs 0 through 6, are freely accessible through
the multiplexer dialog. See the upper section of the dialog shown in Figure B.5b. Since all of these ADC
channels have identical functions, their control is combined in that upper section. The general functions of the
selection boxes in the dialog are shown in Figure B.5c.
Rev. G
231
Open Hardware
Open Hardware Functions
In standard SPM applications the analog lines (AL) have pre-defined signals on them, generated by the
different boards in the NanoDrive controller. Table B.5d lists the signals on the AL lines per application. It
also lists the default ADC that is used by the SPMLab software to monitor those signals. The actual ADC
number assigned to a certain signal can vary depending on software configuration.
STM Sensor
ADC0
ADC0
ADC4
ADC4
ADC0/
4
ADC0
ADC0
ADC
SCM Phase
AL4
ADC0
ADC0
ADC0/4
Tapping Amplitude
ADC0
ADC0
ADC5
LC-STM Sensor
ADC0/4
C-AFM Signal
ADC5
PRM Signal
ADC6
FMM Phase
AL5
ADC6
ADC6
SEPM Potential
ADC6
AL6
X LIN
ADC3
ADC3
ADC3
ADC3
ADC3
AL7
Y LIN
ADC2
ADC2
ADC2
ADC2
ADC2
ADC2
AL8
Z LIN
ADC1
ADC1
ADC1
ADC1
ADC1
ADC1
AL9
IN 1
ADC5
ADC5
ADC5
ADC5
ADC5
AL10
IN 2
ADC6
ADC6
ADC6
ADC6
ADC6
Tapping Phase
ADC4
ADC3
ADC3
ADC3
ADC2
ADC2
ADC2
ADC1
ADC1
ADC1
ADC3
ADC3
ADC2
ADC2
ADC2
ADC1
ADC1
ADC1
ADC5
PRM Amplitude
ADC5
FMM Amplitude
ADC5
SEPM Error
AL12
ADC5
SCM Detector
ADC3
Tapping Phase
ADC4
AL13
SCM Sensor
AL14
AL15
ADC3
ADC4
SCM Amplitude
AL11
Surface Potential
AL3
ADC5
ADC1
Force Modulation
SUM
ADC0
Piezo Response
AL2
ADC5
SCM
ADC1
C-AFM
ADCO
Lateral Signal
EC-STM
Deflection Signal
AL1
EC-tapping
AL0
EC-contact
Signal
LCSTM
Analo
g
Line
STM
Contact Mode
Tapping Mode
Modes:
T-B Fast
ADC6
ADC6
ADC6
232
Rev. B
Open Hardware
Open Hardware Functions
Description
NR
Description
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
39
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
DACmon0 (VTIP)
DACmon1
DACmon2
DACmon3
DACmon4
DACmon5
DACmon6
DACmon7
DACmon8
DACmon9
DACmon10
DACmon11
DACmon12
3.3 V digital supply
5.0 V digital supply
DGnd digital ground reference
+15V analog supply (div. by 2)
-15V analog supply (div. by 2)
+5V supply of DACs
-5V supply of DACs
5.00V reference for calibration
-5.00V reference for cal. verification
2.50V reference for cal. verification
AGnd analog signal reference
5V supply of ADC0 and ADC1
5V supply of ADC2 and ADC3
5V supply of ADC4 and ADC5
5V supply of ADC6 and ADC7
+15V digital supply (div. by 2)
-15V digital supply (div. by 2)
Unused, open
Unused, open
Rev. G
233
Open Hardware
Open Hardware Functions
Section 2 = 7th ADC muxes: ADC no.7 can be set up to take different signals. This is implemented by
configuring the mux. Additionally, the output of this ADC can also be configured to display on the
oscilloscope window.
Section 3 = 64x1 mux: Multiplexer controls which signal is routed to AL15 (which can then be measured
by an ADC.
