Sei sulla pagina 1di 24

DIGITAL TROUBLESHOOTING PROCEDURES &FAULT LISTING

1. Knurr case removal. 2. Rittal case removal. 3. A/D board, I/O board, and controller board access instructions. 4. Voltage Busline and Crystal Checkout. 5. Open Ground fault indicator lamp lit. Open Ground circuit disable instructions for 6kV, 12kV, 24kV, D-Series testers. 6. Hipot Trip fault indicator lamp lit. 7. Hipot Current Bar display not present. 8. CRT blank when PTT button depressed/surge mode. 9. Hipot voltage bar display not present. 10. Seconds/div, volts/div, A/div, or function display incorrect. 11. Printer does not work properly. 12. No Baker logo displayed. 13. No Hipot voltage at test leads. 14. No surge voltage display. 15. Voltage drops when PTT button released. 16. Zero Start Interlock does not engage. 17. PI Value or current measurement inconsistent. 18. Display locked up after self-test. 19. Display unreadable and flashes on/off, does not initiate self-tests.

1 Baker Instrument Company The Measure of Quality

KNURR CASE REMOVAL


All digital series 6kV and 12kV models manufactured before August 1998 are enclosed in a blue shock mounted Knurr design case. Removal from the case is necessary to access internal components. Removal procedure is as follows. 1. Unlock the rear cover latches and unhinge the rear cover. 2. Locate and remove four (4) Phillips M4 x 10mm screws. Two are located on the left side and two are located on the right side. 3. Reinstall the rear cover. 4. Locate and remove on the front panel four (4) M5 x 16mm black hex head screws. Two are on the extreme left edge on the front panel and two are located on the extreme right edge of the front panel. 5. Grasp the CRT bezel in one hand and the test leads with the other hand and pull straight out of the blue Knurr case.
Handle Assembly for Knurr Case

2 Baker Instrument Company The Measure of Quality

RITTAL CASE REMOVAL


The Rittal case has been used on the Digital 6kV and 12kV models after serial number 327. Removal of this case is as follows. 1. Locate and pry open, with a small slotted screwdriver, six plastic covers that overlap six M4 screws that hold the top cover to the case assembly. 2. Remove these six (6) M4 screws and remove the top cover. 3. Locate and remove four (4) M16 x 16mm screws that secure the front panel to the case assembly. Two screws are located to the left of the CRT bezel and two are located to the right of the test lead set as they exit from the front panel. 4. Locate and remove five (5) M5 x 12mm black hex head screws that secure the bottom chassis to the left and right side case L brackets. Two are located to the left of the CRT shield, two are located to the right of the high voltage transformer, and one is located underneath the test select switch. 5. Locate and unplug the two (2) pin molex connector from the 12 volt fan to the wiring harness. This fan is located directly behind the CRT tube. It may be necessary to remove a tie wrap that secures the molex connector to the side case frame. 6. Slide the front panel and bottom assembly forward to remove from the case assembly. Access will now be gained to all internal components and assemblies except the I/O board, A/D board and the Controller Board. See the following instructions for access to these assemblies. Note: For D12000 and DS212 units that are Power Pack capable, this will involve an additional step for removal of a pot core that is mounted adjacent to the 12 volt fan.

Handle Assembly for Rittal Case 3 Baker Instrument Company The Measure of Quality

A/D BOARD, I/O BOARD, AND CONTROLLER BOARD ACCESS INSTRUCTIONS.


These assemblies can be accessed from the front panel from either the Knurr or Rittal case. They are installed in the card cage portion of the unit. 1. To access the I/O board, locate on the front panel the small panel which has the function knob, volts/div knob and the seconds/div knob. 2. Remove two (2) M3 x 7mm hex head screws located above the function knob and below the seconds/div knob. 3. Grasp the function knob and the volts/div knob and pull away from the front panel. These assemblies are 96-pin plug in type cards that slide on a card guide system and plug into mating 96 pin connectors mounted on the Mother Board. 4. To remove the A/D board, locate the panel with the open ground/hipot trip indicator lamps, intensity, vertical, and horizontal position knobs. 5. Remove two (2) M3 x 7mm hex head screws located above the open ground indicator and below the horizontal position knobs. 6. Position fingertips behind the stenciled panel for this assembly and pull away from the front panel. 7. Access to the Controller card is now gained, it will be the only card left in the card cage. Simply grasp this assembly and pull away from the front panel. Note: These assemblies contain C-MOS logic, which is ESD sensitive. Removal of these cards are recommended to be performed at a grounded station by personnel wearing a grounded wrist strap. Failure to do this may result in integrated circuit damage.

VOLTAGE BUSLINE AND CRYSTAL CHECKOUT


All digital models are equipped with a Power Supply board which supplies +5 volts, +12 volts, -12 volts, and several crystals on the A/D board and Controller board. A loss or loading of any of these supplies will result in the following conditions. To narrow down to which assembly is defective can be done quickly by unplugging sequentially the I/O, A/D, and the Controller boards and the CRT 10 pin connector one at a time and 4 Baker Instrument Company The Measure of Quality

