Sei sulla pagina 1di 47

Adllaneed lnstrurnents 'ne.

IiPR.SOO AIS
ppm Dxygen Trsnsmitter
with Optionsl SSlDple System

Dwner's an
2855 Metropolitan Place, Pomona, California 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, e-mail: info@aii1.com
Adllaneed Instrument5 Ine.
Table 01 Conlenls
Introduction 1
Qua'lity Control Certification 2
Safety 3
Features &. Specifications 4
Oe.eration 5
Maintenance 6
Spare Parts 7
Troubleshooting 8
Warranty 9
Material Safety Data Sheets 10
A/R
Explosion Proofing Electrical Connections Appendix A
Correlating readings - lCO display to 4-20mA signal output Appendix B
H2S SCrubber, Sample System, Media MSOS Appendix F
Maintenance H2S Scrubber &. Coalescing Filter Appendix G
1 InlroduClion
Your new oxygen transmitter incorporated an advanced electrochemical sensor specific to oxygen along with state-of-the-art
digital electronics designed to give you years of reliable precise oxygen measurements in variety of industrial oxygen
applications.
TO obtain maximum performance from your new oxygen transmitter, please read and follow the guidelines provided in this
Owner's Manual.
Every effort has been made to select the most reliable state of the art materials and components, to design the transmitter for
superior performance and minimal cost of ownership. This transmitter was tested thoroughly by the manufacturer prior to
shipment for best performance.
However, modern electronic devices do require service from time to time. The warranty included herein plus a staff of trained
professional technicians to quickly service your transmitter is your assurance that we stand behind every transmitter sold.
The serial number of this transmitter may be found on the inside the transmitter. You should note the serial number ,in the
space provided and retains this Owner's Manual as a permanent record of your purchase, for future reference and for warranty
considerations.
Serial Number:
Advanced Instruments Inc. appreciates your business and pledges to make every effort to maintain the highest possible quality
standards with respect to product design, manufacturing and service.
2
Adl/anced In!itrument!i Inc.
2 Qualitv Control Cenification
Date: 3/25/09 Customer: BP Exploration Co (Colombia) Ud Order No.: 4520050336 Pass
~
Model: GPR-1500AIS ppm Oxygen Transmitter
S/ N 1" O (;, S- b
sensor: ( ) GPR-12-333 ppm Oxygen sensor
(x) XLT-12-333 ppm Oxygen sensor
( ) XLT-12-333-F ppm Oxygen sensor S/N
Accessories: Owner's Manual
Configuration: A-1161-AIS-1 PCB Assembly Main/Dispaly SQftware Ver. ~ r J... -2
Power: ( ) A-1l66-AIS-AC PCB Assembly Alarms/AC Power 1l0V & Interconnection
( x) A-1l66-AIS-DC PCB Assembly Alarms/DC Power 12-24V & Interconnection
( ) A-1153-AIS-DC PeB Assembly Alarms/DC Power 12-24V & Interconnection
SPDT relays (B-3208, A-3267, Ex-proof window cover)
Ranges: 0-10 ppm, 0-100, 0-1000 ppm, 0-1%, 0-25%
Barometric pressure and temperature compensation
NEMA 4X rated wall mount enclosures
Test: Pass Pass
Set default zero v'" Noise level < 1.0% FS ./
Set default span @ 600uA v"" Alarm delay
Zero calibration / Alarm bypass
Span calibration _V--;_ Alarm configurations; ALARM 1, ALARM 2
Analog signal output 4-20mA futl scale _t/'--..,.-- Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within 10-50% FS 7' Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas < 2% FS over 24 hour period 1/ Alarm system tall, dry contact
Overall inspection for physical defects
Options: ( x) A-3393-5 sample System
ENCL-1079 NEMA 4X System Enclosure 23x19x9.8", fully plumbed and pre-drilled for power
Notes: 1 of 1 due by 4/15/09
2
Advaneed 'nstrulDents 'ne.
3 Saletv
General
This section summarizes the essential precautions applicable to bhe GPR-1500AIS ppm Oxygen Transmitter. Additional
precautions specific to individual transmitter are contained in the following sections of this manual. To operate the transmitter
safely and obtain maximum performance follow the basic guidelines outlined in this Owner's Manual.
Caution: This symbol is used throughout the Owner's Manual to Caution and alert the user to recommended safety and/or
operating guidelines.
Danger: This symbol is used throughout the Owner's Manual to identify sources of immediate Danger such as the presence of
hazardous voltages.
Read Instructions: Before operating the transmitter read the instructions.
Retain Instructions: The safety precautions and operating instructions found in the Owner's Manual should be retained for
future reference.
Heed Warnings: Follow all warnings on the transmitter, accessories (if any) and in this Owner's Manual.
Follow Instructions: Observe all precautions and operating instructions. Failure to do so may result in personal injury or
damage to the transmitter.
Pressore and Flow
Inlet Pressure: GPR-1500AIS ppm Oxygen Transmitters are designed for flowing samples, equipped with 1/8" bulkhead tube
fitting connections on the side of the unit (unless otherwise indicated, either fitting can serve as inlet or vent) and are intended
to operate at positive pressure regulated to between 5-30 psig.
Caution: If equipped with an optional H2S scrubber, inlet pressure must not exceed 30 psig.
Outlet Pressure: The sample gas vent pressure should be atmospherk.
InstaJlation
Oxygen sensor: DO NOT open the sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner's Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed of in accordance with local regulations.
Mounting: The transmitter is approved for indoor or outdoor use. Mount as recommended by the manufacturero
Power: Supply power to the transmitter only as rated by the specification or markings on the transmitter enclosure. The wiring
that connects the transmitter to the power source should be installed in accordance with recognized electrical standards and so
they are not pinched particularly near the power source and the point where they attach to the transmitter. Never yank wiring
to remove it from a terminal connection. Power consumption is 30 watts, 40 watts with the optional De powered heater.
Operating Temperature: The maximum operating temperature is 45 e on an intermittent basis unless the user is willing to
accept a dramatic reduction in expected sensor life - refer to analyzer specification where expected sensor Iife is specified at
less than 1000 ppm oxygen at 25e and 1 atmosphere of pressure.
Heat: Situate and store the transmitter away from sources of heat.
4
Adllaneed Instruments Ine.
Liquid and Object Entry: The transmitter should not be immersed in any liquid. Care should be taken so that liquids are not
spilled into and objects do not fall into the inside of the transmitter.
Handling: Do 1l0t use force when using the switches and knobs. Before moving your transmitter be sure to disconnect the
wiringjpower cord and any cables connected to the output terminals ,Iocated on the transmitter.
Maintenance
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the transmitter for the operator to service.
Only trained personnel with the authorization of their supervisor should conduct maintenance.
Oxygen sensor: DO NOT open the sensor. The sensor contains a corrosive Iiquid' electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner's Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should
be disposed ofin accordance with local regulations.
Troubleshooting: Consult the guidelines in Section 8 for advice on the common operating errors before concluding that your
transmitter is faulty. Do not attempt to service the transmitter beyond those means described in this Owner's Manual.
Do not attempt to make repairs by yourself as this will void the warranty as per Section 10 and may result in electrical' shock,
injury or damage. AII other servicing should be referred to qualified service personnel.
Cleaning: The transmitter should be c1eaned only as recommended by the manufacturero Wipe off dust and dirt from the
outside of the unit with a 50ft damp c10th then dry immediately. Do not use solvents or chemicals.
Nonuse Periods: If the transmitter is equipped with a range switch advance the switch to the OFF position and disconnect the
power when the transmitter ,is left unused for a long period of time.
4 Features & Specifications
Specifications and pricing are subject to change without notice. See last page for current specifications.

