Sei sulla pagina 1di 11

SYSTEMATIC REVIEW

In-vitro orthodontic bond strength testing:


A systematic review and meta-analysis
Katrina J. Finnema,a Mutlu Özcan,b Wendy J. Post,c Yijin Ren,d and Pieter U. Dijkstrae
Groningen, The Netherlands, and Zürich, Switzerland

Introduction: The aims of this study were to systematically review the available literature regarding in-vitro
orthodontic shear bond strength testing and to analyze the influence of test conditions on bond strength.
Methods: Our data sources were Embase and Medline. Relevant studies were selected based on predefined
criteria. Study test conditions that might influence in-vitro bond strength were independently assessed by 2
observers. Studies reporting a minimum number of test conditions were included for meta-analysis by
using a multilevel model with 3 levels, with author as the highest level, study as the second level, and
specimens in the study as the lowest level. The primary outcome measure was bond strength. Results: We
identified 121 relevant studies, of which 24 were included in the meta-analysis. Methodologic drawbacks of
the excluded studies were generally related to inadequate reporting of test conditions and specimen
storage. The meta-analysis demonstrated that 3 experimental conditions significantly affect in-vitro bond
strength testing. Although water storage decreased bond strength on average by 10.7 MPa, each second
of photopolymerization time and each millimeter per minute of greater crosshead speed increased bond
strength by 0.077 and 1.3 MPa, respectively. Conclusions: Many studies on in-vitro orthodontic bond
strength fail to report test conditions that could significantly affect their outcomes. (Am J Orthod
Dentofacial Orthop 2010;137:615-22)

O
rthodontic bonding of brackets to teeth is a stan- possible to distinguish the adhesive potential of a specific
dard procedure to align teeth with fixed appli- bonding system independent of many other variables
ances. Orthodontic treatment with brackets that can influence either the quality or the longevity of
generally takes approximately 2 years. Bond failure of bracket bonding to enamel.1 In addition, the methodo-
brackets during this period retards treatment and is logic quality of in-vivo randomized controlled trials
costly in terms of time, material, and patient inconve- (RCTs) evaluating debonding and bracket failure is gener-
nience. Bracket debonding at the end of the treatment ally poor.2 Consequently, it is difficult to draw conclusions
should not damage the enamel. Hypothetically, in-vivo about the effectiveness of specific bonding systems and
testing in controlled trials is the best way to test the their effects on the enamel from in-vivo studies.
effectiveness of a bonding system and any detrimental In-vitro studies possibly allow for more standard-
effects on the enamel. However, clinically, it is almost im- ized procedures for testing a specific bonding system.
However, the various test conditions that are used ham-
a
Postgraduate student, Department of Orthodontics, University Medical Center per the comparison of their results.3,4 Test conditions
Groningen, University of Groningen, Groningen, The Netherlands. suggested to influence bond strength include enamel
b
Professor, University of Zürich, Dental Materials Unit, Center for Dental and origin (ie, bovine vs human), substrate storage (eg,
Oral Medicine, Clinic for Fixed and Removable Prosthodontics and Dental Ma-
terials Science, Zürich, Switzerland. physiologic saline solution or water), and pretreatment
c
Statistician, Department of Epidemiology, University Medical Center of the enamel surface (eg, grinding and means of
Groningen, University of Groningen, Groningen, The Netherlands.
d
cleaning).5-7 In addition, bond strength testing might
Professor and chair, Department of Orthodontics, University Medical Center
Groningen, University of Groningen, Groningen, The Netherlands. also be influenced by the specific test mode used (eg,
e
Professor, Center for Rehabilitation, Department of Oral and Maxillofacial Sur- tensile or shear testing).8 An explanation for the differ-
gery, School for Health Research, University Medical Center Groningen, Uni- ent outcomes between in-vitro bond strength studies
versity of Groningen, Groningen, The Netherlands.
The authors report no commercial, proprietary, or financial interest in the prod- might therefore be that bond strength is not being tested
ucts or companies described in this article. but, rather, an unknown combination of mechanical
Reprint requests to: Katrina J. Finnema, Department of Orthodontics, University properties and factors related to the test surfaces. Lately,
Medical Center Groningen, University of Groningen, Hanzeplein 1, PO Box
30.001, 9700 RB Groningen, The Netherlands; e-mail, k.j.finnema@dmo. more attention has been given to the various test condi-
umcg.nl. tions and their effects on the results. Recent studies have
Submitted, July 2009; revised and accepted, December 2009. evaluated the influence of some of these factors on bond
0889-5406/$36.00
Copyright Ó 2010 by the American Association of Orthodontists. strength, including force location,9 tooth type,10-12
doi:10.1016/j.ajodo.2009.12.021 crosshead speed variations,13,14 and loading mode.8
615
616 Finnema et al American Journal of Orthodontics and Dentofacial Orthopedics
May 2010