measuring if the supply that is defective is restored. If not, replacement of the Power Supply board would be required. If a board is found to be loading a supply then any component on that board that is tied to the supply can be suspect. To narrow to the component level will require a one component at a time isolation from the supply via desoldering. 1. Loss of or loading of the 12 volt supply will cause no output in the surge mode when the PTT button is depressed. The trace will be visible at its most right upper position on the CRT and will be non-adjustable with the vertical or horizontal position potentiometers. Surge trigger will be present, but no surge waveform will be attainable when the PTT is engaged. In the Hipot mode, a Hipot trip LED will illuminate and the voltage bar will show a six (6) division up and the current bar will be extended off the screen. No Hipot output will be attainable and numeric display for voltage and current will stay at zero (0). 2. Loss of or loading of the +12 volt supply will cause a total non-operation of the tester. There will be no display upon power up or self-tests or no change when any of the output controls are used. 3. Loss of or loading of +5 volt supply will cause a total non-operation of the tester. There will be no display upon power up or self-tests or no change when any of the output controls are used. 4. Loss of crystal frequency X1 on the A/D board will cause the loss of positional trace control with the surge waveform in the upper right quadrant. In the Hipot mode, after PTT is depressed hipot voltage and current bars will be the same as in step #1 but numeric display for voltage will show 11950 volts and the current will show 996.0 A. Replace X1. 5. Loss of crystal frequency on U6 on the A/D board will show an erratic unstable sawtoothed pattern once the PTT button is depressed and also will fail the A/D memory self-test upon power up. Replace U6. 6. Loss of crystal frequency on XTA1 on the Controller board will cause total nonoperation of the tester. There will be no display upon power up or self-test, or no change when any of the tester controls are used. Replace XTA1. 7. Loss of frequency on Y1 on the Controller board will cause a blinking effect on the tester CRT with no recognizable lettering or numeric display. Tester will not initiate it s self-tests nor will any controls have any change on the display. Replace Y1.

5 Baker Instrument Company The Measure of Quality

OPEN GROUND FAULT INDICATOR LAMP LIT


1. Check that the tester is being operated with a two wire cord with a ground prong. 2. Check for a broken ground connection in the power cord. 3. Check that if being powered from an isolation transformer that a proper ground is supplied to the tester. 4. Check the mating ground pin receptacle at the power outlet for a bad connection. 5. Check continuity from the ground prong on the input power receptacle to the Power board P2 pin 1 and also to chassis ground. 6. Check MOV 1, MOV 2, MOV 3 for electrical damage. If damage is found also check for burnt traces on both sides of this assembly. Replace Power board if traces are damaged. Also to prevent future damage to this assembly or possible further damage, measure and correct line voltage fluctuations so that they remain consistently with the following parameters.

Unit D6000 D12000 DS206 DS212 Lower 100 Volts AC 100 Volts AC 220 Volts AC 220 Volts AC

Line Voltage Range Upper 120 Volts AC 120 Volts AC 250 Volts AC 250 Volts AC

7. Check signal at Q1 collector on the Power board for a .4 volt ripple, if not replace Q1. 8. Check U9 pin 3 on the I/O board for a 0 volts and U9 pin 4 for five (5) volts DC, if not replace U9. 9. Check Q6 on the I/O board for a collector to emitter short, if so then replace Q6. 10. Check for a shorted Open Ground LED.

6 Baker Instrument Company The Measure of Quality

Open Ground Circuit Disable instructions for 6kV, 12kV, and 24kV D Series Testers.
In a testing environment where no ground can be supplied to the surge tester, the following steps can be done so that the tester will operate. Read and follow these instructions prior to proceeding. 1. Locate and remove the two (2) M4 x 7mm black hex head screws on the front panel. The first is located above the function switch knob, and the second is below the seconds/division knob. 2. Grasp the function switch knob in one hand and the seconds/division knob with the other hand and pull away from the tester. By doing this the I/O board will no unplug from the Motherboard. 3. Locate and remove the two (2) M4 x 7mm black hex head screws on the front panel. The first is above the open ground lamp socket and the second is below the horizontal potentiometer knob. 4. With your fingertips, grasp behind the front panel A/D board assembly and pull away from the tester. By doing this the Power board will now unplug from the Motherboard. Caution: DO NOT damage the potentiometers or cause fractured solder joints at the legs of the potentiometers. 5. Remove the remaining printed circuit board, the Controller board. Grasp at the edges of this assembly and pull away from the testers front panel. 6. For the 15kV and 24kV only locate and remove four (4) Phillips screws that secure the blue top cover to the case assembly. Two (2) will be located on the left side and two (2) will be located on the right side. This step applies to 15kV, and 24kV models only. Remove the top cover. Skip to step 10. 7. For 6kV or 12kV models manufactured before August 1998 which are enclosed in a blue shock mounted Knurr design case, removal of the inner chassis from the case will be necessary. If enclosed in a Tan Rittal design case proceed to step 9. 8. Unlock the rear cover latches and unhinge the rear cover. Locate and remove four (4) Phillips M4 x 10mm screws. Two are located on the left side and two are located on the right side. Reinstall the rear cover. Locate and remove the front panel four (4) M5 x 16mm black hex head screws. Two are on the extreme left edge and two are located on the extreme right side of the front panel. Grasp the CRT bezel in one hand and the test leads with the other hand and pull straight out of the Knurr case. Skip to step 10.

