Incubators, Infant
Scope of this Product Comparison
This report covers infant incubators used for intensive, intermediate, and primary care. It also
covers transport infant incubators and one unit
that can be used as either an incubator or a
radiant warmer. Dedicated radiant warmers are
covered in the report titled WARMERS, RADIANT,
conduction (heat loss to cooler surfaces in direct contact with the infant), convection (heat loss to air moving past the infant), radiation (heat loss to cooler
objects not in direct contact with the infant), and water
evaporation (heat loss from the infants lungs and skin
surface). Whereas term neonates naturally regulate
INFANT.
UMDNS information
This Product Comparison covers the following
device terms and product codes as listed in
ECRIs Universal Medical Device Nomenclature
System (UMDNS):
Incubator/Radiant Warming Units, Infant,
Mobile [18-857]
Incubators, Infant, Mobile [17-432]
Incubators, Infant, Transport [12-114]
Purpose
An infant incubator provides a closed, controlled
environment that warms an infant by circulating
heated air over the skin. The heat is then absorbed into
the body by tissue conduction and blood convection.
Ideally, both the skin and core temperatures should be
maintained with only minor variations. Incubators
feature better temperature control than radiant warmers (overhead heating units), which can overheat or
underheat the infant. However, an infant in an incubator is less accessible to the caregiver.
At birth, an infants core and skin temperatures
tend to drop significantly because of heat loss from
211887
424-008
Mobile incubator
E-mail hpcs@ecri.org
Principles of operation
The neonate lies on a mattress in the infant compartment, which is enclosed by a clear plastic hood.
Most incubators have hand access ports with doors
that permit the infant to be handled while limiting the
introduction of cool room air. The clinician can raise or
remove the plastic hood or open a panel to gain greater
access to the infant. Some units feature an air curtain
that causes warm air to sweep past the opening.
Most incubators warm the infant by a forced or
natural flow of heated air. At least one unit supplements air convection by actively warming the incubator walls to reduce radiant heat loss. Another unit uses
a mattress of warm water, rather than a convective
airflow, to warm the infant.
Heating and humidification systems are located beneath the infant compartment. A fan or natural flow
circulates air past the heater and the temperature
measuring device, over a water reservoir used to humidify the air (if desired), and up into the infant
compartment. Most incubators are equipped with proportional heating controls that provide electrical
power to the heating coil in response to the difference
between the actual temperature and the desired temperature. Most units have two modes of operation:
air-temperature control and skin-temperature control.
With the air-temperature (manual) control, the operator sets the temperature of the air in the incubator;
changes in infant body temperature are usually measured periodically with a thermometer, and adjustments in air temperature are made accordingly. In the
skin-temperature control mode, also called the servo
(automatic) mode, a sensor is taped to the infants skin,
and the heater responds to changes in the sensor to
keep the skin temperature at the preset level.
Most units allow the user to vary relative humidity
from either a built-in reservoir or an outside source
(e.g., a humidifier that attaches to one of the inlet
ports). Although increasing the relative humidity in an
incubator can reduce evaporative heat loss, many clinicians avoid supplemental humidification because of
Transport incubator
concern that infectious organisms may proliferate in
the water reservoir.
Many incubators have one or two oxygen inlet ports
and can be equipped with optional oxygen controllers.
The incubators can also provide support and protection for oxygen cylinders when oxygen must be delivered to the infant in the incubator.
Because the room temperature of the nursery is
nearly always lower than that inside the incubator,
radiant heat loss through the incubator walls accounts
for as much as half of the infants total heat loss. In
some nurseries, a plastic heat shield is placed over the
infant inside the incubator to minimize radiant heat
loss. In addition, some incubators have double walls
separated by an air space to prevent excessive heat
loss. However, in a study comparing heat loss from
servo-regulated single- and double-walled incubators,
Bell and Rios (1983) reported that, although the double-walled incubator decreased radiant heat loss, it
increased convective heat loss; total heat loss and
metabolic heat production were the same as in singlewalled incubators.
Incubator/radiant warming units
One manufacturers device combines the thermal
capabilities of an infant incubator with the access
advantages of an infant radiant warming unit. This
hybrid unit provides a centralized control and monitoring system for both the incubator and the warming
unit; it can be converted automatically from a closed
incubator to an open bed radiant warming unit and
vice versa with little effort in a few seconds. This device
eliminates the need to transfer an infant between an
incubator and a radiant warming unit.
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
Transport incubators
During transport within a hospital or to another
facility, infants require thermal support. Transport
incubators can be operated on a variety of power
sources found in hospitals or transport vehicles (e.g.,
120 VAC, 12 VDC, 24 VDC). Most units also carry their
own backup power supplies (e.g., 12 VDC rechargeable
nickel-cadmium [Ni-Cd], gel cell, or lead-acid batteries) that power the unit for short periods in the absence
or failure of other power sources. In addition to having
a reliable portable power source, transport incubators
also need to meet strength requirements for air and
ground travel, provide isolation from external noise
and vibration, and have limited electromagnetic emissions to allow for aircraft use. Transport incubators are
typically smaller and lighter than mobile or stationary
incubators, facilitating easy maneuvering in and out
of emergency vehicles.
Reported problems
Deaths and injuries to neonates in incubators have
been linked to thermostat failure that caused incubator overheating and infant hyperthermia and to malfunctions or design defects that produced fires and
electric shock hazards.
If a surface temperature sensor does not make adequate thermal contact with the infants skin, incorrect
low temperature readings may result, and the incubator could overheat the infant in response to the false
readings. However, using a strongly adherent tape to
fasten the sensor can produce skin irritation or damage. Periodic checks of sensor placement are strongly
recommended.
Inadequate control over the amount of oxygen delivered in an incubator can cause hyperoxia or hypoxia
(excessive or inadequate oxygen, respectively, in body
tissues). Excessively high concentrations of oxygen can
contribute to retrolental fibroplasia (i.e., formation of
fibrous tissue behind the lens, leading to detachment
of the retina and arrested growth of the eye) in premature infants.
Incubators can produce excessive internal noise levels that might adversely affect infant hearing. Slamming a porthole, using a nebulizer, activating an
alarm, and tapping on the incubator hood further
increase the noise level to the infant; levels as high as
90 to 100 dB have been reported. The 1991 Association
for the Advancement of Medical Instrumentation
standard for incubators specifies that the maximum
noise level inside the infant compartment should not
exceed 60 dB during normal use.
Purchase considerations
Infant incubators typically cost between $4,000 and
$14,000, depending on accessories purchased. Higherpriced incubators may include such features as a
mechanism that minimizes the temperature drop
when the front-access panel is opened, a rail system to
mount accessories, a humidification system, and an
electrical height adjustment capability.
Hospitals can purchase service contracts or service
on a time-and-materials basis from the supplier. Service may also be available from a third-party organization. The decision to purchase a service contract should
be carefully considered. Purchasing a service contract
ensures that preventive maintenance will be performed at regular intervals, eliminating the possibility
of unexpected maintenance costs. Also, many suppliers do not extend system performance and uptime
guarantees beyond the length of the warranty unless
the system is covered by a service contract.
ECRI recommends that, to maximize bargaining
leverage, hospitals negotiate pricing for service contracts before the system is purchased.
Stage of development
Manufacturers continue to improve the design of
incubators, focusing on thermal protection. Many
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Bibliography
[standard].
Association of Womens Health, Obstetric, and Neonatal Nurses. Transition to an open crib for the very
low birth weight infant [miscellaneous]. Committee
on Practice. CC1.
British Standards Institution. Specification for nursing incubators [standard]. BS 5724:Section 2.20.
1982.
Specification for transport incubators [standard].
BS 5724:Section 2.21. 1983.
International Electrotechnical Commission. Medical
electrical equipment part 1: general requirements for safety [standard]. IEC 60601-1 (1988-12).
1988.
Medical electrical equipment part 1: general requirements for safety. Amendment 1 [standard].
IEC 60601-1-am1 (1991-11). 1991.
Medical electrical equipment part 1: general requirements for safety. Amendment 2 [standard].
IEC 60601-1-am2 (1995-03). 1995.
Medical electrical equipment part 1: general requirements for safety. Section 2. Collateral standard: electromagnetic compatibility requirements
and tests. IEC 60601-1-2 (1993-04). 1993.
Medical electrical equipment part 2: particular
requirements for the safety of baby incubators
[standard]. IEC 60601-2-19. (1990-12). 1990.
Medical electrical equipment part 2: particular
requirements for the safety of baby incubators.
Amendment 1 [standard]. IEC 60601-2-19-am1
(1996-10). 1996.
Medical electrical equipment part 2: particular
requirements for the safety of transport incubators
[standard]. IEC 60601-2-20 (1990-12). 1990.