At present, there is no overview on bracket bond Table I. Search strategy in Medline and Embase
strength from which general conclusions can be drawn. Search Literature search strategy
Because of the lack of standardization, the growing
number of in-vitro studies being published can only 1 Explode "orthodontic-brackets"/MeSH
be evaluated individually. The aims of this study were all subheadings
2 Bracket*
to systematically review the literature regarding in- 3 Fixed applian*
vitro bond strength and failure mode of the most fre- 4 #1 or #2 or #3
quently used clinical bonding systems and to analyze 5 Explode "dental-bonding"/MeSH all subheadings
by meta-regression the influence of test conditions on 6 Bond*
the bond strength measured. 7 #5 or #6
8 #4 and #7
9 Explode "composite-resins"/MeSH all subheadings
10 Explode "compomers"/MeSH all subheadings
MATERIAL AND METHODS
11 Explode "glass-ionomer-cements"/MeSH
Relevant studies were identified in a literature all subheadings
search and subsequently selected on the basis of inclu- 12 Composite resin
13 Glass ionomer
sion criteria. Studies fulfilling the inclusion criteria
14 Compomer
were assessed for reporting test conditions that could in- 15 #8 and (#9 or #12)
fluence the results of in-vitro bond strength testing. The 16 #8 and (#10 or #14)
studies reporting a minimal number of test conditions 17 #8 and (#11 or #13)
were included for meta-analysis. 18 #8 and bond strength
19 Search Medline/Embase: #15 or #16 or #17 or #18
To identify studies related to in-vitro bond strength,
a search was performed in the databases of Medline #, Search; MeSH, medical subjects heading (a thesaurus word); *, trun-
(1967 to December 2007) and Embase (1950 to Decem- cation of a text word.
ber 2007) (Table I). References of identified studies and
relevant review articles were searched for additional the reporting of the 27 experimental conditions of
studies missed in the initial search. English was the lan- each study in a consensus meeting.
guage restriction.
Initially, the titles and abstracts of the studies iden- Studies included in the meta-analysis
tified in the literature search were prescreened (by The 2 observers independently determined the most
K.J.F.) for relevance to the topic of this study (in-vitro relevant experimental conditions for in-vitro bond
orthodontic shear bond strength testing). The full text strength studies based on the results from previous stud-
of each possibly relevant study was retrieved and as- ies. In a consensus meeting, agreement was reached on
sessed by 2 reviewers (M.Ö., K.J.F.) for inclusion and these required experimental conditions. Studies were
detailed assessment of the experimental conditions. included in the meta-analysis if at least all of the fol-
The Figure outlines the algorithm of the study selection lowing experimental conditions were reported: storage
procedure. solution of teeth,15 cleaning of enamel,16 bracket
Studies regarding bond strength testing were se- type,17-19 etchant type,20 etching time,21 adhesive
lected for detailed assessment of the experimental con- type, photopolymerization device,22 total photopoly-
ditions if they met the following criteria: in-vitro merization time,23 specimen storage time,24 crosshead
investigation, with the shear bond strengths of metal speed,9,14 force location on bracket,13 and blade design
brackets evaluated and expressed in megapascals of the jig of the universal testing machine.25
(MPa), and the sound buccal enamel of human premo- The 2 observers independently extracted the data.
lars used. Case reports, abstracts, letters, and narrative Consensus was reached after discussion in case of dis-
reviews were excluded. agreement.
A list of 27 items was used to assess the experimen- First, for each of the 27 experimental conditions, the
tal conditions, each reflecting an experimental condition number and percentage of included studies describing
that influences the results of in-vitro bond strength test- this specific item were calculated (Table II). Subse-
ing (Table II).3,4 Before the assessment of the studies, 2 quently, for each study included in the meta-analysis,
observers (M.Ö. and K.J.F.) discussed all 27 the following data were presented: thymol storage
experimental conditions to reach consensus about solution (yes/no), fluor-free cleaning (yes/no), mesh
their content. They independently assessed whether brackets (yes/no), phosphoric acid etching (yes/no), etch-
the experimental conditions were reported in the ing time (\30 seconds/$30 seconds), photopolymeriza-
study. After the assessment, the observers agreed on tion time (seconds), photopolymerized composite (yes/
American Journal of Orthodontics and Dentofacial Orthopedics Finnema et al 617
Volume 137, Number 5

Potentially relevant studies Table II. Experimental conditions assessed in the 121
identified and screened for articles
retrieval (n = 918)
Number (%) of studies
adequately reporting
Studies excluded, with
experimental
reasons (n = 752) Experimental condition condition
- not relevant for topic
Studies retrieved for more
detailed evaluation (n = 166) 1. Substrate origin* 121 (100)
2. Type of teeth* 121 (100)
3. Storage time before bonding 38 (31)
Studies excluded, with
4. Storage temperature before bonding 38 (31)
reasons (n = 45)
5. Storage solution before bonding† 108 (89)
- did not meet with
Potentially appropriate studies inclusion criteria 6. Cleaning of specimens† 113 (93)
to be included in the 7. Bracket material* 121 (100)
meta-analysis (n = 121) 8. Type of bracket† 94 (78)
9. Type of etchant† 111 (92)
Studies excluded, with 10. Time of etching† 109 (90)
reasons (n = 93) 11. Adhesive type† 119 (98)
- did not meet threshold 12. Amount of force at bracket placement 18 (15)
Studies included in meta-
analysis (n = 28) 13. Light device type† 75 (62)
14. Total polymerization time† 84 (69)
Studies withdrawn, by
outcome, with reasons 15. Light directions 65 (54)
(n = 4) 16. Sample storage time† 109 (90)
Studies with usable - overlapping data 17. Sample storage solution 103 (85)
information, by outcome or missing data 18. Sample storage temperature 97 (80)
(n = 24) 19. Thermocycling 26 (22)
20. Testing machine 119 (98)
21. Shear testing as test method 121 (100)
Fig. Algorithm of study selection procedure. 22. Crosshead speed† 117 (97)
23. Force location on bracket† 83 (69)
24. Blade design† 73 (60)
25. ARI 93 (77)
no), halogen light (yes/no), water storage (yes/no), stor-
26. Magnification used in determining ARI 70 (58)
age time (hours), thermocycling (yes/no), crosshead 27. Bond strength in MPa* 121 (100)
speed (millimeters per minute), force location on the
tooth-bracket interface (yes/no), and shearing blade *Studies reporting experimental conditions numbers 1, 2, 7 and 27
were included in the systematic review. †Studies reporting experimen-
(yes/no). Of the latter data, the influence on the main out- tal conditions numbers 5, 6, 8, 9, 10, 11, 13, 14, 16, 22, 23 and 24,
come variable (bond strength in megapascals) was evalu- together with those conditions marked *, were selected for meta-
ated in the meta-analysis. Failure mode reported with the analysis.
adhesive remnant index (ARI) was also scored when
reported and is presented as the percentage of specimens with the author. In this way, the correlation within au-
with all adhesive left on the enamel.26 thors between studies was taken into account. Residual
variance was entered in the multilevel model by using
Statistical analysis the within-study variance published in each study.
Different models were compared by using the
Statistical analyses were performed by using the
change in deviance (–2 log likelihood), a likelihood ra-
Statistical Package for the Social Sciences (version
tio test (chi-square distributed). In this way, it was also
16.0, SPSS, Chicago, Ill). The interobserver agreement
possible to test the statistical heterogeneity between
with respect to the reporting of experimental conditions
authors and between articles by making different as-
of the included studies before the consensus meeting
sumptions about (models for) the covariance structure.
was expressed as the Cohen kappa. For descriptive
Effects with P values smaller than 0.05 were considered
statistics, means (standard deviations or medians) and
significant.
interquartile ranges in skewed distributions are re-
ported. In MLwin (version 2.02, Centre for Multilevel
Modelling, Bristol, United Kingdom), a 3-level analysis RESULTS
(random-effects model) was performed. This is a meta- The searches of Medline and Embase yielded 918
regression analysis. The lowest level corresponded to publications. After the first assessment, 166 studies
the specimen level (specimens in studies), the middle were judged to be relevant for this systematic review.
level with the study, and the highest level corresponded We found no additional studies by checking the
618 Finnema et al American Journal of Orthodontics and Dentofacial Orthopedics
May 2010

