Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
# The temporary pacemaker implantation is a therapeutic procedure that can be necessary when the patient has symptoms of low output syndrome due to irreversible and uncontrolled arrhythmia that do not respond to the medical treatment. # Pacemakers provide electrical stimuli to cause cardiac contraction during periods when intrinsic cardiac electrical activity is inappropriately slow or absent.
# Pacemaker output generally stimulates the cavity of the right atrium and or right ventricle.(endocardial pacing).alternatively,epicardial leads can be implanted surgically onto the hearts surface.
Pulse generators
The pulse generator is internal in permanent pacemakers (subcutaneously or submuscularly) and external in temporary pacing. Can be set to a fixed-rate (asynchronous) or demand (synchronous) mode. In the fixed-rate mode, there is a small risk of producing dangerous dysrhythmias if the impulse coincides with the vulnerable period of the T wave. On demand pacemakers detect spontaneous ventricular activity and the output of the pacemaker is either suppressed or discharged in order to make the impulse fall within the safe period of the QRS complex.
Have pacing electrodes in both the right atrium and the right ventricle. They allow maintenance of the physiological relationship between atrial and ventricular contraction.
Dual-site atrial pacing: Newer pacing systems have 2 atrial leads, one in the right atrial appendage and the other either in the coronary sinus or at the os of the coronary sinus. The ventricular lead is in the right ventricle, either at the apex or at the outflow tract. This system has been proposed as a promising treatment option for prevention of paroxysmal atrial fibrillation.
Biventricular pacemakers: Pacemaker leads are placed in the right atrium, right ventricle and left ventricle.
Useful in the management of patients with heart failure who have evidence of abnormal interventricular conduction (most often evident as left bundle branch block on ECG) which causes deranged ventricular contraction or dyssynchrony.
Pacemaker codesThe North American Society of Pacing and Electrophysiology and the British Pacing and Electrophysiology Group have developed a code to describe various pacing modes. It usually consists of three letters, but some systems use four or five: Letter 1: chamber that is paced (A=atria, V=ventricles, D=dual chamber) Letter 2: chamber that is sensed (A=atria, V=ventricles, D=dual chamber, 0=none) Letter 3: response to a sensed event (T=triggered, I=inhibited, D=dual - T and I) Letter 4: rate responsive features Letter 5: Anti-tachycardia facilities
Temporary PacemakersTranscutaneous external pacemaker Epicardial pacemaker Transthoracic thoracic pacemaker Transvenous endocardial pacemaker
Elective:
Support for procedures that may promote bradycardia General anaesthesia with: 2nd or 3rd degree AV block Intermittent AV block 1st degree AV block with bifascicular block 1st degree AV block and LBBB Cardiac surgery : Aortic surgery Tricuspid surgery Ventricular septal defect closure Ostium primum repair Rarely considered for coronary angioplasty (usually to right coronary artery) but may be required for angioplasty-induced bradycardia Temporary application during implantation or exchange of permanent pacemaker.
Insertion
External
Epicardial
- Usually attempted emergently, as a last resort, after other temporary pacing methods have failed. Physician inserts a pericardial needle through the subxyphoid area of the thorax into the right ventricle and advances the lead wire through the needle in order to achieve contact with the endocardium
Transvenous
- May be inserted at the bedside, preferably under fluroscopy. Usually inserted into the subclavian or jugular vein, but can be inserted into the antecubital or femoral vein
SettingsRate
Fixed
(Asynchronous) # Stimulus is provided at a preset rate # Rate is set greater than the patients inherent rate to avoid competition
(Synchronous) # Stimulus is provided when the patients heart rate drops below at predetermined rate # Must have adequate sensing
Demand
Sensitivity Threshold
Sensitivity # Set in millivolts (mV) # Allows pacemaker to detect the patients inherent R wave Sense Indicator # Flashes when inherent R wave is detected # Senses if pacing is in the demand mode Threshold # The minimum R wave amplitude needed to be detected by the pulse generator # Once the sensitivity threshold is determined, the sensitivity is set 2-3 times lower
Stimulation Threshold
Output/mA
# Stimulus current is measured in milliamperes (mA)
# Adjusted based on the amount of current needed to elicit myocardial depolarization and contraction
variables - position of electrode; contact with viable myocardial tissue; level of energy delivered through wire; presence of hypoxia, acidosis or electrolyte imbalances; and other medications being used
Pace indicator
# Flashes each time a pacing stimulus is generated # Does not necessarily indicate that a cardiac contraction occurred
Pacing Threshold
# The minimum amount of mAs needed to achieve 100% capture # The output is then doubled
Pacing Modes
Impulse initiated via the atria Pathway similar to normal conduction Can be initiated via epicardial atrial leads Can result in competition Used for asystole or symptomatic sinus bradycardia Contraindicated for atrial fibrillation or flutter and for person with conduction delays
Impulse is sent to the ventricle when the patients inherent rate drops below the preset rate on the pulse generator No harmony between atria and the ventricle Used as a back up system for sinus bradycardia, heart blocks, atrial fibrillation/atrial flutter with SVR, junctional rhythm
Normal conduction through the heart Used for asystole, symptomatic sinus bradycardia, and for varying degrees of heart block to maintain AV conduction
Impulses sent to atria and ventricles when the patients rated drops below the preset rate on the pulse generator Normal conduction through the heart Used for asytole, symptomatic sinus bradycardia and heart block
Troubleshooting
Failure to fire - cant see pacemaker spikes during periods of asystole or bradycardia
Undersensing
pacemaker fires with no regard to the patients own rhythm. Dangerous because it may lead to ventricular tachycardia and/or ventricular fibrillation
Inadequate QRS signal Myocardial ischemia, fibrosis, electrolyte imbalances, bundle branch block, or a poorly positioned lead
Oversensing
pacemaker thinks it detects a QRS complex, inhibits itself and doesnt fire
Myopotentials (electrical signals produced by skeletal muscle contraction as with shivering or seizures)
Pacemaker
Identify intrinsic rhythm and clinical condition Identify pacer spikes Identify activity following pacer spikes Failure to capture Failure to sense
EVERY PACER SPIKE SHOULD HAVE A PWAVE OR QRS COMPLEX FOLLOWING IT.
Normal Pacing
Atrial Pacing
Normal Pacing
Ventricular pacing
Normal Pacing
A-V Pacing
Atrial & Ventricular pacing spikes followed by atrial & ventricular complexes
Normal Pacing
Abnormal Pacing
Atrial non-capture
Failure to Capture
Causes
Insufficient energy delivered by pacer Low pacemaker battery Dislodged, loose, fibrotic, or fractured electrode Electrolyte abnormalities
Failure to Capture
Solutions
View rhythm in different leads Change electrodes Check connections Increase pacer output (mA) Change battery, cables, pacer Reverse polarity
Late complications
THANK YOU