7 Baker Instrument Company The Measure of Quality

9. If a 6kV or 12kV model was manufactured after August 1998 and is enclosed in a tan Rittal design case locate six (6) gray hinge covers on the top cover. With a small slotted screw3driver pry the covers so that six (6) Phillips screws that secure the top cover are visible. Remove all six (6) screws to remove the top cover. 10. Locate the card cage that houses the Power and Signal Boards. 11. Remove the top cover of the card cage. It is secured by eight (8) M4 x 7mm black hex head screws and two (2) M4 kep nuts. Remove this hardware. Position one hand inside the card cage and lift up at the back end to remove the top cover. 12. Locate on the Motherboard an insulated jumper wire J7 that is positioned between connectors J2 and J3. If you intention is to run the tester without a supplied ground permanently, set step 14. 13. If your intentions is to move the tester between environments that do and do not have a supplied ground, see step 15. 14. Remove jumper at J7. 15. Remove jumper at J7. Install two 22 gauge wires one at each connector hole at J7 and exit these wires from the right side of the card cage. The length of these wires should be long enough to reach the front panel. Tie wrap and secure these wires along the right side of the card cage. Obtain a two-position toggle switch that has a common/ NO and NC contacts. Drill a hole between the leads energized light and the right side of the card cage and mount this toggle switch to the front panel 16. Solder the wires from the J7 connector to the toggle switch. This will now enable you to more easily disable the open ground fault detection circuit. Note: The Open Ground LED will still be lit when the tester does not sense a ground connection, but now the tester will operate.

HIPOT TRIP FAULT INDICATOR LAMP LIT


1. Check that the winding under test is not indeed shorted. 2. Check for a shorted high voltage transformer secondary to ground. Measure with a multimeter from D3 anode on the Stack board to ground. Normal measurements are approximately 12K ohms with the function switch in either the 10 or 100 A position or approximately 20K ohms in the 1 A position. If found to be much lower resistance, replace T1 high voltage transformer. 3. Check for shorted SCR s on the Stack board by measuring the resistance across all cathode/anode junctions of each SCR. These measurements can be obtained easily by 8 Baker Instrument Company The Measure of Quality

measuring across resistors R1 through R10. Good measurements will be 70K ohms or higher. Replace any SCR s found to be shorted. 4. Check U2 pin 7 on the I/O board, this should be a zero (0) volts DC, if at five (5) volts DC then replace U2 and U7. 5. Check Q3 on the I/O board for a collector to emitter short, if so replace Q3. 6. Check for shorted Hipot trip lamp on the A/D board, if so replace D3.

HIPOT CURRENT BAR/DISPLAY NOT PRESENT


1. Check continuity from test lead one (1) alligator clip to deck one (1) of the selector switch. 2. Check continuity from test lead ground alligator clip to the metal spacer adjacent to D1 and D2 on the Discharge board. 3. For the purposes of troubleshooting this circuit, set up will require the tester to be in the Hipot mode with leads one (1) and ground connected to a 10 megohm 35 watt resistor. Use a digital multimeter from these measurements with the negative lead of the meter connected to pin 29C on the I/O board. Set tester to 100A/division. 4. Measure at U11 pin 3 on the I/O board and raise output control to 6 kV. The measurement should increase negatively as more current is drawn through the test load. At 6 kV this should be approximately -.040 volts DC. Measurement at the output U11 pin 6 on the I/O board should be identical. If not replace U11. 5. Measure at U10 pin 6 as the output control is raised to 6kV. At 6kV, this measurement will increase to approximately 2.9 volts DC. If not replace U10. 6. Measure at RPK1 pin 15 and then pin 2 for the output as described in step 5, if not present at RPK1 pin 2, Replace RPK1. 7. Measure at U25 pin 8, use a 10X scope probe, set scope to 1 volt/division and 5 ms and an internal trigger at the 3 volt level. Raise output of tester to 6kV across the 10 megohm load. Measure for approximately 3 volts with a negative going square wave approximately .4 volts at a 5 ms duration. If not present, replace U25. 8. Measure at U22 pin 18 for the same signal as in step 7 and then measure at U22 pin 9, the output of u22 should be doubled to approximately 6 volts with the same negative going square waveform. If not replace U22. 9. Measure at U23 pin 7, if same measurements taken in step 8 is not present, replace U23. 9 Baker Instrument Company The Measure of Quality

CRT BLANK WHEN PTT BUTTON DEPRESSED/SURGE MODE


This particular fault is associated with the loss of one or more of the following signals: line trigger, stack power, real test or sync trigger. 1. Check the stack power signal at P1 pin 2 on the Power board, this signal should be a 360 volt AC sine wave with the negative portion clipped at zero (0) volts. Frequency will be at approximately 60 Hz. If not present remove the P1 connector on the Power board and remeasure at R2. If still not present, check T1, D1, and R2 for defects and replace. If this signal is loaded down upon reconnecting, then check FET Q6 on the Stack board for defects. 2. Check the line trigger signal starting at the junction R1 and T1 pin 5 on the Power board. This signal will be an AC sine wave with an approximate amplitude of 900 volts peak to peak with a 60 Hz frequency. At P1 pin 1 this signal changes due to components tied to this signal line on the I/O board. At his point, this signal will be a 1.2-volt square wave with a frequency of 60 Hz. If this signal is not present or incorrect proceed to the I/O board. Check D1, Q2, and U9 for defects and replace if necessary. Signal at U19 pin 3 should be a 5-volt square wave with a frequency of 60 Hz. 3. Check the real test signal at U29 pin 16 on the Controller board. This signal will be 5-volt logic and should be a positively pulsed waveform with a changing high state duration of approximately 1 to 20ms. Frequency will be 5 Hz. Depressing the PTT button will activate this signal. This signal is then routed to the I/O board at U19 pin 5 and also through an inverter U4 to U19 pin 4, if these signals are missing or incorrect, then any of the three components listed above are suspected to be faulty. 4. Sync trigger will be the output from the flip-flop U19 pin 1. This signal should be a positively pulsed waveform with an approximately 450S high state duration with a frequency of 6 Hz. 5. The PTT button again needs to be depressed to activate this signal. If signal is not present, replace U19 on the I/O board. Check the stack trigger signal at R22 on the A/D board for a 10-volt positively pulsed waveform with a frequency of 5 Hz. If not present, check Q1, Q2, and Q3 for faults. The Stack trigger signal is then routed to J1 pin 4 on the Stack board. Possible loading of this signal can also be attributed to component defects in the Stack board trigger and charging circuit such as C1 or Q6.