references of the included studies and relevant review Table III. Multilevel meta-analysis with bond strength
articles. After we used the specified criteria, 121 of (MPa) as the dependent variable
the 166 studies regarding bond strength testing were in- Lower Upper
cluded for detailed assessment of the experimental con- Independent variable Beta (SE) 95% CI 95% CI
ditions (Fig). Interobserver agreement (Cohen kappa)
for reporting the 27 experimental conditions of the in- Water storage –10.648 (3.541) –17.730 –3.566
(no, 0; yes,1)
cluded studies was 0.86. Disagreements were generally Polymerization time 0.077 (0.030) 0.017 0.137
caused by differences in language interpretation and (per second)
were resolved in the consensus meeting. Crosshead speed 1.302 (0.599) 0.104 2.500
On average, the 121 included studies reported (mm/min)
Constant 20.014 (3.452) 13.110 26.918
a mean of 20.4 (SD, 2.8) experimental conditions with
a minimum of 1227 and a maximum of 26.28,29 The
most poorly reported item was the experimental ported in 57 groups; in 8 experimental groups, artificial
condition of amount of force at bracket placement. saliva was used for storing the test specimens.28 The
This was reported in 18 (15%) of the 121 studies. The storage times of the specimens ranged from 0 to 672
most relevant experimental conditions for in-vitro hours (median, 24; interquartile range, 0-48). Thermo-
bond strength were reported in the following percentage cycling was explicitly reported in 15 of the 65 groups.
of studies: adhesive type, 98%; crosshead speed, 97%; The crosshead speeds when removing the bracket varied
cleaning of enamel, 93%; etchant type, 92%; etching from 0.1 to 5.0 mm per minute (median, 0.5; interquar-
time, 90%; specimen storage time, 90%; storage solu- tile range, 0.5-1.0). In 23 studies, the force location was
tion of teeth before bonding, 89%; bracket type, 78%; at the bracket-enamel interface. In 50 groups, a shearing
total polymerization time, 69%; force location on blade was used for debonding the brackets. In the re-
bracket, 69%; photopolymerization device, 62%; and maining 15 groups, a wire loop was used for the same
blade design, 60% (Table II). purpose. Based on the diversity in reported test condi-
tions, we concluded that there was considerable clinical
Studies included for meta-analysis heterogeneity in these studies. Bond strengths ranged
from 3.5 to 27.8 MPa (mean, 13.4; SD, 5.7). ARI scores
By using the threshold value, 28 studies were in- were reported in 57 experimental groups. Specimens
cluded in the meta-analysis. In 3 of the 9 studies from with all adhesive left on the enamel after bond strength
the same authors, overlapping study data were re- testing varied from 0% to 90%.
ported.30-32 These duplicated data were not entered in
the meta-analysis. One other study was excluded from Meta-analysis
the meta-analysis because the standard deviation of
the main outcome measure (bond strength in megapas- Table III summarizes the results of the meta-
cals) was not reported.33 As a result, 24 studies were in- analysis. Heterogeneity between authors and studies
cluded in the meta-analysis (Appendix).28,34-56 From was still significant after entering the predictor variables
these studies, data regarding experimental groups with (P \0.05). The results of bond strength testing were
a photopolymerized adhesive and nonself-ligating negatively influenced when the teeth were stored in wa-
metal brackets were extracted for further analyses. ter. Water storage on average decreased bond strength
This yielded 65 experimental groups that tested bond by 10.7 MPa, assuming that the other predictors remain
strength in specimen groups ranging from 5 to 40 pre- constant. Analogously, each second of photopolymeri-
molars (mean, 14.1; SD, 7.4) (Appendix). Extracted zation time increased the bond strength by 0.077 MPa;
teeth were stored in distilled water with thymol in 30 ex- when crosshead speed increased by 1 mm per minute,
perimental groups. Fluor-free cleaning was explicitly bond strength increased by 1.3 MPa.
reported in 32 groups. In 58 groups, a metal bracket
with a mesh base was used. Phosphoric acid etching DISCUSSION
was used in 47 groups; in 48 experimental groups, the There was great diversity in the experimental condi-
etching time was $30 seconds. The total polymeriza- tions of studies reporting bond strength testing in ortho-
tion times in the 65 groups varied from 2 to 60 seconds dontics. The results from the meta-analysis in this
(mean, 25.3; SD, 14.8). systematic review indicate that the experimental condi-
A photopolymerized composite was used in 59 tions of water storage, photopolymerization time, and
groups, and a halogen polymerization device was used crosshead speed significantly influenced the results of
to cure the adhesives in 48 groups. Water storage was re- in-vitro bond strength testing.
American Journal of Orthodontics and Dentofacial Orthopedics Finnema et al 619
Volume 137, Number 5