50peration
Principie of Operation
The GPR-1500AIS oxygen transmitter incorporates a variety of ppm
range advanced galvanic fuel cell type sensors. The transmitter is
5
Advanced InstrulDents IDe.
configured in two sections. The signal processing electronics and sensor are housed in a general purpose NEMA 4X rated
enclosure. The terminals for incoming power, power supply, signal output, alarm relay contacts and intrinsic safety barriers are
mounted on a PCB housed in an explosion proaf enclosure.
The two sets of electronics are interconnected using an explosion proof Y-fitting, explosion proof packing fiber and sealing
cement - see Appendix A. Once connected, the intrinsic safety barriers Iimit the amount of voltage that f10ws to and from the
signal processing electronics effectively preventing an explosive condition and meets the intrinsic safety standards required for
use in Class 1, Division 1, Groups B, C, D hazardous areas.
Advanced Galvanic Sensor Technology
The sensors function on the same principie and are specific for oxygen. They measure the partial pressure of oxygen from low
ppm to 100% levels in inert gases, gaseous hydrocarbons, helium, hydrogen, mixed gases, acid gas streams and ambient airo
Oxygen, the fuel for this electrochemical transducer, diffusing into the sensor reacts chemically at the sensing electrode to
produce an el ectricaI current output proportional to the oxygen concentratjon in the gas phase. The sensor's signar output is
linear over all ranges and remains virtually constant over its useful life. The sensor requires no maintenance and is easily and
safely replaced at the end of its useful life.
Proprietary advancements in design and chemistry add significant advantages to an extremely versatile oxygen sensing
technology. Sensors for low ppm analysis recover from air to ppm levels in minutes, exhibit longer Iife and reliable quality. The
expected life of our new generation of percentage range sensors now range to five and ten years with faster response times
and greater stablity. Another significant development involves expanding the operating temperature range for percentage range
sensors from -30C to 50C.
Electronics
The signar generated by the sensor is processed by state of the art low power mcro-processor based digital circuitry. The first
stage amplifies the signa!. The second stage eliminates the low frequency noise. The third stage employs a high frequency filter
and compensates for signal output variations caused by ambient temperature changes. The result is a very stable signa!.
Sample oxygen is analyzed very accurately. Response time of 90% of fuI! scale is less than 10 seconds (actual experience may
vary due to the integrity of sample line connections, dead volume and f10w rate selected) on all ranges under ambient
monitoring conditions. 5ensitivity is typically 0.5% of full scale low range. Oxygen readings may be recorded by an external
device va the 0-lV signal output jack.
The circuit for the 4-20mA signal output and two adjustable alarms is powered by a DC/OC transformer that requires a 12-28
VDC power source and separate wiring for the outputs. A loop power source is not sufficient to power to the circuito The 4
20mA output is also represented on full scale oxygen readings to an externaI device.
Sample System:
The GPR-1500AIS is supplied wthout a sample conditioning system thereby giving users the option of adding their own or
purchasing a factory designed sample conditioning system. Whatever the choice, the sample must be properly presented to the
sensor to ensure an accurate measurement. Users interested in adding their own sample conditioning system should consult the
factory. Advanced lnstruments lne. offers a fuI! line of sample handling, conditioning and expertise to meet your application
requirements. Contact us at 909-392-6900 or e-mail usatinfo@aii1.com.
Accuracv & Callbration
SinglePoint Calibration: As previously described
the galvanic oxygen sensor generates an eledrical
current proportional to the oxygen concentration in
the sample gas. In the absence of oxygen the sensor
exhibits an absolute zero, e.g. the sensor does not
generate a current output in the absence of oxygen.
Given these linearity and absolute zero properties,
single point calibration fs possible.
Pressure: Because sensors are sensitive to the
partial pressure of oxygen in the sample gas their
output is a function of the number of molecules of
-fEMPeR"lURE
Adlliloced lostrument!!!i loc.
oxygen 'per unit volume'. Readouts in percent are permissible only when the total pressure of the sample gas being analyzed
remains constant. The pressure of the sample gas and that of the calibration gasees) must be the same (reality < 1-2 ps).
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise known as an
'oxygen diffusion limiting barrier' and all diffusion processes are temperature sensitive, the fact the sensor's electrical output
will vary with temperature is normal. This variation is relatively constant 2.5% per oc. A temperature compensation circuit
employing a thermistor offsets this effect with an accuracy of S% or better and generates an output function that is
independent of temperature. There is no error if the calibration and sampling are performed at the same temperature or if the
measurement is made immediately after calibration. Lastly, small temperature variations of 10-150 produce < 1% error.
Accuracy: In Iight of the aboye parameters, the overall accuracy of an analyzer is affected by two types of errors: 1) those
producing 'percent of reading errors', illustrated by Graph A below, such as S% temperature compensation circuit, tolerances
of range resistors and the 'play' in the potentiometer used to make span adjustments and 2) those producing 'percent of full
scale errors', iIIustrated by Graph B, such as 1-2% linearity errors in readout devices, which are realty minimal due to today's
technology and the fact that other errors are 'spanned out' during calibration.
Graph C illustrates these 'worse case' specifications that are typically used to develop an transmitter's overall accuracy
statement of < 1% of fuI! scale at constant temperature or < 5% over t-he operating temperature range. QC testing is typically
< 0.5% prior to shipment.
,
/
/
/ '"
/
/
/
'"
,"
/
/
'"
"
"
'"
/ ,
/
/
/
,,'
'
,
/
/
/
/
/
/ '"
/ ,
/
'"
"
'"
/
/
/",'"
/
'",
,..
/
/
A. PER.CENl or REAI>ING
B. ?fRCElil or rUlL SCALE
C. OVEIIAll ACCURACY
ERIIORS
EP.1I01lS
Example 1: As llustrated by Graph A any error, play in the multi-turn span pot or the temperature compensaton circuit,
during a span adjustment at 20.9% (air) of full scafe range woufd be multiplied by a factor of 4.78 (100/20.9) if used for
measurements of 95-100% oxygen concentrations. Conversely, an error during a span adjustment at 100% of fuI! scale range is
reduced proportionately for measurements of lower oxygen concentrations.
Zera Calibration: In theolY, the eleetrochemical galvanic fuel cell type oxygen has an absolute zero meaning it produces no
signal output when exposed to an oxygen free sample gas. In realitv, expect the analyzer to generate an oxygen reading when
sampling oxygen free sample gas due to contaminaton or quality of the zera gas; minor leakage in the sample line connections;
residual oxygen dissolved in the sensor's electrolyte; and, tolerances of the electronic components. The lero Offset capability of
the analyzer is limited to 50% of lowest most sensitive range available with the analyzer.
Recommendation 1: lero calibrabon, see Determining True lero Offset below, is recommended only for online analyzers
performing continuous analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. analysis below
0.5 ppm on the 10 ppm range, or below 0.1% (1000 ppm) with a percent analyzer.
Note 1: Once the zero offset adjustment is made, zero calibration is not required again until the sample system connections are
modified, or, when installing a new oxygen sensor. As a result, zera calibration is not practical and therefore not recommended
for higher ranges or portable analyzers.
Determining True lero Offset: Allow the transmtter approximately 24 hours to stabilize with flowing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device.
Note 2: 24 hours is requred to assure the sensor has consumed the oxygen that has dissolved into the electrolyte insi de the
sensor while exposed to air or percentage levels of oxygen. For optimum accuracy, utilize as much of the actual sample system
as possible.
Span Calibration: lnvolves adjusting the transmitter electronics to the sensor's signal output at a given oxygen standard.
Regardless of the oxygen concentration of the oxygen standard used, a typical span calibration takes appraximately 10 minutes.
7
Adllanced InstrunJents Ine.
Note 3: The amount time required to get the analyzer back on line for normal use is influenced by a.) the level of oxygen
analysis anticipated during normal operation (also determines the initial analyzer seleetion), and, b.) whether the sensor is new
or has been in service tor at least two weeks. General guidelines for analyzers to come online following span calibration and
exposure to a zero/purge!sample gas with an oxygen content below the stated thresholds:
,. measurements aboye 1000 ppm or 0.1% require less than 3 minutes
,. measurements above 100 ppm (parts-per-million analyzer) require less than 10 minutes
;. measurements below 10 ppm (part-per-million analyzer) require 20 minutes if the sensor has been in service at ppm levels
for at least two weeks, and, 60 minutes if the sensor is new assuming the zero/purge!sample gas has an oxygen
concentration below 1 ppm
Recommendation 2: For 'optimum calibration accuracy' calibrate with a span gas approximating 80% of the full scale range
one or two ranges higher than the full scale range of interest (normal use) to achieve the effect illustrated on Graph A and
Example 1. Always calibrate at the sarne temperature and pressure of the sample gas stream.
Note 4: Calibrating with a span gas approximating 10% of the full scale range near the expected oxygen concentraton of the
sample gas is acceptable but less accurate than 'optimum calibration accuracy' method recommended - the rnethod usually
depends on the gas available. calibrating at the same 10% of the full scale range for measurements at the higher end of the
range results in magnification of errors as discussed in Graph A and Exaniple 1 and is not recommended. Of course the user can
(
--always elect at his discretion to accept an accuracy error of 2-3% of full scale range if no other span gas is available.
Air Calibration: Based on the inherent Iinearity of the electrochemical galvanic fuel cell type oxygen sensor enables the user to
calibrate the analyzer with ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest
most sensitive range available with the analyzer - there is no need to recalibrate the anafyzer with span gas containing a lower
oxygen concentration. calibrating either a ppm or percent analyzer with ambient air (with the exception of Oxygen Purity
Analyzers intended to measure elevated oxygen levels ranging from 50-100% oxygen) on the CAL or 0-25% range meets the
80% criteria discussed in Recommendation 2.
Recommendation 3: Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated
oxygen levels ranging from 50-100% oxygen) when operating a percent analyzer, installing and replacing a ppm oxygen sensor,
to verify the oxygen content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer
(immediately following air calibration reintroduce a gas with a low oxygen concentration to expedite the return to ppm level
measurements as per Note 3).
Install'ation Considerations
Gas Sample Stream: Ensure the gas stream composition of the application s consistent with the specifications and review the
application conditions before initiating the installation. Consult the factory if necessary to ensure the sample is suitable for
analysis.
Note: In natural gas applications such as extraction and transmission, a low voltage current s applied to the pipeline itself to
inhibit corrosion. As a result, electronic devices can be affected unless adequately grounded.
Contaminant Gases: A gas scrubber and f10w indcator with integral meterng valve are required upstream of the transmitter
to remove interfering gases such as oxides of sulfur and nitrogen or hydrogen sulfide that can produce false readings, reduce
the expected tife of the sensor and void the sensor warranty if not identified at time of order placement. Installation of a
suitable scrubber is required to remove the contaminant from the sample gas to prevent erroneous analysis readings aod
damage to the sensor or optional components. Consult the factory for recommendations concerning the proper selection and
installation of components.
8
Adllanced Instruments Ine.
Expected Sensor Lite: With reference to the publish specification located as the last page of this manual, the expected life of
all oxygen sensors is predicated on oxygen concentration 1000 ppm or air), temperature (77F/25q and pressure (1
atmosphere) in "normal" applications. Deviations are outside the specifications and will affect the Iife of the sensor. As a rule of
thumb sensor life is inversely proportional to changes in the parameters.
Optimum Accuracy: Determine if lero Calibration is recommended for your application. If it is lera Calibration should
precede Span Calibration and both should be repeated after the analyzer has been allowed to stabilize, typically 24-36 hours
after installation. For Span Calibration use a certified span gas with an oxygen content (balance nitragen) approximating BO% of
the next higher full scale range aboye the intended measuring range is recommended for optimum accuracy, see Calibration
and Accuracy.
Assuming the initial zera is performed according to the procedure described herein, the analyzer should not require lera
Calibraton again until the either the sensor is replaced or a change is made to the sample system or gas lines, and, it should
not require 5pan Calibration again for up to 3 months under "normal" application conditions as described in the published
specifications. One of the unique features of analyzers based on the electrochemical galvanic fuel cell type oxygen sensor is the
fact that t can be field calibrated at the users discretion to whatever standard of certified span gas the user elects to use.
Zero calibration: In theorv, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. In
~ expect the transmitter to generate an oxygen reading when sampling oxygen free sample gas due to contamination or
quality of the zero gas; minor leakage in the sample line connections; residual oxygen dssolved in the sensor's electrolyte; and,
tolerances of the electronic components.
lera calibration, see Determining True lera Offset below, is recommended only for online analyzers performing continuous
analysis below 5% of the lowest most sensitive range available with a ppm analyzer, e.g. anatysis below 0.5 ppm on the 10 ppm
range, or below 0.1% (1000 ppm) with a percent analyzer. Note: Once the zera offset adjustment is made, zera calibration is
-ot required again until the sample system connections are modified, or, when installing a new oxygen sensor. As a result, zero
calibration is not practical and therefore not recommended for hgher ranges or portable analyzers.
Determining True Zero Offset: Allow the transmitter approximately 24 hours to stabilize with f10wing zero gas as evidenced by a
stable reading or horizontal trend on an external recording device. Note: 24 hours is required to assure the sensor has
consumed the oxygen that has dissolved into the electrolyte inside the sensor while exposed to air or percentage levels of
oxygen. For optmum accuracy, utilize as much of the actual sample system as possble.
Span Calibration: Involves adjusting the transmitter electranics to the sensor's signal output at a given oxygen standard, e.g.
a certified span gas with an oxygen content (balance nitragen) approximating 80% of the next higher full scale range aboye the
ntended measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
Recommendation: Based on the inherent linearity of the gafvanic oxygen sensor enables the user to calibrate the analyzer
with ambient air (20.9% oxygen) and operate the analyzer within the stated accuracy spec on the lowest most sensitive range
available with the analyzer - there is no need to recalibrate the analyzer with span gas containing a lower oxygen
concentraton.
calibrating either a ppm or percent analyzer with ambient air (with the exception of Oxygen Purity Analyzers intended to
measure elevated oxygen levels rangng fram 50-100% oxygen) on the CAL or 0-25% range meets the 80% criteria discussed
aboye. Air calibrate the analyzer (with the exception of Oxygen Purity Analyzers intended to measure elevated oxygen levels
rangng from 50-100% oxygen) when operatng a percent analyzer, installing and replacing a ppm oxygen sensor, to verify the
oxygen content of a certified span gas or when a certified span gas is not available to calibrate a ppm analyzer (immediately
following air calibraton reintroduce a gas with a low oxygen concentratan to expedite the return to ppm level measurements).
Materials: Assemble the necessary zera, purge and span gases and optional components such as valves, coalescing or
particulate filters, and, pumps as dictated by the application; stainless steel tubing is essential for maintaining the integrity of
the gas stream for ppm and percentage range (above or below ambient air) analysis; hardware for mounting.
Temperature: The sample must be sufficiently cooled before it enters the transmitter and any optional components. A coiled
10 foot length of '14" stainless steel tubing is sufficient for coolng sample gases as hgh as 1,BOooF to ambiento
Pressure &. Flow
9
Advanced Instruments loc.
AII electrochemical oxygen sensors respond to partial pressure changes in oxygen. lhe sensors are equally
capable of analyzing the oxygen content of a f10wing sample gas stream or monitoring the oxygen
concentration in ambient air (such as a confined space such in a control room or an open area such as a
landfill or bio-pond).
sample systems and/or flowing gas samples are generally required for applications involving oxygen
measurements at levels other than ambient air and when the pressure exceeds ambiento In these
situations, the use of stainless steel tubing and fittings is criticaI to maintaining the integrity of the gas
stream to be sampled and the inlet pressure must always be higher than the pressure at the outlet vent
which is normally at atmospheric pressure.
Flow Through Configuration: The sensor is exposed to sample gas that must f10w or be drawn through
metal tubing insidethe transmitter. lhe internal sample system ineludes 1/8" compression inlet and vent
flttings, a stnless steel sensor housing with an o-ring seal to prevent the leakage of air and stainless
steel tubing.
I
Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, f10w rates aboye 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample gas at
the higher f10w rate. The direction the sample gas flows is not important,'thus either tube fitting can selVe as the inlet or vent
just not simultaneously.
A flow indicator with an integral metering valve upstream of the sensor is recommended as a means of controlling the flow rate
of the sample gas. A flow rate of 2 SCFH or 1 liter per minute is recommended for optimum performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the f10w indicator when gas is f10wing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty). To avoid generating a vacuum on the sensor (as described aboye) during operation, always select and install
the vent fitting first and remove the vent fitting last.
Applcaton Pressure - Positive: A f10w indicator with integral metering valve positioned upstream of the sensor is
recommended for controlling the sample flow rate between 1-5 SCFH. To reduce the possibility of leakage for low ppm
rneasurements, position a rnetering needle valve upstream of the sensor to control the flow rate and position a flow indicator
downstream of the sensor. If necessary, a pressure regulator (with a metallic diaphragm is recommended for optimum
accuracy, the use of diaphragms of more permeable materials may result in erroneous readings) upstream of the flow control
valve should be used to regulate the nlet pressure between 5-30 psig.
Caution: If equipped with a H2S scrubber, inlet pressure must not exceed 30 psig.
Application Pressure - Atmospheric or Slightly Negative: For accurate ppm range oxygen measurements, an optional
external sampling pump should be positioned downstream of the sensor to draw the sample from the process, by the sensor
and out to atmosphere. A f10w meter is generally not necessary te obtain the recommended flow rate with most sampling
pumps.
Caution: If the transmitter is equipped with an optional flow indicator with integral metering valve or a metering flow control
valve upstream of the sensor - open the metering valve completely to avoid drawing a vacuum on the sensor and placing an
undue burden on the pump.
If pump loading is a consideration, a second throttle valve on the pump's inlet side may be necessary to provide a bypass path
so the sample f10w rate is within the aboye parameters.
Recommendations to avoid erroneous oxygen readings and damaging the sensor:
;;.. Do not place your finger over the vent (it pressurizes the sensor) to test the f10w indicator when gas is f10wing to the
sensor. Removing your finger (the restriction) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).
)- Assure there are no restrictions in the sample or vent lines
;;.. Avoid drawing a vacuum that exceeds 14" of water column pressure - unless done gradually
,. Avoid excessive f10w rates aboye 5 SCFH which generate backpressure on the sensor.
'" Avoid sudden releases of backpressure that can severely damage the sensor.
,. Avoid the collection of Iiquids or particulates on the sensor, they block the diffusion of oxvgen into the sensor.
10