Medical electrical equipment part 2: particular
requirements for the safety of transport incubators.
Amendment 1 [standard]. IEC 60601-2-20-am1
(1996-10). 1996.
Medical Devices Agency. Nursing incubators.
Evaluation 1998 May: No. 362.
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
High-risk equipment: outdated incubators. 1988 Feb;
17(2):50.
Reversal of access port gaskets on Hill-Rom AirShields Isolette infant incubators [hazard report].
Health Devices 1998 Sept-Oct;27(9-10):370-1.
Cable failure causes frequent alarms and loss of
control on Hill-Rom Air-Shields C2HS Isolette
infant incubators [User Experience Network].
Health Devices 1999 Aug;28(8):326-7.
Deposits may form in humidifier of Hill-Rom AirShields C2HS Isolette infant incubators [User Experience Network]. Health Devices 2001 Mar;
30(3):94.
patient is not being monitored. The manufacturer initiated a safety alert by letter dated February 28, 2001.
Verify that you have received the letter, prepaid response form, and operation and maintenance manual
replacement pages. Identify any affected product in
your inventory. Inform all caregivers and maintenance
personnel who use or service these units of the safety
alert. If you are a dealer or distributor, or if your office
supplies medical equipment to other locations within
your institution, send a copy of the notification letter
and response form to your customers or other locations
within your facility. To obtain additional copies of the
letter and response form, contact the Ohmeda Medical
safety alert coordinator by telephone at (800) 249-5026
within the U.S. or at (410) 888-5213 outside the U.S.
For further information, contact the Ohmeda Medical
safety alert coordinator by telephone at (800) 249-5026
within the U.S. or at +1 (410) 888-5213 outside the U.S.
Source: Manufacturer.
A4070 FDA has designated Class II Recall Nos. Z234/237-0 certain Hill-Rom Air-Shields K-1 infant
transport incubators, which have a faulty design of the
hand ventilator that does not indicate the true pressure inside the incubator. The manufacturer initiated
a field correction by site visits beginning October 15,
1999, and by letters dated March 3 and April 6, 2000.
Verify that you have received the letters, and identify
and isolate any affected product in your inventory.
Hill-Rom Air-Shields will dispatch service personnel
to your facility to address the subject of product improvement by offering your facility two corrective options. Source: FDA Enforcement Rep 1999 Dec 5;
Manufacturer.
Supplier information
Air-Shields
Hill-Rom Air-Shields
A Hillenbrand Co [339679]
330 Jacksonville Rd
Hatboro PA 19040-2211
Phone: (215) 675-5200, (800) 523-5756
Fax: (215) 674-3497
Internet: http://www.hill-rom.com
Hill-Rom Air-Shields (Australia)
A Hillenbrand Co [339681]
5-9 Devlin St Suite 402 Ground Floor
Ryde NSW 2112
Australia
Phone: 61 (2) 98096622
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Draeger
Draeger Ltd [157747]
The Willows
Mark Rd
Hemel Hempstead, Hertfordshire HP2 7BW
England
Phone: 44 (1442) 213542
Fax: 44 (1442) 240327
Internet: http://www.draeger.co.uk
Draeger Medical Inc [371341]
3135 Quarry Rd
Telford PA 18969
Phone: (215) 721-5400, (800) 462-7566
Fax: (215) 721-9561
Internet: http://www.draegermedical.com
DraegerMedizintechnik GmbH [374044]
Moislinger Allee 53-55
Postfach 1339
D-23542 Luebeck
Germany
Phone: 49 (451) 8820
Fax: 49 (451) 8822080
E-mail: silke.bahr@draeger.com
Internet: http://www.draeger.com
Draeger Southeast Asia Pte Ltd
Medical Div [354511]
73 Science Park Dr #02-01/04
Cintech 1
Singapore
Republic of Singapore
Phone: 65 8729278
Fax: 65 7792165
Fanem
Fanem Fabr de Apars Nac de Eletomedicina Ltda
[174321]
Avenida General Ataliba Leonel 1790
Caixa Postal 2836
02033-020 Sao Paulo-SP
Brazil
Phone: 55 (11) 69791700
Fax: 55 (11) 64122199
E-mail: indl.fanem@uol.com
Internet: http://www.fanem.com
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
Gigante Recem-Nascido
Gigante Recem-Nascido
Industria e Comercio e Representacoes Ltda
[174377]
Avenida Marechal Costa e Silva 3541
14075-610 Ribeirao Preto-SP
Brazil
Phone: 55 (16) 39691000
Fax: 55 (16) 39691001
E-mail: sac@gigante.com
Internet: http://www.gigante.com
Ginevri
Ginevri srl [306412]
via Cancelliera 25/D
I-00040 Cecchina RM
Italy
Phone: 39 (06) 934591
Fax: 39 (06) 93459393
E-mail: mail@ginevri.com
Internet: http://www.ginevri.com
Guido Rayos
Guido Rayos X SA [309753]
Calle Salcedo 5
E-28034 Madrid
Spain
Phone: 34 (91) 3581625
Fax: 34 (91) 3580869
E-mail: guido@terabyte-sl.es
Hero Medical
Hero Medical Pty Ltd [177031]
40 St David St
Fitzroy VIC 3065
Australia
Phone: 61 (3) 94195122
Magna
Magna Medical Systems Inc [377697]
7200 NW 7th St
Miami FL 33126
Phone: (305) 261-2211
Fax: (305) 261-4492
E-mail: magnamedical@magnamedical.com
Internet: http://www.magnamedical.com
Medap
Heraeus Medical GmbH [254159]
Heraeusstrasse 12-14
D-63450 Hanau
Germany
Phone: 49 (6181) 355100
Fax: 49 (6181) 35809
E-mail: public-relations@europe.heraeus.com
Internet: http://www.heraeus.de
Medicor
Medicor Elektronika Rt [277820]
Illatos utca 9
H-1097 Budapest
Hungary
Phone: 36 (1) 2806342
Fax: 36 (1) 2806336
E-mail: medelektronika@mail.datanet.hu
Internet: http://www.datanet.hu/medelektronika
Mediprema
Mediprema SA [151024]
Technopole 4 avenue Marcel Merieux
boite postale 3
F-37913 Tours Cedex 9
France
Phone: 33 (47) 284787
E-mail: mediprema@mediprema.com
Internet: http://www.mediprema.com
Medix
Medix icsa [157008]
Jose Arias 293
1672 Villa Lynch San Martin
Pcia de Buenos Aires
Argentina
Phone: 54 (11) 47545555
Fax: 54 (11) 47541713
E-mail: medix@medix.com.ar
Internet: http://www.medix.com.ar
Ohmeda
Datex-Ohmeda (Finland)
An Instrumentarium Co [351977]
Kuortaneenkat 2
Postilokero 900
FIN-00031 Helsinki
Finland
Phone: 358 (10) 39411
Fax: 358 (9) 1463310
E-mail: webmaster@datex-ohmeda.com
Internet: http://www.datex-ohmeda.com
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Seward
TSE
Ohmeda Medical
An Instrumentarium Co [347601]
8880 Gorman Rd
Laurel MD 20723
Phone: (410) 888-5200, (800) 527-9209
Fax: (410) 888-0545
Internet: http://www.ohmedamedical.com
Parker
Parker Healthcare Pty Ltd [354556]
2 Monomeeth Dr
Mitcham VIC 3132
Australia
Phone: 61 (3) 98735822
Fax: 61 (3) 98735911
E-mail: ecu@parkerhealth.com.au
Internet: http://www.parkerhealth.com.au
Internal noise, dB: The sound level inside the incubator, measured in decibels. The values listed include
only the noise generated by the incubator equipment itself.
Abbreviations:
AAMI Association for the Advancement of Medical Instrumentation
ANSI American National Standards Institute
ARO After receipt of order
Phoenix Medical
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
IEC International Electrotechnical Commission
IMV Intermittent mandatory ventilation
ISO International Organization for Standardization
IV Intravenous
LCD Liquid crystal display
LED Light-emitting diode
SIMV Synchronized intermittent mandatory
breathing
UL Underwriters Laboratories
VLBW Very low birth weight
Note:
The data in the charts derive from suppliers specifications and have not been verified through
independent testing by ECRI or any other agency.
Because test methods vary, different products specifications are not always comparable. Moreover,
About ECRI . . .
ECRI is a nonprofit health services research agency and a Collaborating Center of the World Health
Organization, providing information and technical assistance to the healthcare community to support
safe and cost-effective patient care for more than 25 years. The results of ECRIs research and
experience are available through its publications, information systems, databases, technical assistance
program, laboratory services, seminars, and fellowships.
