Manual de instrucciones OI/FCB300/FCH300-ES Rev.

CoriolisMaster FCB330, FCB350, FCH330, FCH350


Caudalmetro Msico Coriolis

Measurement made easy

CoriolisMaster FCB330, FCB350, FCH330, FCH350


Caudalmetro Msico Coriolis

Manual de instrucciones
OI/FCB300/FCH300-ES
Rev. F
Fecha de edicin: 05.2014
Traduccin de las instrucciones originales

Fabricante
ABB Automation Products GmbH
Process Automation
Dransfelder Str. 2
37079 Gttingen
Germany
Tel: +49 551 905-0
Fax: +49 551 905-777
Customer Center Service
Phone.: +49 180 5 222 580
Fax: +49 621 381 931-29031
automation.service@de.abb.com

2 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from one to tw o c olumns

Contenido
1

4.6.2
4.6.3

Seguridad ........................................................................ 6
1.1
Informaciones generales e indicaciones para la
lectura ................................................................. 6
1.2
Uso conforme al fin previsto ................................. 6
1.3
Uso contrario al fin previsto .................................. 6
1.4
Grupos destinatarios y cualificaciones .................. 6
1.5
Suministro de garanta ......................................... 6
1.6
Etiquetas y smbolos ............................................ 7
1.6.1
Smbolos de seguridad / peligro, smbolos de
informacin .......................................................... 7
1.6.2
Placa de caractersticas ....................................... 7
1.7
Instrucciones de seguridad para el transporte ...... 8
1.8
Instrucciones de seguridad para el montaje ......... 8
1.9
Instrucciones de seguridad para la instalacin
elctrica ............................................................... 8
1.10
Instrucciones de seguridad relativas al
funcionamiento .................................................... 8
1.11
Valores tcnicos lmite.......................................... 9
1.12
Fluidos permitidos................................................ 9
1.13
Instrucciones de seguridad para trabajos de
inspeccin y mantenimiento ................................. 9
1.14
Devolucin de aparatos ....................................... 9
1.15
Sistema de gestin integrado ............................. 10
1.16
Eliminacin de residuos ..................................... 10
1.16.1 Informacin sobre la directiva WEEE 2002/96/CE
(residuos de aparatos elctricos y electrnicos)
(Waste Electrical and Electronic Equipment) ....... 10
1.16.2 Directiva ROHS 2002/95/CE .............................. 10
Diseo y funcin............................................................ 11
2.1
Generalidades.................................................... 11
2.2
Principio de medicin......................................... 11
2.3
Versiones del aparato ........................................ 12
2.3.1
Cuadro sinptico del aparato ATEX, IECEx, NEPSI
.......................................................................... 14
2.3.2
Cuadro sinptico del aparato cFMus.................. 15

Transporte ..................................................................... 16
3.1
Controles ........................................................... 16
3.2
Generalidades.................................................... 16

Montaje .......................................................................... 16
4.1
Generalidades.................................................... 16
4.2
Sensor de caudal ............................................... 16
4.3
Transmisor......................................................... 17
4.3.1
Transmisor de diseo remoto (opcin F1 o F2)... 17
4.3.2
Transmisor de diseo remoto (opcin R1 o R2).. 17
4.4
Orientacin de la caja del transmisor y del indicador
LCD ................................................................... 18
4.4.1
Caja del transmisor ............................................ 18
4.4.2
Indicador LCD ................................................... 18
4.5
Instrucciones para el montaje............................. 19
4.5.1
Requisitos de montaje / instrucciones de
planificacin ....................................................... 19
4.5.2
Tramos de entrada ............................................ 19
4.5.3
Modelos de diseo remoto ................................ 19
4.5.4
Prdida de presin............................................. 19
4.6
Posiciones de montaje ....................................... 19
4.6.1
Montaje en vertical, en una tubera ascendente .. 19

Montaje en vertical, en una tubera descendente 20


Montaje en horizontal para la medicin de lquidos
.......................................................................... 20
4.6.4
Montaje en horizontal para la medicin de gases 20
4.6.5
Puntos de montaje problemticos para la medida
de lquidos ......................................................... 21
4.6.6
Puntos de montaje problemticos para la medida
de gases ............................................................ 21
4.6.7
Ajuste del cero ................................................... 21
4.6.8
Instalacin en funcin de la temperatura del fluido
.......................................................................... 22
4.6.9
Instalacin en caso de uso de la opcin TE1
"Longitud ampliada de la torre" .......................... 22
4.6.10 Notas sobre la conformidad EHEDG .................. 22
5

Conexiones elctricas ................................................... 23


5.1
Instrucciones para conectar la alimentacin
elctrica ............................................................. 23
5.2
Indicaciones para el cableado ............................ 23
5.3
Diseo compacto............................................... 24
5.4
Diseo remoto ................................................... 25
5.4.1
Especificacin del cable ..................................... 25
5.4.2
Colocacin del cable de seal ............................ 25
5.4.3
Conexin del cable de seal .............................. 25
5.5
Comunicacin digital .......................................... 26
5.5.1
Protocolo HART ................................................. 26
5.6
Esquemas de conexin ...................................... 27
5.6.1
Conexin de modelos de transductor a equipos
perifricos .......................................................... 27
5.6.2
Ejemplos de conexin con perifricos ................ 28
5.6.3
Conexin del transductor al sensor de caudal .... 29
5.6.4
Conexin del transductor al sensor de caudal en
Zona 1 / Div. 1 ................................................... 30

Puesta en marcha ......................................................... 31


6.1
Controles antes de la puesta en funcionamiento 31
6.2
Conectar la alimentacin de corriente................. 31
6.2.1
Medidas de control despus de conectar la
alimentacin de corriente ................................... 31
6.3
Ajustes bsicos .................................................. 31
6.4
Configuracin de la salida de impulsos ............... 32
6.5
Interruptor de proteccin contra modificaciones no
autorizadas ........................................................ 32
6.6
Instrucciones para el funcionamiento seguro en
zonas potencialmente explosivas (ATEX) ............ 33
6.6.1
Controles ........................................................... 33
6.6.2
Circuitos elctricos de salida .............................. 33
6.6.3
Contacto NAMUR .............................................. 34
6.6.4
Entradas de cables ............................................ 34
6.6.5
Aislamiento del sensor de caudal ....................... 34
6.6.6
Uso en la zona 2, con la clase de proteccin "a
prueba de vapor" (nR) ........................................ 34
6.6.7
Cambio de tipo de proteccin ............................ 35
6.7
Instrucciones para el funcionamiento seguro en
zonas potencialmente explosivas (cFMus) .......... 36
6.7.1
Controles ........................................................... 36
6.7.2
Entradas de cables ............................................ 36
6.7.3
Conexin elctrica ............................................. 36
6.7.4
Process sealing.................................................. 37
6.7.5
Cambio del tipo de proteccin ........................... 37

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 3

Configuracin, parametracin ...................................... 38


7.1
Manejo .............................................................. 38
7.1.1
Navegacin por mens ...................................... 38
7.2
Niveles del men ............................................... 38
7.2.1
Indicador de procesos ....................................... 39
7.2.2
Cambio al nivel de configuracin (parametrizacin)
.......................................................................... 39
7.2.3
Seleccin y modificacin de parmetros ............ 40
7.3
Resumen de los parmetros en el nivel de
configuracin ..................................................... 41
7.4
Descripcin de los parmetros........................... 45
7.4.1
Men: *Prog. level.............................................. 45
7.4.2
Men: Language ................................................ 45
7.4.3
Men: Mode of operation ................................... 46
7.4.4
Men: Concentration ......................................... 47
7.4.5
Men: Unit ......................................................... 48
7.4.6
Men: Flowmeter primary .................................. 49
7.4.7
Men: QmMax ................................................... 49
7.4.8
Men: Damping ................................................. 49
7.4.9
Men: Low cutoff setting ................................... 49
7.4.10 Men: Field optimization .................................... 49
7.4.11 Men: System Zero adj. ..................................... 50
7.4.12 Men: Alarm ...................................................... 50
7.4.13 Men: Display .................................................... 51
7.4.14 Men: Totalizer .................................................. 52
7.4.15 Men: Pulse Output ........................................... 53
7.4.16 Men: Current output 1 ..................................... 54
7.4.17 Men: Current output 2 ..................................... 55
7.4.18 Men: Switch contacts ...................................... 56
7.4.19 Men: Label....................................................... 56
7.4.20 Men: Interface .................................................. 56
7.4.21 Men: Function test ........................................... 57
7.4.22 Men: Status ..................................................... 59
7.4.23 Men: Versin del software ................................ 59
7.5
Medida de concentraciones DensiMass (slo
disponible para FCB350) ................................... 60
7.5.1
Precisin de la medida de concentraciones ....... 60
7.5.2
Introduccin de la matriz de concentraciones .... 60
7.5.3
Estructura de la matriz de concentraciones ........ 61
7.6
Historial de versiones del software ..................... 62

Mensajes de error ......................................................... 62


8.1
Generalidades ................................................... 62
8.2
Sinopsis............................................................. 63
8.3
Mensajes de error .............................................. 64
8.4
Mensajes de advertencia ................................... 66

Mantenimiento / Reparacin ........................................ 67


9.1
Generalidades ................................................... 67
9.2
Limpieza ............................................................ 67
9.3
Sensor de caudal............................................... 67
9.4
Transmisor......................................................... 67
9.4.1
Cambio.............................................................. 67

10

Datos tcnicos - Sensor de caudal .............................. 68


10.1
Diseos ............................................................. 68
10.2
Dimetro nominal y rango de medicin .............. 68
10.2.1 Rango de caudal recomendado ......................... 68
10.3
Precisin............................................................ 68
10.3.1 Condiciones de referencia.................................. 68

10.3.2
10.3.3
10.3.4
10.3.5
10.4
10.4.1
10.4.2
10.4.3
10.4.4
10.4.5
10.4.6
10.4.7
10.4.8
10.4.9
10.4.10
10.4.11

Desviacin del valor medido ............................... 68


Estabilidad del cero............................................ 69
Influencia de la temperatura del fluido ................ 69
Influencia de la presin operativa........................ 69
Datos tcnicos ................................................... 70
Prdida de presin............................................. 70
Rango de viscosidad ......................................... 70
Lmites de temperatura C (F) ........................... 70
Conexiones a proceso ....................................... 70
Presin nominal ................................................. 70
Carcasa como dispositivo de proteccin (opcional)
70
Directiva de equipos apresin ............................ 70
Materiales para el transmisor.............................. 71
Materiales del sensor de caudal ......................... 71
Cargas del material de las conexiones a proceso 71
Curvas de carga del material de los aparatos
bridados ............................................................ 71

11

Datos tcnicos - Transmisor......................................... 72


11.1
Generalidades.................................................... 72
11.2
Datos tcnicos ................................................... 72
11.2.1 Rango de medida .............................................. 72
11.2.2 Tipo de proteccin ............................................. 72
11.2.3 Conexin elctrica ............................................. 72
11.2.4 Suministro de energa ........................................ 72
11.2.5 Tiempo de reaccin ........................................... 72
11.2.6 Temperatura ambiente ....................................... 72
11.2.7 Versin de caja .................................................. 72
11.2.8 Medicin de caudal de avance / retorno............. 72
11.2.9 Indicador LCD.................................................... 72
11.2.10 Mando ............................................................... 73
11.2.11 Seguridad de datos ........................................... 73
11.3
Especificaciones elctricas................................. 73
11.3.1 Salidas de corriente ........................................... 73
11.3.2 Salida de impulsos ............................................. 74
11.3.3 Salidas de contacto digitales.............................. 74
11.3.4 Entradas de contacto digitales ........................... 74

12

Datos tcnicos relevantes para la proteccin contra


explosiones conforme a ATEX / IECEx / NEPSI........... 75
12.1
Especificaciones elctricas................................. 75
12.1.1 Resumen de las diferentes opciones de salida ... 75
12.1.2 Versin I: salidas de corriente activa / pasiva ...... 75
12.1.3 Versin II: salidas de corriente pasiva / pasiva .... 76
12.1.4 Condiciones especiales de conexin .................. 76
12.2
Modelo de sensor de caudal FCB3xx / FCH3xx . 77
12.2.1 Clase de temperatura ........................................ 77
12.2.2 Homologacin de proteccin contra explosiones
ATEX / IECEx / NEPSI ........................................ 78
12.3
Transductor de diseo remoto Modelo FCT30079
12.3.1 Homologacin de proteccin contra explosiones
ATEX / IECEx / NEPSI ........................................ 79

4 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

13

Datos tcnicos relevantes para la proteccin Ex segn


cFMus ............................................................................ 80
13.1
Resumen de las diferentes opciones de salida ... 80
13.2
Datos elctricos para Div. 2 / Zona 2 ................. 80
13.2.1 Versin I: salidas de corriente activa / pasiva y
versin II: salidas de corriente pasiva / pasiva..... 80
13.3
Datos elctricos para Div. 1 / Zona 1 ................. 81
13.3.1 Versin I: salidas de corriente activa / pasiva ...... 81
13.3.2 Versin II: salidas de corriente pasiva / pasiva .... 81
13.3.3 Condiciones especiales de conexin.................. 81
13.4
Sensor de caudal Modelo FCB300 / FCH300.. 82
13.4.1 Clases de temperatura ....................................... 82
13.4.2 Homologacin de proteccin contra explosiones
cFMus ............................................................... 83
13.5
Transmisor de diseo remoto Modelo FCT300 85
13.5.1 Homologacin Ex cFMus ................................... 85

14

Anexo ............................................................................. 87
14.1
Homologaciones y certificados .......................... 87
14.2
Installation diagram 3KXF002126G0009 ............ 92

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 5

1
1.1

Seguridad

Informaciones generales e indicaciones para la


lectura
Lea atentamente este manual de instrucciones antes de
proceder al montaje y la puesta en marcha!
El manual de instrucciones es una parte integrante esencial
del producto y deber guardarse para su uso posterior.
Para asegurar una orientacin fcil, este manual no puede dar
una informacin exhaustiva sobre todas las versiones del
producto, ni tratar todas las formas posibles de instalacin,
funcionamiento o conservacin.
Si precisa ms informacin o si surgen anomalas no descritas
con detalle en el presente manual de instrucciones, le
rogamos se ponga en contacto con el fabricante para solicitar
ms informacin.
El presente manual de instrucciones ni forma parte ni contiene
una modificacin de un acuerdo, una promesa o relacin
jurdica anterior o existente.
El producto cumple los requisitos de seguridad y los niveles
tecnolgicos actuales. Ha sido examinado y ha salido de
fbrica en condiciones perfectas de seguridad. Para mantener
estas condiciones durante el tiempo de servicio previsto, se
deben observar y seguir las indicaciones del presente manual
de instrucciones.
Las modificaciones y reparaciones estn nicamente
permitidas en la forma descrita en el manual de instrucciones.
Slo cuando se siguen y se observan las indicaciones de
seguridad y todos los smbolos de seguridad y advertencia del
manual de instrucciones, se garantiza que el personal
operador y el medio ambiente estn protegidos contra
peligros posibles y que el aparato funcione correctamente.
Es absolutamente necesario que se observen y sigan los
smbolos e indicaciones que se encuentran en la carcasa del
aparato. Asegrese de que sean perfectamente legibles. No
est permitido eliminarlos.

1.2 Uso conforme al fin previsto


El aparato sirve para:
La conduccin de fluidos lquidos y gaseosos (tambin
fluidos inestables).
La medida directa del caudal msico.
La medida indirecta del flujo volumtrico (mediante la
medida de la densidad y del caudal msico).
La medida de la densidad del fluido.
La medida de la temperatura del fluido.
El uso conforme al fin previsto comprende tambin los
siguientes puntos:
Se deben cumplir y seguir las instrucciones de este
manual.
Se deben mantener los valores lmite indicados. Vase el
captulo "Valores tcnicos lmite".
Se deben utilizar exclusivamente los fluidos permitidos.
Vase el captulo "Fluidos permitidos".
1.3 Uso contrario al fin previsto
No est permitido el uso del aparato para:
utilizarlo como adaptador flexible en tuberas, como p. ej.,
para compensar desviaciones, vibraciones y dilataciones
de las mismas, etc.
utilizarlo como peldao, p. ej., para realizar trabajos de
montage
utilizarlo como soporte para cargas externas, p. ej., como
soporte para tuberas, etc.
recubrirlo con otros materiales, p. ej., por sobrepintar la
placa de caractersticas o soldar piezas
arranque de material, p. ej. mediante perforacin de la
carcasa
1.4 Grupos destinatarios y cualificaciones
La instalacin, puesta en marcha y mantenimiento del
producto slo deben ser llevados a cabo por personal
especializado debidamente instruido que haya sido autorizado
por el propietario del equipo. El personal especializado tiene
que haber ledo y entendido el manual y debe seguir sus
indicaciones.
El usuario debe seguir bsicamente las disposiciones
nacionales vigentes en su pas relacionadas con la instalacin,
verificacin, reparacin y conservacin de productos
elctricos.
1.5 Suministro de garanta
Cualquier forma de uso que se no corresponda con el fin
previsto, as como el incumplimiento de este manual de
instrucciones o el empleo de personal insuficientemente
cualificado y modificaciones arbitrarias del aparato, excluyen
la responsabilidad del fabricante por daos y perjuicios que
resulten de ello. En este caso se extinguir la garanta del
fabricante.

6 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

1.6 Etiquetas y smbolos


1.6.1 Smbolos de seguridad / peligro, smbolos de
informacin

1.6.2

PELIGRO Daos graves para la salud /


peligro de muerte!
Este smbolo indica, en combinacin con el
mensaje "PELIGRO", un peligro inminente. El
incumplimiento de esta indicacin de seguridad
causar la muerte o lesiones gravsimas.

Placa de caractersticas
IMPORTANTE (NOTA)
Las placas de caractersticas representadas son
solo ejemplos. Las placas de caractersticas
colocadas en el aparato pueden divergir de esta
representacin.

PELIGRO Daos graves para la salud /


peligro de muerte!
Este smbolo indica, en combinacin con el
mensaje "PELIGRO", un peligro inminente por
corriente elctrica. El incumplimiento de esta
indicacin de seguridad causar la muerte o
lesiones gravsimas.
ADVERTENCIA Daos fsicos!
El smbolo indica, en combinacin con el mensaje
"ADVERTENCIA", una situacin que puede ser
peligrosa. El incumplimiento de esta indicacin de
seguridad puede causar la muerte o lesiones
gravsimas.
ADVERTENCIA Daos fsicos!
Este smbolo indica, en combinacin con el
mensaje "ADVERTENCIA", una situacin que
puede resultar peligrosa debido a la corriente
elctrica. El incumplimiento de esta indicacin de
seguridad puede causar la muerte o lesiones
gravsimas.
ATENCIN Lesiones leves!
El smbolo indica, en combinacin con el mensaje
"ATENCIN", una situacin que puede ser
peligrosa. El incumplimiento de esta indicacin de
seguridad puede causar lesiones leves o menos
graves. El smbolo puede ser utilizado tambin
para advertir de daos materiales.
AVISO Daos materiales!
El smbolo indica una situacin que puede ser
daina.
El incumplimiento de esta indicacin de seguridad
puede ocasionar daos o la destruccin del
producto y/o de otros componentes del equipo.
IMPORTANTE (NOTA)
El smbolo indica consejos para el usuario o
informaciones muy tiles o importantes sobre el
producto o sus ventajas adicionales. La
indicacin "IMPORTANTE (NOTA)" no advierte de
situaciones peligrosas o dainas.

Fig. 1: Sensor de caudal con forma compacta (ejemplo)


1 Denominacin de modelo completa | 2 Cdigo para pedido |
3 Nmero de pedido | 4 Suministro de energa |
5 Dimetro nominal / modo de proteccin |
6 Conexin a proceso / nivel de presin |
7 Material del tubo de medicin | 8 Coeficiente de calibracin |
9 Caudal mximo | 10 Rango de temperatura media |
11 Versin del software | 12 Fabricante |
13 Ao de construccin (mes / ao) |
14 Rango de temperatura ambiente | 15 Comunicacin |
16 Precisin de calibracin |
17 Marca PED (directiva de equipos a presin) |
18 Consumo de potencia mximo | 19 Nmero de serie del sensor |
20 Marca CE | 21 Homologacin Ex cFMus |
22 Homologacin Ex ATEX / IECEx

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 7

1.7 Instrucciones de seguridad para el transporte


Deben observarse las siguientes indicaciones:
Durante el transporte, no exponer el aparato a humedad.
Embalar el aparato adecuadamente.
Embalar el aparato de tal forma que queda protegido
contra choques o vibraciones (p. ej: embalaje con colchn
de aire).
Segn el tipo de aparato, el centro de gravedad puede no
estar en el centro del equipo.

1.9

1.8 Instrucciones de seguridad para el montaje


Antes de instalar los aparatos hay que asegurarse de que no
presenten daos por transporte inadecuado. Los daos de
transporte deben ser documentados. Todas las reclamaciones
de indemnizacin por daos debern presentarse
inmediatamente, y antes de la instalacin, ante el transportista
competente.

1.10 Instrucciones de seguridad relativas al


funcionamiento
Asegrese, antes de conectar el aparato, de que se cumplen
las condiciones ambientales indicadas en el captulo "Datos
tcnicos" o en la especificacin tcnica.
Cuando sea de suponer que ya no es posible utilizar el
aparato sin peligro, pngalo fuera de funcionamiento y
asegrelo contra arranque accidental.

El sentido de caudal debe corresponder con la


sealizacin (en caso de que exista).
Debe mantenerse el par de apriete mximo en todos los
tornillos de las bridas.
Al montar el aparatos se deben evitar tensiones
mecnicas (torsin, flexin).
Los aparatos bridados deben montarse con contrabridas
planoparalelas.
Los aparatos deben ser aptos para las condiciones de
servicio previstas y deben instalarse con juntas
adecuadas.
En caso de vibraciones de la tubera, utilizar fijaciones
apropiadas para los tornillos y tuercas de las bridas.

Instrucciones de seguridad para la instalacin


elctrica
La conexin elctrica debe efectuarse exclusivamente por
personal tcnico autorizado y de acuerdo con los esquemas
de conexiones.
Deben seguirse las instrucciones para la conexin elctrica
para no deshabilitar el modo de proteccin elctrica.
Poner a tierra el sistema de medida siguiendo las indicaciones
correspondientes.

Fluidos calientes pueden causar quemaduras, por lo que hay


que evitar tocar la superficie del aparato.
Los fluidos agresivos o corrosivos pueden daar las partes
mojadas. Debido a ello, se pueden producir fugas de los
fluidos sometidos a presin.
La fatiga de las juntas de las conexiones a proceso (p. ej.,
racor roscado asptico, Tri-Clamp, etc.) puede provocar fugas
de fluidos a presin.
Juntas planas internas (si existen) pueden fragilizarse por
procesos CIP/SIP.
ADVERTENCIA! Peligro de intoxicacin!
Las tuberas y fluidos pueden contaminarse por
bacterias y sustancias qumicas txicas.
En instalaciones segn EHEDG se deben
observar las indicaciones siguientes:
El vaciado automtico del caudalmetro, necesario segn
el montaje conforme a EHEDG, solo se garantiza con el
montaje en vertical.
Para instalaciones segn la normativa EHEDG, es
absolutamente necesario que la combinacin de conexin
a proceso y juntas realizada por el cliente o propietario
cumpla la normativa EHEDG. Para ello, tenga en cuenta
las indicaciones de la versin actual del documento
EHEDG Position Paper: "Hygienic Process connections to
use with hygienic components and equipment".

8 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

1.11 Valores tcnicos lmite


El aparato se ha concebido para utilizarse exclusivamente
dentro de los valores tcnicos lmite indicados en la placa de
caractersticas y en las especificaciones tcnicas.
Deben mantenerse los siguientes valores tcnicos lmite:
La presin (PS) y temperatura (TS) permitidas del fluido no
deben exceder los valores de presin y temperatura
previstos (p/T-Ratings) (vase el captulo "Datos
tcnicos").
La temperatura de servicio no debe exceder o bajar por
debajo del valor lmite mximo/mnimo previsto.
No debe sobrepasarse la temperatura ambiente permitida.
Durante el uso del aparato deber mantenerse el modo de
proteccin de la caja.
El sensor de caudal no debe instalarse en las
proximidades de campos electromagnticos fuertes, p.
ej., motores, bombas, transformadores, etc. Se debe
mantener la distancia mnima de ~1 m (3,28 ft). Durante el
montaje sobre/en piezas de acero (p. ej., soportes de
acero) debe mantenerse una distancia mnima de 100 mm
(4). (Estos valores se han determinado en conformidad
con la norma IEC801-2 y/o IECTC77B).
1.12 Fluidos permitidos
Para utilizar los fluidos correctamente es necesario que se
observen y sigan las instrucciones siguientes:
Slo deben utilizarse fluidos en los que pueda asegurarse,
segn la tcnica actual o la experiencia de trabajo del
usuario/propietario, que las propiedades fsicas y qumicas
de los materiales en contacto con el fluido no puedan
perjudicarse y, a consecuencia de ello, mermar el tiempo
de servicio previsto.
Por ejemplo, los fluidos que tengan un alto contenido de
cloro pueden causar daos de corrosin invisibles en los
componentes de acero inoxidable, que pueden destruir,
en consecuencia, las partes mojadas y provocar fugas de
fluido. El propietario/usuario deber controlar que los
materiales utilizados sean apropiados para la aplicacin
prevista.
Los fluidos con propiedades desconocidas o los fluidos
abrasivos slo deben utilizarse si el usuario puede
asegurar unas condiciones seguras del aparato mediante
una comprobacin adecuada efectuada con regularidad.
Observe las especificaciones indicadas en la placa de
caractersticas.

