Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
. A bubble at the bottom of the busket may dislodge itself after the basket starts to rotate (Figure 5-4B). Formation and disappearance of this bubble, if it occurs, should he observed by the analyst and docu- ‘mented in the testing, as it may change the dissolution rate. Dissolution Media Variables—Dissolved Gas Liguids are in equilibrium with surrounding gas at the gas-liquid interface. Ata given temperature and pressure, a portion of the gas 1s dissolved in the liquid. The amount of dissolved gas in equilibrium decreases substantially as temperature inereases, The equilibrium value of oxygen in water (measured in mg/L), for example, drops from 8.74 room temperature (22 *C) to 7.31 at 32 °C and 6.73 at 37 °C; 100% saturation at room temperature inereases to 120% at 32°C and 130% at 37 °C, the specified dissolution temperatures. ‘The excess from this supersaturation accumulates as minute bubbles in the med ‘This release of dissolved gas is one of the most annoying variables responsible for distorted dissolution data in all specified or proposed clissolution apparatus. Farly on, seientists at the FDA labs stated that a very appearance occasionally arising on the flask, rotating shaft, or basket or paddle is « warning of the release of dissolved gases that ‘may disturb the test (28). Since then there have been many publications dealing with dissolved gases, The silvery appearance is a result of m croscopic air or a gas bubble released as the medium adjusts to equ librium with the gas. ‘This phenomenon provides a warning 10 check deaeration methods and the influence of the released gas on the solution rate of the test under observation, ‘One may speculate about the many ways in which released affect dissolution, but most probably the small bubbles interfere with fluid dynamics or the area of the liguid-solid interface or both, The Conirolting Variables 87 bubbles may attach to the rotating busket or the sereen in the recipro cating cylinder, thereby altering the effective porosity. They may ac- ccumuilate on or in filers or the glass beads of the flow-theowsh appa- ratus, affecting the flow rate. They may accumulate in flow eells, where they could interfere with absorbance measurements, They may attach to aggregates from disintegrating dosage forms, changing thei eflective liquid-solid interface and flow patterns in Apparatus 1, 2, and 4, They may accumulate on the membranes in transdermal or percutaneous ab= sorption tests, The probable methods of interes ‘The effects of dissolved gases are significant. Studies with fiber optic probes (39) showed an inerease in variability (coefficient of variation) rates of 12% with Prednisone calibrators when the media Was not de- aerated, The presence of dissolved gases is suspected in many failures to meet calibration standards. The dissolution-rate data may be higher of lower, depending on the apparatus and dosage form, The analyst should recognize that no generalizations about the mag- nitude of the effect of dissolved gas in media are valid. The effects vary with the dosage form, and therefore, USP leaves decisions re- garding the presence of dissolved gases to the experimenter, stating only im Dissolution General Chapter <711>, that the presence oF bubbles ‘must not significantly affect the dissolution rate, Deaerating Media Release of dissolved gases can be prevented if the concentration of izases is kept befow the saturation value of the media during the test ‘To avoid problems, a value atleast $% below saturation atthe operating temperature should be obtained, A crude method in use at some labo ratories involves filing the flasks with media at 40 °C and operating the stirring device until the temperature reaches the prescribed 37 °C. This provides a 5% margin below saturation, In 1994, Qureshi and ¢o- workers described the various methods of deaeration (40), He studied the effect of deaeration by the following methods: heating alone, helium sparging, heating and and vacuum, sonication and clusion was that heating along with helium sparging was the best ap- proach. ‘This conclusion was confirmed by a more recent article (41), aand-many others have studied and compared dissolution deacration techniques (42, 43). EDA published in 1978 (28) a satisfactory method that involves sub- ily deaerating the media before its addition to the dissolution88 HANDBOOK OF DISSOLUTION TESTING Vacium Deseratod Media| Masi L } Source Figure 5-5 Deaevation system developed by NCDA. The medium is drawn through a fine spray nozcle By an 18 to 22 inch vacuum into a Jive-galton jug (protect personne! from implosion injury). Gas is re leased from the fine spray and the medium container can be sealed, ajier which it remains satisfactory for dissolution use for 2-3 days apparatus. A later version of this method was published by Moore (44), This method is shown in Figure 5-5, A commercial system, called the Media-Mate™, that uses a variation of the NCDA method, has been available for more than 20 years. Thi system is described in Chapter 8. It combines the spray with thin film to increase the surface area under vacuum. Langenbucher and co-work- cers (33) have reported a similar thin-film technique. Their system achieves a 30% reduction from saturated values (25% with house vac= tuum). The Media-Mate® regularly achieves greater than 10-15% re= duction when tested with a standard dissolved oxygen meter. By com- parison, checks in Bill Hanson's laboratory have achieved a practical 55% reduction through conventional boiling techniques. Subjecting 4 liguid to a vacuum wall result in vaporization as well as the release of dissolved gas. This cools the solution, Tests indicate 1 drop of approximately 1.5 °C when the thin-film method is used. When such a solution is used to directly fill dissolution-fluid systems, the source should be at Teast 1.5 °C higher in temperature than the system ‘There are other automated methods of deaeration, including an in situ deaeration method using the hollow shaft system (45). Diebold and Dressman published a study (46) deseribing in-depth Controtting Variables 89 the use of oxygen to measure de- and re-seration of aqueous media. ‘They describe an oxygen sensor bused on a Clark electrode, They con- cluded that the oxygen measurement method Was reproducible, that the efficiency of deaeration was highly dependent on the method used, and that re-introduction of air into the medium occurs not only during filling of the dissolution vessels but also during the dissolution test. Other studies have suggested that a level of dissolved oxygen with a range of 4-7 me/L produces an equilibrium effect (47). Influence of Released Gases ‘The major influence of gas or air in media seems to be physi js, the bubbles that appear in media may disturb the equi following ways: + Alter low patterns as they tise to the surface. * Associate with aggregate particles, resulting in random eoncentra- tions of the particles in the solvent stream, + Collect atthe sereen on the basket, changing the effective porosity of the mesh, + Collect on the sides of the vessel walls, providing a place for par- ticles to eling, rather than to mix properly + Attach to dosage forms before disintegration, thus altering the dis integration and de-ageregation process by reducing the surface area ‘exposed 0 the solvent siream and/or altering the specific gravity of the mass, resulting in a random, uncontrolled positioning of the ass in the solvent stream. + Contribute to the boundary layer at the solid-liquid interface in a random manner Media Variables—pH ‘Media composition is specified in the individual monographs. Un- butfered media may vary in pH. Typical values for pH in laboratories suggest ustal levels of pH 6.0 for distilled water, pH 6.6 for de-ionized Water (deaerated or not), and pH 7.2 for distilled water deserated by boiling. Some have recommended against water as a media (48), how- ever, others find using water can be practical and ean yield useful results (49). Variations in dissolution rate ut different pH levels is to be ex- pected if the drug has a steep pH/solubility curve, but the plUsolubility curve of various excipients should not be overlooked. tn the method development stage, the test solution pH should be checked at the end90 HANDBOOK OF DISSOLUTION TESTING of the run to see if there has been a pH change. Shorteni ening the disintegration and de-aggregation lag periods (C ures 2-2, 2-3, and 2-4) can exert significant effects on the dissolution rate, even for drugs with relatively flat pHisolubility characteristis, Absorbance may vary: with changes in pH. If the pH is changed uring the (st, for example, when studying delayed release dosage forms, the standard should be checked at each pH used, If pH. varies