IEC 789:1995
STANDARDS AUSTRALIA
1 The Crescent,
Homebush NSW 2140 Australia
STANDARDS NEW ZEALAND
Level 10, Standards House,
155 The Terrace,
Wellington 6001 New Zealand
ISBN 0 7337 0407 7
ii
PREFACE
This Standard was prepared by the Joint Standards Australia/Standards New Zealand Committee
HT/23 on Medical Diagnostic Imaging Equipment, as a Joint Standard.
This Standard is identical with and has been reproduced from IEC 789:1992, Characteristics and
test conditions of radionuclide imaging devices; Anger type gamma cameras.
As this Standard is reproduced from an international Standard, the following applies:
(a) Its number does not appear on each page of text and its identity is shown only on the cover
and title page.
(b) In the source text ‘this International Standard’ should read ‘this Australian/New Zealand
Standard’.
This is a free 6 page sample. Access the full version at http://infostore.saiglobal.com.
(c) A full point substitutes for a comma when referring to a decimal marker.
References to international Standards should be replaced by references to the following Australian
or Joint Australian/New Zealand Standards:
Reference to International Standard Australian/New Zealand Standard
or other publication
IEC AS
788 Medical radiology —Terminology —
NEMA —
NU1 Performance measurements for
scintillation cameras
The term ‘informative’ has been used in this Standard to define the application of the Annex to
which it applies. An ‘informative’ Annex is for information and guidance only.
CONTENTS
Clause Page
3 Test methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3.1 Measurements of SYSTEM SENSITIVITY .. . ... .. ... .. ... .. ... . .. . 1
3.2 LINE SPREAD FUNCTIONS , FWHM, FWTM and EW . . . . . . . . . . . . . . . . . 2
3.3 NON- UNIFORMITY OF RESPONSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3.4 Measurement of INTRINSIC ENERGY SPECTRUM (without COLLIMATOR) . . . 4
3.5 Measurement of INTRINSIC MULTIPLE WINDOW SPATIAL REGISTRATION . . . . 5
This is a free 6 page sample. Access the full version at http://infostore.saiglobal.com.
Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
The object of this Standard is to specify terminology and test methods for declaring the
characteristics of Anger type GAMMA CAMERAS . These consist of a COLLIMATOR, a DETECTOR
SHIELD , and a RADIATION DETECTOR ASSEMBLY , together with recording and display devices.
For the purposes of this Standard, the definitions given in IEC 788 apply, together with the
supplementary definitions given in annex A.
3 Test methods
All measurements shall be performed with the PULSE AMPLITUDE ANALYSER WINDOW settings
specified in table 1.
Unless otherwise specified, measurements shall be carried out at count rates not exceed-
ing 20 000 counts per second.
The measurement shall be carried out using the cylindrical phantom of polymethyl-
methacrylate as specified in figure 2. The plane source shown in figure 1 shall be placed in
the cylindrical hole with the dimensions shown in figure 2, the remainder of the hole shall
then be filled by a cylindrical part the dimensions of which are also shown in figure 2. The
phantom, including the source, shall be placed on the COLLIMATOR (distance d = 0) and
centered on the COLLIMATOR AXIS.
NOTE - Measurements of SYSTEM SENSI TIVITY without scatt er, using the source cuvette of fi gure 1 placed
directly on the COLLIMATOR FRONT FAC E, may be carri ed out.