Advaneed Instruments Ine.
>- If the transmitter is equipped with an optional integral sampling pump (positioned downstream of the sensor) and a fiow
control metering valve (positioned upstream of the sensor), completely open the flow control metering valve to avoid
drawing a vacuum on the sensor and placing an undue burden on the pump.
Moisture &. Particulates: Installation of a suitable coalescing or particulate filter is required to remove condensation, moisture
and/or particulates from the sample gas to prevent erroneous analysis readings and damage to the sensor or optional
components. Moisture and/or particulates do not necessarily damage the sensor, however, coUection on the sensing surface
can block or inhibit the diffusion of sample gas into the sensor resulting in a reduction of sensor signal output - and the
appearance of a sensor failure when in fact the problem is easily remedied by blowing on the front of the sensor. Consult the
factory for recommendations concerning the proper selection and installation of components.
Gas Connections: Inlet and outlet vent gas lines for ppm analysis require 1/8" or Y4" stainless steel compression flttings; hard
plastic tubing with a low permeability factor can be used percentage range measurements.
Power Connection: Locate the appropriate power source to meet the analyzer or transmitter requirements, ensure that is
properly grounded and meets the area classification.
IMounting tbe Transminer
The GPR-1S00AIS consists of two interconnected enclosures (without the optional
sample conditioning system and panel) and measures 8"H x x 7''D. This
configuration is designed ro be mounted directly to any flat vertical surface, wall or
bulkhead plate with eight (8) of the appropriate screws. To facilitate servicing the
interior af the transmitters, pasition it approximately S feet off ground leve!.
Remove the four (4) screws securing the top section of the enclosure, set them
aside for reinstallation and raise the hnged top section 180
0
until it locks in place.
Locate the mounting holes cast into the enclosure.
Secure the enclosure to a vertical surface approximately S feet from the f100r or a level accessible to service personnel. This
requires the user to supply four (4) additional proper sze screws and anchors.
Caution: Do not remove or discard the gaskets from either the enclosure or junction box. Failure to reinstall either gasket will
void the NEMA 4 rating and RFI protection. The transmitters design provides protection from RFI that is maintained by leaving
specific mating areas of the enclosure unpainted to maintain conduetivity the gasket, top and bottom sections of the enclosure.
These unpainted areas are protected by gaskets and contribute to maintaining the NEMA 4 rating. Do not paint these areas.
Panting will negate the RFI proteetion.
Note: If equipped with the optional H25 sample conditioning system, the transmitter and sample system are mounted to a back
panel which has four (4) holes for mounting the 15-3/4"H x lS-3/4'W x 7"D panel to any vertical flat surface.
Gas Connhections
The with its standard flow through configuraton is designed for positve pressure samples and requires
connectons for incoming sample and outgoing vent Iines. Zero and span inlet ports are offered as part of the optional sample
systems. The user is responsible for calibratan gases and the required components, see belaw.
Flow rates of 1-5 SCFH cause no appreciable change in the oxygen reading. However, flow rates aboye 5 SCFH generate
backpressure and erroneous oxygen readings because the diameter of the integral tubing cannot evacuate the sample gas at
the higher flow rateo A fiow indicator with an integral metering valve upstream of the sensor is recommended as a means of
controlling the flow rate of the sample gas. A f10w rate of 2 SCFH or 1 liter per minute is recommended for optimum
performance.
Caution: Do not place your finger over the vent (it pressurizes the sensor) to test the flow indicator when gas is f10wing to the
sensor. Removing your finger (the restrietion) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).
Procedure:
11
Adllaoced lostruments Inc.
1. Caution: Do not change the factory setting until instructed to do in this manual.
2. Designate one of the bulkhead tube fittings as the VENT and the other 5AMPLE.
3. Regulate the pressure as described in Pressure and Flow aboye.
4. Connect a 1/8" vent line to the compression fitting to be used for venting the sample.
5. Connect a 1/8" ZERO, SPAN or SAMPLE line to the fitting designated SAMPLE.
6. If equipped with optional fittings and/or sampie system, connect the ZERO and SPAN gas lines.
7. Allow gas to flow through the transmitters and set the flow rate to 2 SCFH.
8. Note: If equipped with the optional H2S sample conditioning system: Regulate the pressure so that it does not exceed
30 psig use V4" tubing to make the appropriate connections as labeled on the sample panel.
Electrical Connections
Incoming power, power failure and set point alarm, and, output connections are
made to a terminal block mounted on a PeS located in the lower explosion proof
enclosure.
The PeS also includes a transformer to power the alarm relays and intrinsic
safety barriers that limited the amount of voltage going to the upper electronics
enclosure. With proper insulation of the incoming power (see Appendix A), this
configuration the GPR-1500AIS meets the intrinsic safety standards for use in
Class 1, Division 1, Groups A-D hazardous areas.
Caution: The integral 4-20mA converter s internally powered and does not
require external power. DO NOT supply any voltage to either of the two
terminals of the 4-20mA output or the 4-20mA converter will be damaged.
To assure proper grounding, connect the 4-20mA signal output to the externaI device (PLC, DCS, etc.) before attempting any
zero or span adjustments.
Procedure:
Power requirements consist of a two wire shielded cable and a 12-28V DC with negative ground power supply.
1. Unscrew the cone shaped cover from the lower enclosure.
2. 5eparate the shielding from the wires of the cable.
3. Ensure the positive and negative terminals of the power supply are connected to the appropriate terminals of the barrier
strip as marked.
4. Connect the shielding of the cable to the ground screw inside the enclosure. Note: The terminals snap together, making it
possible to detach the section with the ground, install the shielded cable and reinstall.
5. Replace the cover.
6. The 4-20mA current output is obtained by conneeting the current measuring device between the positive and negative
terminalslabeled OUTPUT 4-20mA.
7. To check the signal output of the 4-20mA ElI integrated circuit connect an ammeter as the measuring device and confirm
the output is within O.lmA of 4mA.
8. Caution: To assure proper grounding, connect the 4-20mA signal output to the extemal device (PLC, DCS, etc.) before
attempting any zero or span adjustments.
Installino the Oxvgen Sensor
The GPR-1500AIS ppm Oxygen Transmitter is equipped with an integral oxygen sensor.
They have been tested and calibrated by the manufacturer prior to shipment and are fully
operational from the shipping container.
Caution: AII transmitters must be calibrated once the instaliation has been completed and
periodically thereafter as described below. Following the initial installation and calibration,
allow the transmitters to stabilize for 24 hours and calibrate with certified span gas.
12
Adllanced In!!itruments Ine.
Caution: DO NOT open the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the OWner's Manual appendix. Avoid contact with any liquid or
crystal type powder in or around the sensor or sensor housing, as either could be a form of
electrolyte. Leaking sensors should be disposed of in manner similar to that of a common battery in
accordance with local regulations.
Procedure:
1. The sensor has not been installed at the factory (in standard configuration there are no valves
to isolate the sensor) and it will be necessary to install the sensor in the field.
2. Caution: Do not change the factory settings until instructed to do in this manual.
3. Connect the gas lnes as previously described.
4. Purge the oxygen trapped in the newly connected gas lines for 3-5 minutes.
5. Flow zero gas or sample gas with a low ppm oxygen c:oncentration to the analyzer at the
predetermined flow rate of 2 SCFH.
6. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located under the
sensor housing, see photo.
7. Rotate the upper section of the sensor housing 900 to disengage from the clamp.
8. Remove the upper section by pulling it straight up and place it on a smooth surface.
9. Select the AUTO RANGING option from the SAMPLE menu with gas f10wing to the analyzer.
10. Remove the oxygen sensor from the bag and remove the red shorting device (including the gold ribbon) from the PCS
located at the rear of the sensor. Minimize the time the sensor is exposed to ambient airo
11. Immediately place the sensor in the bottom section of the sensor housing with the PCS facing up.
12. lmmediately place the upper section of the sensor housing over the sensor, gently push the upper section downward and
rotate 90 to engage the clamp.
13. Finger tighten the clamp bolt and then tighten it one full turo with the 5/16 wrench to securely lock the two sections of the
sensor housing.
14. The analyzer will OVER RANGE for a short period of time as indicated by the graphical LCD display.
15. Wait until the display shows a meaningful oxygen reading and begins to approach the expected oxygen content of the
sample gas.
Spao Gas Preparation
Caution: Do not contaminate the span gas cylinder when connecting the regulator. Bleed the air filled regulator (faster and
more reliable than simply flowing the span gas) before attempting the initial calibration of the instrumento
Required
1. Certified span gas cylinder with an oxygen concentration, balance nitrogen, approximating 80% of the fuff scale range
aboye the intended measuring range.
2. Regulator to reduce pressure to between 5 and 30 psig.
3. Flow meter to set the f10w between 1-5 SCFH,
4. Suitable fittings and a 4-6 ft. in length of 1/8" dia. metal tubing to connect the regulator to the f10w meter inlet
5. Suitable fitting and a 4-6 ft. in length of 1/8" dia. metal tubing to connect from the f10w meter vent to tube fitting
designated SAMPLE IN on the GPR-1200.
Procedure:
1. With the span gas cylinder valve c1osed, install the regulator on the cylinder.
2. Open the regulator's exit valve and partially open the pressure regulator's control knob.
3. Open slightly the cylinder va/ve.
4. Loosen the nut connecting the regulator to the cylinder and bleed the pressure regulator.
5. Retighten the nut connecting the regulator to the cylinder
6. Adjust the regulator exit valve and slowly bleed the pressure regulator.
7. Open the cylinder valve completely.
8. Set the pressure between 5-30 psig using the pressure regulator's control knob.
9. Caution: Do not exceed the recommended flow rateo Excessive f10w rate could cause the backpressure on the sensor and
may result in erroneous readings and permanent damage to the sensor.
13
,Adllanced In!!itruments loe.
Establishinl Power to lbe ElecuoniCs
Once the two wires of the shielded cable are properly connected to the terminals insde the junction box as described aboye,
connect the other end of the two wires to a suitable 12-28V DC power supply with negative ground such as a PLC, DCS, etc.
The digital display responds instantaneously. When power is applied, the transmitter performs several diagnostic system status
checks termed "START-UP TEST" as iIIustrated below:
START-UP TESI
ELECTRONICS - PASS
TEMP SENSOR - PASS
BAROMETR'IC SENSOR - PASS
REV.l.6l
Note: The transmitter display defaults to the sampling mode when 30 seconds elapses without user interface.
3.3 PPM
AUTO SAMPUNG
10 PPM, RANGE
24.5 e 100 KPA
LOl 2PPM lOPPM HI2
Menu Navigation
The four (4) pushbuttons located on the front of the transmitter operate the micro-processor:
1. blue ENTER (select)
2. yellow UP ARROW
3. yellow DOWN ARROW
4. green MENU (escape)
MainMeno
Access the MAIN MENU by pressing the MENU key:
MAIN MENU
!ji. l,ti':1Q!3
MANUAL SAMPlE
CAUBRATION
CONFIG ALARMS
BYPASS ALARMS
14
Adllanced In!!itruments Ine.
Banue Selection
The GPR-1500AIS transmitter is equipped with five (5) standard measuring ranges (see specification) and provides users with a
choice of sampling modes. By accessing the MAIN MENU, users may select either the AUTO SAMPUNG (ranging) or MANUAL
SAMPUNG (to lock on a single range) mode.
Note: For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale
MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required - for example, a span gas with an
80 ppm oxygen concentration with the balance nitrogen would dictate the use of the 0-100 ppm full scale range for calibraton
and a 0-10 ppm measuring range.
Auto Sampling:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.
3. Press the ENTER key to select the highlighted menu option.
4. The display returns to the sampling mode:
MAl" MENU
,,ui'li': IQ i ~
MANUAL SAMPLE
CAUSRATION
CONFIG ALAR.MS
BY:PASS AJ.ARMS
3.3 PPM
AUTOSAMPIl.ING,
10 PPM RANGE
24.5,C 100 KPA
LOl 2PPM 10PPM. HI2
The display will shift to the next higher range when the oxygen reading (actually the sensor's signal output) exceeds 99.9% of
the upper limit of the current range. The display will shift to the next lower range when the oxygen reading drops to 85% of the
upper limit of the next lower range.
For example, if the transmitter is reading 1% on the 0-10% range and an upset occurs, the display will shft to the 0-25% range
when the oxygen reading exceeds 9.9%. Conversely, once the upset condition is corrected, the display will shift back to the 0
10% range when the oxygen reading drops to 8.5%.
Manual Sampling:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade OJrsor using the ARROW keys to highlight MANUAL SAMPL'E.
3. Press the ENTER key to select the highlighted menu option.
4. The following display appears:
MAINl MENU
AUTO SAMPLE
1: ',':11,.,./,,':1 Q! 3
CALIBRATION
CONFIG ALARMS
BVPASS ALARMS
>
MANUAL RANGE'
25%
1%,
1000 PPM
100 P,PM
![IIQ:
5. Advance the reverse shade cursor using the ARROW keys to highlight the desired MANUAL RANGE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with the range selected and oxygen concentration of the sample gas:
lS
Advanced Instruments Ine.
MANUAL RANGE

1
%
1000 PPM
100 PPM
>
3.3 PPM
MANUAL SAMPLING
10 PPM RANGE
24.5 e 100 KPA
LO! 2PPM 10PPM HI2
The display will not shift automatically. Instead, when the oxygen readng (actually the sensor's signal output) exceeds 110% of
the upper Iimit of the current range an OVER RANGE warning will be displayed.
[Note: 3.3 PPM displayed on ranges of 1000 PPM and below, 0.0% displayed on ranges of 1% and aboye.
13.3 PPM
[ti14 1f.tij;
MANUAL SA}4PUNG
1000 PPM RANGE
24.5C 100KPA
LOl 1!DPPM HI2
Once the OVER RANGE warning appears the user must advance the transmitter to the next higher range via the menu and
keypad Press MENU, select MANUAL 5AMPLING, press ENTER, select the appropriate MANUAL RANGE and press ENTER again.
Alarms
The CONFIG ALARM5 features a system that can be configured in the field. Two field adjustable alarm relays with dry contacts
operate independently of one another which means the alarms can be set-up as:
,. HI and LO
r LO and LO, LO,
,. HI and HI,HI
,. PERCENT (of full scale range which changes with auto-ranging)
.,. % (oxygen value)
Additional feature ineludes delaying the activation of the local audible alarm and relay contacts for up 99 minutes to enable
users to distinguish between transient occurrences and true upset conditions which is particularly useful on remote applications
16
Advanced InstrulDents loe.
without affecting the 4-20mA signal output. The local audible alarm can be silenced or disabled as well without affeeting the 4
20mA signal output.
Note: A separate feature, BVPASS ALARMS described below, enables the user to disable the local audible alarm and relay
contacts during calibration or servicing. The alarms are enabled when the alarm condition is correeted.
5et Alilrm Values:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONFIG ALARMS.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:
MAIN MENU
AUTO SA'MPLE
MANUAL SAMPLE
CALIBRA110N
[gn:'; (#t-!'';I:k
BYPASS ALARMS
>
MAIN MENU
.';j:l'
SET A!,ARM 2
SET ALARM DELAY
ALARM 1 HIjLO
ALARM 2 HIjlO
ALARM5 AUDIBLE SILENT
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM 1 option.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appears with PERCENT as the default alarm value :
MAIN MENU

SET ALARM 2
SET ALARM. DELAY
ALARM 1 HI/LO
ALARM 2 HI/LO
ALARM5 AUDIBLE/SILENT
>
GAS CONCENTRAll0N
103;1;4:"
PPM
8. Advance the reverse shade cursor using the ARROW keys to highlight the desired option.
9. Press the ENTER key to select the highlighted menu option.
Note: The PERCENT (of FS) alarm value is entered with one decimal, the PPM (oxygen) afarm value is entered as an integer.
01.0
GAS CONCENTRA110N
5Eili ALARM 1 VALUE
>
PRESS UP OR DOWN'
TO CHA'NGE VALUE
ENTER TO SAVE
MENU TO RETURN
GAS CONCENTRATION I
001,
-----.
5ET ALARM i VALUE
1 __ .. _ ..1