significantly, analysts should use multiple standards Media Variables—Volume Iris elementary, of course, that the volume of medium must b tained constant. Volume lost from sampling may be corrected in cal- culations, provided the correction factor is less than 25%, In dissolution testing of extended-release preparations, the volume of samples with= drawn may exceed this limitation, Sample volumes may therefore be replaced (at the same temperature); automated equipment is available to accomplish this task (Chapter 8). The amount of Tiguid lost by evaporation ean be considerable and should be checked. In some labs, the vessel and contents are weighed before and alter the run to determine just the evaporation Loss. In low- humidity environments, up to 15-mL losses from standard dissolution flasks have been measured, Environmental conditions should not be overlooked. There may be a vast difference in evaporation rate in a laboratory in Mexico City or Denver compared to one in Puerto Rico ‘or Thailand, Evaporation control should be included in all dissolution apparatus, Suitable curtain-type covers are supplied in the reciproe; and holder equipment, Tests i sampling equipment should not be considered in the absence of evap- ‘oration control on the sample receptacle unless automated samples are used within minutes of sample leposit. HPLC collectors commonly use septum closures that can be punctured for injection, Test tubes are avail- able with split-plastic throwaway evaporation covers that enable fl tnd emptying with sample probes, ‘The vessel covers specified for Apparatus | and 2 have been some- \what primitive, They do retard evaporation, but they also act as con- densers, accumulating considerable media, The covers could easily in- troduce errors in the 24% range. ‘A common practice is to fill dissolution vessels with accuracy and wait for them to. equilibrate at 37 °C. Considerable evaporation can inder icate they are 99% effective, Automated Controlling Variables a1 ‘occur unless covers are let on du ng this warm-up period, Automated automated equipment avoids this waiting period by loading the flask with prehented deaerated medium, Media Variables—Temperature Solubility is generally linear with temperature, and sometimes the curve is sleep. The effects of temperature variations will, of course, depend upon the temperature/solubility curves of the active ingredient, 1s well as those of the binders and excipients, Dosage forms may differ widely, ranging as high as 5% change in dissolution rate per degree Celsius (12). Because the compendia allow * 0.5 °C tolerance, a con- siderable variation within this range might be anticipated for some dos- age forms and this should be part of the robustness evaluation of the dissolution method ‘The temperature should be monitored at suitable intervals, at least before the test and if possible after the test. This can easily be done by automated systems, and recording deviees are also available, $0 this tay be a routine procedure, However, bath temperature does not ensure Mask content temperature in all situations, The differences might be subtle, but sometimes they are the cause of the puzzling inability (© meet calibration standards, For example, plastie flasks have a heat transfer coefficient that is ap- proximately 30% that of glass (50), In some environments it has been ‘ound impossible (© hold a plastie flask temperature at 37 °C without implementing substantially higher (38-39 °C) bath temperanures. The cooling from the surface of the medium in the flask ean exeeed the rate of heat transfer through the plastic (0 the extent that stability cannot be maintained. Tt is therefore of vital importance that covers always be used, especially with plastic flasks. When covers are used, there is little substantial differential between the temperature of the contents of glass flasks and the bath temperature, ‘The critical apparatus parameter is therefore a consistent temperature at all paris of the bath, This can be easily held within = 0.4 °C in properly designed and installed bath and should be a part of every manufacturer's warranty. Doing research on automatic sampling for percutaneous absorption systems, Bill Hanson was surprised ( find that there is a disturbing itference in flow rate vs. temperature, In small-diameter PTFE tubing, this can exceed 1%7°C. This suggests that care should be taken to useoy HANDBOOK OF DISSOLUTION TESTING only volumetric pumps in the flow-through apparatus and in sampling systems that depend on flow rate for volume measurement Media Variables—Sink Conditions When a low-solubitity drug is specified ut a dosage level that ca saturation in the dissolution medium, accurate dissolution profiles be- ‘come difficult or impossible (@ obtain, In dissolution testing, the rule of thumb has been that sink conditions fre approximated if the test volume is at least 3 times the saturation volume (Chapter 2), USP, in the General Chapter In Vitro and In Vivo Evaluation <1088>, gives the following description of solubility pa rameters when developing a method, “The quantity of mediam used should be not less than 3 times that required to form a saturated solution of the drug substance"(S1). The flow-through cell apparatus provides an infinite or variable sink and was and may still be the method of preference in Europe for low-solubility drugs (32). An alternative ap- paratus for low-solubility drug testing specifies an inerease in the me- «dium volume in Apparatus 1 or 2: a 4-1 flask has been used success- fully in Europe and North America (52). Spectal modifications of Ap- pparatus 1 and 2 for 2- and 4-L flasks are commercially avallable and ‘are allowed in the USP Dissolution General Chapter <711>, Finally, several suggestions have been put forth for media modifi cations in order to increase the solubility of specific dosage forms. The most common uses a surfactant, sodium Iauryl sulfate, in small amounts. Some have suggested that this is an acceptable alternative to media prepared with ionie bile salts that ane a part of the assimilation mechanism of the human (53-55). This proposal, therefore, is in har- ‘mony with the dissolution movement toward better in viteo in vivo correlations ‘The alternate situation, low concentrations of active ingredient but no sink condition issues, is 1 problem with some dosage forms, partie- lularly transdermal, high potency and extended-release dosage forms. ‘The test may require a greatly reduced media volume in order to obtain detectable concentrations; adaptations of a miniaturized basket have been successful (56). Modifications of Apparatus | and 2 have been made for 100- to 200-mL. beakers (57). Smaller volume media tubes may also be used, Flow Pattern and Vessel Hydrodynamics Uncontrolled variables inflow pattern have been extensively studied with Apparatus | and 2, but much investigation needs to be completed Controlling Variables 93 for Apparatus 3-7. Suggestions about unresolved problems were made in the first paragraphs of this chapter Asa result of Cox's studies at NCDA (28) the contours of dissolution flasks used in Apparatus 1 and 2 have been more rigorously defined, ‘The current commercial flasks are far more consistent than those avail- able 10 years ago—and the price has risen accordingly, however prob- lems still occur (58), Cox and co-workers also cautioned against the permanent intrusion Of sampling probes and thermometers into dissolution flasks because of the probes’ effects on dissolution patterns. Savage and Wells in [982 (59) found significant differences in dissolution rate when 7-ram di- ameter sample probes were used, but they found litle difference with 1.5-mm diameter probes. Automated systems are particularly vuln ble to this defect; some use two probes per vessel. Now, some manu- facturers have reconfigured equipment so that probes are removed ex- cept when sampling is actually taking place. NCDA scientists also commented on the return (replacement) of sub- stantial amounts of medium to the flask and the attendant effects that such returns have in disturbing flow patterns, particularly if the returned medium is at a different temperature, ‘The position of the stirring device from the bottom of the flask is well defined in the compendia, although available evidence suggests that this position is not a significant variable if itis held within rea- sonable Timits (12). This variable needs to be more precisely defined in the transdermal paddle over disk, Apparatus 5, Consistent flow pattems in Apparatus 1 and 2 make possible a reli- able sampling point. Gray and Foster in 1997 (60) published research performed at the Drug Research and Testing laboratories using USP Salicylic Acid calibrators anv! found consisteney at « point about one half-way trom the top of the basket