Our full-time staff includes a wide range of specialists in healthcare technology, hospital administration, financial analysis, risk management, and information and computer science, as well as
hospital planners, attorneys, physicists; biomedical, electrical, electronic, chemical, mechanical, and
registered engineers; physicians; basic medical scientists; epidemiologists and biostatisticians; and
writers, editors, and communications specialists.
Underlying ECRIs knowledge base in healthcare technology are its integrity and objectivity. ECRI
accepts no financial support from medical product manufacturers, and no employee may own stock
in or consult for a medical equipment or pharmaceutical company.
The scope of ECRIs resources extends far beyond technology. ECRI keeps healthcare professionals,
manufacturers, legal professionals, information specialists, and others aware of the changing trends
in healthcare, healthcare standards and regulations, and the best ways to handle environmental and
occupational health and safety issues. ECRI also advises on management issues related to healthcare
cost containment, accreditation, risk management, human resources, quality of care, and other
complex topics.
ECRI has more than 35 publications, databases, software, and services to fulfill the growing need
for healthcare information and decision support. They focus on three primary areas: healthcare
technology, healthcare risk and quality management, and healthcare environmental management.
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
AIRBORNE LIFE
SUPPORT
In-house Infant
Transport Sys 185A
AIRBORNE LIFE
SUPPORT
Multipurpose Infant
Transport Sys 20H
AIR-SHIELDS
AIR-SHIELDS
C300 ISOLETTE
Infant Incubator
C400 QT ISOLETTE
Infant Incubator
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Yes
Yes
Yes
Yes
CE MARK
Yes
Yes
No
Yes
MOBILE OR TRANSPORT
Ground transport
Mobile/transport
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
30-38.9 (0.1)
30-38.9 (0.1)
22-38 (0.1)
20-38.5 (0.1)
No
No
NA
NA
Digital
20-45
Digital
20-45
Digital
17-45
Digital
17-45
Digital
20-45
Digital
20-45
NA
NA
NA
NA
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
No
No
Yes
Yes
Yes
Yes
Yes
No
No
Yes
Yes
Yes
Yes
No
NA
NA
Yes
Yes
Yes
Yes
Yes
NA
NA
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
Low battery
indicator
Low battery
indicator
HAND PORTS
TUBING PORTS
1 respiratory,
4 accessory
1 respiratory,
4 accessory
1 respiratory,
2 accessory
2 respiratory,
4 accessory
O2 INLET PORTS/
CONTROLLERS
Optional/optional
2/1-way check
valves
1 standard/no
1 standard/no
SUPPLEMENTAL
HUMIDITY
Adjustable
Optional
Not specified
Optional
Not specified
Yes
No
Yes
Yes *
PHOTOTHERAPY
Optional
Optional
No
No
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
HEATER POWER
INDICATOR
10
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
AIRBORNE LIFE
SUPPORT
In-house Infant
Transport Sys 185A
AIRBORNE LIFE
SUPPORT
Multipurpose Infant
Transport Sys 20H
AIR-SHIELDS
AIR-SHIELDS
C300 ISOLETTE
Infant Incubator
C400 QT ISOLETTE
Infant Incubator
INTERNAL NOISE, dB
<60
<60
60
Typically 49
105-125/210-250
105-125/210-250
100/120/240
100/120/240
BATTERY, TYPE
NA
DOUBLE WALL
Yes
Yes
No
Yes
22.9 (9)
22.9 (9)
30.5 (12)
40.6 (16)
61 x 30.5
(24 x 12)
61 x 30.5
(24 x 12)
63.5 x 33
(25 x 13)
62.9 x 34.9
(24.75 x 13.75)
22.9 x 51.2
(9 x 20.3)
12.7 (5)
22.9 x 51.2
(9 x 20.3)
12.7 (5)
22.9 x 58.4
(9 x 23)
10.2 (4)
32.5 x 85.1
(12.8 x 33.5)
12.7 (5)
Variable, depends on
configuration
Variable, depends on
configuration
54.4 (120)
76 (167.6)
1 year
1 year
1 year
1 year
30-45 days
Not specified
30-45 days
Not specified
30 days
$7,500-11,400
30 days
$10,000-14,900
Not specified
Not specified
January to December
Not specified
Not specified
January to December
1982
Not specified
January to December
1994
Not specified
January to December
Available with
patient monitor,
ventilator, and
ventilator monitor;
mobile cart; storage
shelves available;
optional ambulance
cart. Meets
requirements of CSA,
ETL, and UL.
Available with
patient monitor,
ventilator, and
ventilator monitor;
detachable accessory
and air/oxygen cart;
capacity for M-, E-,
or Kevlar cylinders;
optional
noncollapsible cart.
Meets requirements
of CSA, ETL, FAA,
UL, and United
States Air Force.
Optional intubation
access door, battery
pack for transport
within a hospital,
IV pole.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
Warranty
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
11
AIR-SHIELDS
AIR-SHIELDS
AIR-SHIELDS
AIR-SHIELDS
C450 QT ISOLETTE
Infant Incubator
ISOLETTE Infant
Incubator
ti 500 Globetrotter
Transport Systems
ti 500 Infant
Transport Incubator
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Yes
Yes
Yes
Yes
CE MARK
Yes
Yes
Yes
Yes
MOBILE OR TRANSPORT
Mobile
Mobile
Transport
Transport
TEMPERATURE CONTROL
RANGE, C
Air (increments)
20-38.5 (0.1)
20-39 (0.1)
22-38 (0.1)
22-38 (0.1)
Skin (increments)
34-37.9 (0.1)
34-38 (0.1)
No
No
Digital
17-45
Digital
17-45
Digital
17-45
Digital
17-45
Digital
17-45
Digital
17-45
Digital
17-45
Digital
17-45
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
Trended display *
HAND PORTS
TUBING PORTS
2 respiratory,
4 accessory
2 respiratory,
4 accessory
1 respiratory,
2 accessory
1 respiratory,
2 accessory
O2 INLET PORTS/
CONTROLLERS
1 standard/no
1 standard/optional
servo-controlled
oxygen
1 standard/NA
1 standard/NA
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Yes **
Yes
Optional servo
humidity
Yes
No
Yes
No
PHOTOTHERAPY
No
No
No
No
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
HEATER POWER
INDICATOR
12
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
AIR-SHIELDS
AIR-SHIELDS
AIR-SHIELDS
AIR-SHIELDS
C450 QT ISOLETTE
Infant Incubator
ISOLETTE Infant
Incubator
ti 500 Globetrotter
Transport Systems
ti 500 Infant
Transport Incubator
INTERNAL NOISE, dB
Typically 49
<47
<56
<56
100/120/240
100/120/220/240
120/220/240
110/220/220/240
BATTERY, TYPE
NA
NA
Sealed lead-acid
gel, 12 V/24 A-h
vented
Sealed lead-acid
gel, 12 V/24 A-h
vented
DOUBLE WALL
Yes
Yes
Yes
Yes
40.6 (16)
43.2 (17)
31 (12.2); 33 (13)
high hood
31 (12.2); 33 (13)
high hood
140 x 116 x 56
(55.1 x 45.7 x 22)
[54.1-56.1] x 56.6
x 95.9 ([21.3-22.1]
22.3 x 37.8) **
[54.1-56.1] x 56.6
x 95.9 ([21.3-22.1]
22.3 x 37.8) **
62.9 x 34.9
(24.75 x 13.75)
40.6 x 80
(16 x 31.5)
63.5 x 33
(25 x 13)
63.5 x 33
(25 x 13)
32.5 x 85.1
(12.8 x 33.5)
12.7 (5)
35.6 x 85.1
(14 x 33.5)
12.7 (5)
[21-26] x 66.5
([8.3-10.2] x 26.2)
12.7 (5)
[21-26] x 66.5
([8.3-10.2] x 26.2)
12.7 (5)
WEIGHT, kg (lb)
76 (167.6)
75.3 (166)
PURCHASE INFORMATION
Warranty
1 year
1 year
30 days
$10,000-14,900
30 days
$10,500-16,900
1994
Not specified
January to December
1996
Not specified
January to December
1995
Not specified
January to December
1994
Not specified
January to December
Servo-controlled O2
and humidity; twin
capable; dual access
doors with dual air
curtain; microprocessor controller
with message center;
trended data display
for air, skin 1/2,
O2, humidity; in-bed
scale; IV pole;
drawer module; remote alarm module;
variable-height
stand.