1.13 Instrucciones de seguridad para trabajos de


inspeccin y mantenimiento
ADVERTENCIA - Peligro por corriente
elctrica!
Cuando la caja est abierta, la proteccin CEM
no funciona y el usuario no est protegido contra
el riesgo de contacto accidental.
Antes de abrir la caja hay que desconectar la
alimentacin elctrica.
Los trabajos de reparacin y mantenimiento slo deben ser
llevados a cabo por personal especializado debidamente
instruido.
Antes de desmontar el aparato hay que despresurizar el
aparato y, si existen, los conductos y recipientes
adyacentes.
Antes de abrir el aparato se debe controlar si han sido
utilizadas sustancias peligrosas. Es posible que el aparato
contenga restos peligrosos que puedan salir cuando se
abra el aparato.
En cuanto est previsto dentro del marco de responsabilidad
del usuario, debern realizarse inspecciones peridicas para
controlar los siguientes puntos:
las paredes expuestas a la presin / el revestimiento del
aparato a presin
la funcin tcnica de medicin
la estanqueidad
el desgaste (corrosin)
1.14 Devolucin de aparatos
Para el envo de aparatos para reparacin o recalibracin
deber utilizarse el embalaje original o un recipiente apropiado
de transporte.
El aparato debe enviarse acompaado del impreso de reenvo
debidamente rellenado (vase el captulo "Anexo").
Segn la Directiva CE sobre Sustancias Peligrosas, los
propietarios de basuras especiales son responsables de su
correcta eliminacin y deben observar las siguientes
instrucciones:
Todos los aparatos que se enven a ABB tendrn que estar
libres de sustancias peligrosas (cidos, lejas, soluciones,
etc.).
Consulte al Servicio de atencin al cliente (direccin en la
pgina 2) para el establecimiento colaborador ms cercano.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 9

1.15 Sistema de gestin integrado


La ABB Automation Products GmbH dispone de un sistema
de gestin integrado compuesto por:
El sistema de gestin de calidad ISO 9001:2008
El sistema de gestin ambiental ISO 14001:2004
El sistema de gestin de salud y salud ocupacional
BS OHSAS 18001:2007 y
El sistema de gestin de proteccin de datos e
informacin.
La preocupacin por el medioambiente forma parte de la
poltica de nuestra empresa.
Durante la fabricacin, el almacenamiento, transporte, uso y la
explotacin y eliminacin de nuestros productos y soluciones
tcnicas, la carga contaminante al medio ambiente y a las
personas debe minimizarse al mximo.
Esto requiere, en particular, que los recursos naturales se
utilicen con la precaucin debida. Nuestros folletos de
informacin sirven para llevar un dilogo abierto con el
pblico.
1.16 Eliminacin de residuos
El presente producto est compuesto por materiales que
pueden reciclarse en plantas de reciclaje especializadas.

1.16.1 Informacin sobre la directiva WEEE 2002/96/CE


(residuos de aparatos elctricos y electrnicos)
(Waste Electrical and Electronic Equipment)
El presente producto no est sujeto a la directiva WEEE
2002/96/CE ni a las leyes nacionales pertinentes (en Alemania,
p. ej., ElektroG).
El producto usado debe entregarse a una empresa de
reciclaje especializada. Por favor, no utilice los puntos de
recogida de basura habituales. Estos deben utilizarse
solamente para productos de uso privado segn la directiva
WEEE 2002/96/CE. La eliminacin adecuada evita
repercusiones negativas sobre el hombre y el medio ambiente
y permite el reciclaje de materias primas valiosas.
Si no existe ninguna posibilidad de eliminar el equipo usado
debidamente, nuestro servicio posventa est dispuesto a
recoger y eliminar el equipo abonando las tasas
correspondientes.
1.16.2 Directiva ROHS 2002/95/CE
Con la ElektroG, Alemania transpone las Directivas europeas
2002/96/CE (WEEE) y 2002/95/CE (RoHS) en Derecho
nacional. Por un lado, la ElektoG regula cules productos
deben recogerse y eliminarse de forma ordenada o reutilizarse
en caso de eliminacin / al final de su duracin til. Por otro
lado, la ElektroG prohibe la comercializacin de equipos
elctricos y electrnicos que contengan cantidades
determinadas de plomo, cadmio, mercurio, cromo
hexavalente, bifenoles polibromurados (PBB) y difenilos
polibromurados (PBDE) (sustancias prohibidas).
Los productos suministrados por ABB Automation Products
GmbH no estn sujetos al mbito de aplicacin de la
prohibicin de sustancias peligrosas, o bien, la directiva sobre
restricciones en el uso de determinadas sustancias peligrosas
en equipos electrnicos y elctricos usados (segn ElektroG).
En el supuesto de que los componentes necesarios estn
disponibles a tiempo en el mercado, podremos renunciar, en
el futuro, a utilizar estas sustancias en desarrollos nuevos.

10 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Diseo y funcin

Funcin para clcular el fuerza de Coriolis

2.1 Generalidades
Los caudalmetros msicos de ABB Automation Products
funcionan segn el principio de Coriolis.
El modelo equipado con las tuberas de medida paralelas
convencionales se distingue especialmente por su diseo
compacto, el dimetro nominal grande y su ventaja econmica
para el cliente.
2.2 Principio de medicin
Cuando a travs de una tubera vibrante fluyen masas, se
producen fuerzas de Coriolis que deforman la tubera. Estas
deformaciones muy pequeas de la tubera de medida se
registran y se evalan por sensores ptimamente dispuestos.
Como el desplazamiento de fase de las seales del sensor es
proporcional al caudal, el caudalmetro msico coriolis puede
medir directamente el caudal msico transportado a travs de
la tubera de medida. Este principio de medida funciona
independientemente de la densidad, temperatura, viscosidad,
presin y conductividad.
Las tuberas de medida vibran siempre en resonancia. Esta
frecuencia de resonancia depende de la forma geomtrica de
la tubera de medida, de las propiedades del material y de la
masa del fluido vibrante en la tubera de medida. As se
obtiene una informacin extacta sobre la densidad del fluido a
medir.
El aparato tiene incorporada un sensor de temperatura que
sirve para corregir los parmetros dependientes de la
temperatura. Resumiendo puede decirse que el caudalmetro
msico coriolis permite medir al mismo tiempo el caudal
msico, la densidad y la temperatura. Partiendo de estos
valores se pueden calcular otras magnitudes de medida
como, p. ej., el flujo volumtrico o la concentracin.

Signo de frmula

Descripcin
Fuerza de Coriolis
Velocidad angular
Velocidad de la masa

Fig. 2:

Masa

Representacin grfica simplificada del efecto dinmico de


la fuerza de Coriolis

Fig. 2

Descripcin

"A"

Movimiento de las tuberas, hacia adentro, sin caudal

"B"

Direccin de las fuerzas de Coriolis, con caudal y movimiento


de las tuberas hacia afuera

"C"

Movimiento de las tuberas, hacia afuera, sin caudal

"D"

Direccin de las fuerzas de Coriolis, con caudal y movimiento


de las tuberas hacia adentro

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 11

Change from two to one c olumn

2.3

Versiones del aparato


Sensor de caudal FCB300 y FCH300

FCB300 para aplicaciones estndar

FCH300 para aplicaciones higinicas

Brida DIN 2501 / EN 1092-1

DN 10 200, PN 40 100

Brida ASME B16.5

DN 1/4 8 PN CL150 CL600

Racor DIN 11851

DN 10 100 (1/4" 4")

DN 25 80 (1" 3")

Tri-Clamp

DIN 32676 (ISO 2852)

DIN 32676 (ISO 2852)

BPE Tri-Clamp

BPE Tri-Clamp

Nmero de modelo
Conexiones a proceso

DN 10 100 (1/4" 4")

DN 20 100 (3/4" 4")

Otras conexiones

Bajo pedido

Bajo pedido

Materiales de partes mojadas

Acero inoxidable

Acero CrNi pulido 1.4404 (AISI 316L) o 1.4435 (AISI

Nickel-Alloy C4 / C22

316L)

IP 65 / 67, NEMA 4X

IP 65 / 67, NEMA 4X

Zona 0, 1, 2, 21, 22

Zona 0, 1, 2, 21, 22

Proteccin contra explosiones cFMus

Class I Div. 1, Class I Div. 2, Zona 0, 1, 2, 20, 21

Class I Div. 1, Class I Div. 2, Zona 0, 1, 2, 20, 21

Proteccin contra explosiones NEPSI

Zona 0, 1, 2, 21, 22

Zona 0, 1, 2, 21, 22

Homologaciones de normas sobre

EHEDG, conforme a FDA

Tipo de proteccin conforme a


EN 60529
Homologaciones y certificados
Proteccin contra explosiones ATEX /
IECEx

higiene
Otras homologaciones

Bajo pedido

Carcasa

Diseo compacto, diseo remoto

Precisin de medicin de lquidos

FCB330

FCB350

FCH330

FCH350

Caudal msico 1)

0,4 % y 0,25 %

0,1 % y 0,15 %

0,4 % y 0,25 %

0,1 % y 0,15 %

Caudal volumtrico 1)

0,4 % y 0,25 %

0,15 %

0,4 % y 0,25 %

0,15 %

Densidad

0,01 kg/l

0,002 kg/l

0,01 kg/l

0,002 kg/l

Temperatura
Precisin de medicin de gases

1)

Temperatura permitida del fluido

0,001 kg/l (opcional)

0,001 kg/l (opcional)

0,0005 kg/l 2)

0,0005 kg/l 2)

1K

0,5 K

1K

0,5 K

1%

0,5 %

1%

0,5 %

-50 160 C

-50 200 C

-50 160 C

-50 200 C

(-58 320 F)

(-58 392 F)

(-58 320 F)

(-58 392 F)

1) Indicacin de la precisin en % del valor medido (% v. M.)


2) Precisin tras ajuste in situ en condiciones de funcionamiento

12 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Transductor FCTXXX

Carcasa

Diseo compacto

Longitud del cable

10 m (33 pies) como mximo, solo para diseo remoto

Suministro de energa

100 230 V AC, 24 V AC/DC

Salida de corriente

Diseo remoto

Salida de corriente 1: activa, 0/4 20 mA o pasiva, 4 20 mA


Salida de corriente 2: pasiva, 4 20 mA

Salida de impulsos

Activa (no disponible para Zona 1 / Div. 1) o pasiva

Desconexin externa de la salida

Puesta a cero externa del contador

Medicin de caudal de avance / retorno

Comunicacin

Protocolo HART

Deteccin de tubera vaca

S, mediante una alarma de densidad preajustada de < 0,5 kg/l

Supervisin automtica y diagnstico

Visualizacin in situ / contador

Optimizacin de campo, para el caudal y

la densidad
Tipo de proteccin conforme a EN 60529

Diseo compacto: IP 65 / IP 67, NEMA 4X


Diseo remoto: IP 67, NEMA 4X

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 13

2.3.1

Cuadro sinptico del aparato ATEX, IECEx, NEPSI


Estndar / sin proteccin contra

Zona 2, 21, 22

Zona 1, 21 (Zona 0)

FCx3xx A2, S2

FCx3xx A1, S1

explosiones
Nmero de modelo

FCx3xx Y0

Diseo compacto
Estndar
Zona 2, 21, 22
Zona 1, 21
Zona 0

Nmero de modelo

FCT3xx Y0

FCx3xx Y0

FCT3xx A2

FCx3xx A2, S2

FCT3xx A2, A1

FCx3xx A1, S1

Diseo remoto
Transductor de
medicin y sensor de
caudal
Estndar
Zona 2, 21, 22
Zona 1, 21
Zona 0
Nmero de modelo

FCT3xx Y0

FCT3xx A2

FCx3xx A1, S1

Diseo remoto
Transductor de
medicin
Estndar
Zona 2, 21, 22
Sensor de caudal
Zona 1, 21
Zona 0

IMPORTANTE (NOTA)
Para ver informacin detallada, consulte el captulo "Datos tcnicos relevantes para la proteccin contra explosiones conforme
a ATEX / IECEx / NEPSI" o el certificado de homologacin.

14 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

2.3.2

Cuadro sinptico del aparato cFMus


Estndar / sin proteccin contra

Class I Div. 2 Zone 2, 21

Class I Div. 1 Zone 0, 1, 20 ,21

FCx3xx F2

FCx3xx F1

explosiones
Nmero de modelo

FCx3xx Y0

Diseo compacto
Estndar
Class I Div. 2
Class I Div. 1
Zone 2, 21
Zone 1, 21
Zone 0, 20
Nmero de modelo

FCT3xx Y0

FCx3xx Y0

FCT3xx F2

FCx3xx F2

FCT3xx F1

FCx3xx F1

Diseo remoto
Transductor de
medicin y sensor de
caudal
Estndar
Class I Div. 2
Class I Div. 1
Zone 2, 21
Zone 1, 21
Zone 0, 20
Nmero de modelo

FCT3xx Y0

FCT3xx F2

FCx3xx F1

Diseo remoto
Transductor de
medicin
Estndar
Class I Div. 2
Zone 2, 21
Sensor de caudal
Class I Div. 1
Zone 1, 21
Zone 0, 20

IMPORTANTE (NOTA)
Para ver informacin detallada, consulte el captulo "Datos tcnicos relevantes para la proteccin Ex segn cFMus" o el
certificado de homologacin.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 15

Change from one to tw o c olumns

Transporte

3.1 Controles
Inmediatamente despus de desembalarlos hay que
asegurarse de que los aparatos no presenten daos por
transporte inadecuado.
Los daos de transporte deben ser documentados.
Todas las reclamaciones de indemnizacin por daos debern
presentarse inmediatamente, y antes de la instalacin, ante el
expedidor competente.

3.2 Generalidades
Durante el transporte del aparato al punto de medicin debe
tenerse en cuenta:
El centro de gravedad no est en el centro del aparato.
Los aparatos bridados no deben levantarse a travs de la
carcasa del transmisor o de la caja de conexin.

Change from two to one c olumn

Change from one to tw o c olumns

Montaje

4.1 Generalidades
Durante el montaje se deben observar los siguientes puntos:
El sentido del caudal debe corresponder con la
sealizacin (en caso de que exista).
Asegrese al montar los tornillos de la brida de no
sobrepasar el par mximo de apriete.
Al montar el aparatos se deben evitar tensiones
mecnicas (torsin, flexin).
Los aparatos de brida/Wafer deben montarse con
contrabridas planoparalelas y solamente con juntas
apropiadas.
Las juntas utilizadas deben ser de un material resistente al
fluido y a la temperatura del fluido. En el caso de equipos
higinicos "Hygienic Design", usar material estanqueizante
adecuado.
Las juntas no deben penetrar en la zona de flujo, porque
se pueden producir turbulencias que afectan la precisin
del aparato.
La tubera no debe ejercer ninguna fuerza o par de torsin
sobre el aparato.
Los tapones de los pasacables no deben desmontarse
antes de que se monten los cables elctricos.
Asegrese de que las juntas de la tapa de la caja queden
asentadas correctamente. Cerrar la tapa correctamente.
Apretar las uniones roscadas de la tapa.
Transmisores externos deben instalarse en un lugar libre
de vibraciones.
Asegrese de que el transmisor y el sensor de caudal no
estn expuestos directamente a los rayos del sol; instalar
un dispositivo de proteccin contra rayos solares, si es
necesario.
Al instalar el transmisor en un armario de distribucin, es
necesario asegurar una refrigeracin suficiente.

4.2 Sensor de caudal


El aparato se puede instalar en cualquier zona de la tubera,
siempre que se cumplan los requisitos de instalacin.
1. Desmontar las placas protectoras montadas en los lados
izquierdo y derecho del sensor de caudal (si existen).
2. Montar el sensor de caudal de manera que se site
planoparalela y cntricamente entre las tuberas.
3. Colocar las juntas entre las superficies de junta.

16 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

4.3 Transmisor
El transmisor debe instalarse en un lugar que, en lo posible, est libre de vibraciones (vase el captulo "Datos tcnicos"). No
deben sobrepasarse los valores lmite de temperatura indicados y la longitud mxima permitida del cable de seal entre el
transmisor y sensor de caudal.
IMPORTANTE (NOTA)
Asegrese al elegir el lugar de montaje de que el transmisor no est expuesto directamente a los rayos del sol.
Si no es posible evitar la radiacin directa del sol, hay que instalar un dispositivo de proteccin contra rayos
solares.
Deben mantenerse los valores lmite de la temperatura ambiente.
Caja de campo
La caja tiene el modo de proteccin IP 65 / 67, NEMA 4X (EN 60529) y debe fijarse con cuatro tornillos. Medidas: vese Fig. 3 y
Fig. 4.
4.3.1

Transmisor de diseo remoto (opcin F1 o F2)

Fig. 3: Medidas en mm (inch)


1 Caja de campo con ventana | 2 Racor atornillado para cables M20 x 1,5 1/2 NPT |
3 Agujeros de fijacin para el kit de montaje para tubos de 2"; kit de montaje bajo demanda (referencia de pedido: 612B091U07) |
4 Modo de proteccin IP 67

4.3.2 Transmisor de diseo remoto (opcin R1 o R2)


IP 65 / 67, NEMA 4X

Fig. 4:

Medidas en mm (inch)

Change from one to tw o c olumns

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 17

4.4

Orientacin de la caja del transmisor y del indicador


LCD
Segn la posicin de montaje, es posible girar la caja del
transmisor compacto/indicador LCD y orientarla
horizontalmente, para poder leer mejor la pantalla.

4.4.2

Indicador LCD
ADVERTENCIA - Peligro por corriente
elctrica!
Cuando la caja est abierta, la proteccin CEM
no funciona y el usuario no est protegido contra
el riesgo de contacto accidental.
Antes de abrir la caja hay que desconectar la
alimentacin elctrica.

4.4.1 Caja del transmisor


Para girar la caja del transmisor, se deben realizar los pasos
siguientes. Un dispositivo de bloqueo en la caja del transmisor
impide su giro en exceso (>330).

Para girar el indicador LCD, se deben realizar los pasos


siguientes.

Fig. 6:

1.
2.
3.

4.
5.

Desconectar la alimentacin elctrica.


Desmontar la tapa de la caja (1).
Aflojar los cuatro tornillos de fijacin (2) del indicador LCD.
A continuacin, el indicador LCD cuelga del mazo de
cables conectado a la unidad electrnica enchufable.
Fijar el indicador LCD en la posicin deseada. Asegrese
de no daar el mazo de cables.
Volver a montar la tapa de la caja (1).

Fig. 5: Orientacin de la caja del transmisor


1 Tornillo de fijacin | 2 Caja del transmisor

1.
2.
3.

Orientacin del indicador LCD

Aflojar los tornillos de fijacin (2 vueltas,


aproximadamente)
Girar cautelosamente la caja del transmisor, hasta
alcanzar la posicin deseada.
Apretar el tornillo de fijacin.
PELIGRO Peligro de explosin!
Posible riesgo de perjudicar la proteccin contra
explosin.
No desmontar el transmisor del sensor de caudal.

18 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

AVISO Prdida del modo de proteccin de


la caja!
Prdida del modo de proteccin de la caja por
asiento incorrecto o dao de la junta (junta
trica).
Antes de cerrar la tapa de la caja, controlar la
junta (junta trica) y cambiarla, si es necesario.
Asegrese al cerrar la tapa de la caja de que la
junta est asentada correctamente.

4.5 Instrucciones para el montaje


4.5.1 Requisitos de montaje / instrucciones de
planificacin
El CoriolisMaster FCB330, FCB350, FCH330, FCH350 puede
instalarse tanto en el interior como en el exterior. El modelo
estndar tiene el tipo de proteccin IP 67. El sensor de caudal
permite la medicin en ambas direcciones de flujo y puede
instalarse en cualquier posicin. El usuario ha de asegurarse
de que las tuberas de medicin puedan llenarse
completamente en cualquier momento. Adems, el fluido no
debe afectar a la resistencia de las partes mojadas.
Durante la instalacin, se deben observar los siguientes
puntos:
En la direccin de montaje preferida, el caudal fluye por el
sensor en el sentido de la flecha. En este caso, se indica
un caudal positivo (opcionalmente, hay disponible un
dispositivo de calibracin para caudal directo/inverso). La formacin de burbujas de gas en el tubo de medicin
puede aumentar el nmero de errores de medicin,
especialmente durante la medicin de la densidad. Por
ello, el sensor de caudal no debe montarse en el punto
ms alto del sistema. Lo ideal es un lugar de montaje que
se encuentre entre el punto ms bajo del sistema y tenga
una tubera en forma de U.
Asegrese de que los gases disueltos en el fluido no se
liberen y de que las tuberas de medicin estn siempre
totalmente llenas. Para garantizar esto, se recomienda
aplicar una contrapresin de 0,2 bar (2,9 psi) como
mnimo.
En la medicin de gases, asegrese de que los gases
estn secos y libres de humedad.
En caso de presin negativa en la tubera de medicin o
en aplicaciones de lquidos con bajo punto de ebullicin,
hay que asegurarse de que la presin de vapor del fluido
no caiga por debajo del valor lmite exigido.
Asegrese de que no se efectan transiciones de fase en
el fluido durante el funcionamiento. En el caso de fluidos
gaseosos, se debe evitar una fase lquida.
Se recomienda no instalar tuberas largas de cada detrs
del sensor de caudal, a fin de impedir que las tuberas de
medicin puedan vaciarse completamente.
Los dispositivos se pueden montar directamente
desde/hacia tubos angulares, vlvulas u otros
componentes, siempre que no causen cavitacin.
El dispositivo est concebido para un uso industrial. Si el
entorno electromagntico del campo de aplicacin cumple
las prcticas recomendadas de las normas de la
"Declaracin de conformidad CE", no se requieren
medidas especiales. En caso de campos
electromagnticos que superan la medida normal, debe
guardarse suficiente distancia.
Asegrese de que el sensor de caudal no entre en
contacto con otros objetos. El sensor de caudal no debe
fijarse en la carcasa.

En principio, no son necesarios dispositivos de apoyo ni


amortiguadores especiales en el dispositivo.
En el caso de equipos industriales y martimos concebidos
segn las prcticas recomendadas, se amortiguan
suficientemente las fuerzas habituales que se generan en
el dispositivo. Esto tambin es vlido para montaje de
dispositivos en serie o paralelo. Para ello, es
imprescindible usar y montar el dispositivo de la forma
prevista.
Para evitar daos en las conexiones a proceso y las
tuberas por las fuerzas transversales, en el caso de
dispositivos con un gran peso se deben utilizar
dispositivos de apoyo adecuados montados por el
propietario.
4.5.2 Tramos de entrada
El sensor de caudal no necesita tramos de entrada. Asegrese
de que las vlvulas, compuertas, mirillas etc. instaladas en la
proximidad del sensor de caudal no se caviten y no estn
expuestas a vibraciones causadas por el sensor de caudal.
4.5.3 Modelos de diseo remoto
Asegrese de que el sensor de caudal y el transmisor sean
compatibles. Las placas de caractersticas de los aparatos
compatibles tienen los mismos nmeros finales, p. ej., X001 y
Y001 o X002 y Y002.
4.5.4 Prdida de presin
La prdida de presin vara en funcin de las propiedades del
fluido y el caudal actual.
Para obtener ayuda para el clculo de la prdida de presin,
descargue el archivo correspondiente de la pgina
www.abb.com/flow.
4.6 Posiciones de montaje
El caudalmetro funciona en cualquier posicin de montaje. La
posicin de montaje ptima es en vertical con flujo en sentido
ascendente.
IMPORTANTE (NOTA)
El vaciado automtico del caudalmetro, necesario segn el
montaje conforme a EHEDG, solo se garantiza con el montaje
en vertical.
4.6.1

Fig. 7:

Montaje en vertical, en una tubera ascendente

Montaje en vertical, con vaciado automtico

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 19

4.6.2 Montaje en vertical, en una tubera descendente


Asegrese de que el sensor de caudal est completamente
lleno durante el proceso de medicin.
Para ello, es necesario instalar un estrangulador u obturador
debajo del sensor de caudal. La seccin transversal del
estrangulador u obturador debe ser inferior a la de la tubera,
para impedir que el sensor de caudal se vace durante el
proceso de medicin.

4.6.3

Fig. 9:

4.6.4

Montaje en horizontal para la medicin de lquidos

Montaje en horizontal (lquidos)

Montaje en horizontal para la medicin de gases

Fig. 10: Montaje horizontal (gases)

Para la medicin de gases, el transductor o la caja de


conexiones deben apuntar hacia abajo.

Fig. 8: Montaje en vertical, en una tubera descendente


1 Depsito de reserva | 2 Sensor de cauda l
3 Estrangulador u obturador | 4 Vlvula | 5 Recipiente de recogida

20 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

4.6.5

Puntos de montaje problemticos para la medida


de lquidos
En la medida de lquidos hay que tener en cuenta que
acumulaciones de aire o burbujas de gas en la tubera de
medida pueden tener gran influencia sobre la precisin de
medida.
Para la medida de lquidos, se recomienda no instalar el
aparato en los puntos siguientes:

4.6.7 Ajuste del cero


Los dispositivos de la serie CoriolisMaster no necesitan
obligatoriamente un ajuste del punto cero. Solo se recomienda
utilizar un ajuste del punto cero en los siguientes casos:
Para mediciones en la zona de flujo inferior (por debajo del
10 % de QmxDN),
Si son necesarias precisiones altas (0,1 % o mejor),
Si las condiciones de funcionamiento (presin y
temperatura) difieren ampliamente de las condiciones de
referencia.

Fig. 11: Puntos de montaje problemticos

Fig. 13: Tubera de derivacin

"A": Si el aparato se instala en el punto ms alto de la


tubera, se pueden formar acumulaciones de aire o
burbujas de gas en la tubera de medida, las cuales
afectan la presin de medida.
"B": Si el sensor de caudal se instala en una tubera
descendente, no est garantizado que la tubera de
medida pueda llenarse completamente durante el proceso
de medicin. Esto puede influir muy negativamente sobre
la precisin de medida.

Para el ajuste del punto cero, es necesario un montaje de los


dispositivos de cierre delante (A) y detrs (B) del dispositivo
Se recomienda el montaje de una tubera de derivacin. As se
puede realizar el ajuste del punto cero durante un proceso en
curso.

4.6.6

Puntos de montaje problemticos para la medida


de gases
En la medida de gases hay que tener en cuenta que las
acumulaciones de lquido o de condensado en la tubera de
medida pueden tener gran influencia sobre la precisin de
medida.
Para la medida de gases, se recomienda no instalar el aparato
en los puntos siguientes:

Requisitos necesarios para el ajuste del cero en condiciones


de servicio:
La tubera de medida est completamente llena.
La tubera de medida no contiene burbujas de gas o aire
(medida de lquidos).
La tubera de medida no contiene condensados (medida
de gases).
La presin y temperatura en la tubera de medida
corresponden a las condiciones de servicio normales y
son estables.
En el caso de un punto cero elevado (> 0,1 %), compruebe si
la instalacin se ha realizado correctamente y asegrese de
que no hay componentes gaseosos en los fluidos, ni fluidos o
partculas en los gases. Asegrese de que el dispositivo est
completamente lleno.

