Sei sulla pagina 1di 112

3400 Series

Particle Counter
Operator Manual
Met One
September 2005

KSS - September 2005 - Edition 3

Model 3400

Operator Manual

Model 3400 - Table of Contents

1 of 110

Table of Contents
1

Introduction
1.1
1.2

Initialization
2.1
2.2

2.3
2.4
2.5

2.6
2.7

3.3

3.4

3.5

KSS - September 2005 - Edition 3

Electrical Connections ........................................................................... 13


Battery Charging .................................................................................... 15
2.2.1
Using the 3400....................................................................... 15
2.2.2
Using the External Battery Charger ....................................... 16
Loading Printer Paper ............................................................................ 16
Getting Started....................................................................................... 17
Screen Orientation ................................................................................. 19
2.5.1
Overview................................................................................ 19
2.5.2
Sliding Menu .......................................................................... 20
2.5.3
Battery Status ........................................................................ 20
Quick Start: Taking a Count................................................................... 21
Main Screen Sliding Menu ..................................................................... 23

Configuration
3.1
3.2

Met One

Operation Conventions .......................................................................... 10


Accessories............................................................................................ 11

Setup Overview...................................................................................... 25
Global Settings....................................................................................... 27
3.2.1
Basic Tab Overview............................................................... 28
3.2.2
Sounds Tab ........................................................................... 30
3.2.3
Users Tab Overview .............................................................. 32
3.2.4
Options and Accessories Tab................................................ 32
Sample Screen ...................................................................................... 34
3.3.1
General Tab Overview........................................................... 35
3.3.2
Timing Tab Overview............................................................. 40
3.3.3
Environmental Units Tab ....................................................... 42
Alarm Setup ........................................................................................... 43
3.4.1
Count Tab Overview .............................................................. 45
3.4.2
Environmental Tab Overview................................................. 46
Group Settings ....................................................................................... 48
3.5.1
Save Current Settings as a Group......................................... 50
3.5.2
Loading a Group Setting........................................................ 51
3.5.3
Deleting a Group Setting ....................................................... 52
3.5.4
Printing the System Settings for a Group Setting .................. 52

Operator Manual

2 of 110

Table of Contents - Model 3400

Operating Procedures
4.1
4.2
4.3
4.4
4.5

4.6

4.7

Password Protection
5.1
5.2
5.3
5.4
5.5

Logging In .............................................................................................. 53
Taking Counts........................................................................................ 55
Reviewing Buffer Records ..................................................................... 56
Clearing Buffer Records......................................................................... 57
Print Center............................................................................................ 58
4.5.1
Printing Automatically ............................................................ 60
4.5.2
Printing Buffer or Average of Count Cycles ........................... 62
4.5.3
Cancelling Print Jobs ............................................................. 63
Generating Reports................................................................................ 64
4.6.1
ISO 14644-1 Calculation Report............................................ 64
4.6.2
FS 209E Calculation Report .................................................. 64
4.6.3
BS 5295 Calculation Reports................................................. 64
Exporting Data ....................................................................................... 66

Password Overview ............................................................................... 71


Administrator Login ................................................................................ 71
Enabling Password Protection ............................................................... 72
Changing a Password............................................................................ 72
Replacing a Forgotten Password........................................................... 74

Maintenance and Diagnostics


6.1
6.2
6.3
6.4
6.5
6.6

Cleaning the Unit ................................................................................... 75


Running a Zero Count............................................................................ 75
Updating the 3400 Software .................................................................. 75
Checking Battery Status ........................................................................ 76
Diagnostics Screen Overview ................................................................ 77
6.5.1
System Status Printout .......................................................... 78
Factory Settings Screen......................................................................... 79
6.6.1
Counter Tab........................................................................... 81
6.6.2
Calibration Tab ...................................................................... 82
6.6.3
Test Tab................................................................................. 82

A.1
A.2

Return Procedures................................................................................. 85
Technical Support Information ............................................................... 85

Appendix B: Technical Data


B.1
B.2
B.3
Operator Manual

Performance Specifications ................................................................... 87


Accessories Included ............................................................................. 88
Recommended Spares Kit ..................................................................... 88

Met One

KSS - September 2005 - Edition 3

Appendix A: Service Procedures

Model 3400 - Table of Contents

B.4

3 of 110

Spare Parts List ..................................................................................... 89

Appendix C: Software Menu Trees


C.1

Menu Flow Diagram............................................................................... 91

Appendix D: Group Settings


D.1

Group Setting Forms.............................................................................. 93

Appendix E: Certifications
E.1

Overview ................................................................................................ 97

Appendix F: Sample Status Bit Mask Definitions


Overview .............................................................................................. 103

KSS - September 2005 - Edition 3

F.1

Met One

Operator Manual

Table of Contents - Model 3400

KSS - September 2005 - Edition 3

4 of 110

Operator Manual

Met One

Model 3400 - Manual Overview

5 of 110

Manual Overview
About This Manual
The information in this manual has been carefully checked and is believed to be accurate.
However, Hach Ultra Analytics assumes no responsibility for any inaccuracies that may be
contained in this manual. In no event will Hach Ultra Analytics be liable for direct, indirect,
special, incidental, or consequential damages resulting from any defect or omission in this
manual, even if advised of the possibility of such damages. In the interest of continued product
development, Hach Ultra Analytics reserves the right to make improvements in this manual and
the products it describes at any time, without notice or obligation.
Published in the United States of America
Hach Ultra Analytics P/N: 701241 Edition 3 September 2005
Copyright 2005 by Hach Ultra Analytics
All rights reserved. No part of the contents of this manual may be reproduced or transmitted in
any form or by any means without the written permission of Hach Ultra Analytics.

Safety Conventions
WARNING
A warning is used to indicate a condition which, if not met, could cause serious personal injury
and/or death. Do not move beyond a warning until all conditions have been met.

CAUTION:
A caution is used to indicate a condition which, if not met, could cause damage to the
equipment. Do not move beyond a caution until all conditions have been met.

Note:
A note is used to indicate important information or instructions that should be considered
before operating the equipment.

General Safety Considerations

KSS - September 2005 - Edition 3

Only factory certified personnel should perform service of the Model 3400.
Make sure the 3400 Series Particle Counter is properly installed and all connections are
correctly installed before operation. All safety guidelines should be observed.
Follow all procedures in Service Procedures on page 85 before shipping a unit to a
service center for repair or recalibration.

WARNING
Attempts by untrained personnel to disassemble, alter, modify or adjust the electronics may
result in personal injury and damage to the Model 3400.
Laser Safety Class 1 Laser Product complies with 21 CFR Chapter 1,
Subchapter J.

Met One

Operator Manual

6 of 110

Manual Overview - Model 3400

WARNING
Use of controls or adjustments, or performance of procedures other than those specified
herein may result in hazardous radiation exposure.

CAUTION:
Observe standard electrostatic discharge (ESD) procedures when servicing or installing
equipment to prevent damage.

Warranty
Hach Ultra Analytics warrants this instrument to be free of defects in materials and
workmanship for a period of two (2) years from the shipping date. If any instrument covered
under this warranty proves defective during this period, Hach Ultra Analytics will, at its option,
either repair the defective product without charge for parts and labor, or provide an equivalent
replacement in exchange for the defective product.
To obtain service under this warranty, the customer must notify the nearest Hach Ultra
Analytics service support center on or before the expiration of the warranty period and follow
their instructions for return of the defective instrument. The customer is responsible for all costs
associated with packaging and transporting the defective unit to the service support center, and
must prepay all shipping charges. Hach Ultra Analytics will pay for return shipping if the
shipment is to a location within the same country as the service support center.
This warranty shall not apply to any defect, failure, or damage caused by improper use or
maintenance or by inadequate maintenance or care. This warranty shall not apply to damage
resulting from attempts by personnel other than Hach Ultra Analytics representatives, or
factory-authorized and trained personnel, to install, repair or service the instrument; to damage
resulting from improper use or connection to incompatible equipment; or to instruments that
have been modified or integrated with other products when the effect of such modification or
integration materially increases the time or difficulty of servicing the instrument.
THIS WARRANTY IS GIVEN BY HACH ULTRA ANALYTICS WITH RESPECT TO THIS
INSTRUMENT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. HACH
ULTRA ANALYTICS AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR NON-CONTRACTUAL PURPOSE.
HACH ULTRA ANALYTICS RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE
PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER
FOR BREACH OF THIS WARRANTY. HACH ULTRA ANALYTICS AND ITS VENDORS WILL
NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES EVEN IF HACH ULTRA ANALYTICS OR ITS VENDORS HAS BEEN GIVEN
ADVANCED NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Edition 1, August 2005. Initial release.


Edition 2, August 2005. Incorporation of Certificate of Conformity.
Edition 3, September 2005. Incorporation of sample status bit mask definitions.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Revision History

Model 3400 - Manual Overview

7 of 110

Acknowledgements

KSS - September 2005 - Edition 3

Hytrel is a trademark of E. I. Du Pont De Nemours & Company Corporation.


Long Life Laser is a trademark or registered trademark of Hach Ultra Analytics.
Windows CE is a trademark of Microsoft Corporation.

Met One

Operator Manual

Manual Overview - Model 3400

KSS - September 2005 - Edition 3

8 of 110

Operator Manual

Met One

Model 3400 - Introduction

9 of 110

Introduction
This manual describes use of the 3400 Series Particle Counters, described in Table 1-1. Refer
to Performance Specifications on page 87 for a complete list of specifications.
Table 1-1 : 3400 Series Particle Counter Model Numbers
Flow Rate
Model Number

Minimum Particle Size Channel (m)


L/min

cubic ft/min

3413

28.3

0.3

3415

28.3

0.5

3423

50

1.77

0.3

3425

50

1.77

0.5

intake valve

touch screen
power button

KSS - September 2005 - Edition 3

battery indicator

Fig 1-1 : 3400 Series Particle Counter

Met One

Operator Manual

10 of 110

Introduction - Model 3400

Hach Ultra Analytics particle counters use a laser-diode light source and collection optics for
particle detection. Particles scatter light from the laser diode. The collection optics focus the
light onto a photo diode that converts the bursts of light into electrical pulses. The pulse height
is proportional to the particle size. Pulses are counted and their amplitude is measured for
particle sizing. Results are displayed as particle counts in the specified size range (differential
count mode) or as total particle counts (cumulative count mode). A microprocessor controls all
instrument functions. Count data is displayed as cumulative counts or differential counts.
The particle counter is thoroughly inspected and tested at the factory and is ready for use upon
receipt. When received, inspect the shipping carton for damage. If the carton is damaged,
notify the carrier and save the carton for carrier inspection. Inspect the counter for broken parts,
scratches, dents, or other damage.

1.1 Operation Conventions


The 3400 is operated by use of a TFT color touchscreen located on the front panel of the unit,
shown in Figure 1-1. All commands may be executed through the touchscreen.
Items on the screen that need to be touched or pressed to invoke actions appear in bold text.

A screen is a window of the Graphical User Interface (GUI) which takes up all visible
space on the touchscreen and can only be exited by pressing a button.
A tab is a subdivision of the GUI easily accessed by touching the tab at the top of the
screen.
An icon is an area of the touchscreen that should be touched to invoke another screen
or execute an action (such as performing a sample run).
A button is an area of the touchscreen that should be touched to invoke another screen
or execute an action (such as clearing the buffer).
A sliding menu is a group of icons hidden from view by default. The sliding menu
appears as a tab at the bottom of the Main screen. Touch the tab to show all the icons
and make a selection. Touching the tab again will hide the sliding menu.
A field indicates an area of the screen where strings of text and/or numbers may be
entered. To enter information in a field, touch the field. Depending on the nature of the
field, a keypad will appear on the touchscreen to allow text or numeric entries. To
complete entries from a keypad entry, touch OK.
A checkbox allows functions to be toggled on or off. Touch the box to make a checkmark
appear and enable the function.
A dropdown menu allows operators to choose from a set list of values. Dropdown menus
can be identified by the arrow on the right side of the field. To use a dropdown menu,
touch the arrow to show the list, drag a finger or stylus down the list to show all items,
then highlight the desired item to select it.
A pop-up box is a warning or note to make sure actions are taken as intended. Touch
OK to continue or Cancel to stop.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Screen items commonly referred to in this manual include:

Model 3400 - Introduction

11 of 110

1.2 Accessories
Several accessories are available to tailor the counters functioning. These accessories can be
ordered from a local Hach Ultra Analytics representative or from the factory by calling either a
local representative or the factory at 800.866.7889 or +1 541.472.6500 during the week from
8:00 a.m. to 5:00 p.m. PT.
Table 1-2 : 3400 Series Particle Counter Accessories

KSS - September 2005 - Edition 3

Accessory

Included/Optional?

Description

PortAll Version 2 Software, Demo Version

Included

PortAll Software controls the counter from a PC; downloads count data
into the PC; sorts, normalizes, calculates for cleanroom classification.
Call 800.866.7889 or +1 541.472.6500 for a free 30-day trial license or
to purchase a permanent license.

Stylus

Included

In cleanroom environments where operators wear protective gloves


that may inhibit movement, a stylus permits easier manipulation of the
touchscreen.

Power cord

Included

An AC-to-DC power supply and cord can be used in addition to battery


power.

Thermal printer paper

Included

Two rolls are included with the Model 3400.

Tubing

Included

This tubing allows connections from the probes to the counter.

Probes

Varies

Isokinetic probe

Included

The isokinetic probe is for use with unidirectional air flows to maximize
correlation between counts and actual particle-size distribution. This
can also be used for handheld spot checks.

RH/Temperature probe

Optional

The probe plugs into the counters left side and monitors relative
humidity and temperature. The results are displayed and can be
printed.

Air velocity probe

Optional

The air velocity probe plugs into the counters left side. The probe
monitors air velocity up to 200 feet/minute (1016 millimeters/second).
The results are displayed and can be printed.

Zero count filter

Included

The zero count filter attaches to sensor inlet and keeps external particles from contaminating the sensor while purging the sensor of internal
particles.

Carrying case

Optional

A carrying case protects the counter during shipment and storage.

Spare batteries

Optional1

Spare batteries allow maximum flexibility with portable use.

External battery charger

Optional

An external battery charger ensures any spare batteries are prepared


for use at any time.

1.

One battery is included standard with the 3400.