--------!"ERCENT ,
17
Advsnced Instruments Ine.
m
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
10. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
11. Press the ARROW keys to enter the alarm value.
12. Repeat until the complete span value has been entered.
13. Note: If an alarm is set as a PERCENT value and subsequently changed te a PPM value, the PERCENT value is not retained
and is reset te 00.0. This holds if the alarm was first set as PPM value and then changed to a PERCENT value.
14. Save the alarm value by pressng the ENTER key or abort by pressing the MENU key.
15. The system returns to the SAMPLlNG mode and displays:
3.3 PPM
AUTO SAMPUNG
10 PPM RANGE
-:
,-------'-=-
Repeat the steps aboye to set the ALARM 2 value:
MAIN MENU MAIN MENU
>
SETALARM 1
MANUAL SAM PLE
AUTO SAMPLE
VMe
l
: fj
CALIBRATlON SET ALARM DELAY
ALARM 1 HI/L.O
BYPASS ALARMS ALARM 2 HI/LO
ALARMS AUDIBLE/SILENT
Set Alarm Defay:
Once the values for ALARM 1 and ALARM 2 have been entered, the user may elect to delay the activation of the local alarms
and relay contacts for up te 99 minutes. This feature allows users to distinguish between transient occurrences and true upset
conditions. This feature can be particularly useful on remote applications without affecting the 4-20mA signal output.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONAG ALARM5.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
MAIN MENU MAIN MENU
>
1
MANUAL SAMPLEi
AUTO SAMIPlE
SET ALARM 2
CALIBRATlON
'1"'" jo' ;1:'.]3 "'i
ALARM lIHI/lO
BVPASS ALARMS
SMijltcf,'!;I:b'"
AlARM 2 IIU/lO
ALARMS_AUDIBLE/SILENIT
18
Adllanced In!!!itruments Inc.
5. Advance the reverse shade cursor using the ARROW keys to highlight the SET ALARM DELAY.
6. Press the ENTER key to select the highlighted menu optian.
7. The following displays appear with last alarm delay value :
99
,-----,
I
MAlNMENU
DE.LAY IN MINUTES
>
SETALARM 1
PRESS UP OR DOWN
TO CHANGE VALUE
2
43 .. '.' '.';1:.'# 'J\7
,ENTER 1'0 SAVE
ALARM 2 Hl/lO
ALARM 1 HI/lO
MENU TO REnlRN
AlARMS AUDIBLE/SIlENT
8. Press the ENTER key to advance the underline cursor right or press the MENLJ key to advance the underline cursor left to
reach to the desired digit of the alarm value.
9. Press the ARROW keys to enter the alarm value.
10. Repeat until the complete span value has been entered.
11. save the alarm value by pressing the ENTER key or abort by pressing the MENU key.
12. The system returns the SAMPUNG mode and displays:
3.3 PPM
AUTO SAMPUNG
10 PPM RANGE
24.5 e 100 KPA
lO1! 2PPM lOPPM HI2
Set, HI/LO Alanns:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONFIG ALARMS.
3. Press the ENTER key to select the highlighted menu option.
4. The following dsplays appears:
MAIN MENU
AUTOSAMPlE
MANUAL SAMPlE
CAUBRATION
[iHli(?t"'J:k
BYPASS ALARMS
>
MAlN
SET ALARM 1
SET ALARM 2
SET AlARM DELAY
fJ"';J:'Urm
ALARM 2 HI/lO
ALARMS AUDIBLE SILENT
5. Advance the reverse shade cursor using the ARROW keys to highlight the ALARM 1 option, which appears as either ALARM
1 HI or ALARM 1 LO.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPUNG mode.
19
Advanced InstrulDents Inc.
3.3 PPM
AUTO SAMPUNG
10 PPM RANGI;
24.5 e 100 KPA
L01 2PPM 10PPM HI2
7. Repeat steps 1 through 6 for the ALARM 2 HI/LO setting.
Set Local Alarms:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CONFIG ALARMS.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:
MAIN MENU
AUTO SAMPLI;
MANUAL SAMPLE
CALIBRATlON
t;U'pcCf"j.;i:k
a'YPASS ALARMS
>
MAIN MENU
SET ALARM 1
SETALARM 2
sET ALARM'DELAY
ALARM 1 HI/LO
AlARM 2 HI/LO
, .,
S. Advance the reverse shade cursor using the ARROW keys to highlight the ALARMS AUDIBLEjSILENT option, which appear
as either ALARMS AUDIBLE or ALARMS SILENT.
6. Press the ENTER key to toggle and change the displayed setting. After 3 seconds, the system returns to SAMPUNG mode.
3.3 PPM
AUTO SAMPLING
10 IPPM RANGE
24.SC 100 KPA
L01 2PPM 10PPM HU
Bypass Alarms:
This feature, separate from CONFIG ALARMS aboye, enables the user to disable the local audible alarm and relay contacts
during calibration or servicing. The alarms are enabled when the alarm condition is corrected.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight BYPASS ALARMS.
3. The following displays appears:
20
Adllanced Instruments loc.
IfAIN MENU
AUTO SAMPLE
I'4ANUAL SAMPLE
CAUBRATlON
CONFlG ALARMS
j ~ S j g.$.$j., l' ;1:&"
4. Press the ENTER key to bypass and disable both the local audible alarm and relay contacts. After 3 seconds, the system
returns to SAMPUNG mode.
S. Note: The approprate alarm settng wll alternately reverse shades ndcatng the alarm condton exsts but the BYPASS
ALARM5 feature has disabled the local audible alarm and relay contacto The alarms are enabled when the alarm conditon is
corrected.
3.3 PPM
AUTO SAMPUNG
10 PPM IRANGE
24.5 e 100 KPA
LOl 2PPM J"
Stan-up is complete ... Dmceed to Calibration
21
Advanced Instruments IDC",
lero, Calibratioo
In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. However, the transmitter
will generate an oxygen reading when sampling oxygen free sample gas due to:
~ Contamination or quality of the zero gas
,. Minor leakage in the sample line connections
,. Residual oxygen dissolved in the sensor's electrolyte
~ Tolerances of the electronic components
Recommendation: lero calibration is recommended only for online analyzers ntended for measurements below 1 ppm on the
10 ppm range and then only when the sample system connections are modified and when installing a new oxygen sensor. lt is
not practical on higher ranges or portable analyzers for the following reasons:
1. Determining the true zero offset reguires the user allow the analyzer approximately 24 hours to stabilize with flowing zero
gas as evidenced by a stable reading with no downward trend on an external recordng device. Note: Approximately 24-36
hours is required to assure the sensor has consumed the oxygen that has dissolved into the electrolyte inside the sensor
while exposed to air or percentage levels of oxygen. For optimum aCcuracy, utilize as much of the actual sample system as
possible.
2. Thus it is not practical to find the true zero offset particularly in the case of applications requiring higher level oxygen
measurements because of the low offset value, normally 50% of the most sensitive range, is not material to the accuracy
of higher level measurements. Nor is it practical to zero a portable analyzer every time it is moved from one sample point
to another.
3. caution: Prematurely zeroing the analyzer can cause a negative reading in both the ZERa and SAMPLE modes.
4. Satisfying users that the zera offset is reasonably acceptable for their application can be accomplished much quicker.
Unless the zero gas is contaminated or there is a significant leak in the sample connections, the analyzer should read less
than 100 ppm oxygen within 10 minutes after being placed on zero gas thereby indicating it is operating normally.
Recommendation: Initiate the DEFAULT ZERa pracedure before performing either alERO or SPAN CAUBRATION.
Procedure:
lero calibration should precede the span calibration and once performed should not have to be repeated with subsequent span
calibrations. Normally, zero calibrations are performed when a new sensor is installed or changes are made in the sample
system connectons.
Refer to Span Calibration below for the detaled procedure. Differences inelude the displays iIIustrated below, substituting a
suitable zero gas for the span gas and the time required to determine the true zera offset of specific oxygen sensor, analyzer
and sample system combination.
The maximum zero calibration adjustment permitted is 50% of the lowest full scale analysis range available. Accordingly, the
analyzers lERa has not been adjusted prior to shipment because the factory conditions are different from the application
condition at the users installation.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CAUBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
MAIN MENO
AUTO SAMJ!liE
MANUAL SAMPLE
:',' l ~ , ;1,. (.1 :-1
CONFIG ALARMS
BYPASS ALARMS
>
CALIBRAnON
SPAN CALIBRATE
'43;11'"'' I':J ;'." ~ .
DEFAULT SPAN
DEfAULT ZERO
OUTPlit SPAN
OUTPUTZERO
5. Advance the reverse shade cursor using the ARROW keys to highlght lERO CAUBRATE.
22
Adllanced Instrurnents IDe.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear:
0.000 PPM
ZERO
CALIBRTlON
~ N T E R TO CAUBRATE
MENU TO A B ~ T
8. Press the ENTER key to calibrate or MENU key to abort and return to SAMPUNG mode.
9. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 60% of the full scale low range.
10. Both the lero Calibrate and Span calibrate functions result in the following displays:
OR
PASSED fAILED
CAUBRATlON CAUBRATlON
The maximum zero calibration adjustment permitted is 50% of the lowest full scale range available. Accordingly, the
transmitter's lERa has not been adjusted prior to shipment because the factory conditions are different from the application
condition at the user's installation.
Default Zero
The software will eliminate any previous zero calibration adjustment and display the actual the signal output of the sensor at a
specified oxygen concentration. For example, assuming a zera gas is introduced, the display will reflect an oxygen reading
representing basically the zero calibration adjustment as described aboye. This feature allows the user to test the sensor's signal
output without removing it from the sensor housing.
Recommendation: Initiate the DEFAULT ZERO procedure before performing either a ZERO or SPAN CAUBRATION.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CAUBRATION.
3. Press the ENlCR key to select the highlighted menu option.
4. The following displays appear:
.MAIN MENU CALIBBATlON
>
AlWTO SAM PLE SPAN CALIBRATE'
MANUAL SAMPLE ZERO CALIBRATE
(t.,'I:! .1' ,r.pl DEFAULT SPAN
CONFIG ALARMS .,,.\IJlf34;t.:
BYPASS ALARMS OUTPUTSPAN
OUTPUT ZERO
5. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT lERa.
6. Press the ENTER key to select the highlighted menu optiofl.
23
Advanced Instruments Ine.
7. The following display appears and after 3 seconds the system returns to the SAMPLlNG mooe:
FAetORY 3.3 PPM
DEFAULTS
AUTO SAMPI,.,lNG
SET
10 ppM RANGE
24.5 e 100 KPA
LOl 2PPM 10PPMI H.IZ
OutputZero
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LeD. This feature enables the user to
adjust the 4mA signal output when the LCD displays 00.00. Note: Adjust the 20mA signal output with the OUTPUT SPAN.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CALlBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appear:
MAlN MENU CALIBRAnON
>
AUTO SAMPlE SPAN CAUBRATE
MANUAL SAMPLE ZERO CALIBRATiE
DEFAUlT SPAN
CONF1G ALARMS
tt'. r:\;IJ I len!
DEFAULT ZERO
BYPASS AURMS OUTPUTSPAN
tui 4I*3lt)
S. Advance the reverse shade cursor usng the ARROW keys to highlight DEFAULT ZERD.
6. Press the ENTER key to select the highlighted menu option.
7. The followng display appears:
OUTPUT ZERO OfFSET
PRESS UP OR DOWN
TO CHANGE VAL,UE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as descrbed in Appendix B or consult the factory. The true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
020.0
OUTPUT ZERO OFF5E-m-
PRESS UP OR DOWN
TO C H A ~ G E VALUE
IENTEcR TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT ZERO OFFSET value.
10. Press the ARROW keys to enter the OUTPUT ZERO OFFSET value.
11. Repeat until the complete OUTPUT ZERO OFFSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. The system returns to the SAMPUNG mode.
24
Adllanced Instruments loe.
Spao Calibradoo
Maximum drift from calibration temperature is approximately 0.11% of reading per oc. The analyzer has been calibrated at the
factory. However, in order to obtain reliable data, the analyzer must be calibrated at the initial start-up and periodically
thereafter. The maximum calibration interval recommended is approximately 3 months, or as determined by the user's
application.
Calibration involves adjusting the analyzer electronics to the sensor's signal output at a given oxygen standard, e.g. a certified
span gas with an oxygen content (balance nitrogen) approximating 80% of the next higher full scale range aboye the intended
measuring range is recommended for optimum accuracy, see Calibration and Accuracy.
Recommendation - based on the inherent linearity of the galvanic oxygen sensor air calibrate the analyzer as described
below when installing and replacing the oxygen sensor (exception UHP and M5 versions of the PiCO Ion Sensor); or, to verify the
oxygen content of a certified span gas; or, when a certified span gas is not available (immediately following air calibration
reintroduce a gas with a low oxygen concentration to expedite the return to ppm level measurements).
Recommendation: lnitiate the DEFAULT SPAN procedure before performing either a ZERO or SPAN CAUBRATlON.
Caution: Prematurely initiating the SPAN CAUBRATlON key before the galvanic fuel cell sensor based analyzer reading has
stabilized can result in erroneous readings. For example, to assure an accurate air calibration when installing a new ppm
oxygen sensor from its packaged oxygen free atmosphere allow the oxygen sensor 2-3 minutes to reach equilibrium with the
oxygen content of the ambient ar surrounding it before pressing the SPAN CAUBRATE key. A ppm Oxygen sensor that has not
been allowed te reach equilibrium will generate a lower current output than a ppm oxygen sensor that has reached equilibrium.
Pressing the SPAN CAUBRATE key before the ppm oxygen sensor has reached equilibrium forces the micro-processor to
prematurely read the (erroneous low) current output of the ppm oxygen sensor and introduce larger (erroneous) than required
electronic gain adjustment and display (also erroneous) CAUBRATlON SUCCESSFUL message to the user. The error wll become
evident when a zero gas with a low oxygen concentrati on is introduced into the ppm analyzer to purge it down below 10 ppm.
The analyzer reading may stop and appear to stabilize as high as 1800 ppm - giving the user the (erroneous) impression there
is a problem with the ppm oxygen sensor when in fact the problem lies with the user's failure to follow the recommended
calibration procedure.
Required Components: Refer to lnstallng Span Gas secton aboye.
,. Certified span gas cylinder with an oxygen concentration, balance nitrogen, approximating 80% of the full scale range
aboye the ntended measuring range.
,. Regulator to reduce pressure te between 5 and 30 psig.
,. Flow meter to set the f10w between 1-5 SCFH,
,. Suitable fittings and a 46 ft. in length of 1/8" dia. metal tubing to connect the regulator to the f10w meter inlet
... Suitable fitting and a 4-6 ft. in length of 1/8" dia. metal tubing to connect to the f10w meter vent
... 1/8" male NPT to tube adapter fitting to connect the 1/8" dia. metal tubing from the flow meter vent to the mating male
quick disconnect fitting supplied with the GPR-1500AIS.
Proc:edure:
The user must ascertain that the oxygen reading (actually the sensor's signal output) has reached a stable value within the
limits entered below before entering the span adjustment. Failure to do so will result in an error.
1. This procedure assumes a span gas under positive pressure and is recommended for a transmitter without an optionaf
sampling pump, which if installed downstream of the sensor should be placed in the OFF position and disconnected so the
vent is not restricted during calibration.
2. To assure an accurate calibration, the temperature and pressure of the span gas must c10sely approximate the sample
conditions.
3. For calibration purposes, use of the AUTO SAMPLE mode is recommended. However, the user can select the full scale
MANUAL SAMPLE RANGE for calibration as dictated by the accuracy of the analysis required - for example, a span gas with
an 80 ppm oxygen concentration with the balance nitrogen would dictate the use of the 0-100 ppm full scale range for
calibration and a 0-10 ppm measuring range. Select as described aboye.
4. Access the MAlN MENU by pressing the MENU key.
S. Advance the reverse shade cursor using the ARROW keys to highlight AUTO SAMPLE.
6. Press the ENTER key to select the highlighted menu option.
7. The following displays appear:
25
Advance'd InstrumeDts Ine.
MAIN MENU
3.3 PPM f"iii.J,.,:IQ.
MANUAL SAMPLE
AUTO SAMPUNG
CAUBRATlON
10 PPM RANGE,
CONFIG AlARMS
24.5 e 100 KPA
IJVPASS ALARMS
LOl lPPM l'OPPM HU.
8. Return to the MAIN MENU by pressing the MENU key.
9. Advance the reverse shade cursor using the ARROW keys to highfight CAUBRATION.
10. Press the ENTER key to select the highlighted menu option.
11. Repeat to select DEFAULT SPAN.
12. Repeat to select SPAN CALIBRATE.
13. The following displays appear:
MAIN MENU CALIBRATlON
>
AUTO SA'MPLE
PI' I ;',5 "
MANUAL SAMPLE ZERO CAUBRATE
DEFAUlT SPAN
CONFIG ALARMS
rmr=hy.c.r:i
DEFAULT ZERO
BVPASS OUTPlJT SPAN
OUTPUTZERO
14. Assure there are no restrictions in vent line.
15. Regulate the pressure and control the flow rate as described aboye at 5-30 psig and a 2 SCFH flow rateo
16. Allow the span gas to flow for 1-2 minutes to purge the air trapped in the span gas line.
17. Disconnect the sample gas line and install the purged span gas line.
18. Caution: Allow the span gas to f10w and wait until the reading is stable before proceeding wjth calibration.