As previously shown in 2 other reviews evaluating The third experimental condition shown to signifi-
orthodontic bond strength studies, there is still great di- cantly affect outcomes of bond strength testing was
versity in test protocols and quality of these studies.3,4 crosshead speed. An increase in crosshead speed of 1
The observed heterogeneity between the studies in mm per minute yielded an increase in average bond
this meta-analysis was clinically and statistically large. strength of 1.3 MPa. The opposite effect was demon-
Not one study described all 27 experimental conditions. strated in 2 previous studies in which increases in cross-
This finding might relate to the fact that some experi- head speed from 0.5 to 5.0 mm per minute and from 1 to
mental conditions—eg, the ARI and thermocycling— 200 mm per minute, respectively, were associated with
were not used in all studies. However, when evaluating significant decreases of in-vitro bond strength.14,59 This
the 121 studies for the most relevant experimental con- phenomenon was suggested to relate to the induction of
ditions, only 28 fulfilled the threshold value. This find- a stiff body response and the elimination of the
ing indicates that most of these in-vitro studies did not viscoelastic properties of the adhesive.59 In another
properly report important confounding factors that af- study, no effect was observed on bond strength with
fect bond strength outcomes. When we finally included crosshead speed variations between 0.1 and 5 mm per
24 of these 28 studies in a meta-analysis, water storage minute.9 In the studies included in this meta-analysis,
of the bonded specimens, photopolymerization time, crosshead speeds ranged from 0.1 to 5 mm per minute
and crosshead speed were shown to be the variables with most using a speed of 0.5 mm per minute
that primarily affected the bond strength outcomes. (Appendix). We have no obvious explanation for the
Water storage decreased bond strength on average discrepancy of our results with those of previous stud-
by 10.7 MPa. Although this was the most pronounced ies; unknown confounders might be responsible.
effect of an experimental condition on in-vitro bond Bond strength values reported in the 24 studies in
strength outcomes, this finding was mainly influenced this meta-analysis ranged from 3.5 to 27.8 MPa. Clini-
by 1 relatively large study sample in which artificial sa- cal implications of in-vitro bond strength values are
liva was used as a storage medium for specimens.28 generally based on the recommendation in a review ar-
Most in-vitro bond strength studies used distilled water ticle from 197560; a ‘‘clinically acceptable’’ value was
for storing the specimens, but 11% of the studies did not defined as an in-vitro bond strength of 6 to 8 MPa. Ac-
report the storage medium. It was previously reported cording to this criterion, 53 of the 65 experimental
that specimen storage in artificial saliva reduces bond groups from our meta-analysis should be considered
strength similarly to the effect of water degradation.57 to have clinically acceptable bond strength values. Be-
Although our study indicates that distilled water has cause it has never actually been tested whether 6 to 8
a different effect on bond strength than artificial saliva, MPa is a sufficient in-vitro bond strength for clinical
future research on the effects of different storage media use, the use of this reference value has been criticized
on bond strength is needed. before.4,61 Since the publication of this reference
The second experimental condition that we found to value, various materials have been used, and the
significantly affect in-vitro bond strength was photopo- effects of various test conditions (eg, pH and
lymerization time. Each additional second of photopoly- temperature variations) have been implicated in the
merization increased bond strength by 0.077 MPa. It was aging of composite resins.61 This implies that interpre-
previously suggested that photopolymerization time has tation of bond strength data should be limited to the rel-
a greater influence on in-vitro bond strength than the ative effectiveness of the adhesives used in a study.
type of photopolymerization device.23,58 The studies in Extrapolation of absolute values and comparing them
this meta-analysis showed considerable variations in with a supposedly ‘‘clinically acceptable’’ reference
photopolymerization time: from 2 to 50 seconds. More- value should be avoided.62
over, 31% of the included studies did not even report Since shear bond strength testing is the most com-
polymerization time (Appendix). Most remaining stud- monly used method for debonding brackets, only stud-
ies used 40 seconds for polymerizing the adhesive; this ies that used this method were included in our review.
corresponds to the routine clinical standard. The fact With respect to shear bond strength testing, there are
that the majority of these studies used a halogen device also some variations in blade design. Most studies
for polymerizing the adhesive most likely explains used a shearing blade for debonding the brackets,
why this polymerization time was used. The results whereas in some studies a wire loop was used.40-43,45-47
from our meta-analysis, however, suggest that a longer Debonding brackets with a wire loop is not a true
polymerization time yields higher bond strengths, form of shear bond strength testing, since it also
although the most optimal time for polymerizing the incorporates a component of tensile stress. Although
adhesive cannot be deduced from our results. blade design could not be identified as an
620 Finnema et al American Journal of Orthodontics and Dentofacial Orthopedics
May 2010

experimental condition that significantly affected in- strength of orthodontic brackets. Experimental condi-
vitro bond strength results in this meta-analysis, differ- tions that significantly influence in-vitro bond strength
ent shearing blades could have an effect. This should be are water storage of the bonded specimens, photopoly-
substantiated by additional studies. merization time, and crosshead speed. Based on the re-
Theoretically, in-vitro studies determine the true sults from this systematic review, we concluded that
strength of a given bonding system to the enamel sub- many studies did not properly report test conditions
strate. Unfortunately, in-vitro studies have not been suc- that could have significantly affected the outcomes. Be-
cessful in predicting in-vivo effectiveness.63,64 An cause of developments in adhesive dentistry and the in-
accurate simulation of the clinical situation seems creasing numbers of bond strength studies, uniform
necessary to obtain clinically relevant results from in- guidelines for standardization of the experimental con-
vitro experiments.4 However, because of the many con- ditions of in-vitro bond strength research is clearly
ditions involved in the in-vivo situation, an accurate indicated.
simulation is at present an unrealistic goal. Although
in-vitro bond strength testing is valuable for initial REFERENCES
screening and selection of materials, it cannot be re-
1. Leloup G, D’Hoore W, Bouter D, Degrange M, Vreven J. Meta-
garded as a substitute for in-vivo testing. Orthodontic analytical review of factors involved in dentin adherence. J Dent
materials that perform well in in-vitro experiments Res 2000;80:1605-14.
should always be tested with in-vivo RCTs. 2. Mandall NA, Millett DT, Mattick CR, Hickman J,
Although we systematically reviewed the current Worthington HV, Macfarlane TV. Orthodontic adhesives: a sys-
tematic review. J Orthod 2002;29:205-10.
literature on in-vitro shear bond strength testing, our
3. Fox NA, McCabe JF, Buckley JG. A critique of bond strength test-
study has some possible limitations. Selection bias ing in orthodontics. Br J Orthod 1994;21:33-3.
could have resulted in inappropriate inclusion or exclu- 4. Eliades T, Brantley WA. The inappropriateness of conventional
sion of studies based on factors other than the inclusion orthodontic bond strength assessment protocols. Eur J Orthod
criteria. This chance was minimized by having 2 ob- 2000;22:13-23.
5. Oesterle LJ, Shellhart WC, Belanger GK. The use of bovine
servers independently assess the articles. Another short-
enamel in bonding studies. Am J Orthod Dentofacial Orthop
coming might have been our language restriction. 1998;114:514-9.
Moreover, only 2 databases were searched. Some rele- 6. Mühleman HR. Storage medium and enamel hardness. Helvetica
vant studies might have been missed by doing so. How- Odontologica Acta 1964;8:112-7.
ever, we believe that our conclusion of poor descriptions 7. Eliades T, Viazis AD, Eliades G. Bonding of ceramic brackets to
enamel: morphologic and structural considerations. Am J Orthod
of test conditions in most in-vitro bond strength studies
Dentofacial Orthop 1991;99:369-75.
would not be altered by studies that were possibly 8. Katona TR, Long RW. Effect of loading mode on bond strength of
missed because of language and database restrictions. orthodontic brackets bonded with 2 systems. Am J Orthod Dento-
This is substantiated, since 93 (77%) of the 121 poten- facial Orthop 2006;129:60-4.
tially appropriate studies were excluded from the 9. Klocke A, Kahl-Nieke B. Influence of crosshead speed in ortho-
dontic bond strength testing. Dent Mater 2005;21:139-44.
meta-analysis because test conditions were not ade-
10. Oztürk B, Malkoç S, Koyutürk AE, Catalbas B, Ozer F. Influence
quately reported. Finally, this is a systematic review of different tooth types on the bond strength of two orthodontic
concerning in-vitro observational studies. Conclusions adhesive systems. Eur J Orthod 2008;30:407-12.
in systematic reviews and meta-analyses are preferably 11. Hobson RS, McCabe JF, Hogg SD. Bond strength to surface
based on results from RCTs. No RCTs were available enamel for different tooth types. Dent Mater 2001;17:184-9.
12. Linklater RA, Gordon PH. An ex vivo study to investigate bond
for the topic of this systematic review and meta-
strengths of different tooth types. J Orthod 2001;28:59-65.
analysis. Selection bias, information bias, and unknown 13. Klocke A, Kahl-Nieke B. Influence of force location in orthodon-
confounders threatened the validity of each study that tic shear bond strength testing. Dent Mater 2005;21:391-6.
we evaluated. In future in-vitro studies, bracket debond- 14. Bishara SE, Soliman M, Laffoon J, Warren JJ. Effect of changing
ing should be evaluated more carefully by considering a test parameter on the shear bond strength of orthodontic
brackets. Angle Orthod 2005;75:832-5.
the items studied in this review. With this approach,
15. Titley K, Caldwell R, Kulkarni G. Factors that affect the shear
the most dominant factor affecting bracket adhesion in bond strength of multiple component and single bottle adhesives
vitro might also be identified and correlated with the to dentin. Am J Dent 2003;16:120-4.
clinical situation. 16. Lindauer SJ, Browning H, Shroff B, Marshall F, Anderson RH,
Moon PC. Effect of pumice prophylaxis on the bond strength of
orthodontic brackets. Am J Orthod Dentofacial Orthop 1997;
CONCLUSIONS 111:599-605.
17. Bishara SE, Soliman MM, Oonsombat C, Laffoon JF, Ajlouni R.
In this systematic review and meta-analysis, a sum- The effect of variation in mesh-base design on the shear bond
mary of factors is given that can affect the in-vitro bond strength of orthodontic brackets. Angle Orthod 2004;74:400-4.
American Journal of Orthodontics and Dentofacial Orthopedics Finnema et al 621
Volume 137, Number 5

