1. 2. 3. 4. 5.
Using the oxygen concentrator. Managing the disposable accessories. Cleaning the oxygen concentrator. Simple Troubleshooting. Recommended Biannual Maintenance for Provider.
source. Use only the recommended disposable accessories provided by the Service Provider. Ensure that the patient adhere to the given prescription given by the doctor Ensure proper ventilation is given when the unit is in use.
To ensure proper ventilation for unit, the unit should be place at least 1 Feet away from any solid object.
1 Feet Distance
Disposable accessories.
Disposable Accessories
All
disposable accessories should be kept clean. All disposable accessories are recommended to be change every 2 weeks or earlier if they are found faulty or too dirty.
a clean damp cloth to wipe the external part of the unit. Clean the Gross particle filter on a weekly basis.
Simple Troubleshooting.
Probable cause
Solution
Unit does not turn ON, continuous beep alarm, compressor not moving.
Check power source. Remove from multi plug or multi adaptor and plug directly to wall socket. Unit does not have any power.
Simple Troubleshooting.
Probable cause
Solution
Restricted Air flow. Ball of flow meter does cannot be adjusted. Little or No bubbles coming out from the humidifier.
Check nasal cannula for kink. Check humidifier, if placed properly onto the unit. Replace nasal cannula or humidifier or both.
Simple Troubleshooting.
Probable cause
Solution
Maintenance Programme
Lifeline
Innovators upon recommendeation of the manufacturer recommends that a regular Bi-annual service maintenace be done on the unit. This will ensure proper working condition of the oxygen concentrator.