234
Rev. B
Open Hardware
Open Hardware Functions
B.5.3
This window allows one to directly set certain DAC outputs and read the ADC values. Table B.5f lists the
DACs that can be set. Some DACs are continuously updated through the regular SPMLab software and will
not retain the value set by this dialog. For example, in the closed-loop XY operation setting, the X-DAC will
momentarily set the X-DAC to a different value, but 10sec later the closed-loop feedback will correct the
position offset and change the X-DAC output correspondingly.
Note: When the tip is engaged or close to the sample, changing the DAC values can result in tip or
sample damage.
DAC#
Description
DAC0 - XDac
DAC1 - YDac
DAC2 - ZDac
DAC3 - VTip Dac
DAC4 - Aux Dac1
DAC5 - Aux Dac2
DAC6 - Aux Dac3
DAC7 - Aux Dac4
DAC8 - Aux Dac5
DAC9 - Aux Dac6
DAC10 - Aux Dac7
DAC11 - Aux Dac8
DAC12 - Aux Dac9
DAC13 - Aux Dac10
DAC14 - Aux Dac11
DAC15 - Aux Dac12
DAC16 - Offset Dac0
DAC17 - Offset Dac1
DAC18 - Offset Dac2
DAC19 - Offset Dac3
DAC20 - Offset Dac4
DAC21 - Offset Dac5
DAC22 - Offset Dac6
DAC23 - Offset Dac7
Low Voltage X
Low Voltage Y
Low Voltage Z
Tip or Sample Bias Voltage
Auxiliary DAC1
Auxiliary DAC2
Auxiliary DAC3
Auxiliary DAC4
Auxiliary DAC5
Auxiliary DAC6
Auxiliary DAC7
Auxiliary DAC8
Auxiliary DAC9
Auxiliary DAC10
Auxiliary DAC11
Auxiliary DAC12
Offset DAC for ADC0
Offset DAC for ADC1
Offset DAC for ADC2
Offset DAC for ADC3
Offset DAC for ADC4
Offset DAC for ADC5
Offset DAC for ADC6
Special purpose
Rev. G
235
Open Hardware
Open Hardware Functions
DAC: Use the drop down list to select the desired DAC and move the slider control to
adjust the DAC output for each DAC of interest. The output range is from 10V to +10V for
each DAC.
ADC: Use the drop down list to select the desired ADC. Enable the ADC readout by
clicking the red LED button
B.5.4
Use this dialog to set the tip or sample bias voltage directly, or to add a modulation voltage to the tip or
sample bias. The VTMOD signal allows one to add a modulation signal to the tip or sample bias. The
VTMOD amplitude and frequency are set in the IOMod Control dialog and the VTMOD signal needs to be
enabled in the IOMod Control dialog. By selecting both Enable and VMOD the bias voltage will have
both the DC and AC components.
A schematic diagram of the controls is shown in Figure B.5h.
VTIP DAC
/100
VTIP
VTMOD
+VMOD
+VMOD/100
Enable
236
Rev. B
Open Hardware
Open Hardware Functions
toolbutton.
Rev. G
237
Open Hardware
Open Hardware Functions
B.5.5
IOMod+ Control
This dialog allows for the full control of the dual channel lock-in. Any changes made in this dialog are
applied instantaneously, however, the parameters are not updated if the window is already open and some of
the values are changed using the regular SPMLab software controls (like the cantilever tuning window).
Use the Get Parameters button when switching back and forth between the IOMod Control dialog and the
regular SPMLab functions.
Open the IOMOD Control window by clicking the
toolbutton.
238
Input BNC Shell is Grounded: When checked, input BNC shell is grounded.
Input Gain: Gain applied to input signal before performing lock-in analysis.
Post Gain: Gain applied to input signal after performing lock-in analysis.
Rev. B
Open Hardware
Open Hardware Functions
Rev. G
Section 4: BNC.
BNC1: Signal to be output through BNC1 on the IOMod+ board (at rear of Nanodrive
controller).