18. Sharma-Sayal SK, Rossouw PE, Kulkarni GV, Titley KC. The in- 37. Martin S, Garcia-Godoy F. Shear bond strength of orthodontic
fluence of orthodontic bracket base design on shear bond strength. brackets cemented with a zinc oxide-polyvinyl cement. Am J Or-
Am J Orthod Dentofacial Orthop 2003;124:74-82. thod Dentofacial Orthop 1994;106:615-20.
19. Knox J, Hubsch P, Jones ML, Middleton J. The influence of 38. McCourt JW, Cooley RL, Barnwell S. Bond strength of light-cure
bracket base design on the strength of the bracket-cement inter- fluoride-releasing base-liners as orthodontic bracket adhesives.
face. J Orthod 2000;27:249-54. Am J Orthod Dentofacial Orthop 1991;100:47-52.
20. De Munck J, Van Landuyt K, Peumans M, Poitevin A, 39. Northrup RG, Berzins DW, Bradley TG, Schuckit W. Shear bond
Lambrechts P, Braem M, et al. A critical review of the durability strength comparison between two orthodontic adhesives and
of adhesion to tooth tissue: methods and results. J Dent Res 2005; self-ligating and conventional brackets. Angle Orthod 2007;77:
84:118-32. 701-6.
21. Gardner A, Hobson R. Variations in acid-etch patterns with differ- 40. Romano FL, Tavares SW, Nouer DF, Consani S, Borges de Araújo
ent acids and etch times. Am J Orthod Dentofacial Orthop 2001; Magnani MB. Shear bond strength of metallic orthodontic
120:64-7. brackets bonded to enamel prepared with self-etching primer. An-
22. Niepraschk M, Rahiotis C, Bradley TG, Eliades T, Eliades G. Ef- gle Orthod 2005;75:849-53.
fect of various curing lights on the degree of cure of orthodontic 41. Sayinsu K, Isik F, Sezen S, Aydemir B. New protective polish ef-
adhesives. Am J Orthod Dentofacial Orthop 2007;132:382-4. fects on shear bond strength of brackets. Angle Orthod 2006;76:
23. Eliades T. Orthodontic materials research and applications: part 1. 306-9.
Current status and projected future developments in bonding and 42. Sayinsu K, Isik F, Sezen S, Aydemir B. Light curing the primer—
adhesives. Am J Orthod Dentofacial Orthop 2006;130:445-51. beneficial when working in problem areas? Angle Orthod 2006;
24. Yamamoto A, Yoshida T, Tsubota K, Takamizawa T, 76:310-3.
Kurokawa H, Miyazaki M. Orthodontic bracket bonding: enamel 43. Schaneveldt S, Foley TF. Bond strength comparison of moisture-
bond strength vs time. Am J Orthod Dentofacial Orthop 2006;130: insensitive primers. Am J Orthod Dentofacial Orthop 2002;122:
435. e1-6. 267-73.
25. Mojtahedzadeh F, Akhoundi MS, Noroozi H. Comparison of wire 44. Scougall Vilchis RJ, Yamamoto S, Kitai N, Hotta M,
loop and shear blade as the 2 most common methods for testing Yamamoto K. Shear bond strength of a new fluoride-releasing or-
orthodontic shear bond strength. Am J Orthod Dentofacial Orthop thodontic adhesive. Dent Mater J 2007;26:45-51.
2006;130:385-7. 45. Signorelli MD, Kao E, Ngan PW, Gladwin MA. Comparison of
26. Årtun J, Bergland S. Clinical trials with crystal growth condition- bond strength between orthodontic brackets bonded with halogen
ing as an alternative to acid-etch enamel pretreatment. Am J Or- and plasma arc curing lights: an in-vitro and in-vivo study. Am
thod 1984;85:333-40. J Orthod Dentofacial Orthop 2006;129:277-82.
27. Sehgal V, Shetty VS, Mogra S, Bhat G, Eipe M, Jacob S, et al. 46. Sunna S, Rock WP. An ex vivo investigation into the bond
Evaluation of antimicrobial and physical properties of orthodontic strength of orthodontic brackets and adhesive systems. Br J Or-
composite resin modified by addition of antimicrobial agents—an thod 1999;26:47-50.
in-vitro study. Am J Orthod Dentofacial Orthop 2007;131:525-9. 47. Tecco S, Traini T, Caputi S, Festa F, de Luca V, D’Attilio M. A
28. Chang WG, Lim BS, Yoon TH, Lee YK, Kim CW. Effects of new one-step dental flowable composite for orthodontic use: an
salicylic-lactic acid conditioner on the shear bond strength of in-vitro bond strength study. Angle Orthod 2005;75:672-7.
brackets and enamel surfaces. J Oral Rehabil 2005;32:287-95. 48. Usxümez S, Büyükyilmaz T, Karaman AI. Effect of light-emitting
29. Tecco S, Traini T, Caputi S, Festa F, de Luca V, D’Attilio M. A diode on bond strength of orthodontic brackets. Angle Orthod
new one-step dental flowable composite for orthodontic use: an 2004;74:259-63.
in-vitro bond strength study. Angle Orthod 2005;75:672-7. 49. Uysal T, Basciftci FA, Us xümez S, Sari Z, Buyukerkmen A. Can
30. Vicente A, Bravo LA, Romero M, Ortız AJ, Canteras M. Bond previously bleached teeth be bonded safely? Am J Orthod Dento-
strength of brackets bonded with an adhesion promoter. Br Dent facial Orthop 2003;123:628-32.
J 2004;196:482-5. 50. Vicente A, Bravo LA, Romero M, Ortiz AJ, Canteras M. Adhe-
31. Vicente A, Bravo LA, Romero M, Ortız AJ, Canteras M. Effects sion promoters: effects on the bond strength of brackets. Am
of 3 adhesion promoters on the shear bond strength of orthodontic J Dent 2005;18:323-6.
brackets: an in-vitro study. Am J Orthod Dentofacial Orthop 2006; 51. Vicente A, Bravo LA, Romero M, Ortız AJ, Canteras M. Shear
129:390-5. bond strength of orthodontic brackets bonded with self-etching
32. Vicente A, Bravo LA, Romero M. Self-etching primer and a non- primers. Am J Dent 2005;18:256-60.
rinse conditioner versus phosphoric acid: alternative methods for 52. Vicente A, Bravo LA, Romero M. Influence of a nonrinse condi-
bonding brackets. Eur J Orthod 2006;28:173-8. tioner on the bond strength of brackets bonded with a resin adhe-
33. Aljubouri YD, Millett DT, Gilmour WH. Laboratory evaluation of sive system. Angle Orthod 2005;75:400-5.
a self-etching primer for orthodontic bonding. Eur J Orthod 2003; 53. Vicente A, Bravo LA, Romero M, Ortiz AJ, Canteras M. A
25:411-5. comparison of the shear bond strength of a resin cement and
34. Buyukyilmaz T, Usumez S, Karaman AI. Effect of self-etching two orthodontic resin adhesive systems. Angle Orthod 2005;
primers on bond strength—are they reliable? Angle Orthod 75:109-13.
2003;73:64-70. 54. Vicente A, Navarro R, Mena A, Bravo LA. Effect of surface treat-
35. Kim MJ, Lim BS, Chang WG, Lee YK, Rhee SH, Yang HC. Phos- ments on the bond strength of brackets bonded with a compomer.
phoric acid incorporated with acidulated phosphate fluoride gel Am J Dent 2006;19:271-4.
etchant effects on bracket bonding. Angle Orthod 2005;75: 55. Vicente A, Bravo LA. Direct bonding with precoated brackets and
678-84. self-etching primers. Am J Dent 2006;19:241-4.
36. Klocke A, Korbmacher HM, Huck LG, Kahl-Nieke B. Plasma arc 56. Wendl B, Droschl H. A comparative in-vitro study of the strength
curing lights for orthodontic bonding. Am J Orthod Dentofacial of directly bonded brackets using different curing techniques. Eur
Orthop 2002;122:643-8. J Orthod 2004;26:535-44.
622 Finnema et al American Journal of Orthodontics and Dentofacial Orthopedics
May 2010

