Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
CONVERTER
STANDARD CONVERTER
ML 110
The manufacturer guarantees only English text. Available on our web site: www.isoil.com
ML 110
TECHNICAL DATA
Housing Material Dimensions Protection rate Connection sensor cable/Cable gland Ambient Temperature LCD Display Programming key board Pulse/frequency/alarm output Current output Bi-directional FS value Diagnostic functions Empty pipe detect. Galvanic isolation Data storage Programming plug in CE certification Measurements tolerance Repeatability Power supply Consumption
Nylon with fiber glass 120 X 120 X 55 mm IP 65 C018 cable ( MAX 20 m ) / Standard N 3 PG 11 0+50C Alphanumerical display 16 characters x 2 lines not back light 3 internal keys programmable function max 1250 Hz, 100mA, 40 Vdc N1 0/420mA RL 800 Yes 0,410m/s Yes Yes All the outputs are galvanically isolated from power supply EEprom stored measuring values on power failure Protected plug in for connection to PC or hand terminal Instrument with CE certificate Flow rate (volume) = 0,1% v.l. Out 4/20 mA = 0,12 % v.l. Frequency Out = 0,12% v.l. +/- 0,2 % 90265 Vac 4566 Hz; 1063Vdc/1545 Vac-4566Hz; 5 VA 4 VA ( AC ) / 3 W max ( DC )
The Manufacturer Guarantee Only English Text Available on our web site: www.isoil.com
Page 2 of 7
ML110_EN_DS016REV00_IS
ML 110
OVERALL DIMENSIONS
COMPACT VERSION
127
124
SEPARATE VERSION
79
65
The Manufacturer Guarantee Only English Text Available on our web site: www.isoil.com
Page 3 of 7
129
ML110_EN_DS016REV00_IS
168
25
ML 110
VISUALIZATION PAGES
Tolalizer values
Alarm ON visualization
The Manufacturer Guarantee Only English Text Available on our web site: www.isoil.com
Page 4 of 7
ML110_EN_DS016REV00_IS
ML 110
ELECTRICAL CONNECTIONS
E C
E C
1 2 3 4 5 6 7 8 9 10
20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1
+
11 12 13 14 15 16 17 18 19 20
4-20mA
ON/OFF OUTPUT
ON/OFF OUTPUT
Terminal 16 18 (Out 1) (Out 2)
43 V
ANALOG OUTPUT
ANALOG OUTPUT
+24V INTERNAL I
Terminal (10-)
The Manufacturer Guarantee Only English Text Available on our web site: www.isoil.com
Page 5 of 7
24V
OUT2
OUT1
24V SC B2
COILS
B1 SC
SC
E2
E1
ELECTRODES
ML110_EN_DS016REV00_IS
ML 110
1.7 AUTOZERO CAL. 6.7 USC.mA2 8.5 CURRENCY 8.6 CUR. DECIM. 8.7 XXX/dm + 8.7 RESET P-
2.7 TPUL2
Vedere la relativa tabella di riferimento nella descrizione delle funzione al menu 6 "USCITE"
The Manufacturer Guarantee Only English Text Available on our web site: www.isoil.com
1.SENSOR 8.DISPLAY
8.1 LANGUAGE 9.1 GG/MM/AA 00:00 9.2 ACQUISITION 9.3 INTERVAL 9.4 DISPLAY DATA 9.5 DISPLAY EVENTS 9.6 DISP. MIN/MAX 9.7 CLEAR DATA 9.8 CLEAR EVENTS 9.9 RESET MIN/MAX 10.2 SELFTEST 10.3 SIMULATION 8.2 D.RATE 8.3 CONTRAST 8.4 RESET T+ 8.5 RESET P+ 8.6 RESET T10.1 CALIBRATION
4.ALARM
11.DATI INTERNI
11.1 L2 KEYCODE 11.2 LOCK LEVEL 11.3 LOAD FACT. PRES. 11.4 LOAD USER PRES. 11.5 SAVE USER PRES. 11.6 HOURS 11.7 IGN.CAL.ERR. 11.8 KS
1.1 ND
2.1 FS1
3.1 T. CONST
1.2 COEFF. KA
2.2 FS2
1.3 S. MODEL
2.3 MU.TOT
2.4 IMP1
3.4 CUT-OFF
FUNCTIONS
Page 6 of 7
2.5 IMP2
3.5 FILTER
2.6 TPUL1
3.6 AUTOCAL
3.7 ATORANGE
2.8 FRQ1
3.8 E. SAVING
2.9 FRQ2
ML110_EN_DS016REV00_IS
ML 110
HOW TO ORDER
ML 110
Display A Blind execution (without display and programming keys) B Complete with 2 line back light display (each of 16 characters) and 3 programming keys Housing material - Protection rate 0 Nylon with glass fiber (IP 65) Version A Compact version with sensor MS.... (liquid maximum temperature 100 C) Separate version for wall monting, complete with mounting accessories B (C018 CABLE) Power supply 1 Power supply : 90 ... 265 V 45/66 Hz 2 Power supply : 18...63 V dc / 15...45 V ac - 45...66 Hz 9 Power supply : other Analogue output A Without analogue output B Analogue output 0/4...20/22 mA
EXAMPLE OF CODE FOR ORDER
ML 110 A 0 A 1 A
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17 18 19 20
90..265VAC
>HV<
>LV<
15...45VAC 10...63VDC
ML 110
COMPACT VERSION
SEPARATE VERSION
INTERNAL VIEW
In accordance with a continuous evolution of the product, the company reserves the right to modify without warning the information contained in this document