Z_BI_MOD: Green = Apply the Drive of Channel A to the Z-BI-MOD line. Use the
Dither Drive Source in the IOI Board control window to connect that line to the dither
piezo. Red = Disabled.
Z-MOD: Green = Apply the Drive of Channel A to the Z-MOD line. Use the Z-Voltage
Sources selector in the IO-HV Control window to connect the Z-MOD line to the Zpiezo, or as addition to the regular Z-DAC voltage. Red = Disabled.
VTMOD: Green = Apply the Drive of Channel A to the VTMOD line. Use the VTIP
Control window to add this modulation voltage to the regular tip or sample bias. Red =
Disabled.
239
Open Hardware
Open Hardware Functions
B.5.6
This dialog controls the Innova Interface Board (IO-I board). It allows for putting certain signals on analog
lines (AL), selecting the sources for generic output BNCs and selecting the sources for the dither and bias
voltages. Always avoid turning on multiple signals for the same AL line. Turn off connections first, before
making new connections.
Open the Innova Interface Board window by clicking the
toolbutton.
240
Input AL Connections: Controls the routing of signals. Mark check boxes of desired
options.
Out BNCs: Controls the routing of signals to the two BNCs on the IO-I board (at rear of
NanoDrive controller).
Dither Drive Sources: Controls the routing of signals to be applied to the dither piezo.
ZMOD BNC to dither: Green = overrides the dither drive source selection and connects
the dither to the ZMOD BNC on the IO-I panel. Red = disabled.
Rev. B
Open Hardware
Open Hardware Functions
B.5.7
This dialog controls the Innova High Voltage board (IO-HV board). It allows selection of the low voltage Z
sources (which is the input of the high voltage amplifier), enabling of monitor lines, selection of the high
voltage gain and enabling or disabling of individual xyz outputs.
Schematic diagrams of the high voltage controls are shown in Figure B.5l through Figure B.5p.
9
='$&
+96XSSO\(QDEOH
+9=(QDEOH
=+9
[ RU
[
=(;7
=0 2'
/RZ *DLQ =
y
+90RQ
=([SDQVLRQ
0RQ(Q
='$&[=02'
*5281'
9
;'$&
+96XSSO\(QDEOH
+9;(QDEOH
;+9
[ RU
[
/RZ*DLQ;
y
+90RQ
0RQ(Q
Rev. G
241
Open Hardware
Open Hardware Functions
9
;'$&
+96XSSO\(QDEOH
+9;(QDEOH
Q;+9
[ RU
[
/RZ*DLQ;
y
+90RQ
0RQ(Q
9
<'$&
+96XSSO\(QDEOH
+9<(QDEOH
<+9
[ RU
[
/RZ*DLQ<
y
+90RQ
0RQ(Q
242
Rev. B
Open Hardware
Open Hardware Functions
9
<'$&
+96XSSO\(QDEOH
+9<(QDEOH
Q<+9
[ RU
[
/RZ*DLQ<
y
+90RQ
0RQ(Q
toolbutton.
Z voltage Source: Selects the voltage source applied to the piezo tube for Z motion. The
selected checkbox source is activated.
Monitor Lines: Selects the signals to be monitored. The checkbox activates the selected
signal (group).
Rev. G
When pressed, enables the high voltage output for all X/Y/Z.
Low Gain: This is to enable / Disable the low gain option for X/Y/Z.
Innova User Manual
243
Open Hardware
Examples
B.6
Get parameters? Updates the status of all the parameters in this window.
Examples
B.6.1
This example shows how to reconfigure an ADC input to another source. It is a way to add external signals
to the measurement channels or look at signals that are not configured by default. This example shows how
to use IN2 on the IO-I board as input during Tapping Mode imaging.
Note: The input signal always has to be between -10V and +10V!