or paddle to the top of the media aand not closer than | em to the side of the flask. This requirement has been in effect for many years, and no data refuting it have appeared, I causes some problems with automated equipment, however, when the media are changed during the test from 750 t© 1000 mL. as proposed for delayed release dosage forms. ‘There have been new studies performed on the hydrodynamics of the vessel flow by Mauger, Healy, and Muzzio (61-63), This is an area of great interest and more research is expected, Perhaps the least documented influence of flow-pattern vatiables ex- ists with respect to Apparatus 3 (reciprocating cylinder), and Apparatus94 HANDBOOK OF DISSOLUTION TESTING 4 (flow-through cell). Apparatus 7 (ceciprocating holler) and pereutae neous absorption cells are discussed in Chapter 4, ‘The low pattern of Apparatus 3 varies with reciprocating rate, the porosity of the top and bottom sereens and holes, and the plunger cou- pling (reflected by the respective diameters of the eylinder and vessel). ‘The elegant possibilities of this method for closely controlling the shear rate of particles (particularly beads) depend on rigorous definitions of these variables, Similar definitions are required for consistency with Apparatus 7. Published suggestions on the flow-through cell (Apparatus 4) vary from a smooth laminar flow to a sinusoidal pulse of defined frequency. It scems that this variable is critical; perhaps it should be closely de~ fined in the monograph for each item. More than any other technique, percutaneous absorption systems clearly depend on predictable diffusion-layer characteristies at the membrane. Without close control, consistent results in collaborative studies will not be possible. Current methods of maintaining a constant flow pattern in the receptor chamber of these cells have been improved lover the last decade by use of more precise speed control devices (such ‘ mixer for more homogeneous as Variomag), and use of a “helix” mixing. Sorption [Experience has shown that active ingredients may sorb excessively on certain materials, particularly some plastics, used in dissolution ap- paratus. This should be considered when one is attempting to validate amy new equipment or procedure, particulary automated systems Cox et al, 28) diseussed the bias from filters, A technique for s trating adsorptive sites on filters was studied by Hill and Snider in 1987 (64, Some filter bias ean be reduced by a preflush oF presatu- ration step—this is usually done by drawing enough sample through the filer to suurate the absorptive sites before collecting the sample ‘This way no drug is lost and the measurement of the sample should be accurate, Fillers should be selected with adsorptive characteristics in sind Potcatial sorption problems arise from tubing used in automated equipment. Cox and Wells (65) resolved tubing sorption problems with digoxin tablets by immediately introducing methyl aleahol inte the sample steam asthe aliquot was withdrawn, There have been aneedotal reports of excessive sorption bias with nitroglycerin and diazepam (up Conirolting Variables 95 to 10%%/foot forthe latter) when flexible laboratory tubing was used on ‘automated sampling devices such as perisaltic pumps. Halstead and ‘Theis (66) have reported similar studies on prostagkandin derivatives ‘The safest procedure isto limit materials of construction for automated systems to poly!luorocarbon, glass. and type 316 stainless steel Checklist for Variables and Following GMP. [No quantity of text on variables can substitute for good laboratory technique! Too many times the authors have helped solve dissolution problems for which the real causes were lack of GMP awareness and poor training. Situations such as failure to calibrate lab thermometers: dirty flow cells; use of theoretical values instead of tue standards; using Random Input Variables Checklist Figiere 5-6 Random input variables cheeklist.96 HANDBOOK OF DISSOLUTION TES’ TING six separate standards instead of six aliquots from one; failure to eali- brate instruments; careless sampling points; inadequate documentation, ‘and many other unacceptable practices have been observed. Dissolution isa precise wet-chemistry technique, and GMP must be observed. ‘A review of some of the random variables that atfeet dissolution rate hhas been presented in this chapter. Some of these are listed in the following checklist (Figure 5-6) with maximum allowable intensity that should be tolerated—along with expected changes in dissolution rates that have been reported in detailed studies. Some practical hin about controlling these variables have also been included for ready reference. GMP training is a Key t0 good laboratory results that are without random or human caused errors. There are many courses availabe from 1 number of organizations; this book is a good reference for any dis- solution training program. The key isto have a formal training program to track outside training courses and in-house training that has ocurred. This is a vital wecord for FDA inspections: as training records are & critical part of a quality system. Internal lab audits are a very good ‘way encourage analyst to keep GMP habits up to date. Sometime the analysts themselves will perform internal audits as a part of perfor- mance objectives (18). Controlling Variables 97 References (1) Skoug, JW, “Calibration of Dissolution Rate Apparatuses: User's Perspective”, 1994, Dissolution Technologies, \(2), 3-5 (2) Grady, LT, “Dissolution Calibrators Used by the United States Pharmacopeia’’, 1994, Pharmacopeial Forum, 20(6), 8567-8570. (3) McCormick, Tl, “Industry Perspective on Dissolution Apparatus Calibration”, 1995, Dissolution Technologies, 24), 12-15. Martin, GP, Reed, DG, Magiso, LE, Griffith, MF and Ip, D, “Tu- torial on Dissolution Calibration: An Industrial Perspective”, 1996, Dissolution Technologies, 3(1), 3-6. @ (5) PARMA, Subcommittee on Dissolution Calibration: Brune. S, Bucko, J, Emr, S, Gray, V, Hippeli, K, Kentrup, A, Whiteman, D, Loranger, M and Oates, M, “Dissolution Calibrator: Recommen- dations For Reduced Chemical Testing and Enhanced Mechanical Calibration”, 2000, Pharmacopeial Forum, 264), 1149-1166. (6) Mirza, T, Grady, LT and Foster, TS, “Merits of Dissolution Sys- lability Testing: Response to PARMA’s Propasal on Me. Calibration”, 2000, Pharmacopeial For, 26(4). 1167~ Brown, W, “United States Pharmacopeia Establishes Projeet Team fon Dissolution Calibration”, 2002, Dissolution Technologies, 92). 10. (8) Moore, TW, Shangrass, RF and Habib, Y, “Dissolution Calibrator Tablets: A Recommendation for New Calibrator Tablets to Re- place Both Current USP Calibrator Tablets”, 1996, Phurmaco- peial Forum, 223), 2423-2428, a (9) Gray, VA, Hubert, BB and Krasowski, JA, “Calibration of Dis- solution Apparatus 1 and 2—What To Do When Your Equipment Fails”, 1994, Pharmacopeial Forum, 20(6), 8571-8573. Identifying Sources of Error in Calibration and Sam- 2002, American Pharmaceutical Review, 5(2), 8-2. (10) Gray, VA ple Testing”, (11) Hanson, W, “Solving the Puzzle of Random Variables in Disso- 30-41. lution Testing”. 1977, Pharmaceutical Teclnology. 1 ¢ (12) Hanson, W, “Studies on the Effects of Some Variables Commonly98 a3) ay (1s) 16) an as) 09 (20) 23) HANDBOOK OF DISSOLUTION TESTING Encountered in the Measurement of Dissolution Rate of Solids”, 1979, Ph.D. Dissertation, California Western University. Gray. ¥, Beggy, M, Brockson, R, Corrigan N and Mullen J, “A Comparison of Dissolution Results using O-ring versus Clipped Basket Shatts", 2001, Dissolution Tecluologies, 84), 8-11 Beyer, W and Smith, D, “Unexpected Variable in the USP/NF Rolating Basket Dissolution Rate Test”, 1971, Journal of Phar- maceutical Sciences, 60, 2350-2351. Hanson, W, “Effect of Vibration on Dissolution Rates”, 1975, Presented at the Becknan Conference an Dissolution, Mountain- side, NI Cartwright, A, “Sources of Variation during Collaborative Eval- uation of In Vitro Dissolution ‘Tests for Two Solid Preparations” 1979, Journal of Pharmacewtical Pharmacology, 31, 434440, Collins, CC, “Vibration: What is it and How Might It Effect Dis solution Testing”, 1998, Dissolution Technologies, 5(4), 16-18. Gray. V and Miller. B. in the Dissolution Laboratory 313), 19-21 ‘Thakker, D, Naik, N, Gray, V and Sun S, “Fine Tuning of Di solution Apparatus”, 1980, Pharmacopetal Foren, 6(2), 177— 18s. Borst I, Ugwu, § and Beckett, AH, “New and Extended Appli cations for USP Drug Release Apparatus 3°, 1997, Dissolution Technologies, 41), 11-18. Rohrs, BR, Burch-Clark, DL, Witt, MJ and Stelzer DJ, “USP Dissolution Apparatus 3 (Reciprocating Cylinder): Instrument Pa- rameter Effects on Drug Release from Sustained Release For- 1995, Journal of Pharmaceutical Sciences, 84(8), Good Manufacturing Pr + 2002, Pharmaceutical Canada, mutations”, 922-926, Yang, LI, Ferguson, SM, Hudson, TJ, Watanabe, S, Katsume, M and Fix, JA, “In Vitro Evaluation of Dissolution Behavior for a Colon-Specifie Drug Delivery system (CODES™) in Multi-pH Media using United States Pharmacopeia Apparatus I and II", 2003, AAPS PharmSciTech, 3(4), Yu, LX, Wang, JT and Hussain, AS, “Evaluation of USP Appa- ey) 25) eo) en 28) ey Go) oD a2 (33) G4) Controlling Variables 99 ratus 3 for Dissolution Testi 2002, AAPS PharmSci, 4(1) of Immediate-Release Products”, Nicklasson, M- and Langenbucher, F, “Description of the Flow Cell Dissolution Apparatus as an Alternative Test Method for Drug Release", 1990, Pharmacopetal Forum, 18(3), 532-37. Looney, TJ, “USP Apparatus 4 (Flow Through Method) Primer” 1996, Dissolution Technologies, (4), 10-12. Nicolaides. E, Hempenstall, JM and Reppas, C, “Biorelev solution Tests with the Flow-Through Apparatus’, 2000, Disso- ution Technologies, 1). 