Met One

Operator Manual

Introduction - Model 3400

KSS - September 2005 - Edition 3

12 of 110

Operator Manual

Met One

Model 3400 - Initialization

13 of 110

Initialization
2.1 Electrical Connections
Connect probes, external power, cables, or USB devices to the Model 3400 to expand the
counters functionality or ease data downloads. Make connections where indicated in the
figures.
From the back of the 3400, shown in Figure 2-1, connect:
I/O port for filter scan
Two battery ports

KSS - September 2005 - Edition 3

supplemental feet

I/O port
for filter scan probe

battery ports

supplemental feet

Fig 2-1 : 3400 Back View

Met One

Operator Manual

14 of 110

Initialization - Model 3400

From the left side of the 3400, shown in Figure 2-2, connect:
DC connector for the power supply
Air velocity probe connector
RH/Temperature probe connector
All probes are plug-and-play and should be recognized by the 3400 as soon as they are
connected.

handle

air velocity probe RH/temp probe

KSS - September 2005 - Edition 3

power connector

Fig 2-2 : 3400 Left Side View

Operator Manual

Met One

Model 3400 - Initialization

15 of 110

From the front of the 3400, shown in Figure 2-3, connect:


USB Client
USB Host

printer

USB client USB host

handle

touchscreen

power button

battery status

Fig 2-3 : 3400 Front View

Note:
Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400;
contact Customer Service for additional support at 800.866.7889 or 541.472.6500.

KSS - September 2005 - Edition 3

2.2 Battery Charging


Batteries may be charged while within the 3400 or from an external battery charger, available
as an optional accessory. Refer to Technical Data on page 87 for ordering information.

2.2.1 Using the 3400


1)
2)
3)

Met One

To open the battery compartment, press down on the upper door latches or pull up on
the lower door latches.
Insert the batteries where shown in Figure 2-1. Do not force the batteries into place.
Attach the power supply to the unit where shown in Figure 2-2.
Operator Manual

16 of 110

Initialization - Model 3400

4)
5)

Connect the power supply to external power. The unit will automatically turn on when
charged batteries and/or AC power is connected.
Once the unit is on, charge the batteries until they are fully charged. The battery status
light indicates the level of power in the battery as described in Table 2-1.

Note:
The batteries cannot overcharge.
Batteries will charge when the AC power cord is connected. An external battery charger is also
available for maximum efficiency; refer to Technical Data on page 87 for ordering information
and contact a local Hach Ultra Analytics agent or Hach Ultra Analytics at 800.866.7889 or
541.472.6500 to order.
Batteries charged in the 3400 will take approximately 10 hours to charge completely.
Batteries charged in an external charger will take approximately 7 hours to charge
completely.
Battery status is indicated by the state of the LED on the front of the unit, as described in
Table 2-1.
Table 2-1 : Battery LED Color Indications
LED State

LED Color

Battery Status

Charging or Not Charging?

Flashing

Orange

Low power

Not charging

Flashing

Green

Low power

Charging

Solid

Green

Charged

Charging

Low Battery Alert


When the total available battery charge drops below 15% of total capacity, a warning will
appear and the system will begin beeping once every 5 seconds.
When the total available battery charge drops below 5%, the system will beep once a second
for 15 seconds before shutting down.

2.2.2 Using the External Battery Charger


Charge the batteries in the external battery charger per the manufacturers instructions.

2.3 Loading Printer Paper

CAUTION:
The printer should not be operated without paper as damage may occur to the print head. To
load paper in the printer, follow the instructions in Loading Printer Paper on page 16. If the
particle counter must be operated without paper in the printer, be sure to set the Print Mode to
None as described in Printing Automatically on page 60.

Operator Manual

Met One

KSS - September 2005 - Edition 3

The 3400 features an integrated printer located on the front of the unit next to the touchscreen
for easy printing of reports and records. Print procedures are described in detail in Print
Center on page 58.

Model 3400 - Initialization

17 of 110

CAUTION:
Paper used in this printer is temperature-sensitive on one side and must go into the printer as
explained in the instructions. Do not substitute other types of paper.
To load printer paper:
1)
2)
3)
4)

Locate the printer on the front of the unit, shown in Figure 2-3.
Pull up the green handle in the center of the paper cover until a click is heard.
Fold the door down and remove the cardboard tube from the previous roll, if present.
Load the new paper roll so the paper feeds the top of the roll. Position the end of the
paper on the black roller at the end of the printer feed door.
5) Push the printer door back until it clicks into place.
6) The printer is now ready to print. Print options include:
Buffer records, described in Printing Buffer or Average of Count Cycles on page 62
Regulatory reports, described in Generating Reports on page 64
Unit settings, described in System Status Printout on page 78

Note:
If the paper does not feed out or no image appears on the paper after a print command has
been sent, double-check paper direction.

2.4 Getting Started


The Met One Model 3400 counter can be used immediately after removing from the shipping
carton with AC power. To take a sample count before setting up the programmable functions,
proceed as follows:

WARNING
Do not attempt to sample reactive gases (such as hydrogen or oxygen) with the particle
counter. Reactive gases create an explosion hazard in the counter. Contact a local Hach Ultra
Analytics representative or the factory at 800.866.7889 or +1 541.472.6500 for more
information on product implementation.
1)

Locate counter in a clean environment.

Note:
The 3400 may be seated either on its bottom feet or its back feet, depending on the needs of
the operator.

CAUTION:
KSS - September 2005 - Edition 3

Always remove the protective cap from the inlet tube before performing counts or damage to
the counter will occur.
2)

Remove protective cap from the inlet tube (on top of counter).

WARNING
Exposure to infrared radiation may cause eye injury. Do not look into counter inlet tube while
counter is on.

Met One

Operator Manual

18 of 110

Initialization - Model 3400

3)

After shipping or storage, the counter batteries may need to be charged. Until the
batteries are completely charged, run the 3400 on AC power by connecting the AC-toDC power supply to the DC connector on the left side of the unit and plugging the unit
into AC power.

CAUTION:
Only use the HUA-furnished power supply with the 3400. Substituting a power supply may
damage the unit and will void the product warranty.
4)

The counter should turn on once AC power is connected. After a splash screen, the
Main screen displays, as shown in Figure 2-4.

KSS - September 2005 - Edition 3

Fig 2-4 : Main Screen

Operator Manual

Met One

Model 3400 - Initialization

19 of 110

2.5 Screen Orientation


2.5.1 Overview
The Main screen displays when power is applied to the counter, providing these indicators:
Real-time particle count information for each size range
Amount of time left to sample (if a count is running)
Number of samples left to take in succession (if set for multiple samples)
Environmental probe readings if probes are connected
Temperature

Air velocity

Relative humidity
Count status (Start, Run, Hold, or Stop)
Flow rate through the sensor

Volume sampled

Location name and number

Number of records in the buffer

Date

Time of day

Battery charge indicator

User messages

Count time remaining (0:00)

Count status
Particle counts

Environmental status panel


Flow rate
Volume sampled

Channel sizes

Temperature
Relative humidity
Air velocity
Location Name

Buffer record displayed


Show next record
Show previous record

Location ID
Last Location Number
Next Location Number

Date
Time
Sliding menu tab

Run icon
Battery status

KSS - September 2005 - Edition 3

Fig 2-5 : 3400 Main Screen

Met One

Operator Manual

20 of 110

Initialization - Model 3400

2.5.2 Sliding Menu


The icons in the sliding menu are shown in Figure 2-6. Touch the tab shown in Figure 2-5 to
invoke this menu.

Setup

Alarms Diagnostics Print Export Lock

Fig 2-6 : Sliding Menu Icons

2.5.3 Battery Status


Battery status is shown by the icon highlighted in Figure 2-5. Status is given for each individual
battery.
A maximum of two batteries may be loaded into the 3400 at one time.
If only one battery is loaded into the 3400, only one icon will appear on the Main screen.
States of battery are listed in Table 2-2 with icons.
Table 2-2 : Battery Status
State

Description

OK

Battery installed and is completely charged

Charging

AC power is present and the battery is being charged

Error

Not installed

Battery needs to be calibrated; contact Technical Support at

800.866.7889 or 541.472.6500

No batteries installed; the 3400 is running on AC power

Batteries will charge when the AC power cord is connected. An external battery charger is also
available for maximum efficiency; refer to Technical Data on page 87 for ordering information
and contact a local Hach Ultra Analytics agent or Hach Ultra Analytics at 800.866.7889 or
541.472.6500 to order.
Touch the Battery Status icon to invoke the Battery Diagnostics screen, shown in Figure 6-3
on page 76.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Icon

Model 3400 - Initialization

21 of 110

2.6 Quick Start: Taking a Count


This procedure is described in greater detail in Taking Counts on page 55.
1)

To start counting particles, press the Run icon, shown in Figure 2-7, on the
touchscreen.
Particle counting begins.

Fig 2-7 : Run Icon

The Run icon becomes a Stop icon, shown in Figure 2-8, until the count is
completed. Refer to Figure 2-9.
The display indicates the counts and any other parameters measured.

Fig 2-8 : Stop


Icon

Fig 2-9 : Main Screen: Sampling

Note:

KSS - September 2005 - Edition 3

Alarms are indicated by a red color change and flashing text at the bottom of the touchscreen,
as shown in Figure 2-10. An audible sound may sound if one has been set as described in
Sounds Tab on page 30; touch the screen to silence the alarm.

Met One

Operator Manual

22 of 110

Initialization - Model 3400

Fig 2-10 : Count Alarm Example

2)

KSS - September 2005 - Edition 3

Fig 2-11 : Stop


Icon

The count will stop automatically. To manually stop counts, press the Stop icon. When
the count is complete, the total number of particles counted during the count time
displays as well as the values for any other parameter measured (relative humidity,
temperature, air velocity).

Operator Manual

Met One

Model 3400 - Initialization

23 of 110

2.7 Main Screen Sliding Menu


Touching the tab in the lower left corner of the screen displays a sliding menu, shown in Figure
2-12, that allows access to the following functions:
Counter setup
Alarm settings
System diagnostics
Print and report functions
Data export functions
Lock: login/logoff functions

Fig 2-12 : Main Screen Sliding Menu

KSS - September 2005 - Edition 3

Refer to Setup Overview on page 25 to adjust the settings such as alarm limits, units of
measurement, time and date.

Met One

Operator Manual

Initialization - Model 3400

KSS - September 2005 - Edition 3

24 of 110

Operator Manual

Met One

Model 3400 - Configuration

25 of 110

Configuration
3.1 Setup Overview
System configuration is accessed via the Counter Setup screen, shown in Figure 3-2.
1)

From the sliding menu on the Main screen, touch the Setup icon, shown in Figure 3-1
to invoke the Counter Setup screen.

Setup icon
Fig 3-1 : Main Screen Sliding Menu, Setup Icon

KSS - September 2005 - Edition 3

2)

The Counter Setup displays as shown in Figure 3-2.

Fig 3-2 : Counter Setup Screen

Met One

Operator Manual

26 of 110

Configuration - Model 3400

3)

When each icon is pressed, it invokes a screen that allows configuration of a certain
aspect of the 3400.
This section describes each subsequent screen in detail; Table 3-1 briefly describes the
functions accessed by each icon.
Table 3-1 : Counter Setup Screen Icon Descriptions
Icon

Functions Accessed

Sample

Location ID
Count Cycles
Run Mode
Count Display

Relevant Section (s)

Count Mode
Sample Time
Hold Time
Delay Time

Units of Measurement
Clear Buffer
Count Limits
Concentration Limits
Environmental Limits

Alarm Setup on page 43


Count Tab Overview on page 45
Environmental Tab Overview on page 46

Alarms

Printer

Sample Screen on page 34


General Tab Overview on page 35
Timing Tab Overview on page 40
Environmental Units Tab on page 42

Print Mode
Print ISO 14644-1 Reports
Print FS209E Reports
Print BS5295 Reports
Print Average
Print Buffer
Export Data

Load Group
Add Group

Delete Group
Print Group

Generating Reports on page 64


Exporting Data on page 66

Group Settings on page 48

Group

Factory

Select language
Require logon
Edit User
Set Sounds

Global Settings on page 27


Setting the Language on page 33
Changing the Date or Time on page 29
Setting Sleep or Backlight Timeout on page 30
Setting Sounds on page 31
Setting Volume on page 32

Only printing information. These screens contain information useful for system diagnostics, including calibration expiration information. These screens are for factory use only. Refer to Factory Settings Screen
on page 79.

Returns to the Main screen

KSS - September 2005 - Edition 3

System

Time
Date
Sleep Time
Screen Timeout
Buffer size

n/a

Return

Operator Manual

Met One

Model 3400 - Configuration

27 of 110

3.2 Global Settings


1)

To access Global Settings, from the Main screen, invoke the sliding menu shown in
Figure 3-1 and select the Setup icon, shown in Figure 3-3.

2)

From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-4.
The Global Settings screen displays as shown in Figure 3-5.

Fig 3-3 : Setup


Icon

3)
Fig 3-4 : System
Icon

Fig 3-5 : Global Settings Screen

The Global Settings screen has four tabs:


Basic, discussed on page 28
Sounds, discussed on page 30
Users, discussed on page 32
Options and Accessories, discussed on page 32

KSS - September 2005 - Edition 3

Met One

Operator Manual

28 of 110

Configuration - Model 3400

3.2.1 Basic Tab Overview


The Basic tab allows users to set the time, date, sleep time, backlight timeout, as well as
establishing whether or not user logon is required for product use. The Basic tab is shown in
Figure 3-6.

Fig 3-6 : Basic Tab

On the Basic tab, set the parameters listed in Table 3-2.

Parameter

Description

Type of Entry

Valid Entries

Time

System time in HH:MM:SS format

Numeric

HH:MM:SS

Date

System date in YYYY/MM/DD format

Numeric

YYYY/MM/DD

Sleep Time

Period of inactivity in minutes before the


counter will go into sleep mode

Numeric

Minimum is 0 minutes
Setting to 0 disables feature
Maximum is 30 minutes

Backlight Time- Period of inactivity in seconds before the


out
touchscreens backlight will be disabled

Numeric

Minimum is 0 seconds
Setting of 0-4 disables feature
Maximum is 300 seconds

User Logon
Required

Checkbox

Only to be applied by an administrator;


refer to Enabling Password Protection
on page 72 for more information.

Operator Manual

Checkbox that enables password protection


when checked

Met One

KSS - September 2005 - Edition 3

Table 3-2 : Basic Tab Parameters

Model 3400 - Configuration

29 of 110

Changing the Date or Time

Fig 3-7 : System


Icon

1)
2)

From the Main screen, select the Setup icon from the sliding menu.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-7.