The wait time will vary depending on the amount oxygen introduced to the sensor when the gas lines were switched.
19. Press the ENTER key to select the SPAN CAUBRATE option.
20. Note: A span gas concentration aboye 1000 ppm dietates the selection of the PERCENT option.
21. Advance the reverse shade cursor using the ARROW keys to highlight the desired GAS CONCENTRATION.
22. Press the ENTER key to select the highlighted menu option.
GAS CONCENTRATlON
PERCENT
m
23. The following displays appear:
,.---------- ...,
PRESS UP OR DOWN
TO CHANGE VALUE
TO SAVE
MIENU TO RETURN
>
"..---------- ..
I
SPAN
CALIBRATlON
ENTER TO CALIBRATE
MENU TO ABORT
24. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the alarm value.
25. Press the ARROW keys to enter the alarm value.
26. Repeat until the complete span value has been entered.
26
Adllaoced IOf'trumeots Ine.
27. Save the adjusbnent value by pressing the ENTER key or abort by pressing the MENU key.
28. Allow approximately 60 seconds for the calibration process while the processor determines whether the signal output or
reading has stabilized within 60% of the fuJl scale low range. Both the Zero Calibrate and Span Calibrate functions result in
the following displays:
OR
J>ASSED FAIlED
CALIBRAmo.N CAUBRA1Ii10N
29. If the calibraton is successful, the transmitter returns to the SAMPUNG mode after 30 seconds.
30. If the calibration is unsuccessful, return to the SAMPUNG mode with span gas flowing through the transmitter, make sure
the reading stabilizes and repeat the calibration before concluding the equipment is defective.
31. Before disconnecting the span gas line and connecting the sample gas line, restart if necessary the f10w of sample gas and
allow it to flow for 1-2 minutes to purge the air inside the line. ,
32. Disconnect the span gas line and replace it with the purged sample gas line.
33. Wait 10-15 minutes to ensure the reading is stable and proceed to sampling.
Default Span
The software will set the SPAN adjustment based on the average oxygen reading (actually the sensor's signal output) at any
specific oxygen concentration. For example, when a span gas is introduced, the micro-processor will display an oxygen reading
within SO% of the span gas value. This feature allows the user to test the sensor's signal output without removing it from the
sensor housing.
Recommendation: Initiate the DEFAULT SPAN procedure before performing either a ZERO or SPAN CAUBRATION.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CAUBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. The following displays appears:
MAIN MENU CALIBRATlON
>
AU;rO SAMPlE SPAN CALIBRATE
MAtoIUAl SAMPLE ZERO CAUBRATE
l.] ,jii
Co.NfIGo ALARM5 DEFAULT ZERO
BYPASS ALARMS OUTPUTSPAN
OUTPUTZERO
S. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
6. Press the ENTER key to select the highlighted menu oplion.
7. The following displays appear and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAUl,TS
SE'!"
3.1 PPM
AUTO SAMPUNG
10 PPM RANGE
24.5 C 100lKPA
LOl 2PPM 10PPM Hll.2
27
AdllaneedInstrulDents ,oc.
Output Span:
In rare instances the 4-20mA signal output may not agree to the reading displayed by the LCD. This feature enables the user to
adjust the 20mA signal output should the LCD display not agree. Note: Adjust the 4mA signal output with the OUTPUT ZERO
option described aboye.
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlight CAUBRATION.
3. Press the ENTER key to select the highlighted menu option.
4. lhe following displays appear:
M'AIN MENU CAUBRATION
>
AUTO SAMPLE SPAN CALlBRAn:
MANUAL SAMPLE 2ERO CALIBRATE
:...,i' :1>'." tl):1 DEFAULT SPAN
CONFlG ALARMS DEFAULT ZERO
BYPASS ALARMS [.111 I a'", i.,:
OUTPUTZERO
S. Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
6. Press the ENTER key to select the highlighted menu option.
7. lhe following display appears
OUTPUT SPAN OFFSET
PRESS UP OR DOWN
TO CH'ANGE VAlUE
ENTER TO SAVE
MENU TO RETURN
8. Compute the adjustment value as described in Appendix B or consult the factory. lhe true adjustment value must be
determined empirically by trial and error. Adjust the initial adjustment value for additional percent errors.
0 9 ~ . O
OVTPVT SPAN OFFSET
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
9. Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor left to
reach to the desired digit of the OUTPUT SPAN OFFSET value.
10. Press the ARROW keys to enter the OUTPUT SPAN OFFSET value.
11. Repeat steps 9 and la until the complete OUTPUT SPAN OffSET value has been entered.
12. Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
13. lhe system returns to the SAMPUNG mode.
28
Adllanced Instruments n c ~
S8mpling
GPR-1500AIS ppm Oxygen Transmitter requires positive pressure to f10w the sample gas by the sensor to measure the oxygen
cancentration in a sample gas. If not available see Pressure & Flow section.
Note: Prematurely initiating the ZERO CAUBRATION procedure can cause the analyzer to display a negative reading in both
the ZERO and SAMPLE modes. Prematurely initiating the SPAN CAUBRATION procedure can cause erroneously high offsets and
inaccurate readings.
Procedure:
1. Following calibration the transmitter returns to the SAMPLE mode after 30 seconds.
2. Select the desired sampling mode - auto or if manual, the range that provides maximum resolution - as described aboye.
3. Use metal tubing to transport the sample gas to the transmitter.
4. lhe main consideration is to eliminate air leaks which can affect oxygen measurements aboye or below the 20.9% oxygen
concentration in ambent air - ensure the sample gas tubing connections fit tightly into the 1/8" male NPT to tube adapter,
and, the NPT end is taped and securely tightened into the mating male quick disconnect fittings which mate with the
female fittings on the transmitter
5. Assure there are no restrictions in the sample line.
6. For sample gases under positive pressure the user must provide a means of controlling the inlet pressure between 5-30
psig and the f10w of the sample gas between 1-5 SCFH, a f10w rate of 2 SCHF is recommended
7. Far sample gases under atmospheric ar slightly negative pressure an optional sampling pump is recommended to draw the
sample into the transmitter. Generally, no pressure regulation or f10w control device is nvolved.
8. Caution: If the transmitter is equipped with an optional sampling pump and is intended for use in both positive and
atmospheric/slightly negative pressure applications where a f10w meter valve is involved - ensure the valve is completely
apen when operating the sampling pump. Refer to the Pressure & Flow section aboye.
9. Assure the sample is adequately vented for optimum response and recovery - and safety.
10. Allow the oxygen reading to stabilize for approximately 10 minutes at each sample point.
To avoid erroneous oxygen readings and damaging the sensor:
? Do not place your finger over the vent (it pressurizes the sensor) to test the f10w indicator when gas is f10wing to the
sensor. Removing your finger (the restrietion) generates a vacuum on the sensor and may damage the sensor (voiding the
sensor warranty).
? Assure there are no restrictions in the sample or vent lines
Avoid drawing a vacuum that exceeds 14" of water column pressure - unless done gradually
,. Avoid excessive f10w rates aboye 5 SCFH which generate backpressure on the sensor.
,. Avoid sudden releases of backpressure that can severely damage the sensor.
,. Avoid the collection of particulates, Iiquids or condensabon callect on the sensor that could block the diffusion of oxygen
into the sensor.
:;. If the transmitter is equipped with an optional integral sampling pump (positioned downstream of the sensor) and a flow
control metering valve (positioned upstream of the sensor), completely open the flow control metering valve to avoid
drawing a vacuum on the sensor and placing an undue burden on the pump.
Standbv
The transmitter has no special storage requirements.
lhe sensor should remain conneeted during storage periods.
Store the transmitter with the power OFF.
If storing for an extended period of time protect the analyzer from dust, heat and moisture.
29
Advanced Instrument Ine.
16 Maintenance
Generally, c1eaning the electrical contacts or replacing filter elements is the extent of the maintenance requirements of this
transmitter.
Sensor Beplacement
Periodically, the oxygen sensor will require replacement. The operating lite is determined by a number of factors that are
influenced by the user and therefore difficult to predict. The Features & Specifications define the normal operating conditions
and expected life of the standard sensor utilized by the GPR-1500AIS transmitter. Expected sensor life is inversely proportional
to changes in oxygen concentration, pressure and temperature.
Serviceability: Except for replacing the oxygen sensor, there are no parts inside the transmitter for the operator to service.
Only trained personnet with the authorization of their supervisor should conduct maintenance.
Cautlon: DO NOT apen the oxygen sensor. The sensor contains a corrosive liquid electrolyte that could be harmful if touched
or ingested, refer to the Material Safety Data Sheet contained in the Owner's Manual. Avoid contact with any liquid or crystal
type powder in or around the sensor or sensor housing, as either could be a form of electrolyte. Leaking sensors should be
disposed of in accordance with local regulations.
Procedure:
1. Remove the four (4) screws securing the transmitter's front panel.
2. Caution: Do not discard the gaskets from the enclosure.
3. Using the 5/16 wrench supplied loosen but do not remove the clamp bolt located in the
center of the housing with the elbows attached.
4. Rotate the upper section of the sensor housing 90to disengage from the clamp.
5. Remove the upper section by pulling it straight up and place it on a smooth surface.
6. Remove the old axygen sensor and dispose of it as you would a battery.
7. Remove the new oxygen sensor from the shipping bago
8. Remove the red {abel and the gold ribbon (shorting device) from the PeB at the rear of the
sensor.
9. Caution: Minimize the time the sensor is exposed to ambient airo
10. Place the new sensor in the bottom section of the sensor housing with the PeB facing up.
11. Place the upper section of the sensor housing over the sensor.
12. Span calibrate the transmitter in 20.9% ambient air
t
once the reading stabilzes - see
aboye.
13. Gently push the upper section downward and rotate 90to engage the clamp.
14. Finger tighten the clamp bolt and one full turn with the 5/16 wrench to compressed the o-ring sea!.
15. Connect zero gas or low oxygen content sample gas line to purge the sensor of oxygen.
30
Adllaneed Intruments Ine.
1SparePans
Recommended spare parts for the GPR-1500AIS ppm Oxygen Transmitter:
Remll.
GPR-12-333
XLT-12-333
Other spare parts:
ltem No.
HTR-10D4
A-1004-2-14
A-1016-A
S-2762-A-2-14
MTR-1011
MTR-1D14
ORNG-1DD7
A-1l51-AIS-1
A-1l66-AIS-Ae
A-1l66-AIS-DC
Descriltien
ppm Oxygen Sensor
ppm Oxygen Sensor
Oescription
Heater Low Power Sensor Housing
Housing Sensor Stainless Steel
Housing Sensor Bottom Assembly Stainless Steel
Housing Sensor Upper Assembly Stanless Steel
Meter Digital Panel LCD Sacklight
Meter Digital Panel LCO Low Temperature
O-ring 3/32 x 1-3/8 x 1-9/16 Viton
PCS Assembly Majn / Display
pes Assembly Interconnection / Ae Power Supply
PeS Assembly Interconnection / DC Power Supp[y
8 Troubl'eshOOlinu
Symptom Possible Cause Recommended Action
Slow recovery or At installation, defective sensor
Air leak in sample system connection(s)
Abnormality in zero gas
Damaged in service - prolonged exposure to
ar, electrolyte leak
Sensor nearing end of life
Replace sensor if recovery unacceptable or O
2
reading fails to reach 10% of lowest range
Leak test the entire sample system: Vary the f10w
rate, if the O
2
reading changes inversely with the
change in f10w rate indicates an air leak - correct
source of leak
Qualify zero gas (using portabfe transmitter)
Replace sensor
Replace sensor
Hgh O
2
reading Transmitter calibrated before sensor stablized Allow O
2
reading to stablize before making the
after installing caused by: span/calbration adjustment
or replacing sensor 1) Prolonged exposure to ambient air, worse Continue purge with zero gas
if sensor was unshorted
2) Air leak in sample system connection(s) Leak test the entire sample system (above)
3) Abnormality in zero gas Qualify zero gas (using portable transmitter)
High O
2
reading Flow rate exceeds limits Correct pressure and flow rate
Sampling Pressurized sensor Remove restrction on vent line
Improper sensor selection Replace GPRjPSR sensor with XLT sensor when
CO
2
or acid gases are present
31
Ad'llanced 'n!!iitrument!!ii IDe.
Symptom Possible Cause Recommended Action
Response time slow Air leak, dead legs, distance of sample line,
low f10w rate, volume of optional filters and
scrubbers
Leak test (above), reduce dead volume or ncrease
f10w rate
O
2
reading doesn't Pressure and temperature of the sample is Calibrate the transmitter (calibrate at pressure and
agree to expected O
2
different than span gas temperature of sample)
values Abnormality in gas Qualify the gas (use a portable transmitter)
Erratic O
2
reading
or
No O
2
reading
Change in sample pressure
Dirty electrical contacts in upper section of
sensor housing .
Corroded solder joints on sensor PeS from
corrosive sample or electrolyte leakage from
sensor
Corroded spring loaded contact in upper
section of sensor housing from liquid in
sample or electrolyte leakage from sensor
Liquid covering sensing area
Improper sensor selection
Presence of interference gases
Unauthorized maintenance
Sensor nearing end of Iife
Sensors without PCB use mV setting.
Calibrate the transmitter (calibrate at pressure and
temperature of sample)
Clean contacts with alcohol (minimize exposure time
of MS sensor to ambient air to extent possible)
Replace sensor and return sensor to the factory for
warranty determination
Upper section of sensor housing: Clean contacts
with alcohol, f10w sample or zero gas for 2-3 hours
to f1ush sample system and sensor housing
Sensor: Replace if leaking and retum it to the
factory for warranty determination
Wipe with alcohol and Iint free towel or f10w sample
or zero gas for 2-3 hours to f1ush
Replace GPR/PSR sensor with XLT sensor when CO
2
or acid gases are presento Consult factory.
Replace sensor and install scrubber
Consult factory.
Replace sensor
Erratic O
2
reading
or
Negative O
2
reading
or
No O
2
reading
accompanied by
electrolyte leakage
Pressurizing the sensor by flowing gas to the
sensor with the vent restricted or SHUT OFF
valve c10sed and suddenly removing the
restriction draws a vacuum on the sensor
or
partially opening the valves upstream of the
transmitter when using a pump downstream
of the transmitter to draw sample from a
process at atmospheric pressure or a slight
vacuum. Placing a vacuum on the sensor in
excess 4" cf water column is strongly
discouraged.
A premature adjustment of the ZERO OFFSET
potentiometer is a common problem
Zero the transmitter. If not successful replace the
sensor
Avoid drawing a vacuum on the sensor, a
pressurized sensor may not leak but still produce
negative readings.
From MAIN MENU select DEFAULT ZERO
32
Advaneed Instruments Ine.
Purpose: SPAN CALlBRAll0N of digital (reference to analog) ppm O
2
analyzer already in-servce.
TROU BLESHOOT ppm O
2
analyzer te cenfirm response and stability under controlled conditions.
PreUminary Test
1.) COllfrm metal sarnple system components - no pl(lStl:.
L
2.) leak. Test Varo analyzer fIow rate upJdown and observe reading.
I I
1,) TQ,rerl anj c:>nne..'tlon;;
2,) STO? ,i rcad,...;) IS mt ,tatle and
fl.'lQry,
No
<$?
Ves
Set-up 1.) Piare analyzer SAt"l'lEIBYPASS vdlve(s) in eYPASS mode, see (a) below.
2.) Seloct AUT0-RANGING mode or MANUAlly selecl rc.l\le for Spll gas
3.) Perform DEFAUl T ZERO or MANUAlly eliminate previous zelO adjustmellts
4.) Perform DEFAULT SPAN
Note: Zeroing lhe analyzer is OIlly recornmended for continuous analysis < 1 ppm
whidl normally e:<cludes portable Ildlyzers - cOIlstanUy changng gas Unes.
I
,
1.) Comed the span gas lioe or If equipped with a 3-way SAMPlE/SPAN valve
place it in Ihe SPAN lnode,
2.) AJlow Ihespan gas te liow fOl' 5-10 mirutes to purge Ihe air (20.9%
fi'om nside Ihe span gas Une - during connection or leClks during standby.
11 analyzer is not equlpped \\Iilh eVPASS SAMPlE SYSTErv1, see (d) below:
1.) Purge lhe Spll gas line before connecti19 te !he analyzer.
2.) Conn et the gas lineo; eJ5 quickly as Dossible - sorre air \ViII be introduc.ed
3.) AlIow edra time fer !he reading to stobilize - !he sensor 11Im e:<posed te alr.
4.) When off-line, Illdintain gas Aow thru analyzer or cap connections (inlet filsl)
I
to avoid damaging the sensor by eJ<posng it to ait' (209, 0;.
I
1,) Place anCllyzer SAMPLEIBYPASS valve(s) in SAMPLE mode, see (a) 001011I.
2,) Allow Ihe reading to stabilize, normally 2-10 minutes e:<posed lo air.
3,) Once stable - initiate \llBRATlON routne (or unlock and adjust SPAN knob)
fmm MAIN t-1Er-oAJ, select SPAN CALIBRAlE, enter SPAN 'tALUE, press ENTIR.
1.)ConT 99S ..... ,'
analyzer o::alnrteo wlm alTblent ar,