8-U Biclen, N, “Performance of USP Calibrator Tablets. in. Flow- ‘Through Cell Apparatus”, 2002, fternational Journal of Phar Cox, D, Douglas, C, Furman, W, Kirchhoofer, R, Myrick, J and Wells, C. “Guidelines for Dissolution Testing”, 1978, Pharma: ceutical Technology, 2(8), 40-53. Carstensen, J, Lal, T and Prasad, ¥, “USP Dissolution IV, Com- parison of Methods", 1978, Journal uf Pharmaceutical Sciences, 67, 1303-1307. Noyes, A and Whitney, W, “The Rate of Solution of Solid Sub- stances in Their Own Solutions”, 1897, Jounal of the American Chemistry Society, 19, 930. Langenbucher, F. “In Vitro Assessment of Dissolution Kinetics: Description and Evaluation of a Column-Type Method”, 1969, Journal of Pharmacewicat Sciences, 58, 1265-1272. Langenbucher, “Use of Flow-Through Apparatus for Dissolu- 1980, Presented at Journe’s D’Actualites Biopharmaceutiques de Clermont-Ferrand, France Universite’, de Clermont-Ferrand, Langenbucher, F, Benz, D, Kirth, W, Moller, Hand Ov, M, “Standardized Flow-Cell Method as an Alternative to Existing Pharmacopeial Dissolution Testing”, 1989, Pharmaceutical In dusiry, 311), 1276-1281 Adamson, A., 1979, Physical Chemistry of Surfaces, Wiley science, New York,100 HANDBOOK OF DISSOLUTION TESTING (35) Hanson, W., 1968, U. S. Parent # 3.572.648. (36) “Cefaclor Extended-Release Tablets", 2004, United States Phar- macopeia and National Formulary, United States Pharmacopetal Convention, Inc., 27th Edition, 351 37) “Dirithromycin Delayed-Release Tablets", 2004, United States Pharmacopeia and National Formulary, United States Pharma copeial Convention, Inc., 27h Edition, 646-647. (38) Sarapa, A and Clark, L, “Elimination of Air Entrapment Using USP I Rotating Basket Dissolution Apparatus”, 1980, Jounal of Pharmaceutical Sciences, 69, 129. (39) Wunderlich, M, Way, T and Dressman, J, “Practical Consider- ations When Using Fiber Optics of Dissolution Testing”, 2003, Dissolution Technologies, 1064), 17-19. (40) Qureshi, SA and MeGilveray, 1J, “Impact of Different Deaeration ‘Methods on the USP Dissolution Apparatus Suitability Test Cri- 1994, Phurmacopeial Forwn, 26), 8565-8566. teria’ (41) Degenbarat, OS, Waters, B, Rebelo-Cameitao, A. Meyer, A, Brun- ner, H and Tolt, NP, “Comparison of the Effectiveness of Various Deaeration Techniques”, 2004, Dissolution Technologies, 11(1), IL (42) Griffith, ME. Curley, TE and Martin, GP, “Considerations in Choosing « Deaeration Technique for Dissolution Media, 1997, Dissolution Technologies, (1), 16-17. (45) Robrs, BR and Stelzer, DJ, “Deseration Techniques for Dissolu- tion Media, 1995, Dissolution Technologies, 202), 1 and 7, 8. (44) Moore, TW, “Dissolution Testing: A Fast, Efficient Procedure for Degassing Dissolution Medium”, 1996, Dissolution Technologies, 342), 345 Rolli, R, Fiechter, A and Hengst, R, “Jn-Situ Deaeration of Dis- solution Media through a Hollow Shaft® System”, 2000, Disso- ution Technotogies, 1(2), 20-2 (46) Diebold, SM and Dressman, JB, “Dissolved Oxygen as a Measure for De- and Reaeration of Aqueous Media for Dissolution Test- ing”, 1998, Dissolution Technologies, 5(3), 13-16. (47) Curley, T, Forsyth, R, Sun S, Fliszar K, Colletio, M and Martin, (as) (49) (30) Gb 62) 63) G4) (55) (56) (7) Comirolling Variables 101 GP, “Measurement of Dissolved Oxygen as a Determination of n During Dissolution Testing”, 2004, Dissol dion Technologies, 114), 6-11. Noory, C, Tran, N, Ouderkitk, L, Brown, , Perry, J, Lopez, Co- Jon, M, Faberile, M, Henry, K, Rorberg, J, li, SN and Shath, V, ‘Rethinking the Use of Water as a Dissolution Medium”, 1999, Dissolution Technologies, (4), 6-7. Leeson, LJ, “Some Observations on "Rethinking the Use of Water as Dissolution Medium”, 2000, Dissolution Technologies, 72), 16-17, Blanco, Personal Communication to Bill Hanson in 1980. “USP General Chapter on In Vitro and In Vivo Evaluation of Dosage Forms <1088>", 2004, United States Pharmacopeia and Nationat Formulary, United States Pharmacopeial Convention, Edition, 2513-2518, Ine Rothe, W and Schelthorn, Rate for the European Pharmacopei many, 22, 74-80. Shah, VP, Noory, A, Noory, C, McCullough, B, Clarke, S, Ev- crett, R, Navaisky, H, Srinivasan, BN, Fortman, D and Skelly, JP, “In Vitro Dissolution of Sparingly Water-soluble Drug Dosage Forms”, 1995, feternational Journal of Pharmaceutics, 125, 99 106. Noory, C, Tran, N, Ouderkirk, L and Shah, V, "Steps for Devel- ‘opment of a Dissolution Test for Sparingly Water-soluble Drug Products”, 2000, Dissolution Technologies, 7(1), 16-18, 21 “Determination of the Dissolution 1979, Drugs Made Ger. Robs, BR, “Dissolution Method Development for Poorly Soluble Compounds”, 2001, Dissolution Technologies, 8(3), 6-12. Abdou, HM, Ast, TM and Chioffi, Fl, “Chromatographic Deter mination of Dissolution Rate of Fluorocortisone Avetate Tablets”, 1978, Journal of Pharmaceutical Sciences, 67, 1397-1398, rail, DJ, Tunis, A and Danascreau, R, “Is the Use of a 200-mL. Vessel Suitable for Dissolution of Low Dose Drug Products?” 2004, dnternational Journal of Pharmaceutics, 269, 203-208. Scott, PR “Geometric Irregularities Common to the Dissolution Vessel”, 2005, Dissolution Technologies, 12().12 HANDBOOK OF DISSOLUTION TESTING (59) Savage, TS and Wells, CE, “Automated Sampling of In Vitro Dissolution Mediuny; Effect of Sampling Probes on Dissolution Rate of Prednisone Tablets”, 1982, Journal of Pharmaceutical Seiences, 71, 670-673. (60) Gray, VA and Foster, TS, “Utility of USP Salicylic Acid Cali- brator Tablets”, 1997, Pharmacopeial Forum, 23(6), 5360-5363, (61) Mauger, J, Ballard, J, Brockson, R, De, S, Gray, V and Robinson, D, “Intrinsic Dissolution Performance ‘Testing of the USP Dis- solution Apparatus 2 (Rotating Paddle) Using Modified Salicylic ‘Acid Calibrator Tablets: Proof of Principle”, 2003, Dissolution Technologies, 10(3), 6-15. (62) Healy, AM, McCarthy, G, Callagher, KM and Corrigan, O1, “Sen- sitivity of Dissolution Rate to Location in the Paddle Dissolution Apparatus”, 2002, Journal of Pharmacy and Pharmacology, 54 441444, (63) Kukura, J, Arratia, PC, Szalai “Understanding, Pharmaceu Technology, 26(10), 48-72. (64) Hill, RA and Snider, BG, “A Widely Applicable Automated Sam- pling Apparatus for Dissolution Testing”, 1987, tntermarional Jounal of Pharmaceutics, 36, 175-183, Bittorl, KJ and Muzzio, FI, sal Flows”, 2003, Pharmaceutical (65) Cox, D and Wells, C, “In Vitro Dissolution of Digitoxin Tablets 1978, Internal Documeat supplied to Bill Hanson in correspon- dence with Furman, WC, National Center for Drug. Analysi FDA, Si. Louis, MO. (66) Halstead, GW and Theis, GL, 1985, Jounal of Pharmaceutical Sciences, 74(10), 1086-1090. 6 Setting Up for Dissolution Testing Beto pein sitions says sta dlp imate cal cater tea asco espn Since the Sond aon ots handbook was pbs ta an fica USP Appr te on te ginal ts and pal {USE Appenes 1 a) lve on ve fer wig. Thc ce tha aos ss she sec eo. A cup inspect bs nd palit wil oe eaten Otc Ft! oe et de Saeictekbe ni wundzmal pepe el wih Oe te came wih yup th oy eg pas nd SF nein vse! mens ce an sony Treg te Cet rx Dolton Poss! pics a isco ain eau eae ir sonar pram fo ny Oe Ut Appr Ud forms such as 103104 HANDBOOK OF DISSOLUTION TESTING Checklist for the Dissolution Protocol First, record and check the last calibration date and whether the ap- pparatus has been calibrated for either basket or paddle. 1. General Designation of Method A. Apparatus 1, 2, or other B. RPM, flow rate, or stroke/dip rate C. Modifcations—basket-mesh size, paddle coating, sinker type. vessel size, sercen types, ete D. Sampling intervals) E, Dissolved Q specification: (%) or interimvin-house value. FE. Medium composition (w/w or w/¥) and amount in mL, G. Modifications of apparatus to be used (2-liter vessels, peak ves: sel, mini paddles and vessels, flow cell type, holders, sinkers, tc). H, Sampling method, including medium replacement automated. 