3)

The Global Settings screen displays. Select the Basic tab, shown in Figure 3-8.

Fig 3-8 : Global Settings, Basic Tab

4)
5)
6)
7)
8)
9)

Select the field of the parameter to modify (Date or Time).


A numeric entry screen appears.
Enter the time or date by touching the screen.
When the time or date is entered, touch the Enter button on the screen.
To return to the Counter Setup screen, press the Return icon, shown in Figure 3-9, on
the Global Settings tab.
Press the Return icon to return to the Main screen. The new time or date appears at
the bottom of the screen.

KSS - September 2005 - Edition 3

Fig 3-9 : Return


Icon

Met One

Operator Manual

30 of 110

Configuration - Model 3400

Setting Sleep or Backlight Timeout


1)
2)

Fig 3-10 : System


Icon

3)
4)
5)
6)

From the Main screen, select the Setup icon from the sliding menu.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-10.
The Global Settings screen displays. Select the Basic tab, shown in Figure 3-8.
Select the field of the parameter to modify (Sleep Time or Backlight Timeout).
A numeric entry screen appears.
Enter time by touching the screen.
Sleep Time is given in minutes (maximum 30 minutes)
Backlight Timeout is given in seconds (maximum 300 seconds)

7)
8)
9)

When the correct time is entered, touch the Enter button on the screen.
To return to the Counter Setup screen, press the Return icon, shown in Figure 3-11,
on the Basic tab.
To return to the Main screen, press the Return icon.

Fig 3-11 : Return


Icon

3.2.2 Sounds Tab


Use the Sounds tab to establish all sound settings, including choice of sounds and volume.
The Sounds tab is shown in Figure 3-12.

KSS - September 2005 - Edition 3

Fig 3-12 : Sounds Tab

On the Sounds tab, set the parameters listed in Table 3-3.

Operator Manual

Met One

Model 3400 - Configuration

31 of 110

Table 3-3 : Sound Tab Parameters


Parameter

Description

User Feedback Sound made when user invokes an action

Choices

Default
Infintr
Infbeg
Infend

Recend
Asterisk
Close
None

Stop Error

Sound made when sampling stops inadvertently

Close
Empty

Default
None

Warning

Sound made when a warning message


appears, such as a buffer full pop-up warning

Critical
Default
Question

Windmin
Windmax
None

Alarm Limit

Sound made when a count, concentration, or


environmental parameter exceeds a limit

Critical
Exclam
Infend

Windmin
Windmax

Volume

Sets volume level of all sounds

Slider

Setting Sounds
1)
2)

Fig 3-13 : System


Icon

3)
4)

From the Main screen, select the Setup icon from the sliding menu.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-13.
The Global Settings screen displays. Select the Sounds tab, shown in Figure 3-12.
To establish sounds, select the desired sound from the dropdown menus for the four
sound fields:

5)
6)
7)

User Feedback
Stop Error
Warning
Alarm Limit

Press the Test button to test the sound.


To return to the Counter Setup screen, press the Return icon, shown in Figure 3-14,
on the Sounds tab.
To return to the Main screen, press the Return icon.

KSS - September 2005 - Edition 3

Note:

Fig 3-14 : Return


Touch the screen to silence audible alarms once triggered.
Icon

Met One

Operator Manual

32 of 110

Configuration - Model 3400

Setting Volume
1)
2)

Fig 3-15 : System


Icon

3)
4)

From the Main screen, select the Setup icon from the sliding menu.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-13.
The Global Settings screen displays. Select the Sounds tab, shown in Figure 3-12.
Set the volume for the 3400 by dragging the Volume slider to the desired level.
Drag the slider to the right to increase volume.
Drag the slider to the left to decrease volume.

5)
6)
7)

Press the Test button to test the volume.


To return to the Counter Setup screen, press the Return icon, shown in Figure 3-16,
on the Sounds tab.
To return to the Main screen, press the Return icon.

Fig 3-16 : Return


Icon

3.2.3 Users Tab Overview


The Users tab is used by administrators to establish and maintain user accounts. Refer to
Password Protection on page 71 for detailed procedures.

3.2.4 Options and Accessories Tab

Fig 3-17 : Options and Accessories Tab

On the Options and Accessories tab, set the parameters listed in Table 3-4.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Use the Options and Accessories tab to establish the language for the 3400 as well as setting
up a filter scan probe. The I/O port shown in Electrical Connections on page 13 allows a filter
scan probe to be added. The Options and Accessories tab is shown in Figure 3-17.

Model 3400 - Configuration

33 of 110

Table 3-4 : Options and Accessories Tab Parameters


Parameter

Language

Data Buffer Size

Description
Language for all 3400 screens and
print-outs
Requires reboot of the 3400 to take
effect

Type of Entry

Valid Entries

Dropdown
menu

Set to add a filter scan probe


Maximum number of records that
can be held in the buffer
Once full, the buffer must be cleared Numeric
before future sampling can occur.
See Clearing Buffer Records on
page 57.

English
French
German
Spanish
Italian

50-5000

Setting the Language


1)
2)
3)

KSS - September 2005 - Edition 3

Fig 3-18 : System


Icon

4)

From the Main screen, select the Setup icon from the sliding menu.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-18.
The Global Settings screen displays. Select the Options and Accessories tab,
shown in Figure 3-17.
Highlight the Language field to select the desired language from the dropdown as
shown in Figure 3-19.

Fig 3-19 : Global Settings, Language Options

5)

Met One

When a new language is selected, the 3400 will prompt a restart, as shown in
Figure 3-20.

Operator Manual

34 of 110

Configuration - Model 3400

Fig 3-20 : Restart Prompt

6)

7)

Press OK to continue and restart the 3400; press X to cancel and restart the 3400
manually later. To return to the Counter Setup screen, press the Return icon. To return
to the Main screen, press the Return icon.
Once the 3400 restarts, all screens should appear in the selected language.

3.3 Sample Screen


1)

To access Sample Setup, from the Main screen, invoke the sliding menu shown in
Figure 3-1 and select the Setup icon, shown in Figure 3-21.

2)

From the Counter Setup screen shown in Figure 3-2, select the Sample icon, shown
in Figure 3-22.

3)

The Sample Setup screen appears as shown in Figure 3-23.The Sample Setup
screen has three tabs:

Fig 3-21 : Setup


Icon

Fig 3-22 : Sample


Icon

KSS - September 2005 - Edition 3

General, discussed on page 35


Timing, discussed on page 40
Environmental Units, discussed on page 42

Operator Manual

Met One

Model 3400 - Configuration

35 of 110

3.3.1 General Tab Overview

Fig 3-23 : General Tab

On the General tab, set the parameters listed in Table 3-5. The General tab is shown in
Figure 3-23.
Table 3-5 : Sample Tab Parameters
Parameter

Description

Type of Entry

Valid Entries

Location ID

Numeric value for the location

Field

Numeric entry via keypad screen or


by increment/decrement icons
0 through 999

Location Name Text string for the location

Field

Alphanumeric entry via keyboard


screen

Count Cycles

Numeric

Minimum 0, maximum 100


0 indicates cycles until buffer is full

Number of cycles to run before stopping


Auto: automatic start
Manual: manual start

KSS - September 2005 - Edition 3

Run Mode

Count Display

Met One

dropdown menu N/A


Beep: beeps once the count limit is
reached and again for each multiple of the
count limit
Cumulative: counts particles for the
selected size range and all larger size
ranges
Differential: counts particles from the
selected size up to the next size range

dropdown menu N/A

Operator Manual

36 of 110

Configuration - Model 3400

Table 3-5 : Sample Tab Parameters


Parameter

Description

Type of Entry

Count Mode

dropdown menu N/A

Data Buffer

Clear Buffer

Counts
Counts/Cu. Ft.
Counts/L
Counts/Cu. M

Indicates number of records in the buffer

Deletes all records from the buffer

Valid Entries

N/A

Maximum number of records is 5000


Error messages appear when the
buffer reaches 4980 records to indicate that the buffer is approaching
maximum capacity and records
should be cleared.

N/A

Buffer records are not automatically


deleted on download
Once full, the buffer must be cleared
before sampling can occur.

Changing the Location ID


1)

From the Main screen, either press the increment or decrement icons or touch the
Location ID field to change the number.

Fig 3-24 : Increment and Decrement Icons

If the Location ID field was touched, a numeric entry screen appears, as shown in
Figure 3-25.

KSS - September 2005 - Edition 3

2)

Operator Manual

Met One

Model 3400 - Configuration

37 of 110

Fig 3-25 : Numeric Entry Screen

3)
4)
5)
6)

Enter a number for the location by touching the screen.


When the value is entered, touch the Enter button on the screen.
To return to the Counter Setup screen, press the Return icon, shown in Figure 3-26,
on the General tab.
Press the Return icon to return to the Main screen. The location number appears on
the screen where indicated in Figure 3-27.

Fig 3-26 : Return


Icon

KSS - September 2005 - Edition 3

Location ID

Fig 3-27 : Main Screen with Location ID Highlighted

Met One

Operator Manual

38 of 110

Configuration - Model 3400

Entering the Location Name


The Location Name is a text string which appears on the Main screen to help operators
identify the location.
1)
2)
3)
4)
5)

From the Main screen, touch the Location Name, shown in Figure 3-28.
A keyboard entry screen appears.
Enter a name for the location by touching the screen.
When the value is entered, touch the Enter button on the screen.
Press the Return icon to return to the Main screen. The location name appears on the
screen where indicated in Figure 3-28.

Location name

Fig 3-28 : Main Screen with Location Name Highlighted

Changing the Count Cycles


The number of count cycles is the number of times the counter will count then hold before
stopping. The maximum number of cycles is 100. This setting only affects the counter when
Run Mode is set to Auto.
1)
2)
3)
4)

From the Main screen sliding menu, touch the Setup icon > Sample icon to invoke the
Sample Setup screen, shown in Figure 3-23.
From the General tab, touch the Count Cycles field.
A numeric entry screen appears, as shown in Figure 3-25.
Enter a number between 0 and 100 by touching the screen.

Entering 0 causes the 3400 to count until the buffer is full.


5)

Operator Manual

When the count cycle value is entered, touch the Enter button on the screen.

Met One

KSS - September 2005 - Edition 3

Note:

Model 3400 - Configuration

6)
7)
Fig 3-29 : Return
Icon

To return to the Counter Setup screen, press the Return icon, shown in Figure 3-29,
on the General tab.
To return to the Main screen, press the Return icon.

Changing the Run Mode


1)

Fig 3-30 : Sample


Icon

39 of 110

2)
3)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 3-30, to invoke the Sample Setup screen, shown in Figure 3-23.
From the General tab, touch the Run Mode field to invoke a dropdown menu.
A dropdown menu appears with these choices:
Auto: counts, then holds for the set hold time, then repeats the count-hold cycle
for the value entered in Count Cycle, then stops sampling
Manual: counts for one cycle then stops sampling
Beep: beeps once the count limit is reached and again for each multiple of the
count limit

4)
5)
6)

Touch the screen to highlight the desired run mode.


To return to the Counter Setup screen, press the Return icon, shown in Figure 3-31,
on the General tab.
To return to the Main screen, press the Return icon.

Changing the Count Display


1)
2)
3)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 3-30, to invoke the Sample Setup screen, shown in Figure 3-23.
From the General tab, touch the Count Display dropdown menu.
A dropdown menu appears with these choices:
Cumulative: counts particles for the selected size range and all larger size
ranges
Differential: counts particles from the selected size up to the next size range

4)
5)
6)

Touch the screen to highlight the desired count display.


To return to the Counter Setup screen, press the Return icon, shown in Figure 3-31,
on the General tab.
To return to the Main screen, press the Return icon.

KSS - September 2005 - Edition 3

Fig 3-31 : Return


Icon

Met One

Operator Manual

40 of 110

Configuration - Model 3400

Changing the Count Mode


1)

Fig 3-32 : Sample


Icon

2)
3)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 3-32, to invoke the Sample Setup screen, shown in Figure 3-23.
From the General tab, touch the Count Mode dropdown menu.
A dropdown menu appears with these choices:

4)
5)
6)

Counts
Counts/Cu. Ft.
Counts/L
Counts/Cu. M

Touch the screen to highlight the desired count mode


Press the Return icon, shown in Figure 3-33, on the General tab to return to the
Counter Setup screen.
Press the Return icon to return to the Main screen. Counts will now appear in the
selected mode.

Fig 3-33 : Return


Icon

3.3.2 Timing Tab Overview

On the Timing tab, set the parameters listed in Table 3-6. The Timing tab is shown in
Figure 3-34.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Fig 3-34 : Timing Tab

Model 3400 - Configuration

41 of 110

Table 3-6 : Timing Tab Parameters


Parameter

Description

Sample Time

Length of time to sample air.


In manual mode, counter counts for the
sample time, then stops.
In automatic mode, the sample and hold
time repeat for the number of cycles programmed under count cycles.

Hold Time

Delay Time

Type of Entry

Valid Entries

HH:MM:SS format
Default is 00:01:00
Minimum 00:00:01
Maximum 23:59:59

Numeric

HH:MM:SS format
Default is 00:00:00
Minimum 00:00:00
Maximum 23:59:59

Delays the start of the count cycle after Start


Numeric
is pressed (from a complete Stop).

HH:MM:SS format
Default is 00:00:00
Minimum: 00:00:00
Maximum 23:59:59

Length of time between samples.

Numeric

Setting Sample, Hold, or Delay Times


1)

Fig 3-35 : Sample


Icon

2)
3)
4)
5)
6)
7)
8)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 3-35, to invoke the Sample Setup screen, shown in Figure 3-23.
Touch the Timing tab to display the screen shown in Figure 3-34.
Select the field of the parameter to modify (Sample Time, Hold Time, or Delay Time).
A numeric entry screen appears, as shown in Figure 3-25.
Enter time in HH:MM:SS format by touching the screen.
When the correct time is entered, touch the Enter button on the screen.
To return to the Counter Setup screen, press the Return icon, shown in Figure 3-36,
on the Timing tab.
To return to the Main screen, press the Return icon.