10 Baker Instrument Company The Measure of Quality

HIPOT VOLTAGE BAR/DISPLAY NOT PRESENT


1. Check DC voltage at J1 pin 1 on the Stack board. Voltage output will depend on where the output control is set. A normal range will be from zero (0) volts at minimum output setting to approximately 3 volts DC at 6 kV output. If not present check all SCR s Q1 through Q10 for shorts by measuring the resistance across all cathode to anode junctions of each SCR. Replace those SCR s found to be shorted. Readings should be 700K ohms or higher. Bad junctions will read much lower. 2. Check for same rise in DC voltage at U12 pin 3 on the I/O board, if not present, check connection between these two points. 3. Check at U12 pin 6 on the I/O board for the same increase in voltage, if not present, replace U12. 4. Check U3 pin 6 on the I/O board for the same increase, if not present, replace U3. 11 Baker Instrument Company The Measure of Quality

5. Check U25 pin 4 on the A/D board for the same increase in voltage, if not present, check connection between U3 pin 6 on the I/O board and RPK1 pin 16. Then check RPK1 pin 16 to RPK1 pin 1 for 47K ohms resistance. If open replace RPK1. 6. Check U25 pin 8 for a 2.5-volt positively pulsed waveform with a frequency of 200ms with the tester set at 6kV output with no current deflection on the CRT. If not present check U25 pins 1, 15, 16. Only pin 1 should have a signal which should be a 5-volt waveform pulsing negatively to zero (0) volts every 200ms. If these signals are present at pins 1, 15, and 15 with a measured input at U25 pin 4 then replace U25. 7. Check U22 pin 9 for a 5-volt positively pulsed waveform with a frequency of 200ms. Set the tester to 6 kV output with no current deflection on the CRT. If not present replace U22. 8. Check U23 pin 7 for the same signal. If not present, replace U23.

SECONDS/DIV, VOLTS/DIV, A/DIV OR FUNCTION DISPLAY INCORRECT.


1. The seconds/division display is controlled by switch S1 on the I/O board and pull up resistors R40, R41, R16, R2, R3, R4, R19, R20, and R43. This in turn connected to a series of AND, NAND and Inverter IC s set up to control a binary pattern for control lines Sweep A, B, C, and D. See4 binary chart on the I/O board. A typical fault would be seen as a duplication of any one of the possible displayed sweep settings such as CHANGE, 2S, 6S, 10S, 20S, 60S, 100S, 200S, 600S, or 1mS. Possible failures can include a logic gate stuck high or low such as U1, U8, U4, U5 or U6 or mux U18. Check the pull up resistors listed above. Check that S1 grounds sequentially as it is rotated. 2. The volts/division display is controlled by switch S2 on the I/O board and pull up resistors R10, R8, and R11. This in turn is connected to a series of Inverters and/or gates to set up a binary pattern for control lines V DIV 1 and V DIV 2. See binary charge on the I/O board. A typical fault will be seen as a duplication of any one of the possible volts/division settings such as 500 volts, 1000 volts, 2000 volts, or 3000 volts. Possible failures can include a logic gate stuck high or low such as U4 or U5 or mux U17. Check the pullup resistors listed above. Check that switch S2 grounds sequentially as it is rotated. 3. The /A division display is controlled by switch S3:A on the I/O board and pull up resistors R36 and R38. This in turn is connected to a series of Inverters to set up a binary pattern for control lines A/DIV 1 and A/Div 2. See binary chart on the I/O board. A typical fault will be seen as a duplication of any one of the possible A/ 12 Baker Instrument Company The Measure of Quality

division settings such as 100A, 10A, or 1A. Possible failures can include a logic gate stuck high or low such as U6 or mux U17. Check the pull up resistors listed above. Check that switch S3:A grounds sequentially as it is rotated. 4. The function display is controlled by switch S3:A on the I/O board and pull up resistors R42, R12, and R14. This is in turn connected to a series of Inverters at U6. A typical fault will be seen as the CRT display not matching the function control knob. Possible failures can include a logic gate stuck high or low such as U6, U5, mux IC s U17 or U15. Check pull up resistors listed above and also pull up resistors at RPK3 pins 8 and 9. Check that switch S3:A grounds sequentially as it is rotated.

PRINTER DOES NOT WORK PROPERLY


1. Check that the printer cable is plugged into the printer port. 2. Check for faulty or damaged printer cable by checking continuity between all 25 pins in both ends. Try a replacement printer cable. 3. Digital models with a serial number 001 through serial number 370 or ones manufactured before December 1997 must be used with an Epson FX dot matrix printer. And it must be configured properly to interface. Firmware revision 1.35 is used with the Epson dot matrix printer. Digital models after serial number 370 or ones manufactured after December 1997 will have firmware revision 1.36 which will use an Epson DeskJet 600 printer. 4. Check for faulty printer, try a substitute printer. 5. Measure with a 10x-scope probe the following points on the I/O board. To activate these signal lines a printer will need to be connected and the print softkey will need to be depressed. Monitor PRN DB0 through PRN DB7 from U14 on the I/O board for data transfer. Data transfer is defined as a 5-volt logic level that changes in frequency of pulses as the data transfer sequence changes. Set the oscilloscope to 5 volts/division and a 5ms-sweep rate. If this output is not det4ected on one or more printer lines, check the resistance of all resistors in RPK2. Replace if defective. If RPK2 checks good, replace U14, U16, and U9. Check also the resistance of RPK1 and R50 through R54. Replace if necessary.