Fig. 12: Puntos de montaje problemticos

Si el aparato se instala en el punto ms bajo de la tubera, se


pueden formar acumulaciones de lquidos o de condensado
en la tubera de medida, las cuales afectan la presin de
medida.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 21

4.6.8 Instalacin en funcin de la temperatura del fluido


La posicin de montaje del sensor de caudal depende de la
temperatura del fluido Tmedium. Existen las variantes de
montaje siguientes:

4.6.9

Instalacin en caso de uso de la opcin TE1


"Longitud ampliada de la torre"

IMPORTANTE (NOTA)
El sensor de caudal solo se debe aislar en combinacin con la
opcin TE1 "Longitud ampliada de la torre", como se muestra
en Fig. 16.

Fig. 14: Posicin de montaje para Tmedium -50 120 C (58 248 F)

Fig. 16: Posicin de montaje para Tmedium -50 200 C (58 392 F)
1 Aislamiento

4.6.10 Notas sobre la conformidad EHEDG


ADVERTENCIA! Peligro de intoxicacin!
Las tuberas y fluidos pueden contaminarse por
bacterias y sustancias qumicas txicas.
En instalaciones segn EHEDG se deben
observar las indicaciones siguientes:
Fig. 15: Posicin de montaje para Tmedium -50 200 C (58 392 F)

El vaciado automtico del caudalmetro, necesario segn


el montaje conforme a EHEDG, solo se garantiza con el
montaje en vertical.
Para instalaciones segn la normativa EHEDG, es
absolutamente necesario que la combinacin de conexin
a proceso y juntas realizada por el cliente o propietario
cumpla la normativa EHEDG. Para ello, tenga en cuenta
las indicaciones de la versin actual del documento
EHEDG Position Paper: "Hygienic Process connections to
use with hygienic components and equipment".

22 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Conexiones elctricas

5.1

Instrucciones para conectar la alimentacin elctrica


IMPORTANTE (NOTA)
Se deben mantener los valores lmite de la
alimentacin elctrica (vase el captulo
"Datos tcnicos").
Tener en cuenta la cada de tensin, si se
utilizan cables largos con dimetro pequeo.
La tensin conectada a los terminales del
aparato no debe bajar por debajo del valor
mnimo necesario.
Realizar la conexin elctrica siguiendo los
esquemas de conexin.

La placa de caractersticas del transmisor indica la tensin de


conexin y el consumo de corriente.
En la lnea de alimentacin elctrica del transmisor se debe
instalar un cortacircuito automtico con una corriente nominal
mxima de 16 A.
El dimetro del cable de alimentacin y el cortacircuito
automtico utilizado deben cumplir la norma VDE 0100 y
corresponder al consumo de corriente del caudalmetro
instalado. Las lneas deben ser conformes a IEC 227 o IEC
245.
Se recomienda instalar el cortacircuito automtico cerca del
transmisor y marcarlo como parte del aparato.
La conexin de la alimentacin elctrica se realizar segn las
especificaciones indicadas en la placa de caractersticas, a
travs de los terminales L (fase), N (cero) o 1+, 2- y PE.
El transmisor y el sensor de caudal deben conectarse a tierra.

5.2 Indicaciones para el cableado


Al instalar los cables de conexin en el sensor de caudal hay
que preveer un lazo de goteo (trampa de agua).
En caso de montaje vertical del sensor de caudal, orientar
hacia abajo la entrada de cables. Girar la caja del transmisor,
si es necesario.

Fig. 17: Colocacin de los cables de conexin


1 Lazo de goteo

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 23

5.3 Diseo compacto


En los modelos de diseo compacto, los terminales de
conexin se encuentran detrs de la tapa, en la cara posterior
de la caja del transmisor.
La cara interior de la tapa muestra el esquema de conexin.
La configuracin del aparato se marcar.

IMPORTANTE (NOTA)
Para conectar los cables correctamente se deben
utilizar virolas de cable apropiadas.
Conexin del aparato:
1. Desmontar la tapa del compartimento de conexin.
2 Confeccionar los terminales del cable e introducirlos en el
compartimento de conexin (pasndolos por las entradas
de cables).
3. Desmontar la tapa de los terminales y conectar el cable
de alimentacin siguiendo los esquemas de conexin
correspondientes.
4. Volver a montar la tapa de los terminales.
5. Conectar el cable para las entradas y salidas de seal
siguiendo los esquemas de conexin. Conectar los
apantallamientos del cable (si existen) utilizando al efecto
la abrazadera de puesta a tierra correspondiente.
6. Volver a montar la tapa del compartimento de conexin.
AVISO Prdida del modo de proteccin de
la caja!
Prdida del modo de proteccin de la caja por
asiento incorrecto o dao de la junta (junta
trica).
Antes de cerrar la tapa de la caja, controlar la
junta (junta trica) y cambiarla, si es necesario.
Asegrese al cerrar la tapa de la caja de que la
junta est asentada correctamente.

Fig. 18: Terminales de conexin


1 Tapa del compartimento de conexin |
2 Empleo de las conexiones | 3 Entradas de cables |
4 Terminales de conexin para la alimentacin elctrica |
5 Terminales de conexin para las entradas y salidas de seal |
6 Tapa de los terminales

24 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

5.4 Diseo remoto


En caso de los modelos de diseo remoto, el transmisor se
montar por separado. Su conexin al sensor de caudal se
realizara a travs de un cable de seal.
5.4.1

5.4.3

Conexin del cable de seal


IMPORTANTE (NOTA)
Para conectar los cables correctamente se deben
utilizar virolas de cable apropiadas.

Especificacin del cable

Cable de seal
Denominacin

LI2YCY PiMF
5 x 2 x 0,5 mm2

Apantallamiento

Par blindado con conductor de tierra y


pantalla de trenza de cobre

Rango de temperatura

-30 ... 70 C (-22 ... 158 F)

Resistencia de bucle

mximo: 78,4 /km

Inductividad

~ 0,4 mH/km

Longitud mx. del cable

10 m (33 ft)

5.4.2 Colocacin del cable de seal


Durante la colocacin del cable deben observarse los
siguientes puntos:
El cable de seal conduce una seal de tensin de slo
unos milivoltios y, por lo tanto, debe ser tan corto como
sea posible. La longitud mxima permitida del cable de
seal es de 10 m (33 ft).
Evite colocar el cable cerca de mquinas elctricas
grandes y elementos de conmutacin que pueden
producir interferencias, impulsos de conexin e
inducciones. Si esto no es posible, instalar el cable de
seal en un tubo metlico de proteccin y conectar el
tubo al potencial de tierra de la red.
Para aislarlo contra interferencias magnticas, el cable
dispone de una pantalla exterior conectada al potencial de
tierra de la red.
Debe evitarse que el cable de seal discurra cerca de
cajas de unin o regletas de bornes.

Fig. 19: Confeccin del cable de seal, medidas en mm (inch)


1 Pantalla trenzada |
2 Conductores de tierra de las pantallas de lmina (retorcidos) |
3 Pantalla de lmina | 4 Abrazadera de puesta a tierra |
5 Conductor de tierra | 6 Virolas de cable

1.
2.
3.
4
5.

6.
7.

8.
9.

Pelar el cable de seal como se muestra en la figura.


Cortar unos 15 mm (0,59 inch) de la pantalla trenzada.
Separar el alma del cable y la pantalla de lmina de los
pares de conductores.
Pelar los conductores y dotarlos de virolas de cable.
Retorcer los conductores de tierra de las pantallas de
lmina y envolver con ellos la pantalla trenzada. Para la
conexin al aparato, sujetar debajo de la abrazadera de
puesta a tierra la pantalla trenzada y los conductores de
tierra retorcidos.
Conectar al transmisor y sensor de caudal el cable de
seal siguiendo los esquemas de conexin.
Conectar al transmisor el cable para las entradas y salidas
de seal siguiendo los esquemas de conexin. Conectar
los apantallamientos del cable a la abrazadera de puesta a
tierra correspondiente.
Conectar al transmisor el cable de alimentacin siguiendo
los esquemas de conexin.
Volver a montar todas las tapas de los compartimentos de
conexin del transmisor y sensor de caudal.
AVISO Prdida del modo de proteccin de
la caja!
Prdida del modo de proteccin de la caja por
asiento incorrecto o dao de la junta (junta
trica).
Antes de cerrar las tapas de la cajas, controlar la
junta (junta trica) y cambiarla, si es necesario.
Asegrese al cerrar las tapas de que la junta est
asentada correctamente.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 25

5.5 Comunicacin digital


5.5.1 Protocolo HART
El aparato est registrado en la HART Communication
Foundation.

Integracin en el sistema:
En combinacin con el software DTM disponible para el
aparato (Device Type Manager), la comunicacin
(configuracin, parametrizacin) puede realizarse mediante las
aplicaciones de tramas correspondientes segn FDT 0.98
1.2 (DSV401 R2).
Otras formas de integracin en el sistema/herramientas (p.ej.:
Emerson AMS / Siemens SCS7) pueden suministrarse bajo
demanda.
Los archivos DTM y otros archivos necesarios se pueden
descargar de la pgina www.abb.com/flow.

Fig. 20: Comunicacin mediante el protocolo HART


Protocolo HART
Configuracin

Directamente en el aparato
Mediante software DSV401 + HART-DTM

Transmisin

Modulacin FSK sobre la salida de corriente de


4 ... 20 mA, segn estndar Bell 202

Velocidad en baudios

1200 baud

Indicacin

Lgico 1: 1200 Hz

Amplitud mxima de la

1,2 mAss

Lgico 0: 2200 Hz
seal
Carga en la salida de

250 560

corriente

(en la zona Ex: mx. 300 )

Cable
Diseo

Cable de dos conductores AWG 24, retorcido

Longitud mxima

1500 m (4921 ft)

Para informacin detallada, vase la descripcin de la interfaz


correspondiente.

26 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

5.6 Esquemas de conexin


5.6.1 Conexin de modelos de transductor a equipos perifricos
Modelos FCB330, FCB350, FCH330, FCH350, FCT330, FCT350

Fig. 21
1 Suministro de energa | 2 Salida de corriente 1 | 3 Salida de corriente 2 | 4 Salida de impulsos | 5 Salida de contacto digital |
6 Entrada de contacto digital | 7 Conexin equipotencial (PA)

IMPORTANTE (NOTA)
Si el aparato se utiliza en zonas potencialmente explosivas, se deben respetar las indicaciones de conexin adicionales del
captulo "Datos tcnicos relevantes para la proteccin contra explosiones".
Terminal

Funcin

L / N / PE

Suministro de energa, 100 230 V AC, 50/60 Hz

1+ / 2- / PE

Suministro de energa
24 V AC, 50/60 Hz
24 V DC

31 / 32

Salida de corriente 1, activa


0/4 20 mA , (0 R B 560 , FCT300-A1/F1:1 0 RB 300 )
Salida de corriente 1, pasiva
4 20 mA (0 RB 600 ), tensin de fuente 12 Uq 30 V

33 / 34

Salida de corriente 2, pasiva


4 20 mA (0 RB 600 ), tensin de fuente 12 Uq 30 V

51 / 52

Salida de impulsos, pasiva


fmx = 5 kHz, ancho de impulso = 0,1 2000 ms, 0,001 ... 1000 impulsos/unidad
"cerrada": 0 V UCEL 2 V, 2 mA I CEL 220 mA
"abierta": 16 V UCEH 30 V DC, 0 mA I CEH 0,2 mA
Salida de impulsos activa, U = 16 30 V, carga 150 , fmx = 5 kHz

41 / 42

Salida de contacto digital, pasiva


"cerrada": 0 V UCEL 2 V, 2 mA I CEL 220 mA
"abierta": 16 V UCEH 30 V DC, 0 mA I CEH 0,2 mA

81 / 82

Entrada de contacto digital, pasiva


Entrada "activada": 16 V UKL 30 V
Entrada "desactivada": 0 V UKL 2 V

Conexin equipotencial "PA"


Si se conecta el transductor FCT300 al sensor de caudal FCB3xx / FCH3xx, el transductor tambin se debe conectar a la conexin
equipotencial "PA".

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 27

5.6.2 Ejemplos de conexin con perifricos


Salidas de corriente (incl. comunicacin HART)

Fig. 22: Salidas de corriente activa / pasiva


"A" activa | "B" pasiva | I interna | E externa

Salida y entrada de contacto digitales

Fig. 23
"A" Salida del control del sistema, alarmas Mx./Mn. tubera de medida vaca o sealizacin de directo/inverso |
"B" Entrada para la puesta a cero externa del totalizador o desconexin externa de la salida | I interna | E externa

Salida de impulsos

Fig. 24: Salida de impulsos activa / pasiva


"A" activa | "B" pasiva (optoacoplador) | I interna | E externa

28 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

5.6.3 Conexin del transductor al sensor de caudal


Transductor FCT330, FCT350 al sensor de caudal FCB330, FCB350, FCH330, FCH350

Fig. 25
1 Conexin equipotencial (PA)
Terminal

Color del conductor

Funcin

Terminal

Color del conductor

85

Blanco

Sensor A

91

Gris

Funcin
Excitador

86

Marrn

Sensor A

92

Rosa

Excitador

87

Verde

Sensor B

93

No se utiliza

88

Amarillo

Sensor B

94

No se utiliza

89

Negro

Temperatura

95

Azul

Temperatura

90

Violeta

Temperatura

96

Rojo

Temperatura

IMPORTANTE (NOTA)
La posicin exacta de los terminales de conexin equipotencial puede variar segn el tipo de dispositivo. Los terminales estn
marcados de forma acorde. Si se conecta el transductor FCT330, FCT350 al sensor de caudal FCB330, FCT350, FCH330,
FCH350, tambin debe conectarse a la conexin equipotencial "PA".
Estn permitidas las siguientes combinaciones de sensor de caudal y transductor:
Sensor de caudal FCB330, FCH330 con transductor FCT330
Sensor de caudal FCB350, FCH350 con transductor FCT350

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 29

5.6.4 Conexin del transductor al sensor de caudal en Zona 1 / Div. 1


Transductor FCT330, FCT350 al sensor de caudal FCB330, FCB350, FCH330, FCH350

Fig. 26
1 Conexin equipotencial (PA)
Terminal

Color del conductor

Funcin

Terminal

Color del conductor

85

Blanco

Sensor A

91

Gris

Funcin
Excitador

86

Marrn

Sensor A

92

Rosa

Excitador

87

Verde

Sensor B

93

No se utiliza

88

Amarillo

Sensor B

94

No se utiliza

89

Negro

Temperatura

95

Azul

Temperatura

90

Violeta

Temperatura

96

Rojo

Temperatura

IMPORTANTE (NOTA)
Se deben conectar conductores de pares para garantizar la proteccin CEM.
Estn permitidas las siguientes combinaciones de sensor de caudal y transductor:
Sensor de caudal FCB330, FCH330 con transductor FCT330
Sensor de caudal FCB350, FCH350 con transductor FCT350
Change from one to tw o c olumns

30 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Puesta en marcha

6.3

6.1 Controles antes de la puesta en funcionamiento


Antes de la puesta en servicio se debern controlar los
siguientes puntos:
La asignacin correcta del transmisor al sensor de caudal.
El cableado correcto descrito en el captulo "Conexiones
elctricas".
La puesta a tierra correcta del sensor de caudal.
El nmero de serie del mdulo de almacenamiento de
datos (FRAM externo) debe corresponder con el nmero
de serie del sensor de caudal.
El mdulo de almacenamiento de datos (FRAM externo)
debe estar enchufado en la ranura prevista (vase el
captulo "Mantenimiento / Reparacin").
Las condiciones ambientales deben corresponder con los
valores indicados en la especificacin tcnica.
La alimentacin elctrica debe corresponder con los datos
indicados en la placa de caractersticas.
6.2 Conectar la alimentacin de corriente
Conectar la alimentacin de corriente.
Despus de conectar la alimentacin de corriente, los datos
del sensor memorizados en la FRAM externo se compararn
con los valores guardados internamente.
Si los datos no son idnticos, los datos del transmisor se
intercambiarn automticamente. Poco despus aparecer el
mensaje "Ext. Data loaded". El caudalmetro est listo para el
funcionamiento.
La pantalla LCD indicar el caudal actual.
6.2.1

Medidas de control despus de conectar la


alimentacin de corriente
Despus de la puesta en servicio se debern controlar los
siguientes puntos:
La configuracin de los parmetros debe corresponder
con las condiciones ambientales.
Debe haberse realizado un ajuste del cero.
Informaciones generales:
En el caso de que el indicador de caudal seale que el
sentido de flujo sea incorrecto, pueden haberse
confundido los conductores del cable de seal entre el
sensor de caudal y el transmisor.
La posicin y los valores de los fusibles se indican en la
lista de repuestos.

Ajustes bsicos
IMPORTANTE (NOTA)
Para informacin detallada sobre el uso del
indicador LCD, vase el captulo "Configuracin,
parametracin / Manejo".
Para una descripcin detallada de todos los
mens y parmetros, vase el captulo
"Configuracin, parametracin / Descripcin de
los parmetros".

Si el cliente lo desea, el aparato se entregar ajustado a los


valores especificados. Cuando falten estos datos, el aparato
se entregar con los ajustes de fbrica.
Para ajustar el aparato in situ, basta seleccionar o introducir
unos pocos parmetros.
Antes de la puesta en servicio del aparato es necesario que se
controlen y ajusten los parmetros siguientes:
Valor lmite superior del rango de medida
(Parmetro "QmMax" y submen "Unit").
El aparato est ajustado por defecto al valor lmite mximo
posible del rango de medida, siempre que el cliente no
especifique otros valores.
Salidas de corriente
(Submen "Current output 1" y "Current output 2").
Seleccionar el rango de corriente deseado (0 20 mA 4
20 mA).
Salida de impulsos
(Parmetro "Pulse" y submen "Unit").
Para ajustar el nmero de impulsos por cada unidad
volumtrica, hay que seleccionar primero, en el submen
"Unit", la unidad del totalizador (p. ej., kg. o t). A continuacin
hay que introducir, bajo el parmetro "Pulse", el nmero de
impulsos deseado.
Ancho de impulso
(Parametro "Pulse width").
Para el procesamiento externo de los impulsos de contaje se
puede ajustar un ancho de impulso de entre 0,1 y 2000 ms.
Ajuste del cero del sistema
(Submen "System Zero adj.").
Para ello es necesario detener completamente el lquido
contenido en el sensor de caudal. El sensor de caudal debe
estar completamente lleno. Seleccionar el men "System Zero
adj.". Pulsar la tecla ENTER. Pulsar la tecla STEP para llamar
la opcin "System Zero adj. Function automatic?" y pulsar
ENTER para activar el ajuste automtico. Se puede elegir un
mtodo de ajuste lento o rpido. El mtodo lento es
normalmente ms apropiado para obtener un ajuste ms
preciso del cero.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 31

6.4 Configuracin de la salida de impulsos


La configuracin (activa, pasiva) de la salida de impulsos se
establece en el transductor mediante un puente enchufable.
Para modificar la configuracin, se debe sacar de la carcasa la
unidad de enchufe del transductor.

6.5

Interruptor de proteccin contra modificaciones no


autorizadas

Fig. 28: Interruptor de proteccin contra modificaciones no


autorizadas

Fig. 27: Posicin del puente enchufable


1 Puente enchufable para configurar la salida de impulsos
Posicin

Funcin

"A"

Salida de impulsos 51 / 52 pasiva

"B"

Salida de impulsos 51 / 52 activa (no disponible en


modelos con proteccin contra explosiones)

IMPORTANTE (NOTA)
Tenga en cuenta que, en los dispositivos con
certificado para Zona 1 / Div. 1, el puente
enchufable est en la posicin "B" (activa),
aunque la salida de impulsos est configurada
como pasiva.

Adicionalmente a la proteccin por contrasea, hay la


posibilidad de activar la proteccin de hardware contra
escritura.
El bloqueo de programacin puede activarse o desactivarse
girando el interruptor (1) en el sentido o contra el sentido de
las agujas del reloj.
En el caso de que se deban modificar parmetros y el bloqueo
de programacin est activado, aparecer el mensaje de
advertencia: Operating protection" y el programa denegar el
acceso.
En tal caso, incluso es posible sellar la tapa del aparato de
diseo compacto mediante un tornillo de retencin con
agujero, para impedir que modificaciones de los parmetros
relevantes para la calibracin se queden sin descrubrir.

32 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

6.6

Instrucciones para el funcionamiento seguro en


zonas potencialmente explosivas (ATEX)
6.6.1 Controles
PELIGRO Peligro de explosin!
Peligro de explosin al abrir la carcasa.
Antes de abrir la carcasa se deben observar los
puntos siguientes:
El propietario debe disponer de un certificado
que autorice la utilizacin de fuego.
Asegrese de que no haya peligro de
explosin.
Antes de abrir la caja hay que desconectar la
alimentacin elctrica.
ATENCIN - Peligro de quemadura!
Peligro de quemadura por contacto con fluidos
calientes. No tocar el sensor de caudal! La
temperatura superficial depende de la
temperatura del fluido y puede alcanzar ms de
70 C (158 F)!
Asegrese antes de realizar trabajos en el sensor
de caudal de que el aparato se haya enfriado
suficientemente.
La puesta en servicio tendr que realizarse de acuerdo con el
ElexV (reglamento sobre instalaciones elctricas en zonas
potencialmente explosivas) y la norma EN 60079-14 (montaje
de instalaciones elctricas en zonas potencialmente
explosivas) o las disposiciones nacionales pertinentes.
El montaje, la puesta en funcionamiento y los trabajos de
mantenimiento y reparacin en la zona Ex deben ser
realizados, exclusivamente, por personal debidamente
cualificado.
La puesta en funcionamiento descrita en este documento se
realizar despus del montaje y la conexin elctrica del
caudalmetro instalado.
La alimentacin elctrica est desconectada.
El uso de polvos inflamables est sujeto a la norma
EN 61241-0:2006.
Por favor consulte la explicacin "3KXF002126G0009" del
anexo.

6.6.2 Circuitos elctricos de salida


Instalacin con seguridad intrnseca "i" o seguridad
elevada "e"
Los circuitos elctricos de salida estn diseados de manera
que puedan conectarse a circuitos con o sin seguridad
intrnseca.
No est permitido combinanar circuitos elctricos con y sin
seguridad intrnseca.
A lo largo de la seccin de la lnea del circuito intrnsicamente
seguro deber establecerse una conexin equipotencial.
La tensin de clculo de los circuitos elctricos sin seguridad
intrnseca es UM = 60 V.
IMPORTANTE (NOTA)
En el estado de entrega, los racores atornillados
para cables estn diseados en color negro. En el
caso de que las salidas de seal se conecten a
circuitos intrnsicamente seguros, se recomienda
que para la entrada de cable se utilice la tapa de
color azul claro, que se encuentra en el
compartimento de conexin.
IMPORTANTE (NOTA)
Los datos tcnicos de seguridad de los circuitos
elctricos intrnsecamente seguros pueden verse
en el Certificado CE de homologacin de
modelos de construccin.
Asegrese de que la tapa del enchufe de alimentacin de
corriente est cerrada correctamente. Los circuitos
elctricos de salida intrnsecamente seguros permiten
abrir el compartimento de conexin.
Se recomienda que los racores atornillados para cables
que van incluidos en el suministro (no disponible para la
versin de -40 C (-40 F)), se utilicen para los circuitos
elctricos de salida, de acuerdo con el tipo de proteccin
'e' pertinente. Con seguridad intrnseca: azul. Sin
seguridad intrnseca: negro
El sensor y la caja del transmisor deben conectarse a la
conexin equipotencial. A lo largo de los circuitos
elctricos de las salidas de corriente intrnsecamente
seguras deber establecerse una conexin equipotencial.
Despus de desconectar la alimentacin de corriente y
antes de abrir la caja del transmisor es necesario observar
un tiempo de espera de t > 2 min.
Asegrese de que la puesta en servicio se realice de
acuerdo con la norma EN61241-1:2004 para aplicaciones
en zonas con polvo inflamable.
El propietario/usuario tendr que asegurar, cuando
conecte el conductor protector (PE), que en caso de fallo
no se produzcan diferencias de potencial entre el
conductor protector (PE) y la conexin equipotencial (PA).
Durante el uso en las zonas Ex-polvo, la temperatura
superficial mxima es de 85 C (185 F).
Es posible que la temperatura de proceso del conducto
conectado sobrepase 85 C (185 F).

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 33

6.6.3

Contacto NAMUR

6.6.6

Uso en la zona 2, con la clase de proteccin "a


prueba de vapor" (nR)
La carcasa del transductor de medida (rectangular o redonda,
diseo compacto o remoto) est aprobada para el uso en la
zona 2, con la clase de proteccin "a prueba de vapor" (nR).
ADVERTENCIA Prdida del modo de
proteccin
Despus de los trabajos de instalacin o
mantenimiento o despus de abrir la carcasa, el
usuario debe someter el aparato a un control
segn la norma IEC 60079-15 (vase captulo
"Indicaciones importantes para el control del
aparato").

Fig. 29: Posicin de los puentes enchufables


"A" Conexin estndar | "B" Conexin NAMUR
1 Puente enchufable BR902 | 2 Puente enchufable BR901

Puente

Posicin

Funcin

BR902

Configuracin estndar preferida para

BR901

Ex "e" (configuracin por defecto)

BR902

Configuracin NAMUR preferida para Ex "i"

BR901

enchufable

Los puentes enchufables permiten conectar un amplificador


NAMUR, es decir, la salida de contacto y salida de impulsos
(terminales 41, 42 y 51, 52) pueden conectarse internamente
de manera que funcionen como contacto NAMUR.
6.6.4 Entradas de cables
Indicaciones especiales para modelos con certificado
para Amrica del Norte.
Los modelos con certificado para Amrica del Norte slo
estn equipados con roscas 1/2 NPT (sin racor atornillado
para cables).

Indicaciones importantes para el control del aparato


Obsrvense los siguientes puntos conforme a la norma IEC
60079-15, captulo 23.2.3.2.1.2 "Requisitos para el control
rutinario de carcasas a prueba de vapor; aparatos sin
conexin de prueba":
En condiciones de temperatura constante, el periodo de
tiempo en el que se reduce a la mitad la presin de la
carcasa de al menos 0,3 kPa (30 mm.c.d.a.) no puede ser
inferior a 180 segundos.
Como alternativa, se pueden emplear tambin los siguientes
procedimientos de control con el fin de reducir su tiempo:
En condiciones de temperatura constante, el periodo de
tiempo en el que se reduce la presin de la carcasa de
0,3 kPa (30 mm.c.d.a.) a 0,27 kPa (27 mm.c.d.a.) no
puede ser inferior a 27 segundos.
En condiciones de temperatura constante, el periodo de
tiempo en el que se reduce la presin de la carcasa de
3,0 kPa (300 mm.c.d.a.) a 2,7 kPa (270 mm.c.d.a.) no
puede ser inferior a 27 segundos.