Start with the assumption that the system is configured for Tapping Mode. From Table B.5d, it can be seen
that by default the IN2 channel or the IO-I board is not being measured. Another way to see is by opening
up the channels list in the scanning control window, see Table B.6a.
244
Rev. B
Open Hardware
Examples
245
Open Hardware
Examples
7. All that is left if to connect the external signal (between -10V and 10V) to IN2 of the IO-I board.
8. The TM Deflection signal now will show the IN2 signal. However, the regular SPMLab
software is not aware of the changes, so the default names will still be listed as AL0 - TM
Deflection even though it now really is AL10 - IN2.
B.6.2
In Tapping Mode, by default, the system feeds back on the Amplitude of the cantilever oscillation. This can
be easily changed using the Open Hardware panels.
This example assumes that everything was already setup correctly to do regular Tapping Mode imaging
using the SPMLab default settings. Follow the next steps to change the z-feedback from Amplitude to
Amplitude x cos(phase):
1. It is highly recommended to withdraw the tip from the surface. Changing feedback signals can
easily lead to instable feedback loops and result in tip or sample damage.
2. Open the Open Hardware panel if that is not already open.
3. Open the IOMod+ Control window and select Get Parameters to fill in the current parameters.
246
Rev. B
Open Hardware
Examples
5. The quickest way to change this to Amplitude x cos(phase) is to click the selection box in the top
right of the IOMod+ Control dialog and select Signal x COS (phase) as the AL4 line signal.
Figure B.6g IOMod+ Control Dialog Showing the Options for AL4
6. Verify that the feedback loop still has the correct sign, e.g. z is retracted when the signal is smaller
than the setpoint, and re-engage to the sample. Feedback polarity can be switched in the
Feedback Controls dialog, see Section B.5.1.
B.6.3
This example shows how to change the default filter settings in front of the ADC measurement channels. The
first step is to find out which ADC carries the signal on which to apply the extra filtering. Use Table B.5d to
determine which analog line and which (default) ADC carry the signal.
For example, to add a filter to the IN1 signal in contact mode:
1. Open the Open Hardware control panels
2. Open the Multiplexer Control dialog
3. Verify that AL9 (=IN1) is connected to ADC5 by selecting ADC channel 5 and pressing the ?
button. This will also list the default filter settings for that channel.
Rev. G
247
Open Hardware
Examples
4. Now select the filter settings you want to use and hit the arrow button to apply the change.
Figure B.6i ADC5 with Additional 2kHz Low Pass Filter Set
B.6.4
This example shows how to change the z-feedback to a different or external signal. If the switch over has to
happen while the tip is engaged on the surface, it is highly recommended to do a dry run with the tip
disengaged from the surface, in order to make sure all feedback settings are correct.
There are a few options on how to switch feedback signal through the Open Hardware panels:
1. Change the feedback ADC# in the Feedback Control window
2. Change the AL line assignment to the feedback ADC using the Multiplexer Control dialog
3. Change the signal on the AL line, for example through the IOMOD+ control dialog.
B.6.5
248
Rev. B
Open Hardware
Examples
2. Open The Feedback Control dialog and press the Get parameters button.
Rev. G
249
Open Hardware
Examples
b
a, f
c
d
B.6.6
250
Rev. B
Open Hardware
Examples
5. By moving the tip position around you can manually scratch or press on the surface. The setpoint
will determine the mount of force applied.
6. Reversing the steps of section 5.5 will turn the system back into Tapping Mode.
7. Now you can take another image to see the results of the nanomanipulation.
B.6.7
Second-Harmonic Detection
This example shows how to use the 2nd channel of the lock-in to detect the 2nd harmonic component of the
cantilever oscillation.
First, setup the system for regular Tapping mode, and take an image. By default it will use the first lock-in
channel (channel A) to measure the amplitude of the cantilever oscillation. The second channel (channel B) is
not used, so that channel can be configured to look at the second harmonic of the cantilever oscillation.
