Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Vacuumclave
HS-232 1V . HS-304 1V
ITS O.K WITH MINIMUM CYCLE TIME THAT FROM PRE-VACUUM STERILIZATION THROUGH POST-VACUUM DRYING !!
SAFETY
If power is turned off, the door would not open because of the interlock mechanism, the cycle would not start when the door is unlocked and the door would not open with some pressure within the chamber in any case. It has a door cover which enables to protect operator from thermal hazards and open/close mechanism which enables to operate the door with ease. The safety valve releases to protect the chamber from exceeding its rated pressure.
HS-3041V
07 HANSHIN
Vacuumclave
Tabletop Prevacuum STEAM STERILIZER
HS-2321V / HS-3041V
CYCLE GRAPH
It has five (5) kinds of basic cycle and two (2) kinds of test cycle programs. The sterilization time and drying time can be shortened, and the sterility assurance level can be achieved by performance of the vacuum system. Especially, the liquid cycle program of the basic cycles has a unique ability which can solve the problems caused when sterilizing the culture medium or the liquids with a small sterilizer at the clinical examination laboratory.
PRESSURE TEMPERATURE CYCLE No.1 WRAPPED
BOWIE-DICK TEST
CYCLE PROGRAM
PROGRAM
CYCLE NO. LOADS TYPE PRESSURE EXPOSURE TEMPERATURE EXPOSURE TIME DRY TIME COOLING TIME Max. Load HS-2321V HS-3041V
1 2.10 bar 134.0 15 min 30 min 2 1.10 bar 121.0 20 min 30 min -
BASIC CYCLE
3 2.10 bar 134.0 3 min 30 min 4 1.10 bar 121.0 15 min 30 min 5 Liquids 1.10 bar 121.0 40 min 60 min
Solid : 2 kg, Hollow : 2 kg Bottle : 250X 2ea Solid : 3 kg, Hollow : 3 kg Bottle : 500X 2ea
HANSHIN 14
HANSHIN MEDICAL
SPECIFICATIONS
MODEL
OVERALL SIZE CABINET
HS-2321V
Size : H405 X W516 X D714 mm Size : H360 X W516 X D586 mm Size : H225 X W225 X D425 mm Volume : 21.5 Material : SUS 316
HS-3041V
Size : H489 X W614 X D804 mm Size : H441 X W614 X D646 mm Size : H300 X W300 X D450 mm Volume : 40.5 Material : SUS 316
CHAMBER
PRESSURE VESSEL : Druckbehlterverordnung Gruppe I Max. Working Pressure : 2.80 bar / Min. Working Pressure : -0.90 bar Max. Working Temperature : 138.0 / Test Pressure : 3.00 bar Size : H175 X W230 X D125 mm, Water vol. : 3 , Material : SUS 316 Size : H44 X W216 X D405 mm Qty : 2ea 121.0 ~ 134.0 Saturated Steam of max. 2.10 bar Pressure Vacuum / Gravity Flow Rate : 11 /min Flow Rate : 34 /min Less than 1.0% of the amount of moisture within porous loads Min. filtration efficiency : 99.97% for 0.3 Particles AC 230 V, 50/60 Hz 1,600 W Microprocessor 160 X 128 Dot Graphic display, LCD (EL back-light) Model SPP-100LF/HSM : Thermal Printing, 203 X 406 dpi, W58 X 40mm Paper Roll 64 kg 74 kg 91 kg 104 kg 2,000 W Size : H44 X W291 X D432 mm Qty : 2ea
RESERVOIR TRAY EXPOSURE TEMPERATURE STERILIZING AGENT AIR REMOVAL VACUUM PUMP DRYING ABILITY AIR FILTER POWER SOURCE POWER CONSUMPTION CONTROLS DISPLAY PRINTER WEIGHT(Net) WEIGHT (Gross) ENVIRONMENT CONDITIONS
Altitudes up to 2,000 meters. Temperature range of +5 to +40 . Maximum relative humidity of 80 %. Main supply voltage fluctuation of 10 % of normal. Installation Category (Overvoltage Category) II, Pollution Degree 2.
HS-2321V
H405 X W516 X D714 mm 64 kg H821 X W590 X D660 mm 16 kg H1,445 X W1,190 X D1,655 mm Reservoir fill water : 3 kg Contains : 6.2 kg Max. load : 2 kg Water pressure test : 21.5 kg Weight per support area : 2,315 N/m2 (max. load) Weight per support area : 2,495 N/m2 (water press. Test) each 300 mm 400 mm
HS-3041V
H489 X W614 X D804 mm 91 kg H821 X W690 X D760 mm 22 kg H1,527 X W1,290 X D1,855 mm Reservoir fill water : 3 kg Contains : 11.8 kg Max. load : 3 kg Water pressure test : 40.5 kg Weight per support area : 2,465 N/m2 (max. load) Weight per support area : 2,725 N/m2 (water press. Test) each 300 mm 500 mm
15 HANSHIN