Power on/standby
mode; power source
indicator; battery
charge status; skin
temperature display;
2 E- or D-size tank
mounts; IV pole;
accessory shelf;
optional second
battery.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
49.2 (108.5)
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
13
ATOM MEDICAL
FAILED TO RESPOND *
TC-500 Transport
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-85 Infant
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-850 Infant
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-850W Infant
Incubator
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Not specified
Not specified
Yes
Yes
MOBILE OR TRANSPORT
Both
Mobile
Mobile
Mobile
25-38 (0.1)
25-38 (0.1)
25-38 (0.1)
34-38 (0.1)
34-38 (0.1)
34-38 (0.1)
Digital
30-45
Digital
20-42
Digital
20-42
Digital
20-42
NA
NA
Digital
30-42
Digital
30-42
Digital
30-42
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
NA
NA
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HEATER POWER
INDICATOR
1-point
12-point
10-point
10-point
HAND PORTS
2, additional hand
ports optional
TUBING PORTS
O2 INLET PORTS/
CONTROLLERS
1/optional
1/optional
1/optional
1/optional
Yes
Ambient to 70%
relative humidity
Yes
Ambient to 80%
relative humidity
Yes
Ambient to 80%
relative humidity
Yes
Ambient to 70%
relative humidity
Optional
Optional
Optional
Optional
TEMPERATURE CONTROL
RANGE, C
Air (increments)
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
SUPPLEMENTAL
HUMIDITY
Adjustable
PHOTOTHERAPY
14
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
ATOM MEDICAL
FAILED TO RESPOND *
TC-500 Transport
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-85 Infant
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-850 Infant
Incubator
ATOM MEDICAL
FAILED TO RESPOND *
V-850W Infant
Incubator
INTERNAL NOISE, dB
<50
<48
<48
<48
110/240
100/120/220/240
100/120/220/240
100/120/220/240
BATTERY, TYPE
PP3 alarm
Not specified
Not specified
Not specified
DOUBLE WALL
Optional
No
No
Yes
23.5 (9.25)
34.5 (13.4)
34.5 (13.4)
31.5 (12.4)
42.5 x 83.5 x 43
(16.7 x 32.9 x 16.9)
132 x 98 x 56
(52 x 38.6 x 22)
132 x 98 x 56
(52 x 38.6 x 22)
132 x 98 x 56
(52 x 38.6 x 22)
57.5 x 30.5
(22.6 x 12)
64 x 34
(25.2 x 13.4)
64 x 34
(25.2 x 13.4)
64 x 34
(25.2 x 13.4)
64 x 19.5
(25.2 x 7.7) **
Not specified
28.5 x 65.9
(11.2 x 25.9)
Not specified
28.5 x 65.9
(11.2 x 25.9)
Not specified
28.5 x 65.9
(11.2 x 25.9)
Not specified
WEIGHT, kg (lb)
21 (46.3)
76 (168) with
cabinet stand
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
Not specified
Y941,000 (US$8,208)
Not specified
Y880,000 (US$7,676)
1996
Not specified
Not specified
1994
Not specified
Not specified
Not specified
Y1,260,000
(US$10,990)
1986
Not specified
Not specified
Not specified
Y1,300,000
(US$11,339)
1986
Not specified
Not specified
Optional IV pole,
MF rail, in-bed
scale, phototherapy.
MF rail, in-bed
scale; optional IV
pole, variableheight stand,
phototherapy.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
15
ATOM MEDICAL
FAILED TO RESPOND *
V-2100 G Advanced
Infant Incubator
BERTOCCHI
FAILED TO RESPOND **
GB 35 BR
BERTOCCHI
FAILED TO RESPOND **
GB 35 RO
BERTOCCHI
FAILED TO RESPOND **
GB 58 RO
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Yes
Not specified
Not specified
Not specified
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Transport
30-37 (0.1)
30-37 (0.1)
30-37 (0.1)
30-37 (0.1)
30-37 (0.1)
30-37 (0.1)
Digital
20-42
Digital
30-37
Digital
30-37
Digital
30-37
Digital
30-42
Digital
30-37
Digital
30-37
Digital
30-37
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Yes
No
Yes
Yes
Yes
Yes
No
Yes
No
Yes
Yes
Yes
Yes
No
No
No
Yes
Yes
Low battery
BACKUP THERMOSTAT
Yes
Not specified
Not specified
Not specified
HEATER POWER
INDICATOR
10-point
LED
LED
LED
HAND PORTS
TUBING PORTS
O2 INLET PORTS/
CONTROLLERS
1/optional
1/not specified
1/not specified
1/not specified
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Yes
Yes
Yes
Yes
Yes
No
NA
PHOTOTHERAPY
Optional
Optional
Optional
Optional
TEMPERATURE CONTROL
RANGE, C
Air (increments)
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
16
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
ATOM MEDICAL
FAILED TO RESPOND *
V-2100 G Advanced
Infant Incubator
BERTOCCHI
FAILED TO RESPOND **
GB 35 BR
BERTOCCHI
FAILED TO RESPOND **
GB 35 RO
BERTOCCHI
FAILED TO RESPOND **
GB 58 RO
INTERNAL NOISE, dB
<45
<60
<60
<60
100/120/220/240
220
220
220
BATTERY, TYPE
NA
NA
NA
12 V
DOUBLE WALL
Yes
Optional
Optional
Optional
40.3 (16)
34 (13.4)
34 (13.4)
24 (9.4)
135 x 100 x 58
(53.1 x 40 x 22.8)
99 x 193 x 104
(39 x 76 x 41)
Not specified
74 x 36
(30.2 x 14.2)
63 x 34.5
(24.8 x 13.6)
Not specified
24.5 x 56.5
(9.6 x 22.2)
40.5 x 85.1
(15.9 x 33.5)
12.7 (5)
24 x 52
(9.4 x 20.5)
8 (3)
24 x 52
(9.4 x 20.5)
Not specified
20 x 54.5
(8 x 21.5)
10 (4)
110 (242.6)
115 (253.6)
50 (110.3)
Not specified
1 year
1 year
1 year
Not specified
Not specified
6 weeks
Not specified
6 weeks
Not specified
6 weeks
Not specified
Not specified
Not specified
Not specified
1989
Not specified
January to December
1989
Not specified
January to December
1990
Not specified
January to December
Servo-controlled
O2 and humidity;
dual access doors
with dual air
curtain; in-bed
scale; IV pole;
drawer module;
variable-height
stand.
Meets requirements
of IEC.
Meets requirements
of IEC.
Meets requirements
of IEC.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
Warranty
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
17
BERTOCCHI
FAILED TO RESPOND *
GB 58 RVO
BIO MS
BIO MS
BIO MS
NATISSE AR 500
Intensive Care
NATISSE AR 500
Intermediate Care
NATISSE AR 500
Primary Care
WHERE MARKETED
Worldwide
Africa, Asia,
Europe, Middle East,
Russia
Africa, Asia,
Europe, Middle East,
Russia
Africa, Asia,
Europe, Middle East,
Russia
FDA CLEARANCE
Not specified
No
No
No
CE MARK
Not specified
Yes
Yes
Yes
MOBILE OR TRANSPORT
Transport
Mobile
Mobile
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
30-37 (0.1)
28-39 (0.1)
28-39 (0.1)
28-39 (0.1)
Skin (increments)
30-37 (0.1)
35-38 (0.1)
35-38 (0.1)
35-38 (0.1)
Digital
30-37
LED 8 segments
10-45
LED 8 segments
10-45
LED 8 segments
10-45
Digital
30-37
LED 8 segments
10-45
LED 8 segments
10-45
LED 8 segments
10-45
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
No
No
No
Yes
Yes
Low battery
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Not specified
No
No
No
HEATER POWER
INDICATOR
LED
LED
LED
LED
HAND PORTS
TUBING PORTS
O2 INLET PORTS/
CONTROLLERS
1/not specified
Yes/fixed at 40%
Yes/fixed at 40%
Yes/fixed at 40%
SUPPLEMENTAL
HUMIDITY
Adjustable
No
NA
Yes
40-70%, 10%
Yes
40-70%, 10%
Yes
40-70%, 10%
PHOTOTHERAPY
Optional
Optional Futura
Optional Futura
Futura
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
18
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
BERTOCCHI
FAILED TO RESPOND *
GB 58 RVO
BIO MS
BIO MS
BIO MS
NATISSE AR 500
Intensive Care
NATISSE AR 500
Intermediate Care
NATISSE AR 500
Primary Care
<45
<45
<45
INTERNAL NOISE, dB
<60
220
230
230
230
BATTERY, TYPE
12 V
NA
NA
NA
DOUBLE WALL
Optional
No (optional thermal
shield)
No (optional thermal
shield)
No (optional thermal
shield)
24 (9.4)
40 (15.7)
40 (15.7)
40 (15.7)
140 x 105 x 57
(55 x 41 x 22.4)
140 x 105 x 57
(55 x 41 x 22.4)
140 x 105 x 57
(55 x 41 x 22.4)
24.5 x 56.5
(9.6 x 22.2)
60 x 30
(23.6 x 11.8)
60 x 30
(23.6 x 11.8)
60 x 30
(23.6 x 11.8)
20 x 54.5
(8 x 21.5)
10 (4)
16 x 16
(6.3 x 6.3)
12.5 (4.9)
16 x 16
(6.3 x 6.3)
12.5 (4.9)
16 x 16
(6.3 x 6.3)
12.5 (4.9)
WEIGHT, kg (lb)
56 (123.5)
87 (191.8)
84 (185.2)
82 (180.8)
PURCHASE INFORMATION
Warranty
1 year
15 months
15 months
15 months
6 weeks
Not specified
3 weeks
Not specified
3 weeks
Not specified
3 weeks
Not specified
1990
Not specified
January to December
1996
Not specified
Not specified
1996
Not specified
Not specified
1996
Not specified
Not specified
Ventilator is
incorporated into
the incubator.
