Sample
Sample Preparation
The sample is submitted for testing without prior preparation.
SCA_125_009
www.shimadzu.de
Table 1 Thickness Determination of Polyester Films by Background FP Method Sample Chemical Formula Film Density Determined Value 7.2 m Polyester (C8H10O4)n 1.39 g/cm3 14.5 m Films 29.7 m
Reference Value 6 m 12 m 25 m
Fig.3 Profiles of Polyethylene Film for Quantitative Analysis Table 2 Determined Thickness and Concentration of/in Polyethylene Film by Background FP Method Determined Value Standard Value Sample Film Density
Polyethylene Film including Inorganic Compounds 0.92 3 g/cm ((CH2)n) Cr 91 g/30 mm Mn 297 Fe 219 Co 21 Ni 87 Cu 85 Zn 88 Thickness 218 m ( 645 ppm) ( 2096 ) ( 1546 ) ( 149 ) ( 612 ) ( 598 ) ( 620 ) 100 g/30mm 300 200 20 100 100 100 200 m ( 769ppm) ( 2307 ) ( 1538 ) ( 154 ) ( 769 ) ( 769 ) ( 769 )
Analytical Conditions
Instrument: EDX-700 X-ray Tube: Rh target Filter: not used Voltage - Current: 50 kV 77 A (Auto) 50 kV - 446 A (Auto) Atmosphere: Air Measurement Diameter: 10 mm Measuring Time: 100 sec Dead Time: 24 %
References
Shimadzu Application News X-ray Analysis No.175,176 The Application of Background FP Method(1),(2). Shimadzu Application News X-ray Analysis No.190 The Measurement of the Thickness of a Polymer Coat on a Steel Plate Using the Theoretical Intensity of the Compton Scattering X-rays.
www.shimadzu.de
The given specifications serve purely as technical information for the user. No guarantee is given on technical specification of the described product and/or procedures.