?) Rep",..t at \;!,3$t t"":"'.
3.) :en.O<" ",d rep; 3t o:-.:e.
No
4,) 1f LJIlsu=fu! wlt'l senS()',
Ves
STa:> znd o)ntact fa::tor.
Note: SPAN CALIBRATION ls complete and for TROUBLESHOOTING purposes
demollstrates!he sensorjanaly2el' responds norrnally - !.ill.Qer contr..QJ.l.!:.d. . m!ltions.
For TROUBLESHOOT1NG purposes only: Connect one of Ule dnalyzer;s sigrldl oulputs lo
llIl external rec.ordilg devlc:e and conlnue Ihe fIow of span gas for 1-2 holl's.
1.) ",">d rep", once.

L.) """m rr:w sensor,
No STCP 31">d rontact foctory
Ves
1.) Place analyzer SAMPLE/BVPASS valve(s) In BVP\SS mode, see (a) 001011I,
2.) Conoect Gas lile - as aboye for sanlP1e gas.
3,) Place analyzer SAMPLE!BVPASS valve(s) in SAMPlE mode, see (a) below.
T
SHJT OFf Vf,LVE - o;;en fi'st, cbse la,!
Conned GdS Lile
CaUbratiOl1 Procedure
Troubleshooting
Saflllling
: Rev 7f06 Pa'.;e 10' L
Adllanced In,strum,ents Inc.
Purpose: AlR CAUBRATlON of digital (reference to analog) ppm 0
1
analyzer (only use for 0-25% range):
ao) when nstalling a ne...... sensor,
b.) when span gas (nstTument air piped to sensor - address as SPAN CALlBRAnO ') is not available or
c.) when it is advanrageous from a troubleshooting standpoint to employ a portabie analyzer tIlat has been calibrated with
ambient air as a "referee" te conflrm other analyzers or span gas ... alues.
Note 1: The drawback te alr calibrabon is tne time required lor a ppm sensor exposed te air (1-2 minutes for calibration purposes) to
recover te me O-la ppm range and the added requirement for low ppm 0
1
concentration gas (sample. zero, purge or span) to
purge me sensor 01 me oxygen that dissolves nto the sensor's eleetro!yte when exposed te air (20.9% or 209,000 ppm o, )
Note 2: Expected recovery time ro 10 ppm on 1-2 ppm purge gas is < 1 hr at installatiOl1 aOO < 20 minutes for an analyzer that has been
in-se!Vice > 2 weeks. For higher pl?m analvsis, expected recovery time to SO-lOO ppm on 1-2 ppm purge gas is < 10 minutes.
Prdminary
1,) Conf.rm ""mple .ystem eomponents - no plastic_
Set-up
1.) 5elect AUTO-RANGING mode 01 MAlIUAlLY y.lect 0-25% range for alr atco
2.) Perform DEFAULT lERO or MANUAlLY prevlous lero adjustments
flote: leroin; the anal\,%er Is only re:ommended fOI continuous analySls < 1 ppm whleh normally excludes portlltlie
analyzers - due to !he constant changing 01 gas lines.
3.) Perfolm DEFAULT SPAN
4,) InstaU only: Place analyzer SAM?LE/BYPASS valve(s) In SAMPLE mode, see <al 139, 1
S.) In!itllll only: Connect the cas lne 01 11 ecuippe:l with a 3way SAMPlE/S?AJI v.,1ve pl.ce it In the SAMPLE mode.
1;,) Start the IOVII ppm O, gas (sample, 'ero,. purge Of spanl fiowing at 2'SCf"H to purge the
line 01 ar and atbl'r alr ealibration to purge the senSQ1 a.nd expedite reeovery t<:l ppm levels.
To aeeess tI>e senso! fOI repla:ement 01 air calibrallon:
Accessing Sensor
1.) Locate !he SS sensor houslng n.ide the analyzer.
2..) Using a 5/16" wrench, lcosen the bolt in the eentel 01 !he ,,,ttom secuon.
3,) Rotate the upper s""tion 01 the sensor housing 90
4.) PUI1-up genlly to the sensor houslng (00 NOT lose or dry off the o""ng).
1.) Short sensor acros. pes fOil. t,e,,,porarily
CalibratlOn Procedure 2.} CJ.an :ontact plns (damp la;)
3.) Perfonn eontinuty test - cente< pln to
ves black wire, outer pin to red wir.
1.1) Repl.cement: a,) Ren'lO'\'e 'o!d sensor' and dispose of - local regs fOI batteries.
b,) Remove 'new from N, purged shlpplng package
c.) Remove led shortng tabs from the pcs lolls
1.2) ln-service .Ir :ali'C3tion - if necessary, remove shOfting devke alter e.ontl.nuity test.
Ves
3.) A1low the reading to stablllle, normally 210 minutes unless to air pre'\liously,
4,) Once stable "' initiate U.UBRATION routne (or unlock and adjust SPAN knob) from
MAlN MENU, setect SPAN CAUBRATE, enter SPAN VALUE, plesS ENTER.
2,) lnsert PeS end 01 sensor lnto upper sectlon maklng contad wlth pins and hold it steady.
1,) Repeat w,th current sensor,
2.) Replaee sensor and repeal.
3.) 1I unsuccessful with new
sensor, STOP and eontaet fadory,
1.) Place the sensor in th. bottom sectlon and reassemble the sensor housing,
2.) le low ppm O, gas <above) should alrea"r be flowing at 2 SCFH to purge the to purge
the sensor and expedte 'ecovery to 10 ppm rlnge - oontinue for 6090 minutes.
1.) Pelform LEAl( TEST, see pq. 1
l,) If sueeessful, continue plirgc 1 hr.
3.) 11 unsu((essful afuo!r additional
.=:>----.....1 purge, replaee and r.peat.
No 4.) 11 unsuccessflJl with new
senSOr, !>OP and con:act lactory.
1.) Place analyzer 5ANPLE!BYPASS va!ve(s) In 6YPA5S mode, see (a) 139.1
2,) Connect Gas Une see 139. 1 - for sample gas as neeesry
3.) Place anal\,%er SAMPlEjBYPAS5 .... l\Ie($) In SAMPLE mode, see (al pg. 1
Rev 7}06 Page l 01 2
34
Advanced Instr.uflJents Inc.
9 WarralQ
The design and manufacture of GPR Series oxygen analyzers, monitors and oxygen sensors are performed under a certified
Quality Assurance System that conforms to established standards and incorporates state of the art materials and components
for superior performance and minimal cost of ownership. Prior to shipment every analyzer is thoroughly tested by the
manufacturer and documented in the form of a Quality Control Certification that is included in the Owner's Manual
accompanying every analyzer. When operated and maintaned in accordance with the Owner's Manual, the units will provde
many years of reliable service.
Coverage
Under normal operating conditions, the monitor, analyzers and sensor are warranted to be free of defects in materials and
workmanship far the periad specified in accordance with the most recent published specifications, said period begins with the
date of shipment by the manufacturero The manufacturer nformation and serial number of this analyzer are located on the rear
of the analyzer. Advanced Instruments Ine. reserves the right in its sole discretion to invaldate this warranty if the serial
number does not appear on the analyzer.
If your Advanced Instruments lne. monitor, analyzer and/or oxygen sensor is determined to be defective with respect to
material and/or workmanship, we will repair it or, at our option, replace it at no charge to you. If we choose to repair your
purchase, we may use new or reconditioned replacement parts. If we choose to replaee your Advaneed lnstruments lnc.
analyzer, we may replace it with a new or reeonditioned one of the same or upgraded designo This warranty applies to all
montors, analyzers and sensors purchased worldwide. It is the only one we will give and it sets forth all our responsibilities.
There are no other express warranties. This warranty is limited to the first customer who submits a c1aim for a given serial
number and/or the aboye warranty periodo Under no circumstances will the warranty extend to more than one customer or
beyond the warranty periodo
Limitations
Advanced lnstruments Ine. will not pay for: loss of time; inconvenienee; loss of use of your Advanced Instruments lne. analyzer
or property damage eaused by your Advaneed Instruments lne. analyzer or its failure to work; any speeial, incidental or
consequential damages; or any damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized
repair or modifieation of the analyzer; affixing of any attachment not provided with the analyzer or other failure to follow the
Owner's Manual. Some states and provinees do not allow Iimitations on how an implied warranty lasts or the exclusion of
incidental or consequential damages, these exclusions may not apply.
Exclusions
This warranty does not cover installation defects resulting from accidents damage while in transit to our service location;
damage resulting from alterations, misuse or abuse; lack of proper maintenance; unauthorized repair or modification of the
analyzer; affixing of any label or attaehment not provided with the analyzer; fire, flood, or acts of God; or other failure to follow
the Owner's Manual.
Service
Call Advanced lnstruments lne. at 909-392-6900 (or e-mail info@aii1.eom) between 7:30 AM and 5:00 PM Paeific Time Monday
thru Thursday or 8:00 AM to 12:00 pm on Friday. Trained technieians will assist you in diagnosing the problem and arrange to
supply you with the required parts. You may obtain warranty service by returning you analyzer, postage prepaid to:
Advanced Instruments lne.
2855 Metropolitan Plaee
Pomona, Ca 91767 USA
Be sure to pack the analyzer seeurely. lnclude your name, address, telephone number, and a description of the operating
problem. After repairing or, at our option, replacing your Advaneed lnstruments lne. analyzer, we will ship it to you at no cost
for parts and labor.
35
Adllanced Instruments Ine.
liD MSOS- Material SalelV Data Sheet
Product Identificaton
Product Name
Synonyms
Manufacturer
Emergency Phone Number
Preparation / Revision Date
Notes
Specific Generc Ingredients
Cardnogens at levels > 0.1%
Others at levels > 1. 0%
CAS Number
Chemical (Synonym) and Family
General,Requirements
Use
Handling
Storage
Physical Properties
Boling Point Range
Melting Point Range
Freezing Point
Molecular Weight
Spedfic Gravity
Vapor Pressure
Vapor Density
pH
Solubility in H20
% Volatiles by Volume
Evaporation Rate
Appearance and Odor
Fir;e and Explosion Data
Flash and Fire Ponts
Flammable Umts
Extinguishing Method
Spedal Fire Fighting Procedures
Unusual Fire and Explosion Hazards
IReactivity IOilta
Stabilty
Conditions Contributing to Instability
Incompatibility
Hazardous Decomposition Products
Conditions to Avoid
Oxygen sensor Series - PSR, GPR, AH, XLT
Electrochemical sensor, Galvanc Fuel (el!
Advanced Instruments Inc., 2855 Metropolitan Place, Pomona, o.. 91757 USA
909-392-6900
January 1, 1995
Oxygen sensors are sealed, contaln protective coverings and in normal conditions do not present a
health hazard. Information apples to electrolyte unless otherwse noted.
None
Potassium Hydroxide or Acetic Acid, lead
Potassium Hydroxde = KOH 1310-58-3 or Acetic Acid =64-19-7, lead = Pb 7439-92-1
Potassium Hydroxide (KOH) - Base or Acetic Ad (CHJC02H) - Acid, lead (Pb) - Metal
Potassium Hydroxide or Acetic Acid - electrolyte, lead - anode
Rubber or latex gloves, safety glasses
Indefinitely
KOH = 100 to 11S0 e or Acetic Acid = 100 to 117
0
C
KOH -10 to 0
0
e or Acetic Add - NA, lead 327
0
e
KOH = -40 to -100 C or Acetic Acid = -40 to -100 C
KOH = 56 or Acetic Acid - NA, lead =207
KOH = 1.09 @ 20' C, Acetic Add =1.05 @ 20' (
KOH = NA or Acetic Acid = 11.4 @ 20" C
KOH - NA or Acetic Add =2.07
KOH > 14 or Acetic Acid = 2-3
Complete
None
Similar to water
Aqueous solutions: KOH = Colorless, odorless or Acetic Acid = Col orl ess, vinegar-like odor
Not applicable
Not flammable
Not applicable
Not applicable
Not applicable
Stable
None
KOH = Avoid contact with strong adds or Acetic Acid = Avod contact with strong bases
KOH = None or Acetic Acid = Emits toxic fumes when heated
KOH = None or Acetic Add = Heat
36
Advanced 'nstruments Ine.
Spill or Leak
Steps jf material is released
Disposal
Health Hazafid Information
Primary Route(s) of Entry
Exposure Umits
Ingestion
Eye
Skin
Inhalation
Symptoms
Medical Conditions Aggravated
cardnogenic Reference Data
Other
Special Proteetion Information
Ventilation Requirements
Eye
Hand
Respirator Type
Other Special Protection
Special Precautions
Precautions
Transportation
Sensor is packaged in a sealed plastic bag, check the sensor inside for eledrolyte leakage. If the
sensor leales inside the plastic bag or inside an analyzer sensor housing do not remove it without
rubber or latex gloves and safety glasses and a source of water. Flush or wipe all surfaces
repeatedly with water or wet paper towel (ftesh each time).
In accordance with federal, state and local regulations.
Ingestion, eye and skin contact
Potassium Hydroxide - ACGIH TLV 2 mg/cubic meter or Acetic Acid - ACGIH TLV / OSHA PEL 10
ppm (TWA), lead - OSHA PEL .OS mg/cubic meter
Electrolyte could be harmful or fatal if swallowed. KOH = Oral LOSO (RAT) = 2433 mg/kg or Acetic
Acid = Oral LOSO (RAT) = 6620 mg/kg
Electrolyte is corrosive and eye contact could result in permanent 1055 of visiono
Electrolyte is corrosive and skin contact could result in a chemical bum.
Liquid inhalaron is unlikely.
Eye contact - burning sensation. Skin. contact - soapy s1ick feeling.
None
KOH and Acetic Acid = NTP Annual Report on cardnogens - not listed; LARC Monographs - not
listed; OSHA - not listed
Lead is listed as a chemical known to the State of california to cause birth defects or other
reproductive harm.
None
Safety glasses
Rubber or latex gloves
Not applicable
None
Do not remove the sensor's protective Teflon and PeS coverings. Do not probe the sensor with
sharp objects. Wash hands thoroughly after handling. Avoid contact with eyes, skin and clothing.
Empty sensor body may contain hazardous residue.
Not applicable
37
Ad'llanced Instrum'ent!!i 'ne.
Ap,peodix A
Electrical connections require an approved explosion proof sealing fitting and packing around wires and cables (for incoming
power for the analyzer electronics, power failure alarm relays, set point alarm relays and 4-20mA signal output) coming into and
out of the explosion proaf enciosure that houses the interconnection PeS. Further full compJiance with hazardous area electrical
code requires the wires and cables to be protected by conduit. Advanced Instruments recognizes that safety is an important
factor and offers for the convenience of our customers an approved sealing fitting and packing material as options.
ExPlosion Pr1Packina Filler (non-asbestos)
For use as packing at the hub of sealing fittings. Note: These instructions are supplied
from information and data which we believe is reliable and is given in good faith. 5ince
OUr methods of application and conditions under which our products are put to use are
beyond our control, we are not able to guarantee the application and/or use of same. The
user assumes all risks and liability in connection with the application and use of our
products.
Oirections: Tamp packing fiber between and around conductors where they enter fitting
to prevent leakage of the liquid cement. Leave enough space in the fitting for length
equivalent to the inside diameter of the conduit but, not less then 5/8".
Caution:
., Avoid getting in eyes or breathing dust.
., Use barrier cream, gloves and long sleeve shirts if dust or fiber is irritating.
., Prolonged contact may cause lung, eye or skin irritation.
ExPlosion Prool Sealina Cemem
Directions: Tamp packing fiber between and around conductors where they enter the sealing fitting to prevent leakage of
liquid cemento Make sure conductors are not in contact with each other or with fitting wall. Leave space in the fitting for a
sealing length equivalent to the trade size of the conduct seal but not less than 5/8". FiIl marked shipping container with c1ean
cold water to the "water line" (35 mi to be precise). Caution: Do not exceed the required amount of water. Gradually pour
cement from the plastic bag into the water and stir thoroughly for proper mixture. FiII fitting completely within five (5) minutes
after mixing, then tamp with blunt stick to expel any air bubbles. C10se up any opening in the fitting to insure integrity of the
seal. Fittings requiring more than 10 oZ. of cement must be filled from a single mixture of cement and water. DO NOT POUR
IN STAGES. Allow cement at least 72 hours to cure.
Warning: At least flVe threads must engage on all filll plugs.
Caution: Water-mix sealing compound should not be poured or installed at temperattlre below 40F (4e). Maintain
temperature at or aboye 40F for at least 72 hours after pouring. esA certfied when used with any eSA certified sealing fitting.
Adaco No. 1 sealing cement must be used as a part of any Adalet UL listed fitting.
Notice: This instruction is supplied from information and data which we believe is reliable and is given in good faith. 5ince the
methods of application and conditions under which our products are put to use are beyond our control, we are not able to
guarantee the application and/or use of same. The user assumes all risks and Iiability in connection with the application and
use of our products.
Caution:
Prolonged breathing or ingestion may cause internal obstruction, seek medical careo
., Do not get into eyes or on skin - flush for 15 minutes.
~ Large amounts on skin when hardening may burn .
., Use adequate ventilation.
To reorder: 5pecify P/N ENCL-1071-KIT
38
Adllaneed Instrumenm loe.
AppendixB,
Correlatino Readings -leO Displav and 4-20mA or 0-1V Signal OutPUIS,
In rare instances the signal output may not agree to the reading displayed by the Leo. The OUTPUT ZERO and OUTPUT SPAN
features enable the user to adjust the signal output to correlate with the LCD reading.
For optimum accuracy make two separate adjustments as follows:
1. OUTPUT ZERO feature: To adjust the 4mA or OV signal output and reqUires zero gas.
2. OUTPUT SPAN feature: To adjust the 20mA or 1V signal output and requires span gas near full range.
Note: In the field or in the absence of the preferred gases, use the OUTPUT SPAN feature and adjust the 20mA or 1V signal
output using the span gas available.
Guideline:
If the actual signal output value < the theoretical LCD value, the adjustment value will be < 100%.
1f the actual signal output value > the theoretical LCO value, the adjustment value will be > 100%.
Procedure = regardless of type of adjustment with examples: Signal Output
4-20mA O ~ l Y
1) The OlJTPlJT lERO and OlJTPlJT SPAN default values ... 100.0% 100.0%
2) Adjust the ~ a c t u a l sign,,1 output" .. 24mA 1.2V
to the ntheoretiCillI signal output H basedl on the "span gas value" of .. 84 ppm span gas 84% span gas
3) "nd the "actu;lllCO reading" ... 60 ppm 80%
a) "actual LCD reading" divided by the "span gas value" 60/84 =.71 80/84 = .95
b) 16mA (20mA less 4mA) multiplied by \he result in 3a) .71 x 16 = 11.36 NA
e) 4mA plus the result of 3b) = "theoretieal signal output
H
4 + 11.36;; 15.36mA ,95V (from a)
4) "Adjustment Va'lue" '" ("theoretical
H
divided by "actual signal output ") x 100 (15.36/24) x 100 ;; 64 (.95/ 1.2) x 100 ;; 79
5) Proof = ("actual signal output" multiplied by "adjustment value") divided by 100 (24 x 64) / 100;; 15.36mA (1.2 x 79) / 100 = .95V
6) ENTER "adjustment value
H
via OlJTPUT lERO or OlJTPlJT SPAN routines below. 64 79
7) Fine tuning ... after ehecking the new"actual signal output
H
at the PLC 19.5mA .98V
8) Repeat Step #4 and substitute the latest "actual signal output
H
(15.36/19.5) x 100 =78.8 (.95/ .98) x 100;; 96.9
9) Proof;; ("actual signal output
H
multiplied by "adjustment value
H
) divided by 100 (19.5 x 78.8) 1100 = 15.36 (.98 x 96.9) 1100 =95
10) ENTER "adjustment value
H
via OUTPlJT lERO or OlJTPUT SPAN routines below. 75 95
39
Adllanced Instrument Inc.
OutputZero
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARROW keys to highlght CAllBRAnON.
3. Press the ENTER key to select the highlghted menu option.
4. The following displays appear:
MAIN MENU
AUTO SAMPLI;
MANUAL SAMPLE