Use the following steps to set this up:
1. Open the Open Hardware panels.
2. Open the IOMod+ control dialog.
3. The Tapping drive frequency is displayed in the Channel A - frequency box. However, the number
of digits at which the frequency is displayed is less than the accuracy at which was set using the
Tuning window. So in order to accurately set the 2nd harmonic we must first set the Channel A
frequency to an exact value. For example, the cantilever frequency in Figure B.6m was set to
Rev. G
251
Open Hardware
Examples
328.2845 kHz but in order to know the frequency exactly a value of 328.29 was typed into the
frequency box of Channel A.
Figure B.6m IOMOD+ Control Dialog Set-up for 2nd Harmonic Detection
4. Now, multiply the frequency of Channel A by a factor 2 and use that as the frequency of Channel
B.
5. Switch the Source for Channel B to AL14. This is the fast deflection signal and the same source
as Channel A.
6. In the Analog Lines section of the IOMod+ Control dialog, switch the AL12 line to the
Amplitude of Channel B.
7. Open up a Multimeter window and display AL12.
252
Rev. B
Open Hardware
Examples
For this, open up the Multiplexer Control window and find the ADC that is looking at the TM
Deflection signal. By default this is ADC5 connected to AL0.
Rev. G
253
Open Hardware
Examples
254
Rev. B
Index
A
AFM probe cartridge 50
Align Laser 152
Aligning Laser 96
Alignment Knobs 62
anodic oxidation 188
Area scanning 69
Auto Approach 143
B
Bias Line 189
C
Cable Connections 42
Calibration 193, 199
Calibration References 193
cantilever 4
Cantilever Tuning 115, 176
Channel 130
Channel Selection 100, 184
chip carrier 52
Closed Loop 86, 141
Contact AFM principles 93
Contact Imaging 93
Contact Information iv
D
Deflection Sensor 58, 62
E
EFM Imaging 173
Engage 86, 99, 127, 153
F
feedback 4
Feedback Loop 6, 16
Feedback Signal 15
G
Gains 16
H
high gain 81
I
Indentation 188
Installing a Chip Carrier 49
Integral Gain 16, 103, 143
IV curves 82
L
Laser Indicators 64
Laser on/off 57
Leveling 103
LFM Imaging 105
LFM Signal 107
Lift Mode 79, 161
Line scanning 71
Loading a Sample 47
low gain 81
M
MFM 161, 173
MFM Imaging 161
Mode 161
Motor Speed 66
Motor Stage 97
Motor stage 66
Multimeter 87
N
NanoPlot 88, 187
Number of Data Points 10, 104
O
Oscilloscope 87
Other Controls 91
oxidation mode 188
P
Pan/Zoom 117, 165
phase curve 121
phase detection 161, 173
phase setting 166
Point Spectroscopy 81, 87
Point Spectroscopy window 155
Probe Position 87
Probe positioning 78
Index
Probe Positioning Window 83, 158, 160
Profile 100, 133
Proportional Gain 4, 16, 103, 143
S
Scan angle 11
Scan Parameters 102
Scan Range 102
Scan Rate 7, 11, 102
Scan Size 7, 11
Scanner 2, 44
Scanning conditions 81
Scanning Window 141, 158
Select Frequency 120
Set Frequency 166, 180
Setpoint 5, 10, 14, 15, 104
Signal Tracing 87
Signal Tracing window 83
Single Point Spectroscopy 3
Space requirements 40
Spectroscopy 149
SPM 5
spring tool 50
Stage adjustments 56
Stage Reset 66
Start/Stop Frequency Sweep 165
STM Cartridge 137
STM Image 140, 148
STM Tips 135
System Configuration 60, 114
T
Tapping AFM 113
These 11
Tip Bias Voltage 184
V
Veeco Contact Information iv
W
WinTV32 53, 151
X
X and Y offsets 11
Z
Z piezo linearizer 157
Z Position Bar 14
zoom box 117