NO BAKER LOGO DISPLAYED


1. Upon power up if no display is seen on the CRT, the first step to ensure is that the intensity output potentiometer is turned clockwise are enough to see a display, generally at least a 2 O clock position should be sufficient. 13 Baker Instrument Company The Measure of Quality

2. See Voltage Busline and Crystal Checkout procedures and correct any voltage Busline or crystal output and frequency problems. 3. To determine if the CRT tube is defective a check is needed to ensure that +12VDC is present on P1 CRT pin 7 and then pins 1 and 10 are grounded. 4. Check P1 CRT pin 2 with a digital multimeter, this measurement will be adjustable with the intensity potentiometer, it should range from 14VDC to +5.6VDC, if not check R9 on the A/D board and its connections. 5. Check P1 CRT pin 3, this measurement should be approximately 110VDC. 6. Check P1 CRT pin 4, this measurement will be adjustable with the intensity potentiometer and will range from 110VDC to +5.6 VDC, if not check R9 on the A/D board. 7. Check P1 CRT pin 6 for a 4-volt square waveform pulsing negatively to zero (0) volts. Repetition rate will be every 30S and the negative duration will be 4S. If signal is not present trace back to U24 pin 4 and then to U1 on the Controller board. Replace if necessary. 8. Check P1 CRT pin 9 for a 5-volt positively pulsed waveform. Repetition rate will be every 17mS and the positive duration should be 500S. If not present, trace signal back to U1 on the Controller board. Replace if necessary. 9. Check P1 CRT pin 8 for 2.4 volt data entering from the Controller board U27 pin 13, replace if necessary. 10. If above signals and voltages are correct, replace CRT tube.

NO HIPOT VOLTAGE AT TEST LEADS


1. Ensure the tester function switch is in the hipot mode at 100A/Division and that the test lead select switch is in the Hipot selection. 2. Measure at T2 pin 2 with a digital multimeter, connect positive lead at this point and the negative lead to chassis ground. Set the meter to AC volts. This point should be a line voltage upon turning the on/off switch to the on position. 3. Measure at T2 pin 3, this point will be variable depending on the position of the output control knob. At the full counter-clockwise position this voltage will be less than 3 volts AC. At approximately output turning clockwise this voltage will be line voltage. If not present, check the neutral connection to the variac, if this connection is present replace T2. 14 Baker Instrument Company The Measure of Quality

4. Check at the terminal block adjacent to the high voltage transformer T1 at the orange wire for the same AC voltage starting with the output control to full counterclockwise position and than at output. If not present, see the section on the Zero Start Interlock does not engage. 5. Locate the secondary T1 high voltage transformer wire connected to D3 anode on the Stack board and disconnect. Connect a P6015A high voltage scope probe or equivalent to the T1 secondary wire with the probe negative lead connected to chassis ground. Set the volts/div on the oscilloscope to 2 volts and the timebase to 5ms. Turn the output control approximately to output. This measurement should be 14kV AC peak to peak. 6. Verify that all other primary and secondary wires of T1 are connected to their respective destinations with a digital multimeter. Ensue that power is off and disconnect when making these measurements. If measurement in step 5 is not present and all connections are verified, then replace the high voltage transformer T1. 7. Reconnect T1 secondary wire and lift D3 cathode on the Stack board. Connect the P6015A probe to this point with the negative lead to chassis ground and turn output control to output. Set scope settings to the same as in step 5. This measurement should be approximately 7kV AC with the negative portion clipped by the diode. If not present replace D3 on the Stack board. 8. Check the SCR cathode to anode resistance for Q1-Q10. Ensue that the power is off and disconnected. These measurements can be obtained by measuring across R1 through R10. Replace any SCR s that are shorted. Good measurements will be 700K ohms or higher while defective SCR s will have much lower readings. 9. Check the resistance of the 25K-ohm 25-watt resistor that connects to the J5 wire from the Stack board. Replace if open. 10. Check continuity from lead 1 alligator clip to the 25K 25 watt resistor. If open, replace test lead 1. Check continuity from the ground lead alligator clip to the Discharge board ground stud. If open, replace the ground lead.

NO SURGE VOLTAGE DISPLAY


1. Ensure the tester function switch is in the surge mode and that the test lead select switch is in a test lead selection. Set the tester to 500 volts/division and at 6S/Division. Connect test load across test leads. 2. Measure at T2 pin 2 with a digital multimeter, connect positive lead at this point and the negative lead to chassis ground. Set the meter to AC volts. This point should be at line voltage upon turning on/off switch to the on position. 15 Baker Instrument Company The Measure of Quality