1. Sample analysis protocol manual or Inspection of Equipment A. Check shalt straightness (2 0,005-inch total indicator reading (TIRD. B, Examine paddles for peeling or eracks in the coati . Check paddles for compendial dimensions, particularly blade tip dimension from axis. D. Mount paddles or baskets, anc! check eccentricity. Check speed control—muist be constant and without surges. E. Check vibration levels at vessels—if possible, with a vibration meter + Eliminate outside sources of vibration to keep readings under 1.05 mils displacement. If using baskets mount them on shalts and hand -adjust + Look for holes in sereens or tangs of wire extending into the basket at the seams, + Check that basket clips (USP refers to these as “tangs” are seated firmly in place + Verity chat basket is not malformed and screen is not frayed. + Keep eccentricity within 0.010-inch TR H. Inspect paddles or baskets and all portions of the apparatus in contact with solutions for cleanliness. s Setting Up for Dissolution Testing 103 + Look particularly for erevices or lines in the paddle and along the basket welds for adh + Look for residue build-up on se shaft joint 12 contaminants, 1. Ensure that the water bath is clean and transparent, so the dis- integration and deaggregation process can be visually monitored. The presence of air bubbles or air films can be detected only through a transparent bath, J, Determine when the system was last calibrated with USP cali Drators and if itis still within the calibration interval. 3. Dissolution Bath A. Adjust bath temperature to mainvain 37 °C in vessels—use the appropriate bath temperature for glass oF plastic vessels, as plas- tie vessels take longer to heat B. Insert shafis in the drive, and adjust each vessel with a entering ‘gauge if necessary, . Inspect vessels for abnormalities in dimensions and excessive bumps on the internal radius of the bottom of the vessel . Ensure that the bath fluid level is above the top level of the ‘medium in the vessel 4. Proparation of Medium A, Preheat medium to 37 °C or slightly higher—use NIST-raceable calibrated thermometers. B. Deuerate medium by a suitable method at 37 °C unl the testis performed, C. Check medium pH to two decimal places with a properly cali- brated pH meter: 1d equilibrate and keep 5. Select and Check Analytical Methods ‘A. Check for adsorption or interference because of filtering. B, Cheek for adsorption on tubing (or leaching by plastic tubing) used in automated equipment . Check for interference with sampling probes, . Include information on the standard, is preparation, and placebo interference in the calculations; check to see that it has the same pH as the test medium; and check the effect of water-miscible solvents. E, [absorption with spectrophotomettie methods is used, check to see if the data point at the tip of the curve has a minimum standard deviation (SD) from normal or if the SD is excessive ‘on the slope,106 HANDBOOK OF DISSOLUTION TESTING 6. Getting Ready for the Test A. Insert baskets or paddles —adjust to proper distance from bottom of yessel (2.5 em) when unit is in its operating position. B, Check to see if the water bath is level (OK if system has been brated! within a specified reiable time). C. Check the tilt and eccentricity ofall shafts (visual checks OK if system bs been calibrated within a specified reliable time). Recheck centering of vessels with a gauge if necessary. Set vertical frame limits so the unit ean be dropped atthe proper c, so that all baskets or paddles remain at the proper test hheight—or use markers or eollars on the top of the baskets so that each can be individually deopped to the proper depth at the start af the test F. Add the deaerated medium to the proper level and check the temperature of each vessel—check the average error of measur ing volume © within * 1%. G. Inser the dosage units into the baskets (be sure the baskets are dry) if baskets will be used, or have dosage units ready to drop into the vessels if paddles will be used. H. Chock to see thar the speed is stable to within * 4g without ‘automated sampling-automated detection is being used, adjust the specttophotometer or detector system 0 50% or 100% stan- dard and zero the base line, then calculate carry-over (if present) by running two solvents followed immediately by (wo standards, followed azain by two solvents; the frst pass in each ease will present information about the percentage of carry-over 7. Sampling Procedure A, Determine samp pler to match, + From the sampling procedure, decide whether staggered or simultaneous start of the six tests is needed. B. If manval sampling will be used, check to see that syringes and prepared with filters ready for 2 intervals, and program any automated sam- cannulas, if used, are clean a C. Determine whether the intervals between the sampling of vessels are adequate (if manual sampling will be performed) for the sampling time required by the sequential start of the test. (Is ne For filtration, replacement, simultaneous sam- there enous ple pulls?) 8. E Run Setting Up for Dissoluion Testing 107 Ensure that sampling location, whether manual or automated, is consistent and properly located, Determine whether the procedure is such that any effect of the filter on analytical results is either known or avoided Prepare test tubes or sample collection tubes if needed. ring the Test Place the dosage Forms in the baskets or, if the paddle is being used, have them ready for placement in vessels, Examine the system for air bubbles to check deacration of the ‘medium Start the test using a stopwatch if sequential sampling is to be uused—do not rotate the baskets until they are in their final po- sition nor start paddle rotation until dosage form is introduced, For simultaneous sampling: a For Apparatus 1, drop the baskets, using the frames then start rotation, For Apparatus 2, drop the dosage form into each vessel as rapidly as is possible, and then start paddle rotation. , As soon as baskets are in place, note any basket with a float- ing dosage (orn entrapped in an ait bubble (lates, disead 16> sults from that basket). For sequential sampling: a Drop the individual basket (using the elutches on the drive) against the collar or marks then engage the clutch € stat its rotation, repeating that procedure for baskets 2, 3, and the rest at stopwateh-timed intervals: proceed as in D (b), above the cluteh: drop the dosage form as close as is possible w the center of vessel 1; use the clutch to start rotation; then follow with 2, 3, and so on, b. Use the stopwatch as at E (a) with the paddles stopped by the clutch drop the dosage form as close as possible to the center Of the flask 1; use the elutch to start rotation; then follow with 2.3, and so on. Observe the dissolution test at certain intervals and record ob- servations for each vessel, especially in methoxl development or R&D settings. At a minimum, observations should be made when samples are withdrawn, ishing the Test ‘Check the temperatures of the flasks and record any aberrations from the * 0.5 °C tolerance.108 HANDBOOK OF DISSOLUTION TESTING B, Check and recon! speed. C. Be sure data are printed or recorded before discarding samples it is a good practice to save the samples (iF possible) until all data are processed, because there may be questions that ean be answered only by subsequent analysis, D. Examine the baskets to determine if air has trapped the dosage form—diseard (with proper documentation) results from that flask, E. Note whether any silvery sheen or air bubbles because they indicate the release of dissolved gas, F Note the condition of any remaining mass of the dosage form, its position, whether clogging of baskets has occurred, the nature of coning, and so on; such information may be valuable in at- to explain bizarre results from one flask that might ntly be disearded, G, Check the volume of one or two flasks to ensure that no evap- oration has been overlooked, because such evaporation could significantly affect the analytical data. This is especially true in extended release products that are tested over many hours. appeared ‘The dissolution protocol or test method should follow the standard ‘operating procedures that are unique to the quality assurance program developed by individual laboratories as a part of good laboratory and manufacturing practice (1, 2) Although there has been considerable concern over the adequacy of calibrator tablets, experience has shown them to be useful in standard- izing Apparatus 1 and 2. When the apparatus is regularly calibrated, normally at six-month intervals and when moved, many of the setup procedures (such as checking tilt) may be omitted from the checklist, Undoubtedly, similar standard calibration methods will be offered for all new proposed methods. Details of some of the methods of establishing the correct perfor mance of Apparatus 1 and 2 from the first and second edition ate repeated here. They may be helpful when setting up to run a test with calibration tables, Inspecting Paddles and Shafts Every laboratory should have an inspection and straightening fixture such as the one shown in Figure 6-L, This unit is commercially avail- able. but it may be easily fabricated by an in-house machine shop. Tt consists of two V-blocks firmly mounted on a base with two posts for Setting Up for Dissoluion Testing 109 Pas ” UZ LIE LEA LE LE A KE OOK. (Sor iY ao SY Figure 6-14 Use of machinist’s indicator for straightening shafts mounting a simple machinis’s dial indicator, The dial indicator should hhaye a small ball tip, read a toral travel of * 0.015 inches (* 0.4 mm), and be graduated in not more than 0,001-inch (0,02-mom) intervals, The figure shows three separate tests that should be made on paddles and baskers, View A of the fig ness for either baskets or paddles. The tip of the indicator just touches the shaft, which is slowly rotated by hand. The indicator will measure the degree of shaft crookedness. t should not read a total (TIR) greater than about 0.005 inch (0.1 mm). 