The high-voltage measurement instruments HVM 2001 / HVM 2003 consist of a measuring amplifier with digital display for universal use, a built-in high-voltage divider for direct measurement up to 10 kV / optional 20 KV and an external input for an external highvoltage divider, HVT***RCR, for measurement and display of voltages up to 200 kV.. The high-voltage measurement system HVM 2001 is designed for measuring ac- and dcvoltages. The high-voltage measurement system HVM 2003 allows in addition measurement of the peak value of lightning and switching pulse voltages. Optionally the unit can be delivered with a DKD calibration certificate. Technical specification: Direct high-voltage-input Direct high-voltage-input, optional External input, Rs = 1000 k Calibration possibility of the preamplifier Analogue output, Rs = 5 k Weighting of input voltage: DC AC AC Pulse, LI & SI Pulse, LI & SI HVM 2001 10 kV=/ 7kV 20 kV=/14kV 100V100% 2% 5V100% 1mA100% Average + Peak/2 - Peak/2 HVM 2003
Accuracy, AC/DC Accuracy, Pulse Bandwidth Measuring input internal/external, switch able, BCD-output in multiplex mode, optically isolated Mains power supply Dimensions Weight
Technical specifications subject to change, HVM2003e.doc, 10/12
10 kV=/ 7kV 20 kV=/14kV 100V100% 2% 5V100% 1mA100% Average + Peak/2 - Peak/2 + Peak - Peak 1 % 5 Digit 1 % 5 Digit 2.5 % 5 Digit 0 - 10 kHz -1000 kHz Range: 19.99 / 199.9 15-way-D-connector 230 V/50 Hz/10 VA 485*90*300 mm 3 kg
Page 1/2
optional with:
Technical specification max. Input voltage : DC voltage AC voltage, 50 Hz Pulse voltage, 1.2/50s Dielectric Strength: AC voltage 50 Hz, 0.3 sec Pulse voltage, 1.2/50s Test voltage: 10 pulses 1.2/50 s, pos./neg, each High voltage impedance Nominal load of the low voltage arm of the divider Measuring cable with termination network
HVT 120 RCR 120 kV= 90 kV 200 kV 120 kV 300 kV 250 kV 270 M //100 pF+2k 1.0 M, 10-30 pF 10 m
HVT 240 RCR 240 kV= 180 kV 360 kV 250 kV 600 kVs 500 kV 360 M //180pF+3.6k 1.0 M, 10-30 pF 20 m long
Divider ratio (or customer specific) Short time stability Proximity effect, for wall distance > 2 * height of divider Temperature drift in the range of 5 - 40C Ambient temperature for normal operation Rise time Over shoot Dimensions: 35 ns 20% < 5% 1.2 m 360 mm ca. 25 kg
2000 : 1 1 % 0.1% 0.2% 100 ppm/K 17 C - 23 C 50 ns 20% < 5% 2.15 m 1200X1200 mm 4 wheels, 2 looka. ca. 40 kg
The dividers can also be used for high-voltage measurement of ac-, dc- and pulse voltages with an oscilloscope.
Technical specifications subject to change, HVM2003e.doc, 10/12 Page 2/2