KSS - September 2005 - Edition 3

Fig 3-36 : Return


Icon

Met One

Operator Manual

42 of 110

Configuration - Model 3400

3.3.3 Environmental Units Tab

Fig 3-37 : Environmental Units Tab

On the Environmental Units tab, set the parameters listed in Table 3-7. The Environmental
Units tab is shown in Figure 3-37. All parameters are set using dropdown menus.
Table 3-7 : Environmental Units Tab Parameters
Parameter

Description

Type of Entry

Temperature

Deg C
Deg F

Dropdown

Air Velocity

mm/Sec
ft/Min

Dropdown

Flow Rate

LPM
CFM

Dropdown

1)

Fig 3-38 : Sample


Icon

Operator Manual

2)
3)
4)
5)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 3-38, to invoke the Sample Setup screen, shown in Figure 3-23.
Touch the Environmental Units tab to invoke the screen shown in Figure 3-37.
Select the dropdown menu of the parameter to modify (Temperature, Air Velocity, or
Flow Rate).
A dropdown menu appears, listing the choices available.
Highlight a choice to select it.

Met One

KSS - September 2005 - Edition 3

Setting Units of Measurement

Model 3400 - Configuration

6)
7)

43 of 110

To return to the Counter Setup screen, press the Return icon, shown in Figure 3-39,
on the Environmental Units tab.
To return to the Main screen, press the Return icon.

Fig 3-39 : Return


Icon

3.4 Alarm Setup


Count, concentration, or environmental probe alarm limits may be set or changed from these
screens. Alarms consist of a flashing notice on the screen as well as an audible sound as soon
as any alarm limit is exceeded. Touch the screen to silence an alarm.
To access alarm setup:
1)

From the sliding menu on the Main screen, touch the Alarms icon, shown in
Figure 3-40 to invoke the Alarm Settings screen.

Alarms icon

KSS - September 2005 - Edition 3

Fig 3-40 : Main Screen Sliding Menu, Alarm Icon

Met One

Operator Manual

44 of 110

Configuration - Model 3400

Fig 3-41 : Alarm Settings Screen

2)

On the Alarm Settings screen, shown in Figure 3-41, alarm functions are organized
into two tabs:
Count, discussed on page 45
Environmental, discussed on page 46

Note:
The Alarm Settings screen may also be accessed from the Alarms icon, shown in
Figure 3-42, on the Counter Setup screen (refer to Figure 3-2 on page 25).

KSS - September 2005 - Edition 3

Fig 3-42 : Alarms


Icon

Operator Manual

Met One

Model 3400 - Configuration

45 of 110

3.4.1 Count Tab Overview


Alarms may be set to occur when the count or concentration exceeds a specific value. Ranges
may be set from zero (no alarm set) up to 9,999,999. Each particle size range has a different
alarm limit. Alarms activate as soon as a limit is exceeded. Touch the screen to silence an
audible alarm.

Fig 3-43 : Count Alarm Tab

Setting Count or Concentration Alarms


1)
2)
Fig 3-44 : Alarms
Icon

3)

From the sliding menu on the Main screen, touch the Alarms icon, shown in
Figure 3-44, to invoke the Alarm Settings screen.
The Count tab, shown in Figure 3-43, displays both count and concentration limits by
particle size.
Select the particle size limit to modify by touching the existing limit.
Select a limit from the Count Limit column to modify alarm count limits.
Select a limit from the Conc. Limit column to modify alarm concentration limits.

4)
5)
6)

Press Edit.
A numeric entry screen appears, as shown in Figure 3-25 on page 37.
Enter the number by touching the screen.

Note:
KSS - September 2005 - Edition 3

The maximum limit for any particle size is 9,999,999.


7)

Met One

When the correct limit is entered, touch the Enter button on the screen.

Operator Manual

46 of 110

Configuration - Model 3400

8)
9)

To return to the Counter Setup screen, press the Return icon, shown in Figure 3-45,
on the Count tab.
To return to the Main screen, press the Return icon.

Fig 3-45 : Return


Icon

When alarms are triggered, the screen will flash red, the out of range value will be highlighted
on the Main screen, and an Alarm! will appear at the bottom of the screen, as shown in the
example in Figure 3-46. If an audible alarm has been set as described in Sounds Tab on
page 30, a sound will be heard. Touch the screen to silence the alarm.

Red screen

Out of range value

Alarm! message

Fig 3-46 : Count Alarm Example

3.4.2 Environmental Tab Overview


From the sliding menu on the Main screen, touch the Alarms icon, shown in
Figure 3-47, to invoke the Alarm Settings screen.
2) Touch the Environmental Alarm tab to invoke the screen shown in Figure 3-48.
From the Environmental Alarm tab, shown in Figure 3-48, set a count alarm to occur when
Fig 3-47 : Alarms
environmental parameters exceed a given range. Audible alarms may be set as described in
Icon
Sounds Tab on page 30; touch the screen to silence an alarm. Environmental alarms only
activate when the 3400 is sampling.

KSS - September 2005 - Edition 3

1)

Operator Manual

Met One

Model 3400 - Configuration

47 of 110

Fig 3-48 : Environmental Alarm Tab

Setting Environmental Alarms


CAUTION:
Be sure that the appropriate probe is installed before attempting to monitor an environmental
parameter.
1)
2)
Fig 3-49 : Alarms
Icon

The High value indicates the maximum acceptable range of the measurement.
If this alarm is enabled, the alarm will activate when the measurement is higher
than this value.
The Low value indicates the minimum acceptable range of the measurement. If
this alarm is enabled, the alarm will activate when the measurement drops
below this value.

KSS - September 2005 - Edition 3

3)
4)
5)
6)
7)
8)
9)
10)
11)

Met One

From the sliding menu on the Main screen, touch the Alarms icon, shown in
Figure 3-49, to invoke the Alarm Settings screen.
Touch the Environmental Alarm tab to invoke the screen shown in Figure 3-48. The
Environmental Alarm tab displays both high and low values for each of the possible
environmental measurements.

Select the High measurement limit to modify by touching the existing limit.
A numeric entry screen appears, similar to the one shown in Figure 3-25.
Enter the number by touching the screen.
When the correct limit is entered, touch the Enter button on the screen.
Select the Low measurement limit to modify by touching the existing limit.
A numeric entry screen appears, similar to the one shown in Figure 3-25.
Enter the number by touching the screen.
When the correct limit is entered, touch the Enter button on the screen.
Enable the alarm by touching the Enabled checkbox next to the limit.

Operator Manual

48 of 110

Configuration - Model 3400

12)

Fig 3-50 : Return


Icon

13)
14)

To return to the Counter Setup screen, press the Return icon, shown in Figure 3-50,
on the Environmental Alarm tab.
Press the Return icon to return to the Main screen.
When alarms are triggered, the screen will flash red, the out of range value will be
highlighted on the Main screen, and an Alarm! will appear at the bottom of the screen,
as shown in Figure 3-51. If an audible alarm has been set as described in Sounds Tab
on page 30, a sound will be heard. Touch the screen to silence the alarm.

Red screen

Out of range value

Alarm! message

Fig 3-51 : Environmental Alarm Example

3.5 Group Settings


A unique set of parameters can be saved in memory as a group setting, then loaded to the
counter for use by selecting the correct group via the Group Setting Management screen.
This may be useful when the same settings are used to monitor a location, number of rooms
repeatedly. As many as 20 groups may be stored by the 3400.
Table 3-8 lists the parameters controlled by group settings and the screens where those
parameters are modified.

Operator Manual

Parameter

Screen Access Name

Tab Name

Air Velocity

Sample Setup

Environmental Units

Alarm Concentration Limit


(per channel)

Alarm Settings

Count

Alarm Count Limit (per channel)

Alarm Settings

Count

Alarm: Air Velocity

Alarm Settings

Environmental

Alarm: Flow Rate

Alarm Settings

Environmental

Alarm: Relative Humidity Low/High Alarm Settings

Environmental

Met One

KSS - September 2005 - Edition 3

Table 3-8 : Group Setting Parameters

Model 3400 - Configuration

49 of 110

Table 3-8 : Group Setting Parameters


Parameter

Screen Access Name

Tab Name

Alarm: Temperature Low/High

Alarm Settings

Environmental

Count Cycles

Sample Setup

General

Count Display

Sample Setup

General

Count Mode

Sample Setup

General

Delay Time

Sample Setup

Timing

Flow Rate

Sample Setup

Environmental Units

Hold Time

Sample Setup

Timing

Location ID

Sample Setup

General

Print Cycles

Print Center

n/a

Run Mode

Sample Setup

General

Sample Print Mode

Print Center

n/a

Sample Time

Sample Setup

Timing

Temperature

Sample Setup

Environmental Units

Refer to Group Settings on page 93 for forms that may be copied to keep track of group
setting parameters.

Note:
The Default setting may be used to return the counter to its factory configuration. The default
setting may not be deleted or modified.
1)

To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-52.

2)

From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon, shown in Figure 3-53.
The Group Setting Management screen displays as shown in Figure 3-54.

Fig 3-52 : Setup


Icon

KSS - September 2005 - Edition 3

3)
Fig 3-53 : Group
Icon

Met One

Operator Manual

50 of 110

Configuration - Model 3400

Fig 3-54 : Group Setting Management Screen

There are four functions from the Group Setting Management screen:

Saving current settings as a group, discussed on page 50


Loading a group setting, discussed on page 51
Deleting a group setting, discussed on page 52
Printing the last run from a group setting, discussed on page 52

3.5.1 Save Current Settings as a Group


If settings are not saved as a group setting before power to the 3400 is turned off, the 3400 will
restore the last loaded group setting on reboot and the last samples settings will be lost.
Unsaved sample settings are temporarily saved when the 3400 enters sleep mode; these
settings remain when the 3400 resumes function after sleep mode.
Before beginning this procedure, make sure all parameters for the counter are set as desired.

Note:

1)
2)
Fig 3-55 : Setup
Icon

Operator Manual

3)
4)

To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 on page 25 and select the Setup icon, shown in Figure 3-55.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon.
The Group Setting Management screen displays as shown in Figure 3-54.
In the Defined Groups list, highlight the <New> item by touching it. A keyboard entry
appears, as shown in Figure 3-56.

Met One

KSS - September 2005 - Edition 3

To make data collection easier, a form which may be photocopied and used to record group
settings appears in Group Settings on page 93.

Model 3400 - Configuration

51 of 110

Fig 3-56 : New Group Keyboard Entry Screen

5)

Fig 3-57 : Return


Icon

6)
7)
8)

Enter a new group name by touching the keyboard, then press Enter to return to the
Group Setting Management screen.
Press Save to save the 3400s current settings as the new groups settings.
To return to the Counter Setup screen, press the Return icon shown in Figure 3-57.
To return to the Main screen, press the Return icon.

3.5.2 Loading a Group Setting


1)

KSS - September 2005 - Edition 3

Fig 3-58 : Group


Icon

To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-55.
2) From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon, shown in Figure 3-58.
3) The Group Setting Management screen displays as shown in Figure 3-54.
4) Highlight a group then press Load to save the group as the active group.
5) To return to the Counter Setup screen, press the Return icon on the Group Setting
Management screen.
6) To return to the Main screen, press the Return icon.
This group is now the active group.

Met One

Operator Manual

52 of 110

Configuration - Model 3400

3.5.3 Deleting a Group Setting


Note:
A group may not be deleted if it is the either the default group or active group; see Figure 3-60.
1)
2)
Fig 3-59 : Setup
Icon

3)
4)
5)
6)

To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-59.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon, shown in Figure 3-61.
The Group Setting Management screen displays as shown in Figure 3-54.
Highlight a group then press Delete to delete the group.
To return to the Counter Setup screen, press the Return icon on the Group Setting
Management screen.
To return to the Main screen, press the Return icon.

Fig 3-60 : Delete Group Error Messages

1)
2)

Fig 3-61 : Group


Icon

Operator Manual

3)
4)
5)
6)

To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-59.
From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon, shown in Figure 3-61.
The Group Setting Management screen displays as shown in Figure 3-54.
Highlight a group, then press Print to print the system settings for that group.
To return to the Counter Setup screen, press the Return icon on the Sounds tab.
To return to the Main screen, press the Return icon.

Met One

KSS - September 2005 - Edition 3

3.5.4 Printing the System Settings for a Group Setting

Model 3400 - Operating Procedures

53 of 110

Operating Procedures
4.1 Logging In
If password protection has been enabled, login to use the 3400. If password protection is not
enabled, no login is required.
Instrument access can be shown by the Lock icon on the Main screen, as described in
Table 4-1.
Table 4-1 : Lock Icon Status

Indicates that the instrument is unlocked. When password protection is enabled, touching
this icon invokes a login screen.

Indicates that the instrument is locked. When password protection is enabled, touching this
icon logs the previous user out.

Note:
For information on enabling password protection, see Enabling Password Protection on
page 72.
1)

There are two methods to login, depending on whether the 3400 has first been turned
on or if another user is already logged in:
To log in after the Model 3400 is initialized, touch the screen. When password
protection is enabled, users must log in to proceed.
To login to the unit with another user already logged in, touch the Lock icon to
log out the previous user, then touch the Lock icon again to login.
A login screen appears like the one shown in Figure 4-1. Touch the Name field to enter
the user name and a keyboard entry screen appears.

KSS - September 2005 - Edition 3

2)

Met One

Operator Manual

54 of 110

Operating Procedures - Model 3400

Fig 4-1 : Login Screen

3)
4)
5)

Enter the user name by touching the screen.


When the name is entered, touch the Enter button on the screen.
Touch the Password field to invoke the keyboard screen and enter the password via
the touchscreen shown in Figure 4-2.

6)
7)

Operator Manual

Press Enter.
If login is successful, the Main screen should appear, similar to the one shown in
Figure 4-6. If login is unsuccessful, a Login Failed message appears and the
password should be re-entered.

Met One

KSS - September 2005 - Edition 3

Fig 4-2 : Keyboard Entry Screen

Model 3400 - Operating Procedures

55 of 110

4.2 Taking Counts


CAUTION:
Always remove the protective cap from the inlet tube before performing counts or damage to
the counter will occur.
1)

If still in place, remove protective cap from the inlet tube (on top of counter).

WARNING
Exposure to infrared radiation may cause eye injury. Do not look into counter inlet tube while
counter is on.
2)

Make sure the Main screen displays, as shown in Figure 4-3.