1F it does, the shaft should be rotated shows the proceduse for testing shaft stuaight- Figure 6-1B Use of machinists indicator for checking basket concen- icity.0 HANDBOOK OF DISSOLUTION TESTING Figure 6-1C Use of machinis’s indicator for checking paddle blades. the indicator reads high; a gentle push of the heel of the hand in this position will straighten it. The procedure should be continued until the maximum TIR is 0.005 ine! ‘The procedure may seem difficult, but itis really quite simple. Once the operator has learned the required amount of pressure, shafts can quickly be straightened, and the method works on shafts that are erook~ ed over their whole length. If a shaft has iwo high points, it cannot successfully be straightened, and should be discarded. Finally, the an- alyst should use care to see that the indicator tip does not rest on the polyfiuorocarbon surface of the paddle, if present, where it may indicate ‘surface roughness that has nothing to do with shaft straightness. View B illustrates the procedure for checking baskets. The basket should fit snugly and squarely against the drive-rod disk and should be held tightly in place by the clips. The indicator tip is held lightly against the bottom ring of the basket, and the whole assembly is slowly rotated by hand. The TIR can generally be kept to within approximately 0,010 inch (0.25 mm). The basket (not the shaft) may be bent by hand to bring it within that reading. Oller baskets that have seen much service in acid media tend to be very weak and flexible, and it may be im- Possible to straighten them; they should be discarded, Finally, take great care in attaching the properly aligned basket to the drive. Hold it by the top ring, not by the bottom, in order t avoid distortion, View C shows a procedure for ensuring that the tips of the paddle blades are equidistant from the axis of rotation, Care must be used in this test, because the blade tips have a radius, and one must be careful to measure each tip at the most extended point of that radius. ‘The blade Seating Up for Dissolution Testing mm Figure 6-2 Use of machinist’s indicator to check shaft runout tips should be equidistant from the axis of rotation within a maximum TIR of 0.080 inch (2 mm). This maximum is too generous: the blade lips should be kept to within a ‘TIR of 0.030 ineh (1mm). Checking the Kecentricity of Paddles or Baskets “The paddles or shafts should be inserted into the chucks of the drive spindles and mounted in approximately the vertical position that they will occupy during the test. The drive should be lifted to the extent that the paddles or baskets will rise above the Mask covers. A-ma- chinist’s dial indicator is then used fo determine runout ar eccentricity ‘The indicator must be mounted on a heavy base, of the sort available at a machine-shop supply house or from a dissolution equipment ven- dor. (Magnetic bases will not work because most dissolution drive bases fre not magnetic, A small block of seft iron or steel may be purchased and the standard magnetic base attached t0 it) This method of measuring eccentricity is illustrated in F set the contro! for minimum rpm (not move than 25 rpm), or manually42 HANDBOOK OF DISSOLUTION TESTING rotate the shalt with the clutches disengaged. Each shaft should be checked individually. The procedure requires only a few minutes, and it ensures that no errors are present because of improper or worn chuck grips. USP allows + 2 mm (0.160 inches TIR) if such an eccentricity does not alter the results of the test, Nonetheless, the authors feel that eccentricity should be held under a TIR of 1 mm (0.030 inches) or 0.5 mm for minimum disturbance, Methods of correcting excessive eccentricity are discussed in Chapter 5 Checking the Speed Control Digital tachometer readouts are available on al sate-of-the-art speed controls. The control may also offer a means of calibrating the digital readout against line frequency (50 Hz or 6D Hz), which in most coun- tries is Kept within tight limits. Checking the digital tachometer should bbe a part of the maintenance protocol for the speed control but itis not generally necessary for each dissolution test run. Do not use mechanical tachometers, They impose « load on the sys- tem and may change the speed, (The Toad is relatively constant in solution, and speed controls desisned for such purposes may be se sitive to load changes). If you wish to check the speed, put a piece of tape on the shaft, use a stopwatch, and, with your finger close to the shaft, count the revolutions as the tape tab snaps past your finger. To- day, optical tachometers (light or laser) are most conveniently used to cheek speed rotation, Surges in speed generally can be heard. They are sometimes present in older types of speed conteos, although in rare instances this ean be 4 malfunction in newer equipment. If the control produces surges, it should be replaced Checking Vibration Vibvation is one of the most common random input variables pro- ducing variation in dissolution data, The most satisfactory measurement requires the use of a vibration meter. Vibration exerts most of is infu ence at the vessel, and that is where it should be checked, Figure 5-2 iMustrates the use of a vibration meter that measures displacement in ins, Steps to correct random vibration in excess ofthe O.t-mil limit are diseussed in Chapter 5 TF no vibration meer is available, other techniques may serve in its place, With «litle experience, an analyst ean lean to place his or her Setting Up for Dissolution Testing 113 fingertips on a piece of apparatus and estimate the vibration rate. A visual inspection oF the surrounding area is paramount in a dissolution Jab, especially in lab that shares space with analysts performing other tests. Look for shakers, sonicators, centrifuges, and other equipment that may cause vibration, Install mechanical door closers on the labo- ratory doors, because slamming doors jolt the baths and affeet the dis- solution results, Centering Flasks to the Stirring Drive ‘This problem is discussed in Chapter 5. Various types of centeri gauges are commercially available: the simplest is n plastic disk with ‘taper that ean be located inside any vessel top and has « slotted hole, the center of which coincides with the stirring deviee. ‘When one uses such a gauge, the disk is centered on the shaft, and the vessel is moved until the disk drops into its inside diameter: if centering lugs are provided, they are then adjusted and tightened. Ves- sels are then marked and repliced in the same position from test to test. Modern dissolution test stands have various self-centering options, a discussed ins Chapter §. They may, however, be checked by this method. Analytical Methods and Filtration Special care must be taken to guard against errors resulting from sorption onto filters. Guidelines (3, 4) have outlined specific tests 10 determine the effects of filtration, When performing manual filtration the authors recommend discarding the fist portion of the filtrate (filter check validation establishes the amount to be discarded), They suggest that tests be run on both & 100% standard dissolved in the dissolution medium and on an aliquot from actual dissolution of the tablets in question at the end of the specified sampling time. Each should be subjected to filtration, and sequential S-inL aliquots should be collected. Each should be analyzed until final, reproducible values are obtained. Such a procedure establishes the portion of the sample that should be discarded before the actual analysis. The typical