57. Kitasako Y, Burrow MF, Nikaido T, Tagami J. The influence of 61. Eliades T. Comparing bond strengths. Am J Orthod Dentofacial
storage solution on dentin bond durability of resin cement. Dent Orthop 2002;122(6):13-15A.
Mater 2000;16:1-6. 62. Eliades T, Bourauel C. Intraoral aging of orthodontic materials:
58. Ilie N, Hickel R. Correlation between ceramics translucency and the picture we miss and its clinical relevance. Am J Orthod Den-
polymerization efficiency through ceramics. Dent Mater 2008;24: tofacial Orthop 2005;127:403-12.
908-14. 63. Murray SD, Hobson RS. Comparison of in-vivo and in-vitro
59. Eliades T, Katsavrias E, Zinelis S, Eliades G. Effect of loading shear bond strength. Am J Orthod Dentofacial Orthop 2003;123:2-9.
rate on bond strength. J Orofac Orthop 2004;65:336-42. 64. Sunna S, Rock WP. Clinical performance of orthodontic brackets
60. Reynolds IR. A review of direct orthodontic bonding. Br J Orthod and adhesive systems: a randomized clinical trial. Br J Orthod
1975;2:171-8. 1998;25:283-7.
American Journal of Orthodontics and Dentofacial Orthopedics Finnema et al 622.e1
Volume 137, Number 5