Meets requirements
of IEC.
Optional variable
height, accessories.
Optional
accessories.
Optional storage
(right and left),
IV pole.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
19
DRAEGER
FANEM
GIGANTE RECEMNASCIDO
ALG
GIGANTE RECEMNASCIDO
LUX
Caleo
C-186/TS Infant
Incubator
WHERE MARKETED
Worldwide
Not specified
Not specified
FDA CLEARANCE
CE MARK
Yes
Not specified
Not specified
Not specified
Yes
Not specified
Not specified
Not specified
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
20-39 (0.1)
25-37/39 (0.1)
28-37 (0.2)
28-37 (0.1)
Skin (increments)
34-38 (0.1)
34-37/38 (0.1)
35-37 (0.2)
NA
Digital
13-42
Digital
21-45
Digital
28-39
Digital
28-37
Not specified
13-43
Digital
21-45
Digital
28-37
NA
NA
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
NA
NA
Yes
Yes
No
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HEATER POWER
INDICATOR
No
Vertical bar
Vertical bar
HAND PORTS
TUBING PORTS
10
2 respiratory,
2 accessory
1 respiratory,
2 accessory
Servo-controlled/
optional
1/not specified
Yes/not specified
1/not specified
Yes
30-99%, increments
of 1%
Yes
Yes
Yes
Yes
Yes
No
Optional PT4000
Optional
Optional
Optional
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
O2 INLET PORTS/
CONTROLLERS
SUPPLEMENTAL
HUMIDITY
Adjustable
PHOTOTHERAPY
20
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
DRAEGER
FANEM
GIGANTE RECEMNASCIDO
ALG
GIGANTE RECEMNASCIDO
LUX
Caleo
C-186/TS Infant
Incubator
INTERNAL NOISE, dB
Not specified
50-55
40-50
40-50
110/127/230/240
110-120, 220-240
110/120/220/240
Lithium-ion
For alarm
NA
NA
DOUBLE WALL
Optional, removable
inner walls
Yes
Optional
Optional
38 (15)
38 (15)
36 (14.2)
36 (14.2)
[122-152] x 109 x 68
([48-59.8] x 42.9 x
26.8)
136 x 59 x 100
(53.5 x 23.2 x 39.4)
136 x 59 x 100
(53.5 x 23.2 x 39.4)
64.5 x 50
(25.4 x 19.7)
63 x 34
(24.8 x 13.4)
36 x 80
(14.2 x 31.5)
36 x 80
(14.2 x 31.5)
22 x 64
(8.7 x 25.2)
10 (3.9)
25 x 84
(9.8 x 33.1)
10.2 (4)
15 x 15
(6 x 6)
Not specified
15 x 15
(6 x 6)
Not specified
WEIGHT, kg (lb)
137 (302.1)
65 (143.3)
69 (152.1)
67 (147.7)
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
Not specified
Not specified
Not specified
$4,800-8,500
2 weeks
$5,500
Not specified
$2,847
2000
Not specified
Not specified
1992 (September)
Several thousand
Not specified
1997
83
Not specified
Not specified
Not specified
Not specified
EL display; trend
analysis; integrated
x-ray tray; weighing
scale; thermomonitoring;
adjustable height;
Kangaroo Mode;
1-year calibrationfree O2 sensor;
self-cleaning mode;
Twincubator for cobedding; electronic
bed tilt; all-round
access in
emergencies.
Access door;
electronic scale;
IV pole; guard rail;
emergency oxygen
system; humidifier;
oxygen hood; 4
(3-wire) outlets;
optional shelves,
cabinet stand
(3 kinds), and
phototherapy unit.
Meets requirements
of IEC and ISO 9001.
Optional double
hood, phototherapy
unit, body stand,
monitor stand,
sensor adhesive box,
microfilter box, and
oxygen tent.
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
21
GIGANTE RECEMNASCIDO
SMJ
GIGANTE RECEMNASCIDO
TI
GINEVRI
GINEVRI
BABY SHUTTLE
TRANSPORT INCUBATOR
WHERE MARKETED
Not specified
Not specified
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Not specified
Not specified
Yes
Yes
MOBILE OR TRANSPORT
Mobile
Mobile
Transport
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
28-37 (0.1)
28-37 (0.1)
20-38 (0.1)
20-38 (0.1)
Skin (increments)
28-37 (0.1)
NA
25-38 (0.1),
optional
NA
Digital
28-39
Digital
28-37
Digital
10-45
Digital
10-45
Digital
28-37
NA
NA
Digital
10-45
NA
NA
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
No
No
Yes
Yes
No
Yes
No
NA
NA
Yes
Yes
No
Yes
Yes
Optional
Optional
Yes
Yes
Yes
Yes
Yes
NA
NA
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HEATER POWER
INDICATOR
Vertical bar
Vertical bar
HAND PORTS
TUBING PORTS
2 respiratory,
2 accessory
2 respiratory,
2 accessory
2 respiratory,
2 accessory
2 respiratory,
7 accessory
O2 INLET PORTS/
CONTROLLERS
1/not specified
1/not specified
2/digital control
system
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
No
Yes
No
Yes
Yes
Yes
40-85%
PHOTOTHERAPY
Optional
Optional
Optional
Optional
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
22
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
GIGANTE RECEMNASCIDO
SMJ
GIGANTE RECEMNASCIDO
TI
GINEVRI
GINEVRI
BABY SHUTTLE
TRANSPORT INCUBATOR
INTERNAL NOISE, dB
40-50
40-50
<53
<45
110/120/220/240
110/120/220/240
BATTERY, TYPE
NA
NA
12 V, 40 A-h
DOUBLE WALL
Optional
Optional
Optional
Optional
36 (14.2)
36 (14.2)
26 (10.2)
45 (17.7)
136 x 59 x 100
(53.5 x 23.2 x 39.4)
136 x 59 x 100
(53.5 x 23.2 x 39.4)
60 x 118 x 105
(23.6 x 46.5 x 41.3)
127 x 80 x 60
(50 x 31.5 x 23.6)
36 x 80
(14.2 x 31.5)
36 x 80
(14.2 x 31.5)
55 x 25
(21.7 x 9.8)
60 x 35
(23.6 x 13.8)
15 x 15
(6 x 6)
Not specified
15 x 15
(6 x 6)
Not specified
30 x 19
(11.8 x 7.5)
12.5 (4.9)
37 x 65
(14.6 x 25.6)
10 (3.9)
WEIGHT, kg (lb)
67 (147.7)
67 (147.7)
40 (88.2);
60 (132.3) with
trolley
32 (70.6); 37-51
(81.6-112.5) with
trolley
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
Not specified
$4,350
Not specified
$3,278
4-6 weeks
Not specified
4-6 weeks
Not specified
Not specified
Not specified
Not specified
Not specified
Not specified
Not specified
1997
Not specified
January to December
1995
Not specified
January to December
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
23
GINEVRI
GINEVRI
GINEVRI
GUIDO RAYOS
NESTORET 5050
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Yes
Yes
Yes
Not specified
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
20-38 (0.1)
20-38 (0.1)
20-38 (0.1)
Skin (increments)
25-38 (0.1)
25-38 (0.1)
25-38 (0.1)
20-37 (0.1),
override to 39
33-37 (0.1),
override to 38
Digital
10-45
Digital
10-45
Digital
10-45
Digital
10-50
Digital
10-45
Digital
10-45
Digital
10-45
Digital
10-50
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
HAND PORTS
TUBING PORTS
2 respiratory,
7 accessory
2 respiratory,
7 accessory
2 respiratory,
7 accessory
2 respiratory,
2 accessory
2; 1 limited O2
concentration, 1 up
to 100% O2 concentration/integrated
O2 monitoring with
alarms; optional O2
servo-regulator
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
40-85%
Yes
40-85% with digital
control system
Yes