tolerance on the stan~ dard eould be around 99%-100% recovery. When filters are being reused (as in establishing a profile) or are used in automated equipment, more severe test is required and calls for the preparation of a 100% standard, as before, and a blank disso: lution medium, Analyses as in the previous paragraph are then per-14 HANDBOOK OF DISSOLUTION TESTING formed on aliquots from standard, then from blank, then from standard in succession, using the same filter Toleratices should be 99%-101% recovery for the first standard, no more than 54 in the blank, and 959%— 101% recovery in the second analysis of the standard, If those toler- ances are exceeded, there may be excessive dead volun carry-over, or cumulative sorption from some source that should be investigated and corrected Similar tests may be made on automated equipment by feeding 100% standard through the tubing and determi of suc cessive aliquots. That step is followed by feeding blank solution through the tubing and determining the carry-over content of the ali= quots. Repeating those steps in sequence should establish the suitability ‘and sorption characteristics of the tubing and flow materials in the apparatus Particular eare should be used in preparing the standard. ‘The inter- ference of media components or excipients from the dosage form should be considered. Some standards must be made via the addition Of organic solvent, Prednisone, for example, requires some alcohol for the preparation of a completely dissolved standard. Methods listed in the compendia for standard preparation should therefore be carefully followed, Check the ellect of the organic solvent by preparing several solutions of the same concentration per liter of the drug, but use dif- ferent concentrations of the solvent. Each of those should assay to with= in 99%-01% of the theoretical amount; failing that, a different organic should be selected. Finally, the stability of the standard should be considered. Prednisone standards used for calibrating dissolution equipment, for example, sometimes change in value because of (ap- parently) the precipitation of a minute, almost colloidal, form of the drug, Interference of sampling probes in the How dynamics can be easily checked, and generally will be bothersome only in automated equip- ‘ment from which successive aliqu taken. A dissolution test with the probes in place can be compared with a test conducted with the probes manually removed except during the sampling cycle (4). If the resulting mean deviations or RSDs of the tests are significantly differ- ent, the matter should be further explored; either the probe design will need to be ef J method of probe removal (except during sampling eyeles) should be installed. IF the probes return me= dium or are purged with air, they may have t be removed from the medium except when withdrawing samples (5 the recovery raged oF an aut Sesting Up for Dissolution Testing us Figure 6-3 Use of simple height gauge to establish distance from bot- tom of vessel (25 mm). Distance of the Paddle or Basket from the Bottom of the Vessel Because the overall depth of flasks may vary, should be individually adjusted to the compendia distance from the bottom, 2.5 em + 0.2 em. One simple way w accomplish this is 10 preset the drive in its vertical operating position, push the basket to the bottom, mark the top of the shaft where it protrudes from the hollow drive tube, carefully move it up 2.5 em, and mark or attach a collar 10 establish the proper position as located from the top of the drive tube, Several designs of licight gauges have been suggested. One com- ‘mercial type that can be easily fabricated from ¥-inch (about 3-mm) diameter stainless rod is shown in use in Figure 6-3. This gauge is convenient t0 use with either the basket or puddle. The proper proce- dure for setting shalt height requires that the drive first be set in its operating position, Stops or adjustable collars on the mounting posts ch paddle or basket116 HANDBOOK OF DISSOLUTION TESTING fare genetally used to establish this position so that it can be quickly accurately repeated. While they are in this position, the individual baskets or paddies are adjusted inside the hollow drive spindle using the height gauge. When one is using baskets, that operation should be performed without medium in the asks, if medium is present, the baskets must be thoroughly dried hefore the dosage form is ins (Figure 6-3) ‘Some commercial dissolution (est stations have snap locators or pre- positioned shafts so adjustment by the operator is not necessary. This is convenient but does not allow variation in the vertical position of the drive. This ean be a major defect in automation, particulaly with robotics, The whole drive head may then be lifted up out of the way, om if the basket shafis have been marked or collars attached, they may be individually raised, Retaining the collars and the stops for the vertical posts on the drive allows all six paddles or baskets to be lowered either individually or together at any time into their proper operating posi= tions. ‘When operating the paddle, the drive is dropped into its operating position with the paddles ready to start at the proper speed. The paddles should not rotate until the dosage forms are added and preferably not ‘until they setile into a position in the center of the round bottom of the flask. The addition of the dosage forms may then be staggered at the desired sampling intervals (using a stopwatch or an automated time signal) or dropped into all vessels at approximately the same time, Individual clutches are used for staggered operation, and the speed con trol is simply turned on for simultaneous operation, Clutches and speed ccontrol may be made responsive (o RS-232 input commands in robotic or other automated systems. When one is using the baskets, the baskets must be above the vessels when they ate loaded, Be sure the baskets are dry, They should be attached fo the drive by holding the top ring only, in order to avoid basket distortion. The FDA labs recommended that the basket—with its load—be dropped to the proper position without turning (3). This an be accomplished by turning on the drive (but disengaging the clutches) and pushing each basket down into the media agains its collar at the predetermined sampling interval and then immediately engaging the clutch for that drive, Or, with the baskets already in their proper ‘operating position (but above the medium) and the drive turned off, the ‘operator can lower the drive against its stops and engage the motor immediately Seutnyg Up for Dissolution Testing 417 ‘The analyst may comment that a considerable degree of dexterity is requited for these manual operations, particularly the dropping of dos- age forms for simultaneous operation with the paddle. USP allows a * 2% variation in sampling time, With automated samplers withdraw- ing simultaneously, this + 2% still allows a one seconel interval be- toveen each of six vessels, even with a minimum sampling interval of 5 minutes. Calibration of Equipment The dissoluion testi an analytical procedure designed to determine certain physical characteristies of a compound, One would expect any analytical procedure to produce repeatable results with the same com pound. Any variation in results may be classified as experimental ertor and, depending upon the nature of the procedure and on the state of the art, certain limits should be plaeed on the extent of such experi- imental error, Some philosophical questions have been raised with re- spect to dissolution, and, as yet, there is no consensus on their answers During te 1980s these questions involved wheter dissolution proc dures have any meuningful place as analytical quality control proce- ‘dures, but now there is no doubt, as the dissolution testis required by FDA for all solid oral dosage forms. However, problems arise regarding what degree of experimental error ‘or variation from test to test can be tolerated if the test is to be used as a quality contol and as u regulatory criteion, Dissolution is now ‘considered a mature test with coordinated research in its technology of cover 35 years. Sill the question is raised whether dissolution data have any value in QC if such data are not linked to in vivo performance. Indeed, now 4 current tend isto set in vitro QC dissolution its within thera of experimentally demonstrated acceptablefaon acceptable in vivo data, hhence an in vitro and in vivo relationship (6). European analysts have aptly stated that reproducibility and appa- ratus calibration ean best be achieved from “quality of design” (7, 8) As they further indicate, this requires that state Variables of simple gcometry be restricted by elose tolerances, qualified components, com prehensible and unambiguous