Sei sulla pagina 1di 2

Application Note

Measurement of different Sugar based


samples like Sugar, Honey, Syrup with
FTIR Technique

Syrup, made from different sugar, is a mixture


of Fructose, Glucose, Maltose, etc. And in
this configuration the base of soft drinks
ready for dilution and taste ingredients. The
mixture is designed for a designer product.

These are three different samples of sugar


which are under control of food analysis. The
white crystalline sugar for the coffee is a
nearly pure Saccharose prepared on
technical control as refined sugar of industrial
way.

90
%T
75

60

45

30

15

0
4500
4000
ATR Honig 8

3500

3000

2500

2000

1750

1500

1250

1000

750
1/cm

Fig. 1: Infrared transmission mode spectra prepared from crystalline sugar (light green-yellow line) like coffee sweetener, commercial
honey (grey line) and syrup (green line) as base for soft drinks

Totally different is the honey which is in this


case a sample from a natural product.
Designed from bees for bees, Honey includes
a lot of natural parts like polls, enzymes and

different sugars like maltose, turanose,


glucose and fructose, partly also saccharose.

Discussion of results:
One trial to get more details from the syrup
spectrum was the subtraction of a water
spectrum to get more details about the
remaining sugar. In fig. 2 is the result of a

SCA_110_022

subtraction of water from syrup and the sugar


spectrum in one view.

www.shimadzu.de

Abs

Abs

2500

2000

1750

1500

1250

1000

750
1/cm

Fig. 2: Shown are absorbance mode infrared spectra from


different sources green is syrup minus water and grey is the
honey spectrum, they are looking not so different in the
fingerprint range, only small details are different
Her Majesty the Queen in Right of Canada 1986

Abs

0.8

0.6

0.4

0.2

0
4000
3500
3000
DEXTROSE;D-GLUCOSE

2500

2000

1750

1500

1250

1000

750

2500

2000

1750

1500

1250

1000

750

500
1/cm

All three samples have their specific


spectrum. Just in the fingerprint range are to
see a lot of differences in the individual
spectra. One chance to visualize these
specialities is for example the derivative
spectroscopy. The syrup spectrum is
influenced by the water signal. Subtraction of
water is one possibility to get more
information about the content of the syrup.
Another possibility is to use pure sugar
spectra for the subtraction activity. Fig. 4
contains Dextrose, Glucose spectrum from
powder measured with KBr-Pellet technique.
It seems comparing the pure spectrum with
the crystalline sugar they have more
similarities.

Instrumentation
Instruments in charge were the FTIR-8000Series and IRPrestige-21. To be able to measure the
samples different accessories were used and commercial library from Sadtler, the Canadian
Forensic.
1. Condition for syrup sample
FTIR-8000 Series, H-ATR with ZnSe crystal
trough style. The sample was filled into the
trough and immediately measured.
2. Condition for honey sample
IRPrestige-21, Miracle with ZnSe crystal
One drop sample was simply transferred to
the crystal.

Agate mortar and pestle


A piece of sugar was grinded to represent a
very fine sugar particle. The powder was
brought under the anvil of the pressure
device from Miracle accessory. Sample was
pressed to the ZnSe crystal.

3. Condition for sugar


IRPrestige-21
Miracle with ZnSe crystal
Shimadzu Europa GmbH
Albert-Hahn-Str.6-10, 47269 Duisburg

1/cm

Fig. 3: Shown are absorbance mode spectra from different


sources green is syrup minus water and yellow-green is
simple sugar spectrum, they are looking different in the
fingerprint range

Fig. 4: Spectrum from commercial library.

4500
4000
3500
3000
Subtraction of WATER from SAMPLE1

www.shimadzu.de

The given specifications serve purely as technical information for the user.
No guarantee is given on technical specification of the described product and/or procedures.

4500
4000
3500
3000
Subtraction of WATER from SAMPLE1

Potrebbero piacerti anche