40-90% with servocontrol system
Yes (monitored)
Yes
PHOTOTHERAPY
Optional
Optional
Optional
Optional
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
HEATER POWER
INDICATOR
O2 INLET PORTS/
CONTROLLERS
24
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
GINEVRI
GINEVRI
GINEVRI
GUIDO RAYOS
NESTORET 5050
INTERNAL NOISE, dB
<45
<45
<45
<60
BATTERY, TYPE
See footnote *
DOUBLE WALL
Optional
Optional
Optional
Optional
45 (17.7)
45 (17.7)
45 (17.7)
35.5 (14)
127 x 80 x 60
(50 x 31.5 x 23.6)
127 x 80 x 60
(50 x 31.5 x 23.6)
127 x 80 x 60
(50 x 31.5 x 23.6)
131 x 90 x 59
(51.6 x 35.4 x 23.2)
60 x 35
(23.6 x 13.8)
60 x 35
(23.6 x 13.8)
60 x 35
(23.6 x 13.8)
66 x 35
(26 x 13.8)
37 x 65
(14.6 x 25.6)
10 (3.9)
37 x 65
(14.6 x 25.6)
10 (3.9)
37 x 65
(14.6 x 25.6)
10 (3.9)
25.5 x 78.5
(10 x 30.9)
Not specified
32 (70.6); 37-51
(81.6-112.5) with
trolley
32 (70.6); 37-51
(81.6-112.5) with
trolley
32 (70.6); 37-51
(81.6-112.5) with
trolley
77 (169.8)
1 year
1 year
1 year
1 year
4-6 weeks
Not specified
4-6 weeks
Not specified
4-6 weeks
Not specified
30-45 days
Not specified
1995
Not specified
January to December
1996
Not specified
January to December
1996
Not specified
January to December
Not specified
Not specified
January to December
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
Warranty
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
25
MAGNA
MAGNA
NESTOTRANS
HERO MEDICAL
FAILED TO RESPOND *
DMH 2-Stationary
ICT-2000
ICT-2001
WHERE MARKETED
Worldwide
Asia, Europe
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Yes
Yes
CE MARK
Not specified
Not specified
Not specified
Not specified
MOBILE OR TRANSPORT
Transport
Not specified
Mobile
Transport
20-37 (0.1),
override to 39
33-37 (0.1),
override to 38
21-38 (0.1)
20-39 (0.1)
34-39 (0.1)
21-38 (0.1)
20-39 (0.1)
34-39 (0.1)
Digital
10-50
Digital
21-38
Digital
20-42
Digital
20-42
Digital
10-50
Digital
21-38
Digital
20-42
Digital
20-42
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
No
No
LED
HAND PORTS
TUBING PORTS
1 respiratory,
2 accessory
To customer
specifications
2 respiratory,
2 accessory
2 respiratory,
2 accessory
2; 1 limited O2
concentration, 1 up
to 100% O2 concentration/integrated
O2 monitoring with
alarms
Yes/yes
1 standard/control
system
1 standard/control
system
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
No
Not specified
Not specified
Yes
Yes
Yes
Yes
PHOTOTHERAPY
Not specified
Not specified
Optional
Optional
TEMPERATURE CONTROL
RANGE, C
Air (increments)
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
HEATER POWER
INDICATOR
O2 INLET PORTS/
CONTROLLERS
GUIDO RAYOS
26
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
MAGNA
MAGNA
NESTOTRANS
HERO MEDICAL
FAILED TO RESPOND *
DMH 2-Stationary
ICT-2000
ICT-2001
INTERNAL NOISE, dB
<60
55
65
50
110/240
110/220
110/220
BATTERY, TYPE
12 V
4 AA
12 V
12 V
DOUBLE WALL
Optional
Not specified
Optional
Standard
28.5 (11.2)
38 (15)
37 (14.6)
26.5 (10.4)
191 x 57 x 115
(75.2 x 22.4 x 45.3)
137 x 93 x 52
(54 x 36.6 x 20.5)
See footnote **
99 x 61 x 180
(39 x 24 x 70.9)
53 x 26.5
(20.9 x 10.4)
69.5 x 37.5
(27.4 x 14.8)
64 x 35.5
(25.2 x 14)
64 x 35.5
(25.2 x 14)
30 x 94 (11.8 x 37)
26 x 94 (10.2 x 37)
Caster, cm (in)
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
GUIDO RAYOS
Not specified
34 x 70
(13.4 x 27.6)
Not specified
10 (3.9)
10 (3.9)
WEIGHT, kg (lb)
28 (11)
73 (161)
78 (172)
115 (253.6)
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
30-45 days
Not specified
Not specified
$4,050-6,000
Not specified
$7,017
Not specified
$12,849
Not specified
Not specified
January to December
1992
Not specified
Not specified
Not specified
Not specified
Not specified
Not specified
Not specified
Not specified
Optional shelves,
cabinet, IV pole,
integral oxygen
controller, double
microprocessor, and
multi-incubator
monitoring.
None specified.
None specified.
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
27
MEDAP
FAILED TO RESPOND *
T1500
MEDICOR
MEDICOR
MEDICOR
BLF-2001
ICB-145
ICB-145i
WHERE MARKETED
Not specified
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Not specified
Yes
Yes
Yes
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
32-40 (2)
Fluid thermometer
Not specified
Digital
40-50
Digital
40-50
Digital
40-50
NA
NA
Digital
40-50
NA
NA
Digital
40-50
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
No
NA
NA
NA
NA
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
NA
NA
Yes
Yes
Yes
Yes
No
No
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
No
Yes
Yes
Yes
HEATER POWER
INDICATOR
Yes
LED
LED
LED
HAND PORTS
TUBING PORTS
Flowmeter O2
cylinder; space for
spare cylinder
Yes
Yes
Yes
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Not specified
Yes
No
Yes
No
Yes
No
PHOTOTHERAPY
Optional
Optional
Optional
Optional
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
O2 INLET PORTS/
CONTROLLERS
NA
28
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
MEDAP
FAILED TO RESPOND *
T1500
MEDICOR
MEDICOR
MEDICOR
BLF-2001
ICB-145
ICB-145i
INTERNAL NOISE, dB
Not specified
50
60
60
220/240
230 **
230 **
230 **
BATTERY, TYPE
12-24 V
12 V
9V
12 V
DOUBLE WALL
Yes
Yes
No
Yes
30 (11.8)
31.5 (12.4)
35.6 (14)
31.5 (12.4)
90.5 x 80 x 50.5
(35.6 x 31.5 x 19.9)
130 x 85 x 66.5
(51.2 x 33.5 x 26.2)
132 x 105 x 51
(52 x 41.3 x 20.1)
132 x 105 x 51
(52 x 41.3 x 20.1)
60 x 30
(23.6 x 11.8)
58 x 33
(22.8 x 13)
58 x 33
(22.8 x 13)
58 x 33
(22.8 x 13)
Not specified
Caster, cm (in)
12 (4.7)
36 x 72
(14.2 x 28.3)
9.5 (3.7)
38 x 70
(15 x 27.6)
9.5 (3.7)
36 x 72
(14.2 x 28.3)
9.5 (3.7)
25 (55.1), incubator
only
63 (138.9)
60 (132.3)
63 (138.9)
Available
1 year
1 year
1 year
6-8 weeks
$7,000-7,500
45 days
$6,990
45 days
$3,980
45 days
$6,250
Not specified
~1,000
Not specified
Not specified
Not specified
January to December
Not specified
Not specified
January to December
Not specified
Not specified
January to December
Flat heater is
vulcanized on lower
side of patientsupporting tub; no
fan.
Built-in scale;
built-in x-ray bank;
air curtain; up to
90% humidity;
optional head hood,
servo-controlled
O2 unit, storage
cabinet, infusion
stand, apnea alarm.
Meets requirements
of ISO 9001.
Built-in scale; up
to 90% humidity;
optional head hood,
storage cabinet,
infusion stand,
apnea alarm. Meets
requirements of
ISO 9001.