3. Measure at T2 pin 3, this point will be variable depending on the position of the output control knob. At the full counter-clockwise position this voltage will be less than 3 volts AC. At approximately output turning clockwise this voltage will be line voltage. If not present, check the neutral connection to the variac, if this connection is present replace T2. 4. Check at the terminal block adjacent to the high voltage transformer T1 at the orange wire for the same AC voltage starting with the output control to full counterclockwise position then at output. If not present, see the section on the Zero Start Interlock does not engage. 5. Locate the secondary T1 high voltage transformer wire connected to D3 anode on the Stack board and disconnect. Connect a P6015A high voltage scope probe or equivalent to the T1 secondary wire with the probe negative lead connected to chassis ground. See the volts/div on the oscilloscope to 2 volts and the timebase to 5ms. Turn the output control approximately to output. This measurement should be 14kV AC peak to peak. 6. Verify that all other primary and secondary wires of T1 are connected to their respective destinations with a digital multimeter. Ensure that power is off and disconnected when making these measurements. If measurement in step 5 is not present and all connections are verified, then replace the high voltage transformer T1. 7. Reconnect T1 secondary wire and lift D3 cathode on the Stack board. Connect the P6015A probe to this point with the negative lead to chassis ground and turn output control to output. Set scope settings to the same as in step 5. This measurement should be approximately 7kV AC with the negative portion clipped by the diode. If not present replace D3 on the Stack board. 8. Check the SCR cathode to anode resistance for Q1-Q10. Ensure that the power is off and disconnected. These measurements can be obtained by measuring across R1 through R10. Replace any SCR s that are shorted. Good measurements will be 70K ohms or higher while defective SR s will have much lower readings. 9. Check continuity of test leads 1, 2, and 3 back to the selector switch, replace any found to be open. 10. Check continuity from the ground lead alligator clip to the Discharge board ground stud. If open, replace the ground lead. 11. With a 10X scope probe measure at pin 32B on the I/O board with the negative probe lead connected to chassis ground. After engaging the zero interlock turn the output control to approximately output. Set the oscilloscope to 1 volt/div and 10S sweep rate. The signal at this point should be about a 1.8-volt surge waveform. If not check continuity with the tester AC input disconnected from P1 pin 32B & 32C on the I/O board to J1 pins 3 and 4 on the Discharge board. 16 Baker Instrument Company The Measure of Quality

12. Access the Discharge board by removing the Stack board and check resistance measurements of the following components and replace if necessary. Ensure that power is disconnected when making these measurements and disassembling the unit. Check R1 through R14. 13. If the signal in step 11 was noted proceed to check that relay K1 on the I/O board has no contact resistance between pins 2 and 3. If pin 7 of the relay is at a low state and contact resistance is detected, replace relay K1. If pin 7 is at approximately +12 volts DC, then check U6 pin for +5 volts DC, then replace Q1 on the I/O board. If U6 pin 6 is at a low state, replace U6. 14. Replace U19, U20 and U14 on the A/D board.

VOLTAGE DROPS WHEN PTT BUTTON


RELEASED
In all digital models previous to serial number 210, an intermittent problem was detected in that when an operator of the test was to attain the desired test voltage and release the PTT button, the displayed voltage would drop approximately 10%. To correct this problem the addition of K1 and K2 chassis relays were added. In older revisions the AC line voltage was switched directly from the PTT button or the footswitch to the T2 variac. The Zero Start Interlock board would switch the voltage from T2 pin 3 to the primary of the high voltage transformer. To engage the Zero Interlock, the following needs to occur at the Zero Start Interlock board. 1. AC line voltage is connected through the footswitch or PTT button to J1 pin 1. 2. +13 volts DC is supplied to J2 pin 5 supplied from the I/O board J1 pin 2B. 3. /PTT signal at J2 pin 2 needs to switch from +5.6 volts DC to .4 volts DC when the PTT button is depressed. 4. The output control variac needs to be adjusted completely counter-clockwise. After Serial number 210 Baker has changed from revision D to a revision F chassis wiring diagram. Major changes incorporate using a +12 volt DC supply wired through the footswitch and the PTT button. Also K1 and K2 were added. These relays hold the displayed test voltage slightly longer so that the A/D memory has enough time to capture the displayed test voltage. Relay K2 can be fitted to mount o the side of the card cage above the Zero Start Interlock board and K1 can be mounted on the chassis next to the Zero Start Interlock board. Mount by drilling two mounting holes and securing with mounting hardware. Wiring additions or changes involve the following. To perform this modification removal from the case is required. Remove mounting hardware on the 17 Baker Instrument Company The Measure of Quality

variac and selector switch for easier access. Remove the A/D, I/O and the Controller boards from the card cage. 1. Remove the AC Hot (black) wires from the PTT button and the footswitch. Connect the black wire to K1:B common, K2:B common, and K2C common. 2. Connect a 20 gauge red wire to TB2 pin 1 on the Power Supply board and terminate the other end to the footswitch pin 1,one end of the PTT button and one end of K1:A coil. 3. Remove wiring from footswitch pin 2 (red) to the PT button to J1 pin 1 on the Zero Start Interlock board to T2 pin 2. 4. Replace the wiring on the other end of the PTT button with a wht/grn wire and connect to the footswitch pin 2 and also K2:A coil and an added 1N4936 cathode. 5. Connect the other end of K2:A coil to the 1N4936 anode and also to the ground terminal block. 6. Connect a red 20 gauge wire from J1 pin 1 on the Zero Start Interlock board to K2:C NO contact. 7. Connect a red 20-gauge wire from T2 pin 2 to K1:B NO and K1:A NO. 8. Connect a 26-gauge jumper on the rear of the I/O board from U6 to P1 pin 1A. 9. Remove top cover of card cage held by eight (8) M4 x 7mm black hex head screws and two (2) M4 kep nuts. 10. Remove four (4) kep nuts securing the Motherboard to the back of the card cage and solder a 26 gauge wire to the rear of the Motherboard J3 connector pin 1A. Run this wire out of the card to a small breadboard containing a 4.7K ohm watt resistor and a 2N2222 transistor. See schematic for electrical connection of this circuit. This breadboard needs to be mounted on the side of the card cage by drilling one hole and mounting with a nylon spacer and M4 hardware adjacent to the K2 relay. 11. Run a 20 gauge green wire from the 2N2222-transistor emitter and connect to the ground terminal block. 12. Reassemble unit.