description of protocols, and regular checks of dynamic parameters. No cngineer or scientist can dispute these points. The first step, therefore, is (© establish and rigorous! follow the checklist for dissolution protocol provided in this chapter,18. HANDBOOK OF DISSOLUTION TESTING including inspection of equipment, standards for the dissolution bath, ‘media standards, and details of analytical methods, ‘The first step in calibration of a dissolution test station to USP stan- dards must be checking the state variables. This cannot be overemph: sized. Many of these variables are discussed in detail in this chapter. ‘The extent of their influence over results from Apparatus 1 and 2, and suggestions for their correction are presented in Chapter 5, Calibrator Tablets. Calibrator tablets are available from USP Ref erence Standards, 12601 Twinbrook Parkway, Rockville, MD 20852, USA, They may be ordered in non-disintegrating and disintegrating Forms. The certificate of ranges and the applicable test method is sup- plied with the tablets As of this writing, the non-disintegrating tablets consist of salicylic acid, the disintegrating tablets of prednisone, Questions have arisen regarding standards, and some analytical chemists question whether those tablets are standards in the analytical sense. ‘This has been ex- plored in numerous dissolution workshop articles and seminars, Some say the calibrators are misnamed and should be proficiency or system suitability standards, ‘The answer Is that the standards are functional within the limits of today’s technology as physical standards for dissolution, Even such a basic unit as the speed of light has «a tolerance that we all accept. The tolerances of standards are historieally based upon state-of-the-art tech- nology, and it seems nonproductive to criticize the dissolmtion standards because the current state of the art in the physical test of dissolution does not meet the exactitude of more sophisticated analytical proce- dures, In other words, we have no better alternative at this time, but as ‘mentioned in Chapter 5 the quest continues, Questions about the stability of the currently available calibrators hhave been adequately answered. Studies on two separate batches of m-disintegrating Salicylic Acid calibrator tablets have established in- significant variation with different manufacturers and time—over seven yeurs (9, 5). Studies on the disintegrating calibrator indicate similar Stability and uniformity (10). Studies on laetose-based Prednisone eal- ibrator tablets, however, suggest that moisture exposure must be con- sidered to have significant effects on dissolution characteristics (1). ‘The USP laboratory continuously monitors the stability of the calibra- tos as part of the USP Reference Standards Program, Compendial Calibration Requirements. The instructions, methods, and acceptable ranges are supplied with each lot of calibrator tablets. Calibrator dosage forms and equipment suitability tests for Apparatus Setting Up for Dissolution Testing 119 3, Chlospheniramine Maleate Extended Release tablets are available No other calibrator tablets are availible as of the date of this publi tion, Calibrators for Apparatus 4 are under consideration at USP and are being evaluated in the USP laboratory. In 1979, the Hanson Research laboratory studied 400 different six- spindle dissolution drives using the USP disintegrating calibrator (pred= nisone), To test the hypothesis that the published acceptance ranges were too wide, the laboratory arbitrarily established its own, using the paddle at SO rpm, at * 5% of the acceptance mean of 64% dissolution in 30 min (that is, between 60.89% and 67.2%). ‘The laboratory estab- lished a further requirement that the RSD of the six vessels had to be <5%. OF the 400 units, 5% failed 10 meet the requirements (20 drives). ‘The reason for each failure was determined, corrected, and a subsequent test met the specifications. ‘The reasons for failure were: for nine units, problems with the belt, chain, or drive adjustment resulting in excessive vibrations for six units, misalignment or excessive till; for five units, problems with speed con ‘rol, resulting in roughness and variation; for three units, improper procedures, a dirty flow cell, a deteriorating standard, and so nd for one Unit, 2 Vessel temperature out of specification. These deleets add up t0 more than 20 failures because some tests included more than one faulty variable. Even though these tests were performed many years ago, these problems still plague the analysts (12, 13). ‘There was also reported some uiffieulty in meeting calibration ra with the non-disintegrating salicylic acid tablet. Because this isa heavy tablet that tends to settle to the bottom of the rounded vessel in random positions, the alignment of the drive becomes more critical. Failure to foblain laminar flow because of turbulence may cause the tablet to bounce around in a random manner, Such motion can be observed during the calibration test, and careful inspection oF paddle geometry (panicularly the equidistant position of the tips from the axis of rota- tion) and drive alignment may correct the problems. The salicylic acid test is also very sensitive to slight pH ch must be certain to use a pH meter properly calibrated to two decimal places ad, Meeting Calibration Limits + The analyst must always consider the complete dissolution test as aa system; no variable should be overlooked, ineluding the complete analytical protocol120 HANDBOOK OF DISSOLUTION TESTING + The analyst should refer to a checklist similar (© that provided at the beginning of this chapter and shoukl-begin with & rigorous inspection of the equipment, Following it through with each pro- cedure + The analyst must not assume—behind every scientific problem there is probably an invalid assumption, and this assumption must be found and corrected In summary, the concept of uppurstus suitability, using calibrator standard-dosage forms, has been generally accepted 3s a practical meth od of ensuring consistency in dissolution testing, Once the random input variables are defined and their acceptable limits are studied, there is no reason why equipment and methexls cannot qualify consistently under the restraints of calibration, The authors feel that the calibration procedure should be testing the rozal system. Problems arising during calibration can and should be comected by studying all the random variables, Apparatus suitability tests may be forthcoming for the newer pro- posed dissolution techniques suggested for extended-release and irans- dermal preparations. These suitability tests will help stabilize the field of dissolution testing. Equipment should be calibrated every time it is moved or w significant changes ane made in its immediate environment, Son even calibrate before moving so that if the equipment fails calibration after the move it ean be proven that the move did or didnot cause the equipment 10 go out of calibration, Automated systems should also be calibrated whenever changes are made in sampling, tubing, o¢ other adaptations. Although no compendial requirements exist, good labora tory practice would suggest a minimum 6-month calibration for the dissolution station (see Appendix for reference on IQIOQIPQ), Other ‘components of the system should be calibrated according to good lab- ‘orutory practice requirements for the instrument involved, Calibration of Non-Compendial Equipment Because of the rapid development of special dosage forms, other unique apparatus may become used routinely (15). The Franz diffusion cell and Enhancer eell have been used for ereams and ointments, A chewing gum apparatus has been specially designed for this unique dosage form: this apparatus is deseribed in the EP. Inirinsie dissolution is another test that has Several apparatus in use. With the addition of more intricate equipment and testing apparatus Setting Up Jor Dissolution Testing 2 available For dissolution testing, there isan absence of extensive studies fof the random variables that ate inherent 10 a forunately the burden is on the analyst to discover these sources of error and variability. The analyst does however have the equipment manufacturers” specifications as a starting pls (16). First, look for adjustments and moving parts. Set a baseline for ‘operational parameters, such as flow rate or dip rate, emperature, and volume controls. Choose an in-house calibrator tablet, if possible, with a well-characterized and stable product. The issues of alignment, hy drodynamies, and vibration should always be considered. and if pos sible, measured. A metrology document or protocol should be available for equipment operation, slong with logbooks to ack performance and ‘maintenance (17). new technique. Un- se for assessing variables