Fig 4-3 : Main Screen

Table 4-2 describes the function of the buttons on the Main screen.
Table 4-2 : Main screen Button Descriptions

KSS - September 2005 - Edition 3

Icon

Met One

Icon Name

Description

Run

Starts counts; changes to Stop button when count is in progress

Stop

Stops counts; changes to Run once the count is finished

Plus

Increments either the location or buffer number

Operator Manual

56 of 110

Operating Procedures - Model 3400

Table 4-2 : Main screen Button Descriptions


Icon

Icon Name

Description

Minus

Decrements either the location or buffer number

3)

To start counting particles, press the Run icon, shown in Figure 4-4, on the
touchscreen.
Particle counting begins.
The Run icon becomes a Stop icon, shown in Figure 4-5, until the count is
completed.
The display indicates the counts and any other parameters measured.

Fig 4-4 : Run Icon

Note:
Fig 4-5 : Stop
Icon

Count or environmental alarms will be indicated by a red color and flashing text at the bottom
of the touchscreen, as well as an audible alarm if one was set as described in Sounds Tab on
page 30.
4)

The count stops automatically. To manually stop counts, press the Stop icon. When the
count is complete, the total number of particles counted during the count time displays
as well as the values for any other parameter measured (relative humidity, temperature,
air velocity).
Refer to Configuration on page 25 to adjust the settings, such as alarm limits, units of
measurement, and time and date.

4.3 Reviewing Buffer Records


To view a specific buffer record:
From the Main screen, touch the Sample Buffer field, shown in Figure 4-6.

Buffer record displayed

Fig 4-6 : 3400 Main Screen, Sample Buffer Field


Operator Manual

Met One

KSS - September 2005 - Edition 3

1)

Model 3400 - Operating Procedures

2)
3)
4)

57 of 110

A numeric entry screen appears. Enter the buffer record number by touching the
screen.
When the value is entered, touch the Enter button on the screen.
The buffer record information appears on the Main screen.

4.4 Clearing Buffer Records


The 3400 stores a maximum of 5000 data records in memory. Records are not automatically
deleted from the buffer when data is downloaded. When the buffer reaches 4980 records, a
warning message appears on the Main screen.

CAUTION:
When the buffer is full, the buffer must be cleared before further sampling can occur. No
sampling can occur when the buffer is full.
To clear the buffer:
1)
2)

From the Main screen sliding menu, touch the Setup icon > Sample icon, shown in
Figure 4-7 and Figure 4-8, to invoke the Sample Setup screen, shown in Figure 4-9.
From the General tab, touch the Clear Buffer button.

Fig 4-7 : Setup


Icon

Fig 4-8 : Sample


Icon

KSS - September 2005 - Edition 3

Fig 4-9 : Sample Setup Screen

3)

Met One

A pop-up box will ask for confirmation to clear the buffer, as shown in Figure 4-10.

Operator Manual

58 of 110

Operating Procedures - Model 3400

Fig 4-10 : Clear Sample Buffer Pop-Up Box

CAUTION:
Once the buffer is cleared, the records are erased and cannot be retrieved.
4)
5)

Touch Yes to continue.


The buffer is cleared and the number of records returns to zero.

4.5 Print Center


The 3400 includes a built-in printer. All print functions are available through the Print icon
shown in Figure 4-11.

Note:
Fig 4-11 : Print
Icon

The Print icon is accessible from both the sliding menu shown in Figure 2-12 on page 23 and
the Counter Setup screen shown in Figure 3-2 on page 25.
The Print icon invokes the Print Center screen, shown in Figure 4-12, which allows access to
the following functions:

Operator Manual

KSS - September 2005 - Edition 3

Setting up automatic print functions


Printing regulatory reports, including:
ISO 14644-1
FS 209E
BS5295
Printing buffer records or count averages

Met One

Model 3400 - Operating Procedures

59 of 110

Fig 4-12 : Print Center Screen

CAUTION:
The printer should not be operated without paper as damage may occur to the print head. To
load paper in the printer, follow the instructions in Loading Printer Paper on page 16. If the
particle counter must be operated without paper in the printer, be sure to set the Print Mode to
None as described in Printing Automatically on page 60.

CAUTION:
Paper used in this printer is temperature-sensitive on one side and must go into the printer as
explained in the instructions.
Table 4-3 describes the function of the fields and buttons on the Print Center screen.
Table 4-3 : Print Center Functions
Icon

N/A

Icon/Field Name

Sample Print Mode

Description
This dropdown menu allows records to automatically print under the following
conditions:
Cycles: prints results of the first count cycle and multiples of the count cycle programmed.
None: does not print
Alarms: prints results only when any alarm limit has been exceeded.

KSS - September 2005 - Edition 3

All: prints results after completion of each count cycle.


N/A

Cycles

Lists the number of cycles to print

ISO 14644-1

Generates ISO 14644-1 calculations and prints report

ISO 14644-1

Met One

Operator Manual

60 of 110

Operating Procedures - Model 3400

Table 4-3 : Print Center Functions


Icon

Icon/Field Name

Description

FS209E

Generates FS 209E calculations and prints report

BS5295

Generates BS5295 calculations and prints report

Average

Prints an average of all count cycles in each particle size range

Buffer

Prints the complete contents of the buffer


Note: The buffer holds a maximum of 5000 records. Printing the buffer can be a timeconsuming process if the buffer is approaching maximum capacity.

Return

Closes the Print Center screen

Cancel Print

Cancels the current print job

FS209E

BS5295

Average

Buffer

4.5.1 Printing Automatically


From the sliding menu on the Main screen, touch the Print icon shown in Figure 4-11
to invoke the Print Center screen.

KSS - September 2005 - Edition 3

1)

Operator Manual

Met One

Model 3400 - Operating Procedures

61 of 110

Print icon

KSS - September 2005 - Edition 3

Fig 4-13 : Main Screen Sliding Menu, Print Icon

2)

The Print Center displays as shown in Figure 4-12.

3)

Touch the Print Mode dropdown menu to make a selection. Highlight one of the
following choices, shown in Figure 4-14:

Fig 4-14 : Print Mode Choices

Cycles: prints results of the first count cycle and multiples of the count cycle programmed.
None: no counts print automatically
Alarms: prints results only when a count alarm limit has been exceeded.
All: prints results after completion of each count cycle.

Met One

Operator Manual

62 of 110

Operating Procedures - Model 3400

Note:
Printing or exporting data from the buffer does not clear the buffer.
4)

To return to the Main Screen, touch the Return icon. A sample printout is shown in
Figure 4-16
.

**** ALARM CONDITION ****


Fig 4-15 : Return
Icon

S/N ##########

LOCATION ###

DATE YYYY-MM-DD
CYCLES ###

TIME HH:MM:SS

FLOWRATE ##.#LPM

VOLUME #.#FT^3 PERIOD HH:MM:SS


Fig 4-16 : Sample Data Printout

4.5.2 Printing Buffer or Average of Count Cycles


To print records manually:
1)

Fig 4-17 : Print


Icon

2)
3)

From the sliding menu on the Main screen, touch the Print icon shown in Figure 4-17
to invoke the Print Center screen.
The Print Center displays as shown in Figure 4-12.
Press the icon of the item to print:
Average prints the average of each size channel for the last sample run

Average

Buffer prints the contents of the buffer


Buffer

Note:
Printing or exporting data from the buffer does not clear the buffer.
4)

The data prints. Example printouts are shown in Figure 4-18 and Figure 4-19.

Note:
Printing or exporting data from the buffer does not clear the buffer.
To return to the Main Screen, touch the Return icon.

KSS - September 2005 - Edition 3

5)

Operator Manual

Met One

Model 3400 - Operating Procedures

63 of 110

---PRINT BUFFER, #### RECORDS------

PRINT AVERAGES

----

**** ALARM CONDITION ****


S/N ##########

LOCATION ###

DATE YYYY-MM-DD
CYCLES ###

TIME HH:MM:SS

FLOWRATE ##.#LPM

S/N ##########

LOCATION ###

DATE YYYY-MM-DD
CYCLES ###

TIME HH:MM:SS

FLOWRATE ##.#LPM

VOLUME #.#FT^3 PERIOD HH:MM:SS

PERIOD HH:MM:SS

COUNT SCALE:PARTICLES

COUNT SCALE:PARTICLES/CUBIC FT

TEMPERATURE ###.#F

TEMPERATURE ###.#F

RH ###.#%

AIR VELOCITY #.#FT/MIN

AIR VELOCITY #.#FT/MIN

SIZE

SIZE

CUMULATIVE DIFFERENTIAL

RH ###.#%

CUMULATIVE DIFFERENTIAL

0.3m 12345678.9 12345678.9

0.3m 12345678.9 12345678.9

0.5m 12345678.9 12345678.9

0.5m 12345678.9 12345678.9

1.0m 12345678.9 12345678.9

1.0m 12345678.9 12345678.9

3.0m 12345678.9 12345678.9

3.0m 12345678.9 12345678.9

5.0m 12345678.9 12345678.9

5.0m 12345678.9 12345678.9

10.0m 12345678.9 12345678.9


Fig 4-18 : Sample Buffer Printout

10.0m 12345678.9 12345678.9


Fig 4-19 : Sample Averages Printout

4.5.3 Cancelling Print Jobs


1)
2)

From the sliding menu on the Main screen, touch the Print icon shown in Figure 4-20
to invoke the Print Center screen.
The Print Center displays as shown in Figure 4-12.

3)
4)

Touch the Cancel Print icon, shown in Figure 4-21. Printing will stop.
To return to the Main Screen, touch the Return icon.

Fig 4-20 : Print


Icon

KSS - September 2005 - Edition 3

Fig 4-21 : Cancel


Print Icon

Met One

Operator Manual

64 of 110

Operating Procedures - Model 3400

4.6 Generating Reports


ISO 14644-1, FS 209E and BS 5295 specify calculations for airborne particle counter count
data. These documents establish the definitions for level of cleanliness in cleanrooms and
clean zones based on specified concentrations of airborne particles. The printed reports
provide the data to determine the cleanliness level for which that cleanroom qualifies.

Note:
Reports will be generated from all records in the buffer. It may be necessary to clear some
records from the buffer as described in Clearing Buffer Records on page 57 before
generating a report.

4.6.1 ISO 14644-1 Calculation Report

ISO14644-1

ISO 14644-1 sets classification levels to be used for the specification of air cleanliness in
cleanrooms and other controlled environments. ISO 14644-1 establishes nine classes of
cleanliness, all expressed in counts per cubic meter of air. ISO calculations also require
samples to be taken at two or more locations.

Note:
The 3400s Upper Confidence Limit (UCL) specification of 95% is only valid for 2 to 9 locations
per ISO specifications.

4.6.2 FS 209E Calculation Report


Although FS 209E has been superseded by ISO 14644-1, this regulation specifies the
maximum allowable number of 0.1, 0.2, 0.3, 0.5 micron and larger particles per cubic foot (or
cubic meter) of air as determined by statistical methods. Two or more locations must be
sampled for calculations to be made. Six class limits (1, 10, 100, 1000, 10,000, 100,000)
designate the specific concentrations of airborne particles per cubic foot of air.

FS209E

Note:
The 3400s Upper Confidence Limit (UCL) specification of 95% is only valid for 2 to 9 locations
per ISO specifications.

4.6.3 BS 5295 Calculation Reports


Although BS 5295 has been superseded by ISO 14644-1, this regulation describes in detail the
requirements for cleanrooms and air control devices and is still widely used. It establishes ten
classes of cleanliness, all expressed in particles per cubic meter of air.

Note:
To retake the samples at the same location, verify the location number is the same as the first
sample and sample again. The data from the retaken sample will then be used in the
calculations.

Operator Manual

Met One

KSS - September 2005 - Edition 3

BS5295

Model 3400 - Operating Procedures

65 of 110

To create calculation reports:

Note:
Reports will be generated from all records in the buffer. It may be necessary to clear some
records from the buffer as described in Clearing Buffer Records on page 57 before
generating a report.
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
12)

Position the counter at the first sample location; turn power on.
Assign a location number to the first location using the procedure described in Taking
Counts on page 55.
Set the number of cycles to be taken at each location using the procedure described in
Changing the Count Cycles on page 38.
Set the sample time of each cycle using the procedure described in Setting Sample,
Hold, or Delay Times on page 41 and enter a hold time.
Set mode to Counts/cubic meter as described in Changing the Count Mode on
page 40.
Set mode to Automatic or Beep (not Manual) as described in Changing the Run
Mode on page 39.
Press Run to start the first count cycle. The counter stops when samples are complete.
Turn the counter off to avoid contamination between samples and move counter to the
next location.
Assign a location number to the second location, then press Run to start the first count
cycle.
Repeat step 8 and step 9 for the remaining locations.
From the sliding menu on the Main screen, touch the Print icon shown in Figure 4-22
to invoke the Print Center screen.
Press the icon of the report to print:
ISO 14644-1
FS 209E
BS 5295

Fig 4-22 : Print


Icon

The report prints. Example printouts are shown in Figure 4-23 and Figure 4-24.

KSS - September 2005 - Edition 3

13)

Met One

Operator Manual

66 of 110

Operating Procedures - Model 3400

FEDERAL STANDARD 209E STATISTICS


AVERAGE PARTICLES/CU M
ISO 14644-1 STATISTICS

S/N ##########

AVERAGE PARTICLES/CU M
S/N ##########

LOCATION ###
DATE YYYY-MM-DD

LOCATION ###

CYCLES ###

DATE YYYY-MM-DD
CYCLES ###

TIME HH:MM:SS

FLOWRATE ##.#LPM

TIME HH:MM:SS

FLOWRATE ##.#LPM

VOLUME #.#CU FT PERIOD HH:MM:SS


TEMPERATURE ###.#F

VOLUME #.#CU FT PERIOD HH:MM:SS

AIR VELOCITY #.#FT/MIN

TEMPERATURE ###.#F

SIZE

RH ###.#%

RH ###.#%

CUMULATIVE DIFFERENTIAL

AIR VELOCITY #.#FT/MIN

0.3m 12345678.9 12345678.9

SIZE

0.5m 12345678.9 12345678.9

CUMULATIVE DIFFERENTIAL

0.3m 12345678.9 12345678.9

1.0m 12345678.9 12345678.9

0.5m 12345678.9 12345678.9

3.0m 12345678.9 12345678.9

1.0m 12345678.9 12345678.9

5.0m 12345678.9 12345678.9

Fig 4-23 : ISO 14644-1 Printout

10.0m 12345678.9 12345678.9


LOCATION ###
Fig 4-24 : FS 209E Printout

4.7 Exporting Data


1)

Make sure the USB connection is secure or that the USB cable or memory device is
connected before proceeding.