ZERO START INTERLOCK DOES NOT ENGAGE


1. Check that the output control knob is completely adjusted counter-clockwise. 18 Baker Instrument Company The Measure of Quality

2. Ensure that the test lead select switch is not positioned in the leads ground position. 3. Check that the unit is supplied a proper ground. 4. With a digital multimeter set to DC volts, check the voltage at J2 pin 5 on the Zero Start Interlock board with the negative lead connected to chassis ground. This measurement should be approximately +13.5 volts DC and drop to about +12.7 volts when the PTT button is depressed. If this voltage is not present or is low the fault can be in the following circuits. The microswitch S1 which is mounted on the test lead select switch can be open. Check Q5 base on the I/O board for +12.6 volts DC, if present replace Q5. 5. Check at Q4 base on the I/O board, this should be approximately .7 volts DC, if present replace Q4. 6. Check U8 pin 13 for a +5.5 volts DC, if measured low replace U7. 7. Check U8 pin 12 also for a +5.5 volts DC, if at a low state check R13 and U9. 8. Check at U8 pin 11 for low state, if at +5.5 volts DC, replace U8. 9. Measure at J2 pin 2 on the Zero Start Interlock board, this measurement needs to be +5.5 volts DC and change to approximately .4 volts DC when the PTT button is depressed. If this is present check Q1 through Q5 and OC1, OC2, and OC3 for faults. 10. Measure between R9 and KE2 pin 1, this coil voltage should change from 6.4 volts DC to -.7 volts DC when the PTT button is depressed. If this is noted, replace K1 relay on the Zero Start Interlock board. 11. If the measurement at step 9 is not present, check U15, U7, and R18 for defects, replace if necessary.

PI VALUE OR CURENT MEASUREMENT

IS INCONSISTENT

The digital series design tester will produce accurate results to a +/- 5% specifications. The operators of the digital series may notice fluctuations in the current readings, megohms values and PI data due to fluctuations in the input line power. Small variations in the line power specifically at the one-minute and ten-minute intervals may cause inconsistent readings even when the same windings are analyzed. To greatly improve this occurrence, the steps below can be followed. 1. The use of a Patriot line conditioner in series with the power to the tester will reduce line fluctuations. Contact Baker Instrument Company for price and availability. 19 Baker Instrument Company The Measure of Quality

2. Connecting AC power from an isolated outlet does not have other equipment with high current demands. To check the internal current calibration of the tester will require the I/O board to be on an extender card. Follow the steps in the Quality Control and Calibration Procedure. If one or more range readings outside of +/-5% still occur, see the steps below. 1. Further adjustment may be necessary by altering the values of R33, R35, and R37. Slight resistive value changes of no more than 10% of the existing value up or down. 2. Continue monitoring offset value at the I/O board pin 2A to pin 29C for an offset which drifts away from the nominal measurement of 0.000 volts +/-0.010. If drifting is detected, replaced U10 and U11.

DISPLAY LOCKED UP AFTER SELF-TEST


In this fault mode, an operator will see after the on/off switch to the on position that the tester will illuminate the Baker logo and proceed with it s internal memory test. All selftests will pass. After an attempt to begin testing by depressing the PTT button, there will be no change in the testers display. Also no change will take place when a softkey is used. A shorted softkey or a faulty mux integrated circuit on the I/O board is more than likely the cause. 1. Check that all softkeys are not stuck, closed, or shorted. This type is normally open and should have high resistance across the two terminals used. 2. Check at U15 pin 2 for a high state, if this state is not high replace RPK3. Depress softkey 1. At this point it should change to a low state. 3. Check at U15 pin 4 for a high state, if this is not high replace RPK3. Depress softkey 2. At this point it should change to a low state. 4. Check at U15 pin 6 for a high state, if this is not high replace RPK3. Depress softkey 3. At this point it should change to a low state. 5. Check at U15 pin 8 for a high state, if this is not high replace RPK3. Depress softkey 4. At this point it should change to a low state. 6. Change U15.