6.6.5 Aislamiento del sensor de caudal


Si es necesario aislar el sensor de caudal, se debern seguir
las indicaciones del captulo "Montaje / Posiciones de montaje
/ Montaje con uso del componente opcional TE1 "Longitud
ampliada del torre".

34 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

IMPORTANTE (NOTA)
Si durante el control surge algn problema con la
presin ms baja (0,3 kPa (30 mm.c.d.a.)), puede
realizarse el control con la presin 10 veces ms
alta (3,0 kPa (300 mm.c.d.a.)).

Cmo realizar el control


1. Espere al menos dos minutos antes de conectar el
suministro de energa y de abrir la carcasa.
2. Retire un racor de cables no utilizado. Normalmente se
utilizan racores de cables con certificado ATEX o IECEx
(p. ej., M20 x 1,5 o roscas NPT 1/2").
3. Conecte el equipo de ensayo para la prueba de presin en
lugar del racor de cables extrado. Asegrese de que el
equipo de ensayo est instalado y sellado correctamente.
4. Realice el control con el equipo de ensayo (vase captulo
"Indicaciones importantes para el control del aparato").
5. Retire el equipo de ensayo y vuelva a montar
correctamente el racor de cables.
Antes de conectar el suministro de energa, se deber realizar
una inspeccin visual de la carcasa, las selladuras, las roscas
y las boquillas de paso. Asegrese de que no se hayan
producido daos.

IMPORTANTE (NOTA)
Para la impermeabilizacin de la carcasa deben
utilizarse exclusivamente repuestos originales.
Los repuestos pueden adquirirse a travs del
Servicio de ABB:
Consulte al Servicio de atencin al cliente
(direccin en la pgina 2) para el establecimiento
colaborador ms cercano.
IMPORTANTE (NOTA)
Asegrese al elegir el lugar de montaje de que el
transductor de medida no est expuesto
directamente a los rayos del sol.
Si no es posible evitar la radiacin directa del sol,
hay que instalar un dispositivo de proteccin
contra rayos solares.
Deben mantenerse los valores lmite de la
temperatura ambiente.

Change from two to one c olumn

6.6.7 Cambio de tipo de proteccin


Para la instalacin en DIV. 1 / Zona 1 se pueden operar las salidas de seal INPUT / OUTPUT de los modelos FCB330/350,
FCH330/350 y FCT330/350 con modos de proteccin diferentes:
Salida de seales INPUT / OUTPUT en modo autoprotegido ia(ib) / IS
Salida de seales INPUT / OUTPUT en modo no autoprotegido
Instalacin original

Instalacin nueva

Pasos de control necesarios

DIV. 1 / Zona 1:

DIV. 1 / Zona 1:

500 V AC/1 mn. o 500 x 1,414 = 710 V DC/1 mn.

Salida de seales INPUT /

Salida de seales INPUT /

OUTPUT en modo no

OUTPUT en modo

autoprotegido

autoprotegido ia(ib) / IS

Prueba entre los terminales 31 / 32, 33 / 34, 41 / 42, 51 / 52, 81 / 82 y/o 97 / 98 y los
terminales 31, 32, 33, 34, 41, 42, 51, 52, 81, 82, 97, 98 y entre la carcasa.
Inspeccin visual, en especial de las tarjetas electrnicas.
Inspeccin visual: no deben presentarse daos o daos por explosin.

DIV. 1 / Zona 1:

DIV. 1 / Zona 1:

Inspeccin visual: las roscas no deben presentar daos (tapa, racores de cables

Salida de seales INPUT /

Salida de seales INPUT /

NPT 1/2").

OUTPUT en modo

OUTPUT en modo no

autoprotegido ia(ib) / IS

autoprotegido

IMPORTANTE (NOTA)
Para ms detalles sobre la proteccin contra explosin, tipos de proteccin y modelos de los aparatos, consulte el
diagrama de instalacin FCB 3KXF002126G0009 (vase captulo "Anexo").
Change from one to tw o c olumns

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 35

6.7

Instrucciones para el funcionamiento seguro en


zonas potencialmente explosivas (cFMus)
6.7.1 Controles
PELIGRO Peligro de explosin!
Peligro de explosin al abrir la carcasa.
Antes de abrir la carcasa se deben observar los
puntos siguientes:
El propietario debe disponer de un certificado
que autorice la utilizacin de fuego.
Asegrese de que no haya peligro de
explosin.
Deconecte la alimentacin elctrica antes de
abrir la carcasa y mantenga un tiempo de
espera de t > 2 minutos.
ATENCIN - Peligro de quemadura!
Peligro de quemadura por contacto con fluidos
calientes. No tocar el sensor de caudal! La
temperatura superficial depende de la
temperatura del fluido y puede alcanzar ms de
70 C (158 F)!
Asegrese antes de realizar trabajos en el sensor
de caudal de que el aparato se haya enfriado
suficientemente.
Adems, se deben seguir las instrucciones siguientes:
El montaje, la puesta en funcionamiento y los trabajos de
mantenimiento y reparacin en la zona Ex deben ser
realizados, exclusivamente, por personal debidamente
cualificado.
Cuando la carcasa est abierta, la proteccin CEM y la
proteccin contra contacto accidental estn desactivadas.
La conexin a tierra del sensor y del transmisor debe
cumplir los estndares internacionales vigentes.
La conexin entre el sensor de caudal y el transmisor slo
debe realizarse mediante el cable de sealizacin
suministrado por ABB Automation Products.
En la versin con diseo remoto, la longitud del cable de
seal entre el sensor y el transmisor debe ser de
5 m (16,4 ft) como mnimo.
Se debern mantener estrictamente las clases de
temperatura aprobadas, indicadas en el captulo "Datos
tcnicos relevantes para proteccin Ex segn cFMus".
Por favor consulte la explicacin "3KXF002126G0009" del
anexo.

6.7.2 Entradas de cables


Indicaciones especiales para modelos con certificado
para Amrica del Norte.
Los modelos con certificado para Amrica del Norte slo
estn equipados con roscas NPT (sin racor atornillado para
cables).
6.7.3

Conexin elctrica
IMPORTANTE (NOTA)
La caja del transmisor y del sensor de caudal
debe conectarse al conductor de conexin
equipotencial PA. El propietario deber asegurar
que cuando se conecte el conductor protector
PE, no se produzcan diferencias de potencial
entre el conductor protector PE y la conexin
equipotencial PA.

Los clculos Ex se basan en las temperaturas producidas en


la entrada de cables [70 C (158 F)]. Por ello es necesario
que para la alimentacin elctrica y las salidas y entradas de
seal se utilicen cables que cumplan con una especificacin
mnima de 70 C (158 F).
Puesta a tierra

Fig. 30

Segn el estndar NEC, la conexin a tierra entre el


transmisor y el sensor de caudal se puede realizar de la
siguiente forma:
1. Pelar el cable de seal 100 ... 120 mm (3.94 ... 4,72 inch).
2. Deshilar 10 ... 15 mm (0,39 ... 0,59 inch) de la malla de
blindaje del cable de seal. Trenzar los hilos destrenzados
de la malla de blindaje hasta obtener un haz completo.
3. Introducir el haz en el tubo protector flexible suministrado
(verde/amarillo), hasta que en su extremo sobren 10 mm
(0,39 inch) (acortar el tubo protector flexible, si es
necesario).
4. Montar a presin el terminal de cable anular
suministrado (2).
5. Conectar el cable a la toma de tierra (3).

36 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

6.7.4 Process sealing


Segn "North American Requirements for Process Sealing
between Electrical Systems and Flammable or Combustible
Process Fluids".
IMPORTANTE (NOTA)
El aparato es apropiado para uso en Canad.
Si se utiliza en los entornos Class II, Groups E, F
and G, la temperatura superficial no debe
sobrepasar el mximo permitido [165 C
(329 F)].
Todos los conductos para cables (conduits)
deben sellarse dentro de una distancia de 18 inch
(457.20 mm) desde el aparato.
Los caudalmetros de ABB han sido diseados para el
mercado industrial internacional y son apropiados, entre otras
aplicaciones, para la medida de fluidos y lquidos inflamables y
combustibles y para el montaje en tuberas de proceso.
Entre otras cosas, los aparatos que van equipados con
conductos de cables (conduits), estn conectados a la
instalacin elctrica, lo que hace posible que los fluidos de
proceso puedan entrar en el sistema elctrico.

Para impedir que los fluidos de proceso penetren en el


sistema elctrico, los instrumentos van dotados de un juego
de juntas de aislamiento que cumplen los requisitos de la
norma ANSI / ISA 12.27.01. Los caudalmetros msicos de
efecto Coriolis estn diseados como "Single Seal Devices".
Segn los requisitos de la norma ANSI / ISA 12.27.01, es
necesario reducir los valores lmite de funcionamiento para la
temperatura, la presin y las partes presurizadas y ajustarlos a
los valores lmite siguientes:
Valores lmite
Material de las bridas o tuberas

Todos los materiales del modelo

Dimetros nominales

DN 20 ... 150 (1/2 ... 6)

Temperatura de servicio

-50 C ... 200 C (-58 F ... 392 F)

Presin de proceso

PN100 / Class 600

suministrado

Change from two to one c olumn

6.7.5 Cambio del tipo de proteccin


Los modelos FCB330/350, FCH330/350 y FCT330/350 pueden utilizarse en distintos tipos de proteccin contra explosiones:
Como dispositivo con seguridad intrnseca (IS), si se conecta a un circuito con seguridad intrnseca en Div. 1.
Como dispositivo con blindaje antideflagrante (XP), si se conecta a un circuito elctrico sin seguridad intrnseca en Div 1.
Como dispositivo que no produce chispas (NI), si se conecta a un circuito elctrico sin seguridad intrnseca en Div. 2.
En el caso de que un dispositivo instalado deba utilizarse con otro tipo de proteccin contra explosiones, se tendrn que llevar
a cabo las siguientes medidas o comprobaciones de aislamiento, segn la normativa vigente.
1. Tipo de proteccin

2. Tipo de proteccin

contra explosiones

contra explosiones

Housing: XP, Umax = 60 V

Housing: XP

Outputs non IS

Outputs: IS

Medida /comprobacin necesaria


500 V AC/1min o 500 x 1,414 = 710 V DC/1min
Prueba entre los terminales 31 / 32, 33 / 34, 41 / 42, 51 / 52, 81 / 82 y/o 97 / 98 y
entre los terminales 31, 32, 33, 34, 41, 42, 51, 52, 81, 82, 97, 98 y la carcasa.
Durante esta prueba no se deben producir descargas de tensin en el interior o
exterior del aparato.
Inspeccin visual, en especial de las placas electrnicas.
Inspeccin visual: no debe haber daos o explosiones visibles.

Housings: Div 2
Outputs: NI

500 V AC/1min o 500 x 1,414 = 710 V DC/1min


Prueba entre los terminales 31 / 32, 33 / 34, 41 / 42, 51 / 52, 81 / 82 y/o 97 / 98 y
entre los terminales 31, 32, 33, 34, 41, 42, 51, 52, 81, 82, 97, 98 y la carcasa.
Durante esta prueba no se deben producir descargas de tensin en el interior o
exterior del aparato.
Inspeccin visual, en especial de las placas electrnicas.
Inspeccin visual: no debe haber daos o explosiones visibles.

Outputs: IS

Housing: XP

Housing: XP

Outputs: non IS
Housing: XP

Inspeccin visual: las roscas no deben estar daadas (tapa, racores de cable NPT 1/2").
No se requieren medidas especiales

Outputs: NI
Housing: XP, Umax = 60 V

Housing: XP

Outputs: NI

Outputs: IS

500 V AC/1min o 500 x 1,414 = 710 V DC/1min


Prueba entre los terminales 31 / 32, 33 / 34, 41 / 42, 51 / 52, 81 / 82 y/o 97 / 98 y
entre los terminales 31, 32, 33, 34, 41, 42, 51, 52, 81, 82, 97, 98 y la carcasa.
Durante esta prueba no se deben producir descargas de tensin en el interior o
exterior del aparato.
Inspeccin visual, en especial de las placas electrnicas.
Inspeccin visual: no debe haber daos o explosiones visibles.

Housing: XP

Inspeccin visual: las roscas no deben estar daadas (tapa, racores de cable NPT 1/2").

Outputs: non IS
Change from one to tw o c olumns

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 37

Configuracin, parametracin

7.1 Manejo
7.1.1 Navegacin por mens

Control mediante puntero magntico


El puntero magntico permite la programacin cuando la tapa
de la caja est cerrada.
Para activar esta funcin hay que apuntar con el puntero
magntico a los puntos correspondientes de la pantalla LCD.
Los puntos estn marcados con el smbolo
.
Uso de la tecla ENTER en caso de control mediante
puntero magntico
El puntero magntico no puede utilizarse para activar dos
botones al mismo tiempo. En caso de control mediante
puntero magntico, la funcin ENTER se ejecutar activando
durante ms de tres segundos el punto
.
El indicador LCD comenzar a parpadear, para confirmar que
la funcin ENTER se haya ejecutado correctamente.
7.2 Niveles del men
El nivel de configuracin se halla por debajo del indicador de
procesos.

3 Lnea 1
4 Lnea 2
Fig. 31: Indicador LCD
1 Puntos para la programacin mediante el puntero magntico |
2 Botones de control para navegar por el men |
3 1. lnea del indicador LCD | 3 2. lnea del indicador LCD

Indicador de
procesos

Durante la programacin, el transmisor permanecer en lnea,


es decir, las salidas de corriente y de impulsos seguirn
indicando el estado actual de funcionamiento.

Nivel de
configuracin
Language
Mode of operation

Funciones de los botones de control


Las teclas de control
o
sirven para desplazarse por el
men o para seleccionar valores de una lista.
Segn la posicin en el men, los botones de control pueden
tener funciones adicionales.

Concentration
Unit
Flowmeter primary
QmMax
Damping

Smbolo

Significado

Low cutoff setting

Cambio entre el indicador de procesos y el men

Field optimization

Salida del submen

System Zero adj.

Desplazamiento hacia adelante en el men o en una

Alarm
Display

lista de parmetros
Aumento de los valores numricos (incremento)

Totalizer

Desplazamiento hacia atrs en el men o en una lista

Pulse Output
Current output 1 / 2

de parmetros
Reduccin de los valores numricos (decremento)

Switch contacts

Seleccin de la posicin siguiente para introducir

Label
Interface

valores numricos y alfanumricos


Funcin ENTER

Function test

La funcin ENTER se puede ejecutar pulsando

Status

simultneamente las teclas

Versin del software

Selecciona un parmetro que se debe modificar


Confirmacin de un valor / parmetro introducido

Indicador de

El indicador de procesos muestra los valores

Salto al submen

procesos

actuales del proceso.

Nivel de

El nivel de configuracin contiene todos los

configuracin

parmetros necesarios para la puesta en

IMPORTANTE (NOTA)
Los datos introducidos se someten a un control
de plausabilidad. En el caso de que los datos no
se acepten y se rechacen aparecra un mensaje
correspondiente en la pantalla LCD.

38 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

marcha y la configuracin del equipo. Desde


aqu se puede modificar la configuracin del
aparato.

7.2.1 Indicador de procesos


Una vez conectado el aparato, en la pantalla LCD aparecer el
indicador de procesos. All se muestra informacin sobre el
equipo y los valores de proceso actuales.
V 98.14 %
V 12.000 m3

7.2.2 Cambio al nivel de configuracin (parametrizacin)


En el nivel de configuracin se pueden ver y modificar los
parmetros del equipo.
IMPORTANTE (NOTA)
El mensaje "Error bloqueo de programacin" en
la pantalla LCD indica la activacin de la
proteccin de hardware contra escritura a travs
del interruptor de proteccin contra
modificaciones no autorizadas.

1
2

Fig. 32: Indicador de procesos


1 1. lnea del indicador de procesos |
2 2. lnea del indicador de procesos |

Para modificar la representacin en pantalla de los valores de


proceso actuales indicados en las lneas 1 y 2, hay que
cambiar al nivel de configuracin.

V 98.14 %
V 12.000 m3
1.

Smbolo

Descripcin

Indicacin del totalizador de caudal directo

Indicacin del totalizador de caudal inverso

2.

Mensajes de error del indicador LCD


En caso de fallo aparecer un mensaje de error en la pantalla
del indicador de procesos. El texto mostrado da una
indicacin sobre el rea en la que se ha producido el error.

*Prog. level
Blocked
3.

Indicador de procesos
Mensaje de error

4.

IMPORTANTE (NOTA)
Para una descripcin detallada de los errores y
obtener informacin sobre cmo corregir los
errores, vase el captulo 8 "Mensajes de error".

Pulsar
para pasar al nivel de configuracin. A
continuacin, el indicador LCD muestra una opcin
cualquiera del men.
Pulsar
o
para llamar el submen "Nivel. prog."
y, a continuacin, pulsar
(funcin ENTER)
para modificar el ajuste correspondiente.

Pulsar
o
para seleccionar el nivel de acceso
"Specialist".
Pulsar
(funcin ENTER), para confirmar el
ajuste efectuado.

Si se ha definido una contrasea (Prog. Prot. Code), introducir


la contrasea correspondiente.
Prog. Prot. Code
****
5.
6.
7.

Pulsar
para ajustar el valor deseado (la decimal se
incrementa con cada pulsacin de la tecla).
Pulsar
para seleccionar la decimal siguiente.
Pulsar
(funcin ENTER), para confirmar el
ajuste efectuado.

Una vez introducida la contrasea se autorizar el acceso al


nivel de programacin corrrespondiente. Si se ha
seleccionado el nivel de acceso "Service" hay que introducir la
contrasea de servicio.
A continuacin, la pantalla LCD muestra la opcin de men
primera del nivel de configuracin.
8.
9.

Pulsar
o
Pulsar
men elegido.

para elegir un men.


(funcin ENTER), para confirmar el

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 39

Niveles de acceso
IMPORTANTE (NOTA)
Hay cuatro niveles de acceso. Para los niveles de
acceso "Standard" y "Specialist" se puede definir
una contrasea.
No hay una contrasea por defecto.
Nivel de acceso

Descripcin

Blocked

En el nivel "Blocked" todas las opciones estn

Entrada numrica
En la entrada numrica se ajusta un valor al introducir los
decimales.
QmMx
180.00 kg/h
1.
2.

bloqueadas.

Seleccionar en el men el parmetro a ajustar.


Pulsar
(funcin ENTER) para seleccionar el
parmetro a editar. El valor anterior se borra y en la
posicin primera aparece un cursor ( _ ).

Los mens / parmetros slo pueden leerse.


No es posible modificarlos.
Standard

Indicacin y modificacin de todos los mens /


parmetros necesarios para el funcionamiento
del aparato.

Specialist

Indicacin y modificacin de todos los mens /


parmetros accesibles para el cliente.

Service

Visualizacin del men de servicio despus de


introducir la contrasea de servicio (slo para el
personal de servicio de ABB).

7.2.3 Seleccin y modificacin de parmetros


Entrada de datos desde una tabla
En la entrada de datos desde una tabla, los valores se
seleccionan de una lista de valores paramtricos.

QmMx
254.50 kg/h
3.

Pulsar
para ajustar el valor deseado (la decimal se
incrementa con cada pulsacin de la tecla).
4. Pulsar
para seleccionar la decimal siguiente.
5. Si es necesario, seleccionar y ajustar otras decimales
siguiendo los pasos 3 a 4.
6. Pulsar
(funcin ENTER), para confirmar el
ajuste efectuado.
El ajuste del parmetro ha terminado.

Submen
Unidad
1.
2.
3.
4.

Seleccionar en el men el parmetro a ajustar.


Pulsar
(funcin ENTER) para seleccionar el
parmetro a editar.
Pulsar
o
para seleccionar el valor deseado.
Pulsar
(funcin ENTER), para confirmar la
seleccin.

40 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

7.3

Resumen de los parmetros en el nivel de configuracin


IMPORTANTE (NOTA)
En este resumen de parmetros se indican todos los mens y parmetros disponibles del dispositivo. Segn el
equipamiento y la configuracin del dispositivo, puede que en el dispositivo no aparezcan todos los mens y
parmetros.
En este resumen de parmetros, y por motivos de espacio, la funcin ENTER
solo puede representarse
mediante el smbolo .

*Prog. level

Blocked
Standard
Specialist
Service

Language

Mode of operation

Flow indication
Directional display
Signal Level
DensiMass Code
DensiMass
Min Density

Concentration

Medium
Variable Matrix

Configuration

Number matrices
Number temp.

Label

Number conc.

Unit name

Enter conc. in %

Min. concentr.

Reset matrix

Max. concentr.
Matrix 1 / 2
Unit

Unit Qv

Temperature
Concentration in unit

Unit Qm
Enter matrix finish

Concentration in percent

Density unit

Density

Totalizer Unit Volume

Matrix calculation

Totalizer Unit Mass


Prog. Unit Qm

Unit name
Units factor

Prog. Unit Qv

Unit name
Units factor

Temp. unit
Unit concentration
Flowmeter primary

Meter tube
QmMax meter tube
Order no.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 41

QmMax
Damping
Low cutoff setting

Field optimization

D correction
Qm correction
C correction tab. 1
C correction tab. 2

System Zero adj.

System Zero adj.

System Zero adj. Slow?

Function automatic?

System Zero adj. Fast?

System Zero adj.

System Zero adj.

manual?
Alarm

Min Alarm Qm
Max Alarm Qm
Min Alarm density
Max Alarm density
Min Alarm Temp.
Max Alarm Temp.
Min Alarm Concentr.
Max Alarm Concentr.

Display

1st Line
2nd Line
1st Line Multiplex
2nd Line Multiplex

Totalizer

Totalizer mass

Counter > F
Overflow > F

Totalizer volume

Totalizer < R
Overflow < R

Totalizer net. mass


Totalizer Reset
Pulse Output

Output of
QvMax pulse
Qm% max pulse
Pulse
Pulse width

42 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Current output 1

Output of
Qv l = 100 %
Density l = 0 %
Density l = 100 %
Temp l = 0 %
Temp l = 100 %
Qm% l = 100 %
Concentr. l = 0 %
Concentr. l = 100 %
Current output
Iout for Alarm
Low Alarm
High Alarm

Current output 2

Output of
Qv l = 100 %
Density l = 0 %
Density l = 100 %
Temp l = 0 %
Temp l = 100 %
Qm% l = 100 %
Concentr. l = 0 %
Concentr. l = 100 %
Iout for Alarm
Low Alarm
High Alarm

Switch contacts

Contact input
Contact output

Label

TAG number
Descriptor
Date
Unit number
TAG number

Interface

Communication
Unit address

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 43

Function test

Function test pulse

Pulse Output

Output
Function test Iout 1

Iout 1

Function test Iout 2

Iout 2

Function test Contact

Contact input

input
Function test Contact

Contact output

output
Function test HART
Transmitter
Function test HART
Command
Function test Memory
Simulation

Simulation
Qm
Qm
Density
Density
Temperature
Temp. housing
Temperature
Temp. housing

Status

Error log Number: 3

7 (Set) Temp. Measurement


9a (current) Density

Warning Register

measurement

Number: 1
Power outage
Status Reset
4 (Set) Ext. Cut-off

CoriolisMaster 03.2012
FCB FW

44 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4 Descripcin de los parmetros


7.4.1 Men: *Prog. level
/ *Prog. level
Mscaras

Rango de valores

*Prog. level

Blocked

Descripcin
Seleccin del nivel de acceso.

Standard

Si se ha definido una contrasea (cdigo de proteccin del programa), introducir a continuacin

Specialist

la contrasea correspondiente.

Service

Despus de introducir la contrasea, seleccionar a continuacin el nivel de acceso deseado. Si la


contrasea es "0000" (valor por defecto), el nivel de acceso puede seleccionarse sin introducir la
contrasea.
"Blocked": En el nivel "Bloqueado" todas las opciones estn bloqueadas. Los mens /
parmetros slo pueden leerse. No es posible modificarlos.
"Standard": Indicacin y modificacin de todos los mens / parmetros necesarios para el
funcionamiento del aparato.
"Specialist":Indicacin y modificacin de todos los mens / parmetros accesibles para el
cliente.
"Service":Indicacin adicional del men de servicio despus de introducir la contrasea de
servicio
(slo para el personal de servicio de ABB).
Seleccin del submen "Prog. Prot. Code".

Prog. Prot. Code

/ *Prog. level / Prog. Prot. Code


Mscaras

Rango de valores

Descripcin

Old Prog. Prot. (PS)

0 9999

Cmo cambiar la contrasea


Para poder modificar la contrasea es necesario introducir primero la contrasea actual.

code?

Confirmar la entrada pulsando


New Prog. Prot. (PS)

0 9999

Introducir una contrasea nueva. Confirmar la entrada pulsando

code?

7.4.2

Men: Language

/ Language
Mscaras

Rango de valores

Descripcin

Language

German

Seleccin del idioma del men.

English

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 45

7.4.3

Men: Mode of operation

/ Mode of operation
Interfaz

Valores permitidos

Descripcin

Flow indication

Forward / Return

Seleccin del sentido de flujo.

Forward

"Forward / Return": Indicacin de la medicin del caudal en sentido de avance y de retorno


"Forward": Indicacin de la medicin del caudal solo en sentido de avance. Cuando el sensor
de caudal mide el caudal de retorno, en el indicador LCD parpadea el smbolo R y el caudal
indicado es de 0 %. Adems, aparecer el mensaje de advertencia "Q de retorno".
IMPORTANTE (NOTA)
En el modo de funcionamiento "Forward / Return", la salida de impulsos transmite
impulsos para ambas direcciones de flujo.

Directional display

Normal, Reverse

Inversin del sentido de flujo indicado. En este punto debe tenerse en cuenta que la precisin de
la medicin de caudal depende de si el dispositivo se ha calibrado nicamente en sentido de
avance o en ambos sentidos.
"Normal": Indicacin del sentido de flujo de avance
"Reverse": Indicacin del sentido de flujo de retorno.

Signal Level

automatic, High, Low

Seleccin del comportamiento del nivel de seal.


"automatic": Segn la densidad.
"High": Densidad < 0,4 kg/i -> cambio del nivel de seal a "High".
"Low": Densidad > 0,5 kg/l -> cambio del nivel de seal a "Low".
IMPORTANTE (NOTA)
Nivel de seal "Low" = o,5 * nivel de seal "High".