Note:
It may take up to 60 seconds for the 3400 to recognize a USB memory device.
From the sliding menu on the Main screen, touch the Export icon shown in Figure
4-25 to invoke the Export Sample Data screen.

KSS - September 2005 - Edition 3

2)

Operator Manual

Met One

Model 3400 - Operating Procedures

67 of 110

Export icon
Fig 4-25 : Main Screen Sliding Menu, Export Icon

3)

The Export Sample Data screen displays as shown in Figure 4-26.

KSS - September 2005 - Edition 3

Fig 4-26 : Export Sample Data screen

4)
5)

6)

Met One

From the Export Sample Data, touch to select either a Comma Separated File or a
Tab Separated File.
To change the default file name, touch the File name: field to invoke a keyboard entry
screen. When finished entering the file name, press Enter to return to the Export
Sample Data screen.
Touch Export. A status bar appears at the bottom to indicate how much of the
download has been completed, as shown in Figure 4-27.
Operator Manual

68 of 110

Operating Procedures - Model 3400

Fig 4-27 : Sample Data Exporting

7)

A pop-up box appears to indicate when the export is complete; see Figure 4-28. Press
OK to close this pop-up.

8)
9)

Click Cancel to exit the Export Sample Data screen.


When using a USB memory device:
Wait at least 60 seconds for the 3400 to recognize the USB drive.
Wait until the USB status light has stopped flashing before removing the
memory device.

An example of a data file appears in Figure 4-29.

Operator Manual

Met One

KSS - September 2005 - Edition 3

Fig 4-28 : Sample Data Exported Pop-up Box

Model 3400 - Operating Procedures

69 of 110

Exported Data Status Bit Masks


An example of a data file appears in Figure 4-29.

Fig 4-29 : Sample Data Output File

The Status value represents several elements of unit status; environmental, count, and
concentration alarms can be indicated with one value. The definitions of these values appear
in Table 4-5.
To determine the alarm conditions present, take the Status value and subtract the largest
possible value listed in Table 4-5. Continue to subtract the largest possible value until the value
equals zero (0).
For example, a Status value of 16512 represents a concentration alam, as shown in Table 4-4.
Table 4-4 : Status Value Example: 16512
Status
Value

Minus Largest
Possible Value

Meaning

Value Remaining

16512

- 16382

Channel 1 concentration alarm

128

128

- 128

Concentration alarm

Table 4-5 : Sample Status Bit Mask Definition

KSS - September 2005 - Edition 3

Bit

Met One

Value

Meaning

1 Calibration

2 Flow

4 Temperature

8 Relative Humidity

16 Air Velocity

32 System Alarm

64 Count Alarm

128 Concentration Alarm

256 Channel 1 Count Alarm

512 Channel 2 Count Alarm

Operator Manual

70 of 110

Operating Procedures - Model 3400

Table 4-5 : Sample Status Bit Mask Definition


Bit

Value

Meaning

10

1024 Channel 3 Count Alarm

11

2048 Channel 4 Count Alarm

12

4096 Channel 5 Count Alarm

13

8192 Channel 6 Count Alarm

14

16384 Channel 1 Concentration Alarm

15

32768 Channel 2 Concentration Alarm

16

65536 Channel 3 Concentration Alarm

17

131072 Channel 4 Concentration Alarm

18

262144 Channel 5 Concentration Alarm

19

524288 Channel 6 Concentration Alarm


1048576 Unassigned

21

2097152 Unassigned

22

4194304 Unassigned

23

8388608 Unassigned

24

16777216 Unassigned

25

33554432 Unassigned

26

67108864 Unassigned

27

134217728 Unassigned

28

268435456 Unassigned

29

536870912 Unassigned

30

1073741824 Unassigned

31

2147483648 Unassigned

32

4294967296 Unassigned

KSS - September 2005 - Edition 3

20

Operator Manual

Met One

Model 3400 - Password Protection

71 of 110

Password Protection
5.1 Password Overview
The password function allows a system administrator to prevent users from modifying the
counters settings. Once the password is enabled, the counter allows three types of access
levels:
The Administrator login has permissions to change all settings on the particle counter.
The User login only allows the operator to change the counters location ID. All other
counter settings are only adjustable by an administrator.
The Factory login is for factory use only.
To enable password protection:
1) Establish user accounts and passwords.
2) Enable password protection for the 3400.
If password protection is not enabled, any user may access all the functionality of the 3400.

5.2 Administrator Login


1)

First, from the sliding menu from the main screen, touch the Lock icon, shown in
Figure 5-1, to invoke a login screen, shown in Figure 5-2.

Fig 5-1 : Lock


Icon

KSS - September 2005 - Edition 3

Fig 5-2 : Login Screen

2)
3)
4)
5)
6)
7)

Met One

Touch the Name field to enter the user name and a keyboard entry screen appears.
Enter Admin by touching the screen.
When the name is entered, touch the Enter button on the screen.
Touch the Password field to invoke the keyboard screen.
Enter the default administrator password, 123456.
Press Enter.
Operator Manual

72 of 110

Password Protection - Model 3400

CAUTION:
To ensure system security, change the default administrator password following the procedure
described in Changing a Password on page 72.

5.3 Enabling Password Protection


1)

To access Global Settings, from the Main screen, invoke the sliding menu and select
the Setup icon, shown in Figure 5-3.

2)
3)

From the Counter Setup screen, select the System icon, shown in Figure 5-4.
In the Global Settings screen, select the Basic tab, shown in Figure 5-5.

Fig 5-3 : Setup


Icon

Fig 5-4 : System


Icon

Fig 5-5 : Basic Tab

4)
5)
6)

Touch the checkbox next to User Logon Required to enable password protection.
To save changes and return to the Counter Setup screen, press the Return icon.
To return to the Main screen, select the Return icon.

The Login screen allows users to change their passwords. Refer to Logging In on page 53 for
information about the logon process.
To change a password:
1)

Operator Manual

From the login screen shown in Figure 5-2, touch the Change Password button to
invoke the Change Password screen shown in Figure 5-6.

Met One

KSS - September 2005 - Edition 3

5.4 Changing a Password

Model 3400 - Password Protection

73 of 110

Fig 5-6 : Change Password Screen

2)
3)
4)
5)

Touch the Old Password field to invoke a keyboard entry screen.


Enter the existing password by touching the screen.
Touch the New Password field to invoke a keyboard entry screen.
Enter the new password by touching the screen.

Note:
User names and passwords can be no longer than 10 characters.
When the password is entered, touch the OK button on the screen.
Touch the Confirm Password field to invoke a keyboard entry screen.
Re-enter the new password by touching the screen.
When the password is re-entered, touch the OK button on the screen.
Press OK on the Change Password screen and the password is changed.

KSS - September 2005 - Edition 3

6)
7)
8)
9)
10)

Met One

Operator Manual

74 of 110

Password Protection - Model 3400

5.5 Replacing a Forgotten Password


If the System Administrator password is forgotten or lost, contact Hach Ultra Analytics at
800.866.7889 or +1 541.472.6500 to receive a new password.
Technical support requires the following information to issue a new password:

KSS - September 2005 - Edition 3

The counters serial number


The counters current date setting in the format MMDDYYYY, where MM is the two digit
month, DD is the two digit day, and YYYY is the four digit year

Operator Manual

Met One

Model 3400 - Maintenance and Diagnostics

75 of 110

Maintenance and Diagnostics


6.1 Cleaning the Unit
WARNING
If a hazardous chemical comes in contact with the touchscreen, clean the touchscreen
immediately to prevent human contact with the chemical.
Cap the air intake tube before cleaning.
The units exterior may be cleaned by spraying a mild cleaning solution on a soft cloth
and wiping the unit.
To clean the touchscreen, first wipe it with a soft cloth. If debris remains, wipe with a soft
cloth moistened with a mild cleaning solution or alcohol.

CAUTION:
Do not spray the touchscreen with solution as the solution may penetrate the electronics and
damage the 3400.

6.2 Running a Zero Count


After particularly high counts, especially surprisingly high particle counts, perform a zero count
test on the 3400 to ensure counts are registering accurately and to clear the counter of any
residual particles.
To purge the unit:
1)
2)
3)

Place the zero count filter on the intake tube of the 3400.
Turn unit on, login (if necessary), and press the Run icon, shown in Figure 6-1.
Repeat this process until the counts decrease to zero.

Fig 6-1 : Run Icon

6.3 Updating the 3400 Software


CAUTION:

KSS - September 2005 - Edition 3

Do not execute the following procedure without receiving upgrade files and explicit direction
from Hach Ultra Analytics.
1)

Download a self-extracting .ZIP software update file from Hach Ultra Analytics.

2)
3)
4)
5)

Extract the files to a compatible USB flash drive1. Do not move or delete any files.
Remove the batteries and AC power from the Model 3400.
Plug the USB flash drive into the USB port.
Apply AC power to the Model 3400.

1.

Met One

Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400; contact
Customer Service for additional support at 800.866.7889 or 541.472.6500.
Operator Manual

76 of 110

Maintenance and Diagnostics - Model 3400

6)
7)

8)

Upon successful completion of the update process, the instrument will restart.
To verify the success of the update process, go to the Diagnostics screen as
described in Diagnostics Screen Overview and verify the version number located on
the lower left-hand portion of the screen matches the version on the upgrade.
Once a successful update is confirmed, delete the files from the USB drive.

6.4 Checking Battery Status


From the Main screen, touching the Battery Status icon shown in Figure 6-2 invokes the
Battery Diagnostics screen shown in Figure 6-3.

Battery Status Icons

KSS - September 2005 - Edition 3

Fig 6-2 : Main Screen, Battery Status Icons

Fig 6-3 : Battery Diagnostics Screen

Operator Manual

Met One

Model 3400 - Maintenance and Diagnostics

77 of 110

Press the Return icon to return to the Main screen.

6.5 Diagnostics Screen Overview


From the Main screen, invoke the sliding menu and touch the Diagnostics icon to view
information about the unit.

Diagnostics icon
Fig 6-4 : Main Screen Sliding Menu, Diagnostics Icon

KSS - September 2005 - Edition 3

The Diagnostics screen, shown in Figure 6-5, provides information that may be helpful in
troubleshooting connections or calibration data.

Fig 6-5 : Diagnostics Screen

Met One

Operator Manual

78 of 110

Maintenance and Diagnostics - Model 3400

If any diagnostics have failed, the word FAIL appears in red, as shown in Figure 6-6.

Fig 6-6 : Diagnostics Screen: Clock Battery Failure

6.5.1 System Status Printout


Technical support may ask for a fax copy of the system status printout. To generate this:
1)

To access Diagnostics, from the Main screen, invoke the sliding menu shown in
Figure 6-4 and select the Diagnostics icon, shown in Figure 6-7.

2)
3)

Touch the Print icon.


The printout appears similar to the one shown in Figure 6-9.

Fig 6-7 :
Diagnostics Icon

---- SYSTEM STATUS ---Fig 6-8 : Print


Icon

S/N ##########
DATE YYYY-MM-DD

TIME HH:MM:SS

KSS - September 2005 - Edition 3

Fig 6-9 : System Diagnostics Printout

Operator Manual

Met One

Model 3400 - Maintenance and Diagnostics

79 of 110

6.6 Factory Settings Screen


Note:
These screens are view-only screens and are only included here for reference. An
administrator-level login may be required to view these screens.
The Factory Settings screen shows the units identifying information, including model, serial
number, nominal flow, and calibration information. This information is view-only and may not be
edited.
To display the Factory Settings screen:
1)

From the sliding menu on the Main screen, touch the Login icon, shown in Figure 6-10
to invoke a Login screen.

Setup icon

Login icon

Fig 6-10 : Main Screen Sliding Menu, Setup and Login Icons

2)
3)

KSS - September 2005 - Edition 3

4)

Login as an administrator, if a login is required.


From the sliding menu on the Main screen, touch the Setup icon, shown in Figure 6-10
to invoke the Counter Setup screen.
The Counter Setup displays as shown in Figure 6-11.

Met One

Operator Manual

80 of 110

Maintenance and Diagnostics - Model 3400

Fig 6-11 : Counter Setup Screen

5)

Touch the Factory icon

to display the Factory Settings screen, shown in

Figure 6-12.

Fig 6-12 : Factory Settings Screen

KSS - September 2005 - Edition 3

The Factory Settings screen has three tabs:


Counter, discussed on page 81
Calibration, discussed on page 82
Test, discussed on page 82

Operator Manual

Met One

Model 3400 - Maintenance and Diagnostics

81 of 110

6.6.1 Counter Tab


The Counter tab displays the information shown in Figure 6-13.

Fig 6-13 : Counter Tab

KSS - September 2005 - Edition 3

The Model 3400 must be calibrated every year; the calibration expiration date appears on the
Counter tab. For information on returning the 3400 to the factory for recalibration, follow all
procedures listed in Return Procedures on page 85 or contact Hach Ultra Analytics at
800.866.7889 or 541.472.6500.

Met One

Operator Manual

82 of 110

Maintenance and Diagnostics - Model 3400

6.6.2 Calibration Tab


The Calibration tab, shown in Figure 6-14, provides information on the calibration thresholds
for each particle size channel in the unit. The Add, Remove, and Edit functions are inactive and
for factory use only.

Fig 6-14 : Calibration Tab

6.6.3 Test Tab

Fig 6-15 : Test Tab

From the Test tab, print the factory settings for records, use in troubleshooting, or to assist
Technical Support.
Operator Manual

Met One

KSS - September 2005 - Edition 3

The Test tab displays the information shown in Figure 6-15.

Model 3400 - Maintenance and Diagnostics

83 of 110

1)

Touch the Print icon

2)

The factory settings print similarly to the example shown in Figure 6-16.

---- FACTORY SETUP ---DATE YYYY-MM-DD

TIME HH:MM:SS

MODEL NUMBER

3415

SERIAL NUMBER

#######

NOMINAL FLOW

50LPM

CALIBRATION DATE
CAL DUE DATE

2004/12/12
2005/12/12

SENSOR CALIBRATION

KSS - September 2005 - Edition 3

Fig 6-16 : Factory Setup Printout

Met One

Operator Manual

Maintenance and Diagnostics - Model 3400

KSS - September 2005 - Edition 3

84 of 110

Operator Manual

Met One

Model 3400 - Service Procedures

85 of 110

Appendix A: Service Procedures


A.1 Return Procedures
The 3400 Series Particle Counter has a one-year calibration cycle. The Model 3400 must be
returned to an authorized service center for recalibration after one year of the date of
calibration, listed on the decal on the back of the Model 3400.
To return the 3400 Series Particle Counter for repair or calibration, first obtain a returned
material authorization number (RA#). The RA# number is necessary for any instrument that
requires repair or calibration by an authorized service center. Include the RA# number on the
shipping label when the instrument is returned.
For the most up-to-date RA# process information, including copies of all required forms, call
Hach Ultra Analytics at 800.866.7889 or +1 541.472.6500.
To return an instrument for credit, please contact the local sales representative.

