Sei sulla pagina 1di 8

C122-E057

AA - 7000 Series

S h ima d z u
A t o m ic A b s o r p t i on S p e c tro p ho to meter s
Reaching even greater heights Unprecedented sensitivity,
thanks to a newly developed 3D optical system

Advanced safety technology

System configuration evolving to suit the needs of the user

The newly developed 3D optical system is capable of flame:


Pb 0.015 ppm detection, and furnace: Pb 0.00005 ppm detection.

Equipped with the world’s first*1 vibration sensor, a multimode automatic


gas leak check function, and other advanced safety technology.

Evolving from the basic flame model to suit the needs of the user.
High Performance The dual atomizer system*2 is the world’s smallest*1, and a furnace
model is also available.

Easy-to-understand, user-friendly WizAArd software supports FDA


21 CFR Part 11 compliance. QA/QC functionality and hardware
validation software are also provided.
*1 Survey by Shimadzu in November 2008.
*2 The optional auto atomizer changer is required.
FLAME MODEL
Highly stable, enhanced flame model
And it doesn’ t stop there...
FURNACE MODEL
Cutting-edge furnace model, achieving unprecedented sensitivity
And beyond…
DUAL ATOMIZER SYSTEM
Dual atomizer system, compatible with a variety of analyses
Basics Flame (continuous mode / micro sampling method *1)
Flame continuous method, flame micro sampling method *1, furnace method Burner head Titanium 10¬cm slot (for Air-C2H2 flame)
Optics Optical double-beam, automatic switching to single beam during furnace measurement Nebulizer Pt-lr capillary with PTFE orifice, ceramic impact bead
Aberration-corrected Czerny-Turner spectrometer, photomultiplier tube (capable of handling hydrofluoric acid)
Wavelength range 185.0 to 900.0¬nm, automatic wavelength setting via line search Chamber Engineering plastics
Bandwidth 0.2, 0.7, 1.3, 2.0¬nm (4-step automatic switching) Positioning Lateral/vertical manual adjustment,with option mounted,
Lamp mode EMISSION, NON-BGC, BGC-D2, BGC-SR automatic atomizer switching / automatic search of optimum burner height

Background correction BGC-SR (high-speed self-reversal method) and BGC- D2 (D2 lamp method) Flow rate control Automatic fuel gas flow rate setting, automatic search of optimum gas flow rate

Number of HC lamps 6-lamp turret, 2 lamps simultaneously lit (1 for measurement, 1 warming up for next measurement) Safety measures Automatic flame extinction via flame vibration sensor, automatic gas leak check, etc.

Atomizer position adjustment Manual (standard), automatic (with option mounted)


Furnace
Concentration computation Calibration curve method, standard addition method, simple standard addition method
Heating temperature range Ambient to 3,000¬°C
Repeat analysis Up to 20 repetitions, automatic exclusion of deviant values by setting SD and %RSD
Heating control system Drying: Digital current control with automatic temperature calibration function
Analog output 2 channels (atomic absorption/energy signal, background signal) Ashing, Atomization: Digital temperature control via optical sensor
Output range: Both 4-level switching *2
Setting heating conditions Maximum 20 stages Inner gas type: Dual automatic switching type
Data processing Calculation of final concentrations, QA/QC function High-sensitivity mode setting Enrichment in furnace: Maximum 20 times
Re-analysis Automatic dilution and re-analysis of unknown samples (flame micro sampling method, furnace method) Optimum temperature program search support function

Electronic Records Compatible with FDA 21 CFR Part 11 Compliance Power requirements Choose from 200¬VAC, 220¬VAC, 230¬VAC or 240¬VAC. ±5%, 6¬kVA, 50/60¬Hz combined

Power requirements *3 Choose from 100¬VAC, 120¬VAC, 220¬VAC, or 230¬VAC. ±10%, 230¬VA, 50/60¬Hz combined Dimensions and weight 260¬W¬×¬555¬D¬×¬527¬H¬mm, 46¬kg
(power supply unit)
Dimensions and weight *4 AA-7000F: 700¬W¬×¬588¬D¬×¬714¬H¬mm, 72¬kg
AA-7000G: 700¬W¬×¬580¬D¬×¬538¬H¬mm, 65¬kg (excluding protrusion and options) Personal Computer / Monitor
Ambient temperature / humidity 10 to 35¬°C, 20 to 80% (less than 70% when temperature is higher than 30¬°C) Operating system Microsoft Windows Vista Business or XP Professional
CPU Intel ® Celeron 420 (1.60¬GHz) or higher
RAM 1 GB or higher (Vista) or 512¬MB or higher (XP)
Monitor XGA (1024¬×¬768 dots) or higher
Storage device One CD-ROM drive (for installing software)
Requires 60¬MB min. hard disk space for installation
Founded in 1875, Shimadzu Corporation, a leader in the I/O port One serial port (for AA control)
development of advanced technologies, has a distinguished
*1 Optional autosampler and micro sampling kit are required for the flame micro sampling method.
history of innovation built on the foundation of contributing to *2 1V F.S. fixed in EMISSION mode.
society through science and technology. We maintain a global *3 Power is required separately for the personal computer.
network of sales, service, technical support and applications *4 Protruding parts and optional equipment are not included.
centers on six continents, and have established long-term *Windows is a registered trademark of Microsoft Corp., U.S.A.
relationships with a host of highly trained distributors located
in over 100 countries. For information about Shimadzu, and Dimensions (Dual Atomizer System)
to contact your local office, please visit our Web site at
www.shimadzu.com

588

714
870

SHIMADZU CORPORATION. International Marketing Division

390
3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan
Phone: 81(3)3219-5641 Fax. 81(3)3219-5710
280

URL http://www.shimadzu.com
340
Printed in Japan 3295-01910-30ANS 960