Sei sulla pagina 1di 3

Measuring range:

Separation and detection EPA10 conventional anionic species: F-, Cl-, NO2-, PO43-, Br-, NO3-, SO42-,
ClO2-, BrO3-, ClO3Separation and detection of other anions: Te O42-, SeO42-, CO32-, HCO3-, ASO43-, SiO32-, BO33, Conventional separation and detection of cations: Li +, Na +, NH4 +, K +, Mg2 +, Ca2 +, Sr2 +, Ba2 +
Separation and detection of other metals: Zn, Fe, Mg, Al, etc.
Separation and detection of organics: betaine, choline chloride, organic amine
Separation and detection of organic acids: formic acid, acetic acid, propionic acid, oxalic acid, tartaric
acid
The instrument uses the latest technology, a new generation of products, reflecting the trend of
contemporary ion chromatography, and in the application of the method is also done in-depth
research.
CIC-100 Series Ion chromatography has been widely used in environmental monitoring, health and
epidemic prevention, geology, animal feed, food and beverage, power, research and teaching, etc.

Main features:
Instrument features
integrated machine, digital control, 5.7-inch large-screen LCD display, membrane keypad
operation, with memory function. Full screen display voltage, conductance, current, flow, range,
pressure and other operating parameters, to facilitate real-time monitoring instrument.
advection pump assembly of imported high-pressure, full-peek passage, resistant to acid, alkali,
corrosion-resistant, 100% compatible with the organic solvent, the entire flow system without
leakage.
realize the whole instrument recrimination, by workstation traffic, current, range, zero baseline
immediate adjustment.
Rapid Determination: rapid analysis of conventional components within 10min.
based on user characteristics, package design tailored to customers detection scheme.

Technical parameters:
Work Environment
Power: 220V, 50HZ ambient temperature: 15 ~ 30 Humidity: 5 to 85%;
Host computer
Integrated machine, digital control, 5.7-inch LCD with large-screen LCD display, touch button
operation. Voltage, conductance, current, flow range, pressure and other operating parameters fullscreen display for easy instrument for real-time monitoring.
Ion chromatography pump
Type: double-piston high-pressure pump series advection
Pump Pressure Range: 0 Mpa -40Mpa
Flow range: 0.001 mL / min-9.999mL / min
Flow Accuracy (setting error): RSD0.5%
Flow Stability: RSD 0.2%
Pressure Display Accuracy: 0.4Mpa
Overvoltage protection: overpressure automatic alarm and stop the pump to protect themselves on
their own
Conductivity detector
Structure: Five pole ring passivated 316 stainless steel conductivity cell with electronic thermostat,
temperature accuracy: 5 ~ 50 0.05
Cell volume: 1.0l
Output voltage: -1300 ~ 3000mv adjustable
Range: 10 stalls available, with manual and automatic gear selector function, -1500mv ~ + 1500mv
adjustable
Detection range: 0 ~ 1000us / cm
Resolution: 0.05ns
Baseline Noise: 1.0% FS
Baseline drift: 2.0% FS
Analysis reproducibility: 1.5% (SO42- meter); 1.5% (Na + - meter)
Linear range: 102
The minimum detectable concentration: 2ppb (0.002mg / L) (CI- dollars)
Flow system
High pressure piping, fittings whole PEEK material; US Rheodyne six-port valve, the maximum
pressure 7000psi;

Column chromatography
Efficient anion column; imported cation column; guard column.
Suppression system
Automatic regeneration anion micro membrane suppressor, not-for-acid, online automatic
regeneration, uninterrupted use, simple operation, long life;
0 ~ 120mA constant current source current is adjustable in increments 0.1mA;
Gas-liquid separator
1) can be eliminated from the eluent apparent bubbles, and during operation (pump stop state),
followed by evacuation or replace eluent.
2) In the shower when less lotion, automatic alarm, to prevent damage to the instrument.
Online degassing
Former type SHT-2-line degasser, automatic line degasser, without the machine offline eluent
degassing.
Selection of imported Teflon-AF off the trachea, gas emission rate 200-300 times the ordinary PTFE
tube. High gas removal efficiency.
Chromatography Software
Chinese display, equipped with an electronic version of the operating demonstration,
chromatography software dedicated work, have spectra timing curve shows the spectrum shift,
working conditions and other select memory function, automatically gives the results of the analysis
and a variety of different correction has Chinese display, print, spectrum, calibration curve, analysis
and other functions;
Separation capacity
High-performance anion column can be analyzed F-, Cl-, NO_2, PO4-, Br-, NO_3, SO42-, SCN-, ClO2-,
BrO3-, ClO3- and other anions.
Direct conductivity analysis betaine, Li +, Na +, NH4 +, K +, Mg2 +, Ca2 +, Sr2 +, Ba2 +, choline
chloride and other cations

Potrebbero piacerti anche