APPENDIX REFERENCES 13. Signorelli MD, Kao E, Ngan PW, Gladwin MA. Comparison of
1. Buyukyilmaz T, Usumez S, Karaman AI. Effect of self-etching bond strength between orthodontic brackets bonded with halogen
and plasma arc curing lights: an in-vitro and in-vivo study. Am
primers on bond strength—are they reliable? Angle Orthod
2003;73:64-70. J Orthod Dentofacial Orthop 2006;129:277-82.
2. Chang WG, Lim BS, Yoon TH, Lee YK, Kim CW. Effects of 14. Sunna S, Rock WP. An ex vivo investigation into the bond
strength of orthodontic brackets and adhesive systems. Br J Orthod
salicylic-lactic acid conditioner on the shear bond strength of
brackets and enamel surfaces. J Oral Rehabil 2005;32:287-95. 1999;26:47-50.
3. Kim MJ, Lim BS, Chang WG, Lee YK, Rhee SH, Yang HC. Phos- 15. Tecco S, Traini T, Caputi S, Festa F, de Luca V, D’Attilio M. A
phoric acid incorporated with acidulated phosphate fluoride gel new one-step dental flowable composite for orthodontic use: an
in-vitro bond strength study. Angle Orthod 2005;75:672-7.
etchant effects on bracket bonding. Angle Orthod 2005;75:678-84.
4. Klocke A, Korbmacher HM, Huck LG, Kahl-Nieke B. Plasma arc 16. Usxümez S, Büyükyilmaz T, Karaman AI. Effect of light-emitting
curing lights for orthodontic bonding. Am J Orthod Dentofacial diode on bond strength of orthodontic brackets. Angle Orthod
Orthop 2002;122:643-8. 2004;74:259-63.
17. Uysal T, Basciftci FA, Us xümez S, Sari Z, Buyukerkmen A. Can
5. Martin S, Garcia-Godoy F. Shear bond strength of orthodontic
brackets cemented with a zinc oxide-polyvinyl cement. Am previously bleached teeth be bonded safely? Am J Orthod Dento-
J Orthod Dentofacial Orthop 1994;106:615-20. facial Orthop 2003;123:628-32.
6. McCourt JW, Cooley RL, Barnwell S. Bond strength of light-cure 18. Vicente A, Bravo LA, Romero M, Ortiz AJ, Canteras M.
Adhesion promoters: effects on the bond strength of brackets.
fluoride-releasing base-liners as orthodontic bracket adhesives.
Am J Orthod Dentofacial Orthop 1991;100:47-52. Am J Dent 2005;18:323-6.
7. Northrup RG, Berzins DW, Bradley TG, Schuckit W. Shear bond 19. Vicente A, Bravo LA, Romero M, Ortız AJ, Canteras M. Shear
bond strength of orthodontic brackets bonded with self-etching
strength comparison between two orthodontic adhesives and self-
ligating and conventional brackets. Angle Orthod 2007;77:701-6. primers. Am J Dent 2005;18:256-60.
8. Romano FL, Tavares SW, Nouer DF, Consani S, Borges de Araújo 20. Vicente A, Bravo LA, Romero M. Influence of a nonrinse condi-
Magnani MB. Shear bond strength of metallic orthodontic tioner on the bond strength of brackets bonded with a resin
adhesive system. Angle Orthod 2005;75:400-5.
brackets bonded to enamel prepared with self-etching primer.
Angle Orthod 2005;75:849-53. 21. Vicente A, Bravo LA, Romero M, Ortiz AJ, Canteras M.
9. Sayinsu K, Isik F, Sezen S, Aydemir B. New protective polish effects A comparison of the shear bond strength of a resin cement
on shear bond strength of brackets. Angle Orthod 2006;76:306-9. and two orthodontic resin adhesive systems. Angle Orthod
2005;75:109-13.
10. Sayinsu K, Isik F, Sezen S, Aydemir B. Light curing the primer–
beneficial when working in problem areas? Angle Orthod 2006; 22. Vicente A, Navarro R, Mena A, Bravo LA. Effect of surface treat-
76:310-3. ments on the bond strength of brackets bonded with a compomer.
Am J Dent 2006;19:271-4.
11. Schaneveldt S, Foley TF. Bond strength comparison of moisture-
insensitive primers. Am J Orthod Dentofacial Orthop 2002;122: 23. Vicente A, Bravo LA. Direct bonding with precoated brackets and
267-73. self-etching primers. Am J Dent 2006;19:241-4.
12. Scougall Vilchis RJ, Yamamoto S, Kitai N, Hotta M, 24. Wendl B, Droschl H. A comparative in-vitro study of the strength
of directly bonded brackets using different curing techniques.
Yamamoto K. Shear bond strength of a new fluoride-releasing
orthodontic adhesive. Dent Mater J 2007;26:45-51. Eur J Orthod 2004;26:535-44.
622.e2 Finnema et al American Journal of Orthodontics and Dentofacial Orthopedics
May 2010

Appendix table. Overview of the experimental conditions and outcomes of the studies in the meta-analysis
Reference Number Phosphoric Etch
number of per Thymol Fluor-free Mesh acid time Polymerization Light-cured
study group storage cleaning base etch $30 s time (s) composite

1 20 N N Y Y Y 40 Y
1 20 N N Y N N 40 Y
1 20 N N Y N N 40 Y
1 20 N N Y N Y 40 Y
2 9 Y Y Y Y N 30 Y
2 9 Y Y Y Y N 4 Y
2 9 Y Y Y Y N 50 Y
2 9 Y Y Y Y N 4 Y
2 9 Y Y Y N Y 30 Y
2 9 Y Y Y N Y 4 Y
2 9 Y Y Y N Y 50 Y
2 9 Y Y Y N Y 4 Y
3 10 Y Y Y Y Y 40 Y
3 10 Y Y Y Y Y 40 Y
3 10 Y Y Y Y Y 40 Y
4 15 N N Y Y Y 6 Y
4 15 N N Y Y Y 2 Y
4 15 N N Y Y Y 6 Y
4 15 N N Y Y Y 2 Y
4 15 N N Y Y Y 20 Y
5 10 N N Y Y N 60 Y
6 10 N N Y Y Y 20 Y
6 10 N N Y Y Y 20 Y
6 10 N N Y Y Y 20 N
6 10 N N Y Y Y 20 N
7 20 N N Y Y N 20 Y
8 10 Y Y N Y Y 40 Y
8 10 Y Y N N N 40 Y
8 10 Y Y N N N 40 Y
9 20 N N N Y Y 40 Y
10 15 N N N Y Y 40 Y
10 15 N N N Y Y 40 Y
11 40 Y Y Y Y Y 40 Y
12 35 Y Y Y Y Y 30 Y
12 35 Y Y Y N Y 30 Y
13 15 Y N Y Y Y 20 Y
13 15 Y N Y Y Y 20 Y
13 15 Y N Y Y Y 6 Y
13 15 Y N Y Y Y 6 Y
14 10 N N Y Y Y 20 Y
14 10 N N N Y Y 20 Y
14 10 N N Y Y Y 20 Y
15 15 Y Y Y Y Y 40 Y
16 20 N Y Y Y Y 40 Y
16 20 N Y Y Y Y 10 Y
16 20 N Y Y Y Y 20 Y
16 20 N Y Y Y Y 40 Y
17 20 N Y Y Y Y 20 Y
18 25 Y N Y Y Y 20 Y
19 25 Y N Y N N 20 Y
20 15 Y N Y N N 20 Y
21 25 Y N Y Y Y 30 Y
22 15 Y N Y Y Y 40 Y
22 15 Y N Y N N 40 Y
23 15 Y N Y Y Y 20 Y
23 15 Y N Y N N 20 Y
24 5 N Y Y Y Y 40 Y
24 5 N Y Y Y Y 10 Y
24 5 N Y Y Y Y 2 Y
24 5 N Y Y Y Y 3 Y
24 5 N Y Y Y Y 24 Y
24 5 N Y Y N N 40 N
24 5 N Y Y N N 10 N
24 5 N Y Y N N 6 N
24 5 N Y Y N N 24 N