>
CONFlG ALARMS
BYPASS ALA.RMs
CALIBRADON
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULli SPAN
DEFAUlT ZERO
OUTPUTSPAN
[.11 Al)
5. Advance the reverse shade cursor using the ARRDW keys to highlight DUTPUT ZERD.
6. Press the ENTER key to select the highlighted menu option. .
7. The following display appears:
..
I '
: 100.0 t
I - 1
""------,
OUTPU'f ZERO OFFSET
PRI;SS 'UP- OR DOWN
TO CHANGE VALUE
ENTER TO SAV:E
MENU TO RE1\JRN
8. Enter the OUTPUT ZERD adjustment value as follows ...
9. Selecting digits of the adjustment value to change - Press the ENTER key to advance the underline cursor right or press
the MENU key to advance the underline cursor left to reach to the desired digit.
10. Changing the digits of the adjustment value ... Press the UP/DDWN ARRDW keys to change a digt of the adjustment
value.
11. Repeat steps 9 and 10 until the complete zero value has been entered and the following display appears.
OUTPUl" ZERO OFFSET
PRE$ UP OR DOWN
TO CHANGE VALUE
I;NTER TO SAVE
MENU TO RE1\JRN
12. Save the adjusbnent value by pressing the ENTER &<ey or abort by pressing the MENU key.
13. The system returns to the SAMPllNG mode.
14. Note: After the initial adjustment is made, check the actual signal output at the PLC and if necessary calculate a further
fine tuning adjustment value as illustrated above.
15. Enter subsequent fine tuning adjustments as described above,
Output Span
Procedure:
1. Access the MAIN MENU by pressing the MENU key.
2. Advance the reverse shade cursor using the ARRDW keys to highlight CAllBRATlON.
3. Press the ENTER key to select the highlghted menu option.
4. The following displays appear:
40
Adllanced Instruments Inc.
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
t!J 11 :!;,., ,t.):1
CONFIG ALARMS
BYPASS ALARMS
>
CAUBRATlON
SPA'N CAUBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULTZERO
r:m ,Ji) ..14':1
OUTPUTZERO
5. Advance the reverse shade cursor using the ARROW keys to highlight OUTPUT 5PAN.
6. Press the ENTER key to select the highlighted menu option.
7, The following display appears:
, 100.0:
,-----
OUTPUT SPAN OFfSET
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO REnJRN
8. Enter the adjustment value as follows ...
9. Selecting digits of the adjustment value to change - Press the ENTER key to advance the underline cursor right or press
the MENU key to advance the underline cursor left to reach to the desired dgito
10. Changing the digits of the adjustment value ... Press the UP/DOWN ARROW keys to change a digit of the adjustment
value.
11. Repeat steps 9 and 10 until the complete span value has been entered and the following display appears.
0 6 ~ . O
OUliPUli SPAN OF,FSET
PRESS UP OR DOWN
TO CHANGE VALUE
EN:rER TO SAVE
MENU TO RE1lJRN
12. saye the adjustment value by pressing the E:NTER key or abort by pressing the MENUI key.
13. The system returns to the SAMPUNG mode.
14. Note: After the initial adjustment is made, check the actual sgnal output at the PLC and if necessary calculate a further
fine tuning adjustment value as iIIustrated above.
15. Enter subsequent fine tuning adjustments as described above.
0 7 ~ . O
'OUTPUT SPN OFFSET
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER To SAVE
MENU TD RETURN
16. saye the adjustment value by pressing the ENTER key or abort by pre$Sing the MEi'iU Ikev.
17. The system returns to the SAMPUNG mode.
41
Adllanced InstrunJents Ine.
AppendiK F
H2S.Scrubbers &Sample Systems
Advaneed lnstruments lne. offers a complete line of efficient hydrogen sulfide (HzS)
scrubbers !md sample coooltioning systems for:
Prepllring a gas strellm conUlining HzS fer oxygen analysis
Zero gas generation for HzS tr"zlnsmitters
The scrubber medill selectively removes HzS and other oxidizing gases from the gas
stream that can interiere with the oxygen measurement. As an iOOication the
scrubber is nearing the end of its useful lfe the media changes color from purple te
orange te bro",,'O to white as it tS consumed.
The information included herein is based on data sheets published by the
manufacturer of the scrubbng media as follow: the media not only adsorbs gases but
1IIso chemically transforms them into harmless end products that remain trapped in
the media. Unlike adsorption, chemisorption is an jnstanteous an irreversible process
that perrnanently removes unwanted gases from the environment. The polZlssium
permanganllte (KMn04), the purple eolored media inside the scrubber, tums into
mangllnese oxide aOO elemental sulfur salt which are stllble non-Ieachable solids.
Scrubbers ane available in refillable:
)- 1.0" x 6" (P/N B-2734-6) plexi-glass container with SS connections
)- 1.5" x 12" (P/N B-2734) plexi-glass conUliner with SS connections
)- 2.0" x 24" (P/N A-2839) stllinless steel container with SS connections
)- 2.5" x6" (P/N 8-3247) plexi-glass container with SS connections
)- 2.5" x 24" (P/N B-3247-24) plexi-glass container with SS eonnections
They can be installed in-line or as plIrt of a complete serubber sal11'le coOOitioning
system. More elaborate custom designed scrubber sllmple conditioning systems can
inelude dual scrubbers aOO vlIlve system that can eliminlltc downtime. Opel"lltors
smply switch the gas flow to the fresh scrubber whle servicing Ihe other one.
Speclflcations:
lnlet connections: 1/4" SS tube fitting (refill port)
Outlet connections: 1/4" SS tube fitting
Pressure rating: 30 psig maximum
Temperllture: 200c te 450C (-40F te llJOF)
Application conditions: Free of moisture (ITllIY require cOlllescing filter)
Materials of construction: clear acrylic lInd/or stainless steel
Scrubber life: V!lries with flow rate, media volume, H
2
S concentration
iDlmenslons H
2
S Flow Service
partNo. Concentratlon Bate UfeOays
1.5- x 12'" 1.0% (10,000 ppm) 1{3 SCFH (150 sccm) 3.6
(P/N 8-2734) 0.1% ( 1,000 ppm) same 36.8
0.01% ( 100 ppm) same 368.2
2.Sx 6 1.0% (10,CXXl ppm) 1/3 SCFH (150 sccm) 6.0
(P/N B-3247) 0.1% ( 1,000 ppm) same 59.6
0.01% ( 100 ppm) same 596.4
42
8-2734-6
8-2734
8-3247
1x6"
1.5x12"
2.Sx6"
8-3251 H2S Scrubber Sy5tem
GPR-1500 AlS wIth
8-3247 H2S Scrubber
A-3393 H2S Scrubber System
Advanced InstrulDentli
@@
0@
SPAN GAS IN
o
'7
H2S Sample SJstem
A-3393 STANDARD CONFlGURATION
B-33S9 S4M'I.f P,<;ld IN:n 9-OM'>I)
1 A..TR1OO2_ F1LTB<CO\I3aI<G
2. . AlIJE SS:;W
3 FMlJ'. ] 7'1 r-.'TEl3W. VALIJE
4 71 SOl.UllS\ M2S C1.EAA 'O.YeIlll8 2 5" O
S. 6--3330-A18 -uJ5IN3 SS
FI1N-IOlt> 1I0'TO lj,!"
s.o.M'U:, VB' CCN-CTlO'IS
A-3393-2 COHTIlIIUOUS ORAlN OPTlON:
.00 {) VA:..Y100'1 V VE SS 2WAY l-EI'Bl. .13
R5'l.Al:f 7 RoT1H024 Fll'l15t CQ'\LfSCJ!'fj, I D'W
/O) VAl. 1633 VAlVE "'""me 2.... I
A-3393-4 PR.fSSURE REGlAATOR .. GAUGE OPTlON:
R.5'lXE IH3S8 SA"Ili "Ala (R6'l.ACES 1>-3:3':<9)
N:J) 9 RE(;. J 1.3 ;'-'-fS:U'.f Rro..LATO'l. SS
N:J) 1O.11EG-10CJ?
PRESSURE RfGlAATOR .. GAUGE WITH
CONTINUOUS ORAlN OPTlON:
CONTINUOUS ORAlN SET-tJP:
A.
B. QOSE YALYE
C. $El REGU.ATOR., 20 30 PSIG
D. el'L"N 'iAlVE 10 3 SCfH ON FlOW Io'ETER
E. CPEN VAl\1 TO 1.5 SCfH ON FlOW 1-ETm.
F. SETVAlYEONFlOWt-t:lffi TOlSCFH
FOR Sl'AN GoIS, AF1ER Sl'AN, OPEN YAl\
FOR SAM'l