WARNING
The following actions must be performed when returning any unit for any reason to prevent
personal injury and/or damage to the unit.
When the sensor is not in use, or is being stored, verify that the flow cell is clean. Nonvolatile compounds can coat the flow cell windows, making cleaning very difficult.

A.2 Technical Support Information


Technical Support Engineers are available to provide advice and recommendations for
applications, product operation, measurement specifications, hardware and software, factory
and customer site training.
Please provide name, company, phone number, fax number, model number, serial number and
comment or question.

KSS - September 2005 - Edition 3

Call +1 541.472.6500
Toll Free 800.866.7889 (US/CA)
Fax +1 (541) 474-7414
6:30 AM to 5:00 PM Pacific Time
Monday through Friday
Email: TechSupportGP@hachultra.com

Met One

Operator Manual

Service Procedures - Model 3400

KSS - September 2005 - Edition 3

86 of 110

Operator Manual

Met One

Model 3400 - Technical Data

87 of 110

Appendix B: Technical Data


B.1 Performance Specifications
Number of Size Ranges

Particle Size Ranges1

3413 and 3423


3415 and 3425

0.3, 0.5, 1.0, 3.0, 5.0, 10.0


0.5, 1.0, 2.0, 3.0, 5.0, 25.0

Flow Rate

3413 and 3415


3423 and 3425

28.3 L/min (1.00 cfm)


50 L/min (1.77 cfm)

Zero Count

Conforms to JIS. 1 count or less in 5 minutes, 95% confidence level


3413 and 3415

5% at 14,126,000 particles/m3 (400,000 particles/ft3)

3423 and 3425

5% at 4,000,000 particles/m3 (113,266 particles/ft3)

Counting Efficiency

3413 and 3423


3415 and 3425

50% for 0.3 m


50% for 0.5 m

Light Source

Long Life Laser diode with 10-year Mean Time To Failure (MTTF)

Pump Type

Air vacuum, rated for continuous use

Count Display

Color 1/4 VGA TFT touch screen

Interface

Windows CE-based

Maximum Count Displayed

9,999,999 displayed

Sample Time

1 second to 23 hours 59 minutes 59 seconds

Delay/Hold Times

1 second to 23 hours 59 minutes 59 seconds

Count Alarms

1 to 9,999,999 counts

Data Storage

5,000 samples, scrollable on front panel

Count Cycles

Up to 100 while in automatic mode

Locations

0 to 999, appears on printout

Outputs

USB Client (Version 1.1)


USB Host (Version 1.1)

Battery Type

Lithium ion smart battery; rechargeable, ejectable, and hot-swappable

Quantity Included

Operating Time (Battery)2

3413 and 3415


3423 and 3425

Battery Recharge Time

6.75 hours minimum, 10 hours maximum

Power

100-240 VAC, 50/60 Hz

Size (W x D x H)

31.8 x 25.4 x 20.3 cm (12.5 x 10 x 8)

Weight (without battery)

3413 and 3415


3423 and 3425

KSS - September 2005 - Edition 3

Coincidence Loss

6 hours
7 hours

6.58 kg (14.5 lbs)


7.35 kg (16.2 lbs)

1 Channel selections can be selected at time of order in a range from 0.3 m to 25.0 m; however, 0.3 m and 25.0 m cannot be configured
together.
2 Battery life is estimated with the following conditions: Two fully-charged batteries placed in a 50 L/min unit sampling for 20 minutes (1 m3 sample),
print record, a 5 minute hold time (simulating move to new location), then repeating this cycle. The backlight time-out set to two minutes.

Met One

Operator Manual

88 of 110

Technical Data - Model 3400

Battery Weight

0.66 kg (1.45 lbs)

Enclosure material

Stainless steel

Environment

Operating:

0 to 40C (32 to 104F)


10 to 90% relative humidity, non-condensing

Storage:

-40 to 50C (-40 to 122F)


10 to 90% relative humidity, non-condensing

Instrument:
Long Life Laser diode:

Warranty

2 years
3 years

B.2 Accessories Included

Isokinetic probe (aluminum)

Zero count filter

Tripod for isokinetic probe

Stylus for touch screen

PortAll Version 2 trial software

Flash memory drive1

AC power cord and power supply

Nozzle cleaning brush

2 rolls printer paper

Operator Manual

3413 and 3415


3423 and 3425

0.825 cm (0.325) I.D. Hytrel tubing


1.27 cm (0.5) I.D. Hytrel tubing

LiIon Smart Battery, quantity 1


1.

Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with
the 3400; contact Customer Service for additional support at 800.866.7889 or
541.472.6500.

B.3 Recommended Spares Kit


Hach Ultra Analytics Technical Support Department recommends stocking the items listed in
Table B-1 for each Model 3400 counter.
Table B-1 : Contents of Recommended Spare Parts Kit, PN 2087919-0
Quantity

Spare Lithium Ion Smart Batteries

External battery charger

2-roll pack thermal printer paper

2 (4 rolls total)

KSS - September 2005 - Edition 3

Item

Operator Manual

Met One

Model 3400 - Technical Data

89 of 110

B.4 Spare Parts List


Table B-2 : Spare Parts
Part Number

Description

280-300-5000

Smart Battery Charger

280-120-2024

LiIon Smart Battery

2087966-01

Isokinetic Probe, 1 CFM, AL

2087966-02

Isokinetic Probe, 1 CFM, SS

2087967-01

Isokinetic Probe, 50 LPM, AL

2087967-02

Isokinetic Probe, 50 LPM, SS

2087968-01

Isokinetic Probe Clamp Assembly, AL

2087968-02

Isokinetic Probe Clamp Assembly, SS

970650

Tripod

460519

Thermal Printer Paper

2087939-01

Zero Count Filter

480-100-0004

1/2" ID Hytrel Tubing

701241

Manual

210-400-5171

Stylus

2088028

Demonstration case with foam insert and wheels

70124-01

IQOQ Validation Document

2088010-01

Flash memory drive1

995240

Nozzle cleaning brush


Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400;
contact Customer Service for additional support at 800.866.7889 or 541.472.6500.

KSS - September 2005 - Edition 3

1.

Met One

Operator Manual

Technical Data - Model 3400

KSS - September 2005 - Edition 3

90 of 110

Operator Manual

Met One

Model 3400 - Software Menu Trees

91 of 110

Appendix C: Software Menu Trees


C.1 Menu Flow Diagram
Main Screen
Counter Setup

Sample

Sample Setup
General
Timing
Environmental Units

Alarms

Alarm Settings
Count

KSS - September 2005 - Edition 3

Environmental

Printer

Print Center

Group

Group Setting Management

System

System Diagnostics

Factory

Factory Settings
Counter
Calibration
Test

Met One

Operator Manual

92 of 110

Software Menu Trees - Model 3400

Alarm Settings
Count
Environmental

System Diagnostics

Battery Diagnostics

Print Center

KSS - September 2005 - Edition 3

Export Sample Data

Operator Manual

Met One

Model 3400 - Group Settings

93 of 110

Appendix D: Group Settings


D.1 Group Setting Forms
This appendix contains two forms for customer use. The forms list all parameters controlled by
the group settings function described in Group Settings on page 48.

KSS - September 2005 - Edition 3

These pages may be reproduced to facilitate record keeping and maintenance of the 3400.

Met One

Operator Manual

Group Settings Record Form

Screen

Tab

Parameter

Group Name:

CH1 Low:

High:

CH2 Low:

High:

CH3 Low:

High:

CH4 Low:

High:

CH5 Low:

High:

CH6 Low:

High:

CH1 Low:

High:

CH2 Low:

High:

CH3 Low:

High:

CH4 Low:

High:

CH5 Low:

High:

CH6 Low:

High:

Air Velocity

Low:

High:

Flow Rate

Low:

High:

Relative Humidity

Low:

High:

Temperature

Low:

High:

Count

Alarm Count Limit (per channel)

Environmental

Value

Date Created:

Alarm Concentration Limit

Alarm Settings

Value

Print Cycles
Sample Print Mode
Print Center
Location ID
Count Cycles
General

Count Display
Count Mode
Run Mode
Air Velocity

Sample Setup

Environmental
Units

Flow Rate
Temperature

KSS - September 2005 - Edition 3

Sample Time
Timing

Hold Time
Delay Time

This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.

Group Settings Record Form

Group Name:

Created by/date:

Sample Setup

Location ID:

Count Display:

Sample Time:

Count Cycles:

Count Mode:

Hold Time:

Run Mode:

Delay Time:

KSS - September 2005 - Edition 3

Temperature:
Air Velocity:
Flow Rate:

This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.

Group Settings Record Form

Group Name:

Created by/date:

Alarm Settings

Count Limit

Conc. Limit

Low

Ch1

Temp

Ch2

RH

Ch3

Air Velocity

Ch4

Flow rate

High

Enabled

Ch5
Ch5

Print Center

Print mode:

KSS - September 2005 - Edition 3

Cycles:

This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.

Model 3400 - Certifications

97 of 110

Appendix E: Certifications
E.1 Overview

KSS - September 2005 - Edition 3

This appendix contains the Certificate of Conformity and the WEEE statement.

Met One

Operator Manual

Certifications - Model 3400

KSS - September 2005 - Edition 3

98 of 110

Operator Manual

Met One

Certifications - Model 3400

KSS - September 2005 - Edition 3

100 of 110

Operator Manual

Met One

Electrical equipment marked with this symbol may not be disposed of in European
public disposal systems after 12 August of 2005. In conformity with European local
and national regulations (EU Directive 2002/96/EC), European electrical equipment
users must now return old or end-of life equipment to the Producer for disposal at
no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment for proper disposal.
Important document. Retain with product records.

GERMAN
Elektrogerte, die mit diesem Symbol gekennzeichnet sind, drfen in Europa nach dem
12. August 2005 nicht mehr ber die ffentliche Abfallentsorgung entsorgt werden. In
bereinstimmung mit lokalen und nationalen europischen Bestimmungen (EU-Richtlinie
2002/96/EC), mssen Benutzer von Elektrogerten in Europa ab diesem Zeitpunkt alte bzw.
zu verschrottende Gerte zur Entsorgung kostenfrei an den Hersteller zurckgeben.
Hinweis: Bitte wenden Sie sich an den Hersteller bzw. an den Hndler, von dem Sie das
Gert bezogen haben, um Informationen zur Rckgabe des Altgerts zur ordnungsgemen
Entsorgung zu erhalten.
Wichtige Informationen. Bitte zusammen mit den Produktinformationen aufbewahren.

FRENCH
A partir du 12 aot 2005, il est interdit de mettre au rebut le matriel lectrique marqu de ce
symbole par les voies habituelles de dchetterie publique. Conformment la rglementation
europenne (directive UE 2002/96/EC), les utilisateurs de matriel lectrique en Europe
doivent dsormais retourner le matriel us ou prim au fabricant pour limination, sans
frais pour lutilisateur.
Remarque : Veuillez vous adresser au fabricant ou au fournisseur du matriel pour les
instructions de retour du matriel us ou prim aux fins dlimination conforme.
Ce document est important. Conservez-le dans le dossier du produit.

ITALIAN
Le apparecchiature elettriche con apposto questo simbolo non possono essere smaltite
nelle discariche pubbliche europee successivamente al 12 agosto 2005. In conformit alle
normative europee locali e nazionali (Direttiva UE 2002/96/EC), gli utilizzatori europei di
apparecchiature elettriche devono restituire al produttore le apparecchiature vecchie o a fine
vita per lo smaltimento senza alcun costo a carico dellutilizzatore.
Nota: Per conoscere le modalit di restituzione delle apparecchiature a fine vita da riciclare,
contattare il produttore o il fornitore dellapparecchiatura per un corretto smaltimento.
Documento importante. Conservare con la documentazione del prodotto.

DANISH
Elektriske apparater, der er mrket med dette symbol, m ikke bortskaffes i europiske
offentlige affaldssystemer efter den 12. august 2005. I henhold til europiske lokale og
nationale regler (EU-direktiv 2002/96/EF) skal europiske brugere af elektriske apparater
nu returnere gamle eller udtjente apparater til producenten med henblik p bortskaffelse
uden omkostninger for brugeren.
Bemrk: I forbindelse med returnering til genbrug skal du kontakte producenten eller
leverandren af apparatet for at f instruktioner om, hvordan udtjente apparater bortskaffes
korrekt.
Vigtigt dokument. Opbevares sammen med produktdokumenterne.

SWEDISH
Elektronikutrustning som r mrkt med denna symbol kanske inte kan lmnas in p europeiska
offentliga sopstationer efter 2005-08-12. Enligt europeiska lokala och nationella freskrifter
(EU-direktiv 2002/96/EC) mste anvndare av elektronikutrustning i Europa nu terlmna
gammal eller utrangerad utrustning till tillverkaren fr kassering utan kostnad fr anvndaren.
Obs! Om du ska terlmna utrustning fr tervinning ska du kontakta tillverkaren av
utrustningen eller terfrsljaren fr att f anvisningar om hur du terlmnar kasserad
utrustning fr att den ska bortskaffas p rtt stt.
Viktigt dokument. Spara tillsammans med dina produktbeskrivningar.

SPANISH
A partir del 12 de agosto de 2005, los equipos elctricos que lleven este smbolo no debern
ser desechados en los puntos limpios europeos. De conformidad con las normativas europeas
locales y nacionales (Directiva de la UE 2002/96/EC), a partir de esa fecha, los usuarios
europeos de equipos elctricos debern devolver los equipos usados u obsoletos al fabricante
de los mismos para su reciclado, sin coste alguno para el usuario.
Nota: Srvase ponerse en contacto con el fabricante o proveedor de los equipos para solicitar
instrucciones sobre cmo devolver los equipos obsoletos para su correcto reciclado.
Documento importante. Guardar junto con los registros de los equipos.