DensiMass Code

Si el dispositivo integra la opcin DensiMass, aqu aparecer el cdigo especfico del dispositivo.
Si posteriormente se desea utilizar esta opcin, debe ponerse en contacto con el servicio
posventa o el representante de ABB.

DensiMass

On, Off, Code invalid

Indicador de estado de la funcin DensiMass. Si aparece "Code invalid", se ha introducido un


cdigo de desbloqueo incorrecto.

Min Density

0 3,5 kg/l

El parmetro Min Density establece un lmite de densidad inferior para casos en los que las
tuberas de medicin estn vacas o se produzca una gran formacin de burbujas. Dado que el
modelo CoriolisMaster tambin puede medir gases, este valor est ajustado de serie en
0=desconectado.
Si no se alcanza este lmite de densidad, se emitir la alarma "9b" y se pausarn la medicin de
caudal y de densidad.
La respuesta detallada del contador y de la salida se encuentra en la tabla del captulo "Mensajes
de error / Resumen". Este error es ms preciso que la alarma de densidad del submen "Alarma".
La histresis de este error es de 0,001kg/l.

46 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.4

Men: Concentration

/ Concentration
Mscaras

Rango de valores

Descripcin

Medium

Variable Matrix

Seleccin de la matriz para el clculo de la concentracin

Sodium hydro

Para ms informacin al respecto, vase el captulo "Software de medida de concentracin

Alcohol in water

DensiMass".

Wheat starch
Corn starch
Sugar in H2O (BRIX)
Seleccin del submen "Variable Matrix".

Variable Matrix

/ Concentration / Variable Matrix


Mscaras

Rango de valores

Descripcin

Configuration

Seleccin del submen "Configuration".

Label

Aqu se puede definir un nombre para la matriz.

Unit name

Alfanumrico, con un

Introduccin del nombre de la unidad para la matriz variable.

mx. de 20 caracteres
Min. concentr.

0 100 %

Introduccin de la concentracin mnima permitida para la matriz variable.

Max. concentr.

0 100 %

Introduccin de la concentracin mxima permitida para la matriz variable.

Matrix 1 / 2

Seleccin del submen "Matrix 1 / 2".

Enter matrix finish

Termina la introduccin de matrices. Los valores introducidos se almacenan o se borran


automticamente.

/ Concentration / Variable Matrix / Configuration


Mscaras

Rango de valores

Descripcin

Number matrices

1, 2

Introduccin del nmero de matrices. Se pueden crear 2 matrices diferentes, como mximo.

Number temp.

2 20

Introduccin del nmero de valores de temperatura de una matriz.

Number conc.

2 20

Introduccin del nmero de valores de concentracin de una matriz.

Enter conc. in %

S, no

Seleccin de la unidad para la concentracin


En el caso de que la unidad de la concentracin sea distinta a %, se puede seleccionar
adicionalmente la unidad %.
Esto es absolutamente necesario para poder calcular los caudales msicos netos!

Reset matrix

Resetea la matriz introducida y restablece los valores por defecto.

/ Concentration / Variable Matrix / Matrix 1 / 2


IMPORTANTE (NOTA)
Para informacin detallada sobre cmo introducir las matrices, vase el captulo "Software de medida de
concentracin DensiMass".
Mscaras

Rango de valores

Descripcin

Temperature

Introduccin de los valores de temperatura para la matriz seleccionada (1 2).

Concentration in unit

Introduccin de los valores de concentracin para la matriz seleccionada (1 2).

Concentration in

Introduccin de los valores de concentracin [en %] para la matriz seleccionada (1 2).

percent
Density

Introduccin de los valores de densidad para la matriz seleccionada (1 2).


Los valores entrados se marcan con "E"; los valores calculados por interpolacin o extrapolacin
se marcan con "B".

Matrix calculation

Clculo de la matriz a base de las entradas efectuadas. Los valores que falten se calcularn por
interpolacin o extrapolacin.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 47

7.4.5 Men: Unit


/ Unit
Interfaz

Valores permitidos

Descripcin

Unit Qm

g/s, g/min, g/h, kg/s,

Seleccin de la unidad del caudal msico.

kg/min, kg/h, kg/d, t/min, La unidad elegida es vlida para los parmetros "QmMx" y "QmMx tubera de medicin", y para
t/h, t/d, lb/s, lb/min, lb/h,

la indicacin del caudal msico actual.

lb/d, abc/s, abc/min,


abc/h, abc/d
Unit Qv

l/s, l/min, l/h, m3/s,

Seleccin de la unidad del caudal volumtrico.

m3/min, m3/h, m3/d,

La unidad elegida indica el caudal volumtrico y aparece, p. ej, durante la indicacin del caudal

ft3/s, ft3/min, ft3/h, ft3/d, volumtrico o al introducir los lmites mn. y mx. de la salida de corriente, siempre que el valor
ugl/s, ugl/min, ugl/h,

del caudal volumtrico deba indicarse a la salida de corriente.

mgl/d, igps, igpm, igph,


igpd, bbl/s, bbl/min,
bbl/h, bbl/d, abc/s,
abc/min, abc/h, abc/d
Density unit

g/ml, g/l, g/cm3, kg/l,

Seleccin de la unidad de la densidad.

kg/m3, lb/ft3, lb/ugl


Totalizer Unit Volume

abc, l, m3, ft3, ugl, igl,

Seleccin de la unidad del contador volumtrico.

bbl
Totalizer Unit Mass

g, kg, t, lb, abc

Seleccin de la unidad del contador msico.


Seleccin del submen "Prog. Unit Qm" (unidad de masa programada).

Prog. Unit Qm

Seleccin del submen "Prog. Unit Qv" (unidad de volumen programada).

Prog. Unit Qv
Temp. unit

C, K, F

Seleccin de la unidad de la temperatura.

Unit concentration

%, BRIX, Baum,

Seleccin de la unidad de la concentracin. Consulte tambin el men "Concentration".

/ Unit / Prog. Unit Qm


Este men permite seleccionar una unidad de masa definida (programada) por el usuario. Esta unidad se puede seleccionar en
los mens y parmetros correspondientes.
Interfaz

Valores permitidos

Descripcin

Unit name

ASCII, 3 caracteres

Introduccin del nombre de la unidad de masa definida por el usuario o la unidad del volumen

como mximo

normal (p. ej. "Nm3").

0,00001 5000000 kg

Introduccin del factor kg/unidad para la unidad de masa definida por el usuario.

Units factor

Este factor tambin puede utilizarse en la Prog. Unit Qm para indicar el volumen normal durante
la medicin de gases.
Introduzca la densidad de referencia como factor en este submen. Introduzca la densidad de
referencia con la unidad [kg/x], sustituyendo "x" por la unidad de volumen normal deseada.
Ejemplo:
Gas con una densidad de referencia de 1,28kg/m3
Units factor = 1,28; Unit name = Nm3
Para poder indicar el volumen normal, seleccione esta nueva unidad en el submen "Unit" y
seleccione "Qm" o "Masa" en el men de salida deseado.

/ Unit / Prog. Unit Qv


Este men permite seleccionar una unidad de volumen definida (programada) por el usuario. Esta unidad se puede seleccionar
en los mens y parmetros correspondientes.
Interfaz

Valores permitidos

Descripcin

Unit name

ASCII, 3 caracteres

Introduccin del nombre de la unidad de volumen definida por el usuario.

como mximo
Units factor

0,00001 5000000 l

Introduccin del factor l/unidad para la unidad de volumen definida por el usuario.

48 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.6

Men: Flowmeter primary

/ Flowmeter primary
Mscaras

Rango de valores

Meter tube

slo visualizacin

Descripcin
Indicacin del dimetro nominal del aparato.

QmMax meter tube

slo visualizacin

Indicacin del valor QmMx (caudal msico mximo) para el dimetro nominal del aparato

Order no.

slo visualizacin

Indicacin del nmero de pedido. El nmero de pedido indicado corresponde a los datos
indicados en la placa de caractersticas y la memoria de datos externa.

7.4.7

Men: QmMax

/ QmMax
Mscaras

Rango de valores

Descripcin

QmMax

0.01 1.0 x QmMax

Introduccin del rango de medida.

meter tube

Para determinar los lmites del rango de medida se pueden entrar valores de entre 0,01 y 1,0
QmMxTubo de medida. El rango de medida ajustado es vlido para ambas direcciones de flujo.
QmMx es el valor al cual se refieren el valor de corriente Qm, el caudal bajo y las alarmas Qm.
(QmMx = 20 mA para la salida corriente Qm).
IMPORTANTE (NOTA)
Al introducir un dimetro nominal nuevo, el parmemtro QmMax se pondr
automticamente al valor 1,0 x QmMax meter tube.

7.4.8

Men: Damping

/ Damping
Mscaras

Rango de valores

Descripcin

Damping (5 tau)

1 100 s

Introduccin del valor de amortiguacin


El valor de amortiguacin indica el tiempo dentro el cual el transmisor reacciona a un salto de
escaln unitario (un 99% del valor final).

7.4.9

Men: Low cutoff setting

/ Low cutoff setting


Mscaras

Rango de valores

Descripcin

Low cutoff setting

0 10 %

Introduccin del valor para el caudal bajo, en % de QmMax.


El caudal bajo mximo es del 10 %. La histresis de conmutacin es del 0,1 %. Si se introduce
un caudal bajo de un 0 %, se desactivar tambin la histresis de conmutacin.

7.4.10 Men: Field optimization


/ Field optimization
Mscaras

Rango de valores

D correction

-50 50 g/l

Descripcin
Introduccin del factor de correccin para la densidad.
Para obtener en la medida de densidad una precisin que equivalga aproximadamente a la
reproducibilidad de 0.0001 g/ml, ste factor puede ser utilizado para realizar una optimizacin en
el campo. Los lmites de esta correccin se situan entre 0,50 g/ml (0,05 g/ml).

Qm correction

-5 5 %

Introduccin del factor de correccin para la medida de caudal.


Para obtener en la medida de caudal una precisin que equivalga aproximadamente o sea incluso
superior a una reproducibilidad de un mnimo del 0,1 % del valor medido, ste factor puede ser
utilizado para realizar una optimizacin en el campo. Este valor tiene el efecto de corregir el valor
medido del caudal msico actual. Se entra en forma de porcentajes del valor actual medido.
Los valores lmite correspondientes se situan entre un 5 % del valor medido.

C correction tab. 1
C correction tab. 2

-1000 1000 %

Introduccin del factor de correccin para la medida de concentracin.


Para obtener en la medida de concentracin una precisin que equivalga aproximadamente o sea
incluso superior a la reproducibilidad, ste factor puede ser utilizado para realizar una
optimizacin en el campo. Este valor tiene el efecto de corregir el valor medido de la
concentracin actual.
Se introduce en forma de la unidad de concentracin actualmente ajustada. El valor de
correccin depende de la matriz de concentracin actualmente seleccionada. Si se trata de una
matriz fija, est disponible un solo valor de correccin. Las matrices variables soportan dos.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 49

7.4.11 Men: System Zero adj.


IMPORTANTE (NOTA)
Asegure los puntos siguientes antes de iniciar el ajuste del punto cero del sistema:
La tubera de medida del sensor de caudal debe estar completamente llena.
Asegrese de que no circule ningn fluido por la tubera de medida del sensor (cierre las vlvulas, dispositivos
de cierre, etc.).
No se deben producir golpes de presin en el fluido de medida.
Asegrese de que el sensor de caudal no est expuesto a vibraciones.
En el fluido de medida no deben encontrarse burbujas de gas.
El ajuste del cero debe realizarse en condiciones normales de operacin (temperatura de funcionamiento,
presin de servicio etc.).
/ System Zero adj.
Mscaras

Rango de valores

Descripcin
Seleccin del submen "System Zero adj. Function automatic?".

System Zero adj.


Function automatic?

Seleccin del submen "System Zero adj. manual?".

System Zero adj.


manual?

/ System Zero adj. / System Zero adj. Function automatic?


Mscaras

Rango de valores

Descripcin

System Zero adj. Slow?

Inicio del ajuste lento del cero del sistema.

System Zero adj. Fast?

Inicio del ajuste rpido del cero del sistema.

/ System Zero adj. / System Zero adj. manual?


Mscaras

Rango de valores

Descripcin

System Zero adj.

x.xxx %

Introduccin del valor para el ajuste manual del cero.

7.4.12 Men: Alarm


Introduccin de los valores lmite (mnimo y mximo) para las magnitudes Caudal msico, Densidad, Concentracin y
Temperatura. El contacto de salida digital 41 / 41 permite sealar si se sobrepasan los valores lmite (mn., mx.) ajustados. La
configuracin se realiza a travs del men " / Switch contacts / Contact output".
/ Alarm
Mscaras

Rango de valores

Descripcin

Min Alarm Qm

0 105 %

Introduccin del valor lmite inferior del caudal msico. El valor debe ser inferior a "Max Alarm

Max Alarm Qm

0 105 %

Introduccin del valor lmite superior del caudal msico. El valor debe ser superior a "Min Alarm

Qm".
Qm".
Min Alarm density

0.5 3.5 kg/l

Introduccin del valor lmite inferior de la densidad. El valor debe ser inferior a "Max Alarm
density".

Max Alarm density

0.5 3.5 kg/l

Min Alarm Temp.

-50 200 C

Introduccin del valor lmite superior de la densidad. El valor debe ser superior a "Min Alarm
density".
Introduccin del valor lmite inferior de la temperatura. El valor debe ser inferior a "Max Alarm
Temp.".

Max Alarm Temp.

-50 200 C

Introduccin del valor lmite superior de la temperatura. El valor debe ser superior a "Min Alarm
Temp.".

Min Alarm Concentr.

-5 105,0 %

Max Alarm Concentr.

-5 105,0 %

Introduccin del valor lmite inferior de la concentracin. El valor debe ser inferior a "Max Alarm
Concentr.".
Introduccin del valor lmite superior de la concentracin. El valor debe ser superior a "Min Alarm
Concentr.".

50 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.13 Men: Display


Configuracin del indicador de proceso del transmisor. Hay dos lneas para poder indicar por separado dos valores diferentes.
Adems, es posible activar el modo Multiplex. Cuando el modo Multiplex est activado, en cada lnea se indican dos valores
que se alternan automticamente (cada 3 segundos).
/ Display
Mscaras

Rango de valores

Descripcin

1st Line

Vese la tabla siguiente

Seleccin del valor indicado.

2nd Line
1st Line Multiplex
2nd Line Multiplex
Valor

Descripcin

Q [Bargraph]

Indicacin del caudal en forma de barras.

Qm

Indicacin del caudal msico [en forma de la unidad definida].

Qv

Indicacin del flujo volumtrico [en forma de la unidad definida].

Q [%]

Indicacin del caudal msico [en porcentajes].

Temperature

Indicacin de la temperatura del fluido [en forma de la unidad definida].

Density

Indicacin de la densidad [en forma de la unidad definida].

Concentr. Unit

Indicacin de la concentracin [en forma de la unidad definida].

Concentr. Percent

Indicacin de la concentracin [en porcentajes].

Qm Concentration

Indicacin del caudal msico neto en funcin de la concentracin actual.

TAG Nummer

Indicacin del nmero TAG definido.

Totalizer Mass

Indicacin de los totalizadores de caudal msico directo/inverso, en funcin del sentido de flujo

Totalizer Mass>F

Indicacin del totalizador de caudal msico directo.

Totalizer Mass<R

Indicacin del totalizador de caudal msico inverso.

Totalizer Volumes

Indicacin de los totalizadores de flujo volumtrico directo/inverso, en funcin del sentido de flujo.

Totalizer Vol.>V

Indicacin del totalizador de flujo volumtrico directo.

Totalizer Vol.<R

Indicacin del totalizador de flujo volumtrico inverso.

Totalizer Net Mass

Indicacin del totalizador de caudal msico neto en funcin del caudal masco neto.

Total. Net Mass >F

Indicacin del totalizador de caudal msico neto directo.

Total. Net Mass <R

Indicacin del totalizador de caudal msico neto inverso.

Pipe frequency

Frecuencia de la tubera de medida.

Blank

Off

Modo Multiplex desactivado (slo para modo Multiplex).

IMPORTANTE (NOTA)
Las unidades de los valores indicados se corresponden con las indicadas en el men " / Unit".

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 51

7.4.14 Men: Totalizer


Todos los cuatro/seis totalizadores cuentan hasta 10 milliones (en forma de la unidad elegida para el totalizador). Una vez que el
totalizador llegue a 10 milliones, el valor del totalizador se trasladar al contador de desbordamiento correspondiente y el
totalizador volver a cero.
Para visualizar est operacin en la pantalla del indicador de procesos aparecer en la pantalla un aviso correspondiente. Se
pueden contar hasta 65535 desbordamientos por cada totalizador.
Cada totalizador puede activarse o ponerse a cero por separado (entrando 0). Al activar (o poner a cero) un totalizador, el
contador de desbordamiento correspondiente se pondr automticamente a cero.
Si en el men " / Mode of operation / Flow indication" se ha seleccionado la opcin "Forward", en el men "Totalizer"
slo aparecern los parmetros de los totalizadores de caudal directo.
/ Totalizer
Mscaras

Rango de valores

Descripcin

Totalizer mass

Seleccin del submen "Totalizer mass".

Totalizer volume

Seleccin del submen "Totalizer volume".

Totalizer net. mass

Seleccin del submen "Totalizer net. mass".

Totalizer Reset

Puesta a cero de todos los totalizadores (despus de confirmar la advertencia pulsando


. Todos los totalizadores se pondrn a cero. Para poner a cero un totalizador individual hay
que poner a cero el parmetro correspondiente.

/ Totalizer / Totalizer mass


/ Totalizer / Totalizer volume
/ Totalizer / Totalizer net. mass
Mscaras

Rango de valores

Counter > F

0 10.000.000

Descripcin
Introduccin e indicacin del valor indicado por el totalizador de caudal directo.

Overflow > F

Slo visualizacin

Indicacin de los debordamientos del totalizador de caudal directo correspondiente. Un


desbordamiento equivale a la cifra 10.000.000. Se pueden indicar 65.636 desbordamientos,
como mximo.

Totalizer < R

0 10.000.000

Overflow < R

Slo visualizacin

Introduccin e indicacin del valor indicado por el totalizador de caudal inverso.


Indicacin de los debordamientos del totalizador de caudal inverso correspondiente. Un
desbordamiento equivale a la cifra 10.000.000. Se pueden indicar 65.636 desbordamientos,
como mximo.

52 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.15 Men: Pulse Output


/ Pulse Output
Mscaras

Rango de valores

Descripcin

Output of

Mass

Seleccin del valor medido emitido en la salida de impulsos.

Volume

Mass: Indicacin del caudal msico

Qm Concentration

Volume: Indicacin del flujo volumtrico


Qm Concentration: Indicacin del caudal msico neto

QvMax pulse

Indicacin del caudal por impulso (slo cuando se indica el flujo volumtrico).

Qm% max pulse


Pulse

Indicacin de la masa por impulso (slo cuando se indica el caudal msico).


0.001 1000
imp/unidad

Introduccin de los impulsos por unidad (valor de impulso).


Si es necesario, el valor introducido se corregir por el transmisor, ya que la frecuencia lmite de
la salida de impulsos es de 5000 Hz.
IMPORTANTE (NOTA)
Un aumento del valor de impulso puede reducir el ancho de impulso!

Pulse width

0.1 2000 ms

Introduccin del ancho de impulso. El valor de impulso y ancho de impulso dependen uno del
otro y se corregirn por el transmisor, si es necesario.

Ejemplos
Ejemplo 1
Ajustes

Introduccin

Resultado

QmMx = 24 kg/min = 0,4 kg/s

Ancho de impulso nuevo:

0,4 kg/s x 100 impulsos/kg = 40 impulsos/s = 40 Hz

Unidad del totalizador: kg

10 ms

Duracin del perodo = 25 ms

Valor de impulso: 100 impulsos/kg

Ancho de impulso mximo = duracin del perodo / 2 = 12,5 ms


Resultado: Se acepta el ancho de impulso introducido [10 ms].

Ejemplo 2
Ajustes

Introduccin

Resultado

QmMx = 6 kg/min = 0,1 kg/s = 100 g/s

Valor de impulso nuevo:

100g/s x 60 impulsos/g = 6000 impulsos/s = 6000 Hz

Unidad del totalizador: g

60 impulsos/g

Se ha sobrepasado la frecuencia lmite de 5000 Hz. El

Ancho de impulso: 10 ms

transmisor pondr el valor de impulso automticamente a 50


impulsos/g y la duracin del perodo a 0,2 ms (5 kHz), porque
sto corresponde exactamente a 5000 Hz.
Ancho de impulso mximo = duracin del perodo /2 = 0,1 ms
Resultado: Es necesario reducir el valor de impulso introducido
y, adicionalmente, el ancho de impulso.

IMPORTANTE (NOTA)
Si se utiliza un totalizador mecnico recomendamos utilizar un ancho de impulso de 30 ms y una frecuencia
lmite [fmx] de 3 kHz.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 53

7.4.16 Men: Current output 1


/ Current output 1
Mscaras

Rango de valores

Descripcin

Output of

Qm, Qv, Density,

Seleccin del valor medido emitido en la salida de corriente 1 (31 / 32).

Temperature,
Concentration
Qv l = 100 %

0.1 10.000.000

Introduccin del caudal volumtrico a partir del cual la salida de corriente alcanza su valor
mximo (20 mA) (slo visible si el caudal volumtrico se transmite a travs de la salida de
corriente).

Density l = 0 %

0.5 3.5 g/cm3

Introduccin de la densidad a partir de la cual la salida de corriente alcanza su valor mnimo (0


mA)
(slo visible si la densidad se transmite a travs de la salida de corriente).

Density l = 100 %

0.5 3.5 g/cm3

Introduccin de la densidad a partir de la cual la salida de corriente alcanza su valor mximo (20
mA)
(slo visible si la densidad se transmite a travs de la salida de corriente).

Temp l = 0 %

-50 200 C

Introduccin del valor de temperatura a partir del cual la salida de corriente alcanza su valor
mnimo (0 mA)
(slo visible si el valor de temperatura se transmite a travs de la salida de corriente).

Temp l = 100 %

-50 200 C

Introduccin del valor de temperatura a partir del cual la salida de corriente alcanza su valor
mximo (20 mA)
(slo visible si el valor de temperatura se transmite a travs de la salida de corriente).

Qm% l = 100 %

Introduccin del caudal msico neto a partir del cual la salida de corriente alcanza su valor
mximo (20 mA) (slo visible si el caudal msico neto se transmite a travs de la salida de
corriente).

Concentr. l = 0 %

Introduccin de la concentracin a partir de la cual la salida de corriente alcanza su valor mnimo


(0 mA)
(slo visible si la concentracin se transmite a travs de la salida de corriente).

Concentr. l = 100 %

Introduccin de la concentracin a partir de la cual la salida de corriente alcanza su valor mximo


(20 mA)
(slo visible si la concentracin se transmite a travs de la salida de corriente).

Current output

0 20 mA, 4 20 mA

Seleccin del modo de operacin de la salida de corriente. Si est activada la comunicacin


HART, se debe elegir '4 ... 20 mA'.
IMPORTANTE (NOTA)
La comunicacin HART se realiza a travs de la salida de corriente 1. Para poder
utilizar la comunicacin HART, hay que elegir el modo de operacin 4 20 mA.
Si se utiliza el modo de operacin 0 20 mA y se activa la comunicacin HART,
aparecer un mensaje de error y el tipo de comunicacin no puede modificarse.
Si se ha activado la comunicacin HART y se activa el modo de operacin 0
20 mA, aparecer un mensaje de error y la comunicacin HART se desactivar
automticamente.

Iout for Alarm

Low, High

Seleccin del estado de la salida de corriente 1 en caso de fallo. El ajuste de la corriente emitida

Low Alarm

2 3.6 mA

Seleccin de la corriente para alarma Low.

"Low" o "High" se realizar en los menes siguientes.


El valor depende del rango de salida de corriente elegido.
Para el rango de salida de corriente 0 ... 20 mA, la corriente de alarma es de 0 mA.
Para el rango de salida de corriente 4 ... 20 mA, la corriente de alarma Low puede modificarse
dentro de los lmites 2 ... 3,6 mA.
Si se modifica el rango de salida de corriente, el transmisor modificar automticamente la
corriente de alarma Low adaptndola al rango de salida de corriente nuevo (el rango 0 ... 20 mA
cambiar a 0 mA y el rango 4 ... 20 mA a 2 mA).
High Alarm

21 26 mA

Seleccin de la corriente para alarma High.


El valor es independiente del rango de salida de corriente elegido, ya que el lmite superior de
ambos rangos es de 20 mA. El rango de corriente de alarma High comienza en 21 mA y acaba en
26 mA.

54 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.17 Men: Current output 2


A diferencia de la salida de corriente 1, la salida de corriente 2 no es apta para HART y dispone unicamente de un solo rango
de salida de corriente (4 ... 20 mA).
/ Current output 2
Mscaras

Rango de valores

Descripcin

Output of

Qm, Qv, Density,

Seleccin del valor medido emitido en la salida de corriente 1 (31 / 32).

Temperature,
Concentration
Qv l = 100 %

0.1 10.000.000

Introduccin del caudal volumtrico a partir del cual la salida de corriente alcanza su valor
mximo (20 mA)
(slo visible si el caudal volumtrico se transmite a travs de la salida de corriente).

Density l = 0 %

0.5 3.5 g/cm3

Introduccin de la densidad a partir de la cual la salida de corriente alcanza su valor mnimo (0


mA)
(slo visible si la densidad se transmite a travs de la salida de corriente).

Density l = 100 %

0.5 3.5 g/cm3

Introduccin de la densidad a partir de la cual la salida de corriente alcanza su valor mximo (20
mA)
(slo visible si la densidad se transmite a travs de la salida de corriente).

Temp l = 0 %

-50 200 C

Introduccin del valor de temperatura a partir del cual la salida de corriente alcanza su valor
mnimo (0 mA)
(slo visible si el valor de temperatura se transmite a travs de la salida de corriente).

Temp l = 100 %

-50 200 C

Introduccin del valor de temperatura a partir del cual la salida de corriente alcanza su valor
mximo (20 mA)
(slo visible si el valor de temperatura se transmite a travs de la salida de corriente).

Qm% l = 100 %

Introduccin del caudal msico neto a partir del cual la salida de corriente alcanza su valor
mximo (20 mA) (slo visible si el caudal msico neto se transmite a travs de la salida de
corriente).