Built-in scale;
built-in x-ray bank;
up to 90% humidity;
optional head hood,
storage cabinet,
infusion stand,
apnea alarm. Meets
requirements of ISO
9001.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
Warranty
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
29
MEDIPREMA
MEDIPREMA
MEDIPREMA
MEDIPREMA
NITE
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Worldwide
FDA CLEARANCE
Not specified
Not specified
Not specified
Not specified
CE MARK
Yes
Yes
Yes
Yes
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Transport
28-37 (0.1),
override to 39
35-37 (0.1),
override to 38
28-37 (0.1),
override to 39
35-37 (0.1),
override to 38
28-37 (0.1),
override to 39
35-37 (0.1),
override to 38
28-37 (0.1),
override to 39
35-37 (0.1),
override to 38
Digital
5-45
Digital
5-45
Digital
5-45
Digital
5-45
Digital
5-45
Digital
5-45
Digital
5-45
Digital
5-45
Yes
Yes
Yes
Yes
Yes
Yes
Yes (and system
failure)
Yes
Yes
Yes
Yes
Yes
Yes
Yes (and system
failure)
Yes
Yes
Yes
Yes
Yes
Yes
Yes (and system
failure)
Yes
Yes
Yes
Yes
NA
Yes
Yes (and low battery
and system failure)
Yes
Yes
Yes
Yes
LED or microprocessor
LED or microprocessor
LED or microprocessor
Horizontal bar or
microprocessor
HAND PORTS
TUBING PORTS
13
4 respiratory,
6 accessory
O2 INLET PORTS/
CONTROLLERS
Yes/not specified
Yes/not specified
Yes/optional
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
40/60/80%
Yes
40/60/80%
Yes
40 to 90% optional
No
NA
PHOTOTHERAPY
Optional
Optional
Optional
No
TEMPERATURE CONTROL
RANGE, C
Air (increments)
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
BACKUP THERMOSTAT
HEATER POWER
INDICATOR
30
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
MEDIPREMA
MEDIPREMA
MEDIPREMA
MEDIPREMA
NITE
INTERNAL NOISE, dB
<45
<45
<45
100-260;50/60/400 Hz
BATTERY, TYPE
NA
NA
NA
12-38 V
DOUBLE WALL
Optional
Optional
Optional
No
35 (13.8)
35 (13.8)
35 (13.8)
29 (11.4)
130 x 104 x 58
(51 x 41 x 22.8)
134 x 104 x 60
(53 x 41 x 24)
134 x 104 x 60
(53 x 41 x 24)
54 x 96 x 58
(21.3 x 37.8 x 22.8)
61 x 38.5
(24 x 15.2)
61 x 38.5
(24 x 15.2)
61 x 38.5
(24 x 15.2)
59 x 35
(23.2 x 13.8)
67 x 32
(26.4 x 12.6)
12.5 (5)
67 x 32
(26.4 x 12.6)
12.5 (5)
67 x 32
(26.4 x 12.6)
12.5 (5)
67 x 24
(26.4 x 9.4)
NA
WEIGHT, kg (lb)
80 (176.4)
85 (187)
87 (192)
36.5 (80)
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
8 weeks
Not specified
8 weeks
Not specified
8 weeks
Not specified
8 weeks
Not specified
1997
Not specified
Not specified
1997
Not specified
Not specified
1997
Not specified
Not specified
1995
Not specified
Not specified
Convection heated;
temp adjust control;
electronic
surveillance system;
alarm/communications
display on 68000
Series microprocessor; permanent
auto verification of
probes and control
devices; microprocessor control;
RS232 and RS485
serial links; 1-week
memory of defaults/
temp; x-ray facilities.
Convection heated;
temp adjust control;
electronic
surveillance system;
alarm/communications
display on 68000
Series microprocessor; permanent
auto verification of
probes and control
devices; microprocessor control;
RS232 and RS485
serial links; 1-week
memory of defaults/
temp; x-ray facilities; optional
scale and height.
Convection heated;
temp adjust control;
electronic
surveillance system;
alarm/communications
display on 68000
Series microprocessor; permanent
auto verification of
probes and control
devices; microprocessor control;
RS232 and RS485
serial links; 1-week
memory of defaults/
temp; x-ray facilities; optional
scale and height,
autoclavable water
tank; servo humid.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
31
MEDIX
MEDIX
MEDIX
MEDIX
PC-301 : PC-305
TR-1
TR-200
TR-306
WHERE MARKETED
Africa, Latin
America, Middle
East, Western
Europe
Africa, Latin
America, Middle
East, Western
Europe
Africa, Latin
America, Middle
East, Western
Europe
Africa, Latin
America, Middle
East, Western
Europe
FDA CLEARANCE
Yes
Not specified
Not specified
Not specified
CE MARK
Not specified
Not specified
Not specified
Not specified
MOBILE OR TRANSPORT
Mobile
Transport
Transport
Transport
TEMPERATURE CONTROL
RANGE, C
Air (increments)
20-39 (0.1)
20-38 (0.25)
20-39 (0.1)
20-39 (0.1)
NA : 34-38 (0.1)
NA
34-38 (0.1)
34-38 (0.1)
Digital
20-42
Thermometer
20-42
Digital
20-42
Digital
20-42
NA : Digital
NA : 20-42
NA
NA
No
NA
Digital
20-42
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
NA : Yes
NA : Yes
Yes
Yes
Yes
NA
NA
NA
NA
NA
NA
NA
Yes
Yes
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HAND PORTS
TUBING PORTS
1; 2 positions
(maximum 40% and
maximum 75%)/not
specified
1; 2 positions
(maximum 40% and
maximum 75%)/not
specified
1; 2 positions
(maximum 40% and
maximum 75%)/not
specified
1; 2 positions
(maximum 40% and
maximum 75%)/not
specified
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Optional
Yes
No
Yes
Not specified
Yes
Optional
PHOTOTHERAPY
Optional
No
No
No
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
HEATER POWER
INDICATOR
O2 INLET PORTS/
CONTROLLERS
32
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
MEDIX
MEDIX
MEDIX
MEDIX
PC-301 : PC-305
TR-1
TR-200
TR-306
INTERNAL NOISE, dB
<60
<60
<60
<60
110/220
110/220
110/220
110/220
BATTERY, TYPE
NA
12 V
12 V
12 V
DOUBLE WALL
Optional
Yes
Yes
Yes
34.9 (13.75)
25 (9.8)
30 (11.8)
25 (9.8)
132 x 94 x 53.3
(52 x 37 x 21)
110 x 54 x 130
(43.3 x 21.3 x 51.2)
115 x 54 x 130
(45.3 x 21.3 x 51.2)
132 x 94 x 53.3
(52 x 37 x 21)
63.5 x 35.1
(25 x 13.8)
56 x 25
(22 x 9.8)
56 x 25
(22 x 9.8)
63.5 x 35.1
(25 x 13.8)
26.7 x 83.8
(10.5 x 33)
10 (4)
23 x 64
(9 x 25.2)
Not specified
30 x 70
(11.8 x 27.6)
10.2 (4)
26.7 x 83.8
(10.5 x 33)
10.2 (4)
WEIGHT, kg (lb)
75.3 (166)
19-70 (41.9-154.4)
21-72 (46.3-158.8)
40-59 (90-130)
PURCHASE INFORMATION
Warranty
1 year
1 year
1 year
1 year
4 weeks
$6,200 : $6,500
6 weeks
$3,300-6,400
6 weeks
Not specified
8 weeks
Not specified
1988
450 : 1,300
November to October
1978
1,700
November to October
2000
120
November to October
1997
150
November to October
Meets requirements
of AAMI/ANSI II36,
IEC 601-2-19, and
ISO 9000.
Optional folding
cart, fixed cart,
battery and charger,
and oxygen cylinder/
regulator; ISO 9000.
Optional folding
cart and 2 cylinder/
regulators. Meets
requirements of IEC
601-2-20, ISO 9000.
Optional folding
cart and oxygen
cylinder/regulator.