DUTCH
Elektrische apparatuur die is voorzien van dit symbool mag na 12 augustus 2005 niet meer
worden afgevoerd naar Europese openbare afvalsystemen. Conform Europese lokale en
nationale wetgegeving (EU-richtlijn 2002/96/EC) dienen gebruikers van elektrische apparaten
voortaan hun oude of afgedankte apparatuur kosteloos voor recycling of vernietiging naar de
producent terug te brengen.
Nota: Als u apparatuur voor recycling terugbrengt, moet u contact opnemen met de producent
of leverancier voor instructies voor het terugbrengen van de afgedankte apparatuur voor een
juiste verwerking.
Belangrijk document. Bewaar het bij de productpapieren.

POLISH
Sprzt elektryczny oznaczony takim symbolem nie moe by likwidowany w europejskich
systemach utylizacji po dniu 12 sierpnia 2005. Zgodnie z europejskimi, lokalnymi i pastwowymi
przepisami prawa (Dyrektywa Unii Europejskiej 2002/96/EC), uytkownicy sprztu elektrycznego
w Europie musz obecnie przekazywa Producentowi stary sprzt lub sprzt po okresie
uytkowania do bezpatnej utylizacji.
Uwaga: Aby przekaza sprzt do recyklingu, naley zwrci si do producenta lub dostawcy
sprztu w celu uzyskania instrukcji dotyczcych procedur przekazywania do utylizacji sprztu
po okresie uytkowania.
Wany dokument. Zachowa z dokumentacj produktu.

PORTUGESE
Qualquer equipamento elctrico que ostente este smbolo no poder ser eliminado atravs
dos sistemas pblicos europeus de tratamento de resduos slidos a partir de 12 de Agosto
de 2005. De acordo com as normas locais e europeias (Directiva Europeia 2002/96/EC), os
utilizadores europeus de equipamentos elctricos devero agora devolver os seus
equipamentos velhos ou em fim de vida ao produtor para o respectivo tratamento sem
quaisquer custos para o utilizador.
Nota: No que toca devoluo para reciclagem, por favor, contacte o produtor ou fornecedor
do equipamento para instrues de devoluo de equipamento em fim de vida para a sua
correcta eliminao.
Documento importante. Mantenha junto dos registos do produto.

Model 3400 - Sample Status Bit Mask Definitions

103 of 110

Appendix F: Sample Status Bit Mask Definitions


F.1 Overview
The Status value represents several elements of unit status; environmental, count, and
concentration alarms can be indicated with one value. The definitions of these values appear
in Table F-1.
To determine the alarm conditions present, take the Status value and subtract the largest
possible value listed in Table F-1. Continue to subtract the largest possible value until the value
equals zero (0).
For more information, refer to Exported Data Status Bit Masks on page 69.

Table F-1 : Sample Status Bit Mask Definitions

KSS - September 2005 - Edition 3

Bit

Value

Meaning

Bit

Value

Meaning

1 Calibration

17

131072 Channel 4 Concentration Alarm

2 Flow

18

262144 Channel 5 Concentration Alarm

4 Temperature

19

524288 Channel 6 Concentration Alarm

8 Relative Humidity

20

1048576 Unassigned

16 Air Velocity

21

2097152 Unassigned

32 System Alarm

22

4194304 Unassigned

64 Count Alarm

23

8388608 Unassigned

128 Concentration Alarm

24

16777216 Unassigned

256 Channel 1 Count Alarm

25

33554432 Unassigned

512 Channel 2 Count Alarm

26

67108864 Unassigned

10

1024 Channel 3 Count Alarm

27

134217728 Unassigned

11

2048 Channel 4 Count Alarm

28

268435456 Unassigned

12

4096 Channel 5 Count Alarm

29

536870912 Unassigned

13

8192 Channel 6 Count Alarm

30

1073741824 Unassigned

14

16384 Channel 1 Concentration Alarm

31

2147483648 Unassigned

15

32768 Channel 2 Concentration Alarm

32

4294967296 Unassigned

16

65536 Channel 3 Concentration Alarm

Met One

Operator Manual

Sample Status Bit Mask Definitions - Model 3400

KSS - September 2005 - Edition 3

104 of 110

Operator Manual

Met One

Model 3400 - Annex

105 of 110

Annex
Tables and illustrations

KSS - September 2005 - Edition 3

Table 1-1
Fig. 1-1
Table 1-2
Fig. 2-1
Fig. 2-2
Fig. 2-3
Table 2-1
Fig. 2-4
Fig. 2-5
Fig. 2-6
Table 2-2
Fig. 2-7
Fig. 2-8
Fig. 2-9
Fig. 2-10
Fig. 2-11
Fig. 2-12
Fig. 3-1
Fig. 3-2
Table 3-1
Fig. 3-3
Fig. 3-4
Fig. 3-5
Fig. 3-6
Table 3-2
Fig. 3-7
Fig. 3-8
Fig. 3-9
Fig. 3-10
Fig. 3-11
Fig. 3-12
Table 3-3
Fig. 3-13
Fig. 3-14
Fig. 3-15
Fig. 3-16
Fig. 3-17
Table 3-4
Fig. 3-18
Fig. 3-19
Fig. 3-20
Fig. 3-21

Met One

3400 Series Particle Counter Model Numbers .................................... 9


3400 Series Particle Counter .............................................................. 9
3400 Series Particle Counter Accessories ........................................ 11
3400 Back View................................................................................. 13
3400 Left Side View .......................................................................... 14
3400 Front View ................................................................................ 15
Battery LED Color Indications ........................................................... 16
Main Screen ...................................................................................... 18
3400 Main Screen ............................................................................. 19
Sliding Menu Icons ............................................................................ 20
Battery Status.................................................................................... 20
Run Icon ............................................................................................ 21
Stop Icon ........................................................................................... 21
Main Screen: Sampling ..................................................................... 21
Count Alarm Example ....................................................................... 22
Stop Icon ........................................................................................... 22
Main Screen Sliding Menu ................................................................ 23
Main Screen Sliding Menu, Setup Icon ............................................. 25
Counter Setup Screen....................................................................... 25
Counter Setup Screen Icon Descriptions .......................................... 26
Setup Icon ......................................................................................... 27
System Icon....................................................................................... 27
Global Settings Screen...................................................................... 27
Basic Tab .......................................................................................... 28
Basic Tab Parameters....................................................................... 28
System Icon....................................................................................... 29
Global Settings, Basic Tab ................................................................ 29
Return Icon........................................................................................ 29
System Icon....................................................................................... 30
Return Icon........................................................................................ 30
Sounds Tab ....................................................................................... 30
Sound Tab Parameters ..................................................................... 31
System Icon....................................................................................... 31
Return Icon........................................................................................ 31
System Icon....................................................................................... 32
Return Icon........................................................................................ 32
Options and Accessories Tab ........................................................... 32
Options and Accessories Tab Parameters........................................ 33
System Icon....................................................................................... 33
Global Settings, Language Options................................................... 33
Restart Prompt .................................................................................. 34
Setup Icon ......................................................................................... 34

Operator Manual

Annex - Model 3400

Fig. 3-22
Fig. 3-23
Table 3-5
Fig. 3-24
Fig. 3-25
Fig. 3-26
Fig. 3-27
Fig. 3-28
Fig. 3-29
Fig. 3-30
Fig. 3-31
Fig. 3-32
Fig. 3-33
Fig. 3-34
Table 3-6
Fig. 3-35
Fig. 3-36
Fig. 3-37
Table 3-7
Fig. 3-38
Fig. 3-39
Fig. 3-40
Fig. 3-41
Fig. 3-42
Fig. 3-43
Fig. 3-44
Fig. 3-45
Fig. 3-46
Fig. 3-47
Fig. 3-48
Fig. 3-49
Fig. 3-50
Fig. 3-51
Table 3-8
Fig. 3-52
Fig. 3-53
Fig. 3-54
Fig. 3-55
Fig. 3-56
Fig. 3-57
Fig. 3-58
Fig. 3-59
Fig. 3-60
Fig. 3-61
Table 4-1
Fig. 4-1

Operator Manual

Sample Icon ...................................................................................... 34


General Tab ...................................................................................... 35
Sample Tab Parameters ................................................................... 35
Increment and Decrement Icons ....................................................... 36
Numeric Entry Screen ....................................................................... 37
Return Icon........................................................................................ 37
Main Screen with Location ID Highlighted......................................... 37
Main Screen with Location Name Highlighted................................... 38
Return Icon........................................................................................ 39
Sample Icon ...................................................................................... 39
Return Icon........................................................................................ 39
Sample Icon ...................................................................................... 40
Return Icon........................................................................................ 40
Timing Tab ........................................................................................ 40
Timing Tab Parameters..................................................................... 41
Sample Icon ...................................................................................... 41
Return Icon........................................................................................ 41
Environmental Units Tab ................................................................... 42
Environmental Units Tab Parameters ............................................... 42
Sample Icon ...................................................................................... 42
Return Icon........................................................................................ 43
Main Screen Sliding Menu, Alarm Icon ............................................. 43
Alarm Settings Screen....................................................................... 44
Alarms Icon ....................................................................................... 44
Count Alarm Tab ............................................................................... 45
Alarms Icon ....................................................................................... 45
Return Icon........................................................................................ 46
Count Alarm Example ....................................................................... 46
Alarms Icon ....................................................................................... 46
Environmental Alarm Tab.................................................................. 47
Alarms Icon ....................................................................................... 47
Return Icon........................................................................................ 48
Environmental Alarm Example .......................................................... 48
Group Setting Parameters................................................................. 48
Setup Icon ......................................................................................... 49
Group Icon......................................................................................... 49
Group Setting Management Screen.................................................. 50
Setup Icon ......................................................................................... 50
New Group Keyboard Entry Screen .................................................. 51
Return Icon........................................................................................ 51
Group Icon......................................................................................... 51
Setup Icon ......................................................................................... 52
Delete Group Error Messages........................................................... 52
Group Icon......................................................................................... 52
Lock Icon Status................................................................................ 53
Login Screen ..................................................................................... 54

Met One

KSS - September 2005 - Edition 3

106 of 110

Model 3400 - Annex

107 of 110

KSS - September 2005 - Edition 3

Fig. 4-2
Fig. 4-3
Table 4-2
Fig. 4-4
Fig. 4-5
Fig. 4-6
Fig. 4-7
Fig. 4-8
Fig. 4-9
Fig. 4-10
Fig. 4-11
Fig. 4-12
Table 4-3
Fig. 4-13
Fig. 4-14
Fig. 4-15
Fig. 4-16
Fig. 4-17
Fig. 4-18
Fig. 4-19
Fig. 4-20
Fig. 4-21
Fig. 4-22
Fig. 4-23
Fig. 4-24
Fig. 4-25
Fig. 4-26
Fig. 4-27
Fig. 4-28
Fig. 4-29
Table 4-4
Table 4-5
Fig. 5-1
Fig. 5-2
Fig. 5-3
Fig. 5-4
Fig. 5-5
Fig. 5-6
Fig. 6-1
Fig. 6-2
Fig. 6-3
Fig. 6-4
Fig. 6-5
Fig. 6-6
Fig. 6-7
Fig. 6-8

Met One

Keyboard Entry Screen ..................................................................... 54


Main Screen ...................................................................................... 55
Main screen Button Descriptions....................................................... 55
Run Icon ............................................................................................ 56
Stop Icon ........................................................................................... 56
3400 Main Screen, Sample Buffer Field............................................ 56
Setup Icon ......................................................................................... 57
Sample Icon ...................................................................................... 57
Sample Setup Screen ....................................................................... 57
Clear Sample Buffer Pop-Up Box...................................................... 58
Print Icon ........................................................................................... 58
Print Center Screen........................................................................... 59
Print Center Functions....................................................................... 59
Main Screen Sliding Menu, Print Icon ............................................... 61
Print Mode Choices ........................................................................... 61
Return Icon........................................................................................ 62
Sample Data Printout ........................................................................ 62
Print Icon ........................................................................................... 62
Sample Buffer Printout ...................................................................... 63
Sample Averages Printout................................................................. 63
Print Icon ........................................................................................... 63
Cancel Print Icon ............................................................................... 63
Print Icon ........................................................................................... 65
ISO 14644-1 Printout......................................................................... 66
FS 209E Printout ............................................................................... 66
Main Screen Sliding Menu, Export Icon ............................................ 67
Export Sample Data screen .............................................................. 67
Sample Data Exporting ..................................................................... 68
Sample Data Exported Pop-up Box .................................................. 68
Sample Data Output File ................................................................... 69
Status Value Example: 16512 ........................................................... 69
Sample Status Bit Mask Definition .................................................... 69
Lock Icon ........................................................................................... 71
Login Screen ..................................................................................... 71
Setup Icon ......................................................................................... 72
System Icon....................................................................................... 72
Basic Tab .......................................................................................... 72
Change Password Screen................................................................. 73
Run Icon ............................................................................................ 75
Main Screen, Battery Status Icons .................................................... 76
Battery Diagnostics Screen ............................................................... 76
Main Screen Sliding Menu, Diagnostics Icon .................................... 77
Diagnostics Screen ........................................................................... 77
Diagnostics Screen: Clock Battery Failure ........................................ 78
Diagnostics Icon ................................................................................ 78
Print Icon ........................................................................................... 78

Operator Manual

108 of 110

Annex - Model 3400

System Diagnostics Printout ............................................................. 78


Main Screen Sliding Menu, Setup and Login Icons........................... 79
Counter Setup Screen....................................................................... 80
Factory Settings Screen .................................................................... 80
Counter Tab ...................................................................................... 81
Calibration Tab .................................................................................. 82
Test Tab ............................................................................................ 82
Factory Setup Printout....................................................................... 83
Contents of Recommended Spare Parts Kit, PN 2087919-0 ............ 88
Spare Parts ....................................................................................... 89
Sample Status Bit Mask Definitions ................................................ 103

KSS - September 2005 - Edition 3

Fig. 6-9
Fig. 6-10
Fig. 6-11
Fig. 6-12
Fig. 6-13
Fig. 6-14
Fig. 6-15
Fig. 6-16
Table B-1
Table B-2
Table F-1

Operator Manual

Met One

Model 3400 - User notes

109 of 110

KSS - September 2005 - Edition 3

User notes

Met One

Operator Manual

User notes - Model 3400

KSS - September 2005 - Edition 3

110 of 110

end of the book

Operator Manual

Met One

Potrebbero piacerti anche