Concentr. l = 0 %

Introduccin de la concentracin a partir de la cual la salida de corriente alcanza su valor mnimo


(0 mA)
(slo visible si la concentracin se transmite a travs de la salida de corriente).

Concentr. l = 100 %

Introduccin de la concentracin a partir de la cual la salida de corriente alcanza su valor mximo


(20 mA)
(slo visible si la concentracin se transmite a travs de la salida de corriente).

Iout for Alarm

Low, High

Seleccin del estado de la salida de corriente 1 en caso de fallo. El ajuste de la corriente emitida
"Low" o "High" se realizar en los menes siguientes.

Low Alarm

2 3.6 mA

Seleccin de la corriente para alarma Low.

High Alarm

21 26 mA

Seleccin de la corriente para alarma High.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 55

7.4.18 Men: Switch contacts


/ Switch contacts
Mscaras

Rango de valores

Descripcin

Contact input

No function,

Seleccin de la funcin de la entrada de contacto digital (81 / 82).

Totalizer reset. Concentr.

No function: Entrada de contacto sin funcin.

Table,

Totalizer reset. Concentr. Table: Cambio de las tablas (Marix 1 / 2) para el clculo de la

Ext. output Shut-off,


Totalizer reset.

concentracin.
Ext. output Shut-off: La salida de corriente y salida de impulsos se ponen en cero. Los
totalizadores se pararn.
Totalizer reset.: Todos los totalizadores se ponen en cero.

Contact output

Vese la tabla siguiente

Seleccin de la funcin de la salida de contacto digital (41 / 42).


Para la funcin elegida, la salida puede configurarse opcionalmente como contacto de cierre o
contacto de reposo.

Funcin Contact output

Descripcin

No function

Sin funcin, salida abierta.

F/R-Signal _

Indicacin del sentido de flujo; salida cerrada en caso de flujo inverso.

F/R-Signal /

Indicacin del sentido de flujo; salida abierta en caso de flujo inverso.

General alarm _

Salida cerrada (no hay alarma colectiva). Abre en caso de fallo.

General alarm /

Salida abierta (no hay alarma colectiva). Cierra en caso de fallo.

MAX/MIN Alarm _

Salida cerrada (no hay alarma Mn / Mx). Abre en caso de alarma Mn / Mx.

MAX/MIN Alarm /

Salida abierta (no hay alarma Mn / Mx). Cierra en caso de alarma Mn / Mx.

MIN Alarm _

Salida cerrada (no hay alarma Mn). Abre en caso de alarma Mn.

MIN Alarm /

Salida abierta (no hay alarma Mn). Cierra en caso de alarma Mn.

MAX Alarm _

Salida cerrada (no hay alarma Mx). Abre en caso de alarma Mx.

MAX Alarm /

Salida abierta (no hay alarma Mx). Cierra en caso de alarma Mx.

7.4.19 Men: Label


/ Label
Mscaras

Rango de valores

Descripcin

TAG number

ASCII, 8 caracteres,

Introduccin de un nmero TAG para identificar el punto de medicin (protocolo HART).

como mximo.
Descriptor

ASCII, 16 caracteres,

Introduccin del descriptor HART.

como mximo.
Date

1.1.1900 31.12.2155

Fecha del fichero HART.

Unit number

Slo visualizacin

Indicacin del nmero de aparato.

7.4.20 Men: Interface


/ Interface
Mscaras

Rango de valores

Descripcin

Communication

Off, HART

Seleccin del tipo de comunicacin para la comunicacin digital.


Off: sin comunicacin digital.
HART: comunicacin HART activada, a travs de la salida de corriente 1. Configurar antes el
modo de operacin poniendo la salida de corriente 1 en 4 20 mA.

Unit address

0 15

Introduccin de la direccin HART. El protocolo HART permite la instalacin de un bus con hasta
15 equipos (1 ... 15).
IMPORTANTE (NOTA)
Si la direccin ajustada es superior a 0, el aparato trabajar en modo Multidrop. La
salida de corriente est fijado a 4 mA. A continuacin, se realizar solamente la
comunicacin HART a travs de la salida de corriente.

56 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.21 Men: Function test


/ Function test
Mscaras

Rango de valores

Descripcin
Seleccin del submen "Function test pulse Output".

Function test pulse


Output
Function test Iout 1

Seleccin del submen "Function test Iout 1".

Function test Iout 2

Seleccin del submen "Function test Iout 2".

Function test Contact

Seleccin del submen "Function test Contact input".

input
Function test Contact

Seleccin del submen "Function test Contact output".

output
Function test HART

1200 Hz, 2200 Hz

Seleccin de la frecuencia HART simulada.

Slo visualizacin

Indicacin del ltimo comando HART recibido.

Transmitter
Function test HART
Command
Function test Memory

Activacin de la prueba de funcionamiento de la memoria.

Simulation

Seleccin del submen "Simulation".

/ Function test / Function test pulse Output


Mscaras

Rango de valores

Descripcin

Pulse Output

0.001 5000 Hz

Introduccin de la tasa de impulsos para la prueba funcional de la salida de impulsos. Para


cancelar, pulsar una tecla cualquiera.

/ Function test / Function test Iout 1


/ Function test / Function test Iout 2
Mscaras

Rango de valores

Descripcin

Iout 1

0 26 mA

Introduccin de la corriente de salida para la prueba funcional de la salida de corriente 1. Para


cancelar, pulsar una tecla cualquiera.

Iout 2

3.5 26 mA

/ Function test / Function test Contact input


Mscaras

Rango de valores

Descripcin

Contact input

On, Off

Prueba de funcionamiento de la entrada de contacto. Para cancelar, pulsar una tecla cualquiera.

/ Function test / Function test Contact output


Mscaras

Rango de valores

Descripcin

Contact output

On, Off

Prueba funcional de la entrada de contacto. Para cancelar, pulsar una tecla cualquiera.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 57

/ Function test / Simulation


El men 'Simulacin' contiene opciones que permiten poner en valores programables las magnitudes de medida individuales o
todas las magnitudes de medida del transmisor. Cuando la simulacin est activada aparecen las opciones adicionales del
submen 'Simulacin'.
stas permiten definir para cada magnitud individual si la magnitud se debe medir o simular, as como el valor que la misma
debe tener. Para ello, se puede elegir entre las posibilidades siguientes:
Valor

Descripcin

Medida

Se indicar el valor real medido de la magnitud correspondiente.

Introduccin

El valor de la magnitud se simular. A continuacin, se puede seleccionar un men para ponerlo en un valor fijo.

Step

La magnitud se simular y se podr aumentar o reducir gradualmente a travs de las teclas STEP y DATA siempre que
aparezca el indicador de procesos.

Mscaras

Rango de valores

Descripcin

Simulation

On, Off

Activacin de las funciones de simulacin.

Medicin, introduccin,

Seleccin de la funcin de simulacin.

Qm

Step
Qm

-115 +115 %

Introduccin del caudal msico a simular [en %].

Density

Medicin, introduccin,

Seleccin de la funcin de simulacin.

Step
Density

0,001 g/ml

Introduccin de la densidad a simular [en g/ml].

Temperature

Medicin, introduccin,

Seleccin de la funcin de simulacin.

Step
Temp. housing

Medicin, introduccin,

Seleccin de la funcin de simulacin.

Step
Temperature

-60 210 C

Introduccin de la temperatura a simular [en C]

Temp. housing

-60 210 C

Introduccin de la temperatura a simular [en C]

58 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

7.4.22 Men: Status


/ Status
Mscaras

Rango de valores

Descripcin
Indicacin del nmero de mensajes de error almacenados y seleccin del submen "Memoria de

Error log Number: 3

errores".
Warning Register

Indicacin del nmero de mensajes de advertencia almacenados y seleccin del submen

Number: 1

"Memoria de advertencias".

Power outage

Slo visualizacin

Status Reset

Indicacin del nmero de fallos de red.


Puesta a cero de todos los totalizadores de errores, advertencias y fallos de red (despus de
confirmar la advertencia pulsando

. Todos los totalizadores se pondrn a cero.

/ Status / Memoria de errores


Mscaras

Rango de valores

Descripcin

7 (Set) Temp.

Slo visualizacin

Indicacin del ltimo error ocurrido, con nmero de error y estado (actual o puesto).

Measurement

(ejemplo)

Si hay otros errores, se pueden utilizar las teclas

para hojear en la lista de errores.

9a (current) Density

IMPORTANTE (NOTA)

measurement

El orden en el que los errores aparecen en la lista no responde a su prioridad!

/ Status / Memoria de advertencias


Mscaras

Rango de valores

Descripcin

4 (Set) Ext. Cut-off

Slo visualizacin

Indicacin del ltimo error ocurrido, con nmero de error y estado (actual o puesto).

(ejemplo)

Si hay otros errores, se pueden utilizar las teclas

para hojear en la lista de errores.

IMPORTANTE (NOTA)
El orden en el que los errores aparecen en la lista no responde a su prioridad!

7.4.23 Men: Versin del software


/ Versin del software
Mscaras

Rango de valores

Descripcin

CoriolisMaster 03.2012

Slo visualizacin

Indicacin de la versin del software

FCB FW

(ejemplo)

En la primera lnea se indican el cdigo de identificacin del aparato y la fecha de revisin del
software. En la segunda lnea se indican el nmero de identificacin y nivel de revisin del
software.

Adems de la versin del software indicada en el men, en el lado superior de la unidad de transmisor enchufable se encuentra
una placa indicadora con el nmero de identificacin del software.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 59

Change from one to tw o c olumns

7.5

Medida de concentraciones DensiMass (slo


disponible para FCB350)
El transmisor puede calcular la concentracin actual partiendo
de la densidad y temperatura medidas y utilizando las
matrices de concentracin.
El transmisor viene configurado con las siguientes matrices de
concentracin predefinidas:
Concentracin de hidrxido de sodio en agua
Concentracin de alcohol en agua
Concentracin de hidratos de carbono en agua
Concentracin de almidn de maz en agua
Concentracin de almidn de trigo en agua
Adems, se pueden introducir dos matrices (de hasta 100
valores) especificadas por el usuario.
7.5.1 Precisin de la medida de concentraciones
La precisin de la medida concentraciones depende primero
de la calidad de las matrices introducidas.
Sin embargo, como el clculo se basa en las magnitudes
iniciales Densidad y Temperatura, la precisin de la medida de
concentraciones se determina, en fin de cuentas, por la
precisin de la medida de la temperatura y densidad.
Ejemplo:
densidad: un 0% de alcohol en agua, temperatura:
20 C [68 F] = 998,23 g/l
densidad: un 100 % de alcohol en agua, temperatura:
20 C [68 F] = 789,30 g/l
Concentracin

Densidad

100 %

208,93 g/l

0,48 %

1 g/l

0,69 %

2 g/l

7.5.2 Introduccin de la matriz de concentraciones


La matriz de concentraciones se introduce a travs del men
" / Concentration / Variable Matrix". Los datos de la
matriz deben estar disponibles (vase el captulo "Estructura
de la matriz de concentraciones").
Paso 1:
Men " / Concentration / Variable Matrix".
Introduccin del nombre de la unidad para la matriz variable,
introduccin de los valores lmite superior e inferior para la
concentracin.
Paso 2:
Introduccin de los ajustes bsicos de la matriz en el men
" / Concentration / Variable Matrix / Configuration".
Aqu se fija el nmero de matrices (1 2), el nmero de
valores de temperatura y el nmero de valores de
concentracin.
Paso 3:
Introduccin de los datos de las matrices en el men
" / Concentration / Variable Matrix / Matrix 1 / 2".
Aqu se introducen los valores de temperatura, valores de
concentracin (en la unidad deseada o en %) y los valores de
densidad.
Una vez introducidos los datos, la matriz se calcular a travs
de la opcin de men "Matrix calculation". Los valores que
falten se calcularn por interpolacin o extrapolacin.
Paso 4:
Almacenamiento de las matrices en el men " /
Concentration / Variable Matrix" (a travs de la opcin de
men "Enter matrix finish").
La introduccin de las matrices ha terminado.

Por consiguiente, la clase de precisin de la medida de


densidad determina directamente la precisin de la medida de
concentraciones.

60 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

7.5.3

Estructura de la matriz de concentraciones

Respecto al software, se distinguen dos valores de concentracin diferentes:


1. concentracin indicada en unidades (p. ej.: % o B)
El rango de valores no est limitado; el valor puede transmitirse a la salida de corriente; el valor puede seleccionarse en el
submen Unidades.
2.

Concentracin en porcentajes (%)


El rango de valores est limitado al 0 ... 103,125 %. Este valor slo se utiliza para el clculo interno del caudal msico neto.
El valor del caudal msico neto puede transmitirse a la salida de corriente y la salida de impulsos.

Lmites de concentracin MIN / MAX: -5.0 ... 105,0.


La matriz para el clculo de concentraciones tiene la siguiente estructura:
Temperatura 1
Valor 1 Concentracin en

Valor 1 Concentracin

indicada en unidades (p.

Valor 1,1 Densidad

Temperatura n
Valor n,1 Densidad

ej. % o B)
.

Valor m Concentracin en

Valor m Concentracin

indicada en unidades (p.

Valor 1, m Densidad

Valor n,m Densidad

ej. % o B)

Al entrar los valores en la matriz deben observarse las siguientes reglas:


si se utiliza una sola matriz: 2 N 20; 2 M 20; N * M 100
si se utilizan dos matrices: 2 N 20; 2 M 20; N * M 50
Los valores de densidad de una columna deben estar odenados de forma ascendente debido al algoritmo utilizado en el
software del transmisor.
Density x,1 < < Density x,2 << Density x,M para 1 x M
Los valores de temperatura deben estar ordenados de forma ascendente de la izquierda a la derecha debido al algoritmo
utilizado en el software del transmisor.
Temperatura 1 << Temperatura x << Temperatura N para 1 x N
Los valores de concentracin deben estar ordenados de forma montonamente ascendente o descendiente desde arriba hacia
abajo debido al algoritmo utilizado en el software del transmisor.
Concentr. 1 << Concentr. x < < Concentr. N para 1 x N
o
Concentr. 1 >> Concentr. x > > Concentr. N para 1 x N
Ejemplo:
0%

0 BRIX

10 C (50 F)

20 C (68 F)

30 C (86 F)

0,999 kg/l

0,982 kg/l

0,979 kg/l

10 %

10 BRIX

1,010 kg/l

0,999 kg/l

0,991 kg/l

40 %

30 BRIX

1,016 kg/l

1,009 kg/l

0,999 kg/l

80 %

60 BRIX

1,101 kg/l

1,018 kg/l

1,011 kg/l

Change from one to tw o c olumns

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 61

Change from two to one c olumn

7.6 Historial de versiones del software


Segn la recomendacin NAMUR NE53, ABB ofrece un historial del software claro y fcilmente comprensible.
Versin estndar y versin HART
Software FCB FW
Versin de

Fecha de revisin

Tipo de modificacin

Descripcin

00.01.xx

03/2012

Versin nueva

00.02.xx

06/2013

Actualizacin

Introduccin de una nueva funcin: la densidad mn. genera el error "9b"

software

Change from two to one c olumn

Mensajes de error

8.1 Generalidades
Los listados de alarmas indicados en forma de tabla en las pginas siguientes describen el comportamiento del transmisor en
caso de errores y fallos funcionales.
La tabla lista todos los fallos posibles del transmisor, as como su influencia sobre los valores de las magnitudes medidas, el
comportamiento de las salidas de corriente y la salida de alarma.
Las celdas vacas de la tabla indican que el fallo no tiene efecto sobre la magnitud medida o la sealizacin de alarmas por la
salida correspondiente. Si las celdas de la columna 'Salida de corriente' slo indican 'Alarma', la alarma se sealar solamente
como 'alarma alta' o 'alarma baja', segn la opcin seleccionada en el men de salidas de corriente.
El orden de los errores en la tabla corresponde a su prioridad.
El asiento primero tiene la prioridad ms alta, el ltimo la prioridad mas baja.
Si se producen varios errores al mismo tiempo, el error con la prioridad ms alta determinar el estado de alarma de la
magnitud medida y de la salida de corriente, respectivamente. En cambio, si un error de alta prioridad no tiene influencia sobre
una salida o magnitud medida, el estado de las mismas ser determinado por el error con la prxima prioridad ms baja.
Ejemplo:
Si se produce el error 7a "T Pipe measurement", se puede ver en la tabla que este error cambia la magnitud Temperatura
(temp. constante 20 C [68 F]).
Como la medida de temperatura es elemental para calcular la densidad y, por consiguiente, el valor Qv, las salidas de corriente
asignadas a estos parmetros cambiarn al estado de alarma programado (alarma High o Low).

62 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Sinopsis

Masa neta

20

20

10

DSP communication

20

AD Transmitter

20

11d

Sensor

Sensor amplitude

2a

Driver

2b

Driver current

9a

Density measurement

10

9b

Min Density

2)

2)

2)

11

7a

T Pipe measurement

20

12

7b

T Housing measurement

20

13

Flowrate >103.25 %

Qm =

14

12

Concentration (Percent)

15

Ext. zero return

16

8a

Iout 1 to large

17

8b

Iout 1 to small

18

8c

Iout 2 to large

19

8d

Iout 2 to small

20

6a

Totalizer Mass -> V

1)

21

6b

Totalizer Mass <- R

1)

22

6c

Totalizer Vol.-> V

1)

23

6d

Totalizer Vol.<- R

1)

24

6e

Totalizer Net Mass -> V

25

6f

Totalizer Net Mass <- R

26

11a

Sensor A

27

11b

Sensor B

103 Qm =

N. de error

Prioridad

103

Salida de alarma

Volumen

Caudal msico neto

Masa

Temperatura

Caudal msico neto

External FRAM

Densidad

Concentracin [unidad]

Internal FRAM

5b

Qv

Concentracin [%]

5a

Qm

Temperatura [C]

Texto de error

Densidad [g/cm 3]

Salida de corriente

Qv [%]

Contador

Qm [%]

Magnitudes de medida

Concentracin [unidad]

8.2

1)

1)

103

Los estados actuales de los contadores, de las salidas de corriente y de la salida de alarma se representan con iconos;
consulte la tabla siguiente.
Icono

Descripcin
Parada del contador

Sin modificacin

1)

Al producirse el error, el contador correspondiente vuelve a contar partiendo de 0.

2)

Al producirse el error, la magnitud de medida se restablece al valor de la densidad mnima.


Alarma (general)
Alarma High
Alarma Low

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 63

8.3

Mensajes de error

Mensaje de error
Error: 0

Prioridad
6

Sensor amplitude

Descripcin

Causa posible

Indicaciones para la correccin

La amplitud del sensor, que

El error se produce nicamente

Reducir la concentracin de gas,

depende del dimetro nominal,

cuando el sensor est lleno?

cambiar el fluido

es inferior a 10 mV.

"Fluido que absorbe energa" en el


sensor (p. ej. alta concentracin de
gas, fluidos muy viscosos), por lo
que la corriente de excitacin es
insuficiente

Error: 1

AD Transmitter

El convertidor A/D est saturado

Perturbaciones mecnicas o

Desacoplar el sensor de las fuentes

hidrulicas muy fuertes en la tubera

de perturbaciones

Versin con proteccin para

Utilizar un cable ms corto, reducir la

explosiones y diseo remoto:

resistencia creando una conexin en

La resistencia elctrica del cable del

paralelo o utilizando un cable de bajo

excitador es excesiva

ohmiaje

La tensin del sensor es excesiva

Comprobar las amplitudes del sensor

o no responde

y controlar si la amplitud ajustada es


correcta

Error: 2a

Convertidor A/D averiado

Cambiar la placa DSP

Interrupcin del circuito de

En versiones de diseo remoto:

regulacin; el sensor no es

comprobar el cableado entre el

compatible con el transductor

sensor y el transductor

Consulte el error 0

Consulte el error 0

El valor inferior QmMx ajustado

Rango de medicin ajustado

Aumentar el rango de medicin

se excede en ms del 5 %

insuficiente

(QmMx)

Caudal excesivo

Reducir el caudal

El caudal se pone en cero, por lo

La entrada de contacto externa est

Establecer la entrada de contacto

que los contadores se paran

establecida en "High"

externa en "Low"

Base de datos errnea

Desconectar y reconectar el

El sensor no vibra

Driver
Error: 2b

Driver current

Reaccin del limitador de


corriente del excitador, debido a
una insuficiencia de la corriente
de excitacin

Error: 3

13

Flowrate >103.25 %
Error: 4

14

Ext. zero return

automticamente
Error: 5b

Prdida de la base de datos


externa

External FRAM

dispositivo; realizar una


comprobacin de funcionamiento del
transductor
Falta la memoria de datos externa

Instalar la memoria de datos externa

La memoria de datos externa est

Cargar la memoria de datos externa

vaca
Error: 6a

19

Error: 6b

20

Reprogramar el contador

El contador de caudal msico de

Reprogramar el contador

retorno est averiado

Totalizer Mass <- R


Error: 6c

El contador de caudal msico de


avance est averiado

Totalizer Mass -> V

21

El contador de caudal

Reprogramar el contador

volumtrico de avance est

Totalizer Vol.-> V

averiado
Error 6d
Totalizer Vol.<- R

22

El contador de caudal

Reprogramar el contador

volumtrico de retorno est


averiado

Error: 6e

El contador de caudal msico

Totalizer Net Mass -> V

neto est averiado

Error: 6f

El contador de caudal msico

Totalizer Net Mass <- R

neto est averiado

64 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Reprogramar el contador
Reprogramar el contador

Mensaje de error

Prioridad

Descripcin

Causa posible

Indicaciones para la correccin

Error: 7a

11

Error durante la medicin de

Cableado incorrecto

Revisar el cableado entre el

temperatura

(solo en versiones de diseo remoto)

transductor y el sensor de caudal

Para la compensacin de

El Pt 100 est averiado

Controlar en el sensor la resistencia

T Pipe measurement

temperatura de los valores de

del Pt 100

medicin Qm y Densidad, se parte


ahora de 20 C, es decir, la medicin
se realiza correctamente a una
temperatura del fluido de ~20 C
Error: 7b

12

Error durante la medicin de

Cableado incorrecto

T Housing

temperatura

(solo en versiones de diseo remoto)

transductor y el sensor de caudal

measurement

Para la compensacin de

El Pt 100 est averiado

Controlar en el sensor la resistencia

temperatura de los valores de

Revisar el cableado entre el

del Pt 100

medicin Qm y Densidad, se parte


ahora de 20 C, es decir, la medicin
se realiza correctamente a una
temperatura del fluido de ~20 C
Error: 8a

15

Iout 1 to large

Se ha sobrepasado el rango mximo

Rango ajustado insuficiente

Aumentar el rango

Rango ajustado insuficiente

Aumentar el rango

Rango ajustado insuficiente

Aumentar el rango

Rango ajustado insuficiente

Aumentar el rango

La densidad del fluido medida en el

Este error suele aparecer junto con

Consulte los errores 1 y 9

sensor fuera de la especificacin

los errores 1 y 9.

programado para la salida de


corriente 1

Error: 8b

16

Iout 1 to small

No se ha alcanzado el rango mnimo


programado para la salida de
corriente 1

Error: 8c

17

Iout 2 to large

Se ha sobrepasado el rango mximo


programado para la salida de
corriente 2

Error: 8d

18

Iout 2 to small

No se ha alcanzado el rango mnimo


programado para la salida de
corriente 2

Error: 9a

Density measurement

Consutle los errores 1 y 9


Error: 9b

10

Min Density

Consulte la pgina 46, men "Mode

Parmetro "Min Density" definido.

of operation / Min Density".

Comprobar que el sensor de caudal


no est funcionando vaco y evitar la
formacin de burbujas de gas en el
caso de lquidos
O bien, establecer el parmetro "Min
Density" en "0".

Error: 11a

23

Falta la seal del sensor A

Sensor A

Sensor A averiado o interrupcin del

Medir la resistencia del sensor A

circuito de regulacin de amplitud

En versiones de diseo remoto:


comprobar el cablado entre el sensor
y el transductor

Error: 11b

24

Falta la seal del sensor B

Sensor B

Sensor B averiado o interrupcin del

Medir la resistencia del sensor B

circuito de regulacin de amplitud

En versiones de diseo remoto:


comprobar el cablado entre el sensor
y el transductor

Error: 11d
Sensor

Faltan las seales de al menos dos

Hay al menos dos sensores

Medir la resistencia de los sensores

sensores

averiados o bien una interrupcin en

En versiones de diseo remoto:

el circuito de regulacin de amplitud

comprobar el cablado entre el sensor


y el transductor

Error: 12

Concentracin en porcentajes < 0 %

Concentracin en porcentajes < 0 %

Ajustar los datos de la matriz en el

Concentration

o > 103,125 %

o > 103,125 %

submen Concentracin

Error

No se pueden cambiar los

Est activado el interruptor de

Desactivar el interruptor de

Operating protection

parmetros

proteccin contra modificaciones no

proteccin contra modificaciones no

autorizadas

autorizadas

(Percent)

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 65

8.4

Mensajes de advertencia

Advertencia

Prioridad Descripcin

Causa posible

Indicaciones para la

Advertencia: 1

16

En el submen Prueba de funcionamiento

Desactivar el modo de

est activado el modo de simulacin

simulacin

correccin
El modo de simulacin est activado

**Simulation**
Advertencia: 2

Un contador se ha puesto a cero

Valor Qm por debajo del lmite de alarma

Valor Qm por debajo del lmite de alarma

Reducir el lmite de alarma

MN. ajustado

MN. ajustado

MN.

Valor de densidad por debajo del lmite

Valor de densidad por debajo del lmite de Reducir el lmite de alarma

totalizer reset
Advertencia: 5a
Min Alarm Qm
Advertencia: 5b

Min Alarm Density

de alarma MN. ajustado

alarma MN. ajustado

MN.

Valor de temperatura por debajo del

Valor de temperatura por debajo del lmite

Reducir el lmite de alarma

Min Alarm Temp.

lmite de alarma MN. ajustado

de alarma MN. ajustado

MN.

Advertencia: 5d

Valor de concentracin por debajo del

Valor de concentracin por debajo del

Reducir el lmite de alarma

Min Alarm Conc.

lmite de alarma MN. ajustado La

lmite de alarma MN. ajustado

MN.

Valor Qm por encima del lmite de alarma

Valor Qm por encima del lmite de alarma

Aumentar el lmite de

MX. ajustado

MX. ajustado

alarma MX.

