Sei sulla pagina 1di 10

Titan 2000 DR

Digital Radiology for Veterinarians Product Specifications

Titan 2000 DR - Digital Radiography for Veterinarians

Titan 2000 DR - Digital Radiography for Veterinarians


There are numerous digital x-ray devices for veterinarians on the market; however, the Titan 2000 DR combines in a unique way the best characteristics and features of all products. The Titan 2000 DR was designed in cooperation with a leading veterinarian and combines the best characteristics and features of all other X-ray systems on the market to bring you simple, easy operation; fast and efficient patient throughput; superior image quality; and serious cost reductions. With its small footprint, the Titan 2000 DR requires less space than any conventional X-ray system. The 17 x 17 digital image receptor produces a much larger field of view and simplifies patient positioning with only a two-way float top table. Furthermore, the X-ray beam is always centered to the receptor, making patient positioning fast, easy and precise. The versatile high frequency X-ray generator is powerful, easy to install and easy to use. Power output is 20kW up to 40kW depending on the electrical installation. X-ray technicians can control the generator from the touch-screen preview monitor or from the workstation, whichever is more convenient. Time consuming technique choices are eliminated by precise anatomical programming presets for a perfect exposure every time, while custom techniques require only a fingertip to overwrite default values. The X-ray exposure is initiated by the foot pedal at the table or by mouse click at the workstation. Within just a few seconds the preview image is displayed on both the preview monitor and at the workstation. With only a few mouse clicks your image is processed and ready for diagnosis using Comeds proprietary VetView image processing software. It doesnt get any easier! And VetPACS provides DICOM file transfer; patient search; image comparison; CD/DVD burning; and DICOM film print. VetPACS is customizable to any size veterinary clinical environment, any number of modalities and any number of workstations. The Titan 2000 DR system is shipped in a single crate L 65 x W 61 x H 47, total weight is 1,100 lbs. Installation of the Titan 2000 DR is simple Plug & Play and should take no more than 2 to 3 hours. Assemble the vertical column; mount the X-ray tube and collimator; mount the preview monitor and switch panel; a diagram explains the electrical connections. The generator is thoroughly pre-calibrated by a Comed engineer and an engineer will also log into the preview computer after the hardware installation is complete for a remote check of proper functionality. Verbal communications during this process can be established by using Skype. This support makes the installation fast, easy, and very professional. In no time, you will be ready to start realizing all the benefits that the Titan 2000 DR will be bringing to your practice for years to come.

Product Specifications
CCD X-Ray Receptor and Image Processing Computer
Available cameras Available screens 9 MP or 16 MP (Megapixels) Gadox Screen (Gadolinium Oxide) with 9 MP camera Csi-Screen (Cesium Iodide) with 16 MP camera Field of view 17 x 17 X-Ray Grid No x-ray grid in use Dynamic Range 16-bit Spatial Frequency 16 MP camera Raw image 4.6 lp/mm , processed 3.8 lp/mm Spatial Frequency 9 MP camera Raw image 4.2 lp/mm , processed 3.6lp/mm Image Acquisition Time less than 3 seconds Display of Processed Image within 10 -12 seconds Pre-view Monitor 17 LDC touch screen monitor that also controls x-ray generator Workstation Computer Intel Pentium 4 Dual core processor computer with 20 LCD monitor (or equivalent) Operating System Windows XP Professional External Image Storage 2 TB (optional) Power requirement 220 V AC 1.5 Amperes

Software

Vet-View image processing software that includes several image enhancing filters to choose from and offers viewing functions such as: - Zoom in and out - Brightness / Darkness adjustment - Image rotation to the left and to the right - Horizontal and vertical flip - Invert - ROI function (range of image) - Pan function - Insert markings VetPACS software package (optional)

Titan 2000 DR - Digital Radiography for Veterinarians

Titan 2000 DR - Digital Radiography for Veterinarians

X-Ray Generator
Type Generation method Power output kV output mA output Time steps Generator control Calibration Input Voltage Amp. Service Dimension Weight Single Phase High Frequency X-Ray Generator 40 KHz inverter 20 kW, 30 kW, or 40 kW, adjustable to Ampere service 40 - 125 kV in 1 kV increments 25 mA - 500 mA 1 mS to 5 S RS-232 C Pre-calibrated at Comeds plant 220V/240V AC Single Phase 100 Amperes or higher (slow fuse or slow breaker required) 530 x 410 x 665 mm or 20 x 16.14 x 26.18 88 kg or 193 lbs..

X-Ray Tube
X-Ray Tube Focal Spot Heat units Target angle Toshiba E-7239X (equivalent to RAD-8) 1.0 / 2.0 140 KHU 16

Collimator
Type Max kVp Power Manual collimator 150 kVp 12 V DC No x-ray grid used in current system 20 feet high voltage cables

X-Ray Grid High voltage cables

Total height: 1,884 mm or 74.17 Tube focal spot to table top: 949 mm or 37.36 (SID 40 tube focal spot to camera screen) Table top height: 798 mm or 31.4 Total depth incl. handle in front of table top: 820 mm or 32.28

Titan 2000 DR - Digital Radiography for Veterinarians

Titan 2000 DR - Digital Radiography for Veterinarians

Table length: 1,250 mm or 49.21 Option: Table length of 1,524 mm or 60 Table depth: 600 mm or 23.6 2-way floating to the right and to the left of 200 mm each or 7.80 each

Total height: 1,884 mm or 74.17 Tabletop height: 798 mm or 31.41 Floor to center of pre-view monitor: 1,342 mm or 52.83 Base board width: 960 mm or 37.79

Titan 2000 DR - Digital Radiography for Veterinarians

Titan 2000 DR - Digital Radiography for Veterinarians


Sample Images

Titan 2000 DR - Digital Radiography for Veterinarians

Comed America
3103 Airport Boulevard, Suite 620 Mobile, Alabama 36606 Phone (251) 470 1995 Fax (251) 470 1997 www.comedamerica.com