20 Baker Instrument Company The Measure of Quality

DISPLAY UNREADABLE AND FLASHES ON/OFF, DOES NOT INITIATE SELF TEST
1. A check of the DC supplies and crystal frequency to verify proper conditions, see Voltage Busline and Crystal Checkout procedures. 2. A loss or loading of any one or more of the following lines due to faulty components or connections will cause this occurrence. Since these signal lines are tied to the A/D board, I/O board, and the Controller card it will be necessary to troubleshoot in the manner outlined in step 3. ___ ____ __________ _________ _________ _________ ____ RD, Reset, CRT Enable, 103 Enable, 102 Enable, 101 Enable, SAD, RD/WR, ALE, Reset, DB0 through DB7, AD1 through AD7 and AD8 through AD14. 3. To determine which printed circuit is defective, unplug the A/D board and the I/O board with the power off. Turn on the tester and observe the tester response. If the tester now begins with a Baker logo then the fault is not on the Controller card. The tester will fail the A/D Memory test due to the A/D card not being inserted into the Motherboard. Turn off the testers power. 4. Insert the A/D board into the J1 connector and switch power on. If the display returns to the original fault condition, then a fault will be on the A/D board. Turn off tester power. 5. Insert the I/O card into the J3 connector of the Motherboard and turn on the testers power. If the display returns to the original fault condition, then the fault will be located on the I/O board. 6. After determination as to which card is defective, remove power and the defective card. Install the faulty card on an extender card and apply power. 7. With a 10X-scope probe begin monitoring the data, address and signal lines noted in step 2. It will be helpful to record all lines, which do not have data present. See Motherboard DST Series schematic for pinout contacts of these signal lines to each card. Data is defined as a high frequency 0 to 5-volt amplitude logic level changes determined by the programmed U3 27C256 EEPROM located on the Controller card. 8. After determination as to which signal lines are stuck low or high, it will be necessary to trace the defective signal lines to all components that are tied to the same line and replace one at a time. It may also be necessary to check defective signals lines with a digital multimeter set to ohms for shorts to GND and to +5 volts DC. Also check 21 Baker Instrument Company The Measure of Quality

continuity on defective signal lines from components that are tied to the same signal line. 9. If the fault is located to either A/D board or the Controller card, replacement boards can be sent without affecting the calibration on the unit. An I/O board replacement will require a calibration of the unit.

A/D BOARD, I/O BOARD, CONTROLLER BOARD ACCESS INSTRUCTIONS


When MTA for Windows is installed or an Epson 640 or compatible printer is used, it will be necessary to replace U3 on the Controller Board. For this program to function, install the replacement 08-011R, firmware revision 1.37 for a 12kV tester. Please see the MTA for Windows users manual for further explanation of the procedure. 1. These assemblies can be accessed from the front panel from the Knurr, Rittal or Schroff design cases. These assemblies are installed in the card cage portion of the unit located to the right of the CRT display. 2. To access the I/O board, Baker part 80-108, locate on the front panel which has the function knob, Volts/Division knob and the Seconds/Division knob. 3. Remove two (2) M3 x 7mm hex head screws located above the function knob and below the Seconds/Division knob. 4. Grasp the Function knob and the Volts/Division knob and pull away from the front panel. This assembly is a 96-pin plug in type card that slides on a card guide system and plugs into mating 96 pin connectors mounted on the Motherboard. 5. To remove the A/D board, Baker part 80-106, locate the panel with the Open Ground/Hipot trip indicator lamps Intensity, Vertical and Horizontal Position knobs. 6. Remove two (2) M3 x 7mm hex head screws located above the Open round indicator and below the Horizontal position knob. 7. Position fingertips behind the stenciled panel for this assembly and pull away from the front panel. 8. Access to the Controller board, Baker part 80-107, is now gained. This will be the only board left in the card cage. Simply grasp this assembly and pull away from the front panel. 9. Remove the Controller board and replace U3 with one labeled firmware revision 1.37. 22 Baker Instrument Company The Measure of Quality

10. Install all boards back into the card cage and secure with the original hardware. NOTE: These assemblies contain C-MOS logic which are ES D sensitive. Removal of these boards are recommended to be performed at a grounded station by personnel wearing a grounded wrist strip. Failure to do this may result in integrated circuit damage.

ECO RECORD
Engineering changes have taken place since the manufacturer of D12000 S/D 001. D12000 series after serial number 210 should have these changes installed at the factory. These changes are designed to add more capability and improve the tester s performance. Also they will increase the tester s reliability by protecting components and circuits which have shown over the years to be vulnerable to failure in certain testing applications. The following is a list of these changes we recommend to be checked for installed during servicing. 1. A cooling fan for the CRT circuits was added to eliminate scrolling of the CRT display when internal temperatures reach above 105 degrees. 2. Power board components R3 and C1 were changed to increase reliability of Q1 on the Power board. 3. Discharge board changes include changes in the value of resistor R5 to provide for a greater incoming signal to the I/O board. This allows for a better capability to shape the surge signal so that the displayed surge signal on the CRT display more closely resembles what is shown at the test leads on an oscilloscope. 4. Other changes to this assembly isolation of the surge readout cable from Discharge board ground and ground routing changes. This will improve the noise resolution seen of the first displayed positive peak when surge testing at higher voltages. 5. Changes in the values of Stack board components R1 and R34 were performed to add greater reliability in SCR s Q1-Q5. 6. Changes in the revision of the Zero Start Interlock board form a common neutral design to a common ground design. This revision change allows for the elimination of false hipot trips and adds the capability to disable the Zero Start Interlock when testing armatures when a footswitch is used. 7. The addition of relays K1 and K2 and associated wiring harness changes were added to eliminate the condition where the displayed surge voltage on the CRT would drop approximately 10% when the PTT button was released.

23 Baker Instrument Company The Measure of Quality

8. EEPROM IC U3 is now at revision level 1.37. These firmware revision changes have added the following capability. A Polarization Index clock was added so that the operator can reset the clock after proper test voltage was reached, resulting in a full one or ten minute megohm test. Other changes in the firmware have added the capability to recall in any summary all three-surge test patterns with the test voltages of each lead displayed. This 1.37 revision change is necessary to upgrade to an inkjet Epson 640 printer or compatible printer. It is also necessary to be installed if this tester is used with MTA for Windows operating software system. 9. A/D board grounding changes, clamping diodes and bypass capacitors were added to increase the reliability of IC U14.

24 Baker Instrument Company The Measure of Quality

Potrebbero piacerti anche