Valor de densidad por encima del lmite

Valor de densidad por encima del lmite

Aumentar el lmite de

de alarma MX. ajustado

de alarma MX. ajustado

alarma MX.

Valor de temperatura por encima del

Valor de temperatura por encima del

Aumentar el lmite de

Max Alarm Temp.

lmite de alarma MX. ajustado

lmite de alarma MX. ajustado

alarma MX.

Advertencia: 6d

Valor de concentracin por encima del

Valor de concentracin por encima del

Aumentar el lmite de

Max Alarm Conc.

lmite de alarma MX. ajustado La

lmite de alarma MX. ajustado

alarma MX.

Aparece en el indicador durante 1 min

Se ha cambiado la memoria de datos

tras el encendido

externa (FRAM)

Aparece en el indicador durante 1 min

Se ha realizado una actualizacin del

tras el encendido

software Se ha cambiado la memoria de

Advertencia: 5c

histresis de conmutacin es de 0,1 de


la unidad de concentracin ajustada.
Advertencia: 6a

Max Alarm Qm
Advertencia: 6b

Max Alarm Density


Advertencia: 6c

histresis de conmutacin es de 0,1 de


la unidad de concentracin ajustada.
Advertencia: 7

Ext. Data loaded


Advertencia: 8a

10

Update int. data

datos externa (FRAM)


Advertencia: 8b

11

Update ext. data

Aparece en el indicador durante 1 min

Se ha realizado una actualizacin del

tras el encendido

software Se ha cambiado la memoria de

datos externa (FRAM)


Advertencia: 9a

12

Overflow -> F Mass


Advertencia: 9b

13

Overflow <- R Mass


Advertencia: 9c

14

Overflow -> F Volume


Advertencia: 9d

Desbordamiento del contador de caudal

Poner a cero el contador

msico de avance

msico de avance

Nota: Si se aumental a

Desbordamiento del contador de caudal

Desbordamiento del contador de caudal

unidad, se puede

msico de retorno

msico de retorno

prolongar el tiempo hasta

Desbordamiento del contador de caudal

Desbordamiento del contador de caudal

el prximo

msico de avance

msico de avance

desbordamiento.

Desbordamiento del contador de caudal

Desbordamiento del contador de caudal

Overflow <- R Volume

volumtrico de retorno

volumtrico de retorno

Advertencia: 9e

Desbordamiento del contador de caudal

Desbordamiento del contador de caudal

Overflow -> F %M

msico neto de avance

msico neto de avance

Advertencia: 9f

Desbordamiento del contador de caudal

Desbordamiento del contador de caudal

Overflow <- R %M

msico neto de retorno

msico neto de retorno

El dispositivo trabaja en sentido de

El modo de funcionamiento del dispositivo -

Advertencia: 10
Reverse Q

14

Desbordamiento del contador de caudal

17

retorno

est ajustado en sentido de avance, pero


el fluido en el sensor de caudal fluye en
sentido de retorno

Change from one to tw o c olumns

66 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Mantenimiento / Reparacin

9.1

Generalidades
ADVERTENCIA - Peligro por corriente
elctrica!
Cuando la caja est abierta, la proteccin CEM
no funciona y el usuario no est protegido contra
el riesgo de contacto accidental.
Antes de abrir la caja hay que desconectar la
alimentacin elctrica.
AVISO - Dao de los componentes!
Los componentes electrnicos en las placas de
circuitos impresos pueden daarse por
electricidad esttica (observar las directivas sobre
componentes expuestos a riesgos por
electricidad esttica (ESD)).
Antes de tocar los componentes electrnicos,
asegurarse de que la electricidad esttica de su
cuerpo se descargue.

9.4 Transmisor
9.4.1 Cambio
Todos los parmetros de ajuste se almacenan en una
memoria de datos externa. En caso de cambio de transmisor,
los parmetros de ajuste se extraen por lectura de la memoria
de datos externa y se transfieren al transmisor nuevo.
Los datos del sensor de caudal y los parmetros de ajuste del
cliente se importan automticamente.
Al cambiar el transmisor el usuario deber asegurarse de que
el nmero de serie de la memoria de datos externa
corresponda exactamente con el nmero de serie del sensor
de caudal.
Cuando sea necesario cambiar el transmisor, nuestro servico
tcnico estar siempre a su disposicin para asesorarle y
contestar sus preguntas.
Para cambiar un transmisor por otro con una versin anterior
del software, recomendamos contactar siempre con nuestro
servicio posventa.
Posicin de la memoria de datos externa (FRAM)

Todos los trabajos de reparacin y mantenimiento debern


realizarse, exclusivamente, por el personal tcnico cualificado
del servicio posventa.
Si se cambian o se reparan componentes individuales, se
debern instalar repuestos originales.
9.2 Limpieza
Al limpiar la superficie de las cajas de los aparatos de medida,
asegurarse de que el agente limpiador utilizado no dae las
juntas y la superficie de la cajas.
Para la limpieza nicamente debe utilizarse un pao
humedecido, para impedir cargas electrostticas que daen el
aparato.
9.3 Sensor de caudal
El sensor casi no necesita mantenimiento.
Se recomienda controlar anualmente los siguientes puntos:
Condiciones ambientales (ventilacin, humedad)
Estanqueidad de las conexiones a proceso
Entradas de cables y tornillos de la tapa,
Seguridad funcional de la alimentacin de corriente, la
proteccin contra los rayos y la tierra de la red.
Reparacin del sensor de caudal
En el caso de que sea necesario reparar el sensor de caudal,
se debern observar y seguir las indicaciones del captulo
"Seguridad / Devolucin del aparato".
IMPORTANTE (NOTA)
El uso de medidores desarrollados para zonas
potencialmente explosivas est sujeto a las
normas vigentes en el lugar de destino. Vase al
respecto tambin los captulos 6.6, 6.7, 12 y 13.

Fig. 33: Posicin del mdulo FRAM


1 FRAM (enchufable)

La memoria de datos externa (FRAM) del sensor de caudal se


encuentra en la posicin indicada en la figura (depende del
diseo compacto o remoto).
ADVERTENCIA - Peligro por corriente
elctrica!
Cuando la caja est abierta, la proteccin CEM
no funciona y el usuario no est protegido contra
el riesgo de contacto accidental.
Antes de abrir la caja hay que desconectar la
alimentacin elctrica.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 67

10 Datos tcnicos - Sensor de caudal

10.3 Precisin
10.3.1 Condiciones de referencia

10.1 Diseos
Agua

Medio de calibracin

Temperatura: 25 C (77 F) 5 K
Presin: 2 4 bar (29 58 psi)
Temperatura ambiente

25 C (77 F) +10 K / -5 K

Alimentacin elctrica

Tensin de alimentacin segn la

Fase de calentamiento

30 minutos

Instalacin

Instalacin segn se describe en

placa de caractersticas UN 1 %

"Instrucciones de montaje" y
"Posiciones de montaje"
Sin fase gaseosa visible

Fig. 34: Sensor de caudal FCB3xx / FCH3xx


1 Diseo compacto | 2 Diseo remoto (sin transmisor)

Sin trastornos mecnicos o


hidrulicos exteriores,

10.2 Dimetro nominal y rango de medicin

especialmente: cavitacin
Calibracin de la salida

Salida de impulsos
Igual que la salida de impulsos

Dimetro nominal

Qmax en kg/h (lb/h)

Influencia de la salida

DN 15 (1/2)

0 8.000 (0 17.637)

analgica sobre la precisin de 0,1 % del valor medido

DN 25 (1)

0 35.000 (0 77.162)

medida

DN 50 (2)

0 90.000 (0 198.416)

DN 80 (3)

0 250.000 (0 551.156)

DN 100 (4)

0 520.000 (0 ... 1.146.404)

DN 150 (6)

0 860.000 (0 1.895.975)

10.2.1 Rango de caudal recomendado


Lquidos:
El rango de caudal recomendado es de 5 % 100 % del
valor Q max.
Se deben evitar los caudales < 1 % del valor Qmax.
Gases:
La velocidad del flujo de gases en el tubo de medicin no
debe superar los 0,3 Mach (aprox. 100 m/s [328 pies/s]).
A partir de una velocidad de flujo de aprox. 80 m/s
(262 pies/s), se debe calcular una desviacin mayor en la
repetibilidad.
El rango de caudal mximo de los gases depende de la
densidad operativa. En la pgina www.abb.com/flow hay
disponibles recursos de ayuda para el diseo.

10.3.2 Desviacin del valor medido


La desviacin del caudal medido se calcula de la siguiente
manera:
Caso 1:
Si
Caudal

Estabilidad del cero


(precisin bsica / 100)

se considera que:
Desviacin del valor medido mxima:
precisin bsica en % del valor medido
Repetibilidad:
1/2 x precisin bsica en % del valor medido
Caso 2:
Si
Caudal <

Estabilidad del cero


(precisin bsica / 100)

se considera que:
Desviacin del valor medido mxima:
(estabilidad del cero / valor medido) x 100 % del valor
medido
Repetibilidad:
1/2 x (estabilidiad del cero / valor medido) x 100 % del
valor medido

68 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

10.3.3 Estabilidad del cero


Dimetro nominal

kg/h (lb/h)

DN 15 (1/2)

0,64 (1,41)

DN 25 (1)

2,16 (4,76)

DN 50 (2)

7,20 (15,87)

DN 80 (3)

20 (44)

DN 100 (4)

41,6 (91,7)

DN 150 (6)

68,8 (151,68)

10.3.4 Influencia de la temperatura del fluido


Para el caudal, inferior al 0,0015 % del valor Q max / 1 K.
Para la densidad, inferior a 0,0001 kg/dm3 / 1 K.

Fig. 35: Clculo de la desviacin del valor medido con


FCB350 DN15 (ejemplo)

10.3.5 Influencia de la presin operativa


Dinmica de

Caudal

medicin

Desviacin mxima
del valor medido

Dimetro nominal

Caudal [% del valor

Densidad

medido / bar]

[kg/dm 3 / bar]

100:1

80 kg/h (176,4 lb/h)

0,8 % del valor medido

50:1

160 kg/h (352,7 lb/h)

0,4 % del valor medido

DN 15 (1/2)

-0,002

Ninguna

10:1

800 kg/h (1763,7 lb/h)

0,1 % del valor medido

DN 25 (1)

-0,013

0,00035

2:1

4000 kg/h (8818,5 lb/h)

0,1 % del valor medido

DN 50 (2)

-0,010

0,00027

1:1

8000 kg/h (17637 lb/h)

0,1 % del valor medido

DN 80 (3)

-0,006

0,00019

Desviacin del valor medido y precisin bsica para lquidos

Caudal msico

Caudal volumtrico

FCx330

FCx350

0,4 % del valor

0,15 % del valor

medido

medido

0,25 % del valor

0,1 % del valor

medido

medido (opcional)

0,4 % del valor


medido
0,25 % del valor

DN 100 (4)

-0,009

0,00024

DN 150 (6)

-0,035

0,00045

0,15 % del valor


medido

medido
Densidad

0,010 kg/l 1)

0,002 kg/l 1)
0,001 kg/l 2)
0,0005 kg/l (opcional) 3)

Reproducibilidad
para la densidad

0,002 kg/l 1)
0,002 kg/l

0,001 kg/l 2)
0,00025 kg/l (opcional)
3)

Temperatura

1K

0,5 K

1) Para el rango de densidad 0,5 1,8 kg/dm


2) Igual que la indicada en 1 y para el rango de temperatura de fluido -10 50 C
(14 122 F)
3) Igual que la indicada en 2 y tras un ajuste de campo en condiciones de
funcionamiento

Desviacin del valor medido y precisin bsica para gases

Caudal msico
Temperatura

FCx330

FCx350

1 % del valor

0,5 % del valor

medido

medido

1K

0,5 K

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 69

10.4 Datos tcnicos


10.4.1 Prdida de presin

10.4.4 Conexiones a proceso


Diseo bridado segn EN / ASME / JIS
Tri-Clamp segn DIN 32676 (ISO 2852)
DN 15 100 (1/2 4): lnea 3
BPE Tri-Clamp
DN 15 100 (1/2 4)
10.4.5 Presin nominal
PN 16, PN 40, PN 100, PN 160
CL 150, CL 300, CL 600, CL 900, CL 1500
La presin de servicio mxima permitida depende de la
conexin a proceso utilizada, la temperatura del fluido, los
tornillos y del material de las juntas.
10.4.6 Carcasa como dispositivo de proteccin (opcional)
Estndar:
Presin de estallido 60 bar (870 psi)
Opcional:
Presin de estallido elevada hasta 100 bar (1450 psi),
posible para los dimetros nominales DN15 ... DN100
(1/2" 4").
Presin de estallido elevada hasta 150 bar (2175 psi),
posible para los dimetros nominales DN15 ... DN80
(1/2" 3").
Hay conexiones de lavado bajo pedido.

Fig. 36: Curva de prdida de presin (medida con agua,


viscosidad: 1 mPas)

10.4.2 Rango de viscosidad


En el caso de viscosidades dinmicas 1 Pas (1000 mPas =
1000 cP), pngase en contacto con ABB.

10.4.7 Directiva de equipos apresin


Evaluacin de conformidad segn la categora III, grupo de
fluidos 1, gas
Asegrese de que el material de la tubera de medida sea
resistentes a los efectos corrosivos del fluido.

10.4.3 Lmites de temperatura C (F)


IMPORTANTE (NOTA)
Si el dispositivo se utiliza en zonas potencialmente explosivas,
se debern observar las especificaciones de temperatura
adicionales del captulo "Datos tcnicos relevantes para la
proteccin contra explosiones".
Rango de temperatura del fluido
FCx330: -50 160 C (-58 320 F)
FCx350: -50 200 C (-58 392 F)
Rango de temperatura ambiente
Estndar: -20 60 C (-4 140 F)
Opcional: -40 60 C (-40 140 F)

70 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

10.4.8 Materiales para el transmisor


Caja
Metal ligero fundido, pintado
Color de la carcasa
Parte central:

RAL 7012

Tapa:

RAL 9002

Espesor de la capa de pintura:

80 120 m

10.4.9 Materiales del sensor de caudal


Partes mojadas
Acero inoxidable
1.4404 (AISI 316L) o 1.4435 (AISI 316L)

Fig. 38: Brida ASME de acero inoxidable 1.4571 / 1.4404


(316Ti / 316L) hasta DN 200 (8)

Acero inoxidable pulido


1.4404 (AISI 316L) o 1.4435 (AISI 316L) con certificado de material
del sensor de caudal conforme a EHEDG (AISI 316L)
Nickel-Alloy C4

1)

(2.4610) o Nickel-Alloy C22

1)

(2.4602)

Opcional: fabricacin conforme a NACE MR0175 y MR0103 (ISO 15156)


Carcasa
Acero inoxidable 1.4404 (AISI 316L),1.4301 (AISI 304), 1.4308
(ASTM CF8)
1) Hastelloy C es una marca registrada de Haynes International Nickel-Alloy C4 y
C22 son equivalentes a Hastelloy C4 y Hastelloy C22.

10.4.10 Cargas del material de las conexiones a proceso


Diseo

Dimetro

PSmax

TSmax

TSmin

nominal
Racor roscado (DIN

DN 15 40

40 bar

140 C

-40 C

11851)

(1/2 1 1/2)

(580 psi)

(284 F)

(-40 F)

DN 50 100

25 bar

140 C

-40 C

(2 4)

(363 psi)

(284 F)

(-40 F)

Tri-Clamp

DN 15 50

16 bar

120 C

-40 C

(DIN 32676)

(1/2 2)

(232 psi)

(248 F)

(-40 F)

DN 65 100

10 bar

120 C

-40 C

(2 1/2 4)

(145 psi)

(248 F)

(-40 F)

Fig. 39: Brida DIN de Nickel-Alloy C4 (2.4610) o Nickel-Alloy C22


(2.4602) hasta DN 200 (8)

10.4.11 Curvas de carga del material de los aparatos


bridados

Fig. 40: Brida ASME de Nickel-Alloy C4 (2.4610) o Nickel-Alloy C22


(2.4602) hasta DN 200 (8)

Fig. 37: Brida DIN de acero inoxidable 1.4571 / 1.4404 (316Ti / 316L)
hasta DN 200 (8)

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 71

11 Datos tcnicos - Transmisor


11.1 Generalidades

11.2.7 Versin de caja


Carcasa
Metal ligero fundido, pintado
Color de la carcasa
Parte intermedia:

RAL 7012

Tapa:

RAL 9002

Espesor de la capa de pintura:

80 120 m

11.2.8 Medicin de caudal de avance / retorno


El sentido de flujo se indica mediante flechas (en la pantalla
LCD del transmisor) y la salida de contacto digital (si est
programada).

Fig. 41: Transmisor FCT3xx en caja de campo

11.2 Datos tcnicos


11.2.1 Rango de medida
El rango de medida es ajustable entre 0,01 1 Qmax.
11.2.2 Tipo de proteccin
IP 65 / IP 67, NEMA 4X
11.2.3 Conexin elctrica
Racor atornillado para cables M20 x 1,5 o 1/2 NPT.
En caso del modelo de diseo remoto, la longitud mxima del
cable de seal no debe ser superior a 50 m (164 ft) (cables
ms largos: bajo demanda).

11.2.9 Indicador LCD


Indicador LCD, 2 lneas, con pantalla retroiluminada.
Ambas lneas de la pantalla LCD se pueden configurar
libremente.
Se pueden indicar los siguientes valores:
Caudal msico
Caudal volumtrico
Densidad o temperatura
Totalizacin de caudal, 7 dgitos con contador de
desbordamiento e indicacin de la unidad fsica para la
masa o el volumen.
En el caso de los modelos de diseo remoto, la caja del
transmisor puede girarse unos 180 en cada direccin. El
indicador LCD puede girarse en cuatro posiciones para
optimizar la legibilidad de la pantalla.

11.2.4 Suministro de energa


Voltaje de alimentacin

100 230 V AC, 47 63 Hz


(Tolerancia -15 % / +10 %)
20,4 26,4 V AC, 47 63 Hz
20,4 31,2 V DC
(Ondulacin armnica: 5 %)

Consumo de potencia

S 25 VA

11.2.5 Tiempo de reaccin


Como funcin escalonada del 0 ... 99 % (corresponde a 5 )
1s
11.2.6 Temperatura ambiente
Estndar: -20 60 C (-4 140 F)
Opcional: -40 60 C (-40 140 F)
En el caso de uso a temperaturas inferiores a -20 C (-4 F),
ya no se puede leer la pantalla y se recomienda utilizar el
sistema electrnico de manera que las vibraciones se
reduzcan al minimo posible.
La plena funcionalidad se obtiene a temperaturas superiores a
-20 C (-4 F).

72 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

11.2.10 Mando
El control del transmisor y la introduccin de los paramtros
se efectan a travs de las tres teclas de control del
transmisor. Alternativamente, es posible programar el aparato
con un puntero magntico cuando la caja del transmisor est
cerrada.

Salida de corriente 1 pasiva


Seal de salida

Pasiva, 4 20 mA

Carga

0 R B 600

Tensin de fuente

12 V Uq 30 V

Inseguridad de

< 0,1 % del valor medido

medicin
Terminales

31 / 32

Valores medidos

Caudal msico, flujo volumtrico, densidad y


temperatura (libremente programable
mediante el software)

Salida de corriente 2 pasiva


Seal de salida

Pasiva, 4 20 mA

Carga

0 R B 600

Tensin de fuente

12 V Uq 30 V

Inseguridad de

< 0,1 % del valor medido

medicin
Fig. 42: Manejo con puntero magntico
1 FRAM (enchufable) | 2 Puntero magntico

Terminales

33 / 34

Valores medidos

Caudal msico, flujo volumtrico, densidad y


temperatura (libremente programable
mediante el software)

11.2.11 Seguridad de datos


El almacenamiento de datos se realiza mediante la memoria
FRAM interna del transmisor. Los datos se almacenan durante
un perodo de tiempo de ms de 10 aos, sin alimentacin
elctrica adicional.
El reconocimiento de hardware y software se realiza de
acuerdo con la recomendacin NAMUR NE53.
IMPORTANTE (NOTA)
El aparato cumple la Directiva CEM 2004/108/CE y la
Directiva de baja tensin 2006/95/CE (EN 61010-1).

Fig. 43: Tensin de fuente permitida en funcin de la resistencia de


la carga a I max = 22 mA
1 Rango permitido

11.3 Especificaciones elctricas


11.3.1 Salidas de corriente
Salida de corriente 1, activa
Seal de salida

activa, 0 20 mA 4 20 mA, conmutable

Carga

0 R B 560

Inseguridad de

< 0,1 % del valor medido

medicin
Terminales

31 / 32

Valores medidos

Caudal msico, flujo volumtrico, densidad y


temperatura (libremente programable

Fig. 44

mediante el software)

IMPORTANTE (NOTA)
Informacin de fallo segn la recomendacin NAMUR NE43

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 73

11.3.2 Salida de impulsos


Salida de impulsos normalizada (mx. 5 kHz) con valor de
impulso ajustable entre 0,001 1000 impulsos por unidad. El
ancho de impulso es ajustable entre 0,1 ... 2000 ms. La salida
est separada galvnicamente de las salidas de corriente.

Funcin de

Desconexin externa de la salida

conmutacin

Puesta a cero externa del totalizador

Entrada "On"

16 V UKL 30 V

16 V UCEH 30 V DC 16 V U 30 V DC

Entrada "Off"

0 V UKL 2 V

0 V UCEL 2 V

Carga 150

Resistencia interna

Ri = 2 k

0 mA I CEH 0,2 mA

Terminales

81 / 82

Pasiva
Tensin de servicio
Corriente de servicio

11.3.4 Entradas de contacto digitales


La funcin de conmutacin puede programarse mediante el
software.

Activa

2 mA I CEL 220 mA
fmx

5 kHz

5 kHz

Ancho de impulso

0,1 2000 ms

0,1 2000 ms

Terminales

51 / 52

51 / 52

Todas las entradas y salidas estn separadas galvnicamente


entre s.

IMPORTANTE (NOTA)
Si se utiliza un totalizador mecnico recomendamos utilizar un
ancho de impulso de 30 ms y una frecuencia lmite [fmx] de
3 kHz.
11.3.3 Salidas de contacto digitales
La funcin de conmutacin puede programarse mediante el
software.
Funcin de
conmutacin

Control del sistema (contacto de reposo o


contacto de cierre).
Directo / Inverso (cerrada en el caso de
caudal directo)
Alarma Mn./Mx. (contacto de reposo o
contacto de cierre)

Salida "cerrada"

0 V UCEL 2 V
2 mA I CEL 220 mA

Salida "abierta"

16 V UCEH 30 V DC

Terminales

41 / 42

0 mA I CEH 0,2 mA

74 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

12 Datos tcnicos relevantes para la proteccin contra explosiones conforme a


ATEX / IECEx / NEPSI
12.1 Especificaciones elctricas
12.1.1 Resumen de las diferentes opciones de salida
Versiones

ATEX, IECEx, NEPSI Zona 2

ATEX, IECEx, NEPSI Zona 1

Versin I

Salida de corriente 1: activa

Salida de corriente 1: activa

Opcin de salida A1, A2, H1, H2

Salida de corriente 2: pasiva

Salida de corriente 2: pasiva

en la referencia

Salida de impulsos: activa / pasiva, conmutable

Salida de impulsos: pasiva

Entrada y salida de contacto: pasiva

Entrada y salida de contacto: pasiva

Versin II

Salida de corriente 1: pasiva

Opcin de salida A3, H3

Salida de corriente 2: pasiva


Salida de impulsos: pasiva

en la referencia

Entrada y salida de contacto: pasiva

12.1.2 Versin I: salidas de corriente activa / pasiva


Modelo: FCx3xx-A1, FCT3xx-A1 o FCx3xx-A2, FCT3xx-A2 o FCx3xx-S2, FCT3xx-S2

Salida de corriente 1, activa

Tipo de proteccin

Valores generales

Tipo de proteccin

Tipo de proteccin contra explosiones "ib"

contra explosiones

de

contra explosiones

(Zona 1)

nA (Zona 2)

funcionamiento

"e" (Zona 1)

Ui (V)

I i (mA)

Ub (V)

I b (mA)

U (V)

I (mA)

Terminales 31 / 32
El terminal 32 est
conectado a "PA"

Salida de corriente 2,

30

30

30

30

60

35

Uo

Io

Po

Co

Co pa

Lo

(V)

(mA)

(mW)

(nF)

(nF)

(mH)

20

100

500

217

3,8

Ui

Ii

Pi

Ci

Ci pa

Li

(V)

(mA)

(mW)

(nF)

(nF)

(mH)

60

100

500

2,4

2,4

0,17

30

30

30

30

60

35

30

100

760

2,4

2,4

0,17

30

65

30

65

60

35

15

30

115

2,4

2,4

0,17

30

65

30

65

60

35

15

30

115

2,4

2,4

0,17

30

10

30

10

60

35

30

60

500

2,4

2,4

0,17

pasiva
Terminales 33 / 34
El terminal 34 est
conectado a "PA"
Salida de impulsos, pasiva
Terminales 51 / 52
Salida de contacto, pasiva
Terminales 41 / 42
Entrada de contacto, pasiva
Terminales 81 / 82

Todas las entradas y salidas estn aisladas galvnicamente entre s y del suministro de energa. Solo las salidas de corriente 1 y
2 no estn aisladas galvnicamente entre s en la versin para Zona 1.

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 75

12.1.3 Versin II: salidas de corriente pasiva / pasiva


Modelo: FCx3xx-A1, FCT3xx-A1 o FCx3xx-A2, FCT3xx-A2 o FCx3xx-S2, FCT3xx-S2
Tipo de proteccin
contra explosiones
nA (Zona 2)
Ui (V)
Salida de corriente 1,

I i (mA)

Valores generales
de funcionamiento
Ub (V)

I b (mA)

Tipo de proteccin

Tipo de proteccin contra explosiones "ia"

contra explosiones

(Zona 1)

"e" (Zona 1)
U (V)

I (mA)

Ui

Ii

Pi

Ci

Ci pa

Li

(V)

(mA)

(mW)

(nF)

(nF)

(mH)

30

30

30

30

60

35

60

300

2000

0,47

0,47

0,17

30

30

30

30

60

35

60

300

2000

0,47

0,47

0,17

30

65

30

65

60

35

60

300

2000

0,47

0,47

0,17

30

65

30

65

60

35

60

300

2000

0,47

0,47

0,17

30

10

30

10

60

35

60

300

2000

0,47

0,47

0,17

pasiva
Terminales 31 / 32
Salida de corriente 2,
pasiva
Terminales 33 / 34
Salida de impulsos,
pasiva
Terminales 51 / 52
Salida de contacto,
pasiva
Terminales 41 / 42
Entrada de contacto,
pasiva
Terminales 81 / 82

Todas las entradas y salidas estn aisladas galvnicamente entre s y del suministro de energa.
Change from one to tw o c olumns

12.1.4 Condiciones especiales de conexin


Los circuitos elctricos de salida estn diseados de manera
que puedan conectarse a circuitos con o sin seguridad
intrnseca.
No estn permitidas combinaciones de circuitos elctricos con
y sin seguridad intrnseca. Al cambiar el tipo de proteccin 'e'
se deben seguir las instrucciones del captulo 6.6.7.
A lo largo de la seccin de la lnea de los circuitos
intrnsicamente seguros deber establecerse una conexin
equipotencial.
La tensin de clculo de los circuitos sin seguridad intrnseca
es UM = 60 V.
Para conectar un amplificador NAMUR, la salida de contacto y
la salida de impulsos (terminales 41 / 42, 51 / 52) pueden
conectarse internamente de manera que funcionen como
contacto NAMUR.