FlOW 1.5 SCFH
A-3393-1 NO H2S SCRUBSER OPTION:
A,-3393-3 NO H2S SCRUBSER WlTH CONTlNUOUS
ORAlN OPTION:
CO"eIN" C\?TIO ,- 1 2
A-3393-5 NO H2S SCRU98ER WITH PRfSSURf
UGlLATOR GAUGE OPTION:
A,-3393-7 NO H2S SCRUB8fR WITH PRESSURE
REGUlATOR .. GIWGE WITH CONTINUOUS
DRAlN OPlION:
Advllnced Instruments Inc.

T}Ot'tJ7 1$
,ID "K:.11:;" .AI.;,oAAA;..m

__ ....... m'-JL... If-"'"'--
GPR-1500 AIS/IS SAMPLE SYSTEM A-3393
J:
u..
U
Vl
lI)
....
VENT
43
Adllaoeed lostrumeots loe.
Material Safety Dala Sheet U.S. Department of Labor
May be used lo comply with Occupallonal Salely ana Heallh AdmmlslrallOn
OS-Ws Hazard Communicalion Standard. (Non-Mandatory Form)
. CFR 1910.1200. Sl'andard mus! be orm Approved
Isufted lar sPEf<:lf,e r uirements OBM No. 1218-0D72
ENTln' (As sed on Laoel and List)
Note B!anl<. spa""s am no! pe,mirtod // any Wltm /S not appkcatJle. or ne
UNISORB t1ARK 2
nform.oon is ,.". ce musi blI t1JlIrkfKJO ro indicare In.'
Section I
ManufacltJrer'.. Nome Eme<gency Telephonc Number
7 3) 943-3753 UNISORB CORPORATIQ
1elephone Number for InJOTrrnJ'IOn Addreu (N:Jmbe;. S'ma: CIy. SI.'e. Bnd Z/P Codto)
(7 3)
13 O GENOA STRE5:':'
J."-:'1UARY 7, 2002
Sigoalure or Prepare< (OpI/onal)
SOUTH HOUSTON, T>: 77'.J87
Section 11 - Hazardous Ingredients/ldentity Information
Com:>:X":en1s
(Spe:f,c Chemi<:a lden!;t.,.: Common Name{s)) ACGIH TLV
lSmg/m 10mq/m'
Alumi um OHide: Activatcd Alumina NO:'le
(Th'll.) (T\'IJ\.)

permanganate: Potassw. Salt Non:
('1%'\ or Mn) (T\'IA r
uminum Oxide: CAS - 1344-2B-l
Potassium Permanganate: - 7722-64-7
Section 11I - Physical/Chemical Characteristics
Boillng Point
NI.:"
Spc.dfic Gmvftv
3.2
VaxJl' Pressure (mm Hg )
,lA
MeltIn9Poln1
2050
Vapor Density (Air o ')
NI.'"
EVJporallon Rala
NIA
Solubilily III Water
ln501uble
Appeamnce and OdOl
Purple Spheres, No Oaor
section IV - Fire and Explosion Hazard Data
Flash PoiI\! (M"lhoO flarnma.bll! LJ UEl
No.e Not FlaJTOllable N/A
Exmg . hill9 Media
Foam, Carbon Dioxide, O' 1',. hemical
eclal File Flghllng ProcedUl!!S
Use RespildtOPj Protection
Unus.ua! Fll'e en:! ExpIosIon Haza;ds
None
44
VenULallOn Local Exl>aust RecollUllE'nded
if is
Kcne
OlhPr
None
Adllanced Instrumen'ts loe.
J !SORB MARI< 2
aqents and easi1y materials,
Haxardou!!
Polymcriza\lOn
MayOceur Cond!llons lo AIIOld
None
W,Il Nol Oeeu'
X
Section VI Heatth Hazard Data
ROUie(s) 01 Enlty ? Skln? Ingestion?
'es 'fes Yes
Hea':h Hazard$ (ACtll and Cnrome)
CHRO!\. :
Carcinog<mtclty: NTP?
AlU1l.num Ox e: t o 1
Potassiurn
Sqts and 01 Expowre
Brown disco oration oE th
tissue.
hazards.
OStiA Regulated
OSHA.
MedalCond s
o: open wounds or b rns.
EmergencyandF.-stAldProcedureo; ;:;YE: F:ush ith larga amounts oi water ior al: mi;].
Seek medical hel? immediate_y. j,GEST ON: Give larga quantities of water. Seek medical
imm<:diately. 1 :!.l'.LATIO.": Remove to fresh airo Seek medical help iJm1edi tely.
: Wash ? enty o: water. Seek m dieal tention t irritation perssts.
Section VII - Precautions for Safe Handling and Use
Steps \0 Be Taken 11 case MlJteriallS Released or sp1Je<l
Sweep or shovel material and discard in protected refuse con ainer. '1'0 clean floors,
flush with water 'nto sewcr if oCrnUtted by Feder , State, and loe requlations.
Wasle O,sposaI Method
ew ial may di posed of in tandflll. Jsed that has been exposed to
Loxa chemicals should be examinec and disposed or i aeco_dance with all cegulatons.
lo Be Teken 11 Handling and SIO(,ng
J ae_orcance control measures be1ow. 5tore in sealed contain es a co01,
dcy locaLio..
Oll>er PceocaUlions
Non_
Section VIII - Control Measures
Prclection (specdy rype)
Use NIOSH aporoved orotectioa if are exceedec.
PIOlldi,,,, GJove!,
I
Ey* Prol*<:tlC>n
Rubber gloves -aee or 90991es
O:l\er Prote<:tlve Clothmg or Equpmert
Normal wo:k clothinq ccvec.nq and legs shculd b wc:n.
Wotk Hyglene Practlces
Remove conta, nated e oth_ng a.c wash before re se. Wash with soap nd wal oC befare
3t1n9, drinking, or using t iae afe y shower, eye ath and
acilities shou d be e.
45
Advanced Instruments Ine.
Appendix G
Ilimenance - H2S Scrullller
5ervicing any of the H25 scrubbers will depend on severaI factors as iIIustrated in Appendix F and inelude: the (average) H25
coneentration (average), volume of scrubber media and flow rate through the scrubber (often times maximizing the service Iife
means longer system response time) see Appendix F.
Required equipment:
1. 2x 7/16" open end wrenches
2. lx 9/16" open end wrench
3. 1x 1" open end or adjustable wrench
Procedure:
1. Separate the top connection to the serubber using a 7/16" and the 9/16" open end wrenches on the two top nuts.
2. Hold the second nut with the 9/16" open end wrench.
3. With one of the 7/16" open end wrenches tum the top nut counter c10ckwise until the fitting disengages.
4. 5eparate the bottom connection to the scrubber using both 7/16" open end wrenches.
5. Hold the nut at the bottom of the scrubber with a 7/16" open end wrench.
6. With the other 7/16" open end wrench turn the nut below counter c10ckwise until the fitting disengages.
7. Carefully, do not loose the ferrules inside the fitting, remove the stainless tubing from the top and bottom of the scrubber.
8. Carefully pull the scrubber from its mounting clip which is attaehed to the back panel.
9. Once the scrubber is free, hold the scrubber with one hand and using the 1" open end or adjustable wrench with the other
hand, turn the 1" nut counter c10ckwise and remove the 1" nut at the top of the scrubber.
10. There is no need to remove the 7/16" fitting at the bottom of the scrubber.
11. With the 1" nut removed, empty the spent media through the opening.
12. FiII the scrubber with fresh media (should be rich purple in color).
13. Reverse the aboye steps to re-assemble and install the scrubber.
MainteRance - Coalesc.ng Filter
Servicing the coalescing filter (P/N FLTR-1002-2) depends on the c1eanliness and moisture content of the sample and
maintenance intervals.
Required equipment:
1. Channel locks
2. Damp rag
3. Lubricant (a thin coat applied to the o-ring after eleaning helps ensure a tight seal and extend o-ring Iife)
Procedure:
1. Unscrew the c1ear polycarbonate bowl by turning it counter c1oekwise.
2. Note: It is probably stuck tight - use the damp rag to grip if removing by hand or to prevent damage to the bowl jf using
the channellocks.
3. The bowl seals to the head section with an o-ring, do not lose the o-ringo
4. The filter element screws into the head section, carefully turn it counter clockwise and remove.
5. Using the damp c1oth, c1ean the inside of the bowl and the o-ring before reassembling - apply a very thin coat of lubricant
to the o-ringo
6. Reverse the aboye steps to re-assemble the filter.
46
Technical Specifications
Accuracy:
Analysis:
Application:
Approvals:
Area ClaSSification:
Alarms:
calibration:
Compensation:
Connections:
Controls:
Display:
Endosures:
Flow Sensitivity:
Linearity:
Pressure:
Power:
Recovery Time:
Response Time:
Sample System:
Sensitivity:
Sensor Model:
Sensor Life:
Signal Output:
Temp. Range:
Warranty:
Wetted Parts:
< 1% of F5 range under constant conditions
0-10 ppm, 0-100, 0-1000 ppm, 0-1%, 0-25% FS ranges; auto
ranging or manually lock on single range
Oxygen analysis from 100 ppb to 1% in inert, hydrocarbon,
helium, hydrogen, mixed and acid (COz) gas streams
CE, ATEX ntrinsic Safety certification (pending) Zone 1
Meets recognized intrnsic safety standards fer use in Zone 1;
Class 1, Division 1, Group A to D hazardous areas
Two user configurable alarms: magnetic coil relays rated 3A at
lOOVAC, field programmable time delays, 2 ppm deadband,
bypass for calbration; sensor and system fail.
Certified gas of O
2
balance N
z
approximating SO% of range
aboye analysis range recommended for optimum results.
Barometric pressure and temperature
1/4" compression tube fittings
Water resistant keypad; menu driven and field selectable
range selection, calibration value and alarm functions
Graphical LCD 2.75x1.375"; resoluton .01 ppm; displays real
time ambient temperature and pressure
A1uminum ex-proof, fiberglass NEMA 4X, 16x9x6", 20 lbs
None between 0.5-5 SCFH, 2 SCFH recommended
> .995 over all ranges
Inlet - regulate to 5-30 psig; vent - atmospheric
12-2SVDC (suitable for solar powered systems); 100-240VAC
50 seconds in air to < 10 ppm in < 1 hour on N
2
purge
90% of final F5 reading in 15 seconds
Flow meter; shown with optional A-3393 H25 Sample SY$lem
15x16" panel with coalescing filter, 3-way sample/span valve,
H2S scrubber 240g of media, bypass valve aplional
< 0.5% af FS range
GPR-12-333 for inert gases, HC, He, H2 and mixed gases
XlT-12-333 with > 0.5% COz present - no maintenance
24 months at 25C and 1 atm; average Oz < 1,000 ppm
4-20mA or 1-SV
Sto 45
0
( (GPR sensor), -20 to 4SC (XLT sensor)
12 manths analyzer; 12 manths sensor
Stanless $leel
6PR-'500 AIS
pplIJ Oxygen TranslIJltter
Intrinsically Sate Design
ATEX Certification (pending) Zone 1
Class 1, Division 1, Groups A to O
Tailored tor Natural Gas Applications &
Outdoor Battery Powered Installations
Advanced Sensor Technology
Accuracy < 1% fS Range
fast Recovery to < 10 ppm
No Maintenance, 24 Month Lite
Compatible in 0-100% COz
De or AC Power, Integral Barriers
Field Selectable Alarms with Time Delay
4-20mA or 1-5V Signal Outputs l'
5 Standard Ranges
Auto or Manual Ranging
ISO 9001:2000 Approved QA System
Optional Equipment
Intrinsically safe integral heater for extended temperature range - 40C/F
Expand Sample System to indude bypass valve, pressure regulator w/gauge and/or
continuous liquid drain with regulating valves - consult factory.
2855 Metropalitan Place, Pomana, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, www,aii1.com, e-mail: info@aii1.com Rev 6/07

Potrebbero piacerti anche