SCAIME S.A.S. - BP 501 - F-74105 ANNEMASSE Cedex Tel.

: +33 (0)4 50 87 78 64 - Fax : +33 (0)4 50 87 78 46

SIGNAL AMPLIFIER CMJ-CE SERIES


GENERAL CHARACTERISTICS
Power supply voltage Combined error Noise EMC susceptibility Temperature coefficient of zero Temperature coefficient of span for 0/10 V Temperature coefficient of span for 4/20 mA Capacitive load on 0/10 V output Common mode input voltage Nominal load cell input voltage Zero adjustment (for one 350 load cell) Span adjustment Output voltage Output current Load impedance on 0/10 V output Load impedance on 4/20 mA output Response time (gain min... gain max) Filtering (engaged with a jumper) Supply current (1...4 load cells) Nominal temperature range Storage temperature range 24 4 typ. 0.002 typ. 0.05 <2 typ. 0.005 ... 0.035 typ. 0.02 typ. 0.03 max. 1 >1 5.3 25% -1 ... +5 0.35 ... 3.5 0/10 4/20 2 500 typ. 0.6 ... 4.5 10 50 ... 110 0 ... + 70 -20 ... + 85 VDC % (FS)* % (FS)* % (FS)* % / C (FS)* % / C % / C nF VDC VDC mV/V mV/V VDC mA k ms Hz mA C C

CMJ "CARTE"

CMJ "RAIL DIN"


20

CMJ "BOITIER"

26

SCAIME

30 35 37.5

SCAIME

90.5

SCAIME

+24V Gnd Gnd Iout Vout

81.5 87.3

EX+ SIG- SIG+ EX-

+24V Gnd Gnd Iout Vout

EX+ SIG- SIG+ EX-

+24V Gnd Gnd Iout Vout

EX+ SIG- SIG+ EX-

4.8 7 58 72

2.9

82 16 50 80 4 x 4

4 x R 2.4

All dimensions in mm

* FS : Full Scale

CALIBRATION MODE 1 Configuration of the amplifier:


Place CV1 and CV2 jumpers according to the load-cell specifications and the requested output signal (see board on the right side of this page).
Position of the CV1 jumper according to the load-cell sensitivity to get 0/10 V or 4/20 mA. CV1 Jumper position no. 1 no. 2 no. 3 no. 4 no. 5 no. 6 no. 7 no. 8 X Load-cell sensitivity (mV/V) 0.35 0.45 0.60 0.75 0.95 1.25 1.60 2.10 2.60 to to to to to to to to to 0.45 0.60 0.75 0.95 1.25 1.60 2.10 2.60 3.50
12345678
CV1

2 Connections :
Connect the load-cell, the 24 V power supply and the output 0/10V or 4/20mA according to the sketches hereunder. Switch the power supply on. Wait about 15 minutes before starting the adjustment, in order for the weighing system to reach its nominal operating temperature.

CV1 jumper on position 1


12345678
CV1

CV1 jumper on position 8


12345678
CV1

3 Zero adjustment :
Make sure first that the load-cell doesn't carry any other load than the deadweight to be tared. Adjust the zero (taring) with potentiometers P2 (coarse) and P3 (fine). 0 /10V output : the voltmeter must display 0V. 4 /20mA output : the amperemeter must display 4mA.

4 Span adjustment :
After having applied a known load on the load-cell, adjust the span with potentiometer P1 in order to obtain the requested output.

Example : Load cell capacity = 500 kg Load to be weighed = 200 kg Output signal for 200 kg = (200/500) x 2 = 0.8 mV/V CV1 jumper should be placed on position 4

CV1 jumper on position X (2mV/V sensitivity)

5 Checking :
Remove the reference load from the load-cell and check the zero value. If necessary adjust zero and span one more time following the same procedure.

CV2 jumper position NF F

NF F

Output Signal non filtered filtered (10 Hz)

CV2 jumper on position NF

NF F

CV2

CV2 jumper on position F

Connection sketch for 0/10V output


Power supply + 20V to 28V Voltmeter
+
SCAIME

Connection sketch for 4 /20 mA output


Power supply + 20V to 28V Amperemeter +
R+/Ref+ A+ / Ex+ M-/SigM+/Sig+ A- / ExR-/Ref-

CV2

SCAIME

P1 Span
+24V Gnd Zero P3 P2 Zero (Coarse) Gnd (Fine) Iout 12345678 Vout EX+ SIGSIG+ EX-

P1 Span
+24V Gnd Zero P3 P2 Zero (Coarse) Gnd (Fine) Iout 12345678 Vout

Load-cell(s)

CN1

Load-cell(s)

CN1

Warning : "CARTE" and "RAIL DIN" versions of CMJ-CE being not protected by a housing, it is strongly recommended to pay a special attention while handling the board in order to avoid any accidental electrostatic discharge.

118703-B (GB)

R+/Ref+ A+/ Ex+ M-/SigM+/Sig+ A- / ExR-/Ref-

CV1

CV2

CV2

NF F

EX+ SIGSIG+ EX-

CV1

NF F

108 120

55