En el estado de entrega, los racores atornillados para cables


estn diseados en color negro. En caso de que las salidas de
seal se conecten a circuitos intrnsicamente seguros, se
recomienda que para las entradas de cable correspondientes
se utilicen las tapas de color azul claro suministradas.
IMPORTANTE (NOTA)
Si el conductor protector (PE) se conecta en el espacio de
conexin del caudalmetro, debe asegurarse de que en la zona
potencialmente explosiva no pueda producirse una diferencia
de potencial peligrosa entre el conductor protector (PE) y la
conexin equipotencial (PA).

Change from two to one c olumn

76 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

12.2 Modelo de sensor de caudal FCB3xx / FCH3xx


12.2.1 Clase de temperatura
Modelo FCx3xx-A1Y o FCx3xx-S1Y en Zona 1
40 C (104 F)

50 C (122 F)

60 C (140 F)

T1

200 C (392 F)

200 C (392 F)

200 C (392 F)

T2

200 C (392 F)

200 C (392 F)

200 C (392 F)

T3

185 C (365 F)

180 C (356 F)

180 C (356 F)

T4

125 C (257 F)

120 C (248 F)

120 C (248 F)

T5

85 C (185 F)

85 C (185 F)

75 C (167 F)

T6

65 C (149 F)

65 C (149 F)

60 C (140 F)

40 C (104 F)

50 C (122 F)

60 C (140 F)

T1

200 C (392 F)

200 C (392 F)

180 C (356 F)

T2

200 C (392 F)

200 C (392 F)

180 C (356 F)

T3

180 C (356 F)

180 C (356 F)

180 C (356 F)

T4

115 C (239 F)

115 C (239 F)

115 C (239 F)

Temperatura ambiente
Clase de temperatura

Modelo FCx3xx-A2Y o FCx3xx-S2Y en Zona 2


Temperatura ambiente
Clase de temperatura

T5

80 C(176 F)

80 C(176 F)

75 C (167 F)

T6

60 C (140 F)

60 C (140 F)

60 C (140 F)

Condiciones ambientales y de proceso:


T amb
-20 ... 60 C (-4 ... 140 F)
T amb, optional
-40 ... 60 C (-40 ... 140 F) (solo para dispositivos de diseo compacto)
T medium
-50 ... 200 C (-58 ... 392 F)
Clase de
IP 65, IP 67 y NEMA 4X
proteccin

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 77

12.2.2 Homologacin de proteccin contra explosiones ATEX / IECEx / NEPSI


Segn la versin del sensor de caudal (diseo compacto o remoto), se aplica una designacin especfica conforme a ATEX,
IECEx o NEPSI.
IMPORTANTE (NOTA)
ABB se reserva el derecho de realizar modificaciones a la designacin de proteccin contra explosiones. La codificacin exacta
se indica en la placa de caractersticas de los dispositivos.
Modelo FCx3xx-A2A o FCx3xx-S2A (diseo remoto en Zona 2)
Homologacin

Designacin

Comentarios

ATEX

II 3 G Ex nA IIC T6 T2

II 2 D Ex tD IIIC T85C .. Tmedium


IECEx o NEPSI

Ex nA IIC T6 .. T2 Gc

Ex tb IIIC T85C Tmedium


Modelo FCx3xx-A1A o FCx3xx-S1A (diseo remoto en Zona 1)
Homologacin

Designacin

Comentarios

ATEX

II 1 G Ex ia IIC T6 T2

IECEx o NEPSI

T2 Ga

II 1 D Ex ia IIIC T85C Tmedium


-

Ex ia IIIC T85C .. Tmedium Da


Modelo FCx3xx-A2Y o FCx3xx-S2Y (diseo compacto en Zona 2)
Homologacin

Designacin

Comentarios

ATEX

II 3 G Ex nA nR IIC T6 T2

II 2 D Ex tD IIIC T85C .. Tmedium


IECEx o NEPSI

Ex nA nR IIC T6 .. T2 Gc

Ex tb IIIC T85C Tmedium Db


Modelo FCx3xx-A1Y o FCx3xx-S1Y (diseo compacto en Zona 1)
Homologacin

Designacin

Comentarios

II 1/2 G Ex d e ia IIC T6 .. T2

2 salidas analgicas pasivas, salidas "ia" / "e", segn el modo

II 2 D Ex ia tb IIIC T85C .. Tmedium

de conexin especificado por el usuario

ATEX
Versin II
Versin I

II 1/2 G Ex d e ia ib IIC T6 .. T2 or II 1/2 G Ex d e ia IIC


T6 .. T2

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn el

II 2 D Ex ia ia tb IIIC T85C .. Tmedium or II 2 D Ex ia tb IIIC

modo de conexin especificado por el usuario

T85C .. Tmedium
IECEx o NEPSI
Versin II
Versin I

Ex d e ia IIC T6 .. T2 Ga/Gb

2 salidas analgicas pasivas, salidas "ia" / "e", segn el modo

Ex ia tb IIIC T85C .. Tmedium

de conexin especificado por el usuario

Ex d e ia ib IIC T6 .. T2 Ga/Gb or Ex d e ia IIC T6 .. T2 Ga/Gb

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn el

Ex ia ib tb IIIC T85C .. Tmedium or Ex ia tb IIIC T85C ..

modo de conexin especificado por el usuario

Tmedium

78 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

12.3 Transductor de diseo remoto Modelo FCT300


Condiciones ambientales y de proceso:
T amb
-20 ... 60 C (-4 ... 140 F)
Clase de
IP 65, IP 67 y NEMA 4X / Tipo 4X
proteccin
12.3.1 Homologacin de proteccin contra explosiones ATEX / IECEx / NEPSI
Segn la versin del sensor de caudal (diseo compacto o remoto), se aplica una designacin especfica conforme a ATEX,
IECEx o NEPSI.
IMPORTANTE (NOTA)
ABB se reserva el derecho de realizar modificaciones a la designacin de proteccin contra explosiones. La codificacin exacta
se indica en la placa de caractersticas de los dispositivos.
Modelo FCT3xx-Y0 (transductor montado fuera de la zona de proteccin contra explosiones, sensor de caudal en Zona 0, 1 o 2)
Homologacin

Designacin

Comentarios

ATEX

II (1) G [Ex ia] IIC

IECEx o NEPSI

[Ex ia Ga] IIC

Modelo FCT3xx-A2 o FCT3xx-S2 (transductor en Zona 2, sensor de caudal en Zona 0, 1 o 2)


Homologacin

Designacin

Comentarios

ATEX

II 3(2) G Ex nA nR [ia] IIC T6

II 2 D Ex tb [ia] IIIC T85C


IECEx o NEPSI

Ex nA nR [ia] IIC T6 Gc (Gb)

Ex tb [ia] IIIC T85C Db


Modelo FCT3xx-A1 o FCT3xx-S1 (transductor en Zona 1, sensor de caudal en Zona 0, 1 o 2)
Homologacin

Designacin

Comentarios

II 2 (1) G Ex d e ia IIC T6

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

II 2 (1) D Ex ia tb IIIC T85C

modo de conexin especificado por el usuario

II 2 (1) G Ex d e ib [ia] IIC T6 or II 2 (1) G Ex d e [ia] IIC T6

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn

II 2 (1) D Ex ib tb [ia] IIIC T85C or II 2 (1) D Ex tb [ia] IIIC T85C

el modo de conexin especificado por el usuario

Ex d e ia IIC T6 Gb (Ga)

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

ATEX
Versin II
Versin I
IECEx o NEPSI
Versin II
Versin I

Ex ia tb IIIC T85C Db (Da)

modo de conexin especificado por el usuario

Ex d e ib [ia Ga] IIC T6 Gb or Ex d e [ia Ga] IIC T6 Gb

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn

Ex ib tb [ia Da] IIIC T85C Db or Ex tb [ia Da] IIIC T85C Db

el modo de conexin especificado por el usuario

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 79

13 Datos tcnicos relevantes para la proteccin Ex segn cFMus


13.1 Resumen de las diferentes opciones de salida
Versiones

Clase I Div. 2

Clase I Div. 1

Versin I

Salida de corriente 1: activa

Salida de corriente 1: activa

Opcin de salida A1, A2, H1, H2

Salida de corriente 2: pasiva

Salida de corriente 2: pasiva

en la referencia

Salida de impulsos: activa / pasiva, conmutable

Salida de impulsos: pasiva

Entrada y salida de contacto: pasiva

Entrada y salida de contacto: pasiva

Versin II

Salida de corriente 1: pasiva

Opcin de salida A3, H3

Salida de corriente 2: pasiva


Salida de impulsos: pasiva

en la referencia

Entrada y salida de contacto: pasiva

13.2 Datos elctricos para Div. 2 / Zona 2


13.2.1 Versin I: salidas de corriente activa / pasiva y versin II: salidas de corriente pasiva / pasiva
Modelo FCx3xx-F2, FCT3xx-F2
Tipo de proteccin contra explosiones NI
Umxo (V)

Imxo

[mA]
Salida de corriente 1

30

30

30

30

30

65

30

65

30

10

Terminales 31 / 32
Salida de corriente 2
Terminales 33 / 34
Salida de impulsos
Terminales 51 / 52
Salida de contacto
Terminales 41 / 42
Entrada de contacto
Terminales 81 / 82

Todas las entradas y salidas estn aisladas galvnicamente entre s y del suministro de energa.

80 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

13.3 Datos elctricos para Div. 1 / Zona 1


13.3.1 Versin I: salidas de corriente activa / pasiva
Modelo FCx3xx-F1, FCT3xx-F1
Tipo de proteccin contra

Tipo de proteccin con IS

explosiones sin IS
Umxo (V)

Imxo

Umxo (V)

[mA]
Salida de corriente 1, activa

30

Salida de corriente 2, pasiva

P o (mW)

Co (nF)

Co PA (nF)

L 0 (mH)

[mA]

30

Terminales 31 / 32

Imxo

20

100

500

217

3,8

UMx (V)

I Mx (mA)

P i (mW)

Ci (nF)

Ci PA (nF)

L i (mH)

60

100

500

2,4

2,4

0,17

30

30

30

100

760

2,4

2,4

0,17

30

65

15

30

115

2,4

2,4

0,17

30

65

15

30

115

2,4

2,4

0,17

30

10

30

60

500

2,4

2,4

0,17

Terminales 33 / 34
Salida de impulsos, activa o pasiva
Terminales 51 / 52
Salida de contacto, pasiva
Terminales 41 / 42
Entrada de contacto, pasiva
Terminales 81 / 82

Todas las entradas y salidas estn aisladas galvnicamente entre s y del suministro de energa. Solo las salidas de corriente 1 y
2 no estn aisladas galvnicamente entre s.
13.3.2 Versin II: salidas de corriente pasiva / pasiva
Modelo FCx3xx-F1, FCT3xx-F1
Tipo de proteccin contra

Tipo de proteccin con IS

explosiones sin IS
Salida de corriente 1, pasiva

Umx (V)

Imx (mA)

Umx (V)

Imx (mA)

P i (mW)

Ci (nF)

Ci PA (nF)

L i (mH)

30

30

60

300

2000

0,47

0,47

0,17

30

30

60

300

2000

0,47

0,47

0,17

30

65

60

300

2000

0,47

0,47

0,17

30

65

60

300

2000

0,47

0,47

0,17

30

10

60

300

2000

0,47

0,47

0,17

Terminales 31 / 32
Salida de corriente 2, pasiva
Terminales 33 / 34
Salida de impulsos, activa o pasiva
Terminales 51 / 52
Salida de contacto, pasiva
Terminales 41 / 42
Entrada de contacto, pasiva
Terminales 81 / 82

Todas las entradas y salidas estn aisladas galvnicamente entre s y del suministro de energa.
Change from one to tw o c olumns

13.3.3 Condiciones especiales de conexin


Los circuitos elctricos de salida estn diseados de manera
que puedan conectarse a circuitos con o sin seguridad
intrnseca.
No estn permitidas combinaciones de circuitos elctricos con
y sin seguridad intrnseca. Al cambiar el tipo de proteccin 'e'
se deben seguir las instrucciones del captulo 6.7.5.
A lo largo de la seccin de la lnea de los circuitos
intrnsicamente seguros deber establecerse una conexin
equipotencial.
La tensin de clculo de los circuitos sin seguridad intrnseca
es UM = 60 V.

La seguridad intrnseca se mantiene siempre que durante la


conexin de circuitos elctricos externos sin seguridad
intrnseca no se supere la tensin de clculo UM = 60 V.
IMPORTANTE (NOTA)
La caja del transmisor y del sensor de caudal debe conectarse
al conductor de conexin equipotencial PA. El propietario
debe asegurar que cuando se conecte el conductor protector
PE, no se produzcan diferencias de potencial entre el
conductor protector PE y la conexin equipotencial PA.

Change from two to one c olumn

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 81

13.4 Sensor de caudal Modelo FCB300 / FCH300


13.4.1 Clases de temperatura
Modelo FCx3xx-F1, en Clase I Div. 1
40 C (104 F)

50 C (122 F)

60 C (140 F)

T1

200 C (392 F)

200 C (392 F)

200 C (392 F)

T2

200 C (392 F)

200 C (392 F)

200 C (392 F)

T3

185 C (365 F)

180 C (356 F)

180 C (356 F)

T4

125 C (257 F)

120 C (248 F)

120 C (248 F)

T5

85 C (185 F)

85 C (185 F)

75 C (167 F)

T6

65 C (149 F)

65 C (149 F)

60 C (140 F)

40 C (104 F)

50 C (122 F)

60 C (140 F)

Temperatura ambiente
Clase de temperatura

Modelo FCx3xx-F2, en Clase I Div. 2


Temperatura ambiente
Clase de temperatura
T1

200 C (392 F)

200 C (392 F)

180 C (356 F)

T2

200 C (392 F)

200 C (392 F)

180 C (356 F)

T3

180 C (356 F)

180 C (356 F)

180 C (356 F)

T4

115 C (239 F)

115 C (239 F)

115 C (239 F)

T5

80 C (176 F)

80 C (176 F)

75 C (167 F)

T6

60 C (140 F)

60 C (140 F)

60 C (140 F)

Condiciones ambientales y de proceso:


T amb
-20 ... 60 C (-4 ... 140 F)
T amb, optional
-40 ... 60 C (-40 ... 140 F) (solo para dispositivos de diseo compacto)
T medium
-50 ... 200 C (-58 ... 392 F)
Clase de
IP 65, IP 67 y NEMA 4X / Tipo 4X
proteccin
IMPORTANTE (NOTA)
En la versin de diseo remoto, la longitud del cable de seal entre el sensor de caudal y el transductor debe ser de un mnimo
de 5 m (16,4 ft). Deben colocarse juntas de conducto cada 18 in. (45 cm).

82 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

13.4.2 Homologacin de proteccin contra explosiones cFMus


Segn la versin del sensor de caudal (diseo compacto o remoto), se aplica una designacin especfica conforme a FM.
IMPORTANTE (NOTA)
ABB se reserva el derecho de realizar modificaciones a la designacin de proteccin contra explosiones. La codificacin exacta
se indica en la placa de caractersticas de los dispositivos.
Modelo FCx3xx-F2A (diseo remoto en Zona 2, Div. 2)
Homologacin

Designacin

Comentarios

FM (marking US)

NI: CL I,II,III, DIV 2, GPS ABCDEFG

CL I, ZN2, AEx nA IIC T6 ... T2


ZN 21 AEx tb IIIC T85C ... T165C
FM (marking Canada)

NI: CL I, II, III, Div 2 GPS ABCDEFG

Ex nA IIC T6 T2
Modelo FCx3xx-F2Y (diseo remoto en Zona 2, Div. 2)
Homologacin

Designacin

Comentarios

FM (marking US)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP : CL III, Div 1,2
CL I, ZN 2, AEx nA nR IIC T6 ... T2
ZN 21 AEx tb IIIC T85C ... T165C
FM (marking Canada)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP : CL III, Div 1,2
Ex nA nR IIC T6 T2
Modelo FCx3xx-F1A (diseo remoto en Zona 1, Div. 1)
Homologacin

Designacin

Comentarios

FM (marking US)

CL I, II, III, Div 1, GPS ABCDEFG

CL I, ZN0, AEx ia IIC T6 ... T2


ZN 20 AEx ia IIIC T85C ... T165C
FM (marking Canada)

CL I, II, III, Div 1, GPS ABCDEFG

Ex ia IIC T6 T2

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 83

Modelo FCx3xx-F1Y (diseo compacto en Zona 1, Div. 1)


Homologacin

Designacin

Comentarios

IS: CL I, Div 1, GPS ABCD

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

NI: CL I,II,III, DIV2, GPS ABCDEFG

modo de conexin especificado por el usuario

FM (marking US)
Versin II

XP: CL I, Div 1, GPS ABCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
CL I, ZN1, AEx d ia IIC T6
ZN 21 AEx ia tb IIIC T85C to T165C
Versin I

IS: CL I, Div 1, GPS ABCD

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn

NI: CL I, II, III, DIV2, GPS ABCDEFG

el modo de conexin especificado por el usuario

XP: CL I, Div 1, GPS ABCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
CL I, ZN 1, AEx d ia ib IIC T6 or CL I, ZN 1, AEx d ia IIC
T6
ZN 21 AEx ib ia tb IIIC T85C or ZN21 AEx tb ia IIC T6
FM (marking Canada)
Versin II

IS: CL I, Div 1, GPS ABCD

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

NI: CL I,II,III, Div 2, GPS ABCDEFG

modo de conexin especificado por el usuario

XP: CL I, Div 1, GPS BCD


DIP CL II, Div 1, GPS EFG
DIP CL III, Div 1, 2
Ex d ia IIC T6
Versin I

IS: CL I, Div 1, GPS ABCD

Salidas analgicas activa / pasiva, salidas "ib" / "e", segn

NI: CL I, II, III, Div 2, GPS ABCDEFG

el modo de conexin especificado por el usuario

XP: CL I, Div 1, GPS BCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
Ex d ia ib IIC T6 or Ex d ia IIC T6

84 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

13.5 Transmisor de diseo remoto Modelo FCT300


Condiciones ambientales y de proceso:
T amb
-20 ... 60 C (-4 ... 140 F)
Clase de proteccin IP 65, IP 67 y NEMA 4X / Type 4X
13.5.1 Homologacin Ex cFMus
Segn el diseo del sensor de caudal (compacto o remoto) se aplicar una codificacin especfica segn la norma FM (Factory
Mutual System).
IMPORTANTE (NOTA)
ABB se reserva modificaciones de la codificacin Ex. La codificacin exacta se indica en la placa de caractersticas de los
aparatos suministrados.
Modelo FCT3xx-Y0 (transmisor en general purpose y sensor de caudal en Zona 2, Div 2 o Zona 0, 1 Div 1)
Homologacin

Marca

Observacin

FM (marking US)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP: CL III, Div 1,2
FM (marking Canada)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP: CL III, Div 1,2
Modelo FCT3xx-F2 (transmisor y sensor de caudal instalados en Zona 2, Div 2)
Homologacin

Marca

Observacin

FM (marking US)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP: CL III, Div 1,2
CL I, ZN 2, AEx nA nR [ia] IIC T6
ZN 21 AEx tb [ia] IIIC T85C
FM (marking Canada)

NI: CL I, II, III, Div 2 GPS ABCDEFG

DIP: CL II Div 1 GPS EFG


DIP: CL III, Div 1,2
Ex nA nR [ia] IIC T6

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 85

Modelo FCT3xx-F1 (transmisor instalado en Zona 1, Div 1, sensor de caudal en la zona 0, 1 2, Div 2 o Div 1)
Homologacin

Marca

Observacin

IS: CL I, Div 1, GPS ABCD

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

NI: CL I, II, III, DIV2, GPS ABCDEFG

modo de conexin especificado por el usuario.

FM (marking US)
Versin II

XP: CL I, Div 1, GPS ABCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
CL I, ZN1, AEx d ia IIC T6
ZN 21 AEx ia tb IIIC T85C
Versin I

IS: CL I, Div 1, GPS ABCD

Salidas analgicas activas / pasivas, salidas "ib" / "e",

NI: CL I, II, III, DIV2, GPS ABCDEFG

segn el modo de conexin especificado por el usuario

XP: CL I, Div 1, GPS ABCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
CL I, ZN 1, AEx d ib [ia] IIC T6 or CL I, ZN1, AEx d [ia] IIC T6
ZN21 AEx ib tb [ia] IIIC T85C or ZN21 AEx tb [ia] IIC T6
FM (marking Canada)
Versin II

IS: CL I, Div 1, GPS ABCD

2 salidas analgicas pasivas, salidas "ia" / "e", segn el

NI: CL I, II, III, Div 2, GPS ABCDEFG

modo de conexin especificado por el usuario.

XP: CL I, Div 1, GPS BCD


DIP: CL II, Div 1, GPS EFG
DIP: CL III, Div 1, 2
Ex d ia IIC T6
Versin I

IS: CL I, Div 1, GPS ABCD

Salidas analgicas activas / pasivas, salidas "ib" / "e",

NI: CL I, II, III, Div 2, GPS ABCDEFG

segn el modo de conexin especificado por el usuario

XP: CL I, Div 1, GPS BCD


DIP: CL II, Div 1, GPS EFG
DIP CL III, Div 1, 2
Ex d ib [ia] IIC T6 or Ex d [ia] IIC T6

Change from one to tw o c olumns

Marcas registradas
HART es una marca registrada de HART Communication Foundation
Hastelloy C-4 es una marca registrada de Haynes International
Hastelloy C-22 es una marca registrada de Haynes International
Hastelloy C-276 es una marca registrada de Haynes International
Change from two to one c olumn

86 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Change from two to one c olumn

14 Anexo
14.1 Homologaciones y certificados
Marca CE

El modelo de aparato comercializado por nuestra empresa cumple las normas de las siguientes Directivas CE:
Directiva CEM 2004/108/CE
Directa 2006/95/CE sobre bajas tensiones
Directiva de Equipos a Presin (PED) 97/23/CE
Directiva ATEX 94/9/CE

Proteccin contra
explosin

Marca para indicar el uso conforme al fin previsto en zonas potencialmente explosivas, de acuerdo con:

Directiva ATEX (marca adicional a la marca CE)

IECEx

Normas IEC

NEPSI

cFMus Approvals for Canada and United States

IMPORTANTE (NOTA)
Todas las documentaciones, declaraciones de conformidad y certificados pueden descargarse de la pgina web
de ABB.
www.abb.com/flow

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 87

88 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 89

90 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 91

14.2 Installation diagram 3KXF002126G0009


Page 1 of 4

92 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Page 2 of 4

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 93

Page 3 of 4

94 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Page 4 of 4

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 95

Declaracin sobre la contaminacin de aparatos y componentes


La reparacin y/o el mantenimiento de aparatos y componentes se realizar solamente cuando el impreso de declaracin est
rellenado completamente.
En caso contrario es posible rechazar el envo. Esta declaracin debe ser rellenada y firmada, exclusivamente, por el personal
tcnico autorizado del propietario.
Datos referentes al cliente:
Empresa:
Direccin:
Persona de contacto:
Fax:

Telfono:
Email:

Datos referentes al equipo:


Tipo:
Motivo del envo / descripcin del defecto:

N. de serie:

Ha sido utilizado el aparato para realizar trabajos con sustancias que pueden causar un riesgo o peligro para la salud?
S
No
En el caso afirmativo indique el tipo de contaminacin! (mrquese con una cruz)
biolgico

corrosivo/irritante

inflamable (ligera /altamente inflamable)

txico

explosivo

otras sustancias nocivas

radioactivo

Qu sustancias han estado en contacto con el aparato?


1.
2.
3.

Confirmamos que los aparatos / componentes enviados han sido limpiados y estn libres de cualquier sustancia txica o
peligrosa segn el Reglamento de Sustancias Peligrosas.

Ciudad, fecha

Firma y sello

96 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Notas

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 97

Notas

98 OI/FCB300/FCH300-ES Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350

Notas

CoriolisMaster FCB330, FCB350, FCH330, FCH350 | OI/FCB300/FCH300-ES Rev. F 99

ASEA BROWN BOVERI, S.A.


Process Automation
Divisin Instrumentacin
C/San Romualdo 13
28037 Madrid
Spain
Tel:
+34 91 581 93 93
Fax:
+34 91 581 99 43
ABB S.A.
Process Automation
Av. Don Diego Cisneros
Edif. ABB, Los Ruices
Caracas
Venezuela
Tel:
+58 (0)212 2031676
Fax:
+58 (0)212 2031827
ABB Automation Products GmbH
Process Automation
Dransfelder Str. 2
37079 Goettingen
Germany
Tel:
+49 551 905-0
Fax:
+49 551 905-777
www.abb.com/flow

Nota
Nos reservamos el derecho a realizar cambios
tcnicos o modificar el contenido de este
documento sin previo aviso. En relacin a las
solicitudes de compra, prevalecen los detalles
acordados. ABB no acepta ninguna
responsabilidad por cualquier error potencial o
posible falta de informacin de este documento.
Nos reservamos los derechos de este documento,
los temas que incluye y las ilustraciones que
contiene. Cualquier reproduccin, comunicacin a
terceras partes o utilizacin del contenido total o
parcial est prohibida sin consentimiento previo
por escrito de ABB.
Copyright 2014 ABB
Todos los derechos reservados

3KXF411008R4206
Traduccin de las instrucciones originales

OI/FCB300/FCH300-ES Rev. F 05.2014