Meets requirements
of IEC 601-2-20, ISO
9000.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
33
OHMEDA
OHMEDA
OHMEDA
OHMEDA
Care-Plus
Care-Plus Access
Giraffe OmniBed
WHERE MARKETED
Worldwide
Worldwide
Worldwide
USA
FDA CLEARANCE
Yes
Yes
Yes
Yes
CE MARK
Yes
Yes
Yes
No
MOBILE OR TRANSPORT
Mobile
Mobile
Mobile
Mobile
20-37 (0.1),
override to 39
35-37 (0.1),
optional 35-37.5
20-37 (0.1),
override to 39
35-37 (0.1),
optional 35-37.5
20-37 (0.1),
override to 39
35-37.5 (0.1)
28-37.5 (0.1),
optional 28-39
35-37.5 (0.1)
Digital
20-42
Digital
20-42
Not specified
Not specified
Digital
15-50
Digital
20-42
Digital
20-42
Digital
15-50
Digital
15-50
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
No
Yes
Yes
No
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HEATER POWER
INDICATOR
No *
None
HAND PORTS
4, gumdrop shaped
TUBING PORTS
Yes/separate O2
control system
Yes/separate O2
control system
Servo-controlled
option only;
available August
2001
Not specified
Yes (variable)
Optional servo
controlled
Yes (variable)
Optional servo
controlled
See footnote **
30-95%, increments
of 5%
Yes (variable)
Not specified
Optional
Optional
Optional
TEMPERATURE CONTROL
RANGE, C
Air (increments)
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
O2 INLET PORTS/
CONTROLLERS
SUPPLEMENTAL
HUMIDITY
Adjustable
PHOTOTHERAPY
34
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
OHMEDA
OHMEDA
OHMEDA
OHMEDA
Care-Plus
Care-Plus Access
Giraffe OmniBed
INTERNAL NOISE, dB
<60
<60
40 dBA
<60
100/120/220/240
100/120/220/240
115, 50/60 Hz *
100/120/220/240
BATTERY, TYPE
NA
NA
9 V Ni-MH
NA
DOUBLE WALL
Optional
Optional
Both walls
Yes
38.1 (15)
43.2 (17)
40 (15.7)
34.3 (13.5)
[147-174] x 69 x 112
([57.9-68.5] x 27.2
x 44.1)
34.3 x 64.8
(13.5 x 25.5)
34.3 x 64.8
(13.5 x 25.5)
48 x 66
(18.9 x 26)
38.1 x 62.2
(15 x 24.5)
25.4 x 76.2
(10 x 30)
12.7 (5)
30.5 x 76.2
(12 x 30)
12.7 (5)
27 x 78
(10.6 x 30.7)
12.7 (5)
33 x 76.2
(13 x 30)
12.7 (5)
83.9 (185)
83.9 (185)
129 (284.4)
99.8 (220)
3 years, parts;
1 year, labor
Not specified
$8,475-13,150
1 year
1 year
3 years, parts;
1 year, labor
Not specified
$8,475-13,150
Not specified
$10,125
1988
Not specified
Not specified
1996
Not specified
Not specified
90 days
$34,000, varies with
options
2000
300
Not specified
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
WEIGHT, kg (lb)
PURCHASE INFORMATION
Warranty
OTHER SPECIFICATIONS
Converts from
incubator to radiant
warmer; elevating
hood; closed bed
10 cm above center
of mattress.
1981
Not specified
Not specified
Optional rail,
shelf, radiant
warmer, and
BiliBlanket phototherapy light.
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
35
PARKER
FAILED TO RESPOND *
TC500-1 : TC500-2 :
TC500-3
PHOENIX MEDICAL
SEWARD
TSE
100 A
INCUBATOR 4000
INKA SI-302
WHERE MARKETED
Worldwide
Worldwide
Worldwide
Africa, Europe,
Middle East
FDA CLEARANCE
No
No
Not specified
Not specified
CE MARK
No
No
Yes
Yes
MOBILE OR TRANSPORT
Transport
Mobile
Mobile
Mobile
TEMPERATURE CONTROL
RANGE, C
Air (increments)
30-37 (0.1)
32-38 (0.1)
25-39 (0.1)
30-37.9 (0.1)
None
32-38 (0.1)
32-37 (0.1)
34-37.9 (0.1)
Digital
15-40
Not specified
15-45
Digital
0-99.9
NA
NA
Not specified
15-40
Digital
0-99.9
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
Yes
Yes
HEATER POWER
INDICATOR
Single LED
Not specified
On/off
HAND PORTS
2 or 4
4 or 6
TUBING PORTS
4 respiratory
2 respiratory,
2 accessory
O2 INLET PORTS/
CONTROLLERS
1/not specified
1 standard/monitor
optional
1/Oxycare
Control System
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Not specified
Yes
No
Monitoring only
No
Yes
Yes
PHOTOTHERAPY
Optional
Optional
No
Optional
Skin (increments)
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
36
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
PARKER
FAILED TO RESPOND *
TC500-1 : TC500-2 :
TC500-3
PHOENIX MEDICAL
SEWARD
TSE
100 A
INCUBATOR 4000
INKA SI-302
INTERNAL NOISE, dB
<520 A
60
<54
<52
110/220
230
BATTERY, TYPE
TC515 12 V pack
NA
NA
NA
DOUBLE WALL
Optional
No
Yes
Optional
23 (9)
31 (12.2)
33 (13)
33.5 (13.2)
44 x 43 x 83
(17.3 x 17 x 32.6)
135 x 96 x 55
(53 x 38 x 21)
105 x 71 x 125
(41.3 x 28 x 49.2)
[134-156] x 56 x 105
([52.8-61.4] x 22 x
41.3)
57.5 x 30.5
(22.6 x 12)
61 x 30.5
(24 x 12)
61 x 36
(24 x 14.2)
36 x 67
(14.2 x 26.4)
Not specified
Not specified
Caster, cm (in)
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Not specified
30.5 x 81.2
(12 x 32)
10 (4)
Not specified
36.2 x 84.5
(14.3 x 33.3)
Not specified
WEIGHT, kg (lb)
23 (50.7)
85 (187.4)
110 (242.6)
106 (233.7)
PURCHASE INFORMATION
Warranty
1 year
Not specified
3 years
1 year
6 weeks
$4,000-14,000
Not specified
$3,000
12 weeks
$14,500
6-8 weeks
Not specified
1993
Not specified
Not specified
Not specified
Not specified
Not specified
1997
Not specified
October to September
1996
Not specified
January to December
Crevice-free
interior and message
display alarm. Meets
requirements of
IEC 601-1.
None specified.
Oxygen analyzer;
20-40C
thermometer; oxygen
flowmeter;
bacteriological
filters; 2-position,
bidirectional bed
inclination;
x-ray cassette; 4
wheels with 2
brakes; relative
humidity gauge;
optional storage
case. CE mark. Meets
requirements of
CSN, EN, IEC and
ISO 9001.
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
37
TSE
TSE
INKA SI-303
TI-400 Transport
WHERE MARKETED
Africa, Europe,
Middle East
Africa, Europe,
Middle East
FDA CLEARANCE
Not specified
Not specified
CE MARK
Yes
Yes
MOBILE OR TRANSPORT
Mobile
Transport
TEMPERATURE CONTROL
RANGE, C
Air (increments)
30-37.9 (0.1)
30-38 (0.1)
Skin (increments)
30-37.9 (0.1)
NA
Graphic display
0-99.9
Digital
0-99.9
Graphic display
0-99.9
NA
NA
ALARMS
High air temp
Low air temp
High skin temp
Low skin temp
Fan failure
Sensor failure
Power failure
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
NA
NA
Yes
Yes
Yes
BACKUP THERMOSTAT
Yes
Yes
HEATER POWER
INDICATOR
On/off
HAND PORTS
TUBING PORTS
2 respiratory,
2 accessory
2 respiratory,
2 accessory
1/Oxycare
Control System
1/Oxycare
Control System
SUPPLEMENTAL
HUMIDITY
Adjustable
Yes
Yes
No
NA
PHOTOTHERAPY
Optional
Optional
TEMPERATURE DISPLAY
Air, display type
Range, C
Skin, display type
Range, C
O2 INLET PORTS/
CONTROLLERS
38
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
Incubators, Infant
TSE
TSE
INKA SI-303
TI-400 Transport
INTERNAL NOISE, dB
<52
<60
230
BATTERY, TYPE
NA
12 V
DOUBLE WALL
Yes
Optional
33.5 (13.2)
23.5 (9.3)
[134-156] x 56 x 105
([52.8-61.4] x 22.1
x 41.3)
136.5 x 56 x 86.5
(53.7 x 22.1 x
34.1) w/carriage *
36 x 67
(14.2 x 26.4)
30.9 x 56.3
(12.2 x 22.2)
36.2 x 84.5
(14.3 x 33.3)
Not specified
29.1 x 74.2
(11.5 x 29.2)
Not specified
WEIGHT, kg (lb)
108 (238.1)
58 (128); 78 (172)
with carriage
PURCHASE INFORMATION
Warranty
1 year
1 year
10-12 weeks
Not specified
6-8 weeks
Not specified
1997
Not specified
January to December
1995
Not specified
January to December
Microcomputer
controlled; temperature and oxygen
measurement and control; double cover;
elastic sleeve at
entry port; adjustable bed; pull-out
board for x-ray
cassettes; breath
and pulse control
and weight measuring
optional; IV pole;
shelves. CE mark.
Meets requirements
of CSN, EN, IEC and
ISO 9001.
Temp regulator;
20-40C thermometer; built-in O2
flowmeter; additional external O2
and air supplies;
O2 and air pressure
bottles; polymethacrylate cover; front
side can be lowered;
includes heat unit
fan, electronic control circuits, detachable internal
energy source; rechargeable. Meets
requirements of CSN,
EN, IEC and ISO
9001.
SIZE
Incubator w/stand,
H x W x D, cm (in)
Mattress, W x D,
cm (in)
Hood access door,
H x W, cm (in)
Caster, cm (in)
OTHER SPECIFICATIONS
2001 ECRI. Duplication of this page by any means for any purpose is prohibited.
39