*On bracket-adhesive interface; †all adhesive remained on enamel; Y, yes; N, no; ?, only mean ARI value given; NR, not reported; SEM, standard
error of the mean.
American Journal of Orthodontics and Dentofacial Orthopedics Finnema et al 622.e3
Volume 137, Number 5

Appendix table. (Continued)

Halogen Water Storage Crosshead speed Force Shear ARI† Bond strength
light storage time (h) Thermocycling (mm/min) location* blade (%) (MPa) SEM

Y Y 24 N 0.5 Y Y 85 13.10 0.69


Y Y 24 N 0.5 Y Y 75 16.00 1.01
Y Y 24 N 0.5 Y Y 0 11.50 0.74
Y Y 24 N 0.5 Y Y 0 9.90 0.89
Y N 72 Y 0.5 Y Y 11 24.40 1.84
N N 72 Y 0.5 Y Y 11 18.21 1.34
Y N 72 Y 0.5 Y Y 11 27.76 1.03
N N 72 Y 0.5 Y Y 11 25.09 1.24
Y N 72 Y 0.5 Y Y 0 23.44 1.94
N N 72 Y 0.5 Y Y 0 21.23 1.37
Y N 72 Y 0.5 Y Y 0 22.89 1.19
N N 72 Y 0.5 Y Y 0 16.24 1.05
Y Y 1 N 1.0 Y Y 22 14.05 1.03
Y Y 24 N 1.0 Y Y 30 20.94 0.79
Y Y 0 Y 1.0 Y Y 10 18.03 0.84
N Y 48 N 1.0 N Y 0 12.09 0.89
N Y 48 N 1.0 N Y 0 7.66 0.41
Y Y 48 N 1.0 N Y 7 11.24 0.92
Y Y 48 N 1.0 N Y 0 6.41 0.57
Y Y 48 N 1.0 N Y 7 11.85 1.13
Y Y 24 Y 0.5 Y Y NR 19.60 3.04
Y Y 24 N 0.5 N Y NR 11.35 0.93
Y Y 672 N 0.5 N Y NR 9.19 1.20
Y Y 24 N 0.5 N Y NR 11.58 0.95
Y Y 672 N 0.5 N Y NR 5.39 0.75
N Y 40 N 0.1 N Y 90 15.20 1.01
Y Y 24 N 0.5 N N 80 6.43 0.59
Y Y 24 N 0.5 N N 50 4.61 0.28
Y Y 24 N 0.5 N N 0 4.74 0.4
Y Y 72 N 3.0 N N NR 13.03 0.54
Y Y 72 N 3.0 N N NR 14.45 0.61
Y Y 72 N 3.0 N N NR 14.17 0.54
Y Y 72 Y 2.0 N N 17.5 14.82 0.41
Y Y 360 Y 5.0 Y Y 29 18.10 0.93
Y Y 360 Y 5.0 Y Y 29 12.70 0.56
Y Y 0.5 N 1.0 N N ? 13.6 0.98
Y Y 1 Y 1.0 N N ? 16.1 0.93
N Y 0.5 N 1.0 N N ? 14.2 1.19
N Y 1 Y 1.0 N N ? 18.2 1.19
Y Y 24 N 5.0 N N 0 21.56 1.05
Y Y 24 N 5.0 N N 0 22.32 0.51
Y Y 24 N 5.0 N N 0 17.82 0.5
Y Y 72 N 1.0 N N 65 23.23 1.34
Y Y 24 N 0.5 Y Y 85 13.10 0.69
N Y 24 N 0.5 Y Y 50 9.10 0.69
N Y 24 N 0.5 Y Y 55 13.90 1.07
N Y 24 N 0.5 Y Y 55 12.70 1.16
Y Y 24 N 0.5 Y Y 80 12.90 0.76
Y Y 24 N 1.0 N Y 0 12.27 1.00
Y Y 24 N 1.0 N Y 0 12.20 0.85
Y Y 24 N 1.0 N Y 0 10.45 1.06
Y Y 24 N 1.0 N Y 0 14.93 0.95
Y Y 24 N 1.0 N Y 0 7.19 0.73
Y Y 24 N 1.0 N Y 0 8.34 0.51
Y Y 24 N 1.0 N Y 0 14.13 2.27
Y Y 24 N 1.0 N Y 0 14.28 2.74
Y Y 1 N 0.5 N Y 0 9.50 0.72
Y Y 1 N 0.5 N Y 0 5.80 0.58
N Y 1 N 0.5 N Y 0 3.50 0.49
N Y 1 N 0.5 N Y 0 5.20 0.31
N Y 1 N 0.5 N Y 0 5.80 0.72
Y Y 1 N 0.5 N Y 0 8.20 0.80
Y Y 1 N 0.5 N Y 0 7.50 0.63
N Y 1 N 0.5 N Y 0 9.40 0.85
N Y 1 N 0.5 N Y 0